All rights reserved. No part of this Operator’s Manual may be Code No. 461-2459-3
reproduced in any form or by any means whatsoever without PRINTED IN JAPAN
prior written permission of SYSMEX CORPORATION.Date of Last Revision: March 2001
SF-3000/06.95
Sysmex is a registered trademark of SYSMEX CORPORATION.
CELLPACK, CELLCLEAN, STROMATOLYSER, SULFOLYSER, SF CHECK are
trademarks of SYSMEX CORPORATION.
Canon is a registered trademark of Canon, Inc.
Cubitainer is a registered trademark of Hedwin Corporation.
DeskJet and LaserJet are registered trademarks of Hewlett-Packard Corporation.
EPSON is a registered trademark of Seiko Epson Corporation.
SMI and SMI Micro/Pettor are registered trademarks of Scientific Manufacturing
Industries, Inc.
Teflon is a registered trademark of E.I. du Pont de Nemours & Co., Inc.
VENOJECT is a registered trademark of Terumo Corporation.
OKI and MICROLINE are registered trademarks of Oki Electric Industry Co., Ltd.
Other trademarks referenced are property of their respective owners.
Patient names and doctor names are entered for information and illustration purposes
only, and do not imply real specific persons.
Revised September 1998
RECEIVING INSTRUCTIONS
The SF-3000 has been thoroughly tested before shipment, and has been packaged
carefully to prevent damage from shipping and handling. Reagents and options have
also been sent and will arrive at approximately the same time as the analyzer. Follow
these guidelines when the system arrives:
•Check to see that the arrows on the sides of the packages are pointing up. If the
arrows do not point up, remark this information on the bill of lading.
•Visually inspect the outside of the package for rips, dents, or possible shipping
damage. Document any sign of damage on the bill of lading, regardless of how
insignificant it may appear. This is for your protection!
•Notify your service representative that the SF-3000 system and its components
have arrived.
•Wait for your service representative to unpack the system and open the packages.
•Follow the unpacking and storage instructions provided on the outside of the
package. Special requirements such as refrigeration are clearly marked on the
outside of the carton and will be included in the unpacking instructions and
package inserts.
WARRANTY INFORMATION
All instruments manufactured by Sysmex® are warranted against defective materials or
workmanship for a period of one year commencing on the installation date at the
customer's required location.
This Warranty does not cover any defect, malfunction, or damage due to:
1.Accident, neglect or willful mistreatment of the product
2.Failure to use, operate, service, or maintain the product in accordance with the
applicable Sysmex Operator's Manual
3.Failure to use the appropriate reagents or chemicals specified for the product
Sysmex SF-3000 Operator's Manual -- Revised September 1995
14. SYSTEM SPECIFICATIONS................................................ 1-21
Revised September 1995
INTRODUCTION
1.INTRODUCTION
The Sysmex® SF-3000 is a technically advanced, computerized, fully automated
hematology analyzer with many features, for in vitro diagnostic use in clinical
laboratories. The SF-3000 provides accurate screening for patients whose hematological
abnormalities indicate the need for further testing. In addition, it can contribute in the
diagnosis and therapeutic monitoring of patients.
The SF-3000 has the capability of analyzing up to 80 samples per hour, and display 4
different patterns -- WBC 5 differential, scattergrams for white blood cells and basophils,
and cell size distribution curves for red blood cells and platelets, as well as 23 analysis
data parameters on the color LCD (liquid crystal display) screen.
Chapter 1 is an introduction to this manual, major system components, system operation
and other general considerations. This chapter should be read before operating the
SF-3000. Chapter 1 contains the following subjects:
Manual Organization
A description of the Operator's Manual and explanation of important informational
headings.
System Overview
An introduction to the main elements of the SF-3000 analyzer, including a description of
the options available to enhance the instrument's operation.
Safety Summary
Covers major safety considerations when using and servicing the SF-3000.
Operational Summary
A summary of the operating procedures for each mode, the LCD screen display,
functions of touch panel, etc.
Installation Requirements
Includes space requirements, necessary equipment, installation environment and other
essential information to be reviewed before installation of the SF-3000.
Instrument Specifications
The instrument's specifications are given in a reference table.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-1
INTRODUCTION
2.OPERATOR'S MANUAL ORGANIZATION
The SF-3000 Operator's Manual has been designed with the operator in mind. Each
chapter has its own table of contents and is designed to facilitate rapid location of
appropriate information.
