This instrument is intended for permanent installation, for indoor
use only, in an electrical panel which encloses the rear housing,
exposed terminals and wiring on the back.
Select a location, for instrument mounting, where minimum
vibrations are present and the ambient temperature is within 0
and 50 °C (32 and 122°F).
The instrument can be mounted on a panel up to 15 mm thick
with a cutout 92 x 92.
For outline and cutout dimensions refer to Fig. 2.
The surface texture of the panel must be better than 6,3 µm.
The instrument is shipped with rubber panel gasket (50 to
60 Sh).
To assure the IP65 and NEMA 4X protection, insert the panel
gasket between the instrument and the panel as shown in
fig. 1.
While holding the instrument against the panel proceed as
follows:
1) insert the gasket in the instrument case;
2) insert the instrument in the panel cutout;
3) pushing the instrument against the panel;
4) insert the mounting brackets as shown in fig.1;
5) with a screwdriver, turn the screws with a torque between
0.3 and 0.4 Nm.
Screws
Bracket
Fig. 1
1
Bracket
Gasket
Panel
OUTLINE AND CUT OUT DIMENSIONSOUTLINE AND CUT OUT DIMENSIONS
"
OUTLINE AND CUT OUT DIMENSIONS
OUTLINE AND CUT OUT DIMENSIONSOUTLINE AND CUT OUT DIMENSIONS
+0.03 +0.8
3.62" (92 mm )
+0.03
+0.8
3.62"
(92 mm )
(125 mm)
4.92"
(125 mm)
4.92
Fig. 2. OUTLINE AND CUT-OUT DIMENSIONS
3.78" (96 mm)
0.6" (15mm) max
4.17" (108.5 mm)
5" (128 mm)
0.06" (1.5 mm)
3.8" x 3.8" (97 x 97 mm) gasket
0.63"
(16 mm)
2
CONNECTION DIAGRAMSCONNECTION DIAGRAMS
CONNECTION DIAGRAMS
CONNECTION DIAGRAMSCONNECTION DIAGRAMS
Fig. 3. REAR TERMINAL BLOCK
3
CONNECTIONSCONNECTIONS
CONNECTIONS
CONNECTIONSCONNECTIONS
Connections are to be made when the instrument housing is
NOTE: Any external component (like zener barriers etc.)
NOTENOTE
connected between sensor and input terminals may cause errors
in measurement due to excessive and/or not balanced line
resistance or possible leakage currents.
12
13
+
_
G
Fig. 4 MAIN INPUT WIRING
NOTESNOTES
NOTES:
NOTESNOTES
1) This input is isolated from others measuring inputs
and digital inputs.
2) Don’t run input wires together with power cables.
3) When a shielded cable is used, it should be connected at
one point only.
4
Sensor
A.1) TEMPERATURE (THERMOCOUPLE) INPUTA.1) TEMPERATURE (THERMOCOUPLE) INPUT
A.1) TEMPERATURE (THERMOCOUPLE) INPUT
A.1) TEMPERATURE (THERMOCOUPLE) INPUTA.1) TEMPERATURE (THERMOCOUPLE) INPUT
1
+
_
3
Shield
1
+
_
3
Shield
Fig. 5 TEMPERATURE INPUT WIRING
NOTESNOTES
NOTES:
NOTESNOTES
1) This input is not isolated from others measuring inputs
and digital inputs.
2) Don’t run input wires together with power cables.
3) For TC wiring use proper compensating cable preferable
shielded.
4) When a shielded cable is used, it should be connected at
one point only.
1) This input is not isolated from others measuring inputs
and digital inputs. A double or reinforced insulation between
instrument output and power supply must be assured by the
external instrument.
5
+
mA,
or
_
V
+
mA
or
_
V
G
2) Don’t run input wires together with power cables.
3) Pay attention to the line resistance; a high line resistance
may cause measurement errors.
4) When shielded cable is used, it should be grounded at one
side only to avoid ground loop currents.
