Symmetricom TimeAnalyzer 7500 User Manual

Page 1
TimeAnalyzer 7500
IEEE 1588 Performance Measurement and
Analysis System
User’s Guide
Revision A – February 2010
Part Number 098-00208-000
Page 2
Symmetricom, Inc.
2300 Orchard Parkway
U.S.A.
http://www.symmetricom.com
Copyright © 2010 Symmetricom, Inc.
All rights reserved. Printed in U.S.A.
All product names, service marks, trademarks, and registered trademarks
used in this document are the property of their respective owners.
Page 3
Table of Contents
Chapter 1 Overview
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
TimeAnalyzer 7500 Features & Abilities. . . . . . . . . . . . . . . . . . . . . . . . . . .8
TimeAnalyzer 7100 Features & Abilities. . . . . . . . . . . . . . . . . . . . . . . . . . .9
TimeProvider 5000 PTP Probe Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
TP 5000 Probe Module Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
TimeProvider 500 Client Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Key Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
TP500 Module Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
TimeWatch Probe Module (TA 7500 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
TimeWatch Probe Module Connections (TA 7500 only). . . . . . . . . . . . . .16
Chapter 2 Setting Up
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Site Survey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Tools and Equipment for Set Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Unpacking the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
What You Should Have. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Case Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Opening the Case. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Making Power and Ground Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Connecting DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Connecting the GPS Antenna. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Portable GPS Antenna Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Antenna Placement and Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Connecting LAN and RS-232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
TimeAnalyzer 7x00 Communications Connections . . . . . . . . . . . . . . . . .35
LAN Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
EIA-232 Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Connecting PTP Signals to Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Connecting the E1 Reference to TimeWatch Module. . . . . . . . . . . . . . . . . . . .38
Connecting E1 Signals to Measure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Connecting 1PPS and 10MHz Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Normal Power Up Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 1
Page 4
Table of Contents
Chapter 3 Configuring
Configuring TP5000 Module Using TimeMonitor PDV . . . . . . . . . . . . . . . . . . .46
Overview of TimeMonitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Functions of TimeMonitor PDV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Entering CLI Commands Manually with TimeMonitor PDV . . . . . . . . . . .53
Configuring PDV to Communicate with TP5000 Probe Module . . . . . . . .55
Setting IP Address for MGMT Port of TP5000 Probe Module . . . . . . . . .58
Connecting to the TP5000 Probe MGMT Port with PDV . . . . . . . . . . . . .62
Configuring the TP5000 Probe Module for PDV Measurement . . . . . . . .62
Setting for Free-Run of Internal Rb Clock. . . . . . . . . . . . . . . . . . . . . . . . .64
Managing the TP5000 Module’s User Access List . . . . . . . . . . . . . . . . . .65
Saving and Restoring Configuration Data. . . . . . . . . . . . . . . . . . . . . . . . .68
Probe-Related Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Recovering From Reset to TP5000 Factory Defaults . . . . . . . . . . . . . . . . . . . .72
Enabling the Probe Option on the TP5000. . . . . . . . . . . . . . . . . . . . . . . .73
Enabling Probe Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Setting Up Telnet Firewall. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Provisioning TOD Source for TP5000 Probe Module. . . . . . . . . . . . . . . .74
Provisioning the Reference Output for TimeWatch. . . . . . . . . . . . . . . . . .75
Provisioning the 10MHz & 1PPS Outputs. . . . . . . . . . . . . . . . . . . . . . . . .76
Disabling Battery (-48V) B and Holdover Alarms . . . . . . . . . . . . . . . . . . .77
Provisioning the TP500 PTP Client Module . . . . . . . . . . . . . . . . . . . . . . . . . . .77
TimeWatch Probe Module Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
Connecting to the TimeWatch Probe Module. . . . . . . . . . . . . . . . . . . . . .81
Configuring the TimeWatch Probe Module. . . . . . . . . . . . . . . . . . . . . . . .82
Chapter 4 Operating
PDV Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
Measuring PDV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
E1 Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
E1 Measurements with TP500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
E1 Measurements with External Client. . . . . . . . . . . . . . . . . . . . . . . . . . .98
2 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 5
Table of Contents
Chapter 5 Analyzing
Analysis of PDV Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Analysis with TimeMonitor PDV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Calculations and Masks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Additional Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107
Analysis of E1 Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
Chapter 6 Maintenance and Troubleshooting
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Safety Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
ESD Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Diagnosing the IOC - Reading LED Conditions . . . . . . . . . . . . . . . . . . .116
Diagnosing the IMC - Reading LED Conditions . . . . . . . . . . . . . . . . . . .120
Working With TP5000 Cards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Handling Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Inserting Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Removing Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Repairing the TimeAnalyzer 7x00. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Removing the IOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Replacing the IOC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .124
Replacing the IMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Upgrading the Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
TimeAnalyzer 7x00 Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126
System and Accessory Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . .126
Returning the TimeAnalyzer 7x00. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127
Repacking the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
Equipment Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
User’s Guide Updates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
Obtaining Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
Appendix A Specifications and Factory Defaults
TimeAnalyzer 7x00 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 3
Page 6
Table of Contents
TP5000 Probe Module Specifications & Factory Defaults . . . . . . . . . . . . . . .133
Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Serial Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
LAN Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
Input Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
Output Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135
Clocks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136
TP5000 Probe Module Factory Defaults. . . . . . . . . . . . . . . . . . . . . . . . .136
TP500 Client Module Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141
Serial Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142
Ethernet Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142
T1 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143
1.544 MHz Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143
E1 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144
2.048 MHz Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144
Holdover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144
TimeWatch Probe Module Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .145
4 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 7

Chapter 1 Overview

This chapter provides introductory information for the TimeAnalyzer 7x00.
In This Chapter
OverviewTimeProvider 5000 PTP Probe ModuleTimeProvider 500 Client ModuleTimeWatch Probe Module (TA 7500 only)
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 5
Page 8
Chapter 1 Overview Overview

Overview

Symmetricom's TimeAnalyzer is an all-in-one test and measurement instrument for collecting and analyzing IEEE 1588 packet-timing data, as well as analyzing traditional synchronization signals. This powerful test tool can help you meet the PTP packet sync performance requirements in every stage of your project: engineering, network planning, packet-timing deployment and troubleshooting. With its easy-to-carry portable enclosure, TimeAnalyzer is designed for use in the field, as well as in the lab.
IEEE 1588 packets may traverse a broad range of networks - such as Ethernet, Digital Subscriber Line (xDSL), Gigabit Passive Optical Network (GPON), and Microwave - under a variety of traffic conditions. By measuring Packet Delay Variation (PDV), users can characterize a network and determine its suitability for delivering sync packets. TimeAnalyzer can also measure and analyze traditional synchronization signals, allowing users to validate the synchronization outputs of independent PTP clients. Both types of measurements are essential to ensure reliable synchronization.
Using a hardware time-stamp processor that delivers nanosecond caliber-timing accuracy, and a comprehensive computation engine, TimeAnalyzer collects PTP data and analyzes in real-time and off-line graphical displays. Measurements include Packet Delay Variation (PDV), Minimum Time Deviation (MinTDEV), band TDEV, percentile TDEV, MATIE, MAFE, and Min Time Dispersion. TimeAnalyzer also features a set of masks in the packet domain that correspond to the masks in the synchronization domain, as defined by Telecommunication standards. These masks help determine if the PTP packet flow delivered over the network meets the synchronization criteria for the application, and allow TimeAnalyzer to present a very intuitive and user-friendly PASS/FAIL result for the given conditions.
The six masks available for packetTDEV and packetZTIE analysis cover application requirements ranging from synchronization interface requirements, to traffic interface requirements, to wireless backhaul frequency of fset requirements. Four of the masks apply to the packetZTIE calculation and two of the masks apply to the packetTDEV calculation. The masks are described further in Table 1-1 below:
Table 1-1. Packet-domain Mask Descriptions
Mask Description
TDEVSync.lmt The packetTDEV mask associated with synchronization
interface requirements
TDEVTraffic.lmt The packetTDEV mask associated with traffic interface
requirements
ZTIESync.lmt The packetZTIE mask associated with synchronization
interface requirements
6 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 9
Chapter 1 Overview
Table 1-1. Packet-domain Mask Descriptions
Overview
Mask Description
packetZTIETraf fic.lmt The packetZTIE mask associated with traffic interface
requirements
ZTIEWireless01.lmt The packetZTIE mask associated with wireless backhaul 1
PPB frequency offset requirement
ZTIEWireless15.lmt The packetZTIE mask associated with wireless backhaul 15
PPB frequency offset requirement
TimeAnalyzer can also perform physical-level frequency and phase measurements using frequency and timing signals delivered through T1, E1, 10MHz, and 1PPS. TimeAnalyzer generates various performance data, including MTIE, TDEV, frequency deviation, and fractional frequency offset, which can be compared against Telecordia, ANSI, ETSI, or ITU-T requirement masks.
With an easy-to-use GUI for initiating measurements and analysis, TimeAnalyzer's software enables users to quickly configure a test system and collect PTP timestamp data for PDV analysis. Users can collect data using a wide range of sampling rates (1Hz to 64Hz) to allow for pattern and trending analysis. The timestamp data can also be saved into a file and exported to a network emulator system for traffic simulation and playback. This export function provides a complete integration path to test the PTP network using traffic simulation and verification for R&D development, network planning, and network troubleshooting.
TimeAnalyzer is available in two configurations, the 7500 and the 7100. The TimeAnalyzer 7500 has all the features of the 7100, and also includes the TimeWatch Probe module. Figure 1-1 shows the TimeAnalyzer 7500. Figure 1-2 shows the TimeAnalyzer 7100.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 7
Page 10
Chapter 1 Overview Overview
Figure 1-1. TimeAnalyzer 7500

TimeAnalyzer 7500 Features & Abilities

Measures PTP performance of both the grandmaster and networkIdentifies faults in embedded PTP clientsTests quality of E1 reference output (T1 testing is future abilty)Analyzes packet synchronization quality and network impairements in real-time
and offline
Displays intuitive and user-friendly P ASS/FAIL result for PDV and E1 tests for the
conditions being tested
Provides convenient management connections with internal 5-port Ethernet
switch
Supports field testing with its rugged travel case and internal AC power supply
8 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 11
Figure 1-2. TimeAnalyzer 7100

TimeAnalyzer 7100 Features & Abilities

TimeProvider 5000 PTP Probe Module

Chapter 1 Overview
Measures PTP performance of both the grandmaster and networkIdentifies faults in embedded PTP clientsAnalyzes packet synchronization quality and network impairements in real-time
and offline
Displays intuitive and user-friendly PASS/FAIL result for PDV tests for the
conditions being tested
Provides convenient management connections with internal 5-port Ethernet
switch
Supports field testing with its rugged travel case and internal AC power supply
TimeProvider 5000 PTP Probe Module
The TimeProvider 5000 (TP5000) PTP probe module is a Next Genera tion Network (NGN) packet-based timing and frequency device that is based on the T imeProvider 5000 Grandmaster platform. It combines the functionality of a IEEE 1588 v2 probe with a highly-accurate Rubidium oscillator and E1 and 1PPS/10MHz I/O ports. The TP5000 probe module incorporates hardware-based time stamping to provide the highest level of timing and frequency accuracy for PTP measurements.
The TP5000 probe module features include the following:
Rubidium oscillator
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 9
Page 12
Chapter 1 Overview TimeProvider 5000 PTP Probe Module
Up to 2 Legacy InputsUp to 2 Legacy Outputs 10MHz and 1PPS Outputs2 SFP Connectors on the IOC1 L1 GPS Input on the IMC1 Ethernet Mgmt Port on the IMC1 EIA-232 Serial (Craft) Port on the IMCLED indicators on the IOC and IMCCLI and SNMP Management

TP 5000 Probe Module Connections

Figure 1-3 is a front view of the TP5000 probe module showing connectors and
LEDs.
Figure 1-3. TP5000 PTP Probe Module Connectors and LEDs
PTP Probe Connections
The TP5000 probe module provides two Ethernet 100/1000 Small-form Factor Pluggable (SFP) connections the IOC (see Figure 1-4) module.
Figure 1-4. PTP Output Connections  IOC Module
10 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 13
TimeProvider 5000 PTP Probe Module
Chapter 1 Overview
Ethernet Management Port
The Ethernet Management port is standard 10/100Base-T shielded RJ-45 receptacle. To connect the TP5000 probe module to an Ethernet network, use a shielded twisted pair Ethernet RJ-45 cable.
EIA-232 Serial (Craft) Port
The EIA-232 serial port connection is made through a EIA-23 female connector on the IMC. This port allow you to connect to a terminal or computer using a terminal emulation software package. When connecting to this port, use a shielded EIA-232 direct connect cable.
GPS Connection
The IMC has a GPS input to provide a frequency and time reference (see Figure
1-5).
Figure 1-5. Communication and GPS Connections  IMC Module
Programmable E1 Input Connections
Mini-BNC ports 1 and 2 on the I/O module (see Figure 1-6) are software configurable as traditional telecom input ports with the following signal types:
2.048 Mb/s (G.703/9) 2.048 MHz (G.703/13)
Programmable E1 Output Connections
Two of the four traditional telecom synchronization mini-BNC ports (Port1 and Port2) on the I/O module (Figure 1-6) are software configurable as E1 output ports, with the following signal types:
2.048 Mb/s (G.703/9)2.048 MHz (G.703/13)
10MHz & 1PPS Output Connections
The TP5000 probe module comes equipped with an I/O module featuring 1PPS and 10MHz dedicated output ports. See Figure 1-6.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 11
Page 14
Chapter 1 Overview TimeProvider 500 Client Module
Figure 1-6. Input and Output Connections  I/O Module
UTI Input and Output Connections
The UTI connectors are for future use.
Power and Ground Connections
The TP5000 probe module uses –48 VDC power, which is provided through TimeAnalyzer’s power distribution panel.
Warning: To avoid serious personal injury or death, exercise caution when working near high voltage lines and follow local building electrical codes for grounding the chassis.

