Super Systems AC20 User Manual

Super Systems
Instruction Manual
Model AC20
Part No. 31069 - Standard Atmosphere Controller  with expanded event capability
August 2004
INDEX
MOUNTING REQUIREMENTS ..................................................... 1
OUTLINE AND CUT OUT DIMENSIONS ...................................... 2
CONNECTION DIAGRAMS ......................................................... 3
PRELIMINARY HARDWARE SETTINGS .................................... 14
SECURITY CODE SETTING MODE ........................................... 17
RUN TIME AND CONFIGURATION MODES .............................. 20
General notes about graphic symbols ............................. 20
Keyboard description ....................................................... 20
CONFIGURATIO N MODE.......................................................... 22
RUN TIME MODE ...................................................................... 44
Display function ................................................................ 44
Indicators.......................................................................... 46
Bargraph description ........................................................ 48
Direct access to the set point........................................... 49
Manual function ................................................................ 49
Special output features ..................................................... 50
Burn off procedure ............................................................51
Probe test procedure........................................................ 53
Serial link ......................................................................... 54
Lamp test ......................................................................... 55
SMART function................................................................ 55
COF and H2F adjustment diagram .................................... 57
Parameter protection ........................................................ 58
RUN TIME PARAMETER MODIFICATION ...................................58
ERROR MESSAGES ................................................................. 83
GENERAL INFORMATION ......................................................... 85
MAINTENANCE ......................................................................... 93
DEFAULT PARAMETERS ........................................................ A.1
ALGORITHMS........................................................................ B.1
CALIBRATION PROCEDURE................................................... B.1
MOUNTING REQUIREMENTSMOUNTING REQUIREMENTS
MOUNTING REQUIREMENTS
MOUNTING REQUIREMENTSMOUNTING REQUIREMENTS
This instrument is intended for permanent installation, for indoor use only, in an electrical panel which encloses the rear housing, exposed terminals and wiring on the back. Select a location, for instrument mounting, where minimum vibrations are present and the ambient temperature is within 0 and 50 °C (32 and 122°F). The instrument can be mounted on a panel up to 15 mm thick with a cutout 92 x 92. For outline and cutout dimensions refer to Fig. 2. The surface texture of the panel must be better than 6,3 µm. The instrument is shipped with rubber panel gasket (50 to 60 Sh). To assure the IP65 and NEMA 4X protection, insert the panel gasket between the instrument and the panel as shown in fig. 1. While holding the instrument against the panel proceed as follows:
1) insert the gasket in the instrument case;
2) insert the instrument in the panel cutout;
3) pushing the instrument against the panel;
4) insert the mounting brackets as shown in fig.1;
5) with a screwdriver, turn the screws with a torque between
0.3 and 0.4 Nm.
Screws
Bracket
Fig. 1
1
Bracket
Gasket
Panel
OUTLINE AND CUT OUT DIMENSIONSOUTLINE AND CUT OUT DIMENSIONS
"
OUTLINE AND CUT OUT DIMENSIONS
OUTLINE AND CUT OUT DIMENSIONSOUTLINE AND CUT OUT DIMENSIONS
+0.03 +0.8
3.62" (92 mm )
+0.03
+0.8
3.62" (92 mm )
(125 mm)
4.92" (125 mm)
4.92
Fig. 2. OUTLINE AND CUT-OUT DIMENSIONS
3.78" (96 mm)
0.6" (15mm) max
4.17" (108.5 mm)
5" (128 mm)
0.06" (1.5 mm)
3.8" x 3.8" (97 x 97 mm) gasket
0.63"
(16 mm)
2
CONNECTION DIAGRAMSCONNECTION DIAGRAMS
CONNECTION DIAGRAMS
CONNECTION DIAGRAMSCONNECTION DIAGRAMS
Fig. 3. REAR TERMINAL BLOCK
3
CONNECTIONSCONNECTIONS
CONNECTIONS
CONNECTIONSCONNECTIONS Connections are to be made when the instrument housing is
installed in its proper location.
A) PROBE SENSOR INPUT (MAIN INPUT)A) PROBE SENSOR INPUT (MAIN INPUT)
A) PROBE SENSOR INPUT (MAIN INPUT)
A) PROBE SENSOR INPUT (MAIN INPUT)A) PROBE SENSOR INPUT (MAIN INPUT)
12
13
+
_
Shield
Sensor
MEASURING INPUTSMEASURING INPUTS
MEASURING INPUTS
MEASURING INPUTSMEASURING INPUTS NOTENOTE
NOTE: Any external component (like zener barriers etc.)
NOTENOTE connected between sensor and input terminals may cause errors in measurement due to excessive and/or not balanced line resistance or possible leakage currents.
12
13
+
_
G
Fig. 4 MAIN INPUT WIRING NOTESNOTES
NOTES:
NOTESNOTES
1) This input is isolated from others measuring inputs and digital inputs.
2) Don’t run input wires together with power cables.
3) When a shielded cable is used, it should be connected at one point only.
4
Sensor
A.1) TEMPERATURE (THERMOCOUPLE) INPUTA.1) TEMPERATURE (THERMOCOUPLE) INPUT
A.1) TEMPERATURE (THERMOCOUPLE) INPUT
A.1) TEMPERATURE (THERMOCOUPLE) INPUTA.1) TEMPERATURE (THERMOCOUPLE) INPUT
1
+
_
3
Shield
1
+
_
3
Shield
Fig. 5 TEMPERATURE INPUT WIRING NOTESNOTES
NOTES:
NOTESNOTES
1) This input is not isolated from others measuring inputs and digital inputs.
2) Don’t run input wires together with power cables.
3) For TC wiring use proper compensating cable preferable shielded.
4) When a shielded cable is used, it should be connected at one point only.
A.2) AUXILIARY INPUT (CARBON MONOXIDE)A.2) AUXILIARY INPUT (CARBON MONOXIDE)
A.2) AUXILIARY INPUT (CARBON MONOXIDE)
A.2) AUXILIARY INPUT (CARBON MONOXIDE)A.2) AUXILIARY INPUT (CARBON MONOXIDE)
5
6
Shield
5
6
Fig. 6 AUXILIARY INPUT WIRING NOTESNOTES
NOTES:
NOTESNOTES
1) This input is not isolated from others measuring inputs and digital inputs. A double or reinforced insulation between instrument output and power supply must be assured by the external instrument.
5
+
mA,
or
_
V
+
mA
or
_
V
G
2) Don’t run input wires together with power cables.
3) Pay attention to the line resistance; a high line resistance may cause measurement errors.
4) When shielded cable is used, it should be grounded at one side only to avoid ground loop currents.
5) The input impedance is equal to: < 5 Ω for 20 mA input > 200 k for 5 V input > 400 k for 10 V input
B) LOGIC INPUTSB) LOGIC INPUTS
B) LOGIC INPUTS
B) LOGIC INPUTSB) LOGIC INPUTS
DIG. 1
8
DIG. 2
9
DIG. 3
10
11
IN 1
57
IN 2
58
IN 3
59
IN 4
60
56
Fig. 7.A - LOGIC INPUTS DIG 1 TO DIG 3 AND IN1 TO IN4 WIRING
6
IN 5
46
IN 6
47
IN 7
48
IN 8
49
45
Fig. 7.B - LOGIC INPUTS IN5 TO IN8 WIRING Logic inputs from IN1 to In8 are optional. NOTESNOTES
NOTES:
NOTESNOTES
1) Do not run logic input wiring together with power cables.
2) Use an external dry contact capable of switching 0.5 mA, 5 V DC.
3) The instrument needs 110 ms to recognize a contact status variation.
4) The logic inputs are
NOT NOT
NOT isolated from measuring inputs.
NOT NOT
C.1) RELAY OUTPUTSC.1) RELAY OUTPUTS
C.1) RELAY OUTPUTS
C.1) RELAY OUTPUTSC.1) RELAY OUTPUTS
OUT 1
OUT 2
OUT 3
Fig. 8.A RELAY OUTPUTS 1, 2, 3 WIRING
7
23
24
25
26
27
28
29
NC
C
NO
C
NO
C
NO
OUT 4
30
31
OUT 5
15
14
OUT 10
OUT 11
OUT 12
OUT 13
OUT 14
61 62 63 64
65
66
Fig. 8.B RELAY OUTPUTS 4, 5 and 10 TO 14 WIRING
C
NO
C
NO
NO - OUT 10
NO - OUT 11 NO - OUT 12 NO - OUT 13
NO - OUT 14
C - OUT 10 to 14
NO - OUT 15
OUT 15
OUT 16
OUT 17
OUT 18
OUT 19
Fig. 8.C RELAY OUTPUTS 15 TO 19 WIRING Outputs from OUT 10 to OUT 19 are optional. The contact rating from OUT 1 and OUT 2 is 3A/250V AC on resistive load. The contact rating from OUT 3 and OUT 4 is 2A/250V AC on resistive load. The contact rating of OUT 5 is 1A/250V AC on resistive load. The contact rating from OUT 10 to OUT 19 is 0.5 A/250V AC on resistive load. The number of operations is 1 x 105 at specified rating.
8
50
NO - OUT 16
51
NO - OUT 17
52
NO - OUT 18
53
NO - OUT 19
54
C - OUT 15 to 19
55
NOTES:NOTES:
NOTES: 1) To avoid electrical shock, connect
NOTES:NOTES:
All relay contacts are protected by varistor against inductive load with inductive component up to 0.5 A.
The following recommendations avoid serious problems which may occur, when relay output drives inductive loads.
INDUCTIVE LOADSINDUCTIVE LOADS
INDUCTIVE LOADS
INDUCTIVE LOADSINDUCTIVE LOADS High voltage transients may occur switching inductive loads. Through the internal contacts these transients may introduce disturbances which can affect the performance of the instrument. For all the outputs, the internal protection (varistor) assures a correct protection up to 0.5 A of inductive component. The same problem may occur when a switch is used in series with the internal contacts as shown in Fig. 9.
power line at the end of the wiring procedure.
2) For power connections use No 16 AWG or larger wires rated for at last 75 °C.
3) Use copper conductors only.
4) Don’t run input wires together with power cables.
5) When a relay output is used to drive a low power signal (PLC input, alarm annunciator, etc..) it is necessary to use an external relay with a gilded contact.
C
R
LOAD
Fig. 9 EXTERNAL SWITCH IN SERIES WITH THE INTERNAL
CONTACT In this case it is recommended to install an additional RC network across the external contact as show in Fig. 9. The value of capacitor (C) and resistor (R) are shown in the following table.
LOAD
(mA)
<40 mA <150 mA <0.5 A
Anyway the cable involved in relay output wiring must be as far away as possible from input or communication cables.
9
C
(µF)
0.047
0.1
0.33
(Ω)
100
22 47
R
P.
OPERATING
(W)
VOLTAGE
1/2
260 V AC
2
260 V AC
2
260 V AC
LINE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVEC.2) VOLTAGE OUTPUTS FOR SSR DRIVE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVEC.2) VOLTAGE OUTPUTS FOR SSR DRIVE
+
OUT 1
24
_
+
__
_
__
25
SOLID STATE
RELAY
+
OUT 2
26
_
+
__
_
__
27
SOLID STATE
RELAY
Fig. 10 SSR DRIVE OUTPUT WIRING Logic level 0Logic level 0
Logic level 0: Vout < 0.5 V DC.
Logic level 0Logic level 0 Logic level 1Logic level 1
Logic level 1:
Logic level 1Logic level 1
- 14 V
+ 20 % @ 20 mA
- 24 V
+ 20 % @ 1 mA.
Maximum current = 20 mA
NOTENOTE
NOTE: These outputs are not insulated.
NOTENOTE A double or reinforced insulation between instrument output and power supply must be assured by the external solid state relay.
C.3) ANALOG OUTPUTSC.3) ANALOG OUTPUTS
C.3) ANALOG OUTPUTS
C.3) ANALOG OUTPUTSC.3) ANALOG OUTPUTS
+
OUT 6
16
_
17
+
20 mA
_
Shield
+
16
OUT 6
_
17
Fig. 11.A OUTPUT 6 WIRING
10
+
_
20 mA
G
+
OUT 7
18
_
19
+
_
20 mA20 mA
Shield
+
18
OUT 7
_
19
Fig. 11.B OUTPUT 7 WIRING NOTESNOTES
NOTES:
NOTESNOTES
1) Do not run analog output wires together with AC power cables.
2) Out 6 and 7 are isolated outputs.
3) The maximum load is equal to 600 Ω.
+
_
G
D) SERIAL INTERFACED) SERIAL INTERFACE
D) SERIAL INTERFACE
D) SERIAL INTERFACED) SERIAL INTERFACE RS-485 interface allows you to connect up to 30 devices with one remote master unit.
I N S
T R U M
E N
T
Fig. 12 - RS-485 WIRING The cable length must not exceed 1.5 km at 9600 BAUD. NOTESNOTES
NOTES:
NOTESNOTES
1) This is an isolated RS 485 serial interface.
2) The following report describes the signal sense of the voltage appearing across the interconnection cable as defined by EIA for RS-485. a) The ” A ” terminal of the generator shall be negative with
respect to the ” B ” terminal for a binary 1 (MARK or OFF) state.
11
22
21
20
A/A'
B/B'
COMMON
A'/A
B'/B
M
A S T E R
b) The ” A ” terminal of the generator shall be positive with
respect to the ” B ” terminal for a binary 0 (SPACE or ON).
3) The EIA standard establishes that by RS-485 interface it is possible to connect up to 30 devices with one remote master unit. The serial interface of these instruments is based on “High input impedance” transceivers; this solution allows you to connect up to 127 devices (based on the same transceiver type) with one remote master unit.
E) POWER LINE WIRINGE) POWER LINE WIRING
E) POWER LINE WIRING
E) POWER LINE WIRINGE) POWER LINE WIRING
N (L2)
32
POWER LINE 100 V to
240 V A.C (50/60Hz)
33
L (L1)
Line
Fig. 13 POWER LINE WIRING NOTESNOTES
NOTES:
NOTESNOTES
1) Before connecting the instrument to the power line, make sure that line voltage corresponds to the description on the identification label.
2) To avoid electrical shock, connect power line at the end of the wiring procedure.
3) For supply connections use No 16 AWG or larger wires rated for at last 75 °C.
12
Neutral
4) Use copper conductors only.
5) Don’t run input wires together with power cables.
6) The power supply input is fuse protected by a sub miniature fuse rated T, 1A, 250 V. When fuse is damaged, it is advisable to verify the power supply circuit, so that it is necessary to send back the instrument to your supplier.
7) The safety requirements for Permanently Connected Equipment say:
- a switch or circuit-breaker shall be included in the building
installation;
- it shall be in close proximity to the equipment and within
easy reach of the operator;
- it shall be marked as the disconnecting device for the
equipment.
NOTENOTE
NOTE: a single switch or circuit-breaker can drive more than
NOTENOTE one instrument.
8) When a neutral line is present please connect it to the 32 terminal.
13
PRELIMINARY HARDWARE SETTINGSPRELIMINARY HARDWARE SETTINGS
PRELIMINARY HARDWARE SETTINGS
PRELIMINARY HARDWARE SETTINGSPRELIMINARY HARDWARE SETTINGS
How to remove the instrument from its caseHow to remove the instrument from its case
How to remove the instrument from its case
How to remove the instrument from its caseHow to remove the instrument from its case
1) Switch off the instrument.
2) Push gently the lock A on the right.
3) While the lock A is maintained out, slide out the right side of the instrument (see fig. 14.a)
B
4) Push gently the lock C on the left.
5) While the lock C is maintained out, slide out the instrument (see fig. 14.b)
D
Fig. 14.a
A
B
Fig. 14.b
14
C
D
J103 SETTINGJ103 SETTING
J103 SETTING
J103 SETTINGJ103 SETTING J103 (see fig. 15) must be set as follows:
J103 1-2 open 3-4 open 5-6 open 7-8 open 5-7 close 6-8 close
AUXILIARY INPUT SELECTIONAUXILIARY INPUT SELECTION
AUXILIARY INPUT SELECTION
AUXILIARY INPUT SELECTIONAUXILIARY INPUT SELECTION Set J102 (see fig. 15) according to the desired input type as shown in the following table.
J102 INPUT TYPE
5 V 10 V 20 mA 1-2 close open open 3-4 close close open 5-6 open open close 7-8 open open close 5-7 open close open 6-8 open open open
1 3 5 7
J102
2 4 6 8
1 3 5 7
J103
2 4 6 8
CPU card
Fig.15
15
Operative mode and Hardware lockOperative mode and Hardware lock
Operative mode and Hardware lock
Operative mode and Hardware lockOperative mode and Hardware lock
1) By V101 (see fig. 16) it is possible to select one of the following operative modes: a) run time mode without configuration mode b) run time and configuration modes c) security code setting mode Set V101 according to the following table:
Modes V101.1 V101.2 V101.3 V101.4 a OFF ON ON ON b OFF ON OFF ON c OFF ON OFF OFF
2) When run time mode is selected (mode a or b), V101.3 allows you to activate/deactivate the hardware lock for configuration parameters. If V101.3 is ON, the lock is activated. If V101.3 is OFF, the lock is deactivated. When the lock is activated, no one of the configuration parameter can be modified.
3) All the others switch combinations are reserved.
ON DIP
1 2 3 4
V101
Fig. 16
16
SECURITY CODE SETTING MODESECURITY CODE SETTING MODE
SECURITY CODE SETTING MODE
SECURITY CODE SETTING MODESECURITY CODE SETTING MODE General notesGeneral notes
General notes
General notesGeneral notes
The instrument parameters are divided in two families and each family is divided in groups.
- The first family encompasses all the run time parameters.
- The second family comprises all the configuration parameters. A specific security code enables the parameter modification of each family. For run time parameters, it is possible to select which groups of them will be protected by the security code and in this case it is necessary to set the run time security code before to modify one or more parameters of a protected group. The configuration security code protects all configuration parameters and it will be necessary to set the configuration security code before to start the configuration parameters modification. For configuration parameters an hardware lock is also available.
Security code settingSecurity code setting
Security code setting:
Security code settingSecurity code setting
1) Remove the instrument from its case.
2) Set the internal dip switch V101 as follows:
- V101.1 = OFF - V101.2 = ON
- V101.3 = OFF - V101.4 = OFF
3) Re-insert the instrument.
4) Switch on the instrument. The display will show:
The upper display shows that the security code setting mode is selected while the lower display shows the firmware version.
5) Push the FUNC pushbutton.
17
Run time security codeRun time security code
Run time security code
Run time security codeRun time security code The display will show:
Run time groups protected by security codeRun time groups protected by security code
Run time groups protected by security code
Run time groups protected by security codeRun time groups protected by security code The display will show:
NoteNote
Note: the middle display shows the current status of the run
NoteNote time security code ("0", "1" or "On"). By and push-button, set "S.run" parameter as follows:
0 No protection (it is always possible to modify all run
time parameters);
1 always protected (it is never possible to modify a run
time parameter);
from 2 to 250 security code for run time parameter
protection.
NOTESNOTES
NOTES:
NOTESNOTES
1) The selected value of a security code cannot be displayed anymore and, coming back to the "S.run" parameter, the display will show "On" when "S.run" is different from 0 or 1, "0" when "S.run" is equal to 0, "1" when "S.run" is equal to 1. When the security code is forgotten, a new value can be set.
2) When "S.run" is different from 0 or 1, the "run time default " and the "run time hidden" groups are always protected by security code.
By this parameter it is possible to set if the run time group 1 will be protected or not by the run time security code. By and push-button, set "Gr1" parameter as follows:
nO No protection (it is always possible to modify run time
group 1 parameters).
Yes the run time group 1 parameter modification will be
protected by security code. Push the FUNC push-button; the instrument memorizes the new setting and goes to the next parameter.
NOTESNOTES
NOTES: 1) This selection may be performed only if a run time
NOTESNOTES
security code has been set (from 2 to 250).
2) The above described selection may be repeated for all groups of the run time mode.
18
Configuration security codeConfiguration security code
Configuration security code
Configuration security codeConfiguration security code The display will show:
NoteNote
Note: the middle display shows the current status of the
NoteNote
configuration security code ("0", "1" or "On").
By and push-button, set "S.CnF" parameter as follows:
0 No protection (it is always possible to modify all
configuration parameters);
1 always protected (it is never possible to modify a
configuration parameter);
from 2 to 250 security code for configuration parameter
protection.
NOTESNOTES
NOTES:1) The selected value of a security code cannot be
NOTESNOTES
displayed anymore and, coming back to the "S.CnF" parameter, the display will show "On" when "S.CnF" is different from 0 or 1, "0" when "S.CnF" is equal to 0, "1" when "S.CnF" is equal to 1. When the security code is forgotten, a new value can be set.
2) At the end of the security code setting, set the V101 as described at page16.
RUN TIME AND CONFIGURATION MODESRUN TIME AND CONFIGURATION MODES
RUN TIME AND CONFIGURATION MODES
RUN TIME AND CONFIGURATION MODESRUN TIME AND CONFIGURATION MODES
The hardware selection described in "Operative mode and hardware lock" paragraph allows you to start one of the two following operative modes:
- run time mode
- configuration mode. At power up, the instrument starts in the same mode it was prior
to the power OFF.
General note about graphic symbols used forGeneral note about graphic symbols used for
General note about graphic symbols used for
General note about graphic symbols used forGeneral note about graphic symbols used for mnemonic code visualization.mnemonic code visualization.
mnemonic code visualization.
mnemonic code visualization.mnemonic code visualization. The instrument displays some characters with special symbols. The following table shows the correspondence between the symbols and the characters.
