Getting Started .............................................................................................................................................. 4
Modifying the Default Settings ...................................................................................................................... 5
Description of Menu Items ............................................................................................................................ 5
Parts List and Internal Components ............................................................................................................. 7
Revision History .......................................................................................................................................... 10
Appendix A: Drawings (Electrical, Plumbing, and Mounting) ...................................................................... 11
Exit Program ................................................................................................................................................... 5
Instrument Information .................................................................................................................................. 5
Communication Setup ..................................................................................................................................... 5
Performing a Zero Calibration ................................................................................................................... 7
Performing a Span Calibration .................................................................................................................. 7
Super Systems Inc. Page 3 of 12
Page 4
Basic Nitriding Sampling System – Hydrogen Analyzer
Introduction
Specifications
Electrical Connections / Terminal Assignments
Wire Number
Function
1000
AC Line Power (100-240VAC)
1002
AC Neutral Power
Ground
AC Ground
1121
Analog Output Common (+)
1111
Analog Output #1 (-)
1121
Analog Output Common (+)
1131
Analog Output #2 (-)
1081
RS485 (-)
1091
RS485 (+)
Getting Started
This instrument uses the measurement of Hydrogen to display % Hydrogen (H2), % Dissociation (DA), and
% Ammonia (NH
entered, the instrument can also calculate Nitriding Potential (K
). When the flow rates of Nitrogen, Ammonia, and Dissociated Ammonia are manually
3
).
N
• Power Requirements: 100-240 VAC
• Current Draw: Max. 0.2 Amps
• Sensor Technology: Thermal Conductivity
• User Interface: 3.5” Color QVGA TFT LCD Touch Screen
• Measurement Range: 0-100% H
• Hydrogen Measurement Accuracy: +/- 0.01%
• Hydrogen Measurement Repeatability: +/- 0.01%
• Hydrogen Measurement Resolution: 0.01%
• Analog Outputs: Two Isolated 4-20mA (User Configurable)
• Analog Output Resolution: 0.005mA
• Analog Output Accuracy: +/- 0.01% of Range
• Analog Output Linearity: +/- 0.01%
• Analog Output Load Resistance: Minimum 0 Ohm, Maximum 500 Ohm
• Digital Communications: RS485 Modbus, Ethernet
• Enclosure Size (Without Filter): Approx. 13”L x 13”W x 6”D / 330mmL x 330mmW x 142mmD
Please refer to the enclosed drawing for instructions regarding the proper electrical and mechanical
installation of this instrument. The flow meter on the door of the enclosure should be adjusted to
maintain 1.5 SCFH of process gas flow.
Super Systems Inc. Page 4 of 12
Page 5
Basic Nitriding Sampling System – Hydrogen Analyzer
Default Settings
Modifying the Default Settings
Description of Menu Items
Exit Program
Instrument Information
Communication Setup
Instrument Configuration
When the instrument is turned on, it will display % H2 on the screen. To display additional values (% DA,
, or KN), see the “Instrument Setup” screen. The first 4-20mA output will be set up to retransmit
% NH
3
the %H
Any changes to these default parameters will be stored so they will not need to be re-entered when the
power is shut off to the instrument
value scaled for 0 to 100%, and the second output will be set up for %DA also scaled for 0-100%.
2
To access the operational and setup parameters, press the “menu” button at the lower left section of the
screen. This will allow you to select only two options, Exit Program and Instrument Information. To
prevent unwanted modification to the operation of the instrument, these are the only options available
unless the user logs in. To access the other menu options, press the “Login” button and enter “2”. This
will give the operator access to the setup and operational parameters.
Each individual menu option is described in detail below:
The touch screen display is constantly writing data to its flash card for storage, and it is important that
the instrument not be shut down during this process. Before removing power from the enclosure, select
“Exit Program”. It will ask is you are sure that you want to shut down the interface. By answering yes, it
closes the operating program in an orderly manner. Once the standard Microsoft Windows screen
appears it is safe to remove power from the instrument. If this procedure is not followed, there is a
chance that there could be an error writing to the flash card that could cause problems with the operation
of the instrument.
This screen provides information on any applicable revision levels and serial numbers. It also shows if
the instrument is logging data. There are no functions that can be performed on this screen; it is for
informational purposes only.
The communication methods shown on this screen are for display only and cannot be modified. The baud
rates can be adjusted but they have been optimized for this instrument and modification is not
recommended.
