SYSMEX XE-5000 Instruction Manual

Automated Hematology Analyzer
XE-5000
Instructions for Use
CHAPTER 1 Introduction CHAPTER 2 Safety Information CHAPTER 3 Design and Function CHAPTER 4 Reagents CHAPTER 5 Before Using CHAPTER 6 Operation CHAPTER 7 Quality control CHAPTER 8 Calibration CHAPTER 9 Cleaning and Maintenance CHAPTER 10 Troubleshooting CHAPTER 11 Technical Information CHAPTER 12 Warranty CHAPTER 13 Index
SYSMEX CORPORATION
KOBE, JAPAN
Copyright © 2006 - 2008 by SYSMEX CORPORATION
All rights reserved. No part of this manual may be reproduced in any form or by any means whatsoever without prior written permission of SYSMEX CORPORATION.
Code No. 461-2642-1 PRINTED IN JAPAN Date of Last Revision: December 2008

Table of Contents

Table of Contents
Chapter 1 Introduction .................................1-1
1.1 Hazard information in this manual............................1-3
1.2 Protected names ......................................................1-4
1.3 Measurement parameters ........................................1-4
Chapter 2 Safety Information....................... 2-1
2.1 Specified conditions of use........................... ... ... .... .. 2-1
2.2 General information..................................................2-1
2.3 Installation ................................................................2-2
2.4 Electromagnetic compatibility (EMC) .......................2-3
2.5 Avoidance of infection............................................... 2-3
2.6 Handling of reagents ................................................2-4
2.7 Quality control materials........................................... 2-5
2.8 Laser ........................................................................2-5
2.9 Maintaining the instrument....................................... 2-6
2.10 Disposal of waste fluids, disposables and
instruments...............................................................2-6
2.11 Marking on the instrument........................................2-7
2.12 Personnel ............................................................... 2-12
2.13 Computer viruses................................................... 2-12
Chapter 3 Design and Function ..................3-1
3.1 Overview................................................................... 3-1
3.2 Main Unit ..................................................................3-2
3.3 Pneumatic Unit....................................................... 3-11
3.4 Sampler Unit...........................................................3-13
3.5 Information Processing Unit (IPU).......................... 3-14
3.6 Analysis mode........................................................ 3-17
Chapter 4 Reagents......................................4-1
4.1 General information..................................................4-1
4.2 CELLPACK...............................................................4-1
4.3 CELLSHEATH ..........................................................4-1
4.4 STROMATOLYSER-FB.............................................4-2
4.5 STROMATOLYSER-4DL...........................................4-2
4.6 STROMATOLYSER-4DS ..........................................4-3
4.7 STROMATOLYSER-NR(L)........................................ 4-3
4.8 STROMATOLYSER-NR(S).......................................4-4
4.9 SULFOLYSER ..........................................................4-4
4.10 STROMATOLYSER-IM............................................. 4-5
4.11 RET SEARCH (II) (diluent)
RET SEARCH (II) (dye solution) ..............................4-5
4.12 CELLCLEAN ............................................................4-6
4.13 Control blood (e-CHECK(XE))..................................4-7
4.14 Labeling.................................................................... 4-7
4.15 Symbols used on the labels .....................................4-8
Revised July 2007 EU
Sysmex XE-5000 Instructions for Use I
Table of Contents
Chapter 5 Before Using................................5-1
5.1 Storage prior to transport and installation ................5-1
5.2 Preparation...............................................................5-1
5.3 Peripherals................................................................5-2
5.4 Additional components .............................................5-2
5.5 Basic equipment settings..........................................5-3
Chapter 6 Operation..................................... 6-1
6.1 Summary of Main Unit operation..............................6-1
6.2 Main unit menu tree..................................................6-3
6.3 Summary of Information Processing Unit (IPU)
operation...................................................................6-4
6.4 Information Processing Unit (IPU) menu bar............6-5
6.5 Alarm sound .............................................................6-6
6.6 Operator checks .......................................................6-6
6.7 Turning ON the power...............................................6-8
6.8 Logging on to the Information Processing
Unit (IPU)................ ...... ....... ...... ....... ...... ....... ...... .....6-9
6.9 Self-checks...............................................................6-9
6.10 Pressure and vacuum gauges check......................6-11
6.11 Logging on the Main Unit............................... ...... ...6-11
6.12 Auto output settings check......................................6-11
6.13 Quality control.........................................................6-11
6.14 Conditions for sample analysis...............................6-12
6.15 Analysis mode .................................. ...... ....... ...... ...6-12
6.16 Preparing sample analysis .....................................6-13
