Super Systems HP15 User Manual

Operations Manual
HP15
Self-Heated Oxygen
Measurement and Control
Please read, understand, and follow these instructions before operating this equipment.
Super Systems, Inc. is not responsible for damages incurred due to a failure to comply with
these instructions. If at any time there are questions regarding the proper use of this product,
please contact us at (800) 666-4330 for assistance.
7205 Edington Drive
Cincinnati, OH 45249
513-772-0060 / 800-666-4330
Fax: 513-772-9466
www.supersystems.com
HP 15 Manual Rev A Page 1 of 1
Table of Contents
Table of Contents................................................................................................................. 2
General Information .............................................................................................................3
Electrical Connections........................................................................................................... 4
Plumbing Connections .......................................................................................................... 7
HP 15 Screens ..................................................................................................................... 8
Initial Screen.................................................................................................................... 8
Main Menu .......................................................................................................................8
Set Points..................................................................................................................... 8
Temp Mode .................................................................................................................. 9
CO Factor ..................................................................................................................... 9
Hydrogen Factor ........................................................................................................... 9
Pump Control ............................................................................................................. 10
Auto / Manual............................................................................................................. 10
O2 View ..................................................................................................................... 10
Configuration Menu ........................................................................................................ 10
PVT Type ................................................................................................................... 10
Network Settings ........................................................................................................ 11
PID Settings ............................................................................................................... 11
Loop PID Settings – Page 1 ..................................................................................... 11
Loop PID Settings – Page 2 ..................................................................................... 12
Loop PID Settings – Page 3 ..................................................................................... 12
Loop PID Settings – Page 4 ..................................................................................... 12
Burn Off Setup............................................................................................................ 13
Analog Input Setup ..................................................................................................... 13
Analog Input Settings – Page 1 ................................................................................ 13
Analog Input Settings – Page 2 ................................................................................ 14
Analog Input Settings – Page 3 ................................................................................ 14
Analog Output Setup................................................................................................... 15
Analog Output Settings – Page 1.............................................................................. 15
Alarms ....................................................................................................................... 15
Alarm Settings - Page 1 ........................................................................................... 15
Alarm Settings – Page 2 .......................................................................................... 16
Alarm Settings – Page 3 .......................................................................................... 17
Over Temp Alarm........................................................................................................ 17
SSi Information Screen................................................................................................ 17
Spare Parts List.................................................................................................................. 18
Appendix 1 – Factory Defaults ............................................................................................ 19
Universal Defaults........................................................................................................... 19
Carbon PVT Defaults....................................................................................................... 20
Dewpoint PVT Defaults ................................................................................................... 20
Oxygen PVT Defaults ...................................................................................................... 20
Probe mV PVT Defaults ................................................................................................... 21
Probe Redundancy PVT Defaults...................................................................................... 21
Simple Nitrider PVT Defaults ........................................................................................... 22
Dual Loop PVT Defaults .................................................................................................. 22
Revision History ................................................................................................................. 23
HP 15 Manual Rev A Page 2 of 2
General Information
The HP15 system includes two main enclosures. The first contains the heater and the Gold Probe and is painted with a heat resistant black paint. The second is the blue electrical control panel. Proper operation of this system requires that these two enclosures be wired to each other to allow the heater to be controlled by the electrical enclosure. For detailed information regarding the connections between the enclosures, please refer to the E section of this manual.
lectrical Connections
HP 15 Manual Rev A Page 3 of 3
Electrical Connections
Insert the probe into the coupling at the base of the large box. It is important that the probe is inserted at the correct depth to allow proper heating and prevent damage to the ceramic well inside the unit. There should be between 10.0” and 10.5” between the
bottom of the adjustable 1” NPT gland and the end of the probe (see sketch below).
Once the probe has been inserted into the heater, make the necessary electrical connections at the head of the probe. The green thermocouple wire from the electrical enclosure box should be attached to the thermocouple connection at the head of the probe (Positive (+) = Black and Negative (-) = Red). The sensor wire should also be attached to the probe (Positive (+) = Black and Negative (-) = White).
This unit has been pre-configured and tested prior to shipment. To maximize the accuracy and longevity of the Heated Probe, it is recommended that the unit be operated at 1500°F. It is also possible to configure the HP to match the temperature of your furnace.
