Super Systems MGA6000 User Manual

Operations Manual
Gas Analyzer
MODEL MGA6000
Continuous Multi-Gas NDIR Analyzer
Super Systems, Inc. is not responsible for damages incurred due to a failure to comply with these instructions. If at
any time there are questions regarding the proper use of this analyzer, please contact us at (800) 666-4330 for
assistance.
7205 Edington Drive
Cincinnati, OH 45249
513-772-0060 / 800-666-4330
Fax: 513-772-9466
www.supersystems.com
SSi Manual #4563 Rev. D Page 1 Multi-Gas Model 6000
Table of Contents
WARNING................................................................................................................... 3
Introduction ................................................................................................................................... 4
Possible Configurations.................................................................................................................. 4
Specifications ................................................................................................................................. 4
Part Number Designations ............................................................................................................. 5
Electrical Connections .................................................................................................................... 5
Basic Operating Description........................................................................................................... 5
Keypad Assignments ...................................................................................................................... 6
Analyzer Start-Up Procedure ......................................................................................................... 6
Sampling Criteria............................................................................................................................ 6
Filters.............................................................................................................................................. 7
Condensation / Moisture................................................................................................................ 7
Menu List ........................................................................................................................................ 8
Note about Menu Numbers ..................................................................................................9
Main Page – Menu Page 1 ....................................................................................................9
IR Status – Menu Page 2......................................................................................................9
General Information – Menu Page 3 ..................................................................................11
Pump Control - Menu Page 4..............................................................................................12
Set Display Values – Menu Page 5 .....................................................................................12
Calibration Dates and Run Times - Menu Page 7 ...............................................................13
About/Sign-On – Menu Page 10 ........................................................................................13
Revision Display – Menu Page 11 ......................................................................................13
Set The Date and Time - Menu Page 16 .............................................................................14
Sampling Parameters - Menu Page 17 ...............................................................................14
IR Cell Zero Calibration – Menu Page 18 ...........................................................................16
4-20mA Output Assignment – Menu Page 19 ....................................................................17
Auto Calibration Setup – Menu Page 20 ............................................................................18
Main Display Set – Menu Page 21 ......................................................................................18
Communications Setup – Menu Page 22............................................................................19
Calculation Factors - Menu Page 23...................................................................................22
IR Cell Span Calibration – Menu Page 24 ..........................................................................22
Set Pass Codes - Menu Page 25 .........................................................................................23
Set IP Address- Menu Page 26...........................................................................................24
H2 Cell Calibration – Menu Page 27...................................................................................24
Auto Sequence Setup – Menu Page 28...............................................................................25
Gas or CV Configuration – Menu Page 29 ..........................................................................25
MGA6000 Spare Parts .................................................................................................................. 26
Appendix 1 – MGA Side Vents ...................................................................................................... 27
Appendix 2 – MGA Modbus Registers .......................................................................................... 28
Revision History ........................................................................................................................... 42
SSi Manual #4563 Rev. D Page 2 Multi-Gas Model 6000
WARNING
MOISTURE
WHEN A HOT GAS CONDENSES MOISTURE HAS THE POTENTIAL TO FORM.
WHEN USING THIS INSTRUMENT YOU MUST GUARD AGAINST MOISTURE
FORMING IN THE SAMPLE LINES AND THE BOWL FILTER. THIS CAN BE
ACCOMPLISHED THROUGH SETTING THE PUMP OPERATION TO “AUTO” AND
ENTERING APPROPRIATE SAMPLING PARAMETERS ON SCREEN #17.
THE WARRANTY WILL BE VOIDED IF PROPER SAMPLING TECHNIQUES ARE NOT
SSi Manual #4563 Rev. D Page 3 Multi-Gas Model 6000
MGA6000 Operating Instructions
Introduction
The Model MGA6000 is a continuous Multi-Gas Analyzer using NDIR (Non-Dispersive Infra-Red) technology. It is capable of measuring between one and four gases, depending on how it is configured at the factory prior to shipment. The unit is designed and manufactured for the atmosphere heat treating industry.
Possible Configurations
CO range: 0.00 to 30.00 % *
range: 0.000 to 2.000 %
CO
2
High range: 0.0 to 20.00 %
CO
2
CH
range: 0.00 to 15.00 % *
4
range: 0.1 to 25.0%
O
2
Hydrogen: 0.0 to 100.0% Dew point 0 to 80 F
* Note: The CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown above, however both sensors are capable of measuring gas concentrations of up to 100.00%.