TimeProvider 500 Client Module

The TimeProvider 500 (TP500) client module shown in Figure 1-7 provides a client-clock function with the ability to synchronize to a grandmaster clock. To perform the client-clock function, the TimeProvider 500 uses IEEE-1588 v2 as a frequency reference to steer either T1 or E1 frequency outputs. The TP500 client module can be configured and monitored using the Command-Line Interface (CLI) accessible locally with the RS232 connection or remotely via a Telnet connection.
The TP500 Client module “translates” IEEE 1588v2 packet flow into two T1/E1 signaling outputs (G.823/G.824). The TimeProvider 500 PTP translation function enables legacy Time-Division Multiplexing (TDM)-based devices to transparently send and receive critical timing and synchronization signals over a packet-based, Ethernet backhaul network without requiring any modifications to their existing hardware or software. TimeProvider 500 is fully interoperable with Symmetricom’s TimeProvider 5000, TimeHub, and SSU 2000 IEEE 1588v2 grandmaster products, as well as third-party grandmaster products that comply with the IEEE 1588v2 protocol standard.
12 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 15
TimeProvider 500 Client Module
Chapter 1 Overview
Figure 1-7. TP500 Client Module Front Panel

Key Features

The key features supported in this release of the TP500 IEEE 1588v2 Translator module include:
Hardware Features:
OXCO OscillatorSingle 100 Mbps PTP Network PortDual T1/E1 Output PortsDE-9 Craft PortEthernet Management Port
Software Features:
E1/T1 SignalingPTP to E1/T1 ConversionUnicast CommunicationsVirtual Local Area Network (VLAN)PTP Performance MetricsCommand Line Interface (CLI)

TP500 Module Connections

Figure 1-8 shows the connections for the TP500 Client module.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 13
Page 16
Chapter 1 Overview TimeProvider 500 Client Module
Figure 1-8. TP500 Module Connections
Power Connector
A Molex-style connector is used to provide -48 volts DC (labeled Power). When the unit is successfully powered, the PWR LED near the left side of this panel will illuminate green. If the power connection is reversed, the unit will not be damaged, it simply will not power-up. Startup power will be < 17 watts. Steady-state power (nominal 25° C) <9.5 watts. The TP500 Client module uses –48 VDC power and is not equipped with a power switch. Power to the unit is controlled by TimeAnalyzer’s DC power supply.
Warning: To avoid serious personal injury or death, exercise caution when working near high voltage lines and follow local building electrical codes for grounding the TimeAnalyzer
RJ48C Connectors
The stack of 2 RJ48C connections labeled SPAN1 and SPAN2 (can use standard Ethernet cable) provides two separately driven output signals, configurable as either E1 or T1. These are differential outputs (tip and ring) conforming to signal levels defined in G.703.
Note: The RJ48C physical connector is the same as RJ45 (an 8-pin modular connector). RJ45 is used for Ethernet whereas the term RJ48C is used for T1 lines. The pinouts for RJ45 and RJ48C are different. RJ45 (Ethernet) uses pins 1, 2, 3 and 6 whereas RJ48C uses pins 4 and 5.
14 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 17
TimeProvider 500 Client Module
Chapter 1 Overview
T1/E1 Frequency Output Connectors
The TP500 Client module provides configurable outputs at either E1 (2.048MHz) o r T1 (1.544MHz) frequencies. Specific signal types at each frequency are detailed in the CLI command, set output signaltype, as described in. Both outputs must be the same signal type. There are two independent, separately driven outputs, labeled SPAN1 and SPAN2. For differential connection to the TP500 Client moduleeither output, use the appropriate RJ-48C connection shown in Figure 1-8. SPAN1 is the upper connection and SPAN2 is the lower connection. For single-ended connection to either output, use the BNC connectors shown in Figure 1-8.
Note: The outputs from the left BNC and lower RJ48C connection (both labeled SPAN1) are generated from the same driver circuit, so for proper signal level only one of those connec tio ns sh ou ld be use d at any one time. The same situation applies to the right BNC and upper RJ48C, both labeled SPAN2.
LEDs
At the center are four LEDs:
PWR illuminates green when the unit is powered.SYNC provides high-level status indication. The indications and associated
meanings are described later in this document.
OUT illuminates green when output is being generated on the SPAN
connections.
ALARM illuminates Red to indicate a CRITICAL or MAJOR active alarm, AMBER
to indicate MINOR alarm. Current alarm status can be checked using the show alarm command. For more information, see the CLI command, show alarm, in the TP500 User’s Guide.
RS232 MGMT Serial Port
The DE9 connector labeled RS232 MGMT provides a local serial connection to the TP500 module. Through this connection, a Command Line Interface (CLI) provides configuration and status monitoring capability. The RS-232 serial port connection is made through a DE9S female connector. This port allows you to connect to a computer using terminal emulation software for remote monitoring an control. Whe n connecting to this port, use a shielded direct connect cable. See Provisioning the
TP500 PTP Client Module, on page 77, for default settings.
IEEE 1588 Remote Management Port
The RJ-45 connector labeled “1588” is the connection into the network that is providing the grandmaster clock(s). This same connection is also used for remote management.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 15
Page 18
Chapter 1 Overview TimeWatch Probe Module (TA 7500 only)

TimeWatch Probe Module (TA 7500 only)

TimeAnalyzer 7500 includes the TimeWatch probe module, which is designed to perform an MRTIE measurement between two external clock input s. Th e input s are independent and each can be either a 2.048 MHz clock or an E1 2.048 Mbit/s signal. The probe module uses an internal TIE measurement engine tha t compares two external inputs at 80 samples per second (80Hz) and calculates the relative phase. This data is averaged to produce one TIE value for each second, this is then used to calculate the MRTIE between the two inputs.
The calculated MRTIE is a worst case MRTIE from the start of the measurement period which begins at probe initialization and is restarted if the mask is exceeded. The computed MRTIE is compared against the chosen mask, and an exception is generated if the mask is exceeded. The probe module’s firmware handles communications with the attached PC application, whether T imeMonit or Watch or a terminal, formats and then streams the live TIE measurement data directly to the application. The application can then be used to view, store and further analyze the measurement data.
Figure 1-9. TimeWatch Probe Module Front Panel

TimeWatch Probe Module Connections (TA 7500 only)

The probe module’s front-panel connectors are as follows:
Measure In (BNC)
M(R)TIE is computed on the relative phase between this input and the Ref In input . The interface must be configured to be a full height 2.048Mbit/s (G.703 sec. 9), a 30dB down 2.048Mbit/s signal (G.703 sec. 9) or a 2.048MHz signal (G.703 sec. 13).
Ref In (BNC)
This input is used for the input of a reference clock signal. MRTIE is computed on the relative phase between this input and the Measure In interface. The interface must be configured to be a full height 2.048Mbit/s (G.703 sec. 9), a 30dB down
2.048Mbit/s signal (G.703 sec. 9) or a 2.048MHz signal (G.703 sec. 13).
Aux Out (BNC)
Can optionally provide a buffered copy of the Ref In signal; this interface can on ly be used if the Ref In interface is a 2.048MHz signal.
16 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 19
TimeWatch Probe Module (TA 7500 only)
Chapter 1 Overview
GPS Antenna (SMA)
N/A for implementation in TimeAnalyzer.
GPS Antenna (LOS LED)
N/A for implementation in TimeAnalyzer.
GSM Antenna (SMA)
N/A for implementation in TimeAnalyzer.
PSTN Analogue Modem (RJ11)
This connector is used for connecting the probe to a PSTN network for communication with the server. An optional interna l PSTN modem must be installed to enable this option. N/A for implementation in TimeAnalyzer.
LAN (RJ45)
This connector is used for connecting the probe to an Ethernet network to enable communication with the server. It is available on all TimeWatch variants. The probe will automatically detect the data rate of the network; however it may be necessary to configure other network parameters before normal operation. This port will auto-negotiate and run at speeds of either 10 or 100BASE-T.
Alarm Relay Contact
This interface uses a three pin terminal block with Normally Open, Common and Normally Closed pins. It can optionally be connected to an external alarm management system. If an MTIE exception occurs, the relays will be switched and will remain in the active state until a minimum of 15 minutes after MTIE is within the configured mask. The relay will also be in its alarm state if power fails to the probe or if any hardware fault is detected, for example loss of Measure In signal.
Power (-48vDC)
This is the power supply connector and has four pins. Pin 1 is the positive voltage V+, nominally 0V; pin 2 is the negative voltage, nominally -48V; pin 3 is ground; pin 4 is unused. Power is provided from TimeAnalyzer’s power distribution panel.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 17
Page 20
Chapter 1 Overview TimeWatch Probe Module (TA 7500 only)
18 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 21

Chapter 2 Setting Up

This chapter describes the procedures for setting up the TimeAnalyzer 7x00.
In This Chapter
Getting StartedUnpacking the UnitMaking Power and Ground ConnectionsConnecting the GPS AntennaConnecting LAN and RS-232Connecting PTP Signals to MeasureConnecting the E1 Reference to TimeWatch ModuleConnecting E1 Signals to MeasureConnecting 1PPS and 10MHz Outputs
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 19
Page 22
Chapter 2 Setting Up Getting Started

Getting Started

Before you begin to set up the TimeAnalyzer 7x00, review the information in this section.
If you encounter any difficulties during the set-up process, contact Symmetricom Global Services (SGS). Contact Customer Service for information about you r order, RMAs, and other information.

Site Survey

Before you begin setting up, determine the chassis location and ensure the appropriate power source is available (100 - 240 VAC).
Caution: To avoid interference, you must consider the electromagnetic compatibility (EMC) of nearby equipment when you set up the TimeAnalyzer 7x00.
Electromagnetic interference can adversely affect the operation of nearby equipment.

Tools and Equipment for Set Up

You will need the following tools and equipment to set up the TimeAnalyzer 7x00:
Shielded cabling of the appropriate impedance required by the specific signal
type for signal wiring (including GPS, and Ethernet)
Laptop computer with communications software (e.g., Microsoft Hyperterminal,
ProComm Plus or EmTec ZOC) for setting system parameters

Unpacking the Unit

The TimeAnalyzer 7x00 is packaged to protect it from normal shock, vibration and handling damage.
Caution: To avoid electrostatic discharge (ESD) damage to part s that are packaged with the TimeAnalyzer 7x00, observe the following procedures.
Unpack and inspect the unit as follows:
20 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 23
Chapter 2 Setting Up
Unpacking the Unit
1. Wear a properly grounded protective wrist strap or other ESD device.
2. Inspect the container for signs of damage. If the container appears to be damaged, notify both the carrier and your Symmetricom distributor. Retain the shipping container and packing material for the carrier to inspect.
3. Open the container, being careful to cut only the packaging tape.
4. Locate and set aside the CD, printed information, and paperwork that is included in the container.
5. Locate and set aside small parts which may be packed in the container.
6. Remove the accessories from the container.
7. Remove the anti-static packaging from the unit and accessories.
8. Verify that t he model and item number shown on the shipping list agrees with the model and item number on the equipment. The item number can be found on a label affixed to the top of the case, as shown in Figure 2-1. Contact your Symmetricom distributor if the model or item number do not match.
Figure 2-1. Location of Product Label

What You Should Have

TimeAnalyzer is shipped with accessories and support material in addition to the case and components.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 21
Page 24
Chapter 2 Setting Up Unpacking the Unit
TimeAnalyzer 7500
The TimeAnalyzer 7500 is shipped with the following items:
6 SFP Connectors
– Optical 1000 Base-SX Multi-Mode Ethernet 850nm (2 each) – Optical 1000 Base-LX Single-Mode Ethernet 1310nm (2 each) – RJ45 100/1000/1GbE Base-T Full-Duplex (2 each)
Mini GPS Antenna and cableTimeAnalyzer 7500 Support CDTimeProvider 5000 Support CDTimeProvider 500 Support CDTimeWatch Support CDMini-BNC to BNC 90-degree cableBNC-to-BNC cableAC power cableDC power cables (3)
TimeAnalyzer 7100
The TimeAnalyzer 7100 is shipped with the following items:
3 SFP Connectors
– Optical 1000 Base-SX Multi-Mode Ethernet 850nm (2 each) – Optical 1000 Base-LX Single-Mode Ethernet 1310nm (2 each) – RJ45 100/1000/1GbE Base-T Full-Duplex (2 each)
Mini GPS Antenna and cableTimeAnalyzer 7100 Support CDTimeProvider 5000 Support CDTimeProvider 500 Support CDAC power cableDC power cables (2)

Case Dimensions

The case dimensions for the TimeAnalyzer 7x00 are shown in Figure 2-2 and
Figure 2-3.
22 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 25
Figure 2-2. Case Dimensions - Front View
24.63
24.00
10.7010.00
23.75
Figure 2-3. Case Dimensions - Top View
Chapter 2 Setting Up
Unpacking the Unit

Opening the Case

The TA7x00 case has a front cover and a rear cover. The case covers are each secured by two Transportation Safety Administration (TSA) accepted and recognized locking trigger latches, which allow airport security to perform a contents inspection and then simply re-lock the case. To open a latch, first use the included key to unlock it, then pull the trigger to release the latch. See Figure 2-4 below.
Figure 2-4. Case Latch
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 23
Page 26
Chapter 2 Setting Up Making Power and Ground Connections
Both covers must be removed when using the T A7x00 in order to provide access for testing and to allow air flow for cooling. See Figure 2-5 below.
Figure 2-5. Remove Covers From TimeAnalyzer Case

Making Power and Ground Connections

The TimeAnalyzer 7x00 uses AC po we r to gen erate and distribute DC power to the TP500, TP5000, and TimeWatch (TA 7500 only) through the Power Distribution Panel.