19
symbol character
"
"k " "m " "V " "W " "Z
"J
"
Keyboard descriptionKeyboard description
Keyboard description
Keyboard descriptionKeyboard description MENU = is used to select a parameter group.
FUNC = When the instrument is in "normal display mode" it
changes the indication on the lower display (see "display function").
During parameter modification, it allows you to
memorize the new value of the selected parameter and go to the next parameter (increasing order).
MAN = When the instrument is in "normal display mode",
pushing MAN push-button for more than 1 s, it is possible to enable or disable the manual function.
During parameter modification, it allows you to scroll
back the parameters and groups without memorizing the new setting.
= During parameter modification, it allows you to
increase the value of the selected parameter.
During MANUAL mode, it allows you to increase the
output value.
TST + MAN = By pressing these keys it is possible to start the
BURNOFF function.
TST + FUNC = By pressing these keys it is possible to start the
PROBE TEST function.
+MENU = Are used to start the lamp test function (the
command is accepted when push-buttons are kept depressed for more than 5 s and the instrument is in normal display mode)
+FUNC or +FUNC
During parameter modification they allow you to increase/decrease the value under modification with higher rate.
+MAN or +MAN
During parameter modification they allow you to jump to the max or min programmable value.
= During parameter modification, it allows you to
decrease the value of the selected parameter.
During MANUAL mode, it allows to decrease
the output value.
20
NOTESNOTES
NOTES:
NOTESNOTES
1) All the actions explained above which requires two or more push-buttons must follow exactly the push-button sequence shown.
2) A 10 or 30 seconds time out (see "CnF.6 - t.out" [C.I08]) can be selected for parameter modification during run time mode. If, during parameter modification, no push-button is depressed for more than 10 (30) seconds, the instrument goes automatically to the “normal display mode” and the modification (if carried out) of the last displayed parameter will be lost.
CONFIGURATION MODECONFIGURATION MODE
CONFIGURATION MODE
CONFIGURATION MODECONFIGURATION MODE
Switch on the instrument. The instrument will start in the same way it was prior to the power down (configuration mode or run time mode).
If the instrument starts in configuration mode, push the MENU pushbutton and go to the "Configuration group 1" (see page 25). If the instrument starts in run time mode, by keeping depressed the MENU push-button for more than 5 seconds the instrument will show:
where:
- the upper display shows the selected parameter family;
- the middle display shows the selected action;
- the lower display shows the firmware version. If no push-button is depressed for more than 10 s (or 30 s according to "CnF.6 - "t.out" [time out selection" C.I08] parameter setting), the instrument automatically returns to the normal display mode.
21
By or push-button it is possible to select between:
= (monitor) this selection allows you to monitor but
not to modify the value assigned to the configuration parameters.
= (modify) this selection allows you to monitor and to
modify the value assigned to the configuration parameters.
NOTESNOTES
NOTES:
NOTESNOTES
1) During monitor mode, the instrument continues to operate as in run time mode.
2) When modify mode is started, the instrument stops the control and:
- sets to OFF the control outputs;
- turns to OFF the bargraph displays;
- sets analog retransmissions to the retransmitted initial scale
value;
- sets to OFF the alarms;
- disables the serial link;
- the time out will be disabled.
3) When the modify mode is disabled by V101 (V101.3), the or push-button pressure has no effect.
MONITOR MODEMONITOR MODE
MONITOR MODE
MONITOR MODEMONITOR MODE During the run time mode, it is possible to monitor but not modify all configuration parameters. When it is desired to verify the instrument configuration, proceed as follows:
1) By or push-button select the monitor mode.
2) Push the MENU push-button the display will show:
it shows that configuration group 1 is selected and it
encompasses all the input parameters. The configuration parameter "Monitor mode" follows the "Modify mode" sequence.
NOTESNOTES
NOTES:
NOTESNOTES
1) During monitor mode, the instrument continues to operate as
in run time mode.
2) During monitor mode, if no push-button is depressed for more
than 10 s (or 30 s according to "CnF.6 - t.out" [C.I08]
parameter setting), the instrument returns automatically to the
normal display mode.
22
MODIFY MODEMODIFY MODE
MODIFY MODE
MODIFY MODEMODIFY MODE
1) By or push-button select the modify mode.
2) Push the MENU push-button. If a security code is applied to the configuration parameter, the instrument will show:
3) By and push-button set a value equal to the security code assigned to the configuration mode (see "Configuration security code " at page 19). If the code is different from the security code, the instrument automatically returns to the first configuration display otherwise the display will show:
The modify mode is started. This display allows you to load the default configuration parameter (table 1 or table 2). For more details see chapter "Default parameter" (see Appendix A).
4) By or push-button select the OFF indication and push the MENU push-button. The display will show:
This is the starting display of the first group of configuration parameters.
NOTESNOTES
NOTES:
NOTESNOTES
1) In the following pages we will describe all the parameters of the instrument but the instrument will show only the parameters related with the specific hardware and in accordance with the specific instrument configuration [i.e. setting OUT 6 as "nonE" (not used), all the parameters related with this output will be automatically skipped].
23
2) During configuration parameters modify mode, the upper display shows the selected parameter group, the lower display shows the mnemonic code of the selected parameter while the central display shows the value or status assigned to the selected parameter.
3) For an easy consultation of this manual, a sheet named "Reference parameter guide" with all the parameter visualizations is enclosed. The groups of configuration parameters are identified by the "C" letter followed by A, B etc. The "code" formed by the group and the row (i.e. C.D01 where "C.D" is configuration group1 and "01" is the row 1) is reported, in the user manual, before each parameter description and allows you to quickly find out the respective parameter.
CONFIGURATION GROUP 1 [C.Dxx]CONFIGURATION GROUP 1 [C.Dxx]
CONFIGURATION GROUP 1 [C.Dxx]
CONFIGURATION GROUP 1 [C.Dxx]CONFIGURATION GROUP 1 [C.Dxx] INPUT CONFIGURATION
Push the FUNC push-button C.D01- Line frequencyC.D01- Line frequency
C.D01- Line frequency
C.D01- Line frequencyC.D01- Line frequency Upper display: CnF.1 Lower display: Ln.Fr Range: 50 Hz
60 Hz
C.D02- Controlled variable selectionC.D02- Controlled variable selection
C.D02- Controlled variable selection
C.D02- Controlled variable selectionC.D02- Controlled variable selection Upper display: CnF.1 Lower display: PV.SL Ranges: CP = Carbon potential (span limits within 0.00% to 2.00%). dP = Dew point (span limits within -100 to 100 °F or -75 to 40°C). ñV = Sensor output in mV (span limits within 0 to 1500 mV).
O2 = Oxygen value as primary control variable ( The span limits are 0.0 to 25.0)
24
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) Changing the controlled variable, the following parameters: "CnF.6 - brG.L" [C.I03] and "CnF.6 - brG.H" [C.I04] (Bargraph initial and full scale values); "CnF.2 - O6.Lr" [C.E06] and "CnF.2 - O6.Hr" [C.E07] (Out 6 retransmission initial and full scale values); "CnF.2 - O7.Lr" [C.E11] and "CnF.2 - O7.Hr" [C.E12] (Out 7 retransmission initial and full scale values); "Gr.5 - rL" [R.E10] and "Gr.5 - rH" [R.E11] (set point low and high limits); will be forced to the span limits of the new selected variable.
2) SP, SP2, SP3, SP4 values and alarm thresholds, if out of the limits for the new selected variable, will be forced to the low limit value.
C.D03- Filter on probe sensor input (Main input)C.D03- Filter on probe sensor input (Main input)
C.D03- Filter on probe sensor input (Main input)
C.D03- Filter on probe sensor input (Main input)C.D03- Filter on probe sensor input (Main input) Upper display: CnF.1 Lower display: Pb.FL Ranges: from 0 (no filter) to 8 seconds. NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) This is a first order digital filter applied on probe sensor input value.
2) This filter can affect the control action, the alarms, the SMART algorithm and the process variable retransmission.
C.D04 - Temperature input - TC type selectionC.D04 - Temperature input - TC type selection
C.D04 - Temperature input - TC type selection
C.D04 - Temperature input - TC type selectionC.D04 - Temperature input - TC type selection Upper display: CnF.1 Lower display: tP.In Ranges:
1 = TC K From -100 t o 1370 °C 2 = TC S From -50 to 17 6 0 °C 3 = TC R From -50 to 1760 °C 4 = TC K From -150 to 2500 °F 5 = TC S From -60 to 32 0 0 °F
6 = TC R From -60 to 3200 °F 7= TC B From 0 to 1820 °C 8= TC B From 32 to 3300 °F NOTE:NOTE:
NOTE: When the controlled variable selected is dP, the span
NOTE:NOTE:
limits of the following parameters: "CnF.6 - brG.L" [C.I03] and "CnF.6 - brG.H" [C.I04] (Bargraph initial and full scale values); "CnF.2 - O6.Lr" [C.E06] and "CnF.2 - O6.Hr" [C.E07] (Out 6 retransmission initial and full scale values); "CnF.2 - O7.Lr" [C.E11] and "CnF.2 - O7.Hr" [C.E12] (Out 7 retransmission initial and full scale values); "Gr.5 - rL" [R.E10] and "Gr.5 - rH" [R.E11] (set point low and high limits); will be forced to:
-100 to 100 °F if the temperature input has been changed from a °C to a °F type;
-75 to 40 °C if the temperature input has been changed from a °F to a °C type.
25
C.D05 - Temperature input offset adjustmentC.D05 - Temperature input offset adjustment
C.D05 - Temperature input offset adjustment
C.D05 - Temperature input offset adjustmentC.D05 - Temperature input offset adjustment Upper display: CnF.1 Lower display: OFSt Range: from -500 to 500 NOTENOTE
NOTE: The offset value is algebraically added to temperature
NOTENOTE
input value.
Read-out
OFSt
Real curve
Adjusted curve
Input
C.D06 - Filter on temperature input valueC.D06 - Filter on temperature input value
C.D06 - Filter on temperature input value
C.D06 - Filter on temperature input valueC.D06 - Filter on temperature input value Upper display: CnF.1 Lower display: tP.FL Ranges: from 0 (no filter) to 8 seconds.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) This is a first order digital filter applied to the temperature input value.
2) When "dP" or "CP" or O2 are selected as the controlled variable, this filter can affect the control action, the alarms, the SMART algorithm and the process variable retransmission.
C.D07 - Auxiliary input functionC.D07 - Auxiliary input function
C.D07 - Auxiliary input function
C.D07 - Auxiliary input functionC.D07 - Auxiliary input function
Lower display: A.In.F Range: nonE = Input not used CO = Input used for CO measurement rSP = Input used as remote set point Available: Always Note:The scalable for CO measurement is fixed from 000 to 100 When CO input is not used the CO value will be set to 20 The scalable for rSP measurement is fixed to:
0.00 / 2.00 if carbon potential is selected as primary control variable or
0 /100 °F (-18 / 40 °C) if dew point is selected as primary control variable or 0 / 1500 if sensor output in mV is selected as primary control variable
0.0/ 25.0 if oxygen value is selected as primary control variable
C.D08 - Auxiliary input type selectionC.D08 - Auxiliary input type selection
C.D08 - Auxiliary input type selection
C.D08 - Auxiliary input type selectionC.D08 - Auxiliary input type selection
Lower display: A.In.t Range: 0-20 = 0÷20 mA
4-20 = 4÷20 mA 0- 5 = 0÷ 5 V
1- 5 = 1÷ 5 V 0-10 = 0÷10 V 2-10 = 2÷10 V
Available: When auxiliary input is used
26
C.D09 - TIME CONSTANT FOR FILTER ON REMOTEC.D09 - TIME CONSTANT FOR FILTER ON REMOTE
C.D09 - TIME CONSTANT FOR FILTER ON REMOTE
C.D09 - TIME CONSTANT FOR FILTER ON REMOTEC.D09 - TIME CONSTANT FOR FILTER ON REMOTE SETPOINT VALUE (CnF.1)SETPOINT VALUE (CnF.1)
SETPOINT VALUE (CnF.1)
SETPOINT VALUE (CnF.1)SETPOINT VALUE (CnF.1) Lower display: A.I.FL Range: From 0(filter OFF) to 8 s Available: When auxiliary input is used as remote set point Note: First order filter with selected time constant
C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)
C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)
C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1) Lower display: A.I.Añ Range: norñ = The remote set point is activated by external contact(or is always activated if none contacts are configured for this feature) and the value is limited at its min/max value when out of range Cnd.A = The remote set point activation is controlled by status of auxiliary input. (The device works with local set point when auxiliary input is out of range while it works with remote set point when auxiliary input is in range)
Available: When auxiliary input is used as remote set point
C.D11 - LOCAL / REMOTE SETPOINT OPERATINGC.D11 - LOCAL / REMOTE SETPOINT OPERATING
C.D11 - LOCAL / REMOTE SETPOINT OPERATING
C.D11 - LOCAL / REMOTE SETPOINT OPERATINGC.D11 - LOCAL / REMOTE SETPOINT OPERATING MODE (CnF.1)MODE (CnF.1)
MODE (CnF.1)
MODE (CnF.1)MODE (CnF.1) Lower display: L.r.Oñ Range: ALG = The local setpoint will be aligned to last remote setpoint value when transfer from remote to local setpoint action is performed through external contact
n.ALG = The local setpoint will be not be changed when transfer from remote to local setpoint action is performed (The programmed ramp (Grd1/Grd2) may be activated)
Available: When auxiliary input is used as remote set point and (“A.I.Añ = norñ) Note :If transfer from remote to local setpoint is due to auxiliary input status(“A.I.Añ = Cnd.A) the local setpoint beenbeen
been aligned.
beenbeen
27
will neverwill never
will never
will neverwill never
CONFIGURATION GROUP 2 [C.Exx]CONFIGURATION GROUP 2 [C.Exx]
CONFIGURATION GROUP 2 [C.Exx]
CONFIGURATION GROUP 2 [C.Exx]CONFIGURATION GROUP 2 [C.Exx]
OUTPUTS CONFIGURATION
C.E01 - OUT 1 functionC.E01 - OUT 1 function
C.E01 - OUT 1 function
C.E01 - OUT 1 functionC.E01 - OUT 1 function Upper display: CnF.2 Lower display: O1.Fn Range: nonE = Output not used
ñAin = Time proportional main control output SECn = Time proportional secondary control output ALr.1 = Alarm 1 output
C.E02 - OUT 2 functionC.E02 - OUT 2 function
C.E02 - OUT 2 function
C.E02 - OUT 2 functionC.E02 - OUT 2 function Upper display: CnF.2 Lower display: O2.Fn Range: nonE = Output not used
ñAin = Time proportional main control output SECn = Time proportional secondary control output ALr.2 = Alarm 2 output
C.E03 - OUT 3 functionC.E03 - OUT 3 function
C.E03 - OUT 3 function
C.E03 - OUT 3 functionC.E03 - OUT 3 function Upper display: CnF.2 Lower display: O3.Fn Range: nonE = Output not used
ñAin = Time proportional main control output SECn = Time proportional secondary control output ALr.3 = Alarm 3 output
C.E04 - OUT 6 functionC.E04 - OUT 6 function
C.E04 - OUT 6 function
C.E04 - OUT 6 functionC.E04 - OUT 6 function Upper display: CnF.2 Lower display: O6.Fn Range: nonE = Output not used
ñAin = Main control output (linear) SECn = Secondary control output (linear) PV.rt = Process variable retransmission SP.rt = Operative set point retransmission
C.E05 - OUT 6 rangeC.E05 - OUT 6 range
C.E05 - OUT 6 range
C.E05 - OUT 6 rangeC.E05 - OUT 6 range This parameter is available only when OUT 6 ("O6.Fn" [C.E04]) is different from "nonE". Upper display: CnF.2 Lower display: O6.rn Range: 0-20 = 0-20 mA
4-20 = 4-20 mA
28
C.E06 -C.E06 -
OUT 6 retransmission - initial scale valueOUT 6 retransmission - initial scale value
C.E06 -
OUT 6 retransmission - initial scale value
C.E06 -C.E06 -
OUT 6 retransmission - initial scale valueOUT 6 retransmission - initial scale value This parameter is available only when Out 6 function [C.E04] is configured as “PV.rt” or “SP.rt” Upper display: CnF.2 Lower display: O6.Lr Range: From -1999 to 9999 NOTENOTE
NOTE: The decimal point position will be the same as the
NOTENOTE
“CnF.1 - PV.SL” [C.D02] previously selected.
C.E07 -C.E07 -
OUT 6 retransmission - full scale valueOUT 6 retransmission - full scale value
C.E07 -
OUT 6 retransmission - full scale value
C.E07 -C.E07 -
OUT 6 retransmission - full scale valueOUT 6 retransmission - full scale value This parameter is available only when Out 6 function [C.E04] is configured as “PV.rt” or “SP.rt” Upper display: CnF.2 Lower display: O6.Hr Range: From -1999 to 9999 NOTENOTE
NOTE: The decimal point position will be the same as the
NOTENOTE
“CnF.1 - PV.SL” [C.D02] previously selected.
C.E08 -C.E08 -
OUT 6 filter on the retransmitted valueOUT 6 filter on the retransmitted value
C.E08 -
OUT 6 filter on the retransmitted value
C.E08 -C.E08 -
OUT 6 filter on the retransmitted valueOUT 6 filter on the retransmitted value This parameter is available only when OUT 6 ("O6.Fn" [C.E04]) is configured “PV.rt”. Upper display: CnF.2 Lower display: O6.FL Range: From 0 (no filter) to 8 seconds
NOTENOTE
NOTE: this is a first order digital filter applied to the retransmit-
NOTENOTE ted output value.
C.E09 -C.E09 -
OUT 7 functionOUT 7 function
C.E09 -
OUT 7 function
C.E09 -C.E09 -
OUT 7 functionOUT 7 function Upper display: CnF.2 Lower display: O7.Fn Range: nonE = Output not used
ñAin = Main control output (linear) SECn = Secondary control output (linear) PV.rt = Process variable retransmission SP.rt = Operative set point retransmission
C.E10 - OUT 7 rangeC.E10 - OUT 7 range
C.E10 - OUT 7 range
C.E10 - OUT 7 rangeC.E10 - OUT 7 range This parameter is available only when Out 7 ("O7.Fn" [C.E09]) is different from "nonE". Upper display: CnF.2 Lower display: O7.rn Range: 0-20 = 0 to 20 mA
4-20 = 4 to 20 mA
29
C.E11 -C.E11 -
OUT 7 retransmission - initial scale valueOUT 7 retransmission - initial scale value
C.E11 -
OUT 7 retransmission - initial scale value
C.E11 -C.E11 -
OUT 7 retransmission - initial scale valueOUT 7 retransmission - initial scale value This parameter is available only when Out 7 function [C.E09] is configured as “PV.rt” or “SP.rt” Upper display: CnF.2 Lower display: O7.Lr Range: From -1999 to 9999 NOTENOTE
NOTE: The decimal point position will be the same as the
NOTENOTE
“CnF.1 - PV.SL” [C.D02] previously selected.
C.E12 -C.E12 -
OUT 7 retransmission - full scale valueOUT 7 retransmission - full scale value
C.E12 -
OUT 7 retransmission - full scale value
C.E12 -C.E12 -
OUT 7 retransmission - full scale valueOUT 7 retransmission - full scale value This parameter is available only when Out 7 function [C.E09] is configured as “PV.rt” or “SP.rt” Upper display: CnF.2 Lower display: O7.Hr Range: From -1999 to 9999 NOTENOTE
NOTE: The decimal point position will be the same as the
NOTENOTE
“CnF.1 - PV.SL” [C.D02] previously selected.
C.E13 -C.E13 -
OUT 7 filter on the retransmitted valueOUT 7 filter on the retransmitted value
C.E13 -
OUT 7 filter on the retransmitted value
C.E13 -C.E13 -
OUT 7 filter on the retransmitted valueOUT 7 filter on the retransmitted value This parameter is available only when OUT 7 ("O7.Fn" [C.E09]) is configured as “PV.rt”. Upper display: CnF.2 Lower display: O7.FL Range: From 0 (no filter) to 8 seconds NOTENOTE
NOTE: this is a first order digital filter applied to the retransmit-
NOTENOTE ted output value.
30
General note on configuration group 2General note on configuration group 2
General note on configuration group 2
General note on configuration group 2General note on configuration group 2 Exiting from the configuration group 2 the instrument automati­cally verifies the congruence of all parameters.
If a wrong setting is detected, the device will show:
The parameter group 2 passes the congruence test when:
1) No one of the 5 outputs is configured as control output.
2) Only one of the 5 outputs is configured as main output (“ñAin”).
3) Only one of the 5 outputs is configured as secondary output (“SECn”).
4) If only one control output is configured, it should be the main control output (“ñAin”). When changing menu, the following actions are performed:
1) The "CnF.4 - Añ.UL" [C.G04] parameter will be forced to "buñ" if its value is less than zero and only one control output has been configured.
2) The "Gr.4 - IP" [R.D05] parameter will be forced to 50.0 if only one control output has been configured and its valueis < 0.
CONFIGURATION GROUP 3 [C.Fxx]CONFIGURATION GROUP 3 [C.Fxx]
CONFIGURATION GROUP 3 [C.Fxx]
CONFIGURATION GROUP 3 [C.Fxx]CONFIGURATION GROUP 3 [C.Fxx] CONTROL OUTPUT CONFIGURATION
C.F01 - Main control output conditioningC.F01 - Main control output conditioning
C.F01 - Main control output conditioning
C.F01 - Main control output conditioningC.F01 - Main control output conditioning This parameter is available only when main control output is configured. Upper display: CnF.3 Lower display: ñC.Cn Range: norñ = The control output is as calculated by the PID.