Specifications
Sampling method: Extraction by internal pump (when necessary)
Accuracy and repeatability: ± 1% of full scale
Flow Meter: 0 to 2 SCFH, mounted on enclosure front and also displayed on­screen
Pump Operation: On/Off/Auto Sample
AC Power: 90 to 230 VAC, 50 to 60 Hz, 60 Watts
Communications: Ethernet, RS485
4-20 mA Outputs: Four (4) user-defined
Digital Inputs: Two (2) for stopping pump and inhibiting COF/PF adjustment
Relays: One (1) for alarm or control
Operating Temperature: 32° to 122° F (0° to 50° C)
Dimensions: Approx. 16” x 14” x 8”
Weight: Approx. 20 lbs.
SSi Manual #4563 Rev. D Page 4 Multi-Gas Model 6000
Part Number Designations
The following table references the gases available in the MGA6000s with their applicable part numbers.
Part
Number:
CO CO2 CH4
H2
13404 13391 13392 13393 13394 13395 13396 13397 13398 13399 13402 13403
X X X X X X X X X X X X X X X X X X X
X X X X X X
Oxygen cell option: Part number 13400 will add Oxygen (O2) to the MGA.
Example: Part Number 13399 would contain: Carbon Monoxide (CO), Methane or Natural Gas (CH4), and Hydrogen (H2).
Electrical Connections
1001 + 1002 ­1003 GND 1101 + 1102 ­1103 GND 1211 + 1212 ­1221 + 1222 ­1231 + 1232 ­1241 + 1242 ­1301 + 1302 ­1411 + 1412 ­1421 + 1422 ­1501 + 1502 -
90-265 VAC
RS 485 COMMS
(MASTER)
4-20mA #1
4-20mA #2
4-20mA #3
4-20mA #4
Alarm/Control
Digital Input #1
(Pump Stop)
Digital Input #2
(Inhibit Adjustment)
RS 485 COMMS
(AUTO-CAL BOX)
Basic Operating Description
The Model MGA6000 has been designed for the simultaneous analysis of selected gases in heat-treat furnace atmosphere gases. It has a 16 line by 40-character LCD display and a 4 x 4 keypad for the operator interface. Information is presented to the operator on various screens. The operator selects the appropriate page and enters the number using the keypad.
SSi Manual #4563 Rev. D Page 5 Multi-Gas Model 6000
Keypad Assignments
All of the functions of the MGA6000 are controlled from the keypad. The functions of each button are as follows:
09 are used to enter numeric data that can vary with each page.
“.” is used to enter a decimal point where necessary for data entry.
Enter is used when entering numeric data from the keypad or initiating automatic functions.
Ï and Ð are used to navigate through the menu options or scroll through data on a specific page.
Esc is the escape key. This clears any entered text, and if continuously pressed toggles between the main page and the menu list.
On some screens, This button is used to change a selection or view additional data.
123
456
789
ESC.
0
z
Ï
ENTER
Ð
Analyzer Start-Up Procedure
Turn the power switch ON and allow the instrument to warm up for at least three minutes. During this time, the gas values on the screen will show “********”. While the instrument is warming up, the user will be prevented from accessing the calibration screens since the sensors have not had enough time to provide accurate data. It is recommended that the instrument is powered on for at least five minutes prior to use, although measurements can be taken within only three.
Sampling Criteria
Accurate readings are only possible if the sample is taken from a clean (free of excess carbon buildup) sample port. The current state-of-the-art technology associated with infrared gas detection requires that a clean, soot free sample be available for analysis. This will also increase the life of the filter elements, and reduce the possibility that soot will enter the unit and contaminate the sensors. On a furnace, the ideal port would be found on SSi’s Sample Port (Part Number 20264). This sample port contains a high-temperature filter that contains the same element that is in the filter on the side of the analyzer. This filter will not only prevent contaminants from entering the analyzer, but since the filter is at the furnace it will also protect the sample lines from the furnace to the analyzer.
SSi Manual #4563 Rev. D Page 6 Multi-Gas Model 6000
On a generator, a dedicated sample port should be available. This sample port should be blown out before it is used, which will remove any soot that has accumulated in the line.
