Tektronix TDP1500 User manual

xx
TDP1500
1.5 GHz Differential Probe
ZZZ
Technical Reference
www.tek.com
071-2296-01
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by na
tional copyright laws and international treaty provisions.
Tektronix pro previously published material. S pecications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be r esponsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from a ttempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
Table of Contents
General safety summary .. . .. . .. . . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . . .. . .. . .. . .. . . iii
Preface................................................................................................................................. v
Specications .......................................................................................................................... 1
Warranted Characteristics ....................................................................................................... 1
Typical Cha
Nominal Characteristics..........................................................................................................5
Probe Tip Adapter Specications................................................................................................ 5
Performan
Required Equipment ... . .. . .. . ... . .. . .. . . .. . .. . ... ... . .. . .. . ... . .. . .. . . .. . .. . ... . .. . .. . ... ... . .. . .. . . .. . .. . ... . .. . .. . ... . .. . .. . . 11
Special Adapters................................................................................................................ 12
Preparati
DC Attenuation Accuracy and Differential Signal Range .. . .. . .. . . .. . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . .. . ... . .. . . 15
Analog Bandwidth. . .. . ... . .. . .. . ... . .. . .. . .. . ... . .. . .. . . .. . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . .. . ... . .. . .. . ... . .. . .. . 18
Common M o
Rise Time........................................................................................................................23
Appendix A: Alternate Verication Procedures .. . .. . ... . .. . .. . . .. . .. . ... . .. . .. . ... . .. . .. . . .. . .. . ... . .. . .. . ... . .. . .. . . .. . .. . ... . .. . .. 28
Equipme
Preparation . . .. . ... ... . .. . .. . .. . . .. . .. . .. . ... . .. . .. . .. . ... . .. . .. . .. . ... . .. . .. . .. . ... . .. . .. . .. . ... . .. . .. . .. . . .. . .. . .. . ... . .. . .. . .. 28
Analog Bandwidth. . .. . ... . .. . .. . ... . .. . .. . .. . ... . .. . .. . . .. . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . .. . ... . .. . .. . ... . .. . .. . 29
CMRR (C
Maintenance.......................................................................................................................... 33
Inspection and Cleaning . .. . ... . .. . .. . ... . .. . .. . .. . ... . .. . .. . . .. . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . .. . ... . .. . .. . ... . 33
Replac
Preparation for Shipment . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . ... ... . .. . .. . . .. . .. . ... . .. 33
racteristics ........................................................................................................... 2
ce Verification ............................................................................................................. 10
on ...................................................................................................................... 14
de Rejection Ratio ................................................................................................ 21
nt Required ... . .. . .. . ... . .. . .. . .. . . .. . .. . ... . .. . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . ... . .. . .. . .. . . .. . .. . .. . ... . .. . .. . .. 28
ommon Mode Rejection Ratio) ...................................................................................... 32
ement Parts.............................................................................................................. 33
Table of Content
s
TDP1500 Technical Reference i
Table of Content
s
ii TDP1500 Technical Reference
General safety s
ummary
General safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
To avoid re or personal injury
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected
to a voltag
Ground the
To avoid electric shock, the grounding conductor m ust be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
product m
Connect
Do not ap
e source.
product.
anual for further ratings information before making connections to the p roduct.
the probe reference lead to earth ground only.
ply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
y summary
This product is indirectly grounded through the grounding conductor of the mainframe power cord.
Do not op
Do not o
qualied service personnel.
erate without covers.
perate with suspected failures.
Do not operate this product with covers or panels removed.
If you suspect that there is damage to this product, have it inspected by
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
TDP1500 Technical Reference iii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
iv TDP1500 Technical Reference
Preface
This is the Technical Reference Manual for the TDP1500 differential probe. This manual provides specications and performance verication procedures for the probe.