Table 1-1: Manual Organization
CHAPTER 1
Introduction
CHAPTER 2
Sample Processing
CHAPTER 3
Work List
CHAPTER 4
Result Interpretation/
Reports
CHAPTER 5
Maintenance and Supplies
CHAPTER 6
Quality Control
CHAPTER 7
Calibration
Contains basic information about the instrument; operating procedures,
screen displays, touch panel, installation and safety. This chapter must be
read prior to operating the Sysmex SF-3000.
Explains sampling procedure from start-up to shutdown. All sampling
operation procedures are contained in this chapter.
Explains how to enter order information with the display and management of
the records.
Explains procedures for displaying and processing Stored Data, and
procedures for analyzed result output. Also explains IP messages.
Explains periodic maintenance, and the procedures for replacing
consumables and reagents.
Explains how to run the Quality Control Program. Also explains, for your
reference, the quality control system used in this instrument.
Explains the procedures for auto-calibration and manual calibration.
CHAPTER 8
Troubleshooting
CHAPTER 9
Adjustments
CHAPTER 10
Functional Descriptions
CHAPTER 11
System Setup
APPENDICES A, B, C
Menu Tree, Installation
and Technical Information
Explains error messages and their troubleshooting. Also explains service
data and system test functions.
Explains how to adjust pressure and other adjustments.
Gives the main principles of analysis performed by the instrument. Also
contains system nomenclature.
Explains how to set date, time, units, sample analysis stop and other system
conditions. Also contains data analysis settings -- for abnormal data
judgments, etc.
Contains program menu tree, technical information including unpacking check
lists, installation procedures, bar-code labeling parameters and host
communication protocols.
1-2Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION
3.INFORMATIONAL HEADINGS
3.1 NOTE, CAUTION and WARNING
NOTE, CAUTION and WARNING boxes throughout this manual are meant to get to
attention to important safety and operational information. Always follow the instructions
contained in the boxes. Failure to do so may affect the safety or precision features of the
SF-3000 analyzer. The boxes are defined as follows:
NOTE:Highlights important facts, gives helpful information, and clarifies
procedures.
CAUTION:Provides information on the correct operation of the SF-3000 analyzer.
Information contained in CAUTION boxes is crucial to prevent damage
and to maintain performance of the analyzer.
WARNING! Indicates potentially hazardous situations that could result in serious
injury or infection to laboratory personnel.
3.2 Document Conventions
The SF-3000 Operator's Manual uses the following conventions:
"Quotations": Screen commands, messages, and status information that the SF3000 system displays appear within quotation marks. For example:
"All Samples", "Ready" or "Input Password"
[Square Brackets]: Names of function keys such as control or operational keys,
number keys, or arrow keys that you select on the SF-3000 appear within square
brackets. For example:
[Enter], [Sampler], [2] or [↑]
Italics: Application names and reference material appear in italic form. For
example:
Refer to Chapter 2: Sample Processing.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-3
INTRODUCTION
4.SYSTEM OVERVIEW
The purpose of the SF-3000 is to provide accurate results on 23 parameters and to detect
abnormalities in the sample. To assist the laboratory in the screening of abnormalities,
the SF-3000 analyzes and processes the data obtained from the analyzer unit, and the
LCD displays and outputs either a POSITIVE or a NEGATIVE message for each sample.
A NEGATIVE message signifies that parameter values for the sample are within
acceptable limits and the sample does not require any additional verification of those
values. A POSITIVE message signifies that the sample is not within the acceptable limits
and requires further review and investigation. Proper use of the instrument requires each
laboratory to establish the acceptable limits for results.
The SF-3000 uses the optical detector block incorporating the semiconductor laser to
analyze the WBC population. In the DIFF channel, lymphocytes, monocytes,
eosinophils, and the group of neutrophils and basophils are analyzed to determine the
WBC 4-differentials using flow cytometry. Basophil determination and WBC number
counting are performed in the BASO channel. RBCs and PLTs are analyzed by the RBC
detector block using the DC detection method. HGB determination is performed using
the SLS hemoglobin detection method.
The SF-3000 consists of two major systems.
• Main Unit: Houses the hydraulic and electronic systems and their control
components. The unit analyzes, processes and displays the data
on the LCD screen.