6.17 Sample analysis in manual mode...........................6-17
6.18 Sample analysis in capillary mode .........................6-19
6.19 Sample analysis in sampler mode..........................6-21
6.20 Sample analysis in manual closed mode ...............6-24
6.21 Display analysis results ..........................................6-25
6.22 Analysis result output .................................... ...... ...6-26
6.23 HPC analysis..........................................................6-27
6.24 Body fluid analysis..................................................6-33
6.25 Timer processing....................................................6-38
6.26 Shutdown................................................................6-39
Chapter 7 Quality control.............................7-1
7.1 Quality control materials...........................................7-1
7.2 Quality control methods............................................7-1
7.3 Preparation...............................................................7-2
7.4 Quality control analysis.............................................7-5
7.5 QC data display........................................................7-7
7.6 Troubleshooting ......................................................7-12
7.7 Change lot ..............................................................7-13
7.8 Lot No.....................................................................7-13
7.9 Target/Limit.............................................................7-16
7.10 Display order...........................................................7-21
7.11 File Information....................................... ................7 -22
II Sysmex XE-5000 Instructions for Use
Revised April 2007
Table of Contents
Chapter 8 Calibration...................................8-1
8.1 Calibration execution timing......................................8-1
8.2 Samples used for calibration ....................................8-1
8.3 Reference values...................................................... 8-1
8.4 Automatic calibration ................................................8-2
8.5 Displaying the last sample data................................ 8-6
8.6 Manual calibration .................................................... 8-6
8.7 Calibration history..................................................... 8-9
Chapter 9 Cleaning and Maintenance.........9-1
9.1 Execution of shutdown ............................................. 9-2
9.2 Removing fluid from the Pneumatic Unit
trap chamber ............................................................ 9-3
9.3 Cleaning the sample rotor valve............................... 9-4
9.4 Cleaning the manual rinse cup................................. 9-7
9.5 Cleaning the sample rotor valve tray........................ 9-9
9.6 Cleaning the piercer tray ........................................ 9-10
9.7 Removing clogs (clog removal sequence)..............9-11
9.8 Cleaning the IMI detector aperture.........................9-12
9.9 Cleaning the RBC detector aperture...................... 9-14
9.10 Removing air bubbles from the flow cell in the optical
detector block.........................................................9-17
9.11 Cleaning the flow cell in the optical
detector block.........................................................9-17
9.12 Replacing the waste container (option) ..................9-18
9.13 Replacing and registering reagents........................9-19
9.14 Reagent replacement log display function..............9-31
9.15 Remaining reagent volume function.......................9-36
9.16 Replacing the piercer ............................................. 9-38
9.17 Replacing the hand clipper..................................... 9-41
9.18 Replacing rubber plate No. 39................................ 9-43
9.19 Replacing the fuses................................................9-44
9.20 Replacing the HEPA filter (option).......................... 9-45
9.21 Adjusting pressure and vacuum.............................9-46
9.22 Supplies and replacement parts............................. 9-51
9.23 XE-5000 maintenanc e and ins pe cti on check list....9-5 2
Chapter 10 Troubleshooting........................10-1
10.1 HELP screen .......................................................... 10-1
10.2 Error log .... ...... ...... ....... ...... ..................................... 10-1
10.3 Error message list ..................................................10- 3
10.4 Troubleshooting guide ............................................10-9
10.5 Test....................................................................... 10-30
Revised April 2007 EU
Sysmex XE-5000 Instructions for Use III
Table of Contents
Chapter 11 Technical Information............... 11-1
11.1 Performance/specifications of the XE-5000............11-1
11.2 Possible sample interferences................................11-9
11.3 Interface protocol..................................................11-10
11.4 Program version ...................................................11-10
11.5 Principles..............................................................11-11
11.6 Contents of the packages.....................................11-32
11.7 Check before installation ......................................11-36
11.8 Grounding.............................................................11-36
11.9 Installation environment............. ....... ...... ....... ...... .11-36
11.10 Installati on sp ace.................................... ....... ...... .11-37
Chapter 12 Warranty ....................................12-1
Chapter 13 Index .......................................... 13-1
IV Sysmex XE-5000 Instructions for Use
Revised April 2007