The ideal flow rate for sampling should be between 1.0 and 1.5 Standard Cubic Feet per Hour (SCFH). A visual indication of flow rate can be obtained through the flow meter located on the face of the enclosure, or by the digital representation of flow on the left side of the display screen. The flow meter on the outside of the enclosure also contains a dial that allows the user to restrict the flow, if necessary, to maintain an appropriate flow rate. If the sample gas is not under pressure, the internal pump can be used to extract it. For additional information on the operation of the pump, please refer to the section option 4).
Pump Control
(menu
Filters
The side of the enclosure contains a large bowl filter that is intended to prevent soot and other contaminants from entering the instrument. Periodic inspection of this filter will ensure reliable operation of the MGA6000. When this element is new, it is a white/eggshell color. As this filter becomes dirty it will turn black and the flow rate of the analyzer will begin to diminish. Replacing the filter element is as simple as unscrewing the clear bowl from the filter housing, and unscrewing the retaining plug at the base of the element. Information on replacement filters can be found in the section of this manual entitled “Spare Parts”.
Condensation / Moisture
When a hot gas is cooled rapidly, moisture in the gas can condense and form water. This water can collect in the sample tubing and eventually enter the bowl filter. Care must be taken to ensure that no water enters the unit, as this will cause permanent, irreversible damage to the sensors. The unit should be closely monitored during operation to determine if moisture is collecting in the bowl filter. If this is the case, the bowl filter basin can be removed and emptied by unscrewing it. Although water in the bowl filter will not cause damage to the unit, this filter should not be used as a condensation receptacle. If moisture has collected in the bowl filter, sampling should be stopped, and steps should be taken to prevent this from continuing before operation is resumed.
SSi Manual #4563 Rev. D Page 7 Multi-Gas Model 6000
Menu List
The menu list shows the available pages, displayed six at a time. To access the list, press the Esc key. Depending upon where the user is starting from, it may require pressing this button more than once. The up (Ï) and down (Ð) arrows are used to scroll through the selections, which are repeated below. To go to a specific page, either type in the page number and press Enter, or use the arrow keys to highlight the selection and then press Enter.
Menu options 6, 8, 9, 12, 13, 14, and 15 are reserved for future use and are not currently used. To minimize the possibility of unintended modifications to the instrument, certain menu pages will require the entry of a pass code to access them. Pages 1 through 15 are Operator level screens that do not require any security codes. Pages 16 - 21 are Supervisor screens requiring a level 1 pass code (default = 1). Pages 22 – 28 are Configuration screens requiring a
Operator Level – No Pass Code Required
1. MAIN PAGE
2. IR STATUS (Availability depends on configuration)
3. GENERAL INFORMATION
4. PUMP CONTROL
5. SET DISPLAY VALUES
6.
[NOT CURRENTLY USED]
7. CALIBRATION DATES
8.
[NOT CURRENTLY USED] [NOT CURRENTLY USED]
9.
10. ABOUT / SIGN-ON
11. REVISION DISPLAY
12.
[NOT CURRENTLY USED]
13.
[NOT CURRENTLY USED]
14.
[NOT CURRENTLY USED] [NOT CURRENTLY USED]
15.
Supervisor Level – Level 1 Pass Code Required
16. SET THE DATE AND TIME
17. SAMPLING PARAMETERS
18. IR CELL ZERO CALIBRATION
19. 4 – 20 mA OUTPUT ASSIGNMENT
20. AUTO CALIBRATION SETUP
21. MAIN DISPLAY SETUP
Configuration Level – Level 2 Pass Code Required
22. COMMUNICATIONS SETUP
23. CALCULATION FACTORS
24. IR CELL SPAN CALIBRATION
25. SET PASS CODES
26. SET IP ADDRESS
27. H2 CELL CALIBRATION (Availability depends on configuration)
28. AUTO SEQUENCE SETUP
29. GAS or CV CONFIGURATION
SSi Manual #4563 Rev. D Page 8 Multi-Gas Model 6000
Level 2 pass code (default = 2). Page 29 requires the Super Systems Inc special passcode to access. The default pass codes can be changed by accessing the At the bottom of the Menu Screen is a status bar. This tells the current date and time, and also displays the internal temperature (IT) of the instrument. This internal temperature should never exceed 122°F (50°C).
Set Pass Codes
(menu option 25) menu.
Note about Menu Numbers
Each menu screen has a unique number that will be displayed in the upper left-hand corner of the screen. This number is shown for reference. If you know the menu number of the screen that you would like to go to, this number can be typed in to access it directly from the
Main Page
(Menu option 1) or the Menu List.