CñPL = The control output is complemented (100-PID
calculated value).
C.F02 - Main control output in engineering unitC.F02 - Main control output in engineering unit
C.F02 - Main control output in engineering unit
C.F02 - Main control output in engineering unitC.F02 - Main control output in engineering unit This parameter is available only when main control output is configured. Upper display: CnF.3 Lower display: ñ.SCL Range: nO = Scalable is not required
YES = Scalable is required
31
NOTENOTE
NOTE: This scaling allows you to display the output value in
NOTENOTE engineering units instead of in percent.
C.F03 - Main output decimal point positionC.F03 - Main output decimal point position
C.F03 - Main output decimal point position
C.F03 - Main output decimal point positionC.F03 - Main output decimal point position This parameter will be available only when "ñ.SCL" ("Main control output in engineering units" [C.F02]) is set to "YES". Upper display: CnF.3 Lower display: ñC.dP Range:
---.
= No decimal figure
--.-
= One decimal figure
-.--
= Two decimal figures
C.F04 - Main control output initial scale valueC.F04 - Main control output initial scale value
C.F04 - Main control output initial scale value
C.F04 - Main control output initial scale valueC.F04 - Main control output initial scale value This parameter is available only when "ñ.SCL" ("Main control output in engineering units" [C.F02]) is set to "YES". Upper display: CnF.3 Lower display: ñC.E.L Range: from -199 to 999
C.F05 - Main control output full scale valueC.F05 - Main control output full scale value
C.F05 - Main control output full scale value
C.F05 - Main control output full scale valueC.F05 - Main control output full scale value This parameter is available only when "ñ.SCL" ("Main control output in engineering units" [C.F02]) is set to "YES". Upper display: CnF.3 Lower display: ñC.E.H Range: from -199 to 999
C.F06 - Main output auxiliary conditioningC.F06 - Main output auxiliary conditioning
C.F06 - Main output auxiliary conditioning
C.F06 - Main output auxiliary conditioningC.F06 - Main output auxiliary conditioning This parameter is available only when main control output is configured and “ñC.Cn” ("Main control output conditioning" [C.F01]) is different from “norñ”. Upper display: CnF.3 Lower display: ñC.A.C Range: bEFr = The functions listed at Note (A) are calculated
before before
before to apply the action selected by “ñC.Cn”
before before ("Main control output conditioning" [C.F01]) parameter.
AFtr = The functions listed at Note (A) are calculated
after after
after to apply the action selected by “ñC.Cn”
after after (“Main control output conditioning” [C.F01]) parameter.
Note (A)Note (A)
Note (A)
Note (A)Note (A)
- "Main control output limiters" - for more details see [r.E02] and
32
[r.E03] parameters.
- "Main control output max rate of rise" (see [r.E04]).
- "Control output display value" - for more details see "Display function" paragraph at pag. 43 and [C.F02], [C.F03], [C.F04], [C.F05] parameters.
- The control output value displayed by bargraph.
C.F07 - Secondary control output conditioningC.F07 - Secondary control output conditioning
C.F07 - Secondary control output conditioning
C.F07 - Secondary control output conditioningC.F07 - Secondary control output conditioning This parameter is available only when secondary control output is configured. Upper display: CnF.3 Lower display: SC.Cn Range: norñ = The control output is as calculated by the PID.
CñPL = The control output is complemented (100-PID
calculated value).
C.F08 -C.F08 -
Secondary control output in engineering unitSecondary control output in engineering unit
C.F08 -
Secondary control output in engineering unit
C.F08 -C.F08 -
Secondary control output in engineering unitSecondary control output in engineering unit This parameter is available only when secondary control output is configured. Upper display: CnF.3 Lower display: S.SCL Range: nO = Scalable is not required
YES = Scalable is required
NOTENOTE
NOTE: This scaling allows to display the output value in
NOTENOTE engineering units instead of in percent.
C.F09 -C.F09 -
Secondary control output decimal pointSecondary control output decimal point
C.F09 -
Secondary control output decimal point
C.F09 -C.F09 -
Secondary control output decimal pointSecondary control output decimal point
positionposition
position
positionposition This parameter is available only when "S.SCL" ("Secondary control output in engineering units" [C.F08]) is set to "YES". Upper display: CnF.3 Lower display: SC.dP Range:
---.
= No decimal figure
--.-
= One decimal figure
-.--
= Two decimal figures
C.F10 -C.F10 -
Secondary control output initial scale valueSecondary control output initial scale value
C.F10 -
Secondary control output initial scale value
C.F10 -C.F10 -
Secondary control output initial scale valueSecondary control output initial scale value This parameter will be available only when "S.SCL" ("Secondary control output in engineering units" [C.F08]) is set to "YES". Upper display: CnF.3 Lower display: SC.E.L Range: from -199 to 999
C.F11 -C.F11 -
Secondary control output full scale valueSecondary control output full scale value
C.F11 -
Secondary control output full scale value
C.F11 -C.F11 -
Secondary control output full scale valueSecondary control output full scale value This parameter will be available only when "S.SCL" ("Secondary control output in engineering units" [C.F08]) is set to "YES". Upper display: CnF.3 Lower display: SC.E.H Range: From -199 to 999
33
C.F12 -C.F12 -
Secondary control output auxiliary condition-Secondary control output auxiliary condition-
C.F12 -
Secondary control output auxiliary condition-
C.F12 -C.F12 -
Secondary control output auxiliary condition-Secondary control output auxiliary condition-
inging
ing
inging This parameter will be available only when secondary control output is configured and “SC.Cn” ("Secondary control output conditioning" [C.F07]) is different from “norñ”. Upper display: CnF.3 Lower display: SC.A.C Range: bEFr = The functions listed at Note (B) are calculated
before before
before to apply the action selected by “SC.Cn”
before before ("Secondary control output conditioning" [C.F07]) parameter.
AFtr = The functions listed at Note (B) are calculated
after after
after to apply the action selected by “SC.Cn”
after after (“Secondary control output conditioning” [C.F07]) parameter.
Note (B)Note (B)
Note (B)
Note (B)Note (B)
- "Secondary control output limiters" - for more details see [r.E06] and [r.E07] parameters.
- "Secondary control output max rate of rise" (see [r.E08]).
- "Control output display value" - for more details see "Display function" paragraph at pag. 43 and [C.F08], [C.F09], [C.F10], [C.F11] parameters.
- The control output value displayed by bargraph.
CONFIGURATION GROUP 4 [C.Gxx]CONFIGURATION GROUP 4 [C.Gxx]
CONFIGURATION GROUP 4 [C.Gxx]
CONFIGURATION GROUP 4 [C.Gxx]CONFIGURATION GROUP 4 [C.Gxx] AUXILIARY CONTROL PARAMETER CONFIGURATION
C.G01 - Smart functionC.G01 - Smart function
C.G01 - Smart function
C.G01 - Smart functionC.G01 - Smart function This parameter is available only when at least one control output is configured Upper display: CnF.4 Lower display: Sñ.Fn Range: dIS = Smart function disabled
Enb = Smart function may be enabled
C.G02 - Control action typeC.G02 - Control action type
C.G02 - Control action type
C.G02 - Control action typeC.G02 - Control action type This parameter will be available only when at least one control output is configured. Upper display: CnF.4 Lower display: Cn.tP Range: Pid = The process is controlled by PID actions
Pi = The process is controlled by PI actions
34
NOTENOTE
NOTE: if the control action type is changed, the control
NOTENOTE
parameter values (P, I, D) are updated with values calculated by the last TUNE function (first part of SMART algorithm). If these values are wrong, they are not updated and the E.120 error is displayed (for details see "Error messages" at pag. 83). When run time is activated, it will be necessary to enable a new TUNE procedure.
C.G03 - Manual functionC.G03 - Manual function
C.G03 - Manual function
C.G03 - Manual functionC.G03 - Manual function This parameter will be available only when at least one control output is configured Upper display: CnF.4 Lower display: ñAn.F Range: dIS = Manual function disabled
Enb = Manual function may be enabled
C.G04 -C.G04 -
C.G04 -
C.G04 -C.G04 ­transfertransfer
transfer
transfertransfer
Output value for AUTO to MANOutput value for AUTO to MAN
Output value for AUTO to MAN
Output value for AUTO to MANOutput value for AUTO to MAN
This parameter will be available only when at least one control output is configured and manual function is enabled (“ñAn.F” [C.G03] = “Enb”) Upper display: CnF.4 Lower display: Añ.UL Range: - from 0.0 to 100.0 % of the output span if device is
configured with one control output only
- from -100.0 to 100.0 % of the output span if device is configured with two control outputs.
Above the 100.0 % the display shows “buñ.” meaning that the transfer from AUTO to MANUAL is bumpless (the instrument sets for MANUAL mode the same power output used in AUTO mode)
C.G05 -C.G05 -
C.G05 -
C.G05 -C.G05 -
MANUAL to AUTO transfer typeMANUAL to AUTO transfer type
MANUAL to AUTO transfer type
MANUAL to AUTO transfer typeMANUAL to AUTO transfer type This parameter is available only when at least one control output is configured and manual function is enabled (“ñAn.F” [C.G03] = “Enb”) Upper display: CnF.4 Lower display: ñ.A.t.t Range: buñ. = Bumpless balance transfer.
buñ.b =Bumpless balanceless transfer.
35
NOTESNOTES
NOTES:
NOTESNOTES
1) During the manual to auto transfer, the instrument maintains the same power output as in manual mode.
2) Selecting the "buñ.b", the operative set point will be aligned to the measured value.
3) The “alignment” is not performed if measure is in error condition.
4) The selected local set point will be changed even if it is software protected.
Cnd.b = It starts in the same way it was left prior to
power shut down (if in manual mode the power output will be equal to the last value prior to power shut down).
C.G06 -C.G06 -
C.G06 -
C.G06 -C.G06 -
Device status at start upDevice status at start up
Device status at start up
Device status at start upDevice status at start up This parameter will be available only when at least one control output is configured and manual function is enabled (“ñAn.F” [C.G03] = “Enb”). Upper display: CnF.4 Lower display: St.Fn Range: Auto = It starts always in auto mode
ñan =It starts always in manual mode with power
output set to 0
Cnd.A = It starts in the same way it was left prior to
power shut down (if in manual mode the power output is set to 0)
36
CONFIGURATION GROUP 5 [C.Hxx]CONFIGURATION GROUP 5 [C.Hxx]
CONFIGURATION GROUP 5 [C.Hxx]
CONFIGURATION GROUP 5 [C.Hxx]CONFIGURATION GROUP 5 [C.Hxx] DIGITAL INPUT CONFIGURATION
C.H01 -C.H01 -
C.H01 -
C.H01 -C.H01 -
Digital input 1 functionDigital input 1 function
Digital input 1 function
Digital input 1 functionDigital input 1 function Upper display: CnF.5 Lower display: d1.Fn Range: nonE = Input contact not used
SP1.2 = Input contact used for SP /SP2 set point
selection.
SP3.4 = Input contact used for SP3/SP4 set point
selection.
Au.ñA = Input contact used for Auto/Manual selection
(Manual when logic level is “1”)
O.LIñ = Input contact used for output limiter activation
(Output limited when logic level is “1”).
ñ.rSt = Input contact used to reset (acknowledge) alarm
(Reset when logic level is “1”).
SP.L.r = Input contact used for Local/Remote set
point selection (Remote when logic level is “1”)
NOTENOTE
NOTE:
NOTENOTE
When one logic input is set to “SP.1.2” and no other logic input is set to “SP.3.4”, the relation between the logic level and the selected set point is the following:
Logic level 0 = SP
Logic level 1 = SP2 When one logic input is set to “SP.1.2” and a second logic input is set to “SP.3.4”, the relation between the logic levels and the selected set point is the following:
“SP.3.4” “SP.1.2” selected
level level set point
00SP 0 1 SP2 1 0 SP3
1 1 SP4 When one logic input is set to “SP.3.4” and no other logic input is set to “SP.1.2”, the relation between the logic level and the selected set point is the following:
Logic level 0 = SP Logic level 1 = SP3
37
C.H02 -C.H02 -
C.H02 -
C.H02 -C.H02 -
Digital input 1 contact statusDigital input 1 contact status
Digital input 1 contact status
Digital input 1 contact statusDigital input 1 contact status This parameter is available only when "d1.Fn" [C.H01] is different from “nonE”. Upper display: CnF.5 Lower display: d1.St Range: CLSd = The input is at logic level “1” when contact is
closed.
OPEn = The input is at logic level “1” when contact is
open.
C.H03 -C.H03 -
C.H03 -
C.H03 -C.H03 -
Digital input 2 functionDigital input 2 function
Digital input 2 function
Digital input 2 functionDigital input 2 function Upper display: CnF.5 Lower display: d2.Fn Range: nonE = Input contact not used.
SP1.2 = Input contact used for SP /SP2 set point
selection.
SP3.4 = Input contact used for SP3/SP4 set point
selection.
Au.ñA = Input contact used for Auto/Manual selection
(Manual when logic level is “1”).
O.LIñ = Input contact used for output limiter activation
(Output limited when logic level is “1”).
ñ.rSt = Input contact used to reset (acknowledge) alarm
(Reset when logic level is “1”).
SP.L.r = Input contact used for Local/Remote set point selection (Remote when logic level is
“1”)
NOTENOTE
NOTE:
NOTENOTE
When one logic input is set to “SP.1.2” and no other logic input is set to “SP.3.4”, the relation between the logic level and the selected set point is the following:
Logic level 0 = SP
Logic level 1 = SP2 When one logic input is set to “SP.1.2” and a second logic input is set to “SP.3.4”, the relation between the logic levels and the selected set point is the following:
“SP.3.4” “SP.1.2” selected
level level set point
00SP 0 1 SP2 1 0 SP3
1 1 SP4 When one logic input is set to “SP.3.4” and no other logic input is set to “SP.1.2”, the relation between the logic level and the selected set point is the following:
Logic level 0 = SP Logic level 1 = SP3
38
C.H04 -C.H04 -
C.H04 -
C.H04 -C.H04 -
Digital input 2 contact statusDigital input 2 contact status
Digital input 2 contact status
Digital input 2 contact statusDigital input 2 contact status This parameter will be available only when "d2.Fn" [C.H03] is different from “nonE”. Upper display: CnF.5 Lower display: d2.St Range: CLSd = The input is at logic level “1” when contact is
closed.
OPEn = The input is at lo gic l evel “1” when contact is
open.
General notes on configuration group 5General notes on configuration group 5
General notes on configuration group 5
General notes on configuration group 5General notes on configuration group 5
1) Digital input function selection should be “consistent” with other configuration parameters otherwise this digital input will remain unused.
2) Changing menu the device will check that the same function is not assigned to more than one digital input, otherwise the display will show: Upper display: CnF.5 Middle display: Err Lower display: d.InP
C.H05 -C.H05 -
C.H05 -
C.H05 -C.H05 -
Digital input 3 contact statusDigital input 3 contact status
Digital input 3 contact status
Digital input 3 contact statusDigital input 3 contact status Used to carry out the Burn Off procedure. See on page 50 for details. Upper display: CnF.5 Lower display: d3.St Range: CLSd = The input is at logic level “1” when contact is
closed
OPEn = The input is at lo gic l evel “1” when contact is
open
39
CONFIGURATION GROUP 6 [C.Ixx]CONFIGURATION GROUP 6 [C.Ixx]
CONFIGURATION GROUP 6 [C.Ixx]
CONFIGURATION GROUP 6 [C.Ixx]CONFIGURATION GROUP 6 [C.Ixx] OTHERS CONFIGURATION PARAMETERS
C.I01 -C.I01 -
Green bargraph functionGreen bargraph function
C.I01 -
Green bargraph function
C.I01 -C.I01 -
Green bargraph functionGreen bargraph function Upper display: CnF.6 Lower display: G.brG Range: Pr.Ur = the green bargraph will show the process
variable value.
DEV. = the green bargraph will show the deviation
(process value minus set point).
C.I02 -C.I02 -
Orange bargraph functionOrange bargraph function
C.I02 -
Orange bargraph function
C.I02 -C.I02 -
Orange bargraph functionOrange bargraph function Upper display: CnF.6 Lower display: O.brG Range: OP.SP= The orange bargraph will show the operative set
point value.
P.Out. = The orange bargraph will show the process
output value.
C.I03 -C.I03 -
Bargraph initial scale valueBargraph initial scale value
C.I03 -
Bargraph initial scale value
C.I03 -C.I03 -
Bargraph initial scale valueBargraph initial scale value This parameter is available when “G.brG” [C.I01] is equal to “Pr.Ur” and/or “O.brG” [C.I02] is equal to “OP.SP”. Upper display: CnF.6 Lower display: brG.L Range: span limits of "CnF.1- PV.SL" [C.D02].
C.I04 -C.I04 -
Bargraph full scale valueBargraph full scale value
C.I04 -
Bargraph full scale value
C.I04 -C.I04 -
Bargraph full scale valueBargraph full scale value This parameter is available when “G.brG” [C.I01] is equal to “Pr.Ur” and/or “O.brG” [C.I02] is equal to “OP.SP”. Upper display: CnF.6 Lower display: brG.H Range: span limits of "CnF.1- PV.SL" [C.D02].
C.I05 -C.I05 -
Resolution of the deviation bargraphResolution of the deviation bargraph
C.I05 -
Resolution of the deviation bargraph
C.I05 -C.I05 -
Resolution of the deviation bargraphResolution of the deviation bargraph This parameter is available when “G.brG” [C.I01] is equal to “DEV.”. Upper display: CnF.6 Lower display: brG.d Range: 1-2-5-10-20 or 50 = Digits per segment
40
C.I06 -C.I06 -
Set point display typeSet point display type
C.I06 -
Set point display type
C.I06 -C.I06 -
Set point display typeSet point display type Upper display: CnF.6 Lower display: SP.dS Range: Fn.SP = When device is in normal display mode and it is
performing a set point ramp, the middle display will show the final set point value.
OP.SP= When device is in normal display mode and it is
performing a set point ramp, the middle display will show the operative set point value.
C.I07 -C.I07 -
Burn Off and Purge activation thresholdBurn Off and Purge activation threshold
C.I07 -
Burn Off and Purge activation threshold
C.I07 -C.I07 -
Burn Off and Purge activation thresholdBurn Off and Purge activation threshold Upper display: CnF.6 Lower display: t.t.Ac Range: the instrument will start the Burn Off and the Purge
procedure according with the following selection:
No = without testing the temperature limit (1000 °F) YES= if the temperature test is satisfy.
::
Available
: When mV is selected as primary control variable.
::
C.I08 -C.I08 -
Time out selectionTime out selection
C.I08 -
Time out selection
C.I08 -C.I08 -
Time out selectionTime out selection Upper display: CnF.6 Lower display: t.out Range: tñ.10 = 10 s time out
tñ.30 = 30 s time out
CONFIGURATION GROUP [C.Lxx]CONFIGURATION GROUP [C.Lxx]
CONFIGURATION GROUP [C.Lxx]
CONFIGURATION GROUP [C.Lxx]CONFIGURATION GROUP [C.Lxx] INPUT/OUTPUT TEST
This group is displayed only when configuration modify mode is enabled. This group allows you to verify the correct working of:
- the relay outputs (OUT 1 up to OUT5),
- digital inputs (diG.1 up to diG.3),
- the inputs from In1 up to In8 and the relays of the outputs from OUT10 up to OUT19, if the optional I/O card is fitted.
When an output is in "On" status, the related LED will light up (for OUT 5 is used the "ST" LED)
C.L01 OUT 1 statusC.L01 OUT 1 status
C.L01 OUT 1 status
C.L01 OUT 1 statusC.L01 OUT 1 status Upper display: VErF Lower display: OU.1 Range: On = Output enabled (relay energized)
OFF = Output disabled (relay deenergized)
41
NOTE:NOTE:
NOTE: By pushing the or key it is possible to change
NOTE:NOTE:
the output state. The setting above described is the same for the remaining outputs.
C.L02 OUT 10 statusC.L02 OUT 10 status
C.L02 OUT 10 status
C.L02 OUT 10 statusC.L02 OUT 10 status This parameter is available only if the optional card is fitted. Upper display: VErF Lower display: OU.10 Range: On = Output enabled (relay energized)
OFF = Output disabled (relay deenergized)
NOTE:NOTE:
NOTE: By pushing the or key it is possible to change
NOTE:NOTE:
the output state. The setting above described is the same for the remaining outputs.
C.L04 Digital input In.1 statusC.L04 Digital input In.1 status
C.L04 Digital input In.1 status
C.L04 Digital input In.1 statusC.L04 Digital input In.1 status This parameter is available only if the optional card is fitted. Upper display: VErF Lower display: In.1 Range: Open = Input open
CLSd = Input closed
NOTE:NOTE:
NOTE:The setting above described is the same for the
NOTE:NOTE:
remaining inputs.
General note for configuration group "VErF"General note for configuration group "VErF"
General note for configuration group "VErF"
General note for configuration group "VErF"General note for configuration group "VErF" Returning in the "run time mode", the instrument restarts to control the process, the settings performed into this group do not affect the working settings.