1. Introduction

CHAPTER 1 Introduction
The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The XE-5000 can analyze and output the results for 67 parameters. WBCs, nucleated RBCs and reticulocytes are analyzed by the optical detector block based on the fluorescence flow cytometry method using semiconductor laser. RBCs and platelet count are analyzed by the RBC detector using the Hydro Dynamic Focusing method based on the fluorescence flow cytometry method. Hemoglobin (HGB) is analyzed by the HGB detector based on the SLS hemoglobin determination method. The existence of immature cells are analyzed by the IMI detector based on the RF/DC detection method and the DIFF channel using fluorescence flow cytometry. Analysis data is displayed on the panel keypad LCD screen and the screen of the Information Processing Unit (IPU).
Note:
Data generated by the XE-5000 is not intended to replace professional judgment in the determination of a diagnosis or in monitoring patient therapy.
Note:
Operate the instrument as instructed. Reliability of test results cannot be guaranteed if there are any deviations from the instructions in this manual. If the instrument fails to function properly as a result of either the user's operation not specified in the manual or the user's utilization of a program not specified by Sysmex, the product warranty would not apply.
Revised April 2007
Sysmex XE-5000 Instructions for Use 1-1
CHAPTER 1 Introduction
Contact Address
Manufacturer
Authorized Representative
EC REP
Americas
SYSMEX CORPORATION 1-5-1 Wakinohama-Kaigandori Chuo-ku, Kobe, Hyogo 651-0073 JAP AN
European Representative
SYSMEX EUROPE GmbH Bornbarch 1 D – 22848 Norderstedt, Germany Phone: +49 40 5 27 26-0 Fax: +49 40 5 27 26-100
SYSMEX AMERICA, Inc. 1 Nelson C. White Parkway, Mundelein, IL 60060 U.S.A. Phone: +1-847-996-4500 Fax: +1-847-996-4505
Asia-Pacific
SYSMEX ASIA PACIFIC PTE LTD. 2 Woodlands Sector 1, #01-06 Woodlands Spec trum, Singapore 738068 Phone: +65-6221-3629 Fax: +65-6221-3687
Ordering of Supplies and Replacement Parts
If you need to order supplies or replacement parts, please contact your local Sysmex representative.
Service and Maintenance
Please contact the Service Department of local Sysmex representative.
CE-mark
The IVD-system described in this manual is marked with a CE-mark which confirms the observance of the essential requirements of the following European directive:
98/79/EC IVD Directive
1-2 Sys mex XE-5000 Instructions for Use
Revised July 2007 EU

1.1 Hazard information in this manual

Note, Information, Caution and Warning statements are presented throughout this manual to call attention to important safety and operational information. Non-compliance with this information compromises the safety features incorporated in the analyzer.
This symbol indicates a possible hazardous situation which, if not avoided, may result in infection by pathogens and others.
High risk. Ignoring this warning could result in personal injury to the operator.
CHAPTER 1 Introduction
Risk of infection
Warning!
Caution, Hot!
Indicates a potential risk of burns or other physical damage in the event of incorrect operation or failure to observe the content.
Caution!
Ignoring this warning could result in property damage or incorrect results. To avoid damage and incorrect measuring results, this caution must be followed.
Information
Minor risk. Considerations that should be observed when operating this instrument.
Caution!
Indicates a potential risk of physical damage to functions of the instrument caused by static electricity discharge of the human body, in the event of incorrect operation or failure to observe the content.
Revised April 2007
Sysmex XE-5000 Instructions for Use 1-3
CHAPTER 1 Introduction

1.2 Protected names

Note:
Background information and practical tips.
• Sysmex is a registered trademark of SYSMEX CORPORATION.
• STROMATOLYSER, CELLPACK, SULFOLYSER, RET SEARCH and CELLCLEAN are registered trademarks of SYSMEX CORPORATION.
• CELLSHEATH is a trademark of Sysmex.
• SMI and SMI Micro/Pettor are registered trademarks of Scientific Manufacturing Industries, Inc.
• VENOJECT and VENOJECT II are registered trademarks of TERUMO Corporation.
• Windows is a registered trademark of Microsoft Corporation.
Other trademarks referenced are property of their respective owners. TM and ® are not specified in this manual.