5) The input impedance is equal to:
< 5 Ω for 20 mA input
> 200 kΩ for 5 V input
> 400 kΩ for 10 V input
B) LOGIC INPUTSB) LOGIC INPUTS
B) LOGIC INPUTS
B) LOGIC INPUTSB) LOGIC INPUTS
DIG. 1
8
DIG. 2
9
DIG. 3
10
11
IN 1
57
IN 2
58
IN 3
59
IN 4
60
56
Fig. 7.A - LOGIC INPUTS DIG 1 TO DIG 3 AND IN1 TO IN4
WIRING
6
IN 5
46
IN 6
47
IN 7
48
IN 8
49
45
Fig. 7.B - LOGIC INPUTS IN5 TO IN8 WIRING
Logic inputs from IN1 to In8 are optional.
NOTESNOTES
NOTES:
NOTESNOTES
1) Do not run logic input wiring together with power cables.
2) Use an external dry contact capable of switching 0.5 mA,
5 V DC.
3) The instrument needs 110 ms to recognize a contact status
variation.
4) The logic inputs are
NOT NOT
NOT isolated from measuring inputs.
NOT NOT
C.1) RELAY OUTPUTSC.1) RELAY OUTPUTS
C.1) RELAY OUTPUTS
C.1) RELAY OUTPUTSC.1) RELAY OUTPUTS
OUT 1
OUT 2
OUT 3
Fig. 8.A RELAY OUTPUTS 1, 2, 3 WIRING
7
23
24
25
26
27
28
29
NC
C
NO
C
NO
C
NO
OUT 4
30
31
OUT 5
15
14
OUT 10
OUT 11
OUT 12
OUT 13
OUT 14
61
62
63
64
65
66
Fig. 8.B RELAY OUTPUTS 4, 5 and 10 TO 14 WIRING
C
NO
C
NO
NO - OUT 10
NO - OUT 11
NO - OUT 12
NO - OUT 13
NO - OUT 14
C - OUT 10 to 14
NO - OUT 15
OUT 15
OUT 16
OUT 17
OUT 18
OUT 19
Fig. 8.C RELAY OUTPUTS 15 TO 19 WIRING
Outputs from OUT 10 to OUT 19 are optional.
The contact rating from OUT 1 and OUT 2 is 3A/250V AC on
resistive load.
The contact rating from OUT 3 and OUT 4 is 2A/250V AC on
resistive load.
The contact rating of OUT 5 is 1A/250V AC on resistive load.
The contact rating from OUT 10 to OUT 19 is 0.5 A/250V AC on
resistive load.
The number of operations is 1 x 105 at specified rating.
8
50
NO - OUT 16
51
NO - OUT 17
52
NO - OUT 18
53
NO - OUT 19
54
C - OUT 15 to 19
55
NOTES:NOTES:
NOTES:1)To avoid electrical shock, connect
NOTES:NOTES:
All relay contacts are protected by varistor against inductive load
with inductive component up to 0.5 A.
The following recommendations avoid serious problems which
may occur, when relay output drives inductive loads.
INDUCTIVE LOADSINDUCTIVE LOADS
INDUCTIVE LOADS
INDUCTIVE LOADSINDUCTIVE LOADS
High voltage transients may occur switching inductive loads.
Through the internal contacts these transients may introduce
disturbances which can affect the performance of the instrument.
For all the outputs, the internal protection (varistor) assures a
correct protection up to 0.5 A of inductive component.
The same problem may occur when a switch is used in series
with the internal contacts as shown in Fig. 9.
power line at the end of the wiring procedure.
2) For power connections use No 16 AWG or larger wires
rated for at last 75 °C.
3) Use copper conductors only.
4) Don’t run input wires together with power cables.
5) When a relay output is used to drive a low power
signal (PLC input, alarm annunciator, etc..) it is
necessary to use an external relay with a gilded
contact.
C
R
LOAD
Fig. 9 EXTERNAL SWITCH IN SERIES WITH THE INTERNAL
CONTACT
In this case it is recommended to install an additional RC
network across the external contact as show in Fig. 9.
The value of capacitor (C) and resistor (R) are shown in the
following table.
LOAD
(mA)
<40 mA
<150 mA
<0.5 A
Anyway the cable involved in relay output wiring must be as far
away as possible from input or communication cables.
9
C
(µF)
0.047
0.1
0.33
(Ω)
100
22
47
R
P.
OPERATING
(W)
VOLTAGE
1/2
260 V AC
2
260 V AC
2
260 V AC
LINE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVEC.2) VOLTAGE OUTPUTS FOR SSR DRIVE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVE
C.2) VOLTAGE OUTPUTS FOR SSR DRIVEC.2) VOLTAGE OUTPUTS FOR SSR DRIVE
NOTENOTE
A double or reinforced insulation between instrument output and
power supply must be assured by the external solid state relay.