• Pneumatic Unit : Supplies the required pressure and vacuum for analysis.
Main Unit
Sampler Unit
(Optional)
Pneumatic Unit
Figure 1-1: Sysmex SF-3000 Hematology Analyzer
1-4Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION
5.SYSTEM OPTIONS
Several options are available to enhance system efficiency. Available options for the
SF-3000 are:
• ID Bar Code Reader: Reads the ID numbers of bar code labeled samples,
and inputs the numbers automatically.
• Color Graphics Printer: Prints out analysis results (cell size distribution
or Page Printergraphs, scattergrams) and hard copies of screens in
color or black and white.
Printing can be done both in graphic and list format.
• Data Printer: Prints the analysis data in ticket format.
• (Cap Piercing) Sampler Unit : Consists of a mixer and a cap piercer unit, and loads
up to 5 sample racks (50 sample tubes) that enable an
automated closed sampling system.
• Manual CP Unit: Consists of a mixer and a cap piercer unit which
enables a manual closed sampling system.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-5
INTRODUCTION
6.SAFETY
The operator must follow the instructions in this section. It contains important
information to operate the SF-3000 analyzer safely.
6.1 General Safety
While operating, maintaining, servicing or repairing the system, follow all procedures
contained in this manual. The operator must observe all NOTES, CAUTIONS and
WARNINGS posted anywhere on the instrument or given in this manual.
Do NOT touch the electrical circuits behind the enclosure panels of the SF-3000.
Touching the circuits is dangerous and may give you an electrical shock, especially if
your hands are wet.
Use only designated tools and parts for service and repair. Substitute parts should never
be used, and the SF-3000 should not be modified in any way.
Keep hair, clothing and fingers away from all moving parts.
Install the Sampler Unit on a table or work bench to prevent the system from accidentally
falling.
6.2 Hazardous and Biohazardous Material
All surfaces and components in contact or potentially in contact with patient blood
samples should be considered contaminated. Use of protective garments and gloves is
strongly recommended when operating, maintaining, or servicing the SF-3000.
Do not touch the waste when disposing of it or assembling/disassembling the associated
parts. If accidental contact with waste or any other potentially biohazardous material
occurs, flush the affected area immediately with water and follow your laboratory's
prescribed cleaning and decontamination procedures.
Consider all regulations applicable to your laboratory when operating the Sysmex
SF-3000 instrument.
6.3 Reagent Safety
If reagent comes into contact with skin or clothing, flush it with water. If eyes are
affected, flush with copious amounts of water and obtain medical advice if necessary.
1-6Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION, Safety
6.4 Laser Safety
The SF-3000 uses a semiconductor laser unit. Though this laser unit is placed in the
shield box, the operator must observe all NOTES, CAUTIONS and WARNINGS
concerning laser handling.
WARNING!: DO NOT remove the shield box cover of the Laser Unit.
NOTE: This caution label is located
inside the Laser Unit Cover.
< Top View >
Semiconductor Laser
Unit Cover
< Left Side View >
Figure 1-2: Location of Laser Unit and Labels
(Left Interior of the Main Unit)
Refer to Chapter 5, Section 6 for further information on laser handling.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-7
INTRODUCTION
7.OPERATIONAL SUMMARY
The SF-3000 has two basic operating modes: the Manual Open Mode and Capillary
Mode. Additional three operating modes are available by means of optional units: the
Sampler Mode and Manual Closed Mode by means of Cap Piercing Sampler Unit, and
Manual CP Mode by means of Manual CP Unit.
Table 1-2: Operation Modes
Manual Open
Mode
Select
manual mode
Set manual
sample ID
Set sample at manualPlace sample in rackPlace sample
aspiration pipette
Press the Start Switch
Capillary ModeSampler ModeManual Closed
Mode
Operator checks
Turn ON power
Self-check
Auto rinse
Ready
Select
capillary mode
Prepare
capillary
analysis
sample
(1:5 dilution)
Set capillary
sample ID
Analysis performed
Analysis complete
Post-analysis check
Perform shutdown
Press
[Sampler]
Set sample ID
(or read bar
code label)
Set rack No.
Set tube No.
Set rack on samplerin Manual CP
Press [Start]Press the
Power OFF
Select
closed mode
Set closed
sample ID
Start Switch
Manual CP
closed mode
Set closed
sample ID
Unit
Press the
Start Switch
Mode
Select
Operations performed by the operator are indicated in bold print.