AC Power

The power supply for the TimeAnalyzer 7x00 is 100 - 240 VAC, and is rated for -48 VAC, 6.25 A. The power supply features a male AC power connector on the rear panel, as shown in Figure 2-6 below.
Plug the female end of the AC power cord into the AC power connector
24 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 27
Making Power and Ground Connections
Chapter 2 Setting Up
AC Power Switch
The power supply for TimeAnalyzer features an ON/OFF switch, which is located immediately above the power connector on the rear panel, as shown in Figure 2-6 below.
Warning: To avoid accidental power-up of the chassis, ensure that the switch on the power supply is in the OFF position. Or ensure that the power supply is not plugged into the AC power source until installation is complete.
Figure 2-6. Rear Panel - AC Power Connector and Power Switch

Ground

The TimeAnalyzer 7x00 ground is provided via the grounding prong on the AC power cable.

DC Power

The TimeAnalyzer 7x00 provides -48 VDC from the Powe r Distribution Panel to the measurement elements using external cables, as shown in Figure 2-7. The cables will be three different lengths due to the varying distances between the Power Distribution Panel and the power connectors for the TP5000, TP500, and TimeWatch (see Figure 2-8). The cables will also be unique because the TP5000 , TP500, and TimeWatch each have a different style power connector (see
Figure 2-9). See Table 2-1 for details about the power cables.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 25
Page 28
Chapter 2 Setting Up Making Power and Ground Connections
The connectors on the Power Distribution Panel are each labeled for use with a specific module. (Unused connectors have plugs inserted for safety.) To further ensure that the proper connector is used for each module, the pin-outs are dif ferent for each connector . See Figure 2-10 for a view of the connector pins . See Tab le 2-2 for pin-outs for each module. While the TP5000 cable can be plugged into the TimeW atch connector on the Power Distribution Panel, the connector and cable are wired such that no power will be provided to the device.
Figure 2-7. TimeAnalyzer - Power Connections - Front View
Figure 2-8. DC Power Cables
26 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 29
Table 2-1. Power Cable Details
Making Power and Ground Connections
Chapter 2 Setting Up
Module
Cable
Length
(in.)
Connector
Color
TP500 10.0 White Molex
TP5000 12.0 Green Phoenix
TimeWatch 18.0 Orange Terminal block
Figure 2-9. DC Power Connectors - Modules
Figure 2-10. DC Power Connector - Power Distribution Panel
Connector
Type
Table 2-2. Power Connector Pin Outs
Pin #
1 Power n/c n/c 2 Return Power n/c 3 n/c Return Power 4 n/c n/c Return
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 27
TP500
Client Module
TP5000
Probe Module
TimeWatch
Probe Module
Page 30
Chapter 2 Setting Up Making Power and Ground Connections

Connecting DC Power

Use the following procedure to make the power connections between the Power Distribution Panel and the TP500, TP5000, and TimeWatch modules.
Caution: To avoid damage to the TimeAnalyzer, turn off power using the ON/OFF switch on the rear panel before making power connections between the Power Distribuition Panel and the TP500, TP5000 or TimeWatch modules.
1. Ensure the power supply switch on the back of TimeAnalyzer is turned to the OFF position.
2. Locate the power cable for the TP500 module, which is the shortest cable, with white Molex-type connector (see Figure 2-9). Plug the Molex-type connector into the TP500 module power connector (see Figure 2-11).
Figure 2-11. TP500 Module Power Connector
3. Plug the circular black connector of the TP500 cable into the Power Distribution Panel connector that is labeled “TP500”. Tighten the collar to secure the connector.
4. Locate the power cable for the TP5000 module, which is mid-length with a green Phoenix connector (see Figure 2-9). Plug the Phoenix connector into the TP5000 module power connector (see Figure 2-12).
Figure 2-12. TP5000 Module Power Connector
5. Plug the circular black connector of the TP5000 cable into the Power Distribution Panel connector that is labeled “TP5000”. Tighten the collar to secure the connector.
6. Locate the power cable for the TimeWatch module (TimeAnalyzer 7500 model only), which is the longest cable, with an orange terminal-block connector (see
28 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 31
Chapter 2 Setting Up

Connecting the GPS Antenna

Figure 2-9). Plug the orange connector into the TimeWatch module power
connector (see Figure 2-13).
Notes: A terminal block may already be installed in the TimeWatch module power connector. Remove and discard.
Figure 2-13. TimeWatch Module Power Connector
7. Plug the circular black connector of the TimeWatch cable into the Power Distribution Panel connector that is labeled “TimeWatch”. Tighten the collar to secure the connector.
8. Turn the power switch to the ON position.
The LEDs on the TP5000 probe module, TP500 client module, and TimeWatch probe module should illuminate if power was connected correctly.
Connecting the GPS Antenna
Caution: The GPS Antenna should only be connected while the TA7x00 is connected to AC ground.
The antenna connection for the TimeAnalyzer 7x00 is made at the TP5000 probe module’s IMC module connector labeled “GPS”. See Figure 2-14. Allow at least one hour for the unit to track and lock to GPS once the antenna is connected and power is turned on.
Figure 2-14. TP5000 Probe – IMC Card
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 29
Page 32
Chapter 2 Setting Up Connecting the GPS Antenna

Portable GPS Antenna Option

The TimeAnalyzer 7x00 is available with a portable GPS antenna. This is a premium quality magnetic-mount GPS antenna featuring +28 dB LNA, dual bandpass filters, protection from input RF overload, and double-shielded cable. See
Table 2-3 for GPS antenna parameters.
Table 2-3. GPS Antenna Parameters
Parameter Value
General
Design Features
Frequency Polarization LNA Gain Bandwidth Filtering
Output Impedence VSWR Over-all noise figure
2 Stage LNA 2 stage bandpass filter 10 watt input RF protection Low noise, low drop-out linear regulator Double shielded coax cable
Performance
L1 Band (1575.42 MHz) Right hand circular +28dB typical, less cable 10 MHz
-35dB@+/-50 MHz from f
o
50 ohms
1.5, typical
1.8dB, typical
Electrical
Power input Current draw Over-current
+2.75 to+12VDC with reverse polarity protection 11mA typical, 13mA max Thermal over-current shutdown >+150*C
Physical
Dimensions Mount Construction
30 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
44mm x 34mm x 12mm Magnetic Aluminum base, black polycarbonate radome
Page 33
Table 2-3. GPS Antenna Parameters
Parameter Value
Chapter 2 Setting Up
Connecting the GPS Antenna
Coaxial Cable
5 meter length, double-shielded RG-174 equiv., low loss 0.7 dB/m
Environmental
Operating Temperature Storage Temperature
-30 deg C to +85 deg C
-40 deg C to +90 deg C
Other GPS Antenna Options
In the event that the optional portable GPS antenna is not adequate for specific applications, users can order Symmetricom GPS L1 antenna kits or provide their own GPS antennas. See Appendix C in the TimeProvider 5000 User’s Guide for GPS L1 antenna kits that are compatible with the TP5000 Probe.
Required Gain to GPS Engine
It is important to select the proper gain antenna and coaxial cable that accounts for the insertion loss between the antenna and the radio receiver to avoid under-driving or over-driving the radio receiver antenna input. The gain required at 1575 MHz for a GPS radio receiver input is the manufacturer’s specification for the radio receiver to acquire satellites, with some level of signal degradation allowed.
L1 Antenna gains are usually specified as the minimum effective gain. Such antenna gains can have 4.5 to 10 dB more signal strength at different satellite azimuths and elevations, depending on their reception pattern. The GPS engine of the TP5000 probe requires a gain at the antenna connector input of the chassis to be between
25 dB and 35 dB. This tolerance allows for the signal loss of the internal
coaxial cable and connectors. The optimal gain at the radio receiver input is 30 dB. The total gain is the gain of the GPS antenna less the loss of the transient eliminator, connectors, and cables.

Antenna Placement and Mounting

Mount the antenna outside, preferably on the roof with a clear view of the sky , as the cable length allows. Do not mount the antenna near a wall or other obstruction blocking part of the sky. Mount the antenna well above roads or parking lots.
If fixed mounting is not possible or practical for a Symmetricom GPS antenna kit or user supplied antenna, a tripod can be used for temporary antenna mounting.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 31
Page 34
Chapter 2 Setting Up Connecting the GPS Antenna
Locating the Antenna
The 5 meter cable length may limit some placement options in some locations. Use
Figure 2-15 as a guide to locating the antenna.
Figure 2-15. Locating the GPS Antenna
Warning: To avoid serious personal injury or death, exercise caution when working near high voltage lines. In particular:
Use extreme caution when installing the antenna near, under, or
around high voltage lines.
Follow local building electrical codes for grounding the chass is.Symmetricom does not recommend cutting the antenna cable
provided with the GPS Antenna.
Caution: To avoid damage to the GPS antenna, do not place the antenna where high-power radio signals are beamed directly at the unit. Such signals can damage the preamplifier of the GPS antenna.
32 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 35
Chapter 2 Setting Up
Connecting the GPS Antenna
Recommendation: Symmetricom recommends that you consider the following location and environment influences before installing the GPS antenna:
If possible, provide the antenna with an unobstructed 360-degree
view of the sky from the horizon.
In general, do not allow obstructions that obscure the horizon (as
viewed from the antenna) by more than 10 degrees, as shown in
Figure 2-15.
Locate the antenna well away from, an d pr ef er ab ly in a pl a ne
above electrical equipment such as elevators, air conditioners, or other machinery.
To reduce the risk of lightning damage, do not place the antenna at
the highest point of the building.
Locate the GPS antenna at least 12 feet from metallic objects, if
possible.
Locate the antenna high enough to avoid drifted sn ow.Locate the transient eliminator in a protected area to avoid contact
with standing water.
Locate the antenna within 30 feet of the point at which the antenna
cable enters the building.
Allow at least 10 feet of separation distance between GPS
antennas.
Surfaces above the plane of the unit that are between the antenna
and the horizon can produce reflected (multi-path) signals, which can degrade the performance of the radio receiver.
Developing a Grounding Scheme
In addition to determining where to locate and mount the antenna and cabling, you should develop a grounding scheme. The purpose of the grounding scheme is to provide some protection against voltage surges and static discharge. If transient eliminators are used, they also need to be connected to the perimeter ground system or bulkhead entrance panel that is connected to the perimeter ground system.
Caution: To ensure proper grounding, observe these precautions when installing the antenna:
Allow no sharp bends in the ground conduct ors.Ensure that no painted surface insulates the transient eliminator or
grounding clamps.
Ensure that ground conductors are bonded to the metal enclosure
box (if used) and do not enter through an access hole.
Do not use soldered connections for grounding purposes.Secure all grounding connections with mechanical clamp type
connectors.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 33
Page 36
Chapter 2 Setting Up Connecting the GPS Antenna
In general, follow local building codes when selecting a grounding scheme, wire
size, and installation. Use #6 AWG (13.3 mm) copper ground wire or larger, depending on the distance
to the earth ground electrode. Refer to your local electrical codes for specific details. In most cases, 1/0 AWG (55.0 mm) ground wire will maintain 1/10 the resistance of the coaxial shied..
Connect transient eliminators, if part of the grounding scheme, to earth ground
through a conductor. Symmetricom makes no recommendation as to whether to install transient
eliminators. Symmetricom can provide suitable transient eliminators as an option.
Note: Larger ground conductors provide better transient elimination; that is, the larger the ground conductor, the less likely the chance of transients.
Note: Do not connect the outside transient eliminator ground to the inside equipment rack ground. Doing so can defeat the protection afforded by the transient eliminator.
Never connect antenna systems to the same earth ground connector as heating
and cooling systems, elevator or pump motors, or other motors or machinery which can induce noise in the antenna system.
34 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 37

Connecting LAN and RS-232

The modules in the TimeAnalyzer 7x00 each have a LAN connection and RS-232 connection.

TimeAnalyzer 7x00 Communications Connections

The IMC card allows user control of the TP5000 probe module in the TimeAnalyzer 7x00. The EIA-232 serial port (labeled CRAFT) and Ethernet Management port (labeled “MGMT”) are located on its IMC card, as shown in Figure 2-14.
For the TP500 client module, the EIA-232 serial port is located on its front panel, as shown in Figure 2-11. The RJ-45 connector labeled “1588” is used for remote management in addition to being the connection into the network that is providing the grandmaster clock(s).
Chapter 2 Setting Up
Connecting LAN and RS-232
For the TimeW atch probe manual, the Ethernet Management port (labeled “LAN”) is located on its front panel. The EIA-232 serial port is located on the Power Distribution Panel.