1.3 Measurement parameters

The XE-5000 can analyze the following 67 parameters: 37 Diagnostic Parameters
1WBC 2RBC 3HGB 4HCT 5MCV 6MCH 7 MCHC 8PLT 9RDW-SD 10 RDW-CV 11 PDW 12 MPV 13 P-LCR 14 PCT 15 NEUT% 16 LYMPH% 17 MONO% 18 EO% 19 BASO% 20 NEUT#
Revised April 2007
1-4 Sysmex XE-5000 Instructions for Use
CHAPTER 1 Introduction
21 LYMPH# 22 MONO# 23 EO# 24 BASO# 25 IG% Immature Granulocyte Percent 26 IG# Immature Granulocyte Count 27 HPC# Hematopoietic Progenitor Cell Count 28 NRBC% 29 NRBC#
Note:
XE-5000 automatically analyzes NRBC if an analysis order is assigned to any of the WBC 5 differential counts or percent data.
30 RET% 31 RET# 32 HFR 33 MFR 34 LFR 35 IRF Immature Reticulocyte Fraction 36 RET-He Reticulocyte Hemoglobin Equivalent 37 IPF Immature PLT fraction
6 Body Fluid Mode Parameters
1 WBC-BF WBC body fluid 2MN% Mononuclear cell percent 3MN# Mononuclear cell count 4PMN% Polymorphonuclear cell percent 5PMN# Polymorphonuclear cell count 6 RBC-BF RBC body fluid
24 Pre-Diluted Mode Parameters
1WBC 2RBC 3HGB 4HCT 5MCV 6MCH 7 MCHC 8PLT 9NEUT% 10 LYMPH% 11 MONO% 12 EO% 13 BASO% 14 NEUT# 15 LYMPH# 16 MONO#
Revised April 2007 EU
17 EO#
Sysmex XE-5000 Instructions for Use 1-5
CHAPTER 1 Introduction
18 BASO# 19 IG% Immature Granulocyte Percent 20 IG# Immature Granulocyte Count 21 NRBC% 22 NRBC# 23 RET% 24 RET#
1-6 Sysmex XE-5000 Instructions for Use
Revised April 2007 EU

2. Safety Information

2.1 Specified conditions of use

The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in clinical laboratories. Only human blood, human body fluids or control blood should be run. Any other use is regarded as non-specified. Use only the reagents and rinse solutions mentioned in this manual. The specified conditions of use also entail the observance of the cleaning and maintenance procedures described in these instructions.

2.2 General information

Before operating this instrument, carefully read this manual, and strictly follow its instructions. Keep this manual for future reference after reading.
CHAPTER 2 Safety Information
Warning!
• Unpacking, installation, and confirmation of initial operation must be done by Sysmex technical representative.
• Keep your fingers, clothes and hair etc. away from the instrument while it is in operation. There is a risk of injury from becoming tangled or trapped in the machinery.
• If the instrument emits an abnormal odor or smoke, turn off the power switch immediately and disconnect the power cord from the Main Unit. If the instrument is used continuously in this condition, fire, electrical shock or injury may result. Contact promptly your Sysmex technical representative.
• Take care not to spill any fluids, or drop wire staples or paper clips into the instrument. This might cause a short circuit resulting in smoke emission. If such an incident occurs, turn off the power switch immediately and disconnect the power cord from the Main Unit. Contact your Sysmex technical representative.
• Do not remove the outer cover while the instrument is in use. There is a risk of electrical shock or instrument failure.
• Do not touch the electrical circuits inside the cover, particularly if your hands are wet. Electrical shock may result.
Revised April 2007
Sysmex XE-5000 Instructions for Use 2-1
CHAPTER 2 Safety Information