C.L03 Digital input dIG.1C.L03 Digital input dIG.1
C.L03 Digital input dIG.1
C.L03 Digital input dIG.1C.L03 Digital input dIG.1
statusstatus
status
statusstatus Upper display: VErF Lower display: dIG.1 Range: Open = Input open
CLSd = Input closed
NOTE:NOTE:
NOTE:The setting above described is the same for the
NOTE:NOTE:
remaining inputs.
42
CONFIGURATION GROUP END [C.Mxx]CONFIGURATION GROUP END [C.Mxx]
CONFIGURATION GROUP END [C.Mxx]
CONFIGURATION GROUP END [C.Mxx]CONFIGURATION GROUP END [C.Mxx] END CONFIGURATION This group is displayed only when configuration modify mode is enabled. Upper display: ConF Lower display: End Range: nO =with this selection the instrument comes back
to the first display of the configuration modify mode.
YES =this selection ends the configuration modify
mode; the instrument preforms an automatic
reset and restarts in the run time mode. Pushing ”” or “” push-button select the desired action and then push “MENU” push-button.
RUN TIME MODERUN TIME MODE
RUN TIME MODE
RUN TIME MODERUN TIME MODE If V101 is correctly set (see "Operative mode and hardware lock" at page 16), and the configuration mode has been terminated, the instrument will be in run time mode and in "normal display mode" (see below). During run time mode the instrument performs the loop control and manages all the instrument functions (SMART, ALARMS, etc.).
DISPLAY FUNCTIONDISPLAY FUNCTION
DISPLAY FUNCTION
DISPLAY FUNCTIONDISPLAY FUNCTION When device is powered and it is operating in automatic mode, the upper display shows the process variable, the middle display shows the set point value (final or operative according to “SP.dS” [C.I06] parameter setting). We define the above condition as “normal display mode”.
The lower display will show: a) the control output value.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) If the instrument is configured with one control output (linear or time proportional), the control output value is displayed in the 0 to 100% range.
43
2)if the device is configured with two control outputs; the value of the MAIN output is shown on the two most
significant digits, while the value of the SECONDARY output is shown on the two less significant digits.
The decimal point between the two values will be flashing. The graphic symbol “ “ shows that the specific
control output is > 100%.
b) Pushing the FUNC push-button; it is possible to display the
sensor output in mV, the "mV" LED lights up. NOTE:NOTE:
NOTE: this selection is available:
NOTE:NOTE:
- if "mV" is not selected as controlled variable,
- even if a measure error is shown on the upper display,
- even if Burnoff or Probe test procedure are in progress.
c) Pushing FUNC push-button again, it is possible to display the
sensor temperature, the "TP" LED lights up.
d)Pushing FUNC push-button again, it is possible to display the
Dew point value, the "DP" LED lights up. NOTE:NOTE:
NOTE: this selection is available:
NOTE:NOTE:
- if Dew point or O2 are not selected as controlled variable,
- even if a measure error is shown on the upper display,
- even if Burnoff or Probe test procedure are in progress.
e) Pushing FUNC push-button again, it is possible to display the
Carbon potential, the "CP" LED lights up. NOTE:NOTE:
NOTE: this selection is available:
NOTE:NOTE:
- if Carbon potential or O2 are not selected as controlled variable,
- even if a measure error is shown on the upper display,
- even if a Burnoff or Probe test procedure is in progress.
f)Pushing FUNC push-button again, it is possible to display the
Carbon monoxide factor value, the "COF" LED lights up.
g) Pushing FUNC push-button again, it is possible to display the
Hydrogen factor value, the "H2F" LED lights up.
h) Pushing FUNC push-button again, it is possible to display the
Probe resistance value, the "K" LED lights up (the graphic symbol " " shows that the Probe resistance value is greater than 99.99 KΩ).
i) Pushing FUNC push-button again, it is possible to display the
Probe response time, the "SEC" LED lights up.
l) Pushing FUNC push-button again, it is possible to display the
Probe temperature at last probe test, the "LT" LED lights up.
m)Pushing FUNC push-button again, it is possible to display the
Carbon monoxide value, the display will show "CO." followed by the value measured on the auxiliary input (the graphic symbol " " shows that its value is more than 99).
44
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) The CO. value will be set to 20 if the auxiliary input is out of range.
2) This indication is available only if the auxiliary input is configured as CO input.
n)Pushing FUNC push-button again, the display will show "ñ.",
followed by the MAIN control output value with a range scaled as configured by "CnF.3 - ñ.C.EL" [C.F04] and "CnF.3 ­ñ.C.EH" [C.F05] parameters. NOTE:NOTE:
NOTE: this selection is available only if a scaled MAIN output
NOTE:NOTE: is selected.
o)Pushing FUNC push-button again, the display will show "S.",
followed by the SECONDARY control output value with a range scaled as configured by "CnF.3 - S.C.EL" [C.F10] and "CnF.3 - S.C.EH" [C.F11] parameters. NOTE:NOTE:
NOTE: this selection is available only if two outputs are set
NOTE:NOTE:
and a scaled SECONDARY control output is configured. Pushing FUNC push-button again, the control output value is displayed.
All the visualization types described (exception made for "mV" and "TP") are bound by time out (see “t.out” [C.I08] parameter), at the end of the time out the display will return to show the control output value.
NOTESNOTES
NOTES:
NOTESNOTES
1) The previous information will be displayed only if the relative function has previously been configured.
2) After parameter modification or monitoring, the instrument comes back to the "normal display mode" with, on the lower display, the indication previously selected.
INDICATORSINDICATORS
INDICATORS
INDICATORSINDICATORS Two green LEDs located on the left of the middle display;
CP Lit when the carbon potential is the controlled variable. DP Lit when the dew point is the controlled variable.
Two red LEDs located on the left of the lower display; mV Lit when the sensor output (in mV) is the controlled
variable.
MAN Lit when device is in manual mode.
45
Fourteen red LEDs located down the lower display; DP Lit when the dew point value is shown on the lower
display.
mV Lit when the sensor output (in mV) is shown on the lower
display.
TP Lit when the sensor temperature is shown on the lower
display.
KΩ Lit when the probe resistance value is shown on the lower
display available if O2 is not selected as primary control variable.
CP Lit when the carbon potential value is shown on the lower
display.
SEC Lit when the sensor response time is shown on the lower
display available if O2 is not selected as primary control variable.
COF Lit when the carbon monoxide factor value is shown on the
lower display available if O2 is not selected as primary control variable.
H2F Lit when hydrogen factor value is shown on the lower
display available if O2 is not selected as primary control variable.
LT Lit when the probe temperature of the last probe test is
shown on the lower display available if O2 is not selected as primary control variable.
1 Lit when:
- OUT 1 is used as control output and it is in ON condition;
- alarm 1 is in alarm state and acknowledged.
Flashes when alarm 1 is in alarm state and it is not acknowledged.
2 Lit when:
- OUT 2 is used as control output and it is in ON condition;
- alarm 2 is in alarm state and acknowledged.
Flashes when alarm 2 is in alarm state and it is not acknowledged.
3 Lit when:
- OUT 3 is used as control output and it is in ON condition;
- alarm 3 is in alarm state and acknowledged.
Flashes when alarm 3 is in alarm state and it is not acknowledged.
ST Flashes when the first step of SMART algorithm is
activated. Lit when the second step of SMART algorithm is activated.
RM Lit when the device is controlled by serial link.
46
NOTE:NOTE:
NOTE: an additional indication is available, the decimal point
NOTE:NOTE: shown below is:
- flashing when the set point is controlled by the serial link;
- steady lit when the auxiliary set point (SP2 or SP3 or SP4 or remote set point (RSP)) is used.
This indication will be turned off during monitor and modify parameter procedures.
Decimal point
Optional input/output indicatorsOptional input/output indicators
Optional input/output indicators
Optional input/output indicatorsOptional input/output indicators For optional inputs the LED's In1 to In4 are lit when the contact of the related input is closed. No indication is available for inputs In5 to In8. For optional outputs the LED's from OUT10 to OUT19 are lit when the related relay is energized.
BARGRAPH DESCRIPTIONBARGRAPH DESCRIPTION
BARGRAPH DESCRIPTION
BARGRAPH DESCRIPTIONBARGRAPH DESCRIPTION This device includes two 33 LED’s bargraphs indicators. On the right bargraph, with orange LED’s, it is possible to show:
- the operative set point scaled as configured by “brG.L” [C.I03] and “brG.H” [C.I04]; the “SP” LED, located on the base of the bargraph, is lighted. If the operative set point is out of the range scaled, the first or the last LED's will flash.
- the power output value; the “%” LED, located on the top of the bargraph, is lighted. When the device has two process outputs, the central LED (which is a two-colored LED) becomes green. It divides the bargraph into two parts: the MAIN process output is shown on the upper part while the SECONDARY process output is shown on the lower part.
On the left bargraph, with green LED’s, it is possible to show:
- the process variable scaled as configured by “brG.L” [C.I03] and “brG.H” [C.I04]; the “PV” LED, located on the base of the bargraph, is lighted.
- The deviation error (PV - SP) with resolution configured at “brG.d” [C.I05]; the “DEV” LED, located on the top of the bargraph, is lighted.
47
When the deviation error indication is selected, the central LED (which is a two-colored LED) becomes orange. It devides the bargraph in two parts: the positive error is shown on the upper part while the negative error is shown on the lower part. If the process variable or the deviation error are out of the range, the first or the last LED's will flash.
NOTENOTE
NOTE: The upper or lower bargraph LED's will flash if the
NOTENOTE
displayed value is respectively higher or lower than the range scaled for the specific bargraph visualization.
DIRECT ACCESS TO THE SET POINTDIRECT ACCESS TO THE SET POINT
DIRECT ACCESS TO THE SET POINT
DIRECT ACCESS TO THE SET POINTDIRECT ACCESS TO THE SET POINT When the device is in AUTO mode and in “Normal Display Mode”, it is possible to access directly to the selected set point modification (SP, SP2, SP3 or SP4). Pushing or for more than 2 s, the set point will begin changing. The new set point value becomes operative when the instrument returns in “Normal Display Mode” (at the end of 2 s time out). The direct access to the set point is disabled if the run time group 1 is software protected or RSP is selected.
MANUAL FUNCTIONMANUAL FUNCTION
MANUAL FUNCTION
MANUAL FUNCTIONMANUAL FUNCTION The MANUAL mode function can be accessed (only if configured) by depressing the MAN push-button for more than 1 sec or by acting on a logic input configured as "Au.ñA" (auto/manual selection). The command from keyboard is accepted and executed only if the instrument is in "Normal Display Mode" while the command from logic input is always accepted. When in MANUAL mode the LED's MAN lights up while the middle display shows "ñ." followed by the main output values (from 0.0 to 99.9%) and the lower display shows "S." followed by the secondary output value (from 0.0 to 99.9%). NOTENOTE
NOTE: the graphic symbol " " shows that the related
NOTENOTE control output value is > 100.0%.
The control output value can be modified by using “”and “” keys.
To come back to the AUTO mode, depress “MAN” push-button for more than 1 sec. or toggle the logic input. The transfer from AUTO to MANUAL is bumpless (this function is not provided if an output value has been selected at configuration parameter “Añ.UL” [C.G04]). The transfer from MANUAL to AUTO is bumpless or bumpless balanceless (the bumpless feature is not provided if integral action is excluded).
48
If transfer from AUTO to MANUAL is performed during the first part of SMART algorithm (TUNE) the SMART will be aborted and when returning to AUTO, it will be necessary to start the SMART algorithm again. If transfer from AUTO to MANUAL is performed during the second part of SMART algorithm (ADAPTIVE), when returning to AUTO, the instrument will automatically restore the ADAPTIVE algorithm.
At power up the device starts as configured at “St.Fn” [C.G06] configuration parameter.
NOTESNOTES
NOTES:
NOTESNOTES
1) When auto/manual control mode is selected by logic input, at power up the instrument will start in accordance with the logic input status.
2) If the logic input status requires the manual mode, the instrument will verify the “St.Fn” [C.G06] parameter setting and if “St.Fn” [C.G06] is equal to "Auto", "ñan" or “Cnd.A” the power output will be set to zero otherwise the power output will be set equal to the last value (in manual mode) prior to power shut down.
SPECIAL OUTPUT FEATURESSPECIAL OUTPUT FEATURES
SPECIAL OUTPUT FEATURES
SPECIAL OUTPUT FEATURESSPECIAL OUTPUT FEATURES The control action is deactivated (control outputs forced to zero) and alarms are forced in non alarm condition when: a) The set point is set to minimum value,
0.00 when the controlled variable is Carbon Potential; 0 when the controlled variable is mV;
-100 °F (-75 °C) when the controlled variable is Dew Point.
b) The probe output is lower than 900 mV or greater than
1300 mV and the controlled variable is Carbon Potential or Dew Point.
c) The probe temperature is lower than 900 °F (482 °C) or
greater than 2100 °F (1149 °C) and the controlled variable is Carbon Potential or Dew Point.
NOTE:NOTE:
NOTE: when condition b) or c) occurs (independently from the
NOTE:NOTE:
primary variable selected); the Carbon Potential is forced to 0.00 and the Dew Point is forced to 250 °F (121 °C).
49
BURN OFF PROCEDUREBURN OFF PROCEDURE
BURN OFF PROCEDURE
BURN OFF PROCEDUREBURN OFF PROCEDURE
(Available only if “O2” is not selected as primary control variable)
The Burn off function allows to remove the carbon or soot deposit inside the probe. It is carried out by putting air into the probe; the oxygen present in the air will burn the carbon deposit. The Purge function is used to evacuate the not combusted air present in the probe after the Burn off (for details see the probe instruction manual). The Burn off procedure may be carried out:
- manually (by pressing TST + MAN keys), or
- by external contact (Dig.3), or
- by serial link, or
- automatically, after a period (set by the "bF.tr" [R.H03] parameter) started from the end of the previous Burn off procedure.
The Burn off procedure cannot be carried out if:
1) The first part of the SMART algorithm (TUNE) is in progress.
2) Five minutes are not elapsed from instrument power on or from
the end of another procedure (Burn off or Probe test).
3) The probe temperature is less than 1000 °F (538 °C) or in
error condition. NoteNote
Note: the temperature value will not be taken into account if
NoteNote mV is the selected controlled variable and "t.t.Ac" parameter [C.I07] is equal to "nO".
NOTES:NOTES:
NOTES:
NOTES:NOTES: a) If one or more conditions previously described still exist the
automatic Burn off will be frozen, the state of these conditions are checked by the instrument every minute. During this time, the instrument operates as previously set before the Burn off request. When the checking will be successful, these conditions no longer exist and the automatic Burn off will be immediately performed. The timer programmed to perform the next Burn off will start only at the end of the previous Burn off.
b) If the procedure is requested by keys and it cannot be started,
the following information is displayed for 2 seconds: Upper display: burn Middle display:inh Lower display: Number of the first not complied condition (1, 2 or 3 previously described). Once these conditions no longer exist, repeat manually the Burn off starting command.
During the Burn off procedure, the upper display shows "burn" or "PurG", while the middle display shows the remaining time to end the related procedure. The lower display shows the probe value in mV, the information shown on this display may be changed as described in the "Display function"paragraph (see pag. 43).
50
The Burn off sequence is the following:
- The Burn off output (OUT 4)will be turned ON for a programma­ble period (by "Gr. 8 - "t.bOF" [R.H01] parameter).
- At the end of this period, OUT 4 will be turned OFF and the Purge output (OUT 5) will be turned ON for a programmable period (by "Gr. 8 - "t.PrG" [R.H02] parameter).
- At the end of this period, OUT 5 will be turned OFF, the procedure is completed and the normal control is resumed.
NOTES:NOTES:
NOTES:
NOTES:NOTES: a)During the Burnoff and Purge procedures the control output,
the Adaptive algorithm and alarm status are frozen.
b)During the Burnoff and Purge procedures the input span 0 to
1500mV will be enabled, independently of the primary controlled variable selected.
c)If a power down occurs, the current Burn off status and timings
will be lost (if automatic burn off was set, the timer will start again from the value set by the "bF.tr" [R.H03] parameter).
Status
Purge OUT5
Burnoff OUT4
Controlled variable (% CP)
On
Off On
Off
Burn off and purgeBurn off and purge
Burn off and purge
Burn off and purgeBurn off and purge descriptiondescription
description
descriptiondescription
Burnoff timer
Instrument in AUTO mode
Instrument functions kept frozen
Purge
Time
timer
Instrument in AUTO mode
Automatic Burnoff interval "bF.tr" parameter
Acting by keys, external contact or serial link
51
Automatic Burnoff interval "bF.tr" parameter starts again
Time
PROBE TEST PROCEDUREPROBE TEST PROCEDURE
PROBE TEST PROCEDURE
PROBE TEST PROCEDUREPROBE TEST PROCEDURE
(Available only if “O2” is not selected as primary control variable)
The Probe test may be carried out:
- manually (by pressing TST + FUNC keys), or
- from serial link, or
- automatically, after a period (set by the "Pb.tr" [R.H04] parameter) started from the end of the previous Probe test.
The Probe test cannot be carried out if:
1) the first part of the SMART algorithm (TUNE) is in progress;
2) five minutes are not elapsed from instrument power on or from
the end of another procedure (Burn off, Purge or Probe test);
3) The probe temperature is less than 1000 °F (538 °C) or in
error condition. NoteNote
Note: the temperature value will not be taken into account if
NoteNote mV is the selected controlled variable and "t.t.Ac" parameter [C.I07] is equal to "nO".
4) the probe output is less than 1000 mV;
5) the probe output is unstable (drift more than 10 mV/minute).
If one or more conditions previously described still exist the automatic Probe test will be frozen, the state of these conditions are checked by the instrument every minute. During this time, the instrument operates as previously set before the Burn off request.
When the checking will be successful, these conditions no longer exist and the automatic Probe test will be immediately performed. The timer programmed to perform the next Probe test will start only at the end of the previous Probe test. If the Probe test is requested by keys and it cannot be started, the following information is displayed for 2 seconds: Upper display: Prob Middle display:inh Lower display: Number of the first not complied condition (1 up to
5 previously described).
During the Probe test, the instrument will operate as follows:
1) it shows on the upper display: Prob on the middle display:tESt
2) It connects a 47.5 Kshunt across the probe terminals and
waits for 15 seconds in order to stabilize the measure.
3) It calculates the probe resistance Rp by applying the following
formula: Rp (in KΩ) = (Ec/Es - 1) • 47.5
where:
Ec is the sensor output (in mV) measured before the shunt
connection.
52
Es is is the sensor output (in mV) measured after the shunt
connection.
4) It removes the 47.5 Kshunt resistor.
5) It displays the calculated value on the lower display and stores it in a memory location for later recall.
Probe response timeProbe response time
Probe response time
Probe response timeProbe response time The probe response time is the time which takes the mV measured value (after the shunt is removed) to return to the original value before the shunt connection (+ 4 mV). The probe response time is stored in a memory location for later recall. If the probe response time is greater than 60 seconds, the probe response time calculation will be aborted.
NOTES:NOTES:
NOTES:
NOTES:NOTES: a)During the Probe test, the control output, the Adaptive
algorithm and alarm status are frozen. All functions will be automatically restored at the end of this procedure.
b)If a power down occurs, the current probe test status and
timings will be lost (if automatic probe test was set, the timer will start again from the value set by the "Pb.tr" [R.H04] parameter).
SERIAL LINKSERIAL LINK
SERIAL LINK
SERIAL LINKSERIAL LINK The device can be connected to a host computer by a serial link. The host can put the device in LOCAL (functions and parameters are controlled via keyboard) or in REMOTE (functions and parameters are controlled via serial link). The REMOTE status is signalled by the RM LED. This instrument allows you to modify the operative and configura­tion parameters, via serial link. The necessary conditions to implement this function are the following:
1) the serial parameters (see Run time group 7) should be properly configured.
2) the device must be in the run time mode.
3) when it is desired to modify configuration parameter, the dip switch V101.3 should be set to "OFF" (configuration parameter enabled).
During configuration parameter downloading the device goes in open loop with all output in OFF state. At the end of configuration procedure, the device performs an automatic reset and then returns to close loop control.
53
LAMP TESTLAMP TEST
LAMP TEST
LAMP TESTLAMP TEST When it is desired to verify the display efficiency, put the instrument in "normal display mode" and push + MENU push­buttons for more than 5 s. The instrument will turn ON, with a 50 % duty cycle, all the LEDs of the display (we define this function "LAMP TEST"). No time out is applied to the LAMP TEST. When it is desired to revert to the normal display mode, push any push-button. During the LAMP TEST the instrument continues to control the process.
SMART FUNCTIONSMART FUNCTION
SMART FUNCTION
SMART FUNCTIONSMART FUNCTION It is used to automatically optimize the control action. To enable the SMART function proceed as follows:
1) push the MENU push-button until "Gr. 2" run time parameter group is shown.
2) push the FUNC push-button until "Sñrt" parameter is shown.
3) push or and set the "On" indication on the middle display.
4) push the FUNC push-button.
The ST LED is flashing during the first part of SMART algorithm (TUNE), while is steady lit during the second one (ADAPTIVE). When the smart function is enabled, it is possible to display but not to modify the control parameters. To disable the SMART function proceed as follows:
1) push the MENU push-button until "Gr. 2" run time parameter group is shown.
2) push the FUNC push-button until "Sñrt" parameter is shown.
3) push or and set the "OFF" indication on the middle display.
4) push the FUNC push-button.