HP 15 Manual Rev A Page 4 of 4
Inside Electrical Enclosure
r
A
A
;
6.0 AMP
;
1000
;;
1002
;; ;;
1003
;; ;
1103
;
1601
;
1602
;
1621
;
1622
;
1681
;
1682
;
1060
;;
1070
;; ;
1320
;
1330
;
1340
;
1641
;
1002
;
GND
;;
;
; ; ; ; ; ; ; ;
; ; ; ; ; ;
;
6 Amp Circuit Breaker Incoming Customer 110VAC (Line) Incoming Customer 110VAC (Neutral) 110VAC (Neutral) To Heate
110VAC (Line) to Heater Enclosure
110VAC (Switched Line) to Heater Enclosure Enriching Gas Contact #1 Enriching Gas Contact #2 Dilution Air Contact #1 Dilution Air Contact #2 Alarm Contact #1 Alarm Contact #2 24VDC (Negative)
24VDC (Positive)
Dry Contact for Sample Inhibit Dry Contact for Probe Burn-Off Dry Contact Common Pump Terminal 110VAC (Neutral) - Spare Incoming Customer 110VAC (Ground)
;
1021
;
1031
;
1041
;
1350
;
1360
;
1380
Host Communications connect to the Host Computer.
nalog Output #1 is the 4-20mA signal for Process Variable Retransmission. nalog Output #2 is the 4-20mA signal for Temperature Control (Optional).
; ; ; ; ; ;
RS485 Host Communications Shield RS485 Host Communications RT- (Negative) RS485 Host Communications RT+ (Positive) Analog Output Common (Positive) Analog Output #1 (Negative) Analog Output #2 (Negative)
Inside Heater Enclosure
;
1003
;
1002
;
1103
;; ;; ;;
; ; ;
Incoming 110VAC (Line) from Electrical Enclosure Incoming 110VAC (Neutral) from Electrical Enclosure Incoming 110VAC (Switched Line) from Electrical Enclosure
HP 15 Manual Rev A Page 5 of 5
1. Connect incom ing 120VAC to the Electrical E nclosure: 110 VAC Line Terminal 1000
110 VAC N eutral Terminal 1002 110 VAC G round Terminal GN D
2. Connect Electrical Enclosure to Heater Enclosure:
3. Connect "S" Type Thermocouple wire from P robe to Model 9120 Controller in Electrical Enclosure
4. Connect Millivolt wire from P robe to M odel 9120 Controller in Electrical Enclosure
5.
(Optional)
6.
(Optional)
7.
(Optional)
Electrical Heater Terminal 1003 Terminal 1003 Terminal 1002 Terminal 1002 Terminal 1103 Terminal 1103
Probe M odel 9120
+ T'CPLE (Black) Terminal 30
- T'CP LE (R ed ) Term ina l 29
Probe M odel 9120
+ SENSOR Terminal 32
- SE NS OR Terminal 31
Connect Furnace Thermocouple to Model 9120 Controller in Electrical Enclosure
Probe M odel 9120
+ T'CPLE Term inal 28
- T'CPLE Terminal 27
Connect RS485 Communications to H ost Com puter
Electrical Host Computer Terminal 1021 Shield Terminal 1031 RT­Terminal 1041 RT+
Connect Enriching Gas Contact
Electrical (Custom er) Terminal 1601 T.B.D Terminal 1602 T.B.D
8.
(Optional)
9.
(Optional)
10.
(Optional)
11.
(Optional)
12.
(Optional)
13.
(Optional)
Connect Dilution A ir Contact
Electrical (Custom er) Terminal 1621 T.B.D Terminal 1622 T.B.D
Connect Dilution A ir Contact
Electrical (Custom er) Terminal 1681 T.B.D Terminal 1682 T.B.D
Connect Dry Contact for Sam ple Inhibit
Electrical (Custom er) Terminal 1320 T.B.D Terminal 1340 T.B.D
Connect Dry Contact for Probe Burnoff
Electrical (Custom er) Terminal 1330 T.B.D Terminal 1340 T.B.D
4-20m A O utput for Process Variable Retransm ission
Electrical (Custom er) Terminal 1350 T.B.D + Terminal 1360 T.B.D -
4-20m A O utput for Temperature Control
Electrical (Custom er) Terminal 1350 T.B.D + Terminal 1380 T.B.D -
HP 15 Manual Rev A Page 6 of 6
Plumbing Connections
The sample gas can be transported from the sample port on the furnace to the bottom of the external filter on the side of the electrical enclosure using silicone tubing. The sample flow should be between 4.0 and 5.0 SCFH. Silicone tubing should also connect the ports on the side of the electrical enclosure to the probe in the heater enclosure. Connect tubing from the fitting on the electrical enclosure marked “To Probe Reference” to the “Ref. Air” port on the probe. A connection should also be made between the “To Burnoff Port” fitting and the “B.O.” fitting on the probe. There is a barb fitting on the stainless steel well on the heated probe. This is the vent, which needs to be routed to an appropriately safe location.