2.3 Installation

Warning!
• Never put the power plug in any socket other than the specified voltage. When installing the instrument, be sure to ground it. Otherwise, fire or electrical shock may result.
• Take care not to damage the power cord, not to place a heavy device on it and not to pull it forcibly. Otherwise, the wire may break causing fire or electrical shock.
• When connecting the instrument to a peripheral (host computers and printer etc.), be sure to turn off the power switch beforehand. Otherwise, electrical shock or instrument failure may result.
Caution!
• Install the instrument in a place which is not subject to water splash.
• Install the instrument in a place which is not subject to adverse effects of high temperature, high humidity, dust, direct sunlight, etc.
• Do not give the instrument a strong vibration or impact.
• Install in a well-ventilated area.
• Do not install near devices that cause signal noise, such as radios and centrifugal machines.
• Do not install the instrument near a chemical storage or a place where a gas is generated.
• Do not use this instrument in any operating environment which has electro-conductive or flammable gases, including oxygen, hydrogen and anesthetic.
• This instrument was designed for indoor use only.
2-2 Sysmex XE-5000 Instructions for Use
Revised April 2007

2.4 Electromagnetic compatibility (EMC)

This instrument complies with the following IEC (EN) standards.
• IEC61326-2-6:2005 (EN61326-2 -6:200 5) Electronic equipment for analysis, control and laborat ory use — EMC request.
• EMI (electromagnetic emission (= interference radiation)) Class A requirements are fulfilled.
• EMS (electromagnetic immunity (= resistance against jamming)) For this issue the minimum requirements with regards to immunity are fulfilled.

2.5 Avoidance of infection

Risk of infection
• In principle, all parts and surfaces of the instrument must be regarded as infective.
• Never touch waste or parts having been in contact with waste, with bare hands.
• Should you inadvertently come in contact with potentially infective materials or surfaces, immediately rinse skin thoroughly with water, then follow your laboratory’s prescribed cleaning and decontamination procedures.
• Be careful when handling samples and control blood. Use of personal protective equipment is strongly recommended when operating, maintaini ng, servicing or repairing the instrument.
• If patient samples of control material contact your eyes, wash the area with plenty of water and immediately contact a physician.
• Be careful when handling waste. If you get it on your skin or clothes, wash with disinfectant.
CHAPTER 2 Safety Information
Revised April 2007
Sysmex XE-5000 Instructions for Use 2-3
CHAPTER 2 Safety Information

2.6 Handling of reagents

Warning!
• Do not directly touch reagents. Reagents can cause irritation of the eyes, skin and mucous membranes.
• Should you inadvertently come in contact with reagent, immediately rinse skin thoroughly with water.
• If a reagent should get in your eyes, immediately rinse thoroughly with water and see a physician immediately.
• If reagent is accidentally swallowed, follow recommendations on Material Safety Dade Sheets (MSDS) and see a physician immediately.
• CELLPACK diluent is electrically conductive. If diluent is spilled inadvertently near electrical cables or appliances, there is a risk of electrical shock. Switch the instrument off, unplug it and remove the liquid.
• CELLCLEAN is a strong alkaline detergent. It should not come in contact with skin or clothing. If it happens nevertheless, rinse skin or clothing with plenty of water to avoid injury or damage.
• CELLCLEAN contains sodium hypochlorite. If CELLCLEAN comes in contact with the instrument’s surfaces, there may be a possibility of corrosion. Immediately wipe up CELLCLEAN with a damp cloth.
• Make sure the reagents used with the instruments are kept level or below the Main Unit of the instrument. Do not put reagents on top of the instrument.
Caution!
• Follow directions on reagent containers.
• Avoid letting reagents come in contact with dust, dirt or bacteria especially when installing new cubes.
• Reagents must not be used after their expiration date.
• Handle reagents gently to avoid bubbling. Never shake reagents. Do not use reagents immediately after moving them.
• Take care not to spill reagents. If a reagent is spilled, wipe up with a damp cloth.
2-4 Sys mex XE-5000 Instructions for Use
Revised July 2007

2.7 Quality control materials

2.8 Laser

CHAPTER 2 Safety Information
Caution!
• Do not inject or ingest.
• Follow directions on control ma terial package insert.
• Av oid letting the control material come in contact with dust, dirt or bacteria.
• Control material must not be used after their expiration date.
• Take care not to spill control material. If spilled, follow your laboratory's prescribed cleaning and decontamination procedures.
Warning!
The XE-5000 contains the semiconductor laser unit. This semiconductor laser unit is shielded with the shield box cover, and is provided with an interlock system that prevents laser from oscillation if the cover is removed. Do not open this cover. Otherwise, the laser beam can damage your eyes.
Revised July 2007
Laser Diode
NOTE:This caution label is located
inside the Laser Unit Cover.
Sysmex XE-5000 Instructions for Use 2-5
CHAPTER 2 Safety Information

