Super Systems 7SL User Manual

Instruction Manual PN 31302
7sl-0-00.p65 9/17/02, 2:35 PM1
MODEL 7SL HIGH/LOW LIMITROL
1
®
1/16 DIN, FOUR DIGIT HIGH/LOW LIMITROL
7 S L - 9 1 - 0 0 0 - 0 - 0 0
Field No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Fields 1 through 4. BASE
07SL - Limitrol (High/Low Limit – shipped as High Limit)
Field 5. INPUT
9 - TC types J, K, T, E, N, S, R, B, L, U, G, D, C and
Platinel II; and Pt 100 RTD; 0 to 20 mAdc and 4 to 20 mAdc; 0 to 60 mVdc and 12 to 60 mVdc; 0 to 5 Vdc or 1 to 5 Vdc;
0 to 10 Vdc or 2 to 10 Vdc Note: All inputs are factory calibrated and selectable by jumper. Factory set at Type J.
Field 6. OUTPUT (High/Low Limit)
1 - Relay (Form C)
Fields 7, 8. ALARMS, OPTIONS
00 - None 10 - One alarm – Relay Form A 11 - One alarm – Relay Form A, plus
RS485 and one logic input Note: when code 11 is specified, instrument length is 122 mm.
Field 9. POWER SUPPLY
3 - 100 to 240 Vac 5 - 24 Vac/Vdc
Fields 10 through 15. RESERVED
7sl-0-00.p65 9/17/02, 2:35 PM2
2
Unpack the
Instrument
Wiring
Configuration
Operating
Parameters
SV
Congratulations
Congratulations on your purchase of one of the easiest to configure high/low Limitrols on the market. After a 3 step configuration procedure, your process will be up and running.
Guide to simple set-up
To set-up the Limitrol only 3 steps are required:
1. Wire the instrument (page 7).
2. Configure the instrument (page 14).
3. Check the operating parameters (page 22).
3
7sl-0-00.p65 9/17/02, 2:35 PM3
Table of Contents
1/16 DIN, FOUR DIGIT HIGH/LOW LIMITROL..........2
Guide to simple set-up ......................................3
MOUNTING REQUIREMENTS..................................5
DIMENSIONS AND PANEL CUTOUT .......................6
WIRING GUIDELINES ...............................................7
PRELIMINARY HARDWARE SETTINGS ...............12
CONFIGURATION KEY FUNCTIONS .....................14
CONFIGURATION PROCEDURE ...........................14
OPERATING MODE ................................................20
Normal Display Mode .....................................20
Indicators ........................................................ 20
Key Functions in Normal Display Mode ..........21
Parameter Display Mode................................. 21
Key Functions in Parameter Display Mode .....21
OPERATING PARAMETERS................................... 22
ERROR MESSAGES ...............................................23
DEFAULT PARAMETERS ........................................24
SPECIFICATIONS ................................................... 26
CALIBRATION PROCEDURE.................................. 30
General Guidelines ......................................... 30
Calibration Parameters ................................... 30
Procedure .......................................................31
Entering Calibration Values .............................31
MAINTENANCE .......................................................35
CAUTION:
USE WIRE SUITABLE
FOR 75°C MINIMUM.
NOTES
• For supply connections use No 16 AWG or larger wires rated for at least 75 °C.
• Use copper conductors only.
• Class 2 wiring must be a minimum of 1/4 inch from any Class 1 conductors.
7sl-0-00.p65 9/17/02, 2:35 PM4
4
MOUNTING REQUIREMENTS
Select a mounting location with the following characteristics:
1) Minimal vibration.
2) An ambient temperature range between 0 and 50°C (32 and 122 °F).
3) Easy access to the rear of the instrument.
4) No corrosive gases (sulfuric gas, ammonia, etc.).
5) No water or other fluid (i.e. condensation).
6) Relative humidity of 20% to 80% non condensing.
The instrument can be mounted on a panel up to 15 mm (0.591 in) thick with a square cutout of 45 x 45 mm (1.772 x 1.772 in). For outline refer to Dimensions and Panel Cutout.
Bracket
Gasket
Panel surface texture must be better than 6.3 µmm. The instrument is shipped with a rubber panel gasket (50
to 60 Sh). To assure the IP65 and NEMA 4 protection, insert the panel gasket between the instrument and the panel as shown below.
Install the instrument as follows:
1) Insert the instrument in the gasket.
2) Insert the instrument in the panel cutout.
3) Pushing the instrument against the panel, insert the mounting bracket.
4) Torque the mounting bracket screws between 0.3 and 0.4 Nm (2.66 and 3.54 lbf-in).
5) To insure NEMA 4X/IP65 protection, make sure the instrument does not move within the cutout .
Panel
7sl-1-00.p65 9/17/02, 2:39 PM5
5
DIMENSIONS AND PANEL CUTOUT
48 mm
(1.890 in)
48 mm
(1.890 in)
75 mm
(2.953 in)
10 mm
(0.394 in)
60 mm
(2.362 in)
45 mm, -0, +0.6 mm
(1.772 in, -0, +0.024 in)
45 mm, -0, +0.6 mm
(1.772 in, -0, +0.024 in)
7sl-1-00.p65 9/17/02, 2:39 PM6
122 mm (4.803 in) w/ RS-485
105 mm (4.134 in) w/o RS-485
6
WIRING GUIDELINES
Terminal Layout
+
C
6
OUT 2
NO
7
_
8
RTD
9
_
TC
+
10
11 12 13 14
DIG 1
LINEAR
15
A/A'
B/B' RS485
C
NO
OUT 1
C
NC
PWR LINE
100/240Vac
1 2
3 4
24 Vac/dc
5
A) Measuring Inputs
NOTE: Any external components (like Zener diodes, etc.) connected between sensor and input terminals may cause errors in measurement due to excessive and/or not balanced line resistance or possible leakage cur­rents.