HP 15 Manual Rev A Page 7 of 7
HP 15 Screens
Initial Screen
The first screen to be displayed when the HP is powered up is the initial screen, which will display a combination of: Oxygen, Carbon, millivolts, dew point, temperature, setpoint, percent output, relay status, and pump status depending on the selected process variable type. The first section below the Super Systems Inc logo will display the process variable type (PVT). The PVT can be: Oxygen, Carbon, Dewpoint, or millivolts. See the section information on how to change the PVT type. The second section (Millivolts:) will display the probe’s current millivolt value. The third section will display the probe’s current temperature, as well as the current setpoint. See the section change the setpoints. The last section will display the Loop 1 and Loop 2 output percentages. An A or an M will be displayed next to the output percentage to let the user know if Loop 1 or Loop 2 is in automatic (A) or manual (M) mode. The “Pump” area will let the user know if the pump status is on or off. The “Relays” area, which is directly above the Menu button, will display the energized relays. The number corresponds to the relay in the panel. If a number is displayed, then that specific relay in the panel is energized. If a number is not displayed, then that specific relay in the panel is not energized. The Menu button will display the main menu.
Configuration Menu
Set Points
PVT Type
for more information on how to
for more
Main Menu
The main menu displays eight buttons, from which the user will be able to set up the HP 15 instrument. The Exit button will return the user to the initial screen.
Set Points
Two setpoints are initially displayed on the screen: Loop 1 and Alarm. To change the temperature setpoint or the alarm setpoint, touch the setpoint value, which will display the numeric keypad.
Enter the new setpoint value and press the
ENT button to set the setpoint. Press the ESC button to cancel the change.
The Remote Setpoint Setup button will display the Remote Setpoint screen, which will allow the user to set up a remote setpoint for Loop 2, which is generally the temperature loop. There are six buttons to choose from for a remote setpoint: None, Slave 1 PV, Slave 2 PV, Slave 1 SP, Slave 2 SP, or Input 3 Value. Pressing the desired button will set the setpoint to the specified source. When None is pressed, the setpoint that was set from the previous screen
HP 15 Manual Rev A Page 8 of 8
will be used. To change the hysteresis, click on the Hysteresis button and enter the new hysteresis value from the numeric keypad. The More >>> button will allow the user to set the setpoint for the selected PVT (Dew point, Carbon, Oxygen, or millivolts). The <<< More button will return the user to the previous screen. The Menu button will return the user to the main menu.
Temp Mode
The temperature mode screen displays the current temperature mode (Fahrenheit or Celsius) as well as the current setpoint. To change the mode to Celsius, press the Switch to Celsius button. To change the mode to Fahrenheit, press the Switch to Fahrenheit button. When the temperature mode is changed, the setpoint will be converted
automatically. This is a safety precaution to prevent setting extremely high setpoints when switching to the Celsius scale. The setpoint can always be changed from the
Set Points
menu.
CO Factor
The COF factor allows you to make adjustments that will allow the controller to match the results obtained when measuring shim stock. If the controller set point and the process variable are the same, but you are not attaining the desired surface Carbon, you can make the adjustment with the CO Factor to increase the amount of surface Carbon available to the parts. If you lower the CO factor, it immediately lowers the % Carbon process variable. This will cause the controller to add more enriching gas--- raising the process variable until once again the process variable and the set point match. Conversely, raising the CO Factor will cause the process variable to read
higher, shutting down the enriching gas solenoid (possibly turning on the air dilution solenoid) causing the % Carbon process variable to begin to lower (NOT ADDING ENRICHING GAS) until the set point and process variable match. To change the CO Factor, enter the new factor and press the ENT button. Pressing the ESC button will not set the new CO Factor.
just for display purposes. The CO Factor will not be set to 0.