2.9 Maintaining the instrument

Risk of infection
Always wear protective clothing, gloves and eyeglasses when performing maintenance or inspection to prevent contact with blood contaminated parts. Wash your hands thoroughly following maintenance.
Information
When performing maintenance, use only the tools specially provided for such work. All cleaning and maintenance procedures as described in this manual must be observed for optimal performance.

2.10 Disposal of waste fluids, disposables and instruments

Risk of infection
Use of gloves is strongly recommended when handling waste fluids or instrument consumables. Wash your hands thoroughly following maintenance.
Warning!
Waste fluids, instrument consumables and other waste materials must be disposed of appropriately in accordance with local laws, with due consideration of medical, infectious and industrial wastes.
2-6 Sysmex XE-5000 Instructions for Use
Revised April 2007

2.11 Marking on the instrument

Main Unit - Front View
(1)
(2)
(3)
*1
(4)
(5)
CHAPTER 2 Safety Information
(6)
(7)
(8)
(9)
(Inside)
(1)
WARNING
Be sure to set the stop bar when opening to avoid injury.
(2)
WARNING
Never touch the detector when the power of the Main Unit is turned ON. Otherwise, electrical shock may resu lt.
(3)
CAUTION
1. Dye solution replacement method
• For SNR, always replace dye solution and lyse reagent simultaneously.
• For RED, always replace dye solution and diluent simultaneously.
• For FFS, the message “Replace FFS” is displayed after the analysis of 2000 cycles. When doing so, always press Reagent or Confirm after replacing with new FFS.
2. When the message “Execute Rinse Flowcell” is displayed, always perform Maintenance – 4. Rinse Flowcell.
3. To clean the IMI detector aperture, first run Maintenance – 1. Drain IMI, then turn off the main unit power.
Revised April 2007
Sysmex XE-5000 Instructions for Use 2-7
CHAPTER 2 Safety Information
(4)
When installing this instrument, remove the constant-pressure screw, then take out the anti-sticking spacers which are inserted between the sample rotor valves. At that stage, rinse the surface of each valve cleanly before setting them in place. The fluid channel inside the instruments are rinsed with diluent before shipping from the factory. Background count may be high immediately after unpacking, so rinse thoroughly.
*1: After installing, remove the anti-sticking spacers.
(5)
RISK OF INFECTION
In principle, all parts and surfaces of the instrument must be regarded as infective.
(6)
CAUTION
Always keep the switch in the analysis position.
(7)
WARNING
Never touch the detector when the power of the Main Unit is turned ON. Otherwise, electrical shock may resu lt.
(8)
WARNING
Do not remove this cover when the power to the Main Unit is ON. Doing so may result in injury.
(9)
WARNING
Do not put your fingers inside when the power to the Main Unit is ON. Doing so may result in injury.
2-8 Sysmex XE-5000 Instructions for Use
Revised April 2007
Main Unit - Rear View
(1)
• T o av oid electrical shock, disconnect supply before servicing.
• For the continued protection against risk of fire, replace only with fuse of the specified type and current ratings.
CHAPTER 2 Safety Information
(1)
CAUTION
Fuse Rating
3.15A 250V
Time Lag
Revised July 2007
Sysmex XE-5000 Instructions for Use 2-9
CHAPTER 2 Safety Information
Sampler
(1)
Do not push in the rack by hand.
There is a small risk that such handling could cause a mismatch between sample numbers and analysis results.
Precautions on setting the
sampler
1. Check that the sample tubes fit smoothly into the rack.
2. Check that the rubber stoppers are firmly capped on the sample tube.
3. Insert the sample tube securely all the way to the bottom of the rack.
(1)
Pneumatic Unit - Front View
(1)
(1)
0.25 MPa regulator
(2)
(2)
RISK OF INFECTION
In principle, all parts and surfaces of the instrument must be regarded as infective.
2-10 Sysmex XE-5000 Instructions for Use
Revised April 2007
Pneumatic Unit - Rear View
CHAPTER 2 Safety Information
(1)
(1)
Pneumatic Unit - Right View
(1)
P
V
CAUTION
Do not block the exhaust vent on the rear of the pneumatic unit.
(1)
WARNING
Instrument must be grounded.
Revised April 2007
Sysmex XE-5000 Instructions for Use 2-11
CHAPTER 2 Safety Information

