Super Systems DP2000 User Manual

SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
Model DP 2000
Portable Digital Dew Point Analyzer
For measurement of:
Endothermic Generators
Nitrogen / Methanol Atmosphere
Nitrogen / Hydrogen Atmosphere
Plant Air Systems
SPECIFICATIONS –
Measurement Range: 0 to +80°F (-18 to +27°C) Temperature Range: 0 to +120°F (-18 to +49°C) Power Supply: Factory set to 115 or 240 VAC Display Type: LED Digital Display Resolution: +/- 1°F (+/- 0.1°C) Retransmission Output: 4 –20 mA (range is –50 to +80°F) Battery Specifications: 12 V Rechargeable Battery Life: Approximately 8 hours Size (Closed Case): 11” x 10” x 7” Weight: 8.2 lbs (3.7 kg)
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 1 of 17
Revision Level “F” (800
) 666-4330
SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
TABLE OF CONTENTS –
SPECIFICATIONS –..................................................................................................................................1
TABLE OF CONTENTS –.........................................................................................................................2
INTRODUCTION – ...................................................................................................................................2
WARNINGS –............................................................................................................................................2
STARTUP –................................................................................................................................................3
OPERATION - ...........................................................................................................................................3
I
NSTRUMENT DAMAGE –..............................................................................................................................4
WHAT IS DEW POINT? ...........................................................................................................................5
HOW IT WORKS –....................................................................................................................................5
MAINTENANCE / SAFETY ISSUES –....................................................................................................6
FACTORY CALIBRATION –...................................................................................................................6
FIELD CALIBRATION – ..........................................................................................................................6
RECHARGING THE MODEL DP2000 – ...............................................................................................12
RETURNING THE UNIT TO SS
SPARE PARTS –......................................................................................................................................14
APPENDIX “A” – ....................................................................................................................................15
APPENDIX “B” –.....................................................................................................................................16
I –.........................................................................................................13
INTRODUCTION –
Thank you for selecting Super Systems Inc. and the DP2000 as your source for accurate dew point measurements.
We have taken every precaution to protect this unit during shipment. It has been packed in a customized foam carrier to protect it against mishandling during shipment. Please retain this packing material to use when returning the instrument to SSi for calibration or service. Carefully unpack the Model DP2000 Dew Point Analyzer, and if there are any signs of shipping damage, notify SSi and the shipper immediately.
Keep this instruction book in a secure place and refer to it when there is a question about the analyzer. An electronic version of this manual can be downloaded for free from our website: www.supersystems.com.
WARNINGS –
Although it is intended for use in an industrial environment, the DP2000 is a sensitive piece of analysis equipment. Care should be taken not to drop the analyzer or to operate it in a manner inconsistent with its intended use.
Open all sample ports and remove all soot and/or moisture from the lines prior to
attaching the sample tubing.
The analyzer must be stored at ambient temperature (65-80°F) for at least four hours
prior to operation.
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 2 of 17
Revision Level “F” (800
) 666-4330
SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
When the unit is to be returned to SSi for service or any other reason, it should be
shipped in its original protective packaging. If this packaging is not available, protect the instrument with at least four inches of foam or other impact-absorbing material.
For maximum battery life, do not charge the unit until the “battery low” indicator on
the display is illuminated.
This unit is not designed to measure the dew points in corrosive gasses, such as
Ammonia, S03, Chlorine, and HCL.
Please read and understand this Product Manual before operating the unit.
Failure to comply with these conditions may cause damage to the unit that will not be covered under the warranty. Super Systems, Inc. is not responsible for damage to this unit caused by disregard of these warnings, neglect, or misuse.
STARTUP –
The DP2000 Dew Point Analyzer has been calibrated and fully charged before it was shipped from Super Systems, Inc. You can begin typical operation as soon as the unit has been allowed to stabilize in a temperature similar to the temperature in the heat-treating department. This is particularly important for units that may have been sitting overnight in a delivery van in sub-zero weather, since the rapid temperature change can cause condensation on the sensor which will cause the unit to temporarily display inaccurate readings.