Note: when the ESC button is pressed, the display will show a 0; however, this is
Hydrogen Factor
The Hydrogen factor is very similar to the CO factor, as explained above. The instrument will perform its calculations based on a theoretical amount of carbon or hydrogen. However, the actual amounts of carbon or hydrogen will probably be different from the theoretical amounts that the instrument is using to perform the calculations. Changing the CO factor or the Hydrogen factor will allow the instrument to make its calculations based on the actual amount of carbon or hydrogen, which will make the calculations more accurate. This screen has the same functionality as the CO Factor screen.
HP 15 Manual Rev A Page 9 of 9
Pump Control
This screen will allow the user to manually turn the pump on or off. Press the On button to turn the pump on. Press the Off button to turn the pump off. The “Low Temperature Pump Cutoff” section will allow the user to set a low temperature that will shut the pump off if the specified temperature is reached. Press the value and enter the desired pump cutoff setpoint.
Auto / Manual
The “Loop 1 Out:” section displays the output percentage of Loop 1, along with the current mode of the loop – Auto for automatic, Man for manual. To switch the mode of the loop to automatic, press the Automatic Control button. To switch the mode of the loop to manual, press the Manual Control button. When the loop is in manual mode, the output percentage can be changed by pressing the output percentage value and entering the new value from the numeric keypad that is displayed. The “Loop 2 Out:” section displays the output percentage of Loop 2, along with the current mode of the loop – Auto for automatic, Man for manual. To switch the mode of the loop to automatic, press the Automatic Control
button. To switch the mode of the loop to manual, press the Manual Control button. When the loop is in manual mode, the output percentage can be changed by pressing the output percentage value and entering the new value from the numeric keypad that is displayed.
O2 View
This feature is not yet implemented.
Configuration Menu
The Config Menu button will display the configuration menu. The Configuration menu displays six buttons that will allow the user to configure the HP15. The Exit button will return the user to the main
menu.
PVT Type
This screen will allow the user to set the PVT type. The currently selected PVT type will be listed at the top of the screen. Press the Carbon button to set the PVT type to Carbon. Press the Dew Point button to set the PVT type to Dew Point. Press the Oxygen button to set the PVT type to Oxygen. Press the Millivolts button to set the PVT type to Millivolts. Press the Simple Nitrider button to set the PVT type to Simple Nitriding. Press the Dual Loop button to set the PVT type to Dual Loop. Since changing the PVT type will set
HP 15 Manual Rev A Page 10 of 10
default values for that type, certain settings may be lost. When the PVT type is changed, the user will have to confirm the changes. Pressing the Yes button will change the PVT type, and pressing the No button will not change the PVT type. The Menu button will return the user to the configuration menu.
Network Settings
The Network Settings screen will allow the user to set up the network addresses for the HP15. Each screen – IP Address, Netmask, and Gateway – will function identically. The IP Address button will allow the user to
change the IP address of the device. The Netmask button will allow the user to change the Netmask of the device. The Gateway button will allow the user to change the gateway of the device. To change a specific section of an address, press the value and enter the new value on the numeric keypad.
PID Settings
This menu screen will allow the user to select which loop to change the PID settings for. Press the Loop 1 PID Settings button to change the PID settings for Loop1, or press the Loop 2 PID Settings button to change the PID settings for Loop 2. The Exit button will return the user to the configuration menu screen. The screens for the Loop 1 PID settings and Loop 2 PID settings function identically.
Loop PID Settings – Page 1
From this screen, the user can modify the proportional band, reset, rate, and cycle time. To edit any of these values, press the value and enter the new value from the numeric keypad that is displayed. The More >>> button will display Page 2 of the PID Settings menu. The Menu button will display the
PID Settings
menu.
HP 15 Manual Rev A Page 11 of 11
Loop PID Settings – Page 2
From this screen, the user can modify the control loop output, integral preset, control low limit, control high limit, and the loop mode. To edit the control loop output, integral preset, control low limit, or the control high limit, press the value and enter the new value from the numeric keypad that is displayed. The currently selected loop mode will be displayed in the box. To change the loop mode, press the up arrow next to the displayed mode. This will display the available modes in the form of buttons:
Dual Reverse (Gas/Air, Heat/Cool)
Single Reverse (Heat)
Dual Direct (Dew Point)
Dual Reverse (Cool)
Press the button for the desired mode. The <More
button will display Page 1 of the PID Settings menu. The
More> button will display Page 3 of the PID Settings
menu. The Menu button will display the
PID Settings
menu.