LAN Ports

The LAN ports are standard 10/100Base-T shielded RJ-45 receptacles. To connect a TimeAnalyzer 7x00 module to an Ethernet network, use a twisted pair Ethernet RJ-45 cable.
Note: TimeAnalyzer’s internal 5-port switch does not support VLAN. Symmetricom recommends that the internal switch only be used for management connections. See
The connector pinouts for all LAN ports in the TimeAnalyzer are listed in Table 2-4.
Table 2-4. LAN Ethernet Connector Pin Assignments
Signal Name Pin
TX+ (Transmit positive) 1
Figure 4-1 and Figure 4-2.
TX- (Transmit negative) 2 RX- (Receive negative) 3 RX+ (Receive positive) 6 Not Used 4, 5, 7, 8
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 35
Page 38
Chapter 2 Setting Up Connecting LAN and RS-232

EIA-232 Serial Ports

The EIA-232 serial port connection is made through a EIA-232 fe male connector on the any of the D-Sub 9-pin connectors. This ports allow you to connect to a terminal or computer using a terminal emulation software package for remote monit oring and control. When connecting to this port, use a shielded EIA-232 direct connect cable.
Figure 2-16 shows the EIA-232 male connector pin assignments used for all
TimeAnalyzer 7x00 serial ports.
Figure 2-16. Serial Port Male Connector Pins
Table 2-5 describes the EIA-232 connector pin assignments used for all
TimeAnalyzer 7x00 serial ports.
Table 2-5. Serial Port Connector Pin Assignments
Signal Pin
TXD (Received Data) 2 RXD (Transmitted Data) 3 Ground 5
36 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 39

Connecting PTP Signals to Measure

The two Ethernet 100/1000 Small-form Factor Pluggable (SFP) connections on the IOC provide PTP inputs to the TP5000 probe module. See Figure 2-17. It is possible to monitor two different subnets, one on each port.
Figure 2-17. PTP Input Connections
Connecting PTP Signals to Measure
Chapter 2 Setting Up
Only the SFP transceivers in Table 2-6 are recommended and supported. Other SFP transceivers are not supported and may cause the TP5000 probe module to not comply with regulatory requirements.
Table 2-6. Recommended and Supported SFP Transceivers
Type Transceiver
Optical 1000 Base-SX Multi-Mode Ethernet 850nm (Symmetricom 121-20621-11-2)
Optical 1000 Base-LX Single-Mode Ethernet 131 0nm (Symmetricom 121-20621-10-2)
RJ45 100/1000/1GbE Base-T Full-Duplex Avago ABCU-5710RZ
Avago AFBR-5710PZ Finisar FTLF8519P2BNL
Avago AFCT-5701PZ Finisar FTLF1319P1BTL
Use the following procedue to make the connections for PTP measurements:
1. Plug the SFP transceiver into the ETH1 or ETH2 port on the IOC module of the TP5000 probe module.
2. Plug an Ethernet cable into the SFP transceiver.
3. Plug the other end of the Ethernet cable directly into the network, as shown in
Figure 4-1. Although not recommended, TimeAnalyzer’s 5-port switch can also
be used, but a separate Ethernet cable must be used to connect the switch into the network (unless all devices and the computer are configured as part of the same subnet).
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 37
Page 40
Chapter 2 Setting Up Connecting the E1 Reference to TimeWatch Module

Connecting the E1 Reference to TimeWatch Module

The E1 Reference In signal to the TimeWatch probe module is provided by port 1 of the TP5000 probe module. Use the following procedure to connect the ports with the mini-BNC (right-angle) to BNC cable included in the TimeAnalyzer 7x00 ship kit.
1. Connect the mini-BNC (right-angle) connector of the cable to Port 1 on the TP5000 probe module’s I/O card, as shown in Figure 2-18.
2. Connect the straight BNC connector end of the cable to the REF IN port on the TimeWatch module, as shown in Figure 2-18 .
Figure 2-18. TimeWatch REF IN Connection From TP5000 module
38 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 41

Connecting E1 Signals to Measure

Once the REF IN signal is supplied, as described in the section Connecting the E1
Reference to TimeWatch Module, the E1 signal to be measured can be connected
to the TimeWatch module MEASURE IN port. The E1 MEASURE IN signal to the TimeWatch probe module can be provided by
the SPAN 1 port of the TP500 slave module, or any E1 source. Use the following procedure to connect the ports with the BNC-to-BNC cable included in the TimeAnalyzer 7500 ship kit .
1. Connect the staight BNC connector of the cable to the SPAN 1 BNC port on the TP500 module, as shown in Figure 2-19. If an E1 signal from another source is to be measured, use a cable of appropriate length to connect to the E1 output of that source.
Connecting E1 Signals to Measure
Chapter 2 Setting Up
2. Connect BNC connector on the other end of the cable to the MEASURE IN port on the TimeWatch module, as shown in Figure 2-19. If an E1 signal from another source is to be measured, the cable must have a BNC connector (or BNC adapter) to connect to the TimeWatch module MEASURE IN port.
3. Connect an Ethernet cable into the 1588 port on the TP500. See Figure 1-8.
4. Plug the other end of the Ethernet cable directly into the network, as shown in
Figure 4-2. (Symmetricom recommends that PTP signals be connected directly
to the network rather than through TimeAnalyzer’s 5-port switch.)
Figure 2-19. TimeWatch MEASURE IN Connection From TP500 module
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 39
Page 42
Chapter 2 Setting Up Connecting 1PPS and 10MHz Outputs

Connecting 1PPS and 10MHz Outputs

Two of the four traditional telecom synchronization mini-BNC ports (Port3 and Port4) on the TP5000 module’s I/O card (Figure 2-20) are dedicated output ports, with the following signal types:
1PPS10MHz
These signals can be used to measure phase performance of a PTP client through the use of external customer-provided equipment. (This is a future capability of TimeAnalyzer 7x00).
Use the following procedure to connect the 1PPS or 10 MHz output signals of the TP5000 probe to another device. Users must supply their own cables for connections to these output ports, with a mini-BNC connector on the TP5000 module end of the cable.
1. Connect the mini-BNC connector end of the cable to the 1PPS or 10 MHz port on the TP5000 probe module’s I/O card.
2. Connect the other end of the cable to the desired device.
Figure 2-20. 1PPS and 10MHz Output Connections on I/O Card
40 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 43

Normal Power Up Indications

As the TimeAnalyzer 7x00 powers up and begins normal operation, the IOC card and IMC card LEDs on the TP5000 module probe all turn on. Af ter the card self-test is complete and the card firmware is operational, the LED states may change to indicate the appropriate state or status.
Table 2-7 provides a description of the module LEDs.
Chapter 2 Setting Up
Normal Power Up Indications
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 41
Page 44
Chapter 2 Setting Up Normal Power Up Indications
Table 2-7. TP5000 Module LED Descriptions
LED Label Description
IMC Card
Power Module Power Indicator
-48V A-Bus A-Bus Battery Indicator
-48V B-Bus B-Bus Battery Indicator
System Health Aggregation of system alarms
indicator IMC Alarm
IMC module alarm/fault indicator
GPS Status GPS engine communication and
antenna cable status indicator
PWR Green - Module power available
Off - Module power not available
BT A Off - A-Bus power connected
Red - A-Bus power not connected
BT B Off - B-Bus power connected
Red - B-Bus power not connected
Note: This LED will always be Red because the TP5000 Probe does not require redundant power. The -48V B-Bus is not used.
SYS Off - System operating normally
Amber - System operational Minor Alarm(s) Red - System operational Major or Critical Alarm(s)
ALM Off - IMC operating normally
Amber - IMC operational Minor Alarm(s) Red - IMC operational Major/Critical Alarm(s) Flashing Green - Software starting up
GPS Red - Internal GPS hardware fault
Green - GPS enabled and GPS engine/interface is
operating normally Amber - GPS enabled and GPS engine or antenna
interface has operational issues Off - GPS disabled
TCP/IP Management Port LEDs on the Ethernet connector
MGMT Left Amber Flash - Activity
Left Off - No Activity Right Green - Link has been established Right Off - Link has not been established
IOC Card
Power Module Power Indicator
Active Indicates which IOC is responsible
PWR Green - Module power available
Off - Module power not available
ACT Off - Standby IOC or System Startup
Green - Active IOC for generating Telecom, 10MHz, and 1PPS outputs
42 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 45
Table 2-7. TP5000 Module LED Descriptions (Continued)
LED Label Description
Chapter 2 Setting Up
Normal Power Up Indications
IOC Alarm IOC module alarm/fault indicator
Ethernet Port 1 ETH1 port that is used for timing
distribution Ethernet Port 2
ETH2 port that is used for timing distribution
Local Oscillator State Indicates the current mode of operation of the LO servo
ALM Off - IOC operating normally
Amber - IOC operational Minor Alarm(s)
Red - IOC has Failed or has Major/Critical Alarm(s)
Flashing Green - Software starting up
ETH1 Green - Link has been established
Flashing Green - Link activity
Off - Link has not been established
ETH2 Green - Link has been established
Flashing Green - Link activity
Off - Link has not been established
HOLD Off - LO servo is operating in Locked mode
Amber - LO servo is operating in Holdover or
Acquire modes
Flashing Amber - LO servo is operating in
Warm-up mode
Note: This LED will remain Amber when using the
TP5000 probe module in free-run mode. To avoid a
Red LED for GPS and a resulting alarm, set the
Time of Day source to “sys”.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 43
Page 46
Chapter 2 Setting Up Normal Power Up Indications
44 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 47

Chapter 3 Configuring

This chapter describes the procedures for configuring the components of the TimeAnalyzer 7x00, especially the TP5000 probe module. Use the procedures in this chapter after you have installed and powered up the components of the TimeAnalyzer 7500 (see Chapter 2, Setting Up).
In This Chapter
TP5000 Probe Module
Configuring TP5000 Module Using TimeMonitor PDV
Overview of TimeMonitorFunctions of TimeMonitor PDVEntering CLI Commands Manually with TimeMonitor PDVConfiguring PDV to Communicate with TP5000 Probe Module
Setting IP Address for MGMT Port of TP5000 Probe ModuleConnecting to the TP5000 Probe MGMT Port with PDVConfiguring the TP5000 Probe Module for PDV MeasurementSetting for Free-Run of Internal Rb Clock
Setting the System Date Time and Time Offset
Managing the TP5000 Module’s User Access ListProbe-Related CommandsSaving and Restoring Configuration DataRecovering From Reset to TP5000 Factory Defaults
TP500 PTP Client Module
Provisioning the TP500 PTP Client Module
TimeWatch Probe Module
Connecting to the TimeWatch Probe Module
Connect Via RS-232Connect Via EthernetConfigure the Management PC
Configuring the TimeWatch Probe Module
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 45
Page 48
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV

Configuring TP5000 Module Using TimeMonitor PDV

Overview of TimeMonitor

The TimeAnalyzer PDV Software has three major functions:
Configure the PTP probe in TimeAnalyzer 7x00 for PTP packet-timing
measurement data collection.
Collect packet timing data. The file of collected data contains packet timestamps
between the PTP probe in TimeAnalyzer 7x00 and a PTP grandmaster.
Analyze a probe data file using packetTDEV and packetZTIE, and give a
pass/fail indication using built-in masks.
When TimeMonitor PDV is launched, the main screen will appear.
Anatomy of the Main Screen
The main screen for TimeMonitor PDV is shown in Figure 3-1, with numeric mark ers at the various buttons, fields and boxes. Descriptions for these buttons fields and boxes, corresponding to the numerical markers, are presented in
Table 3-1.
Figure 3-1. Main Screen of TimeMonitor PDV
46 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 49
Configuring TP5000 Module Using TimeMonitor PDV
Table 3-1. Descriptions of Main Screen
Item Label Description
Command Sets TimeMonitor PDV to Command mode. Interaction with
TimeAnalyzer's TP5000 probe module is through the buttons and fields of the GUI. Commands are issued either with buttons or the "Command" text box (see item 26). In Command mode, for example, the Login button can be used to login to TimeAnalyzer's TP5000 probe module or the Probe Address button can be used with the Probe Address/Netmask/Gateway fields (see item 6) to change the address for the selected Probe port.
Terminal Sets TimeMonitor PDV to Terminal mode. The interface will act as
a simplified terminal. Commands and responses are shown in the Main Display (see Item 29). All other fields and buttons will be grayed out, except for Check, Start, and Clear buttons.
Disconnect Sets TimeMon itor PDV to Disco nnect mode. Use as a quick means
to Logout from TimeAnalyzer's TP5000 probe module. Must be in Disconnect mode to change between RS-232 and TCP/IP for communicating with probe, or to change communication settings (see Item 3).
Chapter 3 Configuring
Check Click Check button to verify the test setup before beginning the
PDV measurements for the selected port. Start Click Start button to start PDV measurements for the selected port. Stop Click Stop button to stop PDV measurements for the selected port. RS-232
TCP / IP
Use these radio buttons to select the type of connection for
command and control of the TP5000 probe module. Clicking the
button to the right of the connection type will display the dialog box
to configure the connection. TCP/IP should be used for making
measurements. Either TCP/IP or RS-232 can be used for
configuration. Interval Use these radio buttons to select the desired interval rate for Sync
and Delay Requests from the TP5000 probe module.
Duration Use this field to enter the desired lease duration for the TP5000
probe module
Probe's IP Address Netmask Gateway
Enter the IP address information for the selected TP5000 probe
module port. The Command mode or Disconnect mode (see Item
1) must be selected in order to enter this data. Click the Probe
Address button (see Item 14) to send this address information to
the TP5000 probe module.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 47
Page 50
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
Table 3-1. Descriptions of Main Screen (Continued)
Item Label Description
Grandmaster's IP Address
Enter the Grandmaster connection information for the selected
port. The Command mode (see Item 1) or Disconnect must be
selected in order to enter this data.
(Note: Netmask and Gateway information are not n ecessary for the
Grandmaster.) VLAN Click the check box if the Grandmaster has a VLAN address. Enter
the appropriate VLAN ID in the field.
Clock ID -
Enter the Clock ID of the Grandmaster for the selected port. Master
Clock ID ­Probe
Enter the Clock ID for the selected TP5000 probe module port.
(Note: It is not necessary to fill in the Clock ID field for the TP5000
probe module, but it is recommended. ) Firewall Off Use this button to set the Telnet firewall for the Probe to "Allow".
The default firewall state for Telnet is "Block". The Command mode
(see Item 1) must be selected in order to use this button, and the
user must be logged in to the TP5000 probe module. Eth1
Eth 2
Use these radio buttons to select the TP5000 probe module
Ethernet port for measurements or configuration.
Probe Enable Probe Disable
Use these buttons to Enable or Disable the Probe Mo de for the
selected port. The Command mode (see Item 1) must be selected
in order to use these buttons, and user must be logged in to the
TP5000 probe module. Probe Address This button is used to set the IP address for the selected TP5000
probe module Ethernet port using the info in the Address fields for
the TP5000 probe module (see Item 6). Changes to the TP5000
probe module IP address will not be sent to the TP5000 probe
module until this button is clicked. The Command mode (see
Item1) must be selected in order to use this button, and user must
be logged in to the TP5000 probe module. Probe Restart This button will restart the selected port of the TP5000 probe
module. The Command mode (see Item 13) must be selected in
order to use this button, and user must be logged in to the TP5000
probe module. Show Config Use this button to display the Probe configuration for both ports.
The Command mode (see Item 1) must be se lected in or der to use
this button, and user must be logged in to the TP5000 probe
module.
48 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 51
Configuring TP5000 Module Using TimeMonitor PDV
Table 3-1. Descriptions of Main Screen (Continued)
Item Label Description
Chapter 3 Configuring
Login Logout
Use these buttons to login or logout from the TP5000 probe
module. The Command mode (see Item 1) must be selected in
order to use this button. File The File textbox is used to specify file name and destination
directory. The name need only be specified once because the
software will reuse a file name for new files and add on a date and
time to the file name if the file already exists. Title Enter title for packet delay data file.
Help The Help drop-down box is used to access the Help file (this file),
and get the revision level of the Symmetricom TimeMonitor PDV
software.
All the information about the Help Menu is in this section and can
be read by clicking on one of the subjects below to jump to a topic
directly. Information about other menus can be read by selecting
the menu item in the left pane and jumping to the selected topic
directly. You can also click on the (<=Back) Help button to move to
the most recently viewed topic.
HELP: Opens the HELP file. This file is not context sensitive.
Opening the Help file will always initially display the Help with the
first topic, "Introduction", in view .
README: This file contains important information about the
Symmetricom TimeMonitor Measurement Software. The topics
covered are either not found in other documentation or are judged
to be important enough to warrant restatement.
ABOUT: Displays the Version and Revision number and other
information about the Symmetricom TimeMonitor PDV Software.
SETUP: Selecting this menu item brings up a dialog box which can
be used to adjust the maximum number of samples for analysis.
The default is 1,000,000 samples. If the number of points in the file
exceeds the maximum, the set of data is decimated to a value
equal or lower than the maximum points setting. Hint Provides command hints when manually entering CLI commands.
TDEV ZTIE
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 49
Use these radio buttons to select the type of analysis to be
performed, either packetTDEV or packetZTIE.
Page 52
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
Table 3-1. Descriptions of Main Screen (Continued)
Item Label Description
Analyze Use this button to initiate analysis of the current file.
Clear Click this button to clear the CLI command info from the Main
Display field (see Item 29).
Send Click this button to send the CLI command from the Command field
(see Item 26) to the Probe.
Command This field is used to manually enter CLI commands for the Probe.
TimeMonitor PDV must be in Command mode (see Item 1) to use
this feature. Click the Send button (see Item 25) to send the
command to the Probe. Title Bar Shows the product name. The cursor has a pointer shape in this
area.
Message If live data is being collected, then the Elapsed T ime is shown.
Main Display Probe measurement dat a point s ar e displayed in th is section wh en
the Check button (see Item 2) is pressed, as well as any responses
or messages from the TP5000 probe module during a CLI session.
50 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 53
Configuring TP5000 Module Using TimeMonitor PDV

Functions of TimeMonitor PDV

TimeMonitor PDV has four basic functions:
1. Connect
2. Configure
3. Measure
4. Analyze These functions are shown in Figure 3-2 and described in Table 3-2:
Figure 3-2. TimeMonitor PDV Functionality Groupings
Chapter 3 Configuring
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 51
Page 54
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
Table 3-2. Descriptions of TimeMonitor PDV Functionality
Function Description
Connect to TimeAnalyzer’s TP5000 probe module.
1
a. Select RS-232 or TCP/IP (Disconnect mode).
First time connection to a network requires connection via RS-232 to
change IP address of MGMT port for TP5000 probe module. b. Login to TimeAnalyzer’s TP5000 probe module (Command mode). c. Turn off Telnet firewall of TimeAnalyzer’s TP5000 probe module
(Command mode, logged in).
Configure TimeMonitor and TP5000 probe module address (Command mode).
2
a. Select probe module port.
3
4
b. Enter Grandmaster IP address and Probe IP address. c. Enter Clock ID for Grandmaster. d. Enable probe module for selected port, change probe IP address per
entry in 2b, and restart probe module(Logged in).
Measure packet delay between Grandmaster and TimeAnalyzer’s TP5000 probe module.
a. Enter file name and title for PDV measurement data.
b. Select sync rate for measurement and lease duration. c. Start and Stop measurement (Command mode, logged in)
Analyze data file (Command or Disconnect mode)
a. Set the max number of samples for measurement plot.
b. Select type of measurement plot and start analysis.
52 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 55
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring

Entering CLI Commands Manually with TimeMonitor PDV

The TP5000 probe module can also be configured and operated with Command Line Interface (CLI) commands via a telnet session over TCP/IP. In addition to the GUI, the TimeMonitor PDV software provides two ways to manually enter CLI commands to the TP5000 PTP probe modulein the TimeAnalyzer 7x00 . In Command mode, commands can be entered in the Command field and sent to the TP5000 probe module by clicking the Send button.
In Terminal mode, the main display area of the TM PDV interface will function as a simplified terminal.
Complete commands, with proper syntax, must be used with either approach. The Tab completion feature for TP5000 CLI commands is not available through the TimeMonitor PDV interface in either Terminal or Command mode.
See the TimeProvider 5000 User’s Guide for details about CLI commands and associated syntax.
Command Mode
Use the following procedure to enter CLI commands manually with TimeMonitor PDV in Command mode:
1. Select the radio button for the Command mode.
2. Click the Login button to login to the TP5000 probe module in the TimeAnalyzer 7x00.
3. Enter the Username and Password for the TP5000 probe module when prompted in the main display area of the screen.
4. In the Command field, enter the desired CLI command, as the following example image illustrates.
5. Click the Send button to send the command to the TP5000 probe module for execution. Any response will be shown in the main display area of the screen.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 53
Page 56
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
Terminal Mode
To display the IP address for the MGMT port on the TP5000 probe module’s IMC card, enter the command “show ip status imc”. Use the following pro cedure to ente r CLI commands manually with TimeMonitor PDV in Terminal mode:
1. Select the radio button for the Command mode.
The “Login:” prompt will appear in the main display. All other fields and buttons are greyed out, except for the Check, Start, and Clear buttons.
2. Enter the Username for the TP5000 probe module at the prompt in the main display. Click the Enter button on user’s computer.
3. Enter the Password for the TP5000 probe module at the prompt in the main display. Click the Enter button on the user’s computer. The TP5000 prompt will appear.
tp5000>
4. Enter the command at the TP5000 prompt and click Enter. Example:
54 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 57
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring

Configuring PDV to Communicate with TP5000 Probe Module

The TP5000 PTP probe module in the TimeAnalyzer 7x00 provides two communication interfaces: RS-232 and TCP/IP. TCP/IP should be used for making measurements due to speed limitations on the RS-232 port. Either TCP/IP or RS-232 can be used for configuration.
Configuring RS-232 Parameters
The serial interface for the TP5000 probe module is fixed at the following settings. There is no handshaking.
Baud rate- 57600Stop bit - 1Parity - 0Data length- 8Flow control- none
To configure TimeMonitor PDV and the computer for the TP5000 module serial interface parameters:
1. Make sure that the radio button for the Disconnect mode is selected.
2. Select the RS-232 radio button.
3. Click the button on the right side of the RS-232 radio button to display the RS-232 Settings dialog.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 55
Page 58
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
4. Use the radio buttons to select a Baud Rate of 57600.
5. Select a Parity of None.
6. Select a Data Bits value of 8.
7. Select a Stop Bits value of 1.
8. Select the Com Port that is to be used to communicate between the computer and the TP5000 probe module.
9. Select a Flow Control of None.
10.Click OK to return to the main screen.
11.Click the Login button to login to the TP5000 probe module in the T imeAnalyzer 7x00.
12.Click the Firewall Off button to allow Telnet sessions with the TP5000 probe module via TCP/IP.
13.Click the Logout button.
Configuring TCP/IP Parameters
The default IP address for the MGMT port on the IMC card of the TP5000 probe module is shown in Table 3-3 below, as well as the default login information:
Table 3-3. Default IP Address and Login Info for MGMT Port on TP5000 Module
Parameter Default Values
IP Address 192.168.1.12 Netmask 255.255.255.0 Gateway 0.0.0.0 Username admin Password symmadmin**
56 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 59
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring
See Setting IP Address for MGMT Port of TP5000 Probe Module for a procedure on how to set the IP address for the MGMT port (IMC card).
Use the following procedure to enter the IP address for the MGMT port (IMC card) on the TP5000 probe module into TimeMonitor PDV:
1. Make sure that the radio button for the Disconnect mode is selected.
2. Select the TCP/IP radio button.
3. Click the button on the right side of the TCP/IP radio button to display the TCP/IP Settings dialog.
Selecting this menu item brings up the TCP/IP Settings dialog box.
4. Enter the IP address of the MGMT port (IMC card) on the TP5000 probe module.
5. Enter a value of 0023 for the port.
6. Click the OK button to return to the main screen.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 57
Page 60
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV

Setting IP Address for MGMT Port of TP5000 Probe Module

The TimeProvider 5000 module supports static IP addresses (default) as well as dynamically allocated addresses on the MGMT (Management) port. The dynamically allocated address requires a connection to a DHCP server. When provisioning the unit for static IP address, you can set the Host address, Mask address, and Gateway address, or you can set the addresses to the factory default.
To set up and manage the TP5000 probe module with a terminal or a computer with terminal emulation, you must establish either a serial connection or an Ethernet connection.
Note: Before you can communicate with the TP50 00 prob e mo d ule through an Ethernet connection, you must first configure the Ethernet port using the serial connection.
Communicating Through the Serial Port
An EIA-232 serial port is available on the IMC module for a direct serial connection to a terminal or a computer with terminal emulation. Use the following procedure to connect the TP5000 probe module to a terminal or a computer with terminal emulation via a straight through serial cable:
Note: For information on restricting user access, see Managing the
TP5000 Module’s User Access List, on page 65.
1. Connect one end of a straight-through serial cable to the serial port on the computer or terminal and the other end to the EIA-232 connector on the IMC.
2. Configure the emulation software for 8 dat a bit s, 1 stop bit, no p arity, 57600 baud rate, and no flow control. For details on how to do this with T imeMonitor PDV, see
Configuring RS-232 Parameters, on page 55.
3. Start the terminal emulation software and press Enter. The system prompt should appear. If it does not, recheck each step in this procedure.
4. Type your user name and press Enter. The system prompts for a password.
5. Type your password and press Enter. The system prompt appears.
Note: The unit’s default user name is “admin” and the default password is “
symmadmin**”.
58 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 61
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring
Configuring the Ethernet Port Static IP Mode
The Static IP mode on the MGMT port of the TP5000 module is enabled by default. This allows the user to manually configure a static IP address for the MGMT port. Use the following procedure to configure the TP5000 probe module to enable the Static IP mode on the MGMT port. This procedure also demonstrates how to configure the TP5000 probe module’s Ethernet parameters (IP Address, Subnet Mask, and Gateway Address). Use the dot-decimal notation format xxx.xxx.xxx.xxx to enter the address parameter. See Figure 3-3 for details about the hierarchy for the set IP command set IP command or other CLI commands
Notes: The TP5000 probe module’s default IP address for MGMT port is:
. See the TP5000 User’s Guide for additional details about the
.
IP Address 192.168.1.12 Mask 255.255.255.0 Gateway 0.0.0.0
1. Login at the Admin or Config-user level (see Logging In, on page 65).
2. Type
3. Type
4. Type
set ip-mode imc static and press Enter.
set ip imc address 192.168.0.10 netmask 255.255.255.0 gateway
192.168.0.1
set ip imc restart.
and press Enter.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 59
Page 62
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
Figure 3-3. Set IP Command Hierarchy - TP5000 Probe Module
Configuring the Ethernet Port DHCP Mode
Use the following procedure to configure the TP5000 probe module to enable DHCP on the MGMT port.
1. Login at the Admin or Config-user level (see Logging In, on page 65).
2. Type
3. Type
ip-mode imc dhcp and press Enter. set ip imc restart and press Enter.
Connecting Through an Ethernet LAN
1. Ensure that the TP5000 probe module is connected to a LAN.
2. Telnet from your computer to the IP address assigned to the TP5000 probe module.
Note: The telnet firewall is set to Block by default. Use th e fo llowin g command to set the IMC firewall to Allow for telnet communications:
set firewall imc telnet allow
3. Type your user name and press Enter. If you are not assigned as a user in the system, contact the system administrator or see Adding a User, on page 66.
4. Type your assigned password and press Enter. The system prompt appears.
60 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 63
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring
TimeMonitor PDV can be used to set the IP address for the MGMT port on IMC card in the TP5000 probe module, but CLI commands for the TP5000 module must be entered manually.
St atus info for t he IMC card shows the current IP addre ss being used for the MGMT port. Use the following command to display the status for the IMC card in the TP5000 probe module.
show ip status imc
Use the following command to change the IP address for the MGMT port:
set ip imc address <IP-address> netmask <mask_value>
gateway <gateway_value>
In order for the configuration change to take effect, the interface must be restarted with the following command:
set ip imc restart
Procedure Setting IP Address for MGMT Port
The following procedure shows how to set the IP address for the MGMT port on the TP5000 probe module to a sample value of 192.168.106.16, with a gateway of
192.168.106.255, and a mask of 255.255.255.0. (If no gateway, use value of
0.0.0.0).
1. If using the TimeMonitor PDV software, select the radio button for the Command mode.
The “Login:” prompt will appear in the main display If entering commands directly to the TP5000 probe module via telnet with a
terminal emulator using the serial port, the “Login:” prompt will appear after the terminal emulator opens.
2. Enter the Username for the TP5000 probe module at the prompt in the main display. Click the Enter button on user’s computer.
3. Enter the Password for the TP5000 probe module at the prompt in the main display. Click the Enter button on the user’s computer. The TP5000 probe module prompt will appear.
tp5000>
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 61
Page 64
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
4. Enter the command set ip imc address 192.168.106.16 netmask
192.168.106.255 gateway 255.255.255.0 and press Enter.
5. Enter the command set ip imc restart and press Enter.
6. Follow the steps in
Configuring TCP/IP Parameters, on page 56 to change the IP
address that TimeMonitor PDV uses for conne cting to the TP5000 probe module.

Connecting to the TP5000 Probe MGMT Port with PDV

Use the following procedure to connect to the MGMT port (IMC card) on the TP5000 probe module using TimeMonitor PDV. The IP address for the MGMT port of the probe module must have already been configured with the proced ure in the section
Configuring TCP/IP Parameters in order to successfully connect to the TP5000
module with this procedure.
1. Make sure that the radio button for the Disconnect mode is selected.
2. Select the radio button for the desired type of connection, RS-232 or TCP/IP.
3. Select the radio button for the Command mode.
4. Click the Login button to login to the TP5000 probe module in the TimeAnalyzer.
5. Enter the Username and Password for the TP5000 probe module when prompted in the main display area of the screen.

Configuring the TP5000 Probe Module for PDV Measurement

Use the following procedure to configure TimeMonitor PDV and the TP5000 probe module for PDV measurements:
62 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 65
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring
1. Select the radio button for the Command mode.
2. Select the radio button for the desired Probe port.
3. Enter the Grandmaster’s IP Address, Gateway, and Netmask. If there is no Gateway address for the Grandmaster, use “0.0.0.0” for the Gateway field.
4. If the Grandmaster is configured on a VLAN, check the VLAN box. A box for the ID will appear. Enter the VLAN ID in this box.
Note: TimeAnalyzer’s internal 5-port switch does not support VLAN. Symmetricom recommends that internal switch only be used for management connections. See
Figure 4-1 and Figure 4-2.
5. Enter the TP5000 probe module’s IP Address, Gateway, and Netmask. If there is no Gateway address for the TP5000 probe module, use “0.0.0.0”.
6. Enter the Clock ID for the Grandmaster.
7. Enter the Clock ID for the TP5000 probe module.
8. Select the radio button for the Command mode.
9. Click the Login button to login to the TP5000 probe module in the TimeAnalyzer.
10.Enter the Username and Password for the TP5000 probe module when prompted in the main display area of the screen.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 63
Page 66
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
11.Click the Probe Address button to set the new IP address for the selected Ethernet port on the TP5000 probe module. Skip this step if the IP address for the port was previously configured to the desired value.
12.Click the Logout button.

Setting for Free-Run of Internal Rb Clock

The default Time of Day (TOD) source is GPS, but the TP5000 module can be configured to use the free-running internal Rb clock as reference. The TOD souce must be set to System, with no input signals on the programmable input ports.
1. Login at the Admin or Config-user level (see Logging In, on page 48).
2. Type
3. Type
4. Type
5. Type
6. Type
set tod-source sys and press Enter. show tod-source and press Enter to verify the TOD source. set input state port1 disable and press Enter. set input state port2 disable and press Enter. sync tod-source now and press Enter.
Setting the System Date Time and Time Offset
The TP5000 probe module Time of Day can be set either by an external source, or manually with the “set clock” command. You can set the time and date only in the system tod-source mode, but if the TOD source has been set to GPS (default), it is not possible to set the time manually.
Example
1. Login at the Admin or Config-user level (see Logging In, on page 65).
2. Type
show tod-source and press Enter to check the TOD source.
3. Type
4. Type
set tod-source sys and press Enter. set clock date-time <value> and press Enter. Enter the value in
format yyyy-mm-dd,hh:mm:ss.
5. Type
64 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
sync tod-source now and press Enter.
Page 67
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring

Managing the TP5000 Module’s User Access List

When you are logged in at the Admin level you can add, edit, or delete user names in the user access list. The user list can contain up to 20 names and users can be set to one of three different security levels described in Table 3-4. Users are required to enter a user name and password to log in to the system. Users can access or modify only the parameter settings allowed by their assigned security level.
Caution: To avoid a possible service call, do not lose or misplace the user name and password of the Admin-level user.
Table 3-4. User Level and Access
Security Level Description
User Users can edit their password and show information, or access any
command at the User security level.
Config Users are allowed to change configuration and issue User-level
commands.
Admin Users are allowed to modify the security database or download new
firmware and issue commands at any level.
Use the procedures in this section to manage user access to the TP5000 probe module.
Logging In
Use the following procedure to log in to the system at the admin level.
1. Ensure that the TP5000 probe module is connected to a LAN, or directly connected to a computer through the EIA-232 serial port. See TimeAnalyzer
7x00 Communications Connections, on page 35.
2. If the TP5000 probe module is connected to a LAN, Telnet or SSH from your computer to the assigned IP address.
Note: The telnet firewall is set to Block by default. Use th e fo llowin g command to set the IMC firewall to Allow for telnet communications:
set firewall imc telnet allow
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 65
Page 68
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
If the TP5000 probe module is connected to a computer directly through the serial port, start a Terminal session and press Enter.
3. Type your user name and press Enter. The system prompts for a password.
4. Type your password and press Enter. The system prompt appears.
Note: The unit’s default user name is “admin” and the default password is “ should change the default user name and password.
symmadmin**”. To avoid unauthorized access, you
Adding a User
Use the following procedure to add a user to the system access list.
Note:
User passwords can consist of alphanumeric characters, “~”, “`”, “(“, “)”, “!”, “?”, “-”, “_”, and “.” with a minimum of 8 characters and a maximum of 32 characters.
Usernames can consist of numbers, upper and lower case letters, dash "-", and underscore "_", with a maximum of 20 characters. Usernames cannot use a leading dash “-” as the first character.
For a description of user access levels, see Managing the TP5000
Module’s User Access List, on page 65.
1. Login at the Admin level (see Logging In, on page 65).
2. Type
3. Type
show user and press Enter to view the current list of users. set user add and press TAB twice. The system prompts you for a user
name.
4. Type a user name and press SPACE and then TAB. The name you enter will be the log-in name for the user . The system prompt s you for the access-level for the user.
5. Type in the desire access-level (user, config, or admin) and press Enter. The system prompts you for a password.
6. Type a password and press Enter. The password you enter will be the log-in password for the user. The system responds with the confirm password prompt.
7. Retype the password and press Enter. The system indicates that “
been added: <user>
8. Type
66 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
show user and press Enter to verify that the account exists.
”.
CLI user has
Page 69
Configuring TP5000 Module Using TimeMonitor PDV
Chapter 3 Configuring
9. Repeat steps 3 through 8 of this procedure to add other users.
Deleting A User
Use the following procedure to delete user ABCD1234 from the system access list. Do not delete the default username and password.
1. Login at the Admin level (see Logging In, on page 65).
2. Type
3. Type
show user and press Enter to view the current list of users. set user delete ABCD1234 and press Enter.
Displaying Current Users and User Access Levels
Use the following procedure to show a list of all assigned users and their access levels.
1. Login at the Admin level (see Logging In, on page 65).
2. Type
show user and press Enter. The system displays the current list of users.
Changing a User’s Password and Access Level
To change a user’s password and access level you must delete and then re-create the user. Use the following procedure to change the password and access level.
Note:
User passwords can consist of alphanumeric characters, “~”, “`”, “(“, “)”, “!”, “?”, “-”, “_”, and “.” with a minimum of 8 characters and a maximum of 32 characters.
Usernames can consist of numbers, upper and lower case letters, dash "-", and underscore "_", with a maximum of 20 characters. Usernames cannot use a leading dash “-” as the first character.
For a description of user access levels, see Managing the TP5000
Module’s User Access List, on page 65.
1. Login at the Admin level (see Logging In, on page 65).
2. Type
3. Type
4. Type
show user and press Enter to view the current list of users. set user delete <user name> and press Enter. set user add and press TAB twice. The system prompts you for a user
name.
5. Type a user name and press SPACE and then TAB. The name you enter will be the log-in name for the user . The system prompt s you for the access-level for the user.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 67
Page 70
Chapter 3 Configuring Configuring TP5000 Module Using TimeMonitor PDV
6. Type in the desire access-level (user, config, or admin) and press Enter. The system prompts you for a password.
7. Type a password and press Enter. The password you enter will be the log-in password for the user. The system responds with the confirm password prompt.
8. Retype the password and press Enter. The system indicates that “
been added: <user>
9. Type
show user and press Enter to verify that the account exists.
”.
CLI user has
10.Repeat Steps 3 through 9 to modify other user passwords and access levels.

Saving and Restoring Configuration Data

Symmetricom recommends that you keep an electronic version and/or a written version of the provisioning changes you make to the TP5000 probe module.
The configuration is automatically backed up one minute after a configuration change.
To backup up the current configuration immediately, enter the following:
1. Type
Example
set configuration backup local and press Enter.
To change the current configuration to factory default while preserving user and IP information, enter the following:
1. Type
set configuration default and press Enter.
Example
To change the current configuration to the factory default, enter the following:
1. Type
68 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
set configuration factory and press Enter.
Notes:
Setting the configuration to default will not overwrite the user login information or the system’s IP address. Setting the configuration to factory will overwrite the user login information and return the system’s IP address to the factory set address.
Page 71