Ready:READY LED turns on and "Ready" message is displayed on the LCD screen.
In this state, various operations can be performed, including analysis setting
and data processing.
1-8Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION
8.ANALYSIS PARAMETERS
The SF-3000 uses 4 detection methods and 5 types of reagents to analyze the following
23 parameters:
Table 1-3: Analysis Parameters
ParameterAcronymDetection Method
White Blood Cell CountWBCFlow Cytometry by Semiconductor Laser
Red Blood Cell CountRB CDC
Hemoglobin ConcentrationHGBSLS hemoglobin
Hematocrit (true relative percentage volume
of erythrocytes)
Mean Corpuscular (erythrocyte) VolumeMCVComputed from RBC and HCT
Mean Corpuscular HemoglobinMCHComputed from RBC and HGB
Mean Corpuscular Hemoglobin
Concentration
Platelet CountPLTDC
Neutrophil PercentNEUT%
Lymphocyte PercentLYMPH%
Monocyte PercentMONO%Flow Cytometry by Semiconductor Laser
Eosinophil PercentEO%
Basophil PercentBASO%
Neutrophil CountNEUT#
Lymphocyte CountLYMPH#
Monocyte CountMONO#Flow Cytometry by Semiconductor Laser
Eosinophil CountEO#
Basophil CountBASO#
RBC Distribution WidthRDW-SDComputed
RBC Distribution WidthRDW-CVComputed
Platelet Distribution WidthPDWComputed
Mean Platelet VolumeMPVComputed
Platelet Large Cell RatioP-LCRComputed
H CTRBC Cumulative Pulse Height Detection
MCHCComputed from HCT and HGB
NOTE:
Customers in the United States:
Platelet Distribution Width (PDW) and
Platelet Large Cell Ratio (P-LCR) parameters have not been approved
for reporting by the Food & Drug Administration (FDA). The obtained
values should be excluded from all print forms until reporting approval has
been obtained. See
Chapter 11
for information on excluding these
values.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-9
INTRODUCTION
9.DISPLAY SCREEN
9.1 LCD (Liquid Crystal Display) Screen
The LCD screen is divided into 3 parts: the System Status Area, the Data Processing
Area, and the Menu Area.
1. System Status Area
2. Data Processing Area
3. Menu Area
Figure 1-3: Display Screen
1 .System Status Area
The System Status Area appears as the top two lines of the screen (Figure 1-4).
This area displays Main Unit status information such as analysis status, errors, etc.
It also displays data processing status information such as Peripherals, X M, etc.
The [Sysmex] and [Sampler] function keys also appear in this area.
Sysmex
ReadyManual Mode
DP: 1234567890123XM H C G P D P S U I D
Next No.1234567890123
Sampler
Figure 1-4: System Status Area
• Analysis Status
System status -- "Ready", "Not Ready", etc. is displayed to the left of the top line.
Analysis mode and the next sample ID number --
Analysis mode in which the system is set, and the next sample number to be
analyzed in the selected analysis mode are displayed.
Pressing this keypad displays the Sample No. Setting Screen for the Open
Sampling, Closed Sampling, or Capillary Sampling Manual mode. Using this
setting screen, the analysis mode, next sample ID number for the selected analysis
mode can be set up.
When the Capillary mode is selected, this area is displayed in yellow back-light.
1-10Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION, LCD Screen
NOTE:To set the next Sample ID No., Rack No. and Tube Position No. to be
analyzed in the auto mode by means of Sampler Unit, press [Sampler]
keypad to display the auto mode setting window.
The DP No. is displayed below the system status. If the Data Printer (DP) is the selected
output device, the DP No. displayed indicates the next sample data number to be printed
out.
Errors -- "2.0 kg/cm
2
Error", "Rack Move Error", etc. is displayed below the system
status, in place of the DP No. If 2 or more errors occur simultaneously, they are
displayed in order of importance. When an error is displayed, press [Sysmex] key to
display the action/help messages provided in the Error Log program. Refer to Chapter 8:Troubleshooting for the action message display.
• Data Processing Status
X M : Indicates the status of X M control (one file of the QC program) in the
following way:
No indication:X M control is not in use (Stop).
X M :XM control is in use (Start).