Loop PID Settings – Page 3
From this screen, the user can modify the auto PID setting, as well as whether
or not the high limit shuts down the control for each input. To turn the auto
PID on, press the On button. To turn the auto PID off, press the Off button.
The current status of the auto PID will be displayed. To modify whether the
high limit shuts down the control for Input 1, 2, or 3, press the Toggle button
for the desired input. For example, to have the high limit shut down control
for Input 1, press the top Toggle button; to have the high limit shut down
control for Input 3, press the bottom Toggle button. The <More button will
display Page 2 of the PID Settings menu. The More> button will display Page
4 of the PID Settings menu. The Menu button will display the
menu.
Loop PID Settings – Page 4
From this screen, the user can modify the zero setpoint shuts down control value, as well as the set point change limit value. Whether or not a zero setpoint shuts down control will be displayed above the Yes/No buttons.
To allow a zero setpoint to shut down control, press the Yes button. To not allow a zero setpoint to shut down control, press the No button. The current setpoint change limit is displayed in the box. To change the setpoint change limit, press the up arrow next to the displayed value. This will display eight buttons that will allow the user to change the setpoint change limit: Off
80 % 70 %
PID Settings
HP 15 Manual Rev A Page 12 of 12
60 % 50 % 40 % 30 % 20 %
Press the button for the desired setpoint change limit. The <<< More button will display Page 3 of the PID Settings menu. The Menu button will display the
PID Settings
menu.
Burn Off Setup
From this screen, the user can modify the burn off time in seconds, recovery wait time in seconds, interval in minutes, minimum millivolts, and
the maximum temperature values. The user can also start a manual burnoff from this screen. To modify the burn off time, recovery wait time, interval, minimum millivolts, or the maximum temperature value, press the value and enter the new value from the numeric keypad that is displayed. To start a manual burnoff, press the Start Manual Burn Off button. While a burnoff is in progress, the burnoff in progress screen will be displayed. The current sequence number and the time remaining in seconds will be displayed. The Close button will close down the screen.
Note: While the burnoff in progress screen may be closed, the burnoff will still be in progress, and another manual burnoff may not be started until the current burnoff is finished
.
Analog Input Setup
This menu screen will allow the user to set up the analog input settings for Input 1, 2, or 3. To change the analog inputs settings for Input 1, press the Analog Input 1 Settings button. To change the analog inputs settings for Input 2, press the Analog Input 2 Settings button. To change the analog inputs settings for Input 3, press the Analog Input 3 Settings button. The screens for the analog input settings 1, 2, and 3
function identically. The Exit button will return the user to the configuration menu.
Analog Input Settings – Page 1
From this screen, the user can modify the input type, filter, initial scale, full scale, and decimal point location. The current input type is displayed in the box. To change the input type, press the down arrow next to the displayed value. This action will display twenty buttons that will allow the user to change the input type:
Type B Type C Type E Type J Type K Type N
HP 15 Manual Rev A Page 13 of 13
Type NNM Type R Type S Type T
2.5 Volts
1.25 Volts
78.125 mV
19.53125 mV 4-20 mA 25 Volts
12.5 Volts
781.25 mV
195.3125 mV Close Screen
Press the button for the desired input type. To change the filter, initial scale, full scale, or decimal point location, press the value and enter the new value from the numeric keypad that is displayed. The More >>> button will display Page 2 of the analog input settings menu. The Menu button will display the
Analog Input Setup
menu.
Analog Input Settings – Page 2
From this screen, the user can modify the open input direction, trip 1 value, and the trip 2 value. The current open input direction value is displayed in the box. To modify the open input direction, press the down arrow next to the displayed value. This will display four buttons that will allow the user to change the open input direction: Input X Goes Up Scale
Input X Goes Down Scale One Trip Point Two Trip Points
Press the button for the desired open input direction. To set the Trip 1 value to Above Setpoint, press the Above SP button under the Trip 1 value. To set the Trip 2 value to Above Setpoint, press the Above SP button under the Trip 2 value. To set
the Trip 1 value to Below Setpoint, press the Below SP button under the Trip 1 value. To set the Trip 2 value to Below Setpoint, press the Below SP button under the Trip 2 value. Pressing the <More button will display Page 1 of the analog inputs settings menu. Pressing the More> button will display Page 3 of the analog input settings menu. The Menu button will display the
Analog Input Setup
menu.
