Tanita WB-150MA Instruction Manual

GB
(REMOTE DISPLAY VERSION)
(COLUMN MOUNTED VERSION)
Please read this Instruction Manual carefully and keep it for future reference.
ELECTRONIC SCALE
WB-150MA
Temperature : 5°C – 35°C Relative Humidity : 30% – 80%
(no condensation)
Maximum Altitude : 3,000m ASL Pressure :
86kPa –106kPa
Temperature : -10°C – 60°C Relative Humidity : 10% – 90% (no condensation)
<Storage Conditions><Usage Conditions>
To avoid malfunctions, avoid storing the equipment in a location that is subject to direct sunlight, significant temperature changes, dampness, large amounts of dust, the risk of vibration or impact, or near naked flames.
GB
2
Intended use
Efficiency
This equipment has been calibrated as a precision weighing instrument and can be
used to certify weights and/or business transactions.
It can be used to obtain reference data during medical examinations, such as peri-
odic checkups, and can help the prevention of obesity.
Measurements can be taken quickly and easily, causing minimal inconvenience to the patient during measurements.
GB
3
Contents
For Your Safety 4
1.Product Assembly and Components 8
Product Components
Remote Display Version
Column Mounted Version
Control Panel
Symbols and their Meanings
2.Preparation for Use 10
Level Adjustment 10
10
11
Installation of the Equipment (Remote Display Version)
Installation of the Equipment (Column mounted version)
12Preparing the Power Supply
Before use
3.Operating Instructions 13
Preparation (Automatic power-off to save battery life) 13
Preparation (Weight Lock Mode) 14
Preparation (Confirmation Software Version Number) 14
Measuring Weight 15
How to use
4.Output Data Format 18
5.Troubleshooting 19
Specifications 20
When needed
9
9
8
8
8
GB
4
Before use
(For Your Safety)
For Your Safety
Warning
Warning
Caution
This mark indicates actions that are prohibited.
This mark indicates instructions that must always be followed.
Warning
Do not handle the plug with wet hands.
Doing so may result in electric shock, fire, or leakage.
Do not modify this equipment in any way.
Doing so may cause electric shock or injury, or affect the accuracy of analysis.
Use only a correctly wired outlet.
Only genuine cables and equipment can be used.
Do not use a multi-plug adapter.
Doing so may cause fire.
Failure to follow instructions highlighted with this mark could result in death or severe injury.
Failure to follow instructions highlighted with this mark could result in injury or damage to property.
Keep this equipment away from flammable gas and oxygen-rich environments.
This section explains precautionary measures to be taken to avoid injury to the users of this device and others, and to prevent damage to property. Please familiarise yourself with this contents to ensure safe operation of this equipment.
GB
Before use
(For Your Safety)
5
Caution
Keep away from water.
Do not jump on the equipment.
Do not insert fingers into gaps or holes.
Do not apply force to the display.
The screen panel may break and cause injury.
Firmly hold the bottom of the equipment with both hands when moving the equipment.
Ensure that the batteries are inserted with the correct polarity whenever applicable.
Continually monitor both the user and the equipment for anomalies.
Ignoring this requirement may result in electric shock to the user or malfunctions of the equipment.
If the equipment will be out of use for a prolonged period, unplug the AC Adapter from the wall socket.
If the equipment will be out of use for a prolonged period, remove the batteries before storing the equipment.
The data provided by this equipment should be interpreted by a licensed professional.
Be sure to clean the scale platform with the appropriate disinfectant after each use.
Prohibited
Always…
If an anomaly in the user or equipment is discovered, take the appropriate action such as stopping the equipment while ensuring the safety of the user.
When connecting a computer or peripherals, be sure that compliance with IEC60601-1 (EN60601-1) is maintained. If the power is supplied in compliance with IEC60950(EN60950), power must be supplied from a medical isolating transformer. Keep at a minimum distance of
1.5m between each item of equipment during operation.
6
How to Measure Correctly
For Accurate Measurements
Avoid subjecting the equipment to excessive shocks or vibrations.
Avoid measuring in multiple locations that have greatly differing temperatures.
Place the weighing platform on a level and stable surface.
Measurement may not be possible on a surface that is strongly vibrating.
Avoid using near equipment that emits electromagnetic waves.
To ensure accurate measurement, do not touch the user during measurement.
Always…
Scheduled Maintenance
TANITA recommends that each facility conduct periodic checks for each unit.
1. Check the following at least daily:
• That the unit is on a stable and level surface.
2. Visually inspect the following at least weekly:
• The display for any damage or contamination
• All cables, cords, and connector ends for damage or contamination
• All safety-related labelling for legibility
• All accessories (sensors, electrodes, etc.) for wear or damage
• Documentation for the current equipment model
3. Visually inspect the following at least monthly:
• Screws on equipment
Do not wipe the equipment with corrosive chemicals (benzene, cleaner, etc.).
Please use a neutral detergent to clean the equipment.
The warranty will be invalidated if an unauthorised person attempts to disassemble or repair any of the parts of this equipment.
If the equipment malfunctions, please consult your nearest Tanita sales office or agent.
The equipment should be inspected in accordance with the regulations of your country.
Always…
GB
Before use
(How to Measure Correctly)
Doing so may cause inaccurate measurements. Allow the equipment to stand for at least 2 hours before using if it is moved to a new location with a temperature difference of 20˚C or more.
Update settings, replace items, or call for service as necessary based on the results of the visual inspection. Do not use the unit if you see any signs of damage. Equipment that has been damaged must be checked for proper operation by qualified personnel before using again.