C.3) ANALOG OUTPUTSC.3) ANALOG OUTPUTS
C.3) ANALOG OUTPUTS
C.3) ANALOG OUTPUTSC.3) ANALOG OUTPUTS
+
OUT 6
16
_
17
+
20 mA
_
Shield
+
16
OUT 6
_
17
Fig. 11.A OUTPUT 6 WIRING
10
+
_
20 mA
G
+
OUT 7
18
_
19
+
_
20 mA20 mA
Shield
+
18
OUT 7
_
19
Fig. 11.B OUTPUT 7 WIRING
NOTESNOTES
NOTES:
NOTESNOTES
1) Do not run analog output wires together with AC power
cables.
2) Out 6 and 7 are isolated outputs.
3) The maximum load is equal to 600 Ω.
+
_
G
D) SERIAL INTERFACED) SERIAL INTERFACE
D) SERIAL INTERFACE
D) SERIAL INTERFACED) SERIAL INTERFACE
RS-485 interface allows you to connect up to 30 devices with
one remote master unit.
I
N
S
T
R
U
M
E
N
T
Fig. 12 - RS-485 WIRING
The cable length must not exceed 1.5 km at 9600 BAUD.
NOTESNOTES
NOTES:
NOTESNOTES
1) This is an isolated RS 485 serial interface.
2) The following report describes the signal sense of the voltage
appearing across the interconnection cable as defined by
EIA for RS-485.
a) The ” A ” terminal of the generator shall be negative with
respect to the ” B ” terminal for a binary 1 (MARK or OFF)
state.
11
22
21
20
A/A'
B/B'
COMMON
A'/A
B'/B
M
A
S
T
E
R
b) The ” A ” terminal of the generator shall be positive with
respect to the ” B ” terminal for a binary 0 (SPACE or ON).
3) The EIA standard establishes that by RS-485 interface it is possible
to connect up to 30 devices with one remote master unit.
The serial interface of these instruments is based on “High
input impedance” transceivers; this solution allows you to
connect up to 127 devices (based on the same transceiver
type) with one remote master unit.
E) POWER LINE WIRINGE) POWER LINE WIRING
E) POWER LINE WIRING
E) POWER LINE WIRINGE) POWER LINE WIRING
N (L2)
32
POWER LINE 100 V to
240 V A.C (50/60Hz)
33
L (L1)
Line
Fig. 13 POWER LINE WIRING
NOTESNOTES
NOTES:
NOTESNOTES
1) Before connecting the instrument to the power line, make
sure that line voltage corresponds to the description on the
identification label.
2) To avoid electrical shock, connect power line at the end of
the wiring procedure.
3) For supply connections use No 16 AWG or larger wires rated
for at last 75 °C.
12
Neutral
4) Use copper conductors only.
5) Don’t run input wires together with power cables.
6) The power supply input is fuse protected by a sub miniature
fuse rated T, 1A, 250 V.
When fuse is damaged, it is advisable to verify the power
supply circuit, so that it is necessary to send back the
instrument to your supplier.
7) The safety requirements for Permanently Connected
Equipment say:
- a switch or circuit-breaker shall be included in the building
installation;
- it shall be in close proximity to the equipment and within
easy reach of the operator;
- it shall be marked as the disconnecting device for the
equipment.
NOTENOTE
NOTE: a single switch or circuit-breaker can drive more than
NOTENOTE
one instrument.
8) When a neutral line is present please connect it to the 32
terminal.
AUXILIARY INPUT SELECTIONAUXILIARY INPUT SELECTION
Set J102 (see fig. 15) according to the desired input type as
shown in the following table.
J102INPUT TYPE
5 V10 V20 mA
1-2closeopenopen
3-4closecloseopen
5-6openopenclose
7-8openopenclose
5-7opencloseopen
6-8openopenopen
1 3 5 7
J102
2 4 6 8
1 3 5 7
J103
2 4 6 8
CPU
card
Fig.15
15
Operative mode and Hardware lockOperative mode and Hardware lock
Operative mode and Hardware lock
Operative mode and Hardware lockOperative mode and Hardware lock
1) By V101 (see fig. 16) it is possible to select one of the
following operative modes:
a) run time mode without configuration mode
b) run time and configuration modes
c) security code setting mode
Set V101 according to the following table:
2) When run time mode is selected (mode a or b), V101.3 allows
you to activate/deactivate the hardware lock for configuration
parameters.