Probe-Related Commands

See Chapter 3 of the TP5000 User’s Guide for details about using CLI commands.
show option
This command displays the option mode and key information.
Command Syntax:
show option
Example:
tp5000> show option
Response:
OPTION SERVICE OPTION KEY KEY VALIDITY
-------------- ---------- -----------­SNMP not available invalid PTP PROBE not available invalid
Chapter 3 Configuring
Probe-Related Commands
Level : User, Config, and Admin
set option
This command sets the option mode and key information. The key information is printed on the Symmetricom-issued certificate for the unit.
Command Syntax:
Example:
To enable the probe option with an authentication key of “ABCDEF”,
Note: The SNMP feature can either be ordered in a new system, or added to a previously ordered system as a field upgrade. Contact your Symmetricom representative for more information. The PTP Probe option is only available as part of the TimeAnalyzer 7x00.
set option probe <key>
tp5000> set option probe ABCDEF
Level : Config and Admin only
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 69
Page 72
Chapter 3 Configuring Probe-Related Commands
show ptp-probe
This command is used to begin PTP packet measurements. This command can also be used to display the Probe configuration, which includes Grandmaster Clock ID, IP address and VLAN (if applicable), interval rate, and lease duration.
Command Syntax:
To display the probe configuration data:
To begin gathering data with the probe:
Note: The Time Analyzer option must be purchased to enable the
functionality of all “ptp-probe” commands.
show ptp-probe config ioc1
show ptp-probe data ioc-[y]
where
y= 1|2, the port number on the IOC
Related: To stop gathering data with the probe, press the Enter key.
Example 1:
To display the configuration data of the Probe
tp5000> show ptp-probe config ioc1
Example 2:
To display the test data from port 1 of the Probe
tp5000> show ptp-probe data ioc-1
Response 2:
F,00011,1229541029,099868364,1229541029,099881920,0000000000, B,00011,1229541029,562553320,1229541029,562539700,0000000000, F,00012,1229541030,099868364,1229541030,099881776,0000000000, B,00012,1229541030,562553320,1229541030,562539700,0000000000, F,00013,1229541031,099868364,1229541031,099881936,0000000000, B,00013,1229541031,562553288,1229541031,562539700,0000000000, F,00014,1229541032,099868364,1229541032,099881776,0000000000, B,00014,1229541032,562553288,1229541032,562539700,0000000000,
Level : Config and Admin only
70 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 73
set ptp-probe
Use this command with a TP5000 that is configured in Probe mode to specify the Grandmaster IP address and Clock ID. This command is used to set the interval rate for Sync and Delay Requests and the lease duration for the specified Probe IOC card and port. This command is also used to restart the specified Probe IOC card and port.
Command Syntax:
To set the IP address and Clock ID for the Grandmaster being tested by the
Probe:
Chapter 3 Configuring
Probe-Related Commands
Note: The Time Analyzer option must be purchased to enable the
functionality of all “ptp-probe” commands.
set ptp-probe ioc1-[y] active-probe gm-clock-id <clock-id> gm-ip <addr> [vlan-id <vlan>]
where
y = 1|2, the port number on the IOC <clock-id> is in the form “xx:xx:xx:xx:xx:xx:xx:xx”
To set the interval rate for Sync and Delay Requests from the Probe:
set ptp-probe ioc1-y active-probe interval <int-value>
where
The interval is 2
To set the maximum lease duration for the Probe:
set ptp-probe ioc1-[y] active-probe duration <dur-value>
To restart the specified Ethernet port on the Probe:
set ptp-probe ioc1-[y] restart
z
seconds, where z is <int-value>.
Ranges: Interval = 0.015625 to 128 seconds (int-value = -6 to 7)
Duration = 10 to 1000 seconds
Defaults: Interval = 1 second (int-value = 0)
Duration = 300 seconds
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 71
Page 74
Chapter 3 Configuring Recovering From Reset to TP5000 Factory Defaults
Example 1:
For a Grandmaster Clock ID of 00:B0:AE:FF:FE:00:00:06, GrandmasterIP address of 192.168.1.11, and probe VLAN ID of 35:
TP5000>set ptp-probe ioc1-1 active-probe gm-clock-id 00:B0:AE:FF:FE:00:00:06 gm-ip 192.168.1.11 vlan-id 35
For a Grandmaster Clock ID of 00:B0:AE:FF:FE:00:00:06, GrandmasterIP address of 192.168.1.11, and no VLAN:
TP5000> set ptp-probe ioc1-1 active-probe gm-clock-id 00:B0:AE:FF:FE:00:00:06 gm-ip 192.168.1.11 <enter>
Example 2:
For an interval of 0.25 seconds ( -2 ) on Port 1 of IOC1:
tp5000> set ptp-probe ioc1-1 active-probe interval -2
Example 3:
For a duration of 500 seconds on Port 1 of IOC1:
tp5000> set ptp-probe ioc1-1 active-probe duration 500
Level : Config and Admin only

Recovering From Reset to TP5000 Factory Defaults

The TP5000 Grandmaster factory defaults are different than the initial settings for theTP5000 probe module as part of the TimeAnalyzer 7x00. If the TP5000 is reset to factory defaults, the TP5000 will return to the TP5000 Grandmaster defaults. The following procedures must be used to return the TP5000 to the proper probe settings.
Enabling the Probe Option on the TP5000Enabling Probe ModeSetting Up Telnet FirewallProvisioning TOD Source for TP5000 Probe ModuleProvisioning the Reference Output for TimeWatchProvisioning the 10MHz & 1PPS OutputsDisabling Battery (-48V) B and Holdover Alarms
72 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 75
Recovering From Reset to TP5000 Factory Defaults
Chapter 3 Configuring

Enabling the Probe Option on the TP5000

The probe option must be enabled with the Symmetricom authentication key in order to use the TP5000 module as a probe. The TimeAnalyzer 7x00 is shipped with the TP5000 module enabled for the Probe option. The default condition for the TP5000 is that the Probe option is disabled.
Use the “show option” command to check if the probe option is still enabled for the TP5000. Use the “set option” command to enter the authentication key for the probe option.
Example:
To enable the probe option with an authentication key of “ABCDEF”,
tp5000> set option probe ABCDEF

Enabling Probe Mode

The Probe mode must be enabled to use the TP5000 as a probe. Use the following procedure to configure TimeMonitor PDV and the TP5000 probe
module for PDV measurements:
1. Select the radio button for the Command mode.
2.
Click the Login button to login to the Probe. The Probe will prompt for Username and Password.
3. Select the radio button for the desired Probe port.
4. Click the Probe Enable button to enable the selected port as a Probe.
5. Click the Logout button.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 73
Page 76
Chapter 3 Configuring Recovering From Reset to TP5000 Factory Defaults

Setting Up Telnet Firewall

The firewall for Telnet must be set to “Allow” in order to initiate Telnet sessions with the TP5000 probe module. The TimeAnalyzer 7x00 is shipped with the firewall for Telnet for the TP5000 module set to “Allow”. The default condition for the Telnet firewall for TP5000 is “Block”.
With CLI
Use the “show firewall” command to check if the firewall for Telnet is still set to “Allow” for the TP5000 module. If it is set to “Block”, use the “set firewall imc telnet allow” command to reset the firewall to “Allow”.
Example:
tp5000> set firewall imc telnet allow
With TimeMonitor PDV
Use the following procedure to set the TP5000 probe module’s Telnet firewall to”Allow”:
1. Select the radio button for the Command mode.
2.
Click the Login button to login to the TP5000 probe module. The TP5000 probe will prompt for Username and Password.
3. Click the Firewall Off button in order to allow for Telnet communications with the TP5000 probe module..

Provisioning TOD Source for TP5000 Probe Module

The default TOD source for the TP5000 probe module in the TimeAnalyzer 7x00 is System, with the Rb oscillator in free-run mode. The default TOD source for the TP5000 module is GPS. Use the "set tod-source sys " command to set the TOD source for the TP5000 probe module to System.
"Type "set tod-source sys "
Example:
For a duration of 500 seconds on Port 1 of IOC1:
74 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 77
Recovering From Reset to TP5000 Factory Defaults
tp5000> set tod-source sys
Chapter 3 Configuring

Provisioning the Reference Output for TimeWatch

The two programmable I/O ports on the I/O module, Port1 and Port 2, are software configurable as E1 outputs. The “set io” command is used to set Ports 1 and 2 as outputs. The “set output” command is used to provision the E1 output characteristics. See Figure 3-4.
Figure 3-4. Set Output Commands - TP5000 Probe Module
1. Login at the Admin or Config-user level.
2. Type
3. Type
4. Type
5. Type
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 75
set output frametype port1 2048khz and press Enter. set io port1 output and press Enter. set output generation port1 free-run and press Enter. set output state port1 enable and press Enter.
Page 78
Chapter 3 Configuring Recovering From Reset to TP5000 Factory Defaults

Provisioning the 10MHz & 1PPS Outputs

The I/O card of the TP5000 probe module is equipped with 1PPS and 10MHz dedicated output ports in addition to the two programma ble E1 I/ O ports. The 1PPS and 10MHz output ports are disabled by default for theTimeAnalyzer 7x00. The 1PPS and 10MHz output ports are enabled by default for the TP5000 probe module.
The
set output-10m command is used to provision the 10MHz output. See Figure
3-5. The Figure 3-6
Figure 3-5. Set Output-10M Commands - TP5000 Probe Module
set output-pps command is used to provision the 1 PPS output. See
Figure 3-6. Set Output-PPS Commands
Provisioning 10 MHz Port
1. Login at the Admin or Config-user level.
2. Type
3. Type
set output-10m generation normal and press Enter. set output-10m state enable and press Enter.
76 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 79
Provisioning 1PPS Port
1. Login at the Admin or Config-user level.

Provisioning the TP500 PTP Client Module

Chapter 3 Configuring
2. Type
3. Type
set output-pps generation normal and press Enter. set output-pps state enable and press Enter.

Disabling Battery (-48V) B and Holdover Alarms

The Battery B and Holdover alarms for the TP5000 probe module are disabled by default for the TimeAnalyzer 7x00. These alarms are enabled by default for the TP5000. Use the “set alarm-config state alarm <#> disable” command to disable the Battery B (#15) alarm and Holdover alarms (#29 and #30).
1. Login at the Admin or Config-user level.
2. Type
3. Type
4. Type
set alarm-config state alarm 15 disable set alarm-config state alarm 29 disable set alarm-config state alarm 30 disable
Provisioning the TP500 PTP Client Module
The if0 interface connection is available for a non-VLAN connection. Non-VLAN Telnet
sessions with the TP500 module require that:
Interface if0 is enabled (default) IP address for TP500 module is set using the “set IP” command
The TP500 module's factory default for if0 is enabled which means the unit is readyfor communication with all non-VLAN PTP devices and remote telne t ses sion s
Non-VLAN PTP device communication can begin once a GrandMaster’s (GM1 or GM2) IP address information is configured for connection using the “set ptp” command.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 77
Page 80
Chapter 3 Configuring Provisioning the TP500 PTP Client Module
The TP500 module provides interface connection support for up to four different VLANs, which can be any combination of network devices and remote telnet sessions. Refer to the TimeProvider 500 User’s Guide for procedures on how to provision the TP500 module for VLAN Telnet sessions and VLAN PTP device communication.
Use thefollowing procedure to provision the TP500 PTP client module for non-VLAN Telnet sessions and non-VLAN PTP device communication..
Notes: The Default IP address values for the TP500 module are :
Notes: The Default user name and password for the TP500 module
are:
IP Address 192.168.1.9 Mask 255.255.255.0 Gateway 192.168.1.1
User admin Password symmadmin**
1. Ensure that power is connected to the TP500 module.
2. Use serial-to-USB cable to connect the Craft port to the USB port of the laptop computer.
3. Establish serial communication with TP500 module using terminal emulation, with the following settings: COM1 port: 8 data bits, 1 stop bit, no parity, 57,600 baud rate, no flow control.
All Programs>Accessories>Communication>HyperTerminal
4. Log in to TP500 module
User: admin PW: symmadmin**
5. Configure the Client IP Address The TP500 module only supports static IP addresses in 2.0 release. A choice of
static or DHCP will be available in a future release. Note: The TP500 module has a default static address of 192.168.1.9, netmask 255.255.255.0 and gateway
192.168.1.0 already preconfigured on the system.
78 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 81
Provisioning the TP500 PTP Client Module
Chapter 3 Configuring
5. Set the IP address with the command set ip address {if0 | if1 | if2 | if3 |
if4}<IP-adress> netmask <mask-value> gateway <gateway-value>
Note: The if0 interface connection is used for non-VLAN connections. The if1, if2, if3, and if4 interface connections are available for VLAN
connections for a Grandmaster (gm1 or gm2), a node manager device capable of transporting PTP packet messages, or remote Telnet sessions.
See TimeProvider 500 User’s Guide for procedures on how to configure for VLAN.
Example:
To set the TP500 module IP address to 192.168.5.68, with a mask of
255.255.255.0 and a gateway of 192.168.5.1, for non-VLAN connections (if0):
TP500> set ip if0 address 192.168.5.68 netmask 255.255.255.0
gateway 192.168.1.0
6. Verify that the IP address has been set correctly with the show ip command.
Example:
TP500> show ip
7. Configure the target IP Address for the Grandmaster with the command set ptp
address {gm1 | gm2 | node_mgr} <IP-address>
The TP500 module has a default IP Address for the Grandmaster of 0.0.0.0.
Notes: The TP5000 probe module will not attempt communications with the Grandmaster if this setting is left to the default value of
0.0.0.0.
Example:
To set the target IP address for Grandmaster 1 to 192.168.5.52:
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 79
Page 82
Chapter 3 Configuring Provisioning the TP500 PTP Client Module
TP500> set ptp address gm1 192.168.5.52
Notes: The default interface for gm1 in non-VLAN connection is if0, so the command to assign gm1 to interface if0 is not shown.
If the TA7x000 was used previously with a VLAN connection,attach the Grandmaster to the non-VLAN interace, if0, with the command
set interface attachments {GM1 | GM2 | NODE-MGR}
{if0 | if1| if2 | if3 | if4}.
See the TP500 User’s Guide for additional details about configuring VLAN and interface attachments.
Example: TP500>set interface attachments gm1 if0
8. Confirm network communication between the TP500 module and Grandmaster with the ping command.
Use an Ethernet cable to connect the TP500 module 1588 port to an Ethernet switch that is connected to the same network as the Grandmaster. The small green LED on upper-right corner of "1588" connector should be lit.
Example:
TP500> ping 192.168.1.11 <enter>
Pinging 192.168.1.11 [192.168.1.11]: Reply from 192.168.1.11: time=0ms Reply from 192.168.1.11: time=0ms Reply from 192.168.1.11: time=0ms Reply from 192.168.1.11: time=0ms
80 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 83

TimeWatch Probe Module Configuration

TimeWatch Probe Module Configuration
This section provides information on configuring the TimeWatch probe module.