OPERATION -
The use of the DP2000 is somewhat dependent upon the application. Although they are fundamentally alike, the operation procedures are different for sampling heat treating furnaces and endothermic generators. Regardless of the application, allow the DP2000 to operate for two to three minutes before recording any data. This assures that the sensor has achieved equilibrium in the sample gas. Also, make sure that the in-line filter (P/N 31033 or equivalent) is clean and functional. Not only will this ensure that the sample reading is not abnormally high (since soot tends to trap moisture), but it will also prevent soot and other contaminants from entering the unit and damaging the sensor. The optimum flow rate of the sample gas should be between 1.5 and 2.0 Standard Cubic Feet per Hour (SCFH), although a flow rate as low as 1.0 SCFH is acceptable. If the unit is reading less than 1.0 SCFH, verify that there are no obstructions to the flow such as a clogged sample line or a poorly adjusted knob on the DP2000’s flow meter. If these conditions do not exist, the unit may need to be returned to SSi for replacement of the pump. The pump’s life is dependent upon how frequently it is used as well as the cleanliness of the sample gas. Insufficient filtration will significantly reduce the operating life of the pump.
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 3 of 17
Revision Level “F” (800
) 666-4330
SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
Heat Treat Furnace Sampling: A gas sample may be extracted from a process using the
built-in pump. The sample tube from which the sample is taken out of the furnace should extend into the furnace past the HOT face of the refractory. For accurate results, a designated sample port should be used to extract the sample. SSi offers a sample port assembly (part number 20263) which is ideal for this purpose. If a designated sample port is not available, then a clean “burn-off” port on a Gold Probe™, an industry leading oxygen sensor for atmosphere control, can be used. Readings taken from the burnoff port on a probe may be artificially high due to the presence of soot in the probe sheath.
Endothermic Generator Sampling: For applications under pressure, the pump should be
switched off and the flow controlled by the small restriction valve on the flow meter. A flow rate between 1.5 and 2.0 SCFH is ideal. The sample should be taken from the endothermic gas manifold after the gas has been cooled. NOTE: Allow the sample port to blow out any soot and / or water before connecting the sample tube. Failure to do so will result in inaccurate readings and expose the sensor to potential damage.
Instrument Damage –
The two main causes of damage to the DP2000 are the ingestion of soot and water. Both of these contaminants will cause erroneous readings in the short term, and cause long­term damage to the sensor and internal components.
Soot / Particulate Contaminants
When taking a sample from a furnace or a generator, care should be taken to reduce the amount of soot that enters the instrument. The in-line filter will trap these particles, but cleaning the sample line before attaching the DP2000 will increase the life of the filter. Furnace ports can be burned off by pumping air through them while hot, or by removing them from the heat and mechanically cleaning them. Generator ports should be opened before the instrument is attached to allow any particulate buildup to be blown out. It is also helpful to tap on the port while it is being blown out to eject any loose particles before the instrument is attached. If soot is allowed to collect on the dew point sensor in the instrument, it could result in higher readings. This soot will also retain moisture than can corrode the sensor over time. The sensor tip can be cleaned by carefully removing it from the sample block (see Section 2.3 of the Field Calibration instructions) and rinsing it in isopropyl alcohol. The power should be off while this is done, and the power should remain off for at least 30 minutes after this procedure to allow all of the alcohol to completely evaporate.