Main Page – Menu Page 1
The IR status display shows the current readings of the gases being sampled. Depending on the configuration of your instrument, this screen will show the values from one to four gases or calculations. To change how the gases are displayed, use the menu option (menu option 21). Also shown is the relative flow rate of the sample by a fuel gauge in the left-hand side of the screen. The “Pump = OFF” message in the lower left-hand side shows the status of the pump. When the pump is in auto mode, this message will read “Pump = AUTO”.
Main Display Setup
IR Status – Menu Page 2
Depending on the configuration of the instrument, this screen may not contain all of the information shown in the example. For one- and two-gas configurations, the “Calculated” values will not be shown, since there is not enough information available for the instrument to compute the percent carbon. When the instrument is configured with three gases (CO, CO2, and CH4), carbon percentage can be calculated by the instrument, The IR Status Display provides the user with the calculated carbon percentage (%C) from two different sources (probe and infrared). It provides information to allow the atmosphere controller to be “tuned” to match the information from the 3-gas analyzer.
To obtain the most information from this screen, data from the carbon probe must be entered. This can either be done manually or automatically via RS485 communications. This information is displayed at the right hand side of the screen under the heading
Operator
arrow keys to move the highlighted area up and down, the following data should be entered:
FC TC= The furnace thermocouple value, or the furnace temperature.
PB MV= The millivoltage from the carbon probe.
PB TC= The probe thermocouple value, or the probe temperature.
COF= The CO Factor value read from the SSi, Honeywell, Barber Colman, Yokogawa, or other
atmosphere controller.
PF= The Process Factor value read from the Marathon Sensors atmosphere controller.
. Using the keypad to enter numbers, and the
SSi Manual #4563 Rev. D Page 9 Multi-Gas Model 6000
Temperature Units = This determines the measurement units for temperature. Enter 0 for Fahrenheit (degrees F) or 1 for Celsius (degrees C).
Although it is possible to enter the data manually using the keypad, the atmosphere controller should be utilized to provide the data automatically if possible. When the data is entered automatically, it will change as the composition of the atmosphere changes. This will allow for a more reasonable correlation between the values from the MGA and the values from the probe. The probe information can also be used to ensure that the pump operates only when the conditions are right for sampling. This can be used to prevent damage to the instrument if the pump is left running as the furnace cools or when the conditions deviate from pre-determined parameters. For additional information on operating the pump in “Auto Sample” mode, please refer to the section titled option 17).
Measurement of Infrared % Carbon (IR %C) To accurately measure the % carbon in a furnace atmosphere, the instrument will need to know the values of CO, CO2, and CH4 and the temperature of the gas being measured. At the left side of the screen, under the heading the furnace temperature (FC TC) value, will result in the calculation of the IR % Carbon (IR %C). This is displayed in the center of the screen, under the heading temperature information has not been entered correctly the resulting carbon calculation For best results, it is recommended that the temperature information be entered automatically from either the atmosphere controller or the temperature controller.
Measurement of Probe % Carbon (PB %C)
The measurement of the probe % carbon requires three pieces of information to be entered. These are the probe millivolts (PB MV), probe temperature (PB TC), and either the COF or the PF (depending on the type of atmosphere controller you are using). The probe millivolts and probe temperatures are entered on the right side of the page, under the heading determine the manufacturer of the atmosphere control instrument you are using. If SSi, Honeywell, Barber Colman, Yokogawa, or anyone other than Marathon Sensors manufactures the instrument, then it will contain a CO Factor (COF) adjustment variable. If the user is using a Marathon Sensors instrument, then this variable is called a Process Factor (PF) adjustment. Only one of these (COF or PF) will be used at any one time, and the other will not be relevant. If the instrument has a CO Factor adjustment variable, then use the up or down arrow keys to go to the appropriate field and type in the value that is stored in your atmosphere control instrument. After the value is entered, an asterisk (*) will appear next to the number. This lets the user know that the CO Factor is being used for the calculations, not the Process Factor. If the instrument has a Process Factor adjustment variable, then this value should be entered in the “PF =” slot, which will result in the asterisk appearing by that number entered. The calculation of probe % carbon is no different from the % carbon as displayed on the atmosphere controller. The algorithm used by both instruments in their calculations is identical. The reason the information is entered into the MGA is not to calculate the probe % carbon, but to enable the MGA to compute the suggested COF / PF. Without knowing the current instrument values, it cannot compute the suggested values.