If V101.3 is ON, the lock is activated.
If V101.3 is OFF, the lock is deactivated.
When the lock is activated, no one of the configuration
parameter can be modified.
3) All the others switch combinations are reserved.
SECURITY CODE SETTING MODESECURITY CODE SETTING MODE
General notesGeneral notes
General notes
General notesGeneral notes
The instrument parameters are divided in two families and each
family is divided in groups.
- The first family encompasses all the run time parameters.
- The second family comprises all the configuration parameters.
A specific security code enables the parameter modification of
each family.
For run time parameters, it is possible to select which groups of
them will be protected by the security code and in this case it is
necessary to set the run time security code before to modify one
or more parameters of a protected group.
The configuration security code protects all configuration
parameters and it will be necessary to set the configuration
security code before to start the configuration parameters
modification.
For configuration parameters an hardware lock is also available.
Security code settingSecurity code setting
Security code setting:
Security code settingSecurity code setting
1) Remove the instrument from its case.
2) Set the internal dip switch V101 as follows:
- V101.1 = OFF - V101.2 = ON
- V101.3 = OFF - V101.4 = OFF
3) Re-insert the instrument.
4) Switch on the instrument. The display will show:
The upper display shows that the security code setting mode
is selected while the lower display shows the firmware
version.
5) Push the FUNC pushbutton.
17
Run time security codeRun time security code
Run time security code
Run time security codeRun time security code
The display will show:
Run time groups protected by security codeRun time groups protected by security code
Run time groups protected by security code
Run time groups protected by security codeRun time groups protected by security code
The display will show:
NoteNote
Note: the middle display shows the current status of the run
NoteNote
time security code ("0", "1" or "On").
By and push-button, set "S.run" parameter as follows:
0No protection (it is always possible to modify all run
time parameters);
1always protected (it is never possible to modify a run
time parameter);
from 2 to 250 security code for run time parameter
protection.
NOTESNOTES
NOTES:
NOTESNOTES
1) The selected value of a security code cannot be displayed
anymore and, coming back to the "S.run" parameter, the display
will show "On" when "S.run" is different from 0 or 1, "0" when
"S.run" is equal to 0, "1" when "S.run" is equal to 1.
When the security code is forgotten, a new value can be set.
2) When "S.run" is different from 0 or 1, the "run time default "
and the "run time hidden" groups are always protected by
security code.
By this parameter it is possible to set if the run time group 1 will
be protected or not by the run time security code.
By and push-button, set "Gr1" parameter as follows:
nONo protection (it is always possible to modify run time
group 1 parameters).
Yes the run time group 1 parameter modification will be
protected by security code.
Push the FUNC push-button; the instrument memorizes the new
setting and goes to the next parameter.
NOTESNOTES
NOTES: 1) This selection may be performed only if a run time
NOTESNOTES
security code has been set (from 2 to 250).
2) The above described selection may be repeated for
all groups of the run time mode.
Configuration security codeConfiguration security code
The display will show:
NoteNote
Note: the middle display shows the current status of the
NoteNote
configuration security code ("0", "1" or "On").
By and push-button, set "S.CnF" parameter as follows:
0No protection (it is always possible to modify all
configuration parameters);
1always protected (it is never possible to modify a
configuration parameter);
from 2 to 250 security code for configuration parameter
protection.
NOTESNOTES
NOTES:1) The selected value of a security code cannot be
NOTESNOTES
displayed anymore and, coming back to the "S.CnF"
parameter, the display will show "On" when "S.CnF"
is different from 0 or 1, "0" when "S.CnF" is equal to
0, "1" when "S.CnF" is equal to 1.
When the security code is forgotten, a new value can
be set.
2) At the end of the security code setting, set the V101
as described at page16.
RUN TIME AND CONFIGURATION MODESRUN TIME AND CONFIGURATION MODES
RUN TIME AND CONFIGURATION MODES
RUN TIME AND CONFIGURATION MODESRUN TIME AND CONFIGURATION MODES
The hardware selection described in "Operative mode and
hardware lock" paragraph allows you to start one of the two
following operative modes:
- run time mode
- configuration mode.