Connecting to the TimeWatch Probe Module

On or prior to installation the probe must be configured to setup its IP address. A menu-driven ASCII interface is available on the RS-232 Management Port and telnet/SSH on the Ethernet port.
Connect Via RS-232
First, connect to the Management Port (9-way female D-Type) of the TimeWatch probe module using the serial-to-USB cable. Use a terminal emulator like HyperTerminal to communicate with the probe module; the interface is configured to operate at 115,200 baud with 8 data bits, no parity, 1 stop bit and no flow control. The terminal emulator should be set to emulate VT102; under HyperTerminal this mode is not fully tested so use auto-detect or VT100J mode. In HyperTerminal the boxes will be drawn with other characters but the application will function correctly.
Chapter 3 Configuring
Tera Term Pro can emulate VT102 correctly and works well with the TimeWatch module . It can be downloaded from http://hp.vector.co.jp/authors/VA002416/teraterm.html. On connection to the probe a login prompt will be displayed - press enter if thisdoes not occur. It may be necessary to power-cycle the probe to obtain the prompt, which will appear after startup messages. After connection a login screen will show. Enter the username swconf and when prompted enter the password fr7ebi. This will enter the menu-driven configuration interface as described below.
Notes: The Default user name and password for the TimeWatch module are:
User Password fr7ebi
swconf
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 81
Page 84
Chapter 3 Configuring TimeWatch Probe Module Configuration
Connect Via Ethernet
It is also possible to use a telnet or SSH connection to configure the Probe. A crossover Ethernet cable or Ethernet hub with no network connection is required to connect a PC to the Ethernet port if it is not on an existing network.
Notes: The Windows "telnet" program will not correctly connect to the probe, so an alternative application is required. Symmetricom recommends 'Putty', which is available on the TimeAn alyzerr 7500 support CD. It is also available at:
http://www.chiark.greenend.org.uk/~sgtatham/putty/
Using a telnet/SSH client (not Windows telnet), connect to the probe. After connection a login screen will show. Enter the username swconf and when prompted enter the password fr7ebi. This will enter a menu-driven configuration interface as described in the following section.
Configure the Management PC
The factory default address of the probe is 192.168.1.10. It will be necessary to change the IP address of the PC configuring the probe. Under Windows this can be achieved by opening Network and Dial-up Connections from Control Panel, right-clicking on Local Area Connection, selecting Properties, selecting Internet Protocol, and selecting Properties. Set the IP address to 192.168.1.2 and subnet mask to 255.255.255.0. If the IP address of the probe has been previously changed then the new IP address needs to be entered via telnet or SSH; the PC must be connected to the same network. If the IP address of the probe is not known this can be changed or retrieved as detailed in Connect Via RS-232 above.
Notes: The Default IP address for the TimeWatch module is:
IP Address 192.168.1.10
Notes: To avoid connectivity problems, remember to reset the IP address of the PC before attempting to connect to other networks.

Configuring the TimeWatch Probe Module

This section describes the TimeWatch probe module menus. Only the General and Ethernet menus are needed to configure TImeWatch with the TimeAnalyzer 7500. Other menus may be described but can be ignored. See the TimeWatch Operation and Service Guide for additional details.
Use the General menu to configureTimewatch to operate with TimeMonitor, and
to select the port to be used for communication with TimeMonitor.
82 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 85
TimeWatch Probe Module Configuration
Use the Ethernet menu to set the IP address, netmask, and gateway for
Chapter 3 Configuring
TimeWatch, or to configure the TimeWatch for DHCP.
Navigating the Menu System
Notes: Depending on the application used to access these
menus, the screen may be laid out slightly differently. These example screens are obtained using Putty interfacing via a LAN.
Upon login, the following screen will display for a few seconds:
Figure 3-7. Example of Initial Screen After Login
After this, the main menu screen will display:
Figure 3-8. Main Menu Screen
Use the up and down arrows to select the desired menu item (Probe, Server, Ethernet, Dialup or GPRS) and press enter to select that menu option.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 83
Page 86
Chapter 3 Configuring TimeWatch Probe Module Configuration
Probe Menu
Figure 3-9. Probe Menu Screen
Serial #: This is the serial number of the probe and identifies it as a unique entity
on the TimeWatch Server (n/a for TimeAnalyzer 7x00).
CW-10 GPS installed : T imeWatch Standard module is used in the T imeAnalyzer
7500, so this option does not apply. For reference, a hash [#] indicates that the option is enabled, a blank [ ] indicates it is disabled
[Main Menu]: Use the down arrow to select this, then press enter.
Features Menu
Figure 3-10. Features Menu
Lab License: This license key, when entered, unlocks the TimeMonitor
application functionality . The T imeAnalyzer 7500 is provided with the License key pre-entered.
84 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 87
TimeWatch Probe Module Configuration
Chapter 3 Configuring
General Menu
Figure 3-11. General Menu Screen
Operate With: Use left and right arrows to select the .the TimeMonitor
application. The TimeAnalyzer 7500 is shipped with TimeMonitor Watch selected.
Notes: When TimeMonitor is selected, the only configurable
option is the Lab Configuration port.
Comms: N/A with TimeAnalyzer 7500.Host: N/A with TimeAnalyzer 7500.Port: N/A with TimeAnalyzer 7500.HTTP Proxy Host: N/A with TimeAnalyzer 7500.HTTP Proxy Port: N/A with TimeAnalyzer 7500.Port (Lab Configuration): This is the port number that the TimeWatch probe
module will use to communicate with the TimeMonitor application. Only applicable if TimeMonitor has been specified in the Operate With field.
[Main Menu]: After appropriate changes are made, use the down arro w to select
this, then press enter.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 85
Page 88
Chapter 3 Configuring TimeWatch Probe Module Configuration
Ethernet Menu
Figure 3-12. Ethernet Menu Screen
Notes: When Operate With TimeMonitor is selected - the only
communications medium that is supported is Ethernet.
Use DHCP: If this is enabled then the module will request its Ethernet
configuration over the network. Do not enable unless the network is co nnected to a DHCP server. IP address, subnet mask, gateway and DNS addresses only need to be entered when DHCP is disabled. A hash [#] indicates that DHCP is enabled, a blank [ ] indicates it is disabled. Toggling between enabled and disabled is achieved using the space bar.
IP Address: Use to set the IP address of the probe (e.g. 172.31.2.5). It may only
be entered if DHCP is disabled.
Subnet: Use to enter the subnet mask of the network that the probe is connected
to (e.g. 255.255.255.0). It may only be entered if DHCP is disabled.
Gateway: Use to enter the IP address of the probe's default gateway (e.g.
172.31.2.1). It may only be entered if DHCP is disabled.
DNS 1: The IP address of the primary Domain Name Server. This is not required
if no domain names are used on this screen (i.e. all addresses are IP addresses). It may only be entered if DHCP is disabled.
DNS 2: The IP address of a secondary Domain Name Server. This is optional,
and may only be entered if DHCP is disabled.
[Main Menu]: After appropriate changes are made, use the down arrow to select
this, then press return.
Dial-Up Menu
Ignore for use in TimeAnalyzer 7500.
86 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 89
GPRS Configuration
Ignore for use in TimeAnalyzer 7500.
TimeWatch Probe Module Configuration
Chapter 3 Configuring
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 87
Page 90
Chapter 3 Configuring TimeWatch Probe Module Configuration
88 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 91

Chapter 4 Operating

This chapter provides guidelines for operating the TimeAnalyzer 7x00.
In This Chapter
PDV MeasurementsE1 Measurements
E1 Measurements with TP500E1 Measurements with External Client
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 89
Page 92
Chapter 4 Operating PDV Measurements

PDV Measurements

Setup

Make the following connections shown in Figure 4-1 for PDV measurements:
Figure 4-1. PDV Measurement Configuration (power connections not shown)
Probe GPS Port Connect the GPS antenna to the GPS port on the
IMC module, and mount the antenna. Use the procedure
Setting for Free-Run of Internal Rb Clock, on
page 64 if GPS is not available.
Probe PTP Port Plug an SFP transceiver into IOC-1 or IOC-2, and
connect it to a network switch using an Ethernet cable.
Probe MGMT Port Connect the MGMT Port of IMC to the TimeAnalyzer
switch or directly
90 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 93
Chapter 4 Operating
PDV Measurements

Measuring PDV

Use the following procedure for making PDV measurements with the TP5000 probe module and TimeMonitor PDV.
1. Select the radio button for the Command mode.
2. Select the radio button for the desired sample rate. For rates slower than 1 Hz, check the Reduce Rate box to access additional values. (Reduced rate values are actually values for the interval between samples.)
or
3. Enter the desired value for the Lease Duration.
4. Enter the desired file name and destination directory. The name need only be specified once because the software will reuse a file name for new files and add on a date and time to the file name if the file already exists. Also enter the desired title for the file.
5. Select the radio button for the desired probe module port.
6. To verify the test setup before beginning the measurement, click the Check button.
The top of the main screen will display the message:
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 91
Page 94
Chapter 4 Operating PDV Measurements
The following type of information will appear in the main display if the setup is configured correctly:
7. To stop the Check, click the Stop button.
8. To start the measurement, click the Start button. The top of the main screen will display the elapsed time of the measurement.
9. To stop the measurement, click the Stop button.
92 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 95

E1 Measurements

Setup

Figure 4-2 below shows the configuration for E1 measurements using the
TimeAnalyzer 7500.
Figure 4-2. E1 Measurement Configuration (power connections not shown)
Chapter 4 Operating
E1 Measurements

E1 Measurements with TP500

In addition to the connections shown in Measuring PDV, on page 91, make the following connections to the TimeWatch probe module’s ports:
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 93
Page 96
Chapter 4 Operating E1 Measurements
REF IN Port: Connect the REF IN port to Port 1 on the I/O card of
MEASURE IN Port Connect the MEASURE IN port to the Span1 or S p an2
LAN Connect the LAN port to the internal Ethernet switch
The Standard variant of the TimeWatch module is designed to perform an MRTIE measurement between two external clock inputs. The inputs are independent and each can be either a 2.048 MHz clock or an E1 2.048 Mbit/s signal. The probe module uses an internal TIE measurement engine that compares two external inputs at 80 samples per second (80Hz) and calculates the relative phase. This data is averaged to produce one TIE value for each second, this is then used to calculate the MRTIE between the two inputs.
TimeMonitor Watch Software
The TimeMonitor Watch software is used to interface with the TimeWatch module. See Figure 4-3 below for a view of the Main Screen for this software. See the TimeMonitor Watch User Guide for additional details.
the TP5000 probe module using a BNC-to-mini BNC cable
port of TP500 slave using BNC-to-BNC cable.
using Ethernet cable
94 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 97
Figure 4-3. TimeMonitor Watch - Main Screen
Chapter 4 Operating
E1 Measurements
Procedure
Use the following procedure for making E1 measurements using TimeAnalyzer 7500 and the TimeMonitor Watch software.
1. Connect E1 or 2M Clock signal from TP5000 to the Reference port of TimeW atch. The default output fo rmat from Port 1 of the TP5000 probe module is 2048 kHz.
2. Connect E1 output (Span1 or Span2) from the TP500 module to the Measurement port of TimeWatch.
3. Start the TimeMonitor Watch module on the PC used to communicate with TimeWatch.
4. Click the TCP/IP button.
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 95
Page 98
Chapter 4 Operating E1 Measurements
5. Enter the IP address for the TP500 module to be measured. Click the OK button to return to the main screen.
6. Use the TimeMonitor W atch radial button for Ref to select E1 or 2M as the signal format of the Reference input, as shown in Figure 4-4. Select 2M if using the TP5000 probe module in its default mode for Port 1 output.
Figure 4-4. TimeMonitor Watch - Select Measurement and Ref Signals
7. Use the TimeMonitor Watch radial button for Meas to select E1 or 2M as the signal format from the TP500 client module , as shown in Figure 4-4 above. Select 2M if using the TP500 client module in its default mode for Sp an 1 or S p an 2 output.
8. Use the Sample Rate radial buttons to select the desired sample rate for the E1 measurement, as shown in Figure 4-4 above
9. To begin the measurement, click the Start button.
10.
During a measurement, a plot of the phase collected so far can be produced by clicking
the View button, as shown in
Figure 4-5.
11.By clicking the MTIE button, Live MTIE can be viewed, as shown in Figure 4-6.
12.By clicking the TDEV button, Live TDEV can be viewed, as shown in Figure 4-7.
13.To end the measurement, click the Stop button.
96 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Page 99
Figure 4-5. Live TimeWatch Phase Data
Chapter 4 Operating
E1 Measurements
Figure 4-6. Live TimeWatch MTIE Data
098-00208-000 Revision A – February, 2010 TimeAnalyzer 7x00 User’s Guide 97
Page 100
Chapter 4 Operating E1 Measurements
Figure 4-7. Live TimeWatch TDEV Data

E1 Measurements with External Client

Setup
In addition to the connections shown in connection from TP500 module to the network switch), make the following connections to the TimeWatch module ports:
REF IN Port: Connect the REF IN port to Port 1 on the I/O card of
the TP5000 probe module using a BNC-to-mini BNC cable
MEASURE IN Port Connect the MEASURE IN port to the E1 output of the
external client using the appropriate cable.
LAN Connect the LAN port to the internal Ethernet switch
using Ethernet cable
Procedure
1. Connect E1 or 2M signal from TP5000 module to Reference port of TimeWatch module. The default output from Port 1 of the TP5000 probe module is 2048 kHz.
Measuring PDV, on page 91 (except
2. Connect E1 output from the external client to Measurement port of the TimeWatch probe module.
98 TimeAnalyzer 7x00 User’s Guide 098-00208-000 Revision A – February, 2010
Loading...