Preface
TDP1500 Technical Reference v
Specications
Specication
s
The specications in Tables 1 through 4 apply to a TDP1500 differential probe installed on a Tektronix DPO7000 oscilloscope. When the probe is used with another oscilloscope, the oscilloscope must have an input impedance of 50 and a bandwidth of 1 GHz. The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits described. (See Table 1 on page 1.) The probe calibration routines should b e run on the host instrument before verifying the warranted probe specications. Specications for the TDP1500 and differential probe fall into three categories: warranted, typical, and nominal characteristics.
Warranted Characteristics
Warranted c Table 1.) Warranted characteristics that have checks in the Performance Verication section are marked with the
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. (See Table 1.) Warranted characteristics that have Performance Verication checks are marked with the vericati
Table 1: Warranted electrical characteristics
Specication 850 mV range 8.5 V range
coupled)
Common-mode signal range (DC coupled)
Maximum nondestructive input voltage between signal and common of the same channel
0°Cto+40°C)
Temperature
Humidity
haracteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. (See
on procedures are contained in the TDP1500 Technical Reference, which is Tektronix part number 071-2296-XX.
Differential signal range (DC
DC attenuation accuracy
Bandwidth (both ranges, probe only,
Rise time
Common-mode rejection ratio
symbol. Performance
±0.85 V
±7 V
(DC + peak AC)
±25 V
(DC + peak AC)
(DC + peak AC)
;0.6V
;25V
RMS
RMS
±0.75 V input: ±2% of input±0.85 V input: ±5% of input
±8.5 V
(DC + peak AC)
±7.5 V input: ±2% of input±8.5 V input: ±5% of input
DC to 1.5 GHz
<265 ps
>60 dB at 1 MHz
>38 dB at 100 MHz
>35 dB at 500 MHz
>30 dB at 1 GHz
>45 dB at 1 MHz
>25 dB at 100 MHz
>20 dB at 500 MHz
>18 dB at 1 GHz
Operating: 0 to +40 °C (32 °F to +104 °F)
Nonoperating: –40 to +71 °C (–40 °F to +160 °F)
1
Operating: 5-90% RH, tested at up to +40 °C (+104 °F)
Nonoperating: 5-90% RH, tested at up to +60 °C (+140 °F)
;6V
symbol.
RMS
1
See warning that follows.
WARNING. To avoid a burn hazard at high ambient temperatures, do not touch the probe with bare hands at nonoperating
temperatures above + 50 °C .
1 TDP1500 Technical Reference
Typical Characteristics
Typical characteristics describe typical but not guaranteed performance.
Table 2: Typical electrical characteristics
Differential input resistance, DC coupled 200 k Common mode input resistance 100 k±2%
Differential input capacitance
Common-mode input capacitance
Harmonic distortion
Offset accuracy
Differential offset range ±7 V (both ranges)
Common mode rejection ratio (See Figure 3 on page 3.)
System noise
Referred to probe output
Input impedance
Delay time 5.4 ns ±2 ns
<1.0 pF at 10 MHz
<2.0 pF per side at 10 MHz 1.5% measured using 495 mV
100 MHz
±10 mV
<2mVRMS <50 nV/Hz at 100 MHz
(See Figure 4 on page 4.)
RMS
(or 1.4 V
) output at
P-P
Specications
Figure 1: Typical bandwidth
TDP1500 Technical Reference 2
Specications
Figure 2: Typical rise time
Figure 3: Typical Common-Mode Rejection Ratio
3 TDP1500 Technical Reference
Figure 4: Typical input impedance versus frequency
Table 3: Typical mechanical characteristics
Specications
Dimensions, input connection
Dimensions, control box
Dimensions, probe head
Dimensions, output cable
ight (probe only)
Unit we
Figure 5: Probe dimensions
0.63 mm (0
117 mm × 41
86 mm × 11
1.22 m (
144.5 g
48 in)
(5.1 oz)
.025 in) square pin on 2.54 mm (0.100 in) centers
mm×29mm(4.6in×1.6in×1.15in)
mm × 8.9 mm (3.4 in × 0.45 in × 0.35 in)
TDP1500 Technical Reference 4