TC Input
+
10
_
9
Shield
+
10
_
9
Shield
NOTE: When a relay output is used to drive an inductive load, connect an external snubber network (RC) across the terminals:
RC
in accordance with the following table:
daoL
C
tnerruC
(µ )FR()
Am04<740.00012/1caV062
Am051<1.0222 062caV
pmA5.0<33.0742 062caV
7sl-1-00.p65 9/17/02, 2:39 PM7
P
)W(
dnarotsiseR
NOTE:
1) Do not run input wires with power cables.
2) For TC wiring use proper compensating cable, preferably shielded (see Thermocouple Compen­sating Cable Color Codes).
3) Shielded cable should be grounded at one end only.
egatloVroticapaC
7
Linear Input
+
10
+
mA,
mV
_
9
_
Shield
+
10
_
9
+
_
G
NOTE:
1) Do not run input wires with power cables.
2) High line resistance can cause measurement errors.
3) When shielded cable is used, ground it at one end only to avoid ground loop currents.
4) The input impedance is equal to: Less than 5 for 20 mAdc input Greater than 1 M for 60 mVdc input Greater than 400 K for 5 Vdc and 10 Vdc input
RTD Input
RTD
or
V
mA mV
or
V
8
NOTE:
1) Do not run RTD wires with power cables.
2) Ground shielded cable at one end only.
109
8
3) Use the correct size copper wires.
4) The resistance of the 3 wires must be the same.
B) Logic Input
This input is used as a remote reset. Safety note:
- Do not run logic input wiring with AC power cables.
- Use an external contact with a contact rating greater than 0.5 mA, 5 Vdc.
- The instrument needs 100 ms to recognize a contact status variation.
Logic input
14
RTD
109
7sl-1-00.p65 9/17/02, 2:39 PM8
15
8
Thermocouple Compensating Cable Color Codes.
Thermocouple
Material
T Copper Constantan
J/L Iron Constantan
K Nickel Chromium
Nickel Aluminum
R Platinum/Platinum
13% Rhodium
S Platinum/Platinum
10% Rhodium
E Chromel
Constantan
B Platinum 30% Rh
Platinum 6% Rh
N Nicrosil / Nisil
British
BS 1843
+ White
- Blue Blue
+ Yellow
- Blue Black
+ Brown
- Blue Red
+ White
- Blue Green
+ White
- Blue Green
+ Brown
- Blue Brown – –
American
ANSI MC 96.1
+ Blue
- Red Blue
+ White
- Red Black
+ Yellow
- Red Yellow
+ Black
- Red Green
+ Black
- Red Green
+ Violet
- Red Violet
+ Grey
- Red Grey –
German
DIN 43710
+ Red
- Brown Brown
+ Red
- Blue Blue
+ Red
- Green Green
+ Red
- White White
+ Red
- White White – –
– – – –
French
NFE 18-001
+ Yellow
- Blue Blue
+ Yellow
- Black Black
+ Yellow
- Purple Yellow
+ White
- Green Green
+ White
- Green Green – –
– – – –
7sl-1-00.p65 9/17/02, 2:39 PM9
9
C.1) Relay Outputs
OUT 1
Class 1
OUT 2 Class 1 (Alarm)
NO - OUT 1
1
C - OUT 1
2
NC - OUT 1
3
C
6
NO
7
The cable used for relay output wiring must be as far away as possible from input or communication cables. Relay output: Protected by varistor. OUT 1: Form C contact rating of 3 Amps/250 Vac
resistive load.
OUT 2: Form A Contact rating of 2 Amps/250 Vac
resistive load.
Number of operations: 2 x 10
5
at the specified rating.
NOTES:
1) To avoid shock and possible instrument damage, connect power last.
2) For power connections use 16 AWG or larger wires rated for 75 °C.
3) Use copper conductors only.
4) Do not run input wires with power cables.
C.2) Inductive Loads
High voltage transients may occur when switching induc­tive loads. Through internal contacts these transients may introduce disturbances which can affect the perfor­mance of the instrument. The same problem may occur when a switch is used in series with the internal contacts as shown below.
C
R
POWER
LINE
LOAD
For all the outputs, the internal protection (varistor) assures protection up to 0.5 Amp on inductive loads.
It is good electrical practice to install an additional RC net­work across and as close to the external contacts as possible.
The value of capacitor (C) and resistor (R) are shown in the following table.
daoL
C
tnerruC
(µ )FR()
Am04<740.00012/1caV062
Am051<1.0222 062caV
pmA5.0<33.0742 062caV
P
)W(
dnarotsiseR
egatloVroticapaC
7sl-1-00.p65 9/17/02, 2:39 PM10
10
D) Serial Interface
For units built with optional RS-485 communication
I N S T R U
A/A'
11
B/B'
12
M E
13
N
COMMON
A'/A
B'/B
M A S T E R
T
Maximum cable length: 1.5 km (9/10 mile) at 9600 baud. NOTE: According to EIA specification for RS-485: a) The “A” terminal of the generator shall be negative
with respect to the “B” terminal for a binary 1 (MARK or OFF) state.
b) The “A” terminal of the generator shall be positive
with respect to the “B” terminal for a binary 0 (SPACE or ON) state.
The RS-485 interface can connect up to 31 instruments with the remote master unit (see below).
I N S T R U
A/A'
11
B/B'
12
M E
13
N
A'/A
B'/B
COMMON
M A S T E R
T
I
11
N S T
12
R U M
13
E N T
7sl-1-00.p65 9/17/02, 2:39 PM11
11
Loading...
+ 25 hidden pages