Super Systems 7SL User Manual

Instruction Manual PN 31302
7sl-0-00.p65 9/17/02, 2:35 PM1
MODEL 7SL HIGH/LOW LIMITROL
1
®
1/16 DIN, FOUR DIGIT HIGH/LOW LIMITROL
7 S L - 9 1 - 0 0 0 - 0 - 0 0
Field No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Fields 1 through 4. BASE
07SL - Limitrol (High/Low Limit – shipped as High Limit)
Field 5. INPUT
9 - TC types J, K, T, E, N, S, R, B, L, U, G, D, C and
Platinel II; and Pt 100 RTD; 0 to 20 mAdc and 4 to 20 mAdc; 0 to 60 mVdc and 12 to 60 mVdc; 0 to 5 Vdc or 1 to 5 Vdc;
0 to 10 Vdc or 2 to 10 Vdc Note: All inputs are factory calibrated and selectable by jumper. Factory set at Type J.
Field 6. OUTPUT (High/Low Limit)
1 - Relay (Form C)
Fields 7, 8. ALARMS, OPTIONS
00 - None 10 - One alarm – Relay Form A 11 - One alarm – Relay Form A, plus
RS485 and one logic input Note: when code 11 is specified, instrument length is 122 mm.
Field 9. POWER SUPPLY
3 - 100 to 240 Vac 5 - 24 Vac/Vdc
Fields 10 through 15. RESERVED
7sl-0-00.p65 9/17/02, 2:35 PM2
2
Unpack the
Instrument
Wiring
Configuration
Operating
Parameters
SV
Congratulations
Congratulations on your purchase of one of the easiest to configure high/low Limitrols on the market. After a 3 step configuration procedure, your process will be up and running.
Guide to simple set-up
To set-up the Limitrol only 3 steps are required:
1. Wire the instrument (page 7).
2. Configure the instrument (page 14).
3. Check the operating parameters (page 22).
3
7sl-0-00.p65 9/17/02, 2:35 PM3
Table of Contents
1/16 DIN, FOUR DIGIT HIGH/LOW LIMITROL..........2
Guide to simple set-up ......................................3
MOUNTING REQUIREMENTS..................................5
DIMENSIONS AND PANEL CUTOUT .......................6
WIRING GUIDELINES ...............................................7
PRELIMINARY HARDWARE SETTINGS ...............12
CONFIGURATION KEY FUNCTIONS .....................14
CONFIGURATION PROCEDURE ...........................14
OPERATING MODE ................................................20
Normal Display Mode .....................................20
Indicators ........................................................ 20
Key Functions in Normal Display Mode ..........21
Parameter Display Mode................................. 21
Key Functions in Parameter Display Mode .....21
OPERATING PARAMETERS................................... 22
ERROR MESSAGES ...............................................23
DEFAULT PARAMETERS ........................................24
SPECIFICATIONS ................................................... 26
CALIBRATION PROCEDURE.................................. 30
General Guidelines ......................................... 30
Calibration Parameters ................................... 30
Procedure .......................................................31
Entering Calibration Values .............................31
MAINTENANCE .......................................................35
CAUTION:
USE WIRE SUITABLE
FOR 75°C MINIMUM.
NOTES
• For supply connections use No 16 AWG or larger wires rated for at least 75 °C.
• Use copper conductors only.
• Class 2 wiring must be a minimum of 1/4 inch from any Class 1 conductors.
7sl-0-00.p65 9/17/02, 2:35 PM4
4
MOUNTING REQUIREMENTS
Select a mounting location with the following characteristics:
1) Minimal vibration.
2) An ambient temperature range between 0 and 50°C (32 and 122 °F).
3) Easy access to the rear of the instrument.
4) No corrosive gases (sulfuric gas, ammonia, etc.).
5) No water or other fluid (i.e. condensation).
6) Relative humidity of 20% to 80% non condensing.
The instrument can be mounted on a panel up to 15 mm (0.591 in) thick with a square cutout of 45 x 45 mm (1.772 x 1.772 in). For outline refer to Dimensions and Panel Cutout.
Bracket
Gasket
Panel surface texture must be better than 6.3 µmm. The instrument is shipped with a rubber panel gasket (50
to 60 Sh). To assure the IP65 and NEMA 4 protection, insert the panel gasket between the instrument and the panel as shown below.
Install the instrument as follows:
1) Insert the instrument in the gasket.
2) Insert the instrument in the panel cutout.
3) Pushing the instrument against the panel, insert the mounting bracket.
4) Torque the mounting bracket screws between 0.3 and 0.4 Nm (2.66 and 3.54 lbf-in).
5) To insure NEMA 4X/IP65 protection, make sure the instrument does not move within the cutout .