Analog Input Settings – Page 3
From this screen, the user can modify the offset, trip point 1 set point, trip point 1 force value, trip point 2 setpoint, and the trip point 2 force value. To edit any of these values, press the value and enter the new value from the numeric keypad that is displayed. Pressing the <<< More button will display Page 2 of the analog input settings menu. The Menu button will display the
Analog Input Setup
menu.
HP 15 Manual Rev A Page 14 of 14
Analog Output Setup
This screen will allow the user to set up the analog output settings for output
1 or 2. To change the analog output settings for output 1, press the Analog
Output 1 Settings button. To change the analog output settings for output
2, press the Analog Output 2 Settings button. The screens for the analog
output settings function identically for outputs 1 and 2. The Menu button will
return the user to the configuration menu.
Analog Output Settings – Page 1
From this screen, the user can modify the analog output
assignment, the exponent, and the full-scale values. To
select a new analog output assignment, press the down
arrow next to the assignment. This will display a screen with 18 buttons to choose from:
PV 1 retrans Loop 1 inc Loop1 dec Loop 1 combo PV 2 retrans Loop 2 inc Loop 2 dec Loop 2 combo Disassociation Nit_Pot Hydrogen
- (Invalid Option) Input 1 retrans Loop 2 retrans Loop 3 retrans O2 offset log PV 1 retrans w/ expo range
The Close Screen button will close the screen without selecting a new assignment. Press the new assignment button to select it. To modify either the exponent or the full scale, press the value and enter the new value from the numeric keypad that is displayed. The Back button will display the analog output setup menu screen.
Alarms
This screen will allow the user to set up Alarm 1, Alarm 2, or Alarm 3. To set up Alarm 1, press the Alarm 1 Settings button. To set up Alarm 2, press the Alarm 2 Settings button. To set up Alarm 3, press the Alarm 3 Settings button. The screens for the alarms will function identically for Alarms 1, 2, and 3.
The over temperature alarm will now read “Overtemp Alarm 1”
Note: Alarm 1 is a dedicated over temperature alarm.
.
Alarm Settings - Page 1
This screen will allow the user to set the alarm source, alarm action, and the alarm set point. To set the alarm source, press the up arrow next to
HP 15 Manual Rev A Page 15 of 15
the source. This will display ten buttons from which the user can select the alarm source:
PV1 PV2 PV3 Input 1 Input 2 Input 3 % Output 1 % Output 2 % Output 3 Back Screen
The current alarm source will be displayed at the bottom of the screen. The Back Screen button will display the Alarm Settings page 1 menu screen. To set the alarm action, press the up arrow next to the action. This will display a screen with seven
buttons from which the user can select the alarm action:
Process High Process Low Band Normally Open Band Normally Closed Deviation NO Deviation NC Back Screen
The current alarm action will be displayed at the top of the screen. The Back Screen button will display the Alarm Settings page 1 menu screen. To set the alarm setpoint, press the value and enter the new setpoint on the keypad displayed. Pressing the More >>> button will display page 2 of the Alarm Settings menu. Pressing the Menu button will display the
Alarms
menu.
Alarm Settings – Page 2
This screen will allow the user to set the smart alarm and whether or not a 0 setpoint blocks the alarm. To enable the smart alarm, press the Enable button under the smart alarm status. To disable the smart alarm, press the Disable button under the smart alarm status. To ensure that a 0 setpoint will block the alarm, press the On button below the status. To ensure that a 0 setpoint will not block the alarm, press the Off button below the status. Pressing the <Back button will display the page 1 of the Alarm Settings menu. Pressing the More> button will display page 3 of the Alarm Settings menu. Pressing the Menu button will display the
Alarms
menu.