The ST LED will turn off. The instrument maintains the current set of control parameters and it enables parameter modification. If a power down occurs and the instrument was performing the first part of SMART algorithm (TUNE), at the next power up, all
54
data regarding the TUNE will be lost, so it will be necessary to re-start the SMART function. If a power down occurs during the second part of SMART algorithm (ADAPTIVE), at the next power up, the instrument restarts the ADAPTIVE function.
CONTROL ACTION FEATURECONTROL ACTION FEATURE
CONTROL ACTION FEATURE
CONTROL ACTION FEATURECONTROL ACTION FEATURE The control action is de-activated (control output forced to 0) and alarms forced in no alarm condition when: a) The setpoint is set at minimum value (0.00 for CP/0 for mV/­100°F (-75°C) for Dew Point/0.0 for O2) b) The remote set point( if selected) is out of range and “
A.I.Añ = norñ.” c) The probe output is lower than 900 mV or greater than 1300 mV and “Carbon potential or “Dew Point” are configured as primary control variable. When “O2” is configured as primary control variable the probe mV limits are -15 / 200 mV d) The probe temperature is lower than 900 °F (482 °C) or greater than 2100 °F (1149 °C) and “Carbon potential or Dew Point” are configured as primary control variable. When “O2” is configured as primary control variable the temperature limits are 1200 /2900 °F (650/1590 °C) Note:Note:
Note: When conditions c) or d) occur (independently from
Note:Note: primary control variable selection), the Carbon potential, Dew Point and O2 are not calculated but forced respectively to 0.00 ,250 °F (121 °C) or 0.0
55
COF AND H2F ADJUSTMENT DIAGRAMCOF AND H2F ADJUSTMENT DIAGRAM
COF AND H2F ADJUSTMENT DIAGRAM
COF AND H2F ADJUSTMENT DIAGRAMCOF AND H2F ADJUSTMENT DIAGRAM Others gas present in the atmosphere can affect the carbon potential or the dew point measurement. By this chart it is possible to get the COF or H2F factors able to correct the variation above described. If, for example, the instrument measures 0.4% of carbon with a COF of 200 and the shim stock analysis shows an actual carbon value of
0.35%, draw a line between 0.4% and 200 and mark the intersection on the pivot line. Connect 0.35 of % Carbon with that intersection and extend it till the COF scale (the new COF value will be approximately 175). Correct the COF [R.A05] parameter with the new value. NOTE:NOTE:
NOTE: COF is equal to 10 • %CO and
NOTE:NOTE: H2F is 10 • %H2.
COF or H2F
PIVOT
DEW
POINT
56
% CARBON
Nominal COF
in 40°F endo
Nominal H2F
in 40°F endo
PARAMETER PROTECTIONPARAMETER PROTECTION
PARAMETER PROTECTION
PARAMETER PROTECTIONPARAMETER PROTECTION Moving from one run time parameter group to another, the instrument will show, first, the group number (on the upper display), the mnemonic code (on the lower display) and the security status (on the middle display) of the group. The possible security status are: Enb =the parameters of this group are not protected and they
may be modified;
inh = the parameters of this group are always protected and
they cannot be modified;
- - - - =the parameters of this group are software protected.
In this case, by using ““ and “” push-button, set a value equal to the security code programmed during the "Run time security code " (see page 18) and then push “FUNC”. If the programmed value matches the security code, the parameters of this group can be modified otherwise they will remain protected.
CAUTIONCAUTION
CAUTION: A software key protects one or more run time
CAUTIONCAUTION parameter groups so that when the software key is set, it is possible to modify the run time parameters of all groups. The software protection will be automatically restored when the instrument reverts to the "normal display mode" (by "MENU" push­button pressure or by the time out).
RUN TIME PARAMETER MODIFICATIONRUN TIME PARAMETER MODIFICATION
RUN TIME PARAMETER MODIFICATION
RUN TIME PARAMETER MODIFICATIONRUN TIME PARAMETER MODIFICATION By “MENU” push-button it is possible to start the run time
parameter modification by selecting desired run time parameter group. During the run time parameter modification, the upper display shows the selected parameter group, the lower display shows the mnemonic code of the selected parameter while the middle display shows the value or status assigned to the selected parameter. The parameter monitoring and modification are bound by time out (see “t.out” [C.I08] parameter), after that, the display returns in “Normal display mode” and the modification (if carried out) of the last displayed parameter will be lost.
NOTESNOTES
NOTES:
NOTESNOTES
1) In the following pages we will describe all the parameter of the instrument but it will show only the parameters related with the specific hardware and in accordance with the specific instrument configuration.
2) If all the parameters of a group are not available, the entire group will not be displayed.
57
3) For an easy consultation of this manual, a sheet named "Reference parameter guide" with all the parameter visualizations is enclosed. The groups of run time parameters are identified by the "R" letter followed by A, B etc. The "code" formed by the group and the row (i.e. R.A01 where "R.A" is run time group1 and "01" is the row 1) is reported, in the user manual, before each parameter description and allows you to quickly find out the respective parameter.
Run time group 1 [R.Axx]Run time group 1 [R.Axx]
Run time group 1 [R.Axx]
Run time group 1 [R.Axx]Run time group 1 [R.Axx] Set point and factor values
R.A01 -R.A01 -
R.A01 -
R.A01 -R.A01 -
Main set pointMain set point
Main set point
Main set pointMain set point Upper display: Gr.1 Lower display: SP Range: from "rL" [R.E10] to "rH" [R.E11].
R.A02 - Set point 2R.A02 - Set point 2
R.A02 - Set point 2
R.A02 - Set point 2R.A02 - Set point 2 This parameter is available only when one of the logic inputs is configured for SP/SP2 selection Upper display: Gr.1 Lower display: SP2 Range: from "rL" [R.E10] to "rH" [R.E11].
58
R.A03 - Set point 3R.A03 - Set point 3
R.A03 - Set point 3
R.A03 - Set point 3R.A03 - Set point 3 This parameter is available only when one of the logic inputs is configured for SP3/SP4 selection. Upper display: Gr.1 Lower display: SP3 Range: from "rL" [R.E10] to "rH" [R.E11].
R.A04 - Set point 4R.A04 - Set point 4
R.A04 - Set point 4
R.A04 - Set point 4R.A04 - Set point 4 This parameter is available only when one of the logic inputs is configured for SP/SP2 selection and the other for SP3/SP4 selection. Upper display: Gr.1 Lower display: SP4 Range: from "rL" [R.E10] to "rH" [R.E11].
R.A05 - Carbon monoxide factor settingR.A05 - Carbon monoxide factor setting
R.A05 - Carbon monoxide factor setting
R.A05 - Carbon monoxide factor settingR.A05 - Carbon monoxide factor setting Upper display: Gr.1 Lower display: COF Range: from 1 to 1000
When a primary variable different from “O2” is selected.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) COF is the keyboard selectable correction factor applied to the Carbon algorithm. It is used to bring carbon potential as calculated by the
instrument into alignment with actual carbon potential as determined by shim stock analysis. The default value of 200 corresponds to carbon potentials for plain carbon steel in an atmosphere containing 20% carbonmonoxide (CO). (See COF and H2F adjustment diagram at pag. 56).
2) Changing the A.In.F [C.D07] parameter from "nonE" to "CO"; this parameter is forced to 200.
R.A06 - Hydrogen factor settingR.A06 - Hydrogen factor setting
R.A06 - Hydrogen factor setting
R.A06 - Hydrogen factor settingR.A06 - Hydrogen factor setting Upper display: Gr.1 Lower display: H2F Range: from 1 to 1000.
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE:
NOTE:NOTE: H2F is the keyboard selectable correction factor applied to the dew point algorithm. It is used to bring dew point as calculated by the instrument into alignment with actual dew point as determined by a direct dew point measuring instrument. The default value of 400, corresponds to endothermic gas containing 40% Hydrogen (H2). (See COF and H2F adjustment diagram at pag. 56).
59
R.A07 - Group 1 default data loading.R.A07 - Group 1 default data loading.
R.A07 - Group 1 default data loading.
R.A07 - Group 1 default data loading.R.A07 - Group 1 default data loading. Upper display: Gr.1 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
Run time group 2 [R.Bxx]Run time group 2 [R.Bxx]
Run time group 2 [R.Bxx]
Run time group 2 [R.Bxx]Run time group 2 [R.Bxx] SMART ENABLE/DISABLE
R.B01 - SMARTR.B01 - SMART
R.B01 - SMART
R.B01 - SMARTR.B01 - SMART This parameter is available when at least one control output is configured and SMART function is enabled (“Sñ.Fn” [C.G01]= “Enb”) Upper display: Gr.2 Lower display: Sñrt Range: OFF = SMART algorithm is deactivated
ON = SMART algorithm is activated
Set On or OFF to activate/deactivate the SMART algorithm.
Run time group 3 [R.Cxx]Run time group 3 [R.Cxx]
Run time group 3 [R.Cxx]
Run time group 3 [R.Cxx]Run time group 3 [R.Cxx] ALARM THRESHOLD AND HYSTERESIS VALUE
R.C01 - Manual reset of the alarmsR.C01 - Manual reset of the alarms
R.C01 - Manual reset of the alarms
R.C01 - Manual reset of the alarmsR.C01 - Manual reset of the alarms Upper display: Gr.3 Lower display: ñ.rSt Range: ON/OFF Change to ON and then depress “FUNC” key to reset/ acknowledge the alarm condition. NOTESNOTES
NOTES:
NOTESNOTES
1) The reset/acknowledge function is always enabled but it is possible to protect it by security code.
2) When one of the logic input is configured as alarm reset, both actions (of "ñ.rSt" parameter and of the logic input) are active.
60
R.C02 - Alarm 1 thresholdR.C02 - Alarm 1 threshold
R.C02 - Alarm 1 threshold
R.C02 - Alarm 1 thresholdR.C02 - Alarm 1 threshold This parameter is available when alarm 1 is configured as process or deviation output alarm (“A1.tP” [RF.01] = “Proc” or “dEV”) Upper display: Gr.3 Lower display: AL1 Range: - in engineering units within the span limits for process
alarm;
- from -1000 to 1000 digits for deviation alarm;
NOTENOTE
NOTE: The span limits are related to the controlled variable set
NOTENOTE
by the "PVSL" [C.D02] parameter.
R.C03 -R.C03 -
R.C03 -
R.C03 -R.C03 -
Low threshold used when theLow threshold used when the
Low threshold used when the
Low threshold used when theLow threshold used when the
alarm 1 is a band alarm.alarm 1 is a band alarm.
alarm 1 is a band alarm.
alarm 1 is a band alarm.alarm 1 is a band alarm. This parameter is available only when the alarm 1 is configured as band alarm (“A1.tP” [RF.01] = “bAnd”) Upper display: Gr.3 Lower display: bA1.L Range: from 0 to -1000 digits.
R.C04 -R.C04 -
R.C04 -
R.C04 -R.C04 ­1 is a band alarm.1 is a band alarm.
1 is a band alarm.
1 is a band alarm.1 is a band alarm.
high threshold used when the alarmhigh threshold used when the alarm
high threshold used when the alarm
high threshold used when the alarmhigh threshold used when the alarm
This parameter is available only when the alarm 1 is configured as band alarm (“A1.tP” [RF.01] = “bAnd”) Upper display: Gr.3 Lower display: bA1.h Range: from 0 to 1000 digits. NOTENOTE
NOTE: The "bA1.L" and "bA1.h" values are algebraically added
NOTENOTE to the operative set point in order to obtain the band limits.
R.C05 - Alarm 2 thresholdR.C05 - Alarm 2 threshold
R.C05 - Alarm 2 threshold
R.C05 - Alarm 2 thresholdR.C05 - Alarm 2 threshold This parameter is available when alarm 2 is configured as process or deviation output alarm (“A2.tP” [RF.05] = “Proc” or “dEV”). Upper display: Gr.3 Lower display: AL2 Range: - in engineering units within the span limits for process
alarm;
- from -1000 to 1000 digits for deviation alarm;
NOTENOTE
NOTE: The span limits are related to the controlled variable set
NOTENOTE
by the "PVSL" [C.D02] parameter.
61
R.C06 -R.C06 -
R.C06 -
R.C06 -R.C06 ­is a band alarm.is a band alarm.
is a band alarm.
is a band alarm.is a band alarm.
Low threshold used when the alarm 2Low threshold used when the alarm 2
Low threshold used when the alarm 2
Low threshold used when the alarm 2Low threshold used when the alarm 2
This parameter is available only when the alarm 2 is configured as band alarm (“A2.tP” [RF.05] = “bAnd”) Upper display: Gr.3 Lower display: bA2.L Range: from 0 to -1000 digits.
R.C07 -R.C07 -
R.C07 -
R.C07 -R.C07 ­2 is a band2 is a band
2 is a band
2 is a band2 is a band
high threshold used when the alarmhigh threshold used when the alarm
high threshold used when the alarm
high threshold used when the alarmhigh threshold used when the alarm
alarm.alarm.
alarm.
alarm.alarm. This parameter is available only when the alarm 2 is configured as band alarm (“A2.tP” [RF.05] = “bAnd”) Upper display: Gr.3 Lower display: bA2.h Range: from 0 to 1000 digits. NOTENOTE
NOTE: The "bA2.L" and "bA2.h" values are algebraically added
NOTENOTE to the operative set point in order to obtain the band limits.
R.C08 - Alarm 3 thresholdR.C08 - Alarm 3 threshold
R.C08 - Alarm 3 threshold
R.C08 - Alarm 3 thresholdR.C08 - Alarm 3 threshold This parameter is available when alarm 3 is configured as process or deviation output alarm (“A3.tP” [RF.09] = “Proc” or “dEV”). Upper display: Gr.3 Lower display: AL3 Range: - in engineering units within the span limits for process
alarm;
- from -1000 to 1000 digits for deviation alarm;
NOTENOTE
NOTE: The span limits are related to the controlled variable set
NOTENOTE
by the "PVSL" [C.D02] parameter.
R.C09 -R.C09 -
R.C09 -
R.C09 -R.C09 -
Low threshold used when theLow threshold used when the
Low threshold used when the
Low threshold used when theLow threshold used when the
alarm 3 is a band alarm.alarm 3 is a band alarm.
alarm 3 is a band alarm.
alarm 3 is a band alarm.alarm 3 is a band alarm. This parameter is available only when the alarm 3 is configured as band alarm (“A3.tP” [RF.09] = “bAnd”) Upper display: Gr.3 Lower display: bA3.L Range: from 0 to -1000 digits.
62
R.C10 -R.C10 -
R.C10 -
R.C10 -R.C10 -
High threshold used when theHigh threshold used when the
High threshold used when the
High threshold used when theHigh threshold used when the
alarm 3 is a band alarm.alarm 3 is a band alarm.
alarm 3 is a band alarm.
alarm 3 is a band alarm.alarm 3 is a band alarm. This parameter is available only when the alarm 3 is configured as band alarm (“A3.tP” [RF.09] = “bAnd”) Upper display: Gr.3 Lower display: bA3.h Range: from 0 to 1000 digits. NOTENOTE
NOTE: The "bA3.L" and "bA3.h" values are algebraically added
NOTENOTE to the operative set point in order to obtain the band limits.
R.C11 -R.C11 -
R.C11 -
R.C11 -R.C11 -
Alarm 1 hysteresisAlarm 1 hysteresis
Alarm 1 hysteresis
Alarm 1 hysteresisAlarm 1 hysteresis This parameter is available only when Out 1 is configured as alarm output (“O1.Fn” [C.E01] = “ALr.1”) Upper display: Gr.3 Lower display: HSA1 Range: from 1 to 200 digits
R.C13 -R.C13 -
R.C13 -
R.C13 -R.C13 -
Alarm 3 hysteresisAlarm 3 hysteresis
Alarm 3 hysteresis
Alarm 3 hysteresisAlarm 3 hysteresis This parameter is available only when Out 3 is configured as alarm output (“O3.Fn” [C.E03] = “ALr.3”) Upper display: Gr.3 Lower display: HSA3 Range: from 1 to 200 digits
R.C14 - Group 3 default data loadingR.C14 - Group 3 default data loading
R.C14 - Group 3 default data loading
R.C14 - Group 3 default data loadingR.C14 - Group 3 default data loading Upper display: Gr.3 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
R.C12 -R.C12 -
R.C12 -
R.C12 -R.C12 -
Alarm 2 hysteresisAlarm 2 hysteresis
Alarm 2 hysteresis
Alarm 2 hysteresisAlarm 2 hysteresis This parameter is available only when Out 2 is configured as alarm output (“O2.Fn” [C.E02] = “ALr.2”) Upper display: Gr.3 Lower display: HSA2 Range: from 1 to 200 digits
63
Run time group 4 [R.Dxx]Run time group 4 [R.Dxx]
Run time group 4 [R.Dxx]
Run time group 4 [R.Dxx]Run time group 4 [R.Dxx] CONTROL PARAMETERS
NOTENOTE
NOTE: when SMART function is activated, all the parameters
NOTENOTE calculated by SMART (Pb, Ti, Td and RCG) cannot be modified.
R.D01 - Proportional bandR.D01 - Proportional band
R.D01 - Proportional band
R.D01 - Proportional bandR.D01 - Proportional band This parameter is available when at least one control output is configured. Upper display: Gr.4 Lower display: Pb Range: from 0.5% to 999.0% of the input span.
Set 0.0% for On/OFF control action
NOTESNOTES
NOTES:
NOTESNOTES
1) The Pb resolution is:
0.1% up to 10.0%; 1% from 10% up to 999.0%.
2) When the instrument is working with SMART algorithm, the “Pb” value will be limited as selected by “Pb.Hi” [R.M02] and “Pb.Lo” [R.M01] parameters (if the value is out of limits, the "E.140" indication will be shown on the middle display, for details see "Error messages" at page 83)
R.D02 - Hysteresis (for ON/OFF control)R.D02 - Hysteresis (for ON/OFF control)
R.D02 - Hysteresis (for ON/OFF control)
R.D02 - Hysteresis (for ON/OFF control)R.D02 - Hysteresis (for ON/OFF control) This parameter is available when Pb [r.d01]=0 (On/OFF control action) Upper display: Gr.4 Lower display: HYS Range: from 0.1% to 10.0% of the input span.
R.D03 - Integral timeR.D03 - Integral time
R.D03 - Integral time
R.D03 - Integral timeR.D03 - Integral time This parameter is available when at least one control output is configured and Pb [R.D01] is different from 0. Upper display: Gr.4 Lower display: ti Range: from 00.01 to 20.00 mm.ss
Above this value the display blanks and the integral action is excluded
NOTENOTE
NOTE: When the instrument is working with SMART algorithm
NOTENOTE the “ti” value will be limited as selected by “ti.Hi” [R.M04] and “ti.Lo” [R.M03] parameters (if the value is out of limits, the "E.140" indication will be shown on the middle display, for details see "Error messages at pag. 83).
64
R.D04 - Derivative timeR.D04 - Derivative time
R.D04 - Derivative time
R.D04 - Derivative timeR.D04 - Derivative time This parameter is available when at least one control output is configured, “Cn.tP” [C.G02] is equal to “Pid” and "Pb" [R.D01] is different from 0. Upper display: Gr.4 Lower display: td Range: From 00.00 to 10.00 mm.ss NOTENOTE
NOTE:
NOTENOTE
When device is working with SMART algorithm and “Cn.tP” [C.G02] is equal to “Pid”, the “td” value will be proportional to the “ti” value with a ratio established during the "TUNE" function.
R.D05 - Integral preloadR.D05 - Integral preload
R.D05 - Integral preload
R.D05 - Integral preloadR.D05 - Integral preload This parameter is available when at least one control output is configured and "Pb" [R.D01] is different from 0. Upper display: Gr.4 Lower display: IP Ranges:
- from 0.0 to 100.0 % of the output when device is configured with only one control output.
- from -100.0 to 100.0 % of the output when device is configured with two control outputs.
R.D06 - Relative secondary output gainR.D06 - Relative secondary output gain
R.D06 - Relative secondary output gain
R.D06 - Relative secondary output gainR.D06 - Relative secondary output gain This parameter is available when two control outputs are configured. Upper display: Gr.4 Lower display: r.Gn Range: from 0.20 to 2.00
R.D07 - Overlap/dead band between main andR.D07 - Overlap/dead band between main and
R.D07 - Overlap/dead band between main and
R.D07 - Overlap/dead band between main andR.D07 - Overlap/dead band between main and secondary outputssecondary outputs
secondary outputs
secondary outputssecondary outputs This parameter is available when two control outputs are configured. Upper display: Gr.4 Lower display: OLAP Range: from -20 to 50 NOTENOTE
NOTE: A negative value means a dead band while a positive
NOTENOTE value means an overlap.
65
R.D08 - Group 4 default data loading.R.D08 - Group 4 default data loading.
R.D08 - Group 4 default data loading.
R.D08 - Group 4 default data loading.R.D08 - Group 4 default data loading. This parameter is not available when SMART function is activated (parameters related to SMART function cannot be modified). Upper display: Gr.4 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
Run time group 5 [R.Exx]Run time group 5 [R.Exx]
Run time group 5 [R.Exx]
Run time group 5 [R.Exx]Run time group 5 [R.Exx] AUXILIARY CONTROL PARAMETER
R.E01 - Anti reset wind upR.E01 - Anti reset wind up
R.E01 - Anti reset wind up
R.E01 - Anti reset wind upR.E01 - Anti reset wind up This parameter is available when at least one control output is configured and "Pb" [R.D01] is different from 0. Upper display: Gr.5 Lower display: ArW Range: from 10% to 200% of the input span.