2.12 Personnel

2.13 Computer viruses

Caution!
• This instrument may only be operated by trained personnel having been instructed in its operation.
• Only appropriately trained persons must perform maintenance and repair work. Follow troubleshooting instructions in the manual.
• Unpacking, installation, and confirmation of initial operation must be done by Sysmex technical representative.
Warning!
This system has been tested prior to shipment and is virus free. But it may be infected with virus by improper application on lines of Internet or network. Please investigate and apply the corrective action for computer virus approved by your facility. Sysmex will not be held liable in the case that the customer’s computer is infected with a virus and therefore please take note of the following guidelines in prior .
The following cases are just for your reference. Please study the treatments according to the usage of your computer.
1. Use a virus checker program regularly to check
your system.
2. Do not install any other application program except
a virus protection software.
3. Do not open any file attached to a mail from an
unknown address. Perform a virus check.
4. Do not download from Internet files not concerned
with Sysmex.
5. Files contained in a shared folder should be
subjected to a virus check.
6. Make sure to carry out any virus protection
measures currently used in other computer systems.
2-12 Sysmex XE-5000 Instructions for Use
Revised April 2007
CHAPTER 3 Design and Function

3. Design and Function

3.1 Overview

7 6
1
2
3
5
4
1Main Unit
Analyzes and controls samples.
2 Information Processing Unit (IPU)
Processes data which it receives from the Main Unit.
3 Ledger printer (option)
Prints lists of analysis information or results.
4 Color graphic printer (option)
Prints a hardcopy of analysis results or screen of histograms, scattergrams, etc.
5 Pneumatic unit
Supplies pressure and vacuum used by the Main Unit.
6 Sampler unit
Supplies samples to the Main Unit automatically.
7 Data printer (option)
Prints analysis data in the examination ticket format.
Revised April 2007
Sysmex XE-5000 Instructions for Use 3-1
CHAPTER 3 Design and Function