This instrument is capable of displaying four different parameters. These include:
• Percent Hydrogen (H
• Percent Dissociation (DA)
• Percent Ammonia (NH
• Nitriding Potential (K
The Percent Hydrogen is displayed on the main screen at all times. One additional parameter can be
displayed along with Hydrogen by highlighting it and pressing the “OK” button. A clearer description of
each of these parameters can be found in Appendix “A”, Gas Nitriding Technical paper.
)
2
)
3
)
N
Percent Hydrogen
The percent hydrogen is the amount of Hydrogen that is being detected by the thermal conductivity
sensor inside the instrument. There are no additional calculations being performed to this value.
Super Systems Inc. Page 5 of 12
Page 6
Basic Nitriding Sampling System – Hydrogen Analyzer
Output Configuration
Output Calibration
Sensor Calibration
Percent Dissociation
Dissociation is derived from the amount of Hydrogen in the sample gas.
Percent Ammonia
The amount of Ammonia can also be inferred from the Hydrogen value.
Nitriding Potential
The accurate calculation of Nitriding Potential (KN) requires the flow rates of other gases that are being
introduced into the process. These flow rates are measured in SCFH (Standard Cubic Feet per Hour). K
can only be calculated by entering the flow of Nitrogen (N
(%DA). These values are entered at the bottom of the main screen when K
), Ammonia (NH3) and % Dissociated Ammonia
2
is selected.
N
N
There are two 4-20mA outputs that can be configured for any of the four parameters. For each input, the
operator can select the Source (H
, DA, NH3, KN), the zero value (the value to be represented by 4mA) and
2
the span value (the value to be represented by 20mA).
• Output 1 can be measured from Terminal #1121(+) and 1111 (-).
• Output 2 can be measured from Terminal #1121(+) and 1131 (-).
Accurate retransmission of the selected parameters can only be obtained through calibrating both of the
analog outputs. This is done at the factory prior to shipment, however it is a relatively simple procedure
that can be performed in the field if desired. To perform a calibration, a multimeter with a current input
is required. Please keep in mind that the accuracy of the instrument used to calibrate the outputs is
directly related to the accuracy of the outputs after calibration, so a meter calibrated against NIST
(National Institute of Standards and Technology) standards is preferred. Before performing any
calibrations, all other devices must be disconnected from the analog outputs. Multiple devices on the
outputs will cause inaccurate measurement of current.
Attach the leads of the multimeter to the terminals for Output 1. The positive lead should be attached to
Terminal #1121 and the negative lead to Terminal 1111. Select “Zero Output 1” and press the “Prep for
Cal” button. The low limit of the output is 4mA, so 4.000 is shown as the default measured value. Ideally
the meter connected to the outputs will also show 4.000. If the two values are not close enough to obtain
the desired level of accuracy, a calibration should be performed.
If, for example, the meter connected to the outputs reads 4.216mA, then the value “4.216” should be
entered on the screen as the Measured Value. Once the value has been entered, press the “calibrate”
button. This will offset the mA output of the instrument in an amount that results in an exact output of
4.000mA. When the calibration is complete, the multimeter should be reading 4.000mA (+/- .005).
The procedure can be repeated for the Span of Output 1. When “Span Output 1” is selected and “Prep for
Cal” is pressed, the instrument will output the high output limit, which is 20mA. The multimeter will
display the actual output from the instrument, and if it is not within the desired tolerance it can be
calibrated using the same procedure as above.
To calibrate Output 2, attach the leads of the multimeter to Terminal #1121 (+) and the negative lead to
Terminal 1131 (-) and follow the same procedure that was used for the first output.
A proper calibration of the sensor requires two gases. The first gas should be pure Nitrogen or Argon.
This contains no Hydrogen, and is therefore referred to as the Zero Gas. The second gas is the Span Gas.
The Span Gas should ideally contain a quantity of Hydrogen similar to the amount of Hydrogen in the
process gas. The Span Gas should also include any other gases that are present in the process gas in
Super Systems Inc. Page 6 of 12
Page 7
Basic Nitriding Sampling System – Hydrogen Analyzer
Parts List and Internal Components
Part Number
Description
Full Unit
their respective percentages. The more similar to the process gas the calibration gas is, the more
accurate the calibration will be.
Performing a Zero Calibration
On the Sensor Calibration page, select “Zero Hydrogen”. Turn the valve on the side of the enclosure to
the “Calibration Gas” position, and attach the Zero gas to the “Calibration Gas Inlet” port. Begin the flow
of gas at a rate of 1.5 to 2.0 SCFH as measured on the flow meter on the door of the enclosure. The gas
should not be under any pressure other than the amount required to maintain the appropriate flow
amount. The target Value is shown on the screen. For a Zero Calibration, this will be 0.00 (the amount of
Hydrogen in the Zero Gas). The Measured H
value comes to equilibrium, it will not be showing any upward or downward trends, only the slight
oscillation of the readings. This usually occurs in approximately 30 seconds. When the sensor is at
equilibrium, press the green “Calibrate” button to perform the zero calibration. After the Zero
Calibration is complete, turn off the flow of gas and disconnect it from the enclosure.