At power up, the instrument starts in the same mode it was prior
to the power OFF.
General note about graphic symbols used forGeneral note about graphic symbols used for
General note about graphic symbols used for
General note about graphic symbols used forGeneral note about graphic symbols used for
mnemonic code visualization.mnemonic code visualization.
mnemonic code visualization.
mnemonic code visualization.mnemonic code visualization.
The instrument displays some characters with special symbols.
The following table shows the correspondence between the
symbols and the characters.
19
symbolcharacter
"
"k
" "m
" "V
" "W
" "Z
"J
"
Keyboard descriptionKeyboard description
Keyboard description
Keyboard descriptionKeyboard description
MENU= is used to select a parameter group.
FUNC = When the instrument is in "normal display mode" it
changes the indication on the lower display (see
"display function").
During parameter modification, it allows you to
memorize the new value of the selected parameter
and go to the next parameter (increasing order).
MAN = When the instrument is in "normal display mode",
pushing MAN push-button for more than 1 s, it is
possible to enable or disable the manual function.
During parameter modification, it allows you to scroll
back the parameters and groups without memorizing
the new setting.
= During parameter modification, it allows you to
increase the value of the selected parameter.
During MANUAL mode, it allows you to increase the
output value.
TST + MAN = By pressing these keys it is possible to start the
BURNOFF function.
TST + FUNC = By pressing these keys it is possible to start the
PROBE TEST function.
+MENU= Are used to start the lamp test function (the
command is accepted when push-buttons are
kept depressed for more than 5 s and the
instrument is in normal display mode)
+FUNC or +FUNC
During parameter modification they allow you to
increase/decrease the value under modification with
higher rate.
+MAN or +MAN
During parameter modification they allow you to
jump to the max or min programmable value.
= During parameter modification, it allows you to
decrease the value of the selected parameter.
During MANUAL mode, it allows to decrease
the output value.
20
NOTESNOTES
NOTES:
NOTESNOTES
1) All the actions explained above which requires two or more
push-buttons must follow exactly the push-button sequence
shown.
2) A 10 or 30 seconds time out (see "CnF.6 - t.out" [C.I08]) can
be selected for parameter modification during run time mode.
If, during parameter modification, no push-button is depressed
for more than 10 (30) seconds, the instrument goes
automatically to the “normal display mode” and the
modification (if carried out) of the last displayed parameter will
be lost.
CONFIGURATION MODECONFIGURATION MODE
CONFIGURATION MODE
CONFIGURATION MODECONFIGURATION MODE
Switch on the instrument.
The instrument will start in the same way it was prior to the
power down (configuration mode or run time mode).
If the instrument starts in configuration mode, push the MENU
pushbutton and go to the "Configuration group 1" (see page 25).
If the instrument starts in run time mode, by keeping depressed
the MENU push-button for more than 5 seconds the instrument
will show:
where:
- the upper display shows the selected parameter family;
- the middle display shows the selected action;
- the lower display shows the firmware version.
If no push-button is depressed for more than 10 s (or 30 s
according to "CnF.6 - "t.out" [time out selection" C.I08]
parameter setting), the instrument automatically returns to the
normal display mode.
21
By or push-button it is possible to select between:
= (monitor) this selection allows you to monitor but
not to modify the value assigned to the
configuration parameters.
=(modify) this selection allows you to monitor and to
modify the value assigned to the configuration
parameters.
NOTESNOTES
NOTES:
NOTESNOTES
1) During monitor mode, the instrument continues to operate as
in run time mode.
2) When modify mode is started, the instrument stops the control
and:
- sets to OFF the control outputs;
- turns to OFF the bargraph displays;
- sets analog retransmissions to the retransmitted initial scale
value;
- sets to OFF the alarms;
- disables the serial link;
- the time out will be disabled.
3) When the modify mode is disabled by V101 (V101.3), the or
push-button pressure has no effect.
MONITOR MODEMONITOR MODE
MONITOR MODE
MONITOR MODEMONITOR MODE
During the run time mode, it is possible to monitor but not modify
all configuration parameters.
When it is desired to verify the instrument configuration, proceed
as follows:
1) By or push-button select the monitor mode.
2) Push the MENU push-button the display will show:
it shows that configuration group 1 is selected and it
encompasses all the input parameters.
The configuration parameter "Monitor mode" follows the "Modify
mode" sequence.