X M (with red back-light):X M Limit Error has occurred.
HC: Indicates transmission status of data to the Host Computer.
No indication: The Host Computer is set to "Not Connected".
HC: Ready for transmission to Host Computer.
HC (with green back-light): Now transmitting to Host Computer.
HC (with red back-light): Transmission to Host Computer not occurring
because of abnormality in Host Computer.
HC (with yellow back-light) : Error ("Not Transmissible", "No Ack" or "Ack
Code Err") has occurred between SF-3000 and
Host Computer.
GP: Indicates the status of the Graphic Printer (or Line Printer).
No indication: The Graphic Printer is set to "Not Connected".
GP: The Graphic Printer is ready.
GP (with green back-light): Data being output to Graphic Printer.
GP (with red back-light): Printing not possible because of abnormality in
Graphic Printer.
DP: Indicates the status of the Data Printer.
No indication: The Data Printer is set to "Not Connected".
DP: The Data Printer is ready.
DP (with green back-light): Data being output to Data Printer.
DP (with red back-light): Printing not possible because of abnormality in
Data Printer.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-11
INTRODUCTION, LCD Screen
SU: Indicates the status of the Sampler Unit.
No indication: The Sampler Unit is not connected.
SU: Sampler Unit ready for sampler analysis.
SU (with red back-light): Sampler not operational because of abnormality
in Sampler Unit.
CP: Indicates the status of the Manual Cap Piercing (CP) Unit.
No indication: The Manual CP Unit is not connected.
CP: Manual CP Unit ready for manual closed mode
analysis.
CP (with red back-light): Manual closed mode not operational because of
abnormality in Manual CP Unit.
ID: Indicates the status of the ID Bar Code Reader.
No indication: The Bar Code Reader is set to "Not
Connected".
ID: The Bar Code Reader is ready.
ID (with red back-light): "ID Unit Com. Error" has occurred between
SF-3000 and Bar Code Reader.
HD (with green backlight) : Appears when the system hard disk is being accessed.
• Function Keys
The [Sysmex] and [Sampler] keypads appear at each end of this area.
[Sysmex] -- Press this key to obtain the hard copy of screen and to access the Error Log.
Refer to Section 9.3 in this chapter for hard copy output. Refer to Chapter 8:Troubleshooting for Error Log.
When an error occurs, this key changes to [Alarm Reset] with red backlight. Press to
stop the alarm.
[Sampler] -- Press to display the sampler start check window, where the Sample ID No.,
Rack No. and Tube Position No. for the next auto mode analysis can be set up. This
keypad changes to [Sampler Stop] during auto mode analysis.
2 .Data Processing Area
Analysis and sample information, including analysis data, scattergrams, cell size
distributions, QC menus, action messages, sub-program menu keypads, etc. are
displayed in the center area of the display screen.
1-12Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION, LCD Screen
3 .Menu Area
The available menu choices are displayed in the bottom line of the display screen. Touch
the desired keypad on the LCD to access and proceed for the program. In the Main
Screen, the Root Menus appear in this area. Refer to Section 9.2 in this chapter for the
root menus.
NOTE:The program keys masked as shown below do not function when
pressed.
Figure 1-5: Example of Masked Program Keys
4 .Power Saving Function
The LCD contrast will automatically decrease for the LCD power saving if the Main Unit
does not detect any key operation on the LCD for 5 minutes. Touch anywhere on the
LCD to exit the power saving mode.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-13
INTRODUCTION
9.2 Root Menus
When the instrument power is ON, the Main Screen with the Root Menus is displayed as
the initial screen. By pressing the [Return] key on any screen, the visible display returns
to the previous screen, finally to the Main Screen.
The SF-3000 provides twelve Root Menus. The Main Screen displays seven Root
Menus on the menu line at a time. When you want to change a menu, press the [More]
key to alternate root menu.
The Root Menus is listed below. Each Root Menu consists a second menu(s) or menu
choices to proceed for the corresponding program.
[Status]
Use this menu to check the analyzer condition, analysis process, sensor status, etc. at
any time. (This menu does not display analysis errors.)
[Work List]
Use this menu to display, register, and process the analysis information, including
Sample No., Rack No., Tube Position, Order Information, etc. Refer to Chapter 3:Work List for more information.