R.E02 - Main control output low limitR.E02 - Main control output low limit
R.E02 - Main control output low limit
R.E02 - Main control output low limitR.E02 - Main control output low limit This parameter is available when one control output is configured as main output. Upper display: Gr.5 Lower display: ñ.OLL Range: from 0.0 % (of the output span) to ñ.OLH [R.E03]
66
R.E03 - Main control output high limitR.E03 - Main control output high limit
R.E03 - Main control output high limit
R.E03 - Main control output high limitR.E03 - Main control output high limit This parameter is available when one control output is configured as main output. Upper display: Gr.5 Lower display: ñ.OLH Range: from ñ.OLL [R.E02] to 100.0 % of the output span.
R.E05 - Main control output cycle time.R.E05 - Main control output cycle time.
R.E05 - Main control output cycle time.
R.E05 - Main control output cycle time.R.E05 - Main control output cycle time. This parameter is available when time proportional main control output is configured. Upper display: Gr.5 Lower display: ñC.CY Range: from 1 to 200 s
R.E04 - Main control output max rate of riseR.E04 - Main control output max rate of rise
R.E04 - Main control output max rate of rise
R.E04 - Main control output max rate of riseR.E04 - Main control output max rate of rise This parameter is available when one control output is configured as main output. Upper display: Gr.5 Lower display: ñ.rñP Ranges: from 0.1%/s to 25.0%/s. Above this value the display shows “Inf” meaning that no limit is imposed. NOTESNOTES
NOTES:
NOTESNOTES
1) This parameter will be active even if device is configured for On/OFF control.
2) During the transfer from AUTO to MANUAL, if this action is programmed as a not bumpless transfer, the “ñ.rñP” function will be ignored and the output will step up to the value set by “Añ.UL” [C.G04] parameter.
R.E06 - Secondary control output low limitR.E06 - Secondary control output low limit
R.E06 - Secondary control output low limit
R.E06 - Secondary control output low limitR.E06 - Secondary control output low limit This parameter is available when one control output is configured as secondary control output. Upper display: Gr.5 Lower display: S.OLL Range: From 0.0 % of the output span to S.OLH [R.E07].
R.E07 - Secondary control output high limitR.E07 - Secondary control output high limit
R.E07 - Secondary control output high limit
R.E07 - Secondary control output high limitR.E07 - Secondary control output high limit This parameter is available when one control output is configured as secondary control output. Upper display: Gr.5 Lower display: S.OLH Range: from S.OLL [R.E06] to 100.0 % of the output span.
67
R.E08 -R.E08 -
Secondary control output max rate of riseSecondary control output max rate of rise
R.E08 -
Secondary control output max rate of rise
R.E08 -R.E08 -
Secondary control output max rate of riseSecondary control output max rate of rise This parameter is available when one control output is configured as secondary control output. Upper display: Gr.5 Lower display: S.rñP Ranges: from 0.1%/s to 25.0%/s. Above this value the display shows “Inf” meaning that no limit is imposed NOTESNOTES
NOTES:
NOTESNOTES
1) This parameter will be active even if device is configured for
On/OFF control.
2) During the transfer from AUTO to MANUAL, if this action is
programmed as a not bumpless transfer, the “S.rñP” function will be ignored and the output will step up to the value set by “Añ.UL” [C.G04] parameter.
R.E09 - Secondary control output cycle time.R.E09 - Secondary control output cycle time.
R.E09 - Secondary control output cycle time.
R.E09 - Secondary control output cycle time.R.E09 - Secondary control output cycle time. This parameter is available when a time proportional secondary control output is configured. Upper display: Gr.5 Lower display: SC.CY Range: from 1 to 200 s
R.E10 - Set point low limitR.E10 - Set point low limit
R.E10 - Set point low limit
R.E10 - Set point low limitR.E10 - Set point low limit Upper display: Gr.5 Lower display: rL Range: from input low range (depending on the "PV.SL" [C.D02] setting) to rH [R.E11]. NOTESNOTES
NOTES: Whenever "PV.SL" [C.D02] has been modified, the
NOTESNOTES following occurs:
1) rL will be realigned to the low limit; the realignement will be carried out even if the set point is software protected.
2)If the set point values are out of the new limits, they automati­cally assume the default value even if software protected.
R.E11 - Set point high limit.R.E11 - Set point high limit.
R.E11 - Set point high limit.
R.E11 - Set point high limit.R.E11 - Set point high limit. Upper display: Gr.5 Lower display: rH Range: from rL [R.E10] to input high range (depending on the "PV.SL" [C.D02] setting). NOTESNOTES
NOTES: Whenever "PV.SL" [C.D02] has been modified, the
NOTESNOTES following occurs:
1) rH will be realigned to the low limit; the realignement will be carried out even if the set point is software protected.
2) If the set point values are out of the new limits, they automatically assume the default value even if software protected.
68
R.E12 -R.E12 -
Rate of change for positive set pointRate of change for positive set point
R.E12 -
Rate of change for positive set point
R.E12 -R.E12 -
Rate of change for positive set pointRate of change for positive set point
variationsvariations
variations
variationsvariations Upper display: Gr.5 Lower display: Grd1 Range: from 1 to 200 digits per minutes. Above this value the display shows “Inf” and the transfer will be a step change.
R.E13 -R.E13 -
Rate of change for negative set pointRate of change for negative set point
R.E13 -
Rate of change for negative set point
R.E13 -R.E13 -
Rate of change for negative set pointRate of change for negative set point
variationsvariations
variations
variationsvariations Upper display: Gr.5 Lower display: Grd2 Range: from 1 to 200 digits per minutes. Above this value the display shows “Inf” and the transfer will be a step change.
R.E14 -R.E14 -
External control for AUTO /MANUAL modeExternal control for AUTO /MANUAL mode
R.E14 -
External control for AUTO /MANUAL mode
R.E14 -R.E14 -
External control for AUTO /MANUAL modeExternal control for AUTO /MANUAL mode This parameter is available when one of the digital inputs is configured for Auto/Manual selection. Upper display: Gr.5 Lower display: E.Añ Range: On the Auto/Manual selection is made by digital input
only.
OFF the Auto/Manual selection is made by “Man” key
or serial link only.
R.E15 - Group 5 default data loading.R.E15 - Group 5 default data loading.
R.E15 - Group 5 default data loading.
R.E15 - Group 5 default data loading.R.E15 - Group 5 default data loading. Upper display: Gr.5 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
69
Run time group 6 [R.Fxx]Run time group 6 [R.Fxx]
Run time group 6 [R.Fxx]
Run time group 6 [R.Fxx]Run time group 6 [R.Fxx] ALARM SETTING
R.F01 -R.F01 -
Alarm 1 typeAlarm 1 type
R.F01 -
Alarm 1 type
R.F01 -R.F01 -
Alarm 1 typeAlarm 1 type This parameter is available only when OUT 1 is configured as alarm 1 output (“O1.Fn” [C.E01] = “ALr.1”) Upper display: Gr.6 Lower display: A1.tP Range: Proc = Alarm on process variable
bAnd = Band alarm on process variable dEV = Deviation alarm on process variable
NOTENOTE
NOTE: When alarm type has been changed the alarm threshold
NOTENOTE
will be forced to its default value and alarm status will be cleared.
R.F02 -R.F02 -
Alarm 1 configurationAlarm 1 configuration
R.F02 -
Alarm 1 configuration
R.F02 -R.F02 -
Alarm 1 configurationAlarm 1 configuration This parameter is available only when OUT 1 is configured as alarm 1 output (“O1.Fn” [C.E01] = “ALr.1”) Upper display: Gr.6 Lower display: A1.Cn Range: H.A. = High alarm (outside band) with automatic reset.
L.A. = Low alarm (inside band) with automatic reset. H.A.Ac= High alarm (outside band) with automatic reset
and acknowledge.
L.A.Ac= Low alarm (inside band) with automatic reset
and acknowledge. H.L. = High alarm (outside band) with manual reset. L.L. = Low alarm (inside band) with manual reset.
NOTENOTE
NOTE: When alarm configuration has been changed, the alarm
NOTENOTE status will be cleared.
70
ExampleExample
Example for A1.Cn (A2.Cn, A3.Cn) = H.A.
ExampleExample
Alarm Threshold
ExampleExample
Example for A1.Cn (A2.Cn, A3.Cn) = H.A.A.c
ExampleExample
Alarm Threshold
Alarm Hysteresis
Measured Value
ON Relay out OFF
ON
LED 1(2, 3)
Flash
Flash
OFF
A
A= alarm condition detection
B
A
C
B= automatic reset of alarm C= manual reset; the LED gets steady lit, the alarm remains in
the alarm status until the process variable reaches the alarm threshold minus hysteresis (D).
Steady
D
Alarm Hysteresis
Measured
Value ON Relay out OFF
ON
LED 1(2, 3)
Flash
Flash
OFF
A
B
A
A= alarm condition detection B= automatic reset of alarm C= manual reset; the LED gets steady lit, the relay output goes
OFF but the LED remains steady lit until the process variable reaches the alarm threshold minus hysteresis (D).
71
C
Steady
D
ExampleExample
Example for A1.Cn (A2.Cn, A3.Cn) = H.L.
ExampleExample
Alarm Threshold
Alarm Hysteresis
Measured Value
OFF Relay out
ON
ON
LED 1(2, 3)
Flash
Flash
OFF
A
C
B
A
E
A= alarm condition detection B= the alarm remains in alarm status (even if the measured value
is under the threshold) until a manual reset is performed (C).
E= if a manual reset is performed when the alarm condition still
exist; the LED gets steady lit, the alarm remains in the alarm
Steady
D
status until the process variable reaches the alarm threshold minus hysteresis (D).
NOTESNOTES
NOTES:
NOTESNOTES
1) The alarm status is present when: the relay is energized (A1.Ac, A2.Ac or A3.Ac = dir) the relay is deenergized (A1.Ac, A2.Ac or A3.Ac = rEV)
2) The manual reset (acknowlwdgement) may be performed by
the "ñ.rSt" [R.C01] parameter or by logic input or by serial link.
R.F03 -R.F03 -
Alarm 1 actionAlarm 1 action
R.F03 -
Alarm 1 action
R.F03 -R.F03 -
Alarm 1 actionAlarm 1 action This parameter is available only when OUT 1 is configured as alarm 1 output (“O1.Fn” [C.E01] = “ALr.1”) Upper display: Gr.6 Lower display: A1.Ac Range: dir = direct action (Relay energized or SSr=1 in alarm
condition).
rEV = Reverse action (Relay energized or SSr=1 in non
alarm condition).
72
R.F04 -R.F04 -
Alarm 1 stand by (mask) functionAlarm 1 stand by (mask) function
R.F04 -
Alarm 1 stand by (mask) function
R.F04 -R.F04 -
Alarm 1 stand by (mask) functionAlarm 1 stand by (mask) function This parameter is available only when OUT 1 is configured as alarm 1 output (“O1.Fn” [C.E01] = “ALr.1”) Upper display: Gr.6 Lower display: A1.St Range: OFF = Stand-by function disabled
On = Stand-by function enabled
NOTESNOTES
NOTES:
NOTESNOTES
1) If the alarm is programmed as band or deviation alarm, this
function masks the alarm condition after a set point change or at the instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis. If the alarm is programmed as a process alarm, this function masks the alarm condition at instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis.
2) The change from On to OFF has immediate effect, while the
change from OFF to On has effect at the next start up or set point change.
R.F05 -R.F05 -
Alarm 2 typeAlarm 2 type
R.F05 -
Alarm 2 type
R.F05 -R.F05 -
Alarm 2 typeAlarm 2 type This parameter is available only when OUT 2 is configured as alarm 2 output (“O2.Fn” [C.E02] = “ALr.2”) Upper display: Gr.6 Lower display: A2.tP Range: Proc = Alarm on process variable
bAnd = Band alarm on process variable dEV = Deviation alarm on process variable
NOTENOTE
NOTE: When alarm type has been changed the alarm threshold
NOTENOTE
will be forced to its default value and alarm status will be cleared.
R.F06 -R.F06 -
Alarm 2 configurationAlarm 2 configuration
R.F06 -
Alarm 2 configuration
R.F06 -R.F06 -
Alarm 2 configurationAlarm 2 configuration This parameter is available only when OUT 2 is configured as alarm 2 output (“O2.Fn” [C.E02] = “ALr.2”) Upper display: Gr.6 Lower display: A2.Cn Range: H.A. = High alarm (outside band) with automatic reset.
L.A. = Low alarm (inside band) with automatic reset. H.A.Ac= High alarm (outside band) with automatic reset
and acknowledge.
L.A.Ac= Low alarm (inside band) with automatic reset
and acknowledge.
73
H.L. = High alarm (outside band) with manual reset. L.L. = Low alarm (inside band) with manual reset.
NOTENOTE
NOTE: When alarm configuration has been changed, the alarm
NOTENOTE status will be cleared.
R.F07 -R.F07 -
Alarm 2 actionAlarm 2 action
R.F07 -
Alarm 2 action
R.F07 -R.F07 -
Alarm 2 actionAlarm 2 action This parameter is available only when OUT 2 is configured as alarm 2 output (“O2.Fn” [C.E02] = “ALr.2”) Upper display: Gr.6 Lower display: A2.Ac Range: dir = direct action (Relay energized or SSr=1 in alarm
condition)
rEV = Reverse action (Relay energized or SSr=1 in non
alarm condition)
R.F08 -R.F08 -
Alarm 2 stand by (mask) functionAlarm 2 stand by (mask) function
R.F08 -
Alarm 2 stand by (mask) function
R.F08 -R.F08 -
Alarm 2 stand by (mask) functionAlarm 2 stand by (mask) function This parameter is available only when OUT 2 is configured as alarm 2 output (“O2.Fn” [C.E02] = “ALr.2”) Upper display: Gr.6 Lower display: A2.St Range: OFF = Stand-by function disabled
On = Stand-by function enabled
NOTESNOTES
NOTES:
NOTESNOTES
1) If the alarm is programmed as band or deviation alarm, this function masks the alarm condition after a set point change or at the instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis. If the alarm is programmed as a process alarm, this function masks the alarm condition at instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis.
2) The change from On to OFF has immediate effect, while the change from OFF to On has effect at the next start up or set point change
R.F09 -R.F09 -
Alarm 3 typeAlarm 3 type
R.F09 -
Alarm 3 type
R.F09 -R.F09 -
Alarm 3 typeAlarm 3 type This parameter is available only when OUT 3 is configured as alarm 3 output (“O3.Fn” [C.E03] = “ALr.3”) Upper display: Gr.6 Lower display: A3.tP Range: Proc = Alarm on process variable
bAnd = Band alarm on process variable dEV = Deviation alarm on process variable
NOTENOTE
NOTE: When alarm type has been changed the alarm threshold
NOTENOTE
will be forced to its default value and alarm status will be cleared.
74
R.F10 -R.F10 -
Alarm 3 configurationAlarm 3 configuration
R.F10 -
Alarm 3 configuration
R.F10 -R.F10 -
Alarm 3 configurationAlarm 3 configuration This parameter is available only when OUT 3 is configured as alarm 3 output (“O3.Fn” [C.E03] = “ALr.3”). Upper display: Gr.6 Lower display: A3.Cn Range: H.A. = High alarm (outside band) with automatic reset.
L.A. = Low alarm (inside band) with automatic reset. H.A.Ac= High alarm (outside band) with automatic reset
and acknowledge.
L.A.Ac= Low alarm (inside band) with automatic reset
and acknowledge. H.L. = High alarm (outside band) with manual reset. L.L. = Low alarm (inside band) with manual reset.
NOTENOTE
NOTE: When alarm configuration has been changed, the alarm
NOTENOTE status will be cleared.
R.F11 -R.F11 -
Alarm 3 actionAlarm 3 action
R.F11 -
Alarm 3 action
R.F11 -R.F11 -
Alarm 3 actionAlarm 3 action This parameter is available only when OUT 3 is configured as alarm 3 output (“O3.Fn” [C.E03] = “ALr.3”). Upper display: Gr.6 Lower display: A3.Ac Range: dir = direct action (Relay energized or SSr=1 in alarm
condition).
rEV = Reverse action (Relay energized or SSr=1 in non
alarm condition).
R.F12 -R.F12 -
Alarm 3 standby (mask) functionAlarm 3 standby (mask) function
R.F12 -
Alarm 3 standby (mask) function
R.F12 -R.F12 -
Alarm 3 standby (mask) functionAlarm 3 standby (mask) function This parameter is available only when OUT 3 is configured as alarm 3 output (“O3.Fn” [C.E03] = “ALr.3”) Upper display: Gr.6 Lower display: A3.St Range: OFF = Stand-by function disabled
On = Stand-by function enabled
NOTESNOTES
NOTES:
NOTESNOTES
1) If the alarm is programmed as band or deviation alarm, this
function masks the alarm condition after a set point change or at the instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis. If the alarm is programmed as a process alarm, this function masks the alarm condition at instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis.
2) The change from On to OFF has immediate effect, while the
change from OFF to On has effect at the next start up or set point change
75
R.F13 - Group 6 default data loading.R.F13 - Group 6 default data loading.
R.F13 - Group 6 default data loading.
R.F13 - Group 6 default data loading.R.F13 - Group 6 default data loading. Upper display: Gr.6 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
Run time group 7 [R.Gxx]Run time group 7 [R.Gxx]
Run time group 7 [R.Gxx]
Run time group 7 [R.Gxx]Run time group 7 [R.Gxx] SERIAL LINK PARAMETER
R.G02 -R.G02 -
R.G02 -
R.G02 -R.G02 -
Serial link device addressSerial link device address
Serial link device address
Serial link device addressSerial link device address This parameter is available only when “S.L.Pr” [R.G01] is different from "OFF". Upper display: Gr.7 Lower display: S.L.Ad Range: from 1 to 255 R.G03 -R.G03 -
R.G03 -
R.G03 -R.G03 -
Serial link baud rateSerial link baud rate
Serial link baud rate
Serial link baud rateSerial link baud rate This parameter is available only when “S.L.Pr” [R.G01] is different from "OFF". Upper display: Gr.7 Lower display: S.L.bd Range: from 600 to 19200 baud (19200 baud is shown, on display, as 19.20)
R.G01 -R.G01 -
R.G01 -
R.G01 -R.G01 -
Serial interface protocolSerial interface protocol
Serial interface protocol
Serial interface protocolSerial interface protocol Upper display: Gr.7 Lower display: S.L.Pr Range: OFF = No serial interface
ñbUS = Modbus jbUS = Jbus
R.G04 -R.G04 -
R.G04 -
R.G04 -R.G04 -
Serial link byte formatSerial link byte format
Serial link byte format
Serial link byte formatSerial link byte format This parameter is available only when “S.L.Pr” [R.G01] is different from "OFF". Upper display: Gr.7 Lower display: S.L.bF Range: 8 = 8 bits without parity
8E = 8 bits + even parity 8O = 8 bits + odd parity
76
R.G05 - Group 7 default data loading.R.G05 - Group 7 default data loading.
R.G05 - Group 7 default data loading.
R.G05 - Group 7 default data loading.R.G05 - Group 7 default data loading. Upper display: Gr.7 Lower display: dFLt. Range: OFF = No loading data
ON = Loading data
Run time group 8 [R.Hxx]Run time group 8 [R.Hxx]
Run time group 8 [R.Hxx]
Run time group 8 [R.Hxx]Run time group 8 [R.Hxx] TIMERS SETTING
R.H01 -R.H01 -
R.H01 -
R.H01 -R.H01 -
Burnoff durationBurnoff duration
Burnoff duration
Burnoff durationBurnoff duration Upper display: Gr.8 Lower display: t.bOF Range: from 1.00 to 15.00 mm.ss
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: the time setting may be modified any time but it will be
NOTE:NOTE:
activated at the next operation.
R.H02 -R.H02 -
R.H02 -
R.H02 -R.H02 -
Purge durationPurge duration
Purge duration
Purge durationPurge duration Upper display: Gr.8 Lower display: t.PrG Range: from 1.00 to 15.00 mm.ss
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: the time setting may be modified any time but it will be
NOTE:NOTE:
activated at the next operation.
77
R.H03 -R.H03 -
R.H03 -
R.H03 -R.H03 -
Automatic Burnoff intervalAutomatic Burnoff interval
Automatic Burnoff interval
Automatic Burnoff intervalAutomatic Burnoff interval Upper display: Gr.8 Lower display: bF.tr Range: from 1.00 to 24.00 hh. mm.
Above this 24.00, the display shows "no.tr" (no time
interval is programmed)
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: the time setting may be modified any time but it will be
NOTE:NOTE:
performed at the next Burn off, exception made for the
changement from/to "no.tr" that it will be immediately
activated. R.H04 -R.H04 -
R.H04 -
R.H04 -R.H04 -
Automatic probe test intervalAutomatic probe test interval
Automatic probe test interval
Automatic probe test intervalAutomatic probe test interval Upper display: Gr.8 Lower display: Pb.tr Range: from 1.00 to 24.00 hh. mm.
Above this 24.00, the display shows "no.tr" (no time
interval is programmed)
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: the time setting may be modified any time but it will be
NOTE:NOTE:
performed at the next Probe test, exception made for the
change from/to "no.tr" that it will be immediately
activated.
R.H05 -R.H05 -
R.H05 -
R.H05 -R.H05 ­BurnoffBurnoff
Burnoff
BurnoffBurnoff
Time remaining to the next automaticTime remaining to the next automatic
Time remaining to the next automatic
Time remaining to the next automaticTime remaining to the next automatic
This is a reading parameter only and it is available when "bF.tr" is different from "no.tr". Upper display: Gr.8 Lower display: bF.ñn Range: from 1.00 to 24.00 hh. mm.