Water / Moisture Contamination
When a furnace or generator is being started up or cooled down, the resulting gases will contain unusually high amounts of CO2. When these gases cool, moisture will precipitate out and become condensation inside the sample tubing assembly. Even if the
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 4 of 17
Revision Level “F” (800
) 666-4330
SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
furnace or generator is operating normally, residual moisture may still be present in the sample tube or plumbing system. In the same way that the ports are checked for soot (see above) they should be checked for moisture before attaching the instrument. This is especially important when taking a sample from a generator, since the sample port is usually preceded by a significant amount of plumbing. All traces of moisture should be eliminated before attaching the instrument. Failure to do so will result in erroneous measurements and could result in damage to the analyzer. The first signs of moisture in the instrument will be visible condensation in the sample tubing and an unusually high dew point. The upper range of the sensor is +80°F, so if that value is displayed on the instrument it is probably due to the presence of moisture. If this moisture is not removed, it will cause the sensor tip to corrode and will eventually require the sensor to be replaced. To remove moisture from the instrument, the sample tubing and filter should be removed from the instrument since they will probably be wet. An inert gas such as Nitrogen or Argon should then be flowed through the instrument (with the pump off) for as much time as it takes to dry out. This dry-out time will depend on the amount of moisture present in the instrument. The condition of the sensor can be monitored by periodically reading the dew point from the display and watching the value decrease over time. To test if it is operating properly, verify the ambient dew point against a web-based weather station that will report the ambient dew point for your area. If the displayed reading is within three degrees of the reported dew point when the instrument is taken outside, then all of the moisture has probably been successfully removed. The wet filter and sample tubing can be re-attached after they have been completely dried out. To prevent the possibility of moisture damaging the instrument, be sure that the measured dew point is below ambient levels before it is stored. If necessary, Nitrogen or Argon can be used to dry out the instrument after use.
WHAT IS DEW POINT?
Dew point can be defined as the temperature at which the water vapor pressure of the gas equals the saturated water vapor pressure. In other words, it is the temperature at which condensation will just begin to occur as the gas is cooled. Dew point and relative humidity are not the same measurement. Relative humidity is the amount of water vapor in the air compared to the amount the air could hold if it was totally saturated, and it is expressed as a percentage, not a temperature. To determine dew point, two main variables are required: Relative Humidity and Temperature. The DP2000 measures both variables to compute the displayed dew point.
HOW IT WORKS –
The dew point sensor is a “dielectric ceramic” that varies its electrical capacitance with changes in relative humidity. The sensor is mounted in a short probe, which is installed in a T-fitting that allows the sample gas to flow past the sensor. The tip of this probe
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 5 of 17
Revision Level “F” (800
) 666-4330
SSi Super Systems Inc.
Product Manual - Model DP2000 Portable Dew Point Analyzer
contains the dielectric ceramic relative humidity (RH) sensor, as well as a built in temperature sensor to determine its dry bulb temperature. Information from both of these sensors is used to compute the resultant dew point, which is displayed on the digital LED display.
MAINTENANCE / SAFETY ISSUES –
One of the added features of the DP2000 is the ability to monitor the sensor’s operating temperature through the built-in thermister in the probe tip. The temperature of the sample gas can be determined by pressing the switch labeled “Sensor Temp”. It is spring loaded, so it will automatically return to displaying dew point. Maintaining proper sensor temperature will prevent the premature failure of the sensor. The operating temperature of the sensor should remain below 140° F (60°C) at all times. Periodic checks of the sensor temperature will verify that the sensor is not being exposed to excessive heat. If these periodic checks show a high sensor temperature, then the length of sample tubing should be increased to allow for adequate cooling of the sample before it passes the sensor tip. Continuous operation of the DP2000 Dew Point Analyzer will lead to premature failure, since many the internal components are not designed for uninterrupted use. If continuous monitoring of Dew Points is required, please contact Super Systems, Inc. at (800) 666­4330 to inquire about products intended specifically for this application.
FACTORY CALIBRATION –
Factory calibration is recommended every six months if the unit is used regularly. SSi’s calibration is NIST traceable and includes a numbered “Certificate of Calibration”. This certificate also indicates the accuracy of the analyzer before and after calibration. Please contact Super Systems, Inc. at (800) 666-4330 for more information regarding this service.
FIELD CALIBRATION –
It is also possible to calibrate the DP2000 in the field, which will require the optional calibration kit (Part Number 31030). The instructions for a field calibration are shown here, however please feel free to contact Super Systems at 800-666-4330 if you would like to review the process with us before you begin.
The calibration kit consists of two bottles of saturated salt solution in which each bottle generates a precise relative humidity percentage (R.H.%) value. One bottle is 11.3% R.H., and the other is 75.3% R.H. These two specific calibration points are already pre­programmed into the microprocessor board.
1.0 Open the unit
SSi Manual 4545 – DP2000 Super Systems, Inc. Page # 6 of 17
Revision Level “F” (800
) 666-4330
Loading...
+ 11 hidden pages