3.2 Main Unit

Front View
1
5
6
2
3 4
7
1 Front cover
Can be opened upward by hand. Open the cover to inspect or clean inside the Main Unit.
Warning!
When inspecting inside the Main Unit with the front cover opened, be sure to set the stop bar in advance. Otherwise, the cover can drop down and injure your head.
2READY LED
Lights up when the Main Unit enters Ready status.
3 Manual aspiration pipette
Used to aspirate a sample in manual or capillary mode.
4 START switch
Used to start an analysis in manual, capillary or manual closed mode.
5 LCD screen
Displays the status of the Main Unit, sample ID number and analysis data.
6 Panel Keypad
Used to perform basic operation, such as inputting sample ID number, starting the sampler analysis an d se lec ti ng th e an al ysis par a me te r. For details, see “Panel k eypad” .
7 CP cover
This is the protection cover of cap piercing unit.
3-2 Sysmex XE-5000 Instructions for Use
Revised April 2007
CHAPTER 3 Design and Function
Rear View
1
14 15
2
16 17
3 4
5 6 7 8 9
10 11
12
18 19
20 21
22 23
24
13 25
1 Float switch connector
Connected to the float switch of each reagent.
2 Pneumatic Unit control output connector
Used as output connector for controlling ON/OFF of the Pneumatic Unit power. Connected to the connector on the rear panel of the Pneumatic Unit.
3 Fuse holder
This is a 250V, 3.15A (time lag) fuse.
Warning!
• To avoid electrical shock, disconnect supply before servicing.
• For the continued protection against risk of fire, replace only with a fuse of the specified type and current ratings.
4 AC power supply
Supplies power using the provided power cable.
5 ESE inlet nipple (ESE-1)
CELLSHEATH is aspirated via this nipple. Connected to the lower nipple of the CELLSHEATH float switch.
Revised April 2007 EU
Sysmex XE-5000 Instructions for Use 3-3
CHAPTER 3 Design and Function
6 SIM aspiration nipple (SIM-1)
STROMATOLYSER-IM is aspirated via this nipple. Connected to the lower nipple of the STROMATOLYSER-IM float switch.
7 SLS inlet nipple (SLS)
SULFOLYSER is aspirated via this nipple. Connected to the container of SULFOLYSER.
8 FFD reagent inlet nipple (FFD)
STROMATOLYSER-4DL is aspirated via this nipple. Connected to the container of STROMATOLYSER-4DL.
9 FBA inlet nipple (FBA)
STROMATOLYSER-FB is aspirated via this nipple. Connected to the container of STROMATOLYSER-FB.
10 SNR reagent inlet nipple (SNR)
STROMATOLYSER-NR (L) is aspirated via this nipple. Connected to the container of STROMATOLYSER-NR (L).
11 RED diluent inlet nipple (RED)
RET SEARCH (II) diluent is aspirated via this nipple. Connected to the container of RET SEARCH (II) diluent.
12 SIM air bubbles outlet nipple (SIM-2)
Air bubbles inside the float switch of STROMATOLYSER-IM are discharged via this nipple. Connected to the upper nipple of the STROMATOLYSER-IM float switch.
13 ESE air bubbles outlet nipple (ESE-2)
Air bubbles inside the float switch of CELLSHEATH are discharged via this nipple. Connected to the upper nipple of the CELLSHEATH float switch.
14 Air drier outlet nipple
Outputs pressure after removing dust or moisture by the air drier. Connected to the pressure supply nipple.
15 Air drier inlet nipple
Connected to the pressure outlet nipple of the Pneumatic Unit.
16 Air drier drain nipple
Connected to the drain outlet nipple.
17 Air drier
Removes dust or moisture from the a ir (at PRESSURE sid e) supplied b y th e Pneumatic U nit.
18 EPK inlet nipple (EPK-1)
CELLPACK is aspirated via this nipple. Connected to the lower nipple of the CELLPACK float switch.
19 Drain outlet nipple (D)
Water droplets from the air drier are discharged via this nipple to the waste line.
20 Vacuum supply nipple (V)
Connected to the vacuum outlet nipple of the Pneumatic Unit.
21 Waste level detection nipple (P2) 22 Waste level detection nipple (P1)
When the optional waste monitoring sensor is provided, the waste volume is monitored by detecting differential pressure between the waste level detection nipples P1 and P2. Connected to the detection tube of the waste container.
23 Pressure supply nipple (P)
Connected to the air drier outlet nipple.
24 Waste outlet nipple (W)
Waste fluid is discharged via this nipple. Connected to the sewer or the waste container.
25 EPK air bubbles outlet nipple (EPK-2)
Air bubbles inside the float switch of CELLPACK are discharged via this nipple. Connected to the upper nipple of the CELLPACK float switch.
Revised April 2007 EU
3-4 Sysmex XE-5000 Instructions for Use
CHAPTER 3 Design and Function
Left Side View
1 2 3
4 5
1 0.16 MPa regulator
Adjusts the pressure to 0.16 MPa.
2 0.07 MPa regulator
Adjusts the pressure to 0.07 MPa.
3 0.03 MPa regulator
Adjusts the pressure to 0.03 MPa.
4 Air filter
Prevents dust from entering the bellows unit.
5 Bellows Unit
Adjusts the bellows pressure to -0.04 MPa.
Revised April 2007
Sysmex XE-5000 Instructions for Use 3-5
CHAPTER 3 Design and Function
Right Side View
4
1 Handheld barcode scanner connector
Connected to the optional handheld barcode scanner required to input sample ID numbers.
2 IPU connector
The communication connector with the IPU. Connects to the connector of the IPU using the provided cable.
1 2
3
Caution!
When connecting the communications cable (LAN cable) to the Information Processing Unit (IPU), there is a risk of damage to communications equipment due to static electrical charges on human body . Touch something else to discharge static before connecting the cable.
3 DP connector
Connected to the data printer.
4 Power Supply Switch
Turns the power ON and OFF.
Note:
Do not turn the power ON and OFF repeatedly in a short period of time. This can cause an overload to blow the fuse.
3-6 Sysmex XE-5000 Instructions for Use
Revised April 2007
Loading...
+ 232 hidden pages