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: if a power down occurs, this information will be lost,
NOTE:NOTE:
when power is restored, the countdown will restart from the "bF.tr" [R.H03] value.
R.H06 -R.H06 -
R.H06 -
R.H06 -R.H06 ­Probe testProbe test
Probe test
Probe testProbe test
Time remaining to the next automaticTime remaining to the next automatic
Time remaining to the next automatic
Time remaining to the next automaticTime remaining to the next automatic
This is a reading parameter only and it is available when "pb.tr" is different from "no.tr". Upper display: Gr.8 Lower display: Pb.ñn Range: from 1.00 to 24.00 hh. mm.
When a primary variable different from “O2” is selected.
NOTE:NOTE:
NOTE: if a power down occurs, this information will be lost,
NOTE:NOTE:
when power is restored, the countdown will restart from the "Pb.tr" [R.H04] value.
78
R.H07 - Group 8 default data loading.R.H07 - Group 8 default data loading.
R.H07 - Group 8 default data loading.
R.H07 - Group 8 default data loading.R.H07 - Group 8 default data loading. Upper display: Gr.8 Lower display: dFLt. Range: OFF = No loading data
On = loading data
Run time group 9 [R.Ixx]Run time group 9 [R.Ixx]
Run time group 9 [R.Ixx]
Run time group 9 [R.Ixx]Run time group 9 [R.Ixx] OPTIONAL OUTPUT SETTING
R.I01 - OUT 10 settingR.I01 - OUT 10 setting
R.I01 - OUT 10 setting
R.I01 - OUT 10 settingR.I01 - OUT 10 setting This parameter is available when optional output circuitry is fitted. Upper display: Gr.9 Lower display: Ou.10 Range: OFF = Relay deenergized
On = Relay energized
For OUT 11 up to OUT 19, the instrument will show on the lower display the related output and it is possible to perform the same setting as above described.
R.I02 - All relays deenergizationR.I02 - All relays deenergization
R.I02 - All relays deenergization
R.I02 - All relays deenergizationR.I02 - All relays deenergization This parameter is available when optional output circuitry is fitted. Upper display: Gr.9 Lower display: dEEn Range: OFF = No action
On = All relays will be deenergized
79
Run time group "dF" [R.Lxx]Run time group "dF" [R.Lxx]
Run time group "dF" [R.Lxx]
Run time group "dF" [R.Lxx]Run time group "dF" [R.Lxx] DEFAULT RUN TIME PARAMETER LOADING
R.L01 -R.L01 -
Default run time parameter loading.Default run time parameter loading.
R.L01 -
Default run time parameter loading.
R.L01 -R.L01 -
Default run time parameter loading.Default run time parameter loading. This parameter is not available when SMART function is activated. Upper display: Gr.dF Lower display: dFLt. Range: OFF = No loading data
On = the run time parameters of all groups (exception
made for group 9) will be forced to their default values.
Run time group "Hd" [R.Mxx]Run time group "Hd" [R.Mxx]
Run time group "Hd" [R.Mxx]
Run time group "Hd" [R.Mxx]Run time group "Hd" [R.Mxx] HIDDEN PARAMETERS - SMART LIMITS
NOTENOTE
NOTE: This group is accessible from every group by keeping
NOTENOTE depressed the “MENU” key for 8 seconds.
R.M01 -R.M01 -
R.M01 -
R.M01 -R.M01 -
Minimum value of proportional bandMinimum value of proportional band
Minimum value of proportional band
Minimum value of proportional bandMinimum value of proportional band
calculated by SMART algorithm.calculated by SMART algorithm.
calculated by SMART algorithm.
calculated by SMART algorithm.calculated by SMART algorithm. This parameter is available only when smart function is configured (“Sñ.Fn” [C.G01] = “Enb”) Upper display: Gr.Hd Lower display: Pb.Lo Range: From 2.0% to “Pb.Hi” [R.M02]. NOTENOTE
NOTE: The resolution on “Pb.Lo” value is: 0.1% up to 10.0%;
NOTENOTE
1% up to 999.0%.
80
R.M02 -R.M02 -
R.M02 -
R.M02 -R.M02 -
Maximum value of proportional bandMaximum value of proportional band
Maximum value of proportional band
Maximum value of proportional bandMaximum value of proportional band
calculated by SMART algorithmcalculated by SMART algorithm
calculated by SMART algorithm
calculated by SMART algorithmcalculated by SMART algorithm This parameter is available only when smart function is configured (“Sñ.Fn” [C.G01] = “Enb”) Upper display: Gr.Hd Lower display: Pb.Hi Range: from “Pb.Lo” [R.M01] to 999.0% NOTENOTE
NOTE: The resolution on “Pb.Hi” value is: 0.1% up to 10.0%;
NOTENOTE
1% up to 999.0
R.M03 -R.M03 -
R.M03 -
R.M03 -R.M03 -
Minimum value of integral timeMinimum value of integral time
Minimum value of integral time
Minimum value of integral timeMinimum value of integral time
calculated by SMART algorithmcalculated by SMART algorithm
calculated by SMART algorithm
calculated by SMART algorithmcalculated by SMART algorithm This parameter is available only when smart function is configured (“Sñ.Fn” [C.G01] = “Enb”) Upper display: Gr.Hd Lower display: ti.Lo Range: from 00.01 mm.ss to “ti.Hi” [R.M04].
R.M04 -R.M04 -
R.M04 -
R.M04 -R.M04 -
Maximum value of integral timeMaximum value of integral time
Maximum value of integral time
Maximum value of integral timeMaximum value of integral time
calculated by SMART algorithm.calculated by SMART algorithm.
calculated by SMART algorithm.
calculated by SMART algorithm.calculated by SMART algorithm. This parameter is available only when smart function is configured (“Sñ.Fn” [C.G01] = “Enb”) Upper display: Gr.Hd Lower display: ti.Hi Range: from “ti.Lo” [R.M03] to 20.00 mm.ss
R.M05 -R.M05 -
R.M05 -
R.M05 -R.M05 -
Relative gain of the secondary outputRelative gain of the secondary output
Relative gain of the secondary output
Relative gain of the secondary outputRelative gain of the secondary output
calculated by SMART algorithm.calculated by SMART algorithm.
calculated by SMART algorithm.
calculated by SMART algorithm.calculated by SMART algorithm. This parameter is available only when smart function is configured (“Sñ.Fn” [C.G01] = “Enb”) and a secondary control output is configured. Upper display: Gr.Hd Lower display: rG.CL Range: OFF = Smart algorithm does not calculate “r.Gn”
[R.D06] value.
On = Smart algorithm calculates “r.Gn” [R.D06] value.
R.M06 -R.M06 -
R.M06 -
R.M06 -R.M06 -
Hidden group default data loadingHidden group default data loading
Hidden group default data loading
Hidden group default data loadingHidden group default data loading Upper display: Gr.Hd Lower display: dFLt. Range: OFF = No loading data
On = loading data
81
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGESERROR MESSAGES OVERRANGE, UNDERRANGE AND TEMPERATUREOVERRANGE, UNDERRANGE AND TEMPERATURE
OVERRANGE, UNDERRANGE AND TEMPERATURE
OVERRANGE, UNDERRANGE AND TEMPERATUREOVERRANGE, UNDERRANGE AND TEMPERATURE INPUT LEAD BREAKING DETECTIONINPUT LEAD BREAKING DETECTION
INPUT LEAD BREAKING DETECTION
INPUT LEAD BREAKING DETECTIONINPUT LEAD BREAKING DETECTION The device is capable to detect a fault condition on input
variables (OVERRANGE or UNDERRANGE or TEMPERATURE INPUT LEAD BREAKING). When an error is detected the error code is shown (flashing) on the upper display as follows: a) Prhi Sensor input out of range (OVERRANGE).
It is detected when:
- mV is the primary controlled variable and its value is greater than 1515 mV;
- CP or DP is the primary controlled variable and its value is greater than 1303 mV.
(Greater than 200 mV if “O2” is configured as
primary control variable) b) PrLo Sensor input out of range (UNDERRANGE).
It is detected when:
- mV or O2 are the primary controlled variable and its
value is less than -15 mV.
- CP or DP is the primary controlled variable and its value is less than 997 mV.
c) t.OPn Lead breaking detected on temperature input.
NOTE:NOTE:
NOTE: if this error is detected, the instrument
NOTE:NOTE:
operates as in presence of a temperature input
OVERRANGE. d) tchi Out of range detected on temperature input. e) COFt Out of range detected on carbon monoxide input.
NOTE:NOTE:
NOTE: if this error is detected, the CO value is
NOTE:NOTE:
forced t o 20.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) When "CP" or "DP" is the controlled variable and an error from
a) to d) is detected, the device acts as described at "Special output features" (see page 49).
2) When "mV" is the controlled variable and the error a)
(OVERRANGE) is detected, the "Main" output is forced to 0 and the "Secondary" (if configured) is forced to 100%. The alarms and analog retransmissions operate as in presence of the maximum measurable value.
3) When "mV" is the controlled variable and the error b)
(UNDERRANGE) is detected, the "Main" output is forced to 100% and the "Secondary" (if configured) is forced to 0. The alarms and analog retransmissions operate as in presence of the minimum measurable value.
4) When "mV" is the controlled variable the errors from c) to e)
are displayed only without affecting the control action.
5) On OVER-RANGE on remote setpoint measurement is
82
signalled on lower display as “ On UNDER-RANGE on remote setpoint measurement is signalled on lower display as “- ” (indication only when required . See “DISPLAY FUNCTION”) The sensor leads break can be detected when the range selected has a zero elevation (4-20 mA or 1-5 V or 2-10 V) and signalled as “OPEn”. The sensor leads break acts as an UNDER-RANGE condition. For functioning during out of range see also “A.I.Añ – CnF.1”
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGESERROR MESSAGES At instrument start up in operative mode all the parameters are checked.
If an error is detected, the display will show the parameter group and the mnemonic code of the group with the wrong parameter setting while “Err” is shown on the central display. The device reset automatically after a time out of 6 s (20 s if serial link is enabled)
Following the normal procedure reach the group with the wrong parameter setting and correct it (during run time mode, every keystroke restarts the time out. During modify configuration parameters mode, the time out is disabled). When the error is corrected push “MENU” until the device resets (if in run time mode) or end the modify configuration parameters mode following the normal procedure. Repeat the above described procedure if another error is shown.
Error list for control action setting: E.120 Error on control parameter calculated by SMART
when the control action type has been changed in the configuration mode.
E.130 Error during the TUNE mode. The algorithm was not
able to correctly calculate the control action parameters.
83
This error can be detected also at start up in ADAPTIVE mode when the values calculated by SMART are wrong. In both cases, the instrument will be forced to work as a PI controller.
E.140 Error detected when the control action values
calculated by ADAPTIVE algorithm are out of range by the limits set of the "Hidden group".
NOTE:NOTE:
NOTE: push any key in order to remove the E.130 and E.140
NOTE:NOTE:
error indications if the instrument is in local state. If the instrument is in remote state refer to the Modbus protocol communication.
The instrument is also capable to detect the following errors: E.100 Error during data saving in FRAM E.110 Error in FRAM handling E.500 Error during auto-zero measurement E.501 Error during zero integrator measurement
NOTE:NOTE:
NOTE: when an error E.500 or E.501 is detected, all
NOTE:NOTE: input measurement are forced to OVERRANGE condition.
E.502 Error during reference junction measurement.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) If this error is detected, the temperature input is forced to OVERRANGE condition.
2) This error may be generated by an ambient temperature higher than 70 °C (158 °F) or lower than -20 °C (-4 °F).
3) When "mV" is the controlled variable this error is displayed only without affecting the control action.
When one of these errors is detected, contact your supplier. Two errors, related with the preliminary hardware setting, are detectable:
2. change V101 dip switch or end keyboard/serial link configuration.
8. wrong position of the V101 dip switch.
When one of these error is detected, the display blanks and the error code is shown on most significant digit of the upper display. Correct the V101 setting.
84
GENERAL INFORMATIONGENERAL INFORMATION
GENERAL INFORMATION
GENERAL INFORMATIONGENERAL INFORMATION
GENERAL SPECIFICATIONSGENERAL SPECIFICATIONS
GENERAL SPECIFICATIONS
GENERAL SPECIFICATIONSGENERAL SPECIFICATIONS CaseCase
Case: Polycarbonate black color; self-extinguishing degree: V-0
CaseCase according to UL 94. Front protectionFront protection
Front protection: designed for IP 65 (*) and NEMA 4X (*) for
Front protectionFront protection indoor locations (when panel gasket is installed). (*) Test were performed in accordance with IEC 529, CEI 70-1 and NEMA 250-1991 STD. InstallationInstallation
Installation: panel mounting.
InstallationInstallation Rear terminal blockRear terminal block
Rear terminal block: 54 screw terminals (screw M3, for
Rear terminal blockRear terminal block cables from φ 0.25 to φ 2.5 mm with connections diagram and safety rear cover. DimensionsDimensions
Dimensions: according to DIN 43700
DimensionsDimensions
3.78" x 3.78" (96 x 96 mm), depth 5" (128 mm). WeightWeight
Weight:750 g (full option).
WeightWeight Power supplyPower supply
Power supply:
Power supplyPower supply
- 100 V to 240 V AC 50/60 Hz (-15% to + 10% of the nominal value).
- 24 V AC/DC (
Power consumptionPower consumption
Power consumption:
Power consumptionPower consumption
- 16 VA max. (without optional I/O)
- 20 VA max. (with optional I/O)
Insulation resistanceInsulation resistance
Insulation resistance: > 100 M according to IEC 1010-1.
Insulation resistanceInsulation resistance Dielectric strengthDielectric strength
Dielectric strength: 1500 V rms according to IEC 1010-1.
Dielectric strengthDielectric strength
+ 10 % of the nominal value).
2
or from AWG 22 to AWG 14)
Common mode rejectionCommon mode rejection
Common mode rejection: 120 dB @ 50/60 Hz.
Common mode rejectionCommon mode rejection Normal mode rejectionNormal mode rejection
Normal mode rejection: 60 dB @ 50/60 Hz.
Normal mode rejectionNormal mode rejection Electromagnetic compatibility and safety require-Electromagnetic compatibility and safety require-
Electromagnetic compatibility and safety require-
Electromagnetic compatibility and safety require-Electromagnetic compatibility and safety require­mentsments
ments: This instrument is marked CE.
mentsments Therefore, it is conforming to council directives 89/336/EEC and to council directives 73/23/EEC and 93/68/EEC. Installation categoryInstallation category
Installation category: II
Installation categoryInstallation category Operative temperatureOperative temperature
Operative temperature: from 0 to 50 °C (+32 to 122 °F).
Operative temperatureOperative temperature Storage temperatureStorage temperature
Storage temperature:
Storage temperatureStorage temperature HumidityHumidity
Humidity: from 20 % to 85% RH, non condensing.
HumidityHumidity AltitudeAltitude
Altitude:
This product is not suitable for use above 2000m
AltitudeAltitude (6562ft).
85
-20 to +70 °C (-4 to 158 °F)
INPUTSINPUTS
INPUTS
INPUTSINPUTS A) A)
MAIN INPUTMAIN INPUT
A)
MAIN INPUT
A) A)
MAIN INPUTMAIN INPUT
Type:Type:
Type:
Type:Type:
- 0 to 1500 mV (when mV sensor output or oxygen are selected as primary controlled variable).
- 1000 to 1300 mV (when carbon potential or dew point is selected as primary controlled variable). NOTE:NOTE:
NOTE: if the probe input value is out of the 1000 to 1300 mV,
NOTE:NOTE: the range automatically becomes 0 to 1500 mV.
Input typeInput type
Input type: isolated from others measuring inputs and digital
Input typeInput type inputs.
Carbon range:Carbon range:
Carbon range: 0.00 to 2.00%
Carbon range:Carbon range: Dew point range:Dew point range:
Dew point range: -100 to 100 °F or -75 to 40 °C
Dew point range:Dew point range:
Oxygen range:Oxygen range:
Oxygen range: 0.0 to 25.0 %
Oxygen range:Oxygen range: Resolution:Resolution:
Resolution: 0.1
Resolution:Resolution:
mV range:mV range:
mV range: 0 to 1500 mV
mV range:mV range: Cargon resolution:Cargon resolution:
Cargon resolution: 0.01%
Cargon resolution:Cargon resolution: Dew point resolutionDew point resolution
Dew point resolution: 1°F (1°C)
Dew point resolutionDew point resolution mV resolutionmV resolution
mV resolution: 1 mV
mV resolutionmV resolution Accuracy:Accuracy:
Accuracy:
Accuracy:Accuracy: + 0.02 digits for 0.40 to 1.60 % of the carbon range. + 0.03 digits for the remaining carbon range.
Temperature drift:Temperature drift:
Temperature drift:
Temperature drift:Temperature drift:
- 350 ppm/°C when the range is 1000 to 1300 mV.
- 200 ppm/°C when the range is 0 to 1500 mV. Input impedanceInput impedance
Input impedance: > 100 M
Input impedanceInput impedance Sampling time:Sampling time:
Sampling time: 125 ms (typical).
Sampling time:Sampling time: Display updating timeDisplay updating time
Display updating time: 375 ms.
Display updating timeDisplay updating time Insulation voltage:Insulation voltage:
Insulation voltage: 500 VAC
Insulation voltage:Insulation voltage: Probe resistance:Probe resistance:
Probe resistance: up to 100 K
Probe resistance:Probe resistance: Accuracy on probe resistance measurement:Accuracy on probe resistance measurement:
Accuracy on probe resistance measurement: 2% f.s.v.
Accuracy on probe resistance measurement:Accuracy on probe resistance measurement:
B) TEMPERATURE INPUTB) TEMPERATURE INPUT
B) TEMPERATURE INPUT
B) TEMPERATURE INPUTB) TEMPERATURE INPUT Thermocouple type: Thermocouple type:
Thermocouple type: K, S, R, B
Thermocouple type: Thermocouple type: Input typeInput type
Input type: not isolated from others measuring inputs and
Input typeInput type digital inputs. Sampling time:Sampling time:
Sampling time: 1125 ms if carbon monoxide/remote set point
Sampling time:Sampling time:
input is used.
750 ms otherwise.
AccuracyAccuracy
Accuracy:
AccuracyAccuracy
+ 0.2% f.s.v. + 1 digit @ 25 °C and nominal power
supply voltage.
Temperature drift:Temperature drift:
Temperature drift: < 200 ppm/°C of full span
Temperature drift:Temperature drift: Source impedance: Source impedance:
Source impedance: 100 max.
Source impedance: Source impedance: Current for TC open detection:Current for TC open detection:
Current for TC open detection: -100 nA.
Current for TC open detection:Current for TC open detection: Cold junctionCold junction
Cold junction: automatic compensation from 0 to 50 °C.
Cold junctionCold junction Cold junction accuracy Cold junction accuracy
Cold junction accuracy : 0.1 °C/°C
Cold junction accuracy Cold junction accuracy
86
Input impedanceInput impedance
Input impedance: > 1 M
Input impedanceInput impedance CalibrationCalibration
Calibration : according to IEC 584-1.
CalibrationCalibration STANDARD RANGES TABLE
T/CT/C
T/C
T/CT/C
typetype
type
typetype
C) AUXILIARY INPUT (CARBON MONOXIDE)C) AUXILIARY INPUT (CARBON MONOXIDE)
C) AUXILIARY INPUT (CARBON MONOXIDE)
C) AUXILIARY INPUT (CARBON MONOXIDE)C) AUXILIARY INPUT (CARBON MONOXIDE) Input range: Input range:
Input range: 0/4 - 20 mA, 0/1 - 5 V or 0/2 - 10 V.
Input range: Input range: Input typeInput type
Input type: not isolated from others measuring inputs and digital
Input typeInput type
1
KK
K
KK
2
SS
S
SS
3
RR
R
RR
4
BB
B
BB
5
KK
K
KK
6
SS
S
SS
7
RR
R
RR
8
BB
B
BB
-100 / 1370 °C
- 50 /1760 °C
- 50 /1760 °C 0 /1820 °C
-150 / 2500 °F
- 60 /3200 °F
- 60 /3200 °F
32 / 3300 °F
RangesRanges
Ranges
RangesRanges
IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09 IEC 584-1: 1995-09
inputs. Sampling time:Sampling time:
Sampling time: 1125 ms.
Sampling time:Sampling time: Temperature driftTemperature drift
Temperature drift: < 300 ppm/°C of full span.
Temperature driftTemperature drift Input span:Input span:
Input span: from 0 up to 100.
Input span:Input span:
NoteNote
Note: The type of inputs are keyboard and jumper selectable
NoteNote All type of inputs are factory calibrated
The span is fixedThe span is fixed
The span is fixed from 0 upto 100 for Carbone Monoxide or
The span is fixedThe span is fixed as per Main input for Remote Set point
( 0.0 to 2.00 for carbon potential) ( 0 to 100 °F or -18 to 40 °C for Dew point ) ( 0 to 1500 for mV) )
STANDARD RANGES TABLE
Input typeInput type
Input type
Input typeInput type
0 - 200 - 20
mAmA
0 - 20
mA
0 - 200 - 20
mAmA
4 - 204 - 20
mAmA
4 - 20
mA
4 - 204 - 20
mAmA
0 -0 -
55
VV
0 -
5
V
0 -0 -
55
VV
1 -1 -
55
VV
1 -
5
V
1 -1 -
55
VV
0 -0 -
1010
VV
0 -
10
V
0 -0 -
1010
VV
2 -2 -
1010
VV
2 -
10
V
2 -2 -
1010
VV
87
ImpedanceImpedance
Impedance
ImpedanceImpedance
< 5 > 200 k > 400 k
AccuracyAccuracy
Accuracy
AccuracyAccuracy
0.2 % @ 25°C
+ 1 digit
D) LOGIC INPUTSD) LOGIC INPUTS
D) LOGIC INPUTS
D) LOGIC INPUTSD) LOGIC INPUTS The instrument is equipped with 3 logic inputs. Input typeInput type
Input type: Contact closure (voltage free).