Value can be seen at the bottom of the screen. When this
2
Performing a Span Calibration
To perform a span calibration, select “Span Hydrogen”, attach the Span Gas to the Calibration Gas Inlet
port, and begin the flow of gas at 1.5 to 2.0 SCFH. The Target Value should be set to the exact amount of
Hydrogen that is in the Span Gas cylinder. Then the same procedure should be followed as the Zero
calibration, with the “Calibrate” button being pressed after the readings reach equilibrium. After the
Span Gas calibration is complete, turn off the flow of gas, disconnect the cylinder from the enclosure, and
restore the valve on the side of the instrument to the “Sample Gas” position. This will re-connect the
sensor to the process gas stream.
The following items can be purchased as needed for the Hydrogen Nitrider Analyzer.
31604 2 GB SD card
37026 Fan (24VDC, 60mm)
37029 Fan filter assembly
37150 3-way ball valve
37050 High capacity bowl filter
37051 High capacity filter element
34444 Tubing
33097 Circuit breaker
31135 Power supply
A20624 Hydrogen sensor
31296 3.5” touch screen
15357 Hydrogen Nitrider Analyzer
Super Systems Inc. Page 7 of 12
Page 8
Basic Nitriding Sampling System – Hydrogen Analyzer
The following diagram illustrates the location of important internal components of the Hydrogen Nitrider
Analyzer, along with relevant part numbers.
Super Systems Inc. Page 8 of 12
Page 9
Basic Nitriding Sampling System – Hydrogen Analyzer
Warranty
Limited Warranty for Super Systems Products:
The Limited Warranty applies to new Super Systems Inc. (SSI) products purchased direct from SSI or
from an authorized SSI dealer by the original purchaser for normal use. SSI warrants that a covered
product is free from defects in materials and workmanship, with the exceptions stated below.
The limited warranty does not cover damage resulting from commercial use, misuse, accident,
modification or alteration to hardware or software, tampering, unsuitable physical or operating
environment beyond product specifications, improper maintenance, or failure caused by a product for
which SSI is not responsible. There is no warranty of uninterrupted or error-free operation. There is no
warranty for loss of data—you must regularly back up the data stored on your product to a separate
storage product. There is no warranty for product with removed or altered identification labels. SSI DOES
NOT PROVIDE ANY OTHER WARRANTIES OF ANY KIND, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
WARRANTIES OR CONDITIONS OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
SOME JURISDICTIONS DO NOT ALLOW THE LIMITATION OF IMPLIED WARRANTIES, SO THIS LIMITATION
MAY NOT APPLY TO YOU. SSI is not responsible for returning to you product which is not covered by this
limited warranty.
If you are having trouble with a product, before seeking limited warranty service, first follow the
troubleshooting procedures that SSI or your authorized SSI dealer provides.
SSI will replace the PRODUCT with a functionally equivalent replacement product, transportation prepaid
after PRODUCT has been returned to SSI for testing and evaluation. SSI may replace your product with a
product that was previously used, repaired and tested to meet SSI specifications. You receive title to the
replaced product at delivery to carrier at SSI shipping point. You are responsible for importation of the
replaced product, if applicable. SSI will not return the original product to you; therefore, you are
responsible for moving data to another media before returning to SSI, if applicable. Data Recovery is not
covered under this warranty and is not part of the warranty returns process. SSI warrants that the
replaced products are covered for the remainder of the original product warranty or 90 days, whichever is
greater.
Super Systems Inc. Page 9 of 12
Page 10
Revision History
Rev.
Description
Date
MCO#
NEW
First release
N/A
A
Revised terminal numbers; added product diagrams;
added part numbers; updated format of document.
11/7/2014
2155
Basic Nitriding Sampling System – Hydrogen Analyzer
Super Systems Inc. Page 10 of 12
Page 11
Basic Nitriding Sampling System – Hydrogen Analyzer
Appendix A: Drawings (Electrical, Plumbing, and Mounting)
Super Systems Inc. Page 11 of 12
Page 12
Basic Nitriding Sampling System – Hydrogen Analyzer
Super Systems Inc. Page 12 of 12
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.