[Stored Data]
This menu provides options to display analysis results up to 1,000 samples stored in the
memory. The Stored Data List includes Date and Time analyzed, Sequential No.,
Sample No., Rack No., Tube Position, output status to the peripherals, flag information
and analysis results. Refer to Chapter 4: Result Interpretation/Reports for more
information.
[QC]
The SF-3000 quality control program provides X or L-J, and X M control
applications. The QC Root Menu performs QC analysis, displays and/or outputs the QC
Chart, establishes the QC settings, etc. Refer to Chapter 6:Quality Control for more
information.
[GP Print]
Use this menu to get options to output the data currently displayed on the screen to the
Graphic Printer.
[DP Print]
Use this menu to get options to output the data currently displayed on the screen to the
Data Printer.
[Replace Reagent]
Use this menu after replacing a reagent container to prime the reagent into internal
reservoir chamber.
1-14Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION, Root Menu
[Mainte.]
Use this menu to perform the customer maintenance programs, such as waste chamber
cleaning, clog removal, etc. and to perform the system test programs.
[Settings]
Use this menu to set the system parameters and conditions, such as date, time, units,
analysis stop condition, etc. Also this program contains data analysis settings to
establish certain standards for the SF-3000 system. System's operational parameters can
be set up to meet individual laboratory preferences to customize the operation of the
SF-3000. Refer to Chapter 11: System Setup for the setup procedure.
[Calib.]
Use this menu to calibrate the SF-3000 for specified parameter(s) automatically or
manually. Refer to Chapter 7: Calibration for the procedure.
[Shutdown]
Use this menu to clean the detector chambers and all hydraulic lines in the SF-3000
system with detergent.
[Auto Rinse]
Use this menu to rinse the hydraulic lines with the diluents, and check the background
values after rinsing. If the background values exceed the preset limit values, a
"Background Error" occurs.
Sysmex SF-3000 Operator's Manual -- Revised September 19951-15
INTRODUCTION
9. 3 Hard Copy
The [Sysmex] keypad always appears on the left of the top line of the screen.
Press this key to display the Sysmex submenu window as shown in Figure 1-6.
Sysmex
Status
Ready
DP:1234567890123
Hard
Copy
Error
Log
Cancel
Work
List
Manual Mode Next No.1234567890123
NO.R :
WBC
NEUT
LYMPH
MONO
EO
BASO
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
Q C
G P
Print
Stored
Data
B A S O
40fL
XM HC GP DP SU
3
[x10 /uL]
6
[x10 /uL]
[g/dL]
[%]
[fL]
[pg]
[g/dL]
[fL]
[%]
3
[x10 /uL]
[fL]
[fL]
[%]
D P
Print
Replace
Reagent
[%]
[%]
[%]
[%]
[%]
Sampler
WBC Flag
RBC Flag
PLT Flag
More
Figure 1-6: Sysmex Submenu Window
[Hard Copy]
Pressing this key outputs the hard copy of screen (screen shot), as displayed on the LCD
at the time of pressing the [Hard Copy] keypad, to the Graphic Printer. This function is
available in any status of instrument operation. The window will close automatically
after pressing this key.
[Error Log]
Pressing this key displays the error log and action/help messages for a specified error.
Refer to Chapter 8: Troubleshooting.
[Cancel]
Pressing this key closes the Sysmex submenu window.
1-16Sysmex SF-3000 Operator's Manual -- Revised September 1995
INTRODUCTION
10. PASSWORD
Some programs in Q.C., Settings and Calibration require a password entry to prevent
unauthorized changes of the important settings and values. When such a program is
selected, the Password Input Screen appears with the message "Input Password". Enter
the registered password and press the [Enter] key. If the entered password is correct, the
program will be executed.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No.1234567890123
XM HC GP DP SU
Input Password
Password
***
78
4
123
0ÐEnter
C
.
Cancel
Sampler
9
65
Figure 1-7: Password Entry Screen
The password is important for the management of the SF-3000, so be sure to memorize it
and/or have aids to help you remember it. Should you forget it, consult your Sysmex
service representative.
The password can consist of up to 12 digits. If you set the password to "0", no
password check will be performed, and the programs will be executed without the need
of password input.
For the procedures for setting and changing the password, refer to Chapter 11, "System
Setup".
Sysmex SF-3000 Operator's Manual -- Revised September 19951-17
Loading...
+ 575 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.