SYSMEX XW-100 Start-up Quick Manual

BEYOND A BETTER BOX
XW™-100 AUTOMATED HEMATOLOGY ANALYZER
CLIA WAIVED
START-UP QUICK GUIDE
Code No. AD794857 en-am PRINTED IN JAPAN Date of Issue: 11/2017 Document Version: 2.0 Software Version: Ver.1© SYSMEX CORPORATION 2017
X W-10 0 S TA RT-UP
Start-up
1. Choose a convenient work space:
Near an internet connection (provided Ethernet cable is 6.6 feet long) and a 3-prong power outlet
(provided power cord is 6.6 feet long).
Ensure the space is level, dry, dust-free, and protected from splashing water.
Avoid installation within 1 foot of devices that may cause potential interference, such as radios, computers,
and wireless devices.
Environmental conditions:
Ambient temperature: 15 - 25
Relative humidity: 30% - 85%
2. Unpack the XW-100. Ensure that the Start-up Quick Guide is present and perform a parts check on the boxes that arrive with the analyzer:
o
C or 59 - 77 oF
Box 1: (open rst)
Reagent Tray
Thermal Paper
Waste Bottle
Adapters (Green and White)
Clinician's Quick Guide
Operator's Quick Guide
Shutdown Quick Guide
Technical Specications
3. Remove the XW-100 and Power Cord from Box 2.
4. Connect the Power Cord to the XW-100 by inserting it into the socket in the back.
5. Insert the Power Cord into a 3-prong outlet. (Continued on next page)
SAFETY WARNINGS:
The XW-100 weighs approximately 38 lbs.; DO NOT try to lift it
without help.
Install the instrument with the following clearance, for ventilation:
At least 6 inches at the front and sides.
At least 12 inches at the back.
Box 2:
Power Cord
Ethernet Cable
Barcode Reader and Cable
Reagent Tubing
COIL TUBE SS-10
Instructions for Use
DO NOT place anything on, bend, kink, or pull the Power Cord.
Improper grounding of the instrument can cause electrical shock.
The printer head can get very hot. DO NOT touch! Only use
paper recommended by Sysmex.