NOTESNOTES
NOTES:
NOTESNOTES
1) During monitor mode, the instrument continues to operate as
in run time mode.
2) During monitor mode, if no push-button is depressed for more
than 10 s (or 30 s according to "CnF.6 - t.out" [C.I08]
parameter setting), the instrument returns automatically to the
normal display mode.
22
MODIFY MODEMODIFY MODE
MODIFY MODE
MODIFY MODEMODIFY MODE
1) By or push-button select the modify mode.
2) Push the MENU push-button.
If a security code is applied to the configuration parameter,
the instrument will show:
3) By and push-button set a value equal to the security
code assigned to the configuration mode (see "Configuration
security code " at page 19).
If the code is different from the security code, the instrument
automatically returns to the first configuration display
otherwise the display will show:
The modify mode is started.
This display allows you to load the default configuration
parameter (table 1 or table 2).
For more details see chapter "Default parameter" (see
Appendix A).
4) By or push-button select the OFF indication and push
the MENU push-button.
The display will show:
This is the starting display of the first group of configuration
parameters.
NOTESNOTES
NOTES:
NOTESNOTES
1) In the following pages we will describe all the parameters of
the instrument but the instrument will show only the
parameters related with the specific hardware and in
accordance with the specific instrument configuration [i.e.
setting OUT 6 as "nonE" (not used), all the parameters related
with this output will be automatically skipped].
23
2) During configuration parameters modify mode, the upper
display shows the selected parameter group, the lower display
shows the mnemonic code of the selected parameter while
the central display shows the value or status assigned to the
selected parameter.
3) For an easy consultation of this manual, a sheet named
"Reference parameter guide" with all the parameter
visualizations is enclosed.
The groups of configuration parameters are identified by the
"C" letter followed by A, B etc.
The "code" formed by the group and the row (i.e. C.D01
where "C.D" is configuration group1 and "01" is the row 1) is
reported, in the user manual, before each parameter
description and allows you to quickly find out the respective
parameter.
CONFIGURATION GROUP 1 [C.Dxx]CONFIGURATION GROUP 1 [C.Dxx]
CONFIGURATION GROUP 1 [C.Dxx]
CONFIGURATION GROUP 1 [C.Dxx]CONFIGURATION GROUP 1 [C.Dxx]
INPUT CONFIGURATION
Push the FUNC push-button
C.D01- Line frequencyC.D01- Line frequency
C.D01- Line frequency
C.D01- Line frequencyC.D01- Line frequency
Upper display: CnF.1
Lower display: Ln.Fr
Range: 50 Hz
C.D02- Controlled variable selectionC.D02- Controlled variable selection
Upper display: CnF.1
Lower display: PV.SL
Ranges:
CP = Carbon potential (span limits within 0.00% to 2.00%).
dP = Dew point (span limits within -100 to 100 °F or -75 to 40°C).
ñV = Sensor output in mV (span limits within 0 to 1500 mV).
O2 = Oxygen value as primary control variable
( The span limits are 0.0 to 25.0)
24
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) Changing the controlled variable, the following parameters:
"CnF.6 - brG.L" [C.I03] and "CnF.6 - brG.H" [C.I04] (Bargraph
initial and full scale values);
"CnF.2 - O6.Lr" [C.E06] and "CnF.2 - O6.Hr" [C.E07] (Out 6
retransmission initial and full scale values);
"CnF.2 - O7.Lr" [C.E11] and "CnF.2 - O7.Hr" [C.E12] (Out 7
retransmission initial and full scale values);
"Gr.5 - rL" [R.E10] and "Gr.5 - rH" [R.E11]
(set point low and high limits); will be forced to the span limits
of the new selected variable.
2) SP, SP2, SP3, SP4 values and alarm thresholds, if out of the
limits for the new selected variable, will be forced to the low
limit value.
C.D03- Filter on probe sensor input (Main input)C.D03- Filter on probe sensor input (Main input)
C.D03- Filter on probe sensor input (Main input)
C.D03- Filter on probe sensor input (Main input)C.D03- Filter on probe sensor input (Main input)
Upper display: CnF.1
Lower display: Pb.FL
Ranges: from 0 (no filter) to 8 seconds.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) This is a first order digital filter applied on probe sensor input
value.