HP 15 Manual Rev A Page 16 of 16
Alarm Settings – Page 3
This screen will allow the user to set the hysteresis and the on delay time for the alarm. To set the hysteresis, press the value and enter the new hysteresis value from the calculator. To set the On delay time, press the value and enter the new On delay time from the calculator. Pressing the <<<Back button will display page 2 of the Alarm Settings menu. Pressing the Menu button will display the
Alarms
menu.
Over Temp Alarm
If at any time the alarm setpoint is exceeded, the over temp alarm screen will be displayed on the screen. This screen will remain displayed until the acknowledge button, ACK, is pressed.
SSi Information Screen
On any screen where the Super Systems Inc logo is visible, the user will be able to display the information screen for the HP 15. Press the logo to
display the screen. The information screen will display the
Super Systems Inc address and contact information, as well
as the screen version and the firmware version of the 9120.
The user can close out the screen either by pressing the Super Systems Inc
logo again, or by pressing the Close button.
HP 15 Manual Rev A Page 17 of 17
Spare Parts List
31125 24 Volt / 1.2 Amp Power Supply
13454 Series 9120 Controller
31292 Operator Interface - Touch Screen
37177 Pump - Sample and Burnoff Air
37020 Pump - Reference Air
20059 Filter with Media
37037 Replacement Filter Media (20 grams)
32074 Heater - 110V / 425 Watt
10138 Ceragold Probe 400 zirconia sensor w/ "S" T/C
HP 15 Manual Rev A Page 18 of 18
Appendix 1 – Factory Defaults
Universal Defaults
Temperature in °F
All Baud Rates initially 19200 and set for Modbus
Burnoff time is 90 seconds
Burnoff recovery wait time is 2 minutes
Burnoff interval is 12 hours
Burnoff maximum temperature is 2000 °F
CO Factor is 200
Hydrogen factor for Dewpoint calculation is 40%
Analog to Digital Converter filter time is 8 seconds for all 3 ADCs
For analog input 1 initial scale is 0, and full scale is 1000
For analog input 2 initial scale is 0, and full scale is 10000
For analog input 3 initial scale is 0, and full scale is 10000
First Process Control Setpoint is set to 0
First Process Proportional Band is 20%
First Process Reset for PID control is On
First Process Rate for PID control is 0
First Process Output Cycle Time for Control is 16 milliseconds
First Process Auto/Manual set to Automatic
First Process Control Load Line or Manual Reset (integral preset) set to 0
Second Process Control Setpoint is set to 0
Second Process Proportional Band is 4.4%
Second Process Reset for PID control is 12
Second Process Rate for PID control is 216
Second Process Output Cycle Time for Control is 60 milliseconds
Second Process Auto/Manual set to Automatic
Second Process Control Load Line or Manual Reset (integral preset) set to 0
Start PV for temporary tuning set to 0
Temporary Proportional band for tuning set to 2%
Temporary reset for tuning set to On
Temporary rate for tuning set to 0
Temporary integral preset for tuning set to 0
Relay 1 set for loop 2 heat
Relay 2 set for alarm 1
Relay 3 set for loop 1 increment (gas)
Relay 4 set for loop 1 decrement (air)
Relay 5 fixed by PV type.
Relay 6 set for Probe Burnoff
Relay 7 set for Alarm 3
Relay 8 set for all alarms.
