(Version 2)
XP Series PM Standard Procedure
[Required jigs and controls, etc.]
Code Product Name Remarks
- EIGHTCHECK Control blood
Items to be checked
Operation
time
Procedure
Checking setting
values
Print Setting Value
5 min.
Go to [Menu] -> [Service] -> [Settings] -> [Print
Settings] to print all setting values.
HOST Output Setting OFF
Go to [Menu] -> [Settings] -> [Host Output
Settings] to change "Connection" to "Disconnect".
Data analysis
before
maintenance
Check background count level
10 min.
Analyze normal sample or control material and
print average values or record them using stored
data/QC chart.
Check data using control material
or samples
Perform background check.
Check adjustment
values before
maintenance
HGB background convert
WBC clog level
RBC clog level
0.05MPa
-0.0333MPa
Go to [Menu] -> [Service] -> [Service Data]. Print
out and check service data for the latest samples.
Go to [Menu] -> [Maint.] -> [Status Display] to
check pressure.
Removing covers Remove front and side covers.
SRV
Disassemble and clean SRV
15 min.
Use CELLCLEAN and distilled water to perform
cleaning.
(Refer to Instructions for Use Chapter 12.)
Clean the tray
Replace the aspiration pipette
Replace PIPETTE NO.57 using FITTING NO. 9-
A.
Rinse Cup
Clean Rinse Cup 5 min.
Use CELLCLEAN and distilled water to perform
cleaning.
(Refer to Instructions for Use Chapter 12.)
Replace Tube Pharmed for Waste
Line
5 min.
Replace by using Tube Pharmed 1/4 IN X 1/8 IN.
(Refer to Service Manual "Schematics".)
Waste Chamber
Replace Waste Chamber (Every 5
years)
5 min. Replace WASTE CHAMBER ASSY NO.35.
Disassemble and clean Waste
Chamber (when not replacing
Waste Chamber).
5 min.
Wipe the interior of the chamber and the float
switch using gauze impregnated with
CELLCLEAN.
Replace Waste Chamber Drain
Line silicon tubes.
5 min.
Replace silicon tube F739-8.
(Refer to Service Manual "Schematics".)
Bellows Unit
Disassemble and rinse Bellows Unit 5 min.
Wipe the unit and the needle tip using gauze
impregnated with ethanol.
Clean orifice tube 5 min.
Inject ethanol using an injection syringe to clean.
If the tube is extremely dirty, replace the whole
tube.
TUBE POLYURETHANE 1.8MMX3.4MM
(1400mm)
Pneumatic Unit
Rinse Trap Chamber
30 min.
Clean Trap Chamber using distilled water.
Replace Pneumatic Unit protection
filter
Replace FILTER NO. 9.
Replace Compressor
Replace by using KX-21 Pneumatic Unit (XP
series).
(Refer to Service Manual Parts List.)
Pressure and
vacuum
adjustment
0.5kg/cm2(0.05MPa)
Pressure adjustment
10 min.
Adjust pressure to 0.05Mpa using Regulator.
(Refer to Instructions for Use Chapter 12.)
250 mmHg (-0.0333MPa)
Vacuum adjustment
5 min.
Adjust vacuum to -0.0333MPa using the bellows
unit. (Refer to Instructions for Use Chapter 12.)
Transducer
Clean WBC transducer aperture
10 min.
Apply CELLCLEAN to a brush for clog removal
and clean the aperture. (Refer to Instructions for
Use Chapter 12.)
Clean RBC transducer aperture
Clean inside of Transducer
Remove dust, etc. and wipe with gauze
impregnated with ethanol. Then, wipe with dry
gauze.