tional copyright laws and international treaty provisions.
Tektronix pro
previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of
shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective
product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and
replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced
parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period
and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall
be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and
care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel
other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or
d) to service a product that has been modified or integrated with other products when the effect of such modification or integration
increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN B Y TEKTRONIX WITH RESPEC T TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES,
EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS
IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX
AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
ration for Shipment .......................................................................................................40
Table of Content
s
TDP0500 & TDP1000 Technical Referencei
Table of Content
s
iiTDP0500 & TDP1000 Technical Reference
General safety s
ummary
General safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other
component manuals for warnings and cautions related to operating the system.
To avoid fire or personal injury
Connect and disconnect properly.
to a voltag
Ground the
To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the i nput
or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings.
product m
Connect
Do not ap
e source.
product.
anual for further ratings information before making connections to the product.
the probe reference lead to earth ground only.
ply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
y summary
Do not connect or disconnect probes or test leads while they are connected
This product is indirectly grounded through the grounding conductor of the mainframe power cord.
To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the
Do not op
Do not o
qualified service personnel.
erate without covers.
perate with suspected failures.
Avoid exposed circuitry.
Do not operate this product with covers or panels removed.
If you suspect that there is damage to this product, have it inspected by
Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
TDP0500 & TDP1000 Technical Referenceiii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
ivTDP0500 & T DP 1000 Technical Reference
Preface
This is the Technical Reference Manual for the TDP0500 & TDP1000 differential probes. This manual provides specifications
and performance verification procedures for the probes.
Preface
TDP0500 & TDP1000 Technical Referencev
Specifications
Specification
s
The specifications in Tables 1 through 4 apply to a TDP0500 or TDP1000 differential probe installed on a Tektronix DPO7000
oscilloscope. When the probe is used with another oscilloscope, the oscilloscope must have an input impedance of 50 Ω
and a bandwidth of 1 GHz. The probe must have a warm-up period of at least 20 minutes and be in an environment that
does not exceed the limits described. (See Table 1.) The probe calibration and Autozero routines should be run on the host
instrument before verifying the warranted probe specifications. Specifications for the TDP0500 and TDP1000 differential
probes fall into three categories: warranted, typical, and nominal characteristics.
Warranted Characteristics
Warranted c
Table 1.) Warranted characteristics that have checks in the Performance Verification section are marked with the
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements.
(See Table 1.) Warranted characteristics that have Performance Verification checks are marked with the
Performan
part number 071-1977-XX.
Table 1: W
SpecificationTDP0500TDP1000
Common-mode signal range (DC coupled)
Maximu
between signal and common of the same
channel
(10–90%, + 20 °C to + 30 °C)
Temperature
Humidity
1
haracteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. (See
ce verification procedures are contained in the TCP0500 & TDP1000 Technical Reference, which is Tektronix
arranted electrical characteristics
Differential signal range (DC c oupled)
m nondestructive input voltage
DC attenuation accuracy
dth (4.25 V range, probe only)
Bandwi
Rise time, 4.25 V range, probe only
Common-mode rejection ratio
See warning that follows.
symbol.
±4.25 V
±42 V
±35V
±100 V
(DC + peak AC)
(DC + peak AC)
(DC + peak AC)
(DC + peak AC)
;3V
;30V
;25V
RMS
RMS
RMS
±4.25 V
±42 V
±35V
±100 V
(DC + peak AC)
(DC + peak AC)
(DC + peak AC)
(DC + peak AC)
;3V
;30V
;25V
RMS
RMS
RMS
±5% of input (both ranges)±5% of input (both ranges)
DC to ≥500 MHzDC to ≥1GHz
≤700 ps≤350 ps
42 V range:
>55 dB at 30 kHz
>50 dB at 1 MHz
>18 dB at 250 MHz
42 V range:
>55dBat30kHz
>50 dB at 1 MHz
>18 dB at 250 MHz
Operating: 0 to +40 °C (32 °F to +104 °F)
Nonoperating: –55 to +75 °C (–67 °F to +167 °F)
1
Operating: 0-90% RH, tested at + 30 to +40 °C (+86 °F to +104 °F)
Nonoperating: 0-90% RH, tested at +30 to +60 °C (+86 °F to +140 °F)
symbol.
WARNING. To avoid a burn hazard at high ambient temperatures, do not touch the probe with bare hands at nonoperating
mperatures above +50 °C .
te
1TDP0500 & TDP1000 Technical Reference
Typical Characteristics
Typical characteristics (Tables 2 and 3) describe typical but not guaranteed performance.
Table 2: Typical electrical characteristics
Differential input resistance, DC coupled1 MΩ
Common mode input resistance500 kΩ
Differential input capacitance
Common-mode input capacitance
Harmonic disto rtion
Offset accuracy
Differential offset range±42 V (both ranges)
Common mode rejection ratio(See Figure 5 on page 4.) (See Figure 6 on page 5.)
System noise
Referred to probe output
Referred to probe input
Input impedance
Bandwidth limit-3 dB at 1 MHz
Propagation delay
<1.0 pF at 100 MHz
<2.0 pF per side at 100 MHz
≤1.5% measured using 495 mV
100 MHz
±10 mV
<2.0 mV
<10.0 mV
(4.25 V range), <1.0 mV
RMS
(4.25 V range), <50.0 mV
RMS
(See Figure 7 on page 5.)
-3 dB at 10 kHz
-3 dB at 100 Hz
6.5 ns from probe tip to output
RMS
(or 1.4 V
(42 V range)
RMS
RMS
Specifications
) output at
P-P
(42 V range)
Figure 1: Typical bandwidth (TDP0500)
TDP0500 & TDP1000 Technical Reference2
Specifications
Figure 2: Typical bandwidth (TDP1000)
Figure 3: Typical rise time (TDP0500)
3TDP0500 & TDP1000 Technical Reference
Specifications
Figure 4: Typical rise time (TDP1000)
Figure 5: Typical Common-Mode Rejection Ratio (TDP0500)
TDP0500 & TDP1000 Technical Reference4
Specifications
Figure 6: Typical Common-Mode Rejection Ratio (TDP1000)
Figure 7: Typical in put impedance versus frequency
5TDP0500 & TDP1000 Technical Reference
Loading...