Input typeInput type Input function:Input function:
Input function:
Input function:Input function: DIG 1 and DIG 2 can be programmed as:
- Set point selection (SP - SP2)
- Set point selection (SP3 - SP4)
- Auto/manual selection
- Output limiter activation
- Manual reset of the alarms (acknowledge).
- Locale / Remote set point selection
DIG 3 is used to start the Burn off routine. Active logic level: Active logic level:
Active logic level: Close or open programmable.
Active logic level: Active logic level:
D1) OPTIONAL LOGIC INPUTSD1) OPTIONAL LOGIC INPUTS
D1) OPTIONAL LOGIC INPUTS
D1) OPTIONAL LOGIC INPUTSD1) OPTIONAL LOGIC INPUTS The instrument may be equipped with 8 optional logic inputs (IN1 to IN8). Input typeInput type
Input type: Contact closure (voltage free).
Input typeInput type The input status can be read only through serial link.
SET POINTSSET POINTS
SET POINTS
SET POINTSSET POINTS This instrument allows to use up to 4 set points: SP, SP2, SP3 and SP4. The set point selection is possible only by logic input. Set point transferSet point transfer
Set point transfer:
Set point transferSet point transfer The transfer between one set point to another (or between two different set point values) may be realized by a step transfer or by a ramp with two different programmable rates of change (ramp up and ramp down). Slope valueSlope value
Slope value: 1 - 200 eng. unit/min or step.
Slope valueSlope value Set points limiterSet points limiter
Set points limiter: rL [r.E10] and rH [r.E11] parameters, program-
Set points limiterSet points limiter mable.
88
CONTROL ACTIONSCONTROL ACTIONS
CONTROL ACTIONS
CONTROL ACTIONSCONTROL ACTIONS AlgorithmAlgorithm
Algorithm: PID + SMART
AlgorithmAlgorithm Types:Types:
Types:
Types:Types:
- one control output (digital or analog output)
- two control outputs NOTENOTE
NOTE: the outputs can be freely selected between analog and
NOTENOTE
digital outputs.
Digital output typesDigital output types
Digital output types: Relay or SSR.
Digital output typesDigital output types Digital output action typeDigital output action type
Digital output action type: Proportional time
Digital output action typeDigital output action type Analog output typesAnalog output types
Analog output types: 20 mA.
Analog output typesAnalog output types Proportional BandProportional Band
Proportional Band: programmable from 0.5% to 999.0% of
Proportional BandProportional Band the input span. Setting a PB equal to 0 the control action becomes ON/OFF. Hysteresis (for ON/OFF control action)Hysteresis (for ON/OFF control action)
Hysteresis (for ON/OFF control action): programmable
Hysteresis (for ON/OFF control action)Hysteresis (for ON/OFF control action) from 0.1% to 10.0 % of the input span. Integral timeIntegral time
Integral time: programmable from 1 second to 20 minutes or
Integral timeIntegral time excluded. Derivative timeDerivative time
Derivative time: programmable from 1 second to 10 minutes
Derivative timeDerivative time or excluded. Integral pre-loadIntegral pre-load
Integral pre-load: programmable
Integral pre-loadIntegral pre-load
- for one control output, from 0 to 100% of the output range.
- for two control outputs, from -100 % to +100 % of the output range.
Anti reset windupAnti reset windup
Anti reset windup: from 10 % to 200 % of the input span.
Anti reset windupAnti reset windup Main output cycle timeMain output cycle time
Main output cycle time: from 1 second to 200 seconds.
Main output cycle timeMain output cycle time
Secondary output cycle timeSecondary output cycle time
Secondary output cycle time: from 1 second to 200
Secondary output cycle timeSecondary output cycle time seconds. Relative secondary output gainRelative secondary output gain
Relative secondary output gain: keyboard programmable
Relative secondary output gainRelative secondary output gain from 0.20 to 2.00 referred to the proportional band. Overlap / dead band Overlap / dead band
Overlap / dead band : keyboard programmable from - 20 %
Overlap / dead band Overlap / dead band (dead band) to + 50 % (overlap) of the proportional band. Output limitersOutput limiters
Output limiters.
Output limitersOutput limiters For the main and/or secondary control outputs it is possible to set:
- output high limits
- output low limits
- output max. rate of rise. AUTO/MANUAL modeAUTO/MANUAL mode
AUTO/MANUAL mode: selectable by front push-button or
AUTO/MANUAL modeAUTO/MANUAL mode logic input.
89
REMOTE SET POINTREMOTE SET POINT
REMOTE SET POINT
REMOTE SET POINTREMOTE SET POINT The device can be equipped with a remote set point. This feature must be configured (see “A.In.F / A.In.t / A.I.FL / A.I.Añ / L.r.Oñ “ CnF.1 and “d1.Fn / d2.Fn” CnF.5 ) If “A.I.Añ” = norñ the remote set point can be activated by selected external contact (If none digital input is configured to select from Local / Remote set point, the Remote set point
always active and always active and
will be
always active and will be
always active and always active and present on the device.) If the auxiliary input is out of range or open the control output is de-activated (see “Control Action Feature”)
If “A.I.Añ” = Cnd.A the remote set point is activated by status of auxiliary input (Remote set point if auxiliary input is in range, local set point if it is out of range or open) When remote set point is selected the decimal point at right hand of LSD on middle display is steady lit.(when device is in “Normal display mode”)
The scale for remote set point is fixed and equal to primary control variable; the set point value will be further limited by “rL/ rH”. The sampling rate is 1.125 mS The value can be filtered.
the only SET POINT the only SET POINT
the only SET POINT
the only SET POINT the only SET POINT
onlyonly
only
onlyonly
When transfer from Remote to Locale setpoint is made(by external contact), the local setpoint can be aligned to last remote setpoint value (see L.r.Oñ – CnF.1) When remote set point value is displayed on lower display (see “DISPLAY FUNCTION”) the out of range condition are displayed as follow:
OVER-RANGE ( “****”), UNDER-RANGE ( “-***”), break (“OPEn” ).
90
leads
OUTPUTSOUTPUTS
OUTPUTS
OUTPUTSOUTPUTS Out 1 and 2Out 1 and 2
Out 1 and 2
Out 1 and 2Out 1 and 2 FunctionFunction
Function: singularly programmable as
FunctionFunction
- Control output
- Alarm output TypeType
Type: Relay or SSR
TypeType Out 1 - RelayOut 1 - Relay
Out 1 - Relay
Out 1 - RelayOut 1 - Relay Relay typeRelay type
Relay type: SPDT
Relay typeRelay type Contact ratingContact rating
Contact rating: 3 A @ 250 V on resistive load.
Contact ratingContact rating Out 2 - RelayOut 2 - Relay
Out 2 - Relay
Out 2 - RelayOut 2 - Relay Relay typeRelay type
Relay type: SPST
Relay typeRelay type Contact ratingContact rating
Contact rating: 3 A @ 250 V on resistive load.
Contact ratingContact rating Out 1 and 2 - SSROut 1 and 2 - SSR
Out 1 and 2 - SSR
Out 1 and 2 - SSROut 1 and 2 - SSR Type: notType: not
Type: not isolated voltage output
Type: notType: not
- Logic level 1: 14V
+ 20% @ 20 mA max.
24 V + 20% @ 1 mA.
- Logic level 0: < 0.5 V D.C.
Out 3Out 3
Out 3
Out 3Out 3 FunctionFunction
Function: programmable as:
FunctionFunction
- Control output
- Alarm output TypeType
Type: Relay
TypeType Relay typeRelay type
Relay type: SPST
Relay typeRelay type Contact ratingContact rating
Contact rating: 2 A @ 250 V on resistive load.
Contact ratingContact rating Out 4Out 4
Out 4
Out 4Out 4 Function:Function:
Function: Burn off function
Function:Function: TypeType
Type: Relay
TypeType Relay typeRelay type
Relay type: SPST
Relay typeRelay type Contact ratingContact rating
Contact rating: 2 A @ 250 V on resistive load.
Contact ratingContact rating Out 5Out 5
Out 5
Out 5Out 5 Function:Function:
Function: Purge function
Function:Function: TypeType
Type: Relay
TypeType Relay typeRelay type
Relay type: SPST
Relay typeRelay type Contact ratingContact rating
Contact rating: 1 A @ 250 V on resistive load.
Contact ratingContact rating
91
ANALOG OUTPUTSANALOG OUTPUTS
ANALOG OUTPUTS
ANALOG OUTPUTSANALOG OUTPUTS Out 6 and 7Out 6 and 7
Out 6 and 7
Out 6 and 7Out 6 and 7 FunctionFunction
Function: Programmable as
FunctionFunction
- Linear control output
- Analog retransmission of the measured value
- Analog retransmission of the operative set point. Output typeOutput type
Output type: Isolated output programmable as
Output typeOutput type
- 0-20 mA
- 4-20 mA. ScalingScaling
Scaling: from -1999 to 9999 when used as signal
ScalingScaling retransmission. Maximum loadMaximum load
Maximum load: 600 Ω.
Maximum loadMaximum load AccuracyAccuracy
Accuracy:
AccuracyAccuracy
- 0.1 % when it is used as control output
- 0.05 % when it is used as analog retransmission. FilterFilter
Filter: It is possible to apply a first order digital filter on the
FilterFilter retransmitted value. The time constant of the filter may be programmed within 0 and 8 s.
OPTIONAL OUTPUTSOPTIONAL OUTPUTS
OPTIONAL OUTPUTS
OPTIONAL OUTPUTSOPTIONAL OUTPUTS The instrument may be equipped with 10 optional relay outputs. TypeType
Type: Relay
TypeType Relay typeRelay type
Relay type: SPST
Relay typeRelay type Contact ratingContact rating
Contact rating: 0.5 A @ 250 V on resistive load.
Contact ratingContact rating The status of the optional outputs can be set by keys and by serial link.
ALARMSALARMS
ALARMS
ALARMSALARMS Alarm actionAlarm action
Alarm action: direct or reverse function programmable.
Alarm actionAlarm action Alarm functionsAlarm functions
Alarm functions: each alarm can be configured as process
Alarm functionsAlarm functions alarm, band alarm, deviation alarm. Alarm reset/acknowledgeAlarm reset/acknowledge
Alarm reset/acknowledge: automatic or manual reset
Alarm reset/acknowledgeAlarm reset/acknowledge programmable on each alarm. Alarm maskingAlarm masking
Alarm masking: each alarm can be configured as masked
Alarm maskingAlarm masking alarm or standard alarm. This function allows you to delete false indication at instrument start up and after a set point change.
Process alarmProcess alarm
Process alarm:
Process alarmProcess alarm Operative modeOperative mode
Operative mode: Minimum or maximum programmable.
Operative modeOperative mode ThresholdThreshold
Threshold: programmable in engineering unit within the input
ThresholdThreshold range. HysteresisHysteresis
Hysteresis: programmable in engineering units from 1 to 200
HysteresisHysteresis digits.
92
Band alarmBand alarm
Band alarm
Band alarmBand alarm Operative modeOperative mode
Operative mode: Inside or outside programmable.
Operative modeOperative mode ThresholdThreshold
Threshold: two thresholds programmable:
ThresholdThreshold
- Low - from 0 to -1000 digits.
- High - from 0 to +1000 digits.
HysteresisHysteresis
Hysteresis: programmable in engineering units from 1 to 200
HysteresisHysteresis digits.
Deviation alarmDeviation alarm
Deviation alarm
Deviation alarmDeviation alarm Operative modeOperative mode
Operative mode: high or low programmable.
Operative modeOperative mode ThresholdThreshold
Threshold: programmable from - 1000 to +1000 digits.
ThresholdThreshold HysteresisHysteresis
Hysteresis: programmable in engineering units from 1 to 200
HysteresisHysteresis digits.
SERIAL INTERFACESERIAL INTERFACE
SERIAL INTERFACE
SERIAL INTERFACESERIAL INTERFACE Types: Types:
Types: Optoisolated RS 485
Types: Types: Protocol typeProtocol type
Protocol type: MODBUS, JBUS (RTU mode).
Protocol typeProtocol type Baud rateBaud rate
Baud rate: programmable from 600 to 19200 BAUD.
Baud rateBaud rate Byte formatByte format
Byte format: 8 bit.
Byte formatByte format ParityParity
Parity: even, odd or none programmable.
ParityParity Stop bitStop bit
Stop bit: one.
Stop bitStop bit AddressAddress
Address: from 1 to 255.
AddressAddress Output voltage levelsOutput voltage levels
Output voltage levels: according to EIA standard.
Output voltage levelsOutput voltage levels NOTENOTE
NOTE: The EIA standard establishes that by RS-485 interface it
NOTENOTE is possible to connect up to 30 devices with one remote master unit. The serial interface of these instruments is based on “High input impedance” transceivers; this solution allows you to connect up to 127 devices (based on the same transceiver type) with one remote master unit.
93
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
1) REMOVE POWER FROM THE POWER SUPPLY TERMINALS AND FROM RELAY OUTPUT TERMINALS
2) Remove the instrument from case.
3) Using a vacuum cleaner or a compressed air jet (max. 3 kg/
2
cm
) remove all deposit of dust and dirt which may be present on the louvres and on the internal circuits trying to be careful for not damage the electronic components.
4) To clean external plastic or rubber parts use only a cloth moistened with:
- Ethyl Alcohol (pure or denatured) [C
- Isopropil Alcohol (pure or denatured) [(CH
- Water (H
5) Verify that there are no loose terminals.
O)
2
2H5
OH] or
3)2
CHOH] or
6) Before re-inserting the instrument in its case, be sure that it is perfectly dry.
7) Re-insert the instrument and turn it ON.
94
DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT PARAMETERS
DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT RUN TIME PARAMETERSDEFAULT RUN TIME PARAMETERS
DEFAULT RUN TIME PARAMETERS
DEFAULT RUN TIME PARAMETERSDEFAULT RUN TIME PARAMETERS
A complete and consistent set of run time parameters is stored in the instrument. These data are the typical values loaded in the instrument prior to shipment from factory. This instrument allows you to load the default value of a single run time parameter group or to load all the run time parameters
A ) When it is desired to load the default parameter of a single
group, proceed as follows: A.1) By MENU pushbutton, select the desired run time
parameter group.
NOTE:NOTE:
NOTE: if the selected group is protected by security
NOTE:NOTE: code, set the assigned value by V and W pushbuttons.
A.2) By FUNC pushbutton, select the last parameter of the
selected group. The middle and lower displays will show:
A.3) By V or W pushbuttons, select the "On" indication on
the central display.
A.4) Push the FUNC pushbutton.
The default parameter loading procedure for the selected group is ended. B ) When it is desired to load the default value of all the run
time parameters proceed as follows: B.1)By MENU pushbutton, select the "Run time group dF"
[R.Lxx].
NOTE:NOTE:
NOTE: if all run time parameters are protected by security
NOTE:NOTE: code, set the assigned value by V and W pushbuttons. B.2)Push the FUNC pushbutton.
The middle and lower displays will show:
B.3)By V or W pushbuttons, select the "On" indication on
the central display. B.4)Push the FUNC pushbutton. The central display will show:
The default parameter loading procedure of all run time
parameters is ended.
A. 1
The following is a list of the default run time parameters loaded during the above procedure:
Run time group 1 [R.Axx]Run time group 1 [R.Axx]
Run time group 1 [R.Axx]
Run time group 1 [R.Axx]Run time group 1 [R.Axx] PARAMETER DEFAULT VALUE SP = Set point low limit ("rL" [R.E10]) SP2 = Set point low limit ("rL" [R.E10]) SP3 = Set point low limit ("rL" [R.E10]) SP4 = Set point low limit ("rL" [R.E10]) COF = 200 H2F = 400
Run time group 3 [R.Cxx]Run time group 3 [R.Cxx]
Run time group 3 [R.Cxx]
Run time group 3 [R.Cxx]Run time group 3 [R.Cxx] PARAMETER DEFAULT VALUE ñ.rSt =OFF AL1 = Initial range value (for process alarm)
= 0 (for deviation alarm) bA1.L = -10 bA1.h = 10 AL2 = Initial range value (for process alarm)
= 0 (for deviation alarm) bA2.L = -10 bA2.h = 10 AL3 = Initial range value (for process alarm)
= 0 (for deviation alarm)
bA3.L = -10 bA3.h = 10 HSA1 = 1 HSA2 = 1 HSA3 = 1
Run time group 4 [R.Dxx]Run time group 4 [R.Dxx]
Run time group 4 [R.Dxx]
Run time group 4 [R.Dxx]Run time group 4 [R.Dxx] PARAMETER DEFAULT VALUE Pb = 10.0 % (if PID control is configured)
= 15.0 % (if PI control is configured) HYS = 0.5 % ti = 1.50 mm.ss (if PID control is configured)
= 2.45 mm.ss (if PI control is configured) td = 1.00 mm.ss I P = 50.0 (If only one control output is configured).
0.0 (If two control outputs are configured). r.Gn = 1.00 OLAP =0
Run time group 5 [R.Exx]Run time group 5 [R.Exx]
Run time group 5 [R.Exx]
Run time group 5 [R.Exx]Run time group 5 [R.Exx] PARAMETER DEFAULT VALUE ArW = 100% ñ.OLL = 0.0% ñ.OLH =100.0% ñ.rñP =Inf
A. 2
ñC.CY = 16 s (If relay output)
2 s (If SSR output) S.OLL = 0.0% S.OLH = 100.0% S.rñP = Inf SC.CY = 16 s (If relay output)
2 s (If SSR output) rL = Initial range value rH = Final range value Grd1 = Inf Grd2 = Inf E.Añ = On
Run time group 6 [R.Fxx]Run time group 6 [R.Fxx]
Run time group 6 [R.Fxx]
Run time group 6 [R.Fxx]Run time group 6 [R.Fxx] PARAMETER DEFAULT VALUE A1.tP = Proc A1.Cn = H.A. A1.Ac = rEV A1.St = OFF A2.tP = Proc A2.Cn = H.A. A2.Ac = rEV A2.St = OFF A3.tP = bAnd A3.Cn = H.A.
A3.Ac = rEV A3.St = OFF
Run time group 7 [R.Gxx]Run time group 7 [R.Gxx]
Run time group 7 [R.Gxx]
Run time group 7 [R.Gxx]Run time group 7 [R.Gxx] PARAMETER DEFAULT VALUE S.L.Pr = ñbUS S.L.Ad = 1 S.L.bd = 19.20 S.L.bF = 8
Run time group 8 [R.Hxx]Run time group 8 [R.Hxx]
Run time group 8 [R.Hxx]
Run time group 8 [R.Hxx]Run time group 8 [R.Hxx] PARAMETER DEFAULT VALUE t.bOF = 3.00 mm. ss. t.PrG = 1.00 mm. ss. bF.tr = 12.00 hh.mm pb.tr = 24.00 hh.mm
Run time group Hd [r.Lxx]Run time group Hd [r.Lxx]
Run time group Hd [r.Lxx]
Run time group Hd [r.Lxx]Run time group Hd [r.Lxx] PARAMETER DEFAULT VALUE Pb.Lo = 2.0% Pb.Hi =999.0% ti.Lo =00.01 mm.ss ti.Hi =20.00 mm.ss rG.CL =OFF
A. 3
DEFAULT CONFIGURATIONDEFAULT CONFIGURATION
DEFAULT CONFIGURATION
DEFAULT CONFIGURATIONDEFAULT CONFIGURATION PARAMETERSPARAMETERS
PARAMETERS
PARAMETERSPARAMETERS
A complete and consistent set of configuration parameters is stored in the instrument. These data are the typical values loaded in the instrument prior to shipment from factory. When it is desired to load the default value of all the configura­tion parameters, proceed as follows:
If the instrument starts in configuration mode, push the MENU pushbutton. If the instrument starts in run time mode, by keeping depressed the MENU push-button for more than 5 seconds the instrument will show:
NOTENOTE
NOTE: If no push-button is depressed for more than 10 s
NOTENOTE (or 30 s according to "CnF.6" "t.out" [time out selection" C.I10] parameter setting), the instrument returns automatically to the normal display mode.
By V or W push-button select "ñodF."
NOTESNOTES
NOTES:
NOTESNOTES
1) When modify mode is started, the instrument stops the control and:
- sets to OFF the control outputs;
- turns to OFF the bargraph displays;
- sets analog retransmissions to the retransmitted initial
scale value;
- sets to OFF the alarms;
- disables the serial link;
- removes the time out.
2) When the modify mode is disabled by V101 (SW3), the V or W push-button pressure has no effect.
Push MENU pushbutton again and select the "Default configuration group" [C.Cxx]. NOTE:NOTE:
NOTE: if the configuration parameters are protected by security
NOTE:NOTE:
code, by V or W pushbuttons set the security code
assigned and press the FUNC pushbutton. By V or W push-button select the desired configuration parameter set "tb.1" (european) or "tb.2" (american) Push MENU pushbutton again The central display will show:
A. 4
Loading...