2) This filter can affect the control action, the alarms, the SMART
algorithm and the process variable retransmission.
C.D04 - Temperature input - TC type selectionC.D04 - Temperature input - TC type selection
C.D04 - Temperature input - TC type selection
C.D04 - Temperature input - TC type selectionC.D04 - Temperature input - TC type selection
Upper display: CnF.1
Lower display: tP.In
Ranges:
1 = TC KFrom -100 t o 1370 °C
2 = TC SFrom -50 to 17 6 0 °C
3 = TC RFrom -50 to 1760 °C
4 = TC KFrom -150 to 2500 °F
5 = TC SFrom -60 to 32 0 0 °F
6 = TC RFrom -60 to 3200 °F
7= TC BFrom 0 to 1820 °C
8= TC BFrom 32 to 3300 °F
NOTE:NOTE:
NOTE: When the controlled variable selected is dP, the span
NOTE:NOTE:
limits of the following parameters:
"CnF.6 - brG.L" [C.I03] and "CnF.6 - brG.H" [C.I04]
(Bargraph initial and full scale values);
"CnF.2 - O6.Lr" [C.E06] and "CnF.2 - O6.Hr" [C.E07] (Out
6 retransmission initial and full scale values);
"CnF.2 - O7.Lr" [C.E11] and "CnF.2 - O7.Hr" [C.E12] (Out
7 retransmission initial and full scale values);
"Gr.5 - rL" [R.E10] and "Gr.5 - rH" [R.E11]
(set point low and high limits); will be forced to:
-100 to 100 °F if the temperature input has been
changed from a °C to a °F type;
-75 to 40 °C if the temperature input has been
changed from a °F to a °C type.
25
C.D05 - Temperature input offset adjustmentC.D05 - Temperature input offset adjustment
C.D05 - Temperature input offset adjustment
C.D05 - Temperature input offset adjustmentC.D05 - Temperature input offset adjustment
Upper display: CnF.1
Lower display: OFSt
Range: from -500 to 500
NOTENOTE
NOTE: The offset value is algebraically added to temperature
NOTENOTE
input value.
Read-out
OFSt
Real curve
Adjusted
curve
Input
C.D06 - Filter on temperature input valueC.D06 - Filter on temperature input value
C.D06 - Filter on temperature input value
C.D06 - Filter on temperature input valueC.D06 - Filter on temperature input value
Upper display: CnF.1
Lower display: tP.FL
Ranges: from 0 (no filter) to 8 seconds.
NOTES:NOTES:
NOTES:
NOTES:NOTES:
1) This is a first order digital filter applied to the temperature
input value.
2) When "dP" or "CP" or O2 are selected as the controlled
variable, this filter can affect the control action, the alarms,
the SMART algorithm and the process variable retransmission.
C.D07 - Auxiliary input functionC.D07 - Auxiliary input function
C.D07 - Auxiliary input function
C.D07 - Auxiliary input functionC.D07 - Auxiliary input function
Lower display: A.In.F
Range: nonE = Input not used
CO = Input used for CO measurement
rSP = Input used as remote set point
Available: Always
Note:The scalable for CO measurement is fixed from 000 to 100
When CO input is not used the CO value will be set to 20
The scalable for rSP measurement is fixed to:
0.00 / 2.00 if carbon potential is selected as primary control variable or
0 /100 °F (-18 / 40 °C) if dew point is selected as primary control variable or
0 / 1500 if sensor output in mV is selected as primary control variable
0.0/ 25.0 if oxygen value is selected as primary control variable
C.D08 - Auxiliary input type selectionC.D08 - Auxiliary input type selection
C.D08 - Auxiliary input type selection
C.D08 - Auxiliary input type selectionC.D08 - Auxiliary input type selection
Lower display:A.In.t
Range:0-20 = 0÷20 mA
4-20 = 4÷20 mA
0- 5 = 0÷ 5 V
1- 5 = 1÷ 5 V
0-10 = 0÷10 V
2-10 = 2÷10 V
Available:When auxiliary input is used
26
C.D09 - TIME CONSTANT FOR FILTER ON REMOTEC.D09 - TIME CONSTANT FOR FILTER ON REMOTE
C.D09 - TIME CONSTANT FOR FILTER ON REMOTE
C.D09 - TIME CONSTANT FOR FILTER ON REMOTEC.D09 - TIME CONSTANT FOR FILTER ON REMOTE
SETPOINT VALUE (CnF.1)SETPOINT VALUE (CnF.1)
SETPOINT VALUE (CnF.1)
SETPOINT VALUE (CnF.1)SETPOINT VALUE (CnF.1)
Lower display:A.I.FL
Range:From 0(filter OFF) to 8 s
Available:When auxiliary input is used as remote
set point
Note: First order filter with selected time constant
C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)C.D10 - REMOTE SETPOINT ACTIVITY MODE (CnF.1)
Lower display:A.I.Añ
Range:norñ = The remote set point is
activated by external contact(or is always activated if none
contacts are configured for this feature) and the value is limited
at its min/max value when out of range
Cnd.A = The remote set point
activation is controlled by status of auxiliary input.