Alarm 1 setpoint set for 1800
Alarm 1 is set for analog input 2
Alarm 1 hysteresis is set for 1
Alarm 2 setpoint set for 2000
Alarm 2 is set for analog input 1
HP 15 Manual Rev A Page 19 of 19
Alarm 2 hysteresis is set for 1
Alarm 3 setpoint set for 2000
Alarm 3 is set for process variable 2
Alarm 3 hysteresis is set for 1
Digital to Analog converter 1 assignment is PV1 retransmit
Digital to Analog converter 2 assignment is loop 2 increment
Default IP is 192.168.0.200
Carbon PVT Defaults
Process variable 1 is carbon
Process variable 2 is temperature
Process variable 3 is temperature
Analog output 1 offset is 0
Analog output 1 range is 2000
Analog output 1 offset is 0
Analog output 1 range is 2000
First process control low limit is –100
First process control high limit is 100
First process control mode is such that setpoint of 0 shuts down and it is in dual mode
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such a high analog input 2 shuts down
Analog input 1 full scale is 2500
Analog input 1 type is 2.5 volts
Analog input 2 type is S thermocouple
Analog input 3 type is K thermocouple
Dewpoint PVT Defaults
Process variable 1 is dewpoint
Process variable 2 is temperature
Process variable 3 is temperature
Analog output 1 offset is 0
Analog output 1 range is 2000
Analog output 1 offset is 0
Analog output 1 range is 2000
First process control low limit is –100
First process control high limit is 100
First process control mode is such that dual direct mode
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such a high analog input 2 shuts down
Analog input 1 full scale is 2500
Analog input 1 type is 2.5 volts
Analog input 2 type is S thermocouple
Analog input 3 type is K thermocouple
Oxygen PVT Defaults
Process variable 1 is carbon
HP 15 Manual Rev A Page 20 of 20
Process variable 2 is temperature
Process variable 3 is temperature
Analog output 1 offset is 0
Analog output 1 range is 2000
Analog output 1 offset is 0
Analog output 1 range is 2000
First process control low limit is 0
First process control high limit is 100
First process control mode is such that it is in single reverse
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such a high analog input 2 shuts down
Analog input 1 full scale is 1250
Analog input 1 type is1.25 volts
Analog input 2 type is S thermocouple
Analog input 3 type is S thermocouple
Alarm 3 setpoint is 1.0%
Alarm 3 type PV1 process low
Open TC input1 goes down scale
Relay 3 assignment is loop 1 decrement (gas)
Relay 4 assignment is loop 1 increment (air)
Probe mV PVT Defaults
Process variable 1 is Millivolts
Process variable 2 is temperature
Process variable 3 is temperature
Analog output 1 offset is 0
Analog output 1 range is 2000
Analog output 1 offset is 0
Analog output 1 range is 2000
First process control low limit is –100
First process control high limit is 100
First process control mode is such that setpoint of 0 shuts down and it is in dual mode
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such a high analog input 2 shuts down
Analog input 1 full scale is 2500
Analog input 1 type is 2.5 volts
Analog input 2 type is S thermocouple
Analog input 3 type is K thermocouple
Probe Redundancy PVT Defaults
Process variable 1 is probe 1
Process variable 2 is probe 2
Process variable 3 is temperature
Analog output 1 offset is 0
Analog output 1 range is 200
Analog output 1 offset is 0
HP 15 Manual Rev A Page 21 of 21
Analog output 1 range is 200
First process control low limit is –100
First process control high limit is 100
First process control mode is such that setpoint of 0 shuts down and it is in dual mode
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such that it is in single reverse
Analog input 1 full scale is 2500
Analog input 2 full scale is 2500
Analog input 1 type is 2.5 volts
Analog input 2 type is 2.5 volts
Analog input 3 type is K thermocouple
20 minutes delay time before alarm
Band +/1 for alarm in millivolts is 10
Modbus at 9600 baud
Simple Nitrider PVT Defaults
Process variable 1 is carbon
Process variable 2 is temperature
Process variable 3 is temperature
DAC 1 is set for Hydrogen
DAC2 is set for Dissociation
Analog output 1 offset is 0
Analog output 1 range is 1000
Analog output 1 offset is 0
Analog output 1 range is 1000
First process control low limit is –100
First process control high limit is 100
First process control mode is such that setpoint of 0 shuts down and it is in dual mode
Second process control low limit is 0
Second process control high limit is 100
Second process control mode is such that it is in single reverse
Analog input 1 full scale is 2500
Analog input 1 type is 2.5 volts
Analog input 2 type is S thermocouple
Analog input 3 type is K thermocouple
host RS232
Hitech H2 cell at 9600 baud
Dual Loop PVT Defaults
Alarm 2 setpoint is 2000
Alarm 3 setpoint is 2000
Alarm 2 type is input 1 process high
Alarm 3 type is PV2 process high
HP 15 Manual Rev A Page 22 of 22
Revision History
Rev. Description Date MCO #
- Initial Release 06/14/07 N/A
A Updated logo on title page; Changed screen
shots; Added “O2 View” (although this section is not implemented); Added warning when PVT type is changed; Added “Alarms” section of Configuration Menu; Updated “Revision History” section – Added “MCO #” block; Set Alarm 1 as the dedicated over temperature alarm; Added “Factory Defaults” appendix
12/21/07 2055
HP 15 Manual Rev A Page 23 of 23
Loading...