1 302A050=> Not Applicable 2002/04/18 Registering XT-2000i
2 200219=> None 2002/05/20 Updating XT-2000i Parts
3 302B058=> XT-2000i S/N
4 302C058=> Not fixed yet. 2002/06/24 Upgrading loader program
5 200222=> - 2002/07/11 Updating XT-2000i Service
6 302F006=> Not applicable 2002/08/21 Registering Rinse Cup to
7 200239=> None 2002/10/10 Correcting XT-2000i Parts
8 302G018=> XT-2000i: 11332-
9 302G032=> XT-2000i: 11252-
10 302G038=> S/N: Not Fixed Yet 2002/12/17 Providing the optional
11 200243=> --- 2002/12/17 Updating XT-2000i Service
12 302A034=>
13 302J032=> Not fixed yet 2003/01/07 Changing intake path to
14 302J059=> S/N not fixed. 2003/01/23 Changing cubitaner cap
15 302J070=> Not fixed 2003/01/27 Changing B el l ows With
16 302L002=> XT-2000i: 11474-
17 302L050=> Not applicable 2003/04/01 Setting FCM sheath
18 303B019=> Serial Nos. have
Applicable
S/N
11001-
XT-1800i: 11144-
XT-1800i: 11054-
XT-2000i: 11428XT-1800i: 11218-
XT-1800i: 11243-
not fixed yet.
Released
Date
Barcode Reader
List
2003/01/17 Changing piercer unit
syringe to improve
durability
to version 00-03
Manual
conform to various test
tubes
List wrong description
2002/11/01 Adding water proof sheet
to the valve unit
2002/11/01 Changing aspiration tubing
length to prevent
aspiration error
waste level sensor
Manual
2003/01/07 Enlarging reaction unit
nipple to prevent from
clogging or leakage
avoid MV diaphragm
deforming
opener
Metal
2003/02/20 Changing target value of
the optical alignment
syringe seal replacement
interval in the service
manual
2003/04/04 Inserting check valve
prevent sampling error
Summary Page
A-1-1, A-1-100
XT-2000i Service Manual, pages, A-1-11 (1/2),
A-1-55, A-1-57, A-1-71, A-1-77, A-4-1,
Alphabetical Index, Code No. Index
A-1-36, A-1-37
A-1-93
1-1, 1-2, 1-5 through 1-9, 1-11, 1-12, 1-14, 1-17
through 1-21, 1-30, 1-32, 2-2, 2-4, 2-5, 2-6, 2-8,
2-11, 2-15, 3-16, 3-18, 4-14, 4-15, 4-17, 4-18,
5-17, 5-18, 6-i, 6-ii, 6-11, 6-26 through 6-28,
7-contents, 7-1, 7-4 through 7-27, A-1-6, A-1-12,
A-1-13, A-1-20, B-contents, B-8 through B-11
through B-14, B-16, B-31 through B-33, B-51
through B-54
A-1-38
A-1-57, A-1-71, A-5-7-A-5-8, A-6-1, A-6-17
A-1-14 through A-1-19
A-1-41, A-1-42
7-2, A-1-12, A-1-13, A-1-21
Section 2 contents, 2-11, 2-16 through 2-35,
Section 4 contents, 4-24 through 4-43, Section 7
contents, 7-28 through 7-31
A1-45
A-1-18
A-1-2, A-1-3, A-1-99, B-1, B-2
A-1-12, 13
4-17, 4-18, B-55
2-3
A-1-49, 7-1
XT-Series - 1/5 -
Page 2
Rev
NO.
ECR/TB
NO.
19 303C009=> S/N has not fixed
20 303A064=> S/Ns Not Fixed
21 200319=> 2003/06/13Updating the XT-Series
22 303D038
23 303D042= S/N has not fixed
24 200328
25 303F038 Applicable SNs
26 303F039
27 200329
28 200337
29 303G054
30 303I035
31 200346
32 200417
33 303D025
34 304C055
Applicable
S/N
yet.
Yet (applicable
after depleting the
manufacturer's
inventory)
Not Fixed
yet
S/N not applicable 2003/07/02
have not been
fixed yet
Parts will be
changed when our
stock is depleted.