What is a CO Factor or a Process Factor?
The carbon probe is measuring the amount of oxygen in the atmosphere. Knowing the amount of oxygen, the atmosphere controller can determine the percentage of carbon. The calculation that the instrument uses to translate oxygen concentration into percent carbon is based on a theoretically pure atmosphere being present in the furnace. The composition of this theoretically pure atmosphere is 40% hydrogen (H2), 40%
Measured
, are the real-time values of CO, CO2, and CH4. The values of these three gases, plus
Operator
Calculated
. Before entering either a COF or a PF, you will need to
. Please note that if the furnace
Sampling Parameters
will not
be accurate.
(menu
SSi Manual #4563 Rev. D Page 10 Multi-Gas Model 6000
nitrogen (N2), and 20% carbon monoxide (CO). In many situations, the measured amount of CO is less than the theoretically exact 20%. This can be due to a variety of factors including seasonal changes in natural gas composition and incomplete gas cracking in an endothermic generator. The CO Factor and Process Factor adjustments are intended to make adjustments to the calculation to accommodate differences between theoretical and actual gas compositions.
Suggested COF / PF
Between the two methods of determining % carbon (probe and infrared), the infrared is considered to be more accurate. This is because the infrared analyzer evaluates the levels of three gases (CO, CO2, and CH4) to make its calculation, instead of just using % oxygen like the probe does. At the bottom of the middle display column, making a comparison between where the probe is actually reading and where it should be reading. The suggested COF and PF values can be entered into your atmosphere controller to make it display the same % carbon reading as the MGA6000. By performing periodic evaluations with the MGA6000 and making the suggested modifications to the adjustment factor in the atmosphere controller, the user can be assured that the continuous source of process data (the probe) is as accurate as possible. Of course, large changes in CO Factor or Process Factor should be verified by shim stock analysis or other means to confirm the significance of the change.
Calculated
, you will see suggested COF and PF values. These values are determined by
Example – The following will show how to set the Operator variables for: furnace temperature of 1500 degrees, probe millivolts of 1000, probe temperature of 1500 degrees, a CO factor of 250, and the temperature units are Fahrenheit. When the menu screen first displays, the furnace temperature field is already highlighted. Enter a 1500 and press Enter. Press the down arrow once to highlight the probe millivolts field. Enter a 1000 and press Enter. Press the down arrow key once to highlight the probe temperature field. Enter a 1500 and press Enter. Press the down arrow key once to highlight the CO Factor field. Enter a 250 and press Enter (Note – there will be an asterisk * next to the value entered). Press the down arrow key twice to highlight the temperature units field. Enter a 0 and press Enter (Note – the screen will display degrees F).
General Information – Menu Page 3
The General Information screen displays some general information. This screen is for display purposes only, so no information can be entered on this screen. There are two digital inputs that can be energized to either stop the pump or to temporarily inhibit the adjustment of COF/PF. These can be useful for making sure that the instrument does not operate normally when unusual temporary conditions exist (i.e. probe burnoff, door opening, etc.). Connections for these inputs can be made at the terminal block (See
Input #1 is for stopping the pump. Pump operation will be stopped if the contact between Terminal 1411 and 1412 is closed. Digital Input #2 is for inhibiting the adjustment of COF/PF. The adjustment will be stopped if the contact between Terminal 1421 and 1422 is closed.
Electrical Connections
section). Digital
SSi Manual #4563 Rev. D Page 11 Multi-Gas Model 6000
Pump Control - Menu Page 4
The pump control page is used to set the pump mode. The sample pump mode can be changed from OFF to AUTO
SAMPLE by pressing the circular arrow key. AUTO SAMPLE mode will automatically turn the pump on and off
based on the conditions described in the
Parameters
menu (menu option 17).
screen is the only location to change the mode of the pump
Also shown is the relative flow rate of the sample by a fuel gauge in the left-hand side of the screen. There is also a traditional flow meter located in the door of the MGA. Although the flow indicator on the screen has been calibrated at the factory, the most accurate flow measurements should be taken with the flow meter on the front of the analyzer.
Sampling
Note: This menu
.
Example – the following will show how to change the pump status to Auto Sample. When the screen first displays, the pump status field is already highlighted, displaying the word “OFF”. Press the circular arrow key to change the pump status to Auto Sample.