(The device works with local set point when auxiliary input is
out of range while it works with remote set point when auxiliary
input is in range)
Available: When auxiliary input is used as remote set point
C.D11 - LOCAL / REMOTE SETPOINT OPERATINGC.D11 - LOCAL / REMOTE SETPOINT OPERATING
C.D11 - LOCAL / REMOTE SETPOINT OPERATING
C.D11 - LOCAL / REMOTE SETPOINT OPERATINGC.D11 - LOCAL / REMOTE SETPOINT OPERATING
MODE (CnF.1)MODE (CnF.1)
MODE (CnF.1)
MODE (CnF.1)MODE (CnF.1)
Lower display:L.r.Oñ
Range:ALG = The local setpoint will be
aligned to last remote setpoint value when transfer from remote
to local setpoint action is performed through external contact
n.ALG = The local setpoint will be
not be changed when transfer from remote to local setpoint
action is performed (The programmed ramp (Grd1/Grd2) may be
activated)
Available: When auxiliary input is used as remote set point
and (“A.I.Añ = norñ)
Note :If transfer from remote to local setpoint is due to auxiliary
input status(“A.I.Añ = Cnd.A) the local setpoint
beenbeen
been aligned.
beenbeen
27
will neverwill never
will never
will neverwill never
CONFIGURATION GROUP 2 [C.Exx]CONFIGURATION GROUP 2 [C.Exx]
CONFIGURATION GROUP 2 [C.Exx]
CONFIGURATION GROUP 2 [C.Exx]CONFIGURATION GROUP 2 [C.Exx]
OUTPUTS CONFIGURATION
C.E01 - OUT 1 functionC.E01 - OUT 1 function
C.E01 - OUT 1 function
C.E01 - OUT 1 functionC.E01 - OUT 1 function
Upper display: CnF.2
Lower display: O1.Fn
Range: nonE = Output not used
ñAin = Time proportional main control output
SECn = Time proportional secondary control output
ALr.1 = Alarm 1 output
C.E02 - OUT 2 functionC.E02 - OUT 2 function
C.E02 - OUT 2 function
C.E02 - OUT 2 functionC.E02 - OUT 2 function
Upper display: CnF.2
Lower display: O2.Fn
Range: nonE = Output not used
ñAin = Time proportional main control output
SECn = Time proportional secondary control output
ALr.2 = Alarm 2 output
C.E03 - OUT 3 functionC.E03 - OUT 3 function
C.E03 - OUT 3 function
C.E03 - OUT 3 functionC.E03 - OUT 3 function
Upper display: CnF.2
Lower display: O3.Fn
Range: nonE = Output not used
ñAin = Time proportional main control output
SECn = Time proportional secondary control output
ALr.3 = Alarm 3 output
C.E04 - OUT 6 functionC.E04 - OUT 6 function
C.E04 - OUT 6 function
C.E04 - OUT 6 functionC.E04 - OUT 6 function
Upper display: CnF.2
Lower display: O6.Fn
Range: nonE = Output not used
ñAin = Main control output (linear)
SECn = Secondary control output (linear)
PV.rt = Process variable retransmission
SP.rt = Operative set point retransmission
C.E05 - OUT 6 rangeC.E05 - OUT 6 range
C.E05 - OUT 6 range
C.E05 - OUT 6 rangeC.E05 - OUT 6 range
This parameter is available only when OUT 6 ("O6.Fn" [C.E04]) is
different from "nonE".
Upper display: CnF.2
Lower display: O6.rn
Range: 0-20 = 0-20 mA
4-20 = 4-20 mA
28
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.