Not applicable
S/N not applicable 2003/10/07
XT-2000i S/N
11690 and
thereafter
XT-1800i S/N
11617 and
thereafter
S/N not fixed yet.
For parts ordered
in December 2003
and thereafter.
S/N not applicable 2003/11/25
S/N not applicable 2004/06/10
XT-1800i S/N
11639 and
thereafter
S/N not fixed yet. 2004/10/25
Released
Date
2003/04/08Changing Diaphragm
2003/04/11Changing f an on PU17 A-2-5
2003/06/23Changing s tainless steel
2003/06/24
2003/07/24
2003/07/24
2003/07/25
2003/12/10
2003/11/13
2004/06/29
Summary Page
Pump No.55 to prevent
leakage to vacuum line
Service Manual
used in the Flow Cell to
titanium to avoid optical
axis error
Adapting bottle stand for
common use and updating
OM
Updating XE-Series
Service Manual
Changing material and
shape of Support No.186
to prevent breakage
(Retrofit No.0307)
Changed O-ring used on
Float Switch
Adding DP volume
verification procedure to
XT-Series Service Manual
Updating the XT-Seris
Service Manual
Adding a cover to LD
driver board to prevent
Laser Tube Aged error
Changing Barcode Reader
to improve reading
performance
Digital tool for XT-Series
optical Alignment
Updating the XT-Series
Service Manual
Deleting soleniod valves
for RET
Adapting mechanical parts
in FCM detector block for
common use
A-1-16, A-1-17
Sec 1 TOC, 1-32, 2-16, Sec 4 TOC,
4-44, Sec 5 TOC, 5-20 through 5-65,
Sec 7 TOC, A-4-1, Appen B TOC, B-60
through B-62
A-1-60, A-1-61
A-1-2, A-1-3
A-2-8, A-2-9
A-1-32
A-1-31, A-1-32
4-45, 4-46
1-20 through 1-31
2-7, 4-1, A-1-56, A-1-57
A-1-100
4-1, 2-7
4-21, 4-33, 5-2, 5-6, 5-8 through 5-10, 5-66
through 5-71, 6-5, B-25, B-64 through B-77
A-1-16, A-1-17, A-1-20, A-1-21, 7-1
A-1-57, A-1-65 througj A-1-67
XT-Series - 2/5 -
Page 3
Rev
NO.
ECR/TB
NO.
35 304E029
36 304I080
37 304G043-R
38 304J056
39 200507
40 304J010
41 304I095
42 304J004
43 305A045
44 305C016
45 305B026
46 304K031
47 200517
Applicable
S/N
XT-2000i S/N
12112 and
thereafter
XT-1800i S/N
12322 and
thereafter
S/N not fixed yet 2004/11/16
XT-2000i S/N
12105 and
thereafter
XT-1800i S/N
12294 and
thereafter
XT-2000i S/N
12150, 12151,
12153, 12155,
12157 through
12159, 12162,
12165, 12169,
12201, 12204,
12207, 12209,
12211, 12215,
12259 and
thereafter
XT-1800i S/N
12389 through
12393, 12397,
12398, 12403,
12409 through
12419, 12424,
12426, 12428,
12429, 12433,
12434, 12479,
12484, 12485,
12565 and
thereafter
Not applicable 2005/03/03
XT-1800i S/N
12466 and
thereafter
XT-2000i S/N
12218 and
thereafter
Not applicable 2005/03/11
XT-2000i S/N
12281 and
thereafter
XT-1800i S/N
12613 and
thereafter
S/N not
applicable
The S/Ns are not
fixed yet.
XT-1800i S/N
12847 and
thereafter
XT-2000i S/N
12469 and
thereafter
S/Ns are not fixed
yet.