Set Display Values – Menu Page 5
The Set Display Values screen is used to adjust the display values. The brightness and contrast values are factory set to 51%, which should be appropriate for most viewing conditions. The values can range from 0 to 100(%) if desired. Also shown on this page is the Back light ON Time. This is a power saving feature that will enhance battery life by turning the backlight off if no buttons are pressed within a pre-determined number of minutes. When it turns off, the instrument still fully operates, but the display is more difficult to read. To restore the backlight, press any key. The backlight is factory set to 15 minutes. To disable this function, enter a value of 0 in this field and the backlight will always be on. The maximum
number of minutes allowed is 240.
percentage, while the back light ON time is displayed and entered in number of minutes
The MGA 6000 comes equipped with an internal cooler that can control the internal temperature of the instrument. The Cooler ON temperature and the Cooler OFF temperature can contain values ranging from 0 to 122 degrees when degrees Fahrenheit is selected, or 0 to 52 degrees, when degrees Celsius is selected. The Cooler ON and Cooler OFF temperatures are not updated if the temperature mode is switched from Fahrenheit to Celsius or vice-versa. Therefore, the user will have to manually change the temperatures to reflect the new temperature mode. The Cooler action line will display one of three messages:
,
Cooler Always OFF
ON
and the Cooler OFF temperature. “Hys” is an abbreviation for hysteresis, which is the difference between the on and off temperatures. If the ON temperature is less than or equal to the OFF temperature, then the cooler will always be on and the Cooler action label will read greater than the OFF temperature, then the cooler will turn on when the internal temperature rises above the Cooler ON temperature, and the cooler will turn off when the internal temperature falls below the Cooler OFF temperature. If the ON temperature is set above 120 degrees Fahrenheit, or 50 degrees Celsius, then the cooler will always be off and the Cooler action line will read
, or
Control Hys xx
Note – the brightness and contrast levels are displayed and entered as a
.
Cooler Always
where xx is the difference between the Cooler ON temperature
Cooler Always ON
Cooler Always Off
. If the ON temperature is
.
Note – If the OFF
SSi Manual #4563 Rev. D Page 12 Multi-Gas Model 6000
temperature is higher than the ON temperature (Cooler Always ON), but the ON temperature is above 120 (or 50) degrees (Cooler Always OFF), then the cooler will always be off
between settings, the cooler will always be off. To reduce the frequency of heating and cooling cycles, it is recommended that at least a 10° difference is maintained between the On and Off temperature set points.
. Whenever there is a conflict
Example – The following will show how to set the brightness to 60%, contrast to 40%, the back light ON time to 20 minutes, the Cooler ON temperature to 75 degrees (assume °F), and the Cooler OFF temperature to 50 degrees (assume °F). When the menu screen first displays, the brightness level field is already highlighted. Enter a 60 and press Enter. Press the down arrow key once to highlight the contrast field. Enter a 40 and press Enter. Press the down arrow key once to highlight the back light ON time field. Enter a 20 (for the back light ON time) and press Enter
.
Press the down arrow key once to highlight the Cooler ON temperature field. Enter a 75 and press Enter. Press the down arrow key once to highlight the Cooler OFF temperature field. Enter a 50 and press Enter. The Cooler action field should read “Control Hys 25”.
Note – Any changes made to the brightness or contrast level will take effect immediately.
Calibration Dates and Run Times - Menu Page 7
This page shows the most recent calibration dates, as well as the amount of time that has elapsed since each calibration. Time is shown in days (d), hours (h), and minutes (m). These dates do not need to be set after calibration since they will be set automatically whenever a calibration is performed.
entered, the internal clock must be set correctly
internal clock, use the menu option (menu option 16).
NOTE: For accurate calibration dates to be
. To set the
Set the Date and Time
About/Sign-On – Menu Page 10
This page is the sign-on screen that shows the SSi logo and phone number. Also shown is the unit serial number, the date of the last factory calibration, and the number of days, hours, and minutes that the instrument has been in use since the last factory calibration.
Revision Display – Menu Page 11
This page shows the firmware revision levels for the instrument. Although the user cannot modify this information, it may be helpful to report when contacting Super Systems for support.
SSi Manual #4563 Rev. D Page 13 Multi-Gas Model 6000
Loading...
+ 29 hidden pages