Panel
7sl-1-00.p65 9/17/02, 2:39 PM5
5
DIMENSIONS AND PANEL CUTOUT
48 mm
(1.890 in)
48 mm
(1.890 in)
75 mm
(2.953 in)
10 mm
(0.394 in)
60 mm
(2.362 in)
45 mm, -0, +0.6 mm
(1.772 in, -0, +0.024 in)
45 mm, -0, +0.6 mm
(1.772 in, -0, +0.024 in)
7sl-1-00.p65 9/17/02, 2:39 PM6
122 mm (4.803 in) w/ RS-485
105 mm (4.134 in) w/o RS-485
6
WIRING GUIDELINES
Terminal Layout
+
C
6
OUT 2
NO
7
_
8
RTD
9
_
TC
+
10
11 12 13 14
DIG 1
LINEAR
15
A/A'
B/B' RS485
C
NO
OUT 1
C
NC
PWR LINE
100/240Vac
1 2
3 4
24 Vac/dc
5
A) Measuring Inputs
NOTE: Any external components (like Zener diodes, etc.) connected between sensor and input terminals may cause errors in measurement due to excessive and/or not balanced line resistance or possible leakage cur­rents.
TC Input
+
10
_
9
Shield
+
10
_
9
Shield
NOTE: When a relay output is used to drive an inductive load, connect an external snubber network (RC) across the terminals:
RC
in accordance with the following table:
daoL
C
tnerruC
(µ )FR()
Am04<740.00012/1caV062
Am051<1.0222 062caV
pmA5.0<33.0742 062caV
7sl-1-00.p65 9/17/02, 2:39 PM7
P
)W(
dnarotsiseR
NOTE:
1) Do not run input wires with power cables.
2) For TC wiring use proper compensating cable, preferably shielded (see Thermocouple Compen­sating Cable Color Codes).
3) Shielded cable should be grounded at one end only.
egatloVroticapaC
7
Linear Input
+
10
+
mA,
mV
_
9
_
Shield
+
10
_
9
+
_
G
NOTE:
1) Do not run input wires with power cables.
2) High line resistance can cause measurement errors.
3) When shielded cable is used, ground it at one end only to avoid ground loop currents.
4) The input impedance is equal to: Less than 5 for 20 mAdc input Greater than 1 M for 60 mVdc input Greater than 400 K for 5 Vdc and 10 Vdc input
RTD Input
RTD
or
V
mA mV
or
V
8
NOTE:
1) Do not run RTD wires with power cables.
2) Ground shielded cable at one end only.
109
8
3) Use the correct size copper wires.
4) The resistance of the 3 wires must be the same.
B) Logic Input
This input is used as a remote reset. Safety note:
- Do not run logic input wiring with AC power cables.
- Use an external contact with a contact rating greater than 0.5 mA, 5 Vdc.
- The instrument needs 100 ms to recognize a contact status variation.
Logic input
14
RTD
109
7sl-1-00.p65 9/17/02, 2:39 PM8
15
8
Thermocouple Compensating Cable Color Codes.
Thermocouple
Material
T Copper Constantan
J/L Iron Constantan
K Nickel Chromium
Nickel Aluminum
R Platinum/Platinum
13% Rhodium
S Platinum/Platinum
10% Rhodium
E Chromel
Constantan
B Platinum 30% Rh
Platinum 6% Rh
N Nicrosil / Nisil
British
BS 1843
+ White
- Blue Blue
+ Yellow
- Blue Black
+ Brown
- Blue Red
+ White
- Blue Green
+ White
- Blue Green
+ Brown
- Blue Brown – –
American
ANSI MC 96.1
+ Blue
- Red Blue
+ White
- Red Black
+ Yellow
- Red Yellow
+ Black
- Red Green
+ Black
- Red Green
+ Violet
- Red Violet
+ Grey
- Red Grey –
German
DIN 43710
+ Red
- Brown Brown
+ Red
- Blue Blue
+ Red
- Green Green
+ Red
- White White
+ Red
- White White – –
– – – –
French
NFE 18-001
+ Yellow
- Blue Blue
+ Yellow
- Black Black
+ Yellow
- Purple Yellow
+ White
- Green Green
+ White
- Green Green – –
– – – –
7sl-1-00.p65 9/17/02, 2:39 PM9
9
C.1) Relay Outputs
OUT 1
Class 1
OUT 2 Class 1 (Alarm)
NO - OUT 1
1
C - OUT 1
2
NC - OUT 1
3
C
6
NO
7
The cable used for relay output wiring must be as far away as possible from input or communication cables. Relay output: Protected by varistor. OUT 1: Form C contact rating of 3 Amps/250 Vac
resistive load.
OUT 2: Form A Contact rating of 2 Amps/250 Vac
resistive load.
Number of operations: 2 x 10
5
at the specified rating.
NOTES:
1) To avoid shock and possible instrument damage, connect power last.
2) For power connections use 16 AWG or larger wires rated for 75 °C.
3) Use copper conductors only.
4) Do not run input wires with power cables.
C.2) Inductive Loads
High voltage transients may occur when switching induc­tive loads. Through internal contacts these transients may introduce disturbances which can affect the perfor­mance of the instrument. The same problem may occur when a switch is used in series with the internal contacts as shown below.
C
R
POWER
LINE
LOAD
For all the outputs, the internal protection (varistor) assures protection up to 0.5 Amp on inductive loads.
It is good electrical practice to install an additional RC net­work across and as close to the external contacts as possible.
The value of capacitor (C) and resistor (R) are shown in the following table.
daoL
C
tnerruC
(µ )FR()
Am04<740.00012/1caV062
Am051<1.0222 062caV
pmA5.0<33.0742 062caV
P
)W(
dnarotsiseR
egatloVroticapaC
7sl-1-00.p65 9/17/02, 2:39 PM10
10
D) Serial Interface
For units built with optional RS-485 communication
I N S T R U
A/A'
11
B/B'
12
M E
13
N
COMMON
A'/A
B'/B
M A S T E R
T
Maximum cable length: 1.5 km (9/10 mile) at 9600 baud. NOTE: According to EIA specification for RS-485: a) The “A” terminal of the generator shall be negative
with respect to the “B” terminal for a binary 1 (MARK or OFF) state.
b) The “A” terminal of the generator shall be positive
with respect to the “B” terminal for a binary 0 (SPACE or ON) state.
The RS-485 interface can connect up to 31 instruments with the remote master unit (see below).
I N S T R U
A/A'
11
B/B'
12
M E
13
N
A'/A
B'/B
COMMON
M A S T E R
T
I
11
N S T
12
R U M
13
E N T
7sl-1-00.p65 9/17/02, 2:39 PM11
11
E) Power Line and grounding
N, L2
4
5
R (S,T), L1
NOTE:
1) Before connecting the power line, check that the voltage is correct (see Model Number).
2) For supply connections use 16 AWG or larger wires rated for at least 75 °C.
3) Use copper conductors only.
4) Do not run input wires with power cables.
5) Polarity does not matter for 24 Vdc wiring.
6) The power supply input is Please provide it externally.
When the fuse is damaged the instrument should be returned to your supplier to check the power supply.
7) Safety requirements for permanently connected equipment:
- Include a switch or circuit-breaker in the
installation.
- Place the switch in close proximity to the
equipment and within easy reach of the operator.
- Mark the switch as the disconnecting device for
the equipment. NOTE: A single switch or circuit-breaker can drive more than one instrument.
8) When the NEUTRAL line is present, connect it to terminal 4.
9) To avoid shock and possible instrument damage, connect power last.
POWER SUPPLY
100 to 240 Vac
24 Vac/Vdc
N, L2
R (S,T), L1
not
fuse protected.
ylppuSrewoPepyTtnerruCegatloV
cdV/caV42TAm005V052
caV042/001TAm521V052
PRELIMINARY HARDWARE SETTINGS
1) Remove the instrument from its case.
2) Set J106 according to the following table:
epyTtupnI
2-14-36-58-7
,DTR,C/T
Vm06-21,Vm06-0
esolc
)noitarugifnocdradnats(
V5-1,V5-0
V01-2,V01-0
Am02-4,Am02-0
nepo esolc nepo nepo nepo nepo esolc nepo nepo nepo nepo esolc
K081=291RCTTCtupnI
nepO=101HCdesolC=021HS
12
601J
nepo
nepo nepo
7sl-1-00.p65 9/17/02, 2:39 PM12
Limitrol with RS-485
V101
Limitrol without RS-485
J106
8642
7531
7sl-1-00.p65 9/17/02, 2:39 PM13
13
CONFIGURATION KEY FUNCTIONS
RESET In Configuration Mode, used only to scroll
back parameters
Used in Configuration Mode to decrease the parameter value.
Used in Configuration Mode to increase the parameter value.
FUNC Monitors/Modifies parameters.
+ Loads the default parameters.
+ FUNC or + FUNC
Increases/decreases values at a higher rate when modifying parameters.
+ RESET or + RESET
Jumps to the Maximum or Minimum parameter value when modifying parameters.
CONFIGURATION PROCEDURE
1) Remove the instrument from its case.
2) Open switch V101 (See illustrations under “Prelimi­nary Hardware Settings.”)
3) Re-insert the instrument in its case.
4) Switch on power to the instrument. The upper display will show COnF.
5) Press the “ ” key and the lower display will show the firmware version.
Press the "FUNC" key to start the configuration proce­dure with the first parameter (L1). Press the "RESET" key to start the configuration procedure with the last parameter (d1).
The following is a complete list of parameters. The lower display will show the parameter code (L1 to d1) and the upper display will show the selection code or numerical value. No timeout is applied in the configuration mode.
L1 = Serial Interface Protocol
(Skipped if option is not available.) OFF = No serial interface nbUS= Modbus jbUS = Jbus
L2 = Serial Link Device Address
(Skipped if option is not available or L1 = OFF) From 1 to 255 NOTE: The device can connect up to 31 instruments.
L3 = Baud Rate for Serial Link
(Skipped if option is not available or L1 = OFF) Set value from 600 to 19200 baud. (19200 baud is shown on display as 1920)
L4 = Byte Format for Serial Link
(Skipped if option is not available or L1 = OFF) 8E = 8 bits + even parity 8O = 8 bits + odd parity 8 = 8 bits without parity
7sl-2-00.p65 9/17/02, 2:40 PM14
14
r1 = Input Type and Range Value
0 = TC J From -100 to 1000 °C 1 = TC K From -100 to 1370 °C 2 = TC T From -200 to 400 °C 3 = TC E From -100 to 800 °C 4 = TC N From -100 to 1400 °C 5 = TC S From -50 to 1760 °C 6 = TC R From -50 to 1760 °C 7 = TC B From 0 to 1820 °C 8 = TC L From -100 to 900 °C 9 = TC U From -200 to 600 °C 10= TC G From 0 to 2300 °C 11= TC D From 0 to 2300 °C 12= TC C From 0 to 2300 °C 13= TC Plat. II From -100 to 1400 °C 14= RTD Pt 100 From -200 to 850 °C 15= Linear From 0 to 60 mV 16= Linear From 12 to 60 mV 17= Linear From 0 to 20 mA 18= Linear From 4 to 20 mA 19= Linear From 0 to 5 V 20= Linear From 1 to 5 V 21= Linear From 0 to 10 V 22= Linear From 2 to 10 V 23= TC J From -150 to 1830 °F 24= TC K From -150 to 2500 °F 25= TC T From -330 to 750 °F 26= TC E From -150 to 1470 °F 27= TC N From -150 to 2550 °F 28= TC S From -60 to 3200 °F 29= TC R From -60 to 3200 °F 30= TC B From 32 to 3300 °F 31= TC L From -150 to 1650 °F 32= TC U From -330 to 1110 °F 33= TC G From 0 to 4170 °F 34= TC D From 0 to 4170 °F 35= TC C From 0 to 4170 °F 36= TC Plat. II From -150 to 2550 °F 37= RTD Pt100 From -330 to 1560 °F
r2 = Decimal Point Position
(Available only for linear range r1 = 15 to 22)
_ _ _ _ . = No decimal _ _ _ . _ = One decimal place _ _ . _ _ = Two decimal places _ . _ _ _ = Three decimal places
r3 = Low Scale Range Value
(Available only for linear range r1 = 15 to 22) Range: From -1999 to 9999
r4 = High Scale Range Value
(Available only for linear range r1 = 15 to 22) Range: From -1999 to 9999
r5 = Offset Adjustment
Range: From -500 to 500 Offset value algebraically added to the measured value.
r6 = Time Constant for Filter on Displayed Value
Range: From 0 (filter OFF) to 8 seconds. (First order filter with selected time constant.)
r7 = Alarm Action on Fault
uP = The alarm assumes an upscale reading. doun = The alarm assumes a downscale reading.
C1 = Type of Limit Action
Hi. = High limit (for heating process) Lo. = Low limit (for cooling process) Hi.Lo = Band limit (for special process)
C2 = Acknowledgment Mode
O = The acknowledgment action is ignored if performed
when setpoint is exceeded.
I = The acknowledgment action is recognized also
when the setpoint is exceeded. (In this case, the control output is instantaneously restored when process variable is within setpoint).
7sl-2-00.p65 9/17/02, 2:40 PM15
15
Control Output Function
The relay output operates in fail-safe mode (relay de­energized during reset condition) and latching mode.
The control output turns OFF when the setpoint is ex­ceeded when C1 = Hi, or C1 = Lo. (When C1 = Hi.Lo control output turns off when the process is greater than “Su” or less than “S1”).
The control output remains OFF until the process is within setpoint and the acknowledge action has been performed (the sequence of action may or may not be important depending on parameter C2).
The upper display flashes when the setpoint is exceeded and returns to a steady display when the process is within setpoint.
When the control output is OFF the RESET LED is ON (if C2 = 0) or flashes (if C2 = 1).
When C2 = 1 the RESET LED is steady ON when control output is OFF and acknowledged.
The reset condition can be stored in permanent memory (see C4).
Acknowledgment can be performed by pressing the RESET key, by momentarily closing the external dry contact or by a command from the serial link.
The length of the reset condition and max/min values detected are stored in memory and available for viewing until the next reset condition occurs. The information is lost at power down.
During a reset condition the values are continuously updated and can be monitored.
Example of limit function when C1 = Hi and C2 = O
HS
(Setpoint
threshold
hysteresis)
ON
Relay OUT 1
OFF
ON
"Reset" LED
OFF
Flash
Upper display
Steady
FLASH
ABC
FLASH
Su (Setpoint threshold)
A, B, C = Acknowledgment action by RESET button, digital input or serial link. NOTE: Acknowledgment B has no effect.
Example of limit function when C1 = Hi and C2 = 1
(Hysteresis)
Relay OUT 1
"Reset" LED
Upper display
ON
OFF
ON
OFF
Flash
Steady
FLASH FLASH
FLASH
AB
FLASH
Su (Setpoint threshold)
A, B = Acknowledgment action by RESET button, digital input or serial link.
7sl-2-00.p65 9/17/02, 2:40 PM16
16
C3 = Reset at Power-up
Auto = Automatic reset
= Manual reset
nAn
C4 = Reset Memory
0 = The reset condition will be saved (at next
power up it will be reactivated)
1 = The reset condition will be lost in case of
power down
C5 = Time Constant for Filter on Measured Value for Reset Action
Range: From 0 (filter OFF) to 8 seconds Note: First order filter with selected time Constant.
P1 = Alarm Function
(Skipped if option not available) nonE = Not provided AL.P = Process alarm AL.b = Band alarm AL.d = Deviation alarm When C1 = Hi.Lo, “AL.b” and “AL.d” are not available.
P2 = Alarm configuration
(Skipped if option is not available or P1 = none) H.A. = High alarm with automatic reset L.A. = Low alarm with automatic reset H.A.Ac = High alarm with automatic reset and
acknowledge
L.A.Ac =Low alarm with automatic reset and
acknowledge H.L. = High alarm with manual reset L.L. = Low alarm with manual reset
NOTE: For band alarm, H.A./H.A.Ac/H.L. signifies outside band alarm, while L.A./ L.A.Ac/L.L. signifies inside band alarm.
For every alarm configuration: LED flashes, relay placed in alarm status
.A.H=2P
llits
.A.L=2P
cA.A.H=2P
llits
cA.A.L=2P
llits
.L.H=2P
.L.L=2P
.kcA
mralA
oN
stsixe
.kcA
mralA
sraelc
mralA
stsixe
mralA
sraelc
mralA
stsixe
mralA
sraelc
ffoDEL
.kcA
oN
.kcA
ffoDEL
.kcA
oN
.kcA .kcA
oN
.kcA
gnihsalfDEL
gnihsalfDEL
gnihsalfDEL
FFODEL
gnihsalfDEL
Ack. = Alarm Acknowledgment (For relay status see configuration parameter P3.)
Example for P2 = H.A.
(Hysteresis)
Alarm status*
Relay
No alarm status
ON
ALM LED
OFF
ACK
FLASHFLASH
ACK = Alarm acknowledgment from "AK" parameter
or serial link.
* Alarm Status:Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
17
ydaetstilDEL
sutatsmralanitonyaleR ydaetstilDEL
sutatsmralanitonyaleR ydaetstilDEL
sutatsmralaniyaleR
sutatsmralaniyaleR
sutatsmralaniyaleR
sutatsmralaniyaleR
sutatsmralaniyaleR
sutatsmralaniyaleR
(Alarm threshold)
sutatsmralanitonyaleR
sutatsmralanitonyaleR
7sl-2-00.p65 9/17/02, 2:40 PM17
Example for P2 = H.A.A.c
(Hysteresis)
Alarm status*
Relay
No alarm status
ALM LED
OFF
ON
FLASH FLASH
ACK
(Alarm threshold)
ACK = Alarm acknowledgment from "AK" parameter
or serial link.
* Alarm Status:Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
Example for P2 = H.L.
(Hysteresis)
Alarm status*
Relay
No alarm status
ALM LED
OFF
ON
FLASHFLASH
ACK ACK
(Alarm threshold)
ACK = Alarm acknowledgment from "AK" parameter
or serial link.
* Alarm Status:Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
P3 = Alarm Action
(Skipped if option not is available or P1 = none) dir = Direct action
(Relay energized in alarm condition)
rEV = Reverse action
(Relay energized in non-alarm condition)
P4 = Alarm Standby Function
(Skipped if option is not available or P1= none) OFF = Standby function disabled On = Standby function enabled
If the alarm is programmed as band or deviation, this function masks the alarm condition at start up and at a “Su” setpoint change until the process variable reaches the alarm threshold, plus or minus hysteresis. This standby function masks a Process Alarm condition at start up until the process variable reaches the alarm threshold plus or minus hysteresis.
PF = Time Constant for Filter on Measured Value for Alarm Action
(Skipped if option is not available or P1 = none) Range: From 0 (filter OFF) to 8 seconds (First order filter with selected time constant.)
n 1 = Safety Lock
0 = UNLOCKed. The device is always UNLOCKed
and all parameters can be modified.
l = LOCKed. The device is always LOCKed and
no parameters can be modified
From 2 to 9999 = This number is a password, to be used
in run time (see “nn”), to LOCK/UNLOCK the device.
7sl-2-00.p65 9/17/02, 2:40 PM18
18
t1 = Timeout Selection
tn10 = 10 second timeout tn30 = 30 second timeout
d1 = Digital Input (contact closure)
(This is a read only parameter) Enb = Digital input enabled dlS = Digital input disabled (The digital input is used as a remote process restart.)
The configuration procedure is now complete. The display should show "COnF".
7sl-2-00.p65 9/17/02, 2:40 PM19
19
OPERATING MODE
1) Remove the instrument from its case.
2) Set switch V101 to the closed position.
3) Re-insert the instrument in its case.
4) Switch on the instrument.
Normal Display Mode
On powerup the device starts in the "Normal Display Mode."
By pressing the or key, it is possible to change the displayed information; therefore, one of the following display modes can be selected:
1) The upper display shows the measured value while
the lower display shows the "Pu" (Process variable)
2) The upper display shows the setpoint threshold while
the lower display shows "Su." If this display was active at power down, it will be active at powerup.
3) The upper display shows the setpoint1 threshold
while the lower display shows "S1." This information is available only if C1 = Hi.Lo If this display was active at power down, it will be active at powerup
4) The upper display shows the total time (hh.mm) of the
last reset condition while the lower displays shows “t.” (if no data is available, the upper display will show "- - - -" ). The information is lost at power down and at powerup the device will display the process variable.
5) The upper display shows the maximum value of
process variable detected during the last "reset" con­dition while the lower display shows "Ph." (if no data is available, the upper display will show “- - - -” ). This information is not available if C1 = Lo. The information is lost at power down and at powerup the device will display the process variable.
NOTE: In case the reset condition was generated by a fault condition in the measure variable, the upper display
will indicate
"m.Err"
6) The upper display shows the minimum value of the process variable detected during the last "reset" con­dition while the lower display shows "PL.” (if no data is available, the upper display will show “- - - -”) This information is not available if C1 = Hi. The information is lost at power down and at powerup the device will display the process variable .
NOTE: in case the “reset condition” was generated by a fault condition in the measure variable, the upper
display will indicate
"m.Err"
At powerup the display will show the process variable unless otherwise stated in one of the above display options
If, at power off, the device was in reset condition and configured to save it (C4 = 0), and/or it was programmed for manual reset at startup (C3 = 1), then at the next power up the lower display will be flashing.
Indicators
“RESET“ = Indicates control output status
as follows: a) With configuration parameter C2 = 0 LED ON when Output is OFF LED OFF when Output is ON b) With configuration parameter C2 = 1 LED flashes when Output is OFF LED ON when Output is OFF and acknowledged LED OFF when Output is ON
“ALM” = Indicates alarm status as follows:
Flashes when alarm is ON ON when alarm has been acknowledged OFF when alarm is OFF
7sl-3-00.p65 9/17/02, 2:41 PM20
20
Key Functions in Normal Display Mode
“FUNC” = By pressing it, the display changes from
“Normal Display Mode” to “Parameter Display Mode.”
= Pressing it for more than ten seconds
initiates the Lamp Test. During the Lamp Test the device functions normally while all display segments and LED's are lit with a 50% duty cycle. No timeout is applied to a lamp test.
Press the "FUNC" key again to end the Lamp Test.
" " or “ " = By pressing these keys it is possible to
change the displayed information. See “Normal Display Mode” on previous page.
“RESET” = Press and hold for 1 second to initiate
“reset.”
Parameter Display Mode
The "FUNC" key initiates the Parameter Display Mode when pressed for less than 10 seconds in the "Normal Display Mode."
The lower display shows the parameter identification code while the upper display shows the parameter value. The value of these parameters can be modified with the
and keys.
Press the "FUNC" key again to store the new value and advance to the next parameter.
If no keys are pressed within the timeout period (see t1), the display will automatically return to the "Normal Dis­play Mode" in the previous display and any modifications of the last displayed parameter will be lost.
All parameters (except
) can be modified only when the
device is UNLOCKed. The LOCK/UNLOCK status can be selected in configura-
tion using parameter n1 or during the operating mode with the parameter password.
To switch from LOCKED to UNLOCKED, enter the n1 parameter setting. To switch from UNLOCKED to LOCKED, enter any number other than the n1 parameter setting.
When the device is in remote (the serial link controls the device) no parameters can be modified.
Key Functions in Parameter Display Mode
FUNC = Press the “FUNC” key again and the
instrument stores the new setting (if changed) and goes to the next parameter.
or = Changes the setting of the selected
parameter.
RESET = Press and hold for more than 1 second to
initiate reset.
7sl-3-00.p65 9/17/02, 2:41 PM21
21
OPERATING PARAMETERS
Some of the following parameters may not appear, depending on the configuration.
Alarm Acknowledge
(Available only if P1 = AL.p, AL.b or AL.d) Lower display: Upper display: OFF/ON
Select ON and press the FUNC key in order to acknowledge the alarm. ON = Alarm Acknowledged OFF = Alarm Not Acknowledged
Software Key
(Skipped if n1 = 0 or 1) Lower display: nn Upper display: Numeric password to
LOCK/UNLOCK the device NOTE: The upper display will initially show the "LOCK" status. ON = the device is LOCKED. OFF = the device is UNLOCKED.
Setpoint Threshold
Lower display: Su Upper display: Setpoint threshold value Range: Span limits (From "S1" to high span
limit when C1 = Hi.Lo)
Setpoint1 Threshold
(Available when C1 = Hi.Lo) Lower display: S1 Upper display: Setpoint threshold value Range: From low span limit to "Su"
Setpoint Threshold Hysteresis
Lower display: HS Upper display: Hysteresis value Range: From 0.1% to 10.0% of span value
or 1 LSD
Alarm Threshold (optional)
(Available only if P1=AL.P, AL.b or AL.d.) Lower display: AL Upper display: Alarm threshold value Range:
Span limits (if process alarm P1 = AL.P) From 0 to 500 (if band alarm P1 = AL.b) From -500 to 500 (if deviation alarm P1 = AL.d)
Alarm Hysteresis (optional)
(Available only if P1 = AL.P, AL.b or AL.d) Lower display: HA Upper display: Hysteresis value Range: From 0.1% to 10.0% of span value
or 1 LSD
Serial LInk (optional)
The device can be connected to a host computer via serial link.
The host can put the device in LOCAL (parameters are controlled via keyboard) or in REMOTE (functions and parameters are controlled via serial link).
REMOTE is signified by the decimal point to the left of "RMT" which is on the right side of the numeric display.
Via serial link it is possible to read and/or to modify all the operative and configuration parameters. The following conditions must apply to implement this function:
1) Configure parameters L1 through L4 with the front keyboard
2) The device must be in the Operating mode.
Control output is OFF during the download procedure. At the end of configuration downloading procedure the
device preforms an automatic reset and returns to normal condition.
22
7sl-3-00.p65 9/17/02, 2:41 PM22
ERROR MESSAGES
Overrange, Underrange and Sensor Break Indications
This device detects process variable faults (OVERRANGE, UNDERRANGE OR SENSOR BREAK). When the process variable exceeds the span limits established by configuration parameter r1 an OVERRANGE condition will appear as:
An UNDERRANGE condition will appear as:
A sensor break is signaled as "OPEN”. On the mA/V input, a sensor break can be detected only when the range selected has a zero elevation (4/20 mA, 12/60 mV, 1/5 V or 2/10 V.)
On the RTD input "shrt" is signalled when input resis­tance is less than 15 (short circuit sensor detection).
This device detects reference junction errors or errors on the internal autozero measurement. When a fault is detected the output goes OFF and the alarm assumes an upscale/downscale reading in accordance with r7.
Error Messages
On powerup, the instrument performs a self-diagnostic test. When an error is detected, the lower display shows an "Er" indication while the upper display shows the code of the detected error.
Error List
100 Error in EEPROM writing 150 Short circuit on CPU's outputs 200 Error on "protect register" in EEPROM XXX Configuration parameter error. 301 Error on calibration of selected input. 307 rj input calibration error. 400 Error on operative parameter. 500 Error on autozero measurement. 502 Error on reference junction measurement. 510 Error during calibration procedure.
Dealing with Error Messages
1) When a configuration parameter error is detected, repeat the configuration procedure of that specific parameter.
2) If an error 400 is detected, simultaneously press the
and keys to load the default parameters
and then repeat the control parameter setup.
3) For all other errors, contact your Service Representative.
7sl-4-00.p65 9/17/02, 2:42 PM23
23
DEFAULT PARAMETERS
Loading Default Operating Parameters
The control parameters can be loaded with predeter­mined default values. These are the settings loaded into the instrument prior to shipment from the factory. To load the default values proceed as follows:
a) Press and hold the key and press the key; the
displays will show:
b) Press either the or key; the display will show:
c) Press the "FUNC" key; the display will show:
This indicates that the loading procedure has been initiated. After about 3 seconds the loading procedure is complete and the instrument reverts to the “Normal Display Mode.” The following is a list of the default operating parameters loaded during the procedure:
Default Operating Parameters List
Parameter Default Value Alarm Acknowledge OFF Software Key Unlock Setpoint Threshold Low range value (if low limit)
High range value (if high or high/
low limit) Setpoint1 Threshold Low range value Setpoint Threshold Hysteresis 0.1% Alarm Threshold Low range (if process alarm)
100 (if deviation or band alarm) Alarm Hysteresis 0.1%
Loading Default Configuration Parameters
The configuration parameters can be loaded with prede­termined default values. These are the settings loaded into the instrument prior to shipment from the factory. To load the default values proceed as follows:
a) Internal switch V101 must be open. b) The upper display will show:
c) Press the key; the lower display will show the
firmware version.
7sl-4-00.p65 9/17/02, 2:42 PM24
24
d) Still holding the key, press the key; the display
will show:
e) Press the key to select Table 1 (European) or
Table 2 (American) default parameters; the display will show:
f) Press the FUNC key; the display will show:
This indicates that the loading procedure has been initiated. After about 3 seconds the procedure is com­plete and the instrument reverts to the “COnF” display. The following is a list of the default configuration param­eters loaded during the procedure:
Default Configuration Parameter List
PARA. Table 1 Table 2
European American L1 nbUS nbus L2 1 1 L3 19200 19200 L4 8E 8E r1 Type J Type J
(-100 to 1000 °C) (-150 to 1830 °F) r2 ——. ——. r3 -100 -150 r4 1000 1830 r5 0 0 r6 1 second 1 second r7 uP uP c1 Hi Hi c2 1 0 c3 Auto Auto c4 0 0 c5 1 second 1 second P1 nonE nonE P2 H.A. H.A.Ac P3 rEV rEV P4 OFF OFF PF 1 second 1 second n1 0 0 t1 10 seconds 30 seconds
7sl-4-00.p65 9/17/02, 2:42 PM25
25
SPECIFICATIONS Case RABS Grey dark color (RAL 7043).
Self-extinguishing degree V0 according to UL 94
Front panel Designed and tested for IP65 and NEMA
4X for indoor location (when panel gasket is installed). Tests were performed in accordance with IEC 529, CEI 70-1 and NEMA 250-1991 STD.
Installation Panel mounting Rear Terminal Block
15 screw terminals with safety rear cover
Dimensions Unit with RS-485 communications
DIN 43700 48 x 48 mm x 122 mm (1.890 x 1.890 in. x 4.803 in).
Dimensions Unit without RS-485 communications
DIN 43700 48 x 48 mm x 105 mm (1.890 x 1.890 in. x 4.134 in).
Cutout 45 x 45 mm (1.772 x 1.772 in). Weight 250 g (8.75 oz). Upper Display 4 digits, 7 segment red LED’s with
decimal point, 7.62 mm high (0.3 in).
Lower Display 2 digits, 7 segment red LED’s with
decimal point, 7.62 mm high (0.3 in).
Indicator 2 red LED annunciators for:
“RESET” Indicates control output status
as follows: a) With configuration parameter
C2 = 0 LED ON when Output is OFF LED OFF when Output is ON
b) With configuration parameter
C2 = 1 LED flashes when Output is OFF LED ON when Output is OFF and acknowledged LED OFF when Output is ON
“ALM” indicates alarm status as follows:
Flashes when alarm is ON. ON when alarm has been acknowledged OFF when alarm is OFF
RMT A decimal point on the right side
of the upper display flashes to indicate the device is controlled by serial link
Keyboard Four keys (covered by polyester mem-
brane) labelled RESET, , & FUNC.
Power Supply 100 to 240 Vac 50/60 Hz switching or
24 Vac/Vdc with interruptions less than 100 ms
Power Supply Variation
-15% to 10% (for 100 to 240 Vac)
-10% to 10% (for 24 Vac/Vdc)
Power Consumption 8 VA max Insulation Resistance Greater than 100 M Insulation Strength1500 V rms for 1 minute EC Compatibility
Conforms with 89/336/EEC directives re­garding electro-magnetic compatibility Emission Complies with: Generic emission standard EN50081-2 Immunity Complies with the generic im­munity standard EN50082-2. Conforms with 72/23/EEC and 93/68/EEC directives regarding low voltage. It complies with the safety requirements for electrical equip­ment for measurement, control and laboratory use, Generic Standard EN61010-1, Installation Category II.
Sampling Time 250 ms for linear input
500 ms for TC/RTD input
7sl-5-00.p65 9/17/02, 2:43 PM26
26
Display Update Time
500 ms The display value can be filtered
Reference Accuracy
± 0.2% fsv ±1 digit @ 25 °C and nominal power supply voltage.
Temperature Drift
Less than 200 ppm/°C of full span for mV and TC ranges 1, 2, 4, 5, 9, 14 (CJ excluded). Less than 300 ppm/°C of full span for mA/ V and TC ranges 11, 12, 13 (CJ excluded) Less than 400 ppm/°C of full span for RTD range 15 and TC range 10 (CJ excluded). Less than 500 ppm/°C of full span for TC ranges 3, 6, 7 (CJ excluded). Less than 600 ppm/°C of full span for TC range 8. NOTE: Precision and drift guaranteed (for T>300°C/570°F).
Reference Junction Drift
0.1 °C/°C
Common Mode Rejection Ratio
120 dB @ 50/60 Hz
Normal Mode Rejection Ratio
60 dB @ 50/60 Hz
Ambient Temperature
0 to 50 °C
Storage Temperature
-20 to 70 °C
Humidity Maximum of 85% RH non condensing Input Thermocouple (J, K, T, E, N, S, R, B, L, U,
G, D, C, Platinel II), RTD Pt100, mVdc, Vdc, mAdc. The type of inputs are key­board and jumper selectable.
Range 1) TC type