Countermeasure against
tube disconnection at WC
Registering a replacement
kit as countermeasure
against sticking silicone
tubing
Countermeasure against
leaking Manifold MV No.14
Updating the XT-Series
service manual to add
timing chart
Changing material of
Piston No.63
Changing waster line joint
on FCM Detector Block
Changing non-return valve A-1-49, A-1-50
Registering piercer set for
various collection tubes
Upgrading program
version to 00-12
Changing Air Valves
"WTKV" used on XT and
KX
Countermeasure against
regulator clogging
Updating XT-series
Service Manual
A-1-12
A-1-28
A-1-78, A-1-80
7-TOC, 7-32 through 7-37
A-1-53
A-1-56, A-57 (2/2)
A-1-3
B-33 through B-38, B-78, B-79, B TOC
A-1-15, A-1-17, A-1-19, A-1-78, A-1-82
A-1-12, A-1-13 (1/2)
1-31_32,1-TOC,4-50_52,4-TOC,B-77_78,B-TOC
XT-Series - 3/5 -
Page 4
Rev
ECR/TB
NO.
NO.
48 305F038_1
49 304F036
50 200527
51 305L004
52 305H084
53 306B072
54 306A007
55 305L047
56 306B022
57 306C036
58 305L030
59 305J073
60 306C059
61 306C104-1
62 306H037
63 306E075
64 306E066-1
65 200633
Applicable
S/N
S/N not applicable 2005/09/07
Not Fixed 2005/09/29
S/N not applicable 2005/10/17
XT-1000i S/N: Not
fixed
XT-800i S/N: Not
fixed
XT-2000i S/N
12881 and
thereafter
XT-1800i S/N
13847 and
thereafter
S/N not applicable 2006/03/31
XT-2000iV S/N
A1021, A1032 and
thereafter
XT-1800iV S/N
A1012, A1013,
A1019 through
A1021, A1026 and
thereafter
Not fixed yet 2006/06/06
S/N not applicable 2006/06/27
Not fixed yet 2006/07/05
Not fixed yet 2006/07/10
S/N not fixed Yet 2006/07/18
XT-2000i S/N:
13160 and
therafter
XT-1800i S/N:
14426 and
therafter
S/N not applicable
Parts will be
changed when our
stock is depleted.
S/N not fixed yet 2006/09/04
XT-2000i S/N
13285 and
thereafter
XT-1800i S/N
14662 and
thereafter
S/N: Not
Applicable
Spare parts will be
changed when our
stock is depleted.
S/N not applicable 2006/11/24
Released
Date
2006/01/24
2006/01/26
2006/05/31
2006/07/26
2006/08/11
2006/09/29
2006/11/06
Summary Page
Changing
Photo-Interrupter
Modifying Valve Seal
No.44 for improved
countermeasure against
leakage
Updating XT-series
Service Manual
Eliminating one FAN
CNDC12Z7 from PU-17
Chassis Complete
Updating XT Operator's
Manual (including some
service manual updates)
Registering Solenoid
Valve GA010E1-PL-005W
Upgrading the
XT-2000iV/XT-1800iV
programs to version 00-08
Replacing RBC cylinder
to improve reproducibility
at installation
Changing SSC position
adjustment
Countermeasure against
Rack Move Error 1
Changing Fuses used
on XT Series and KX
Series
Changing Holding
Material No.122
Changing the knob used
in XT RBC-Detector Block
Replacing Pharmed
Tubing due to part being
discontinued
S/N not applicable.
Parts will be
changed when our
stock is depleted.
13334 and
thereafter
XT-1800i S/N
14727 and
thereafter
Spare parts will be
changed when our
stock is depleted.
12731 and
thereafter
XT-1800i S/N
13416 and
thereafter
Released
Date
2007/03/19
2007/04/16
2007/04/17
2007/05/30
2007/06/06
2007/06/06
2007/06/07
Summary Page
Changing RBC isolation
chamber to prevent PLT
sampling error
Changing Power Cord
included in Sysmex supply
parts to conform to the
RoHS directive (European
markets only)
Adding antistatic switch
to Optical Unit No.22 to
prevent electrostatic
damage
Changing blood asp
sensor
Changing slide rail used
on Sysmex products
Changing Flow Cell of
XT-Series
Registering assemblies for
XT-Series Main Chassis
Comp. XT
Changing waste line tubing
used on XT-Series
A-1-80
A-1-5, A-1-100, チ-2-5
2-7
A-1-42
A-1-87, A-1-88
A-1-61
A-1-11
A-1-27,A-1-28, A-4-1
XT-Series - 5/5 -
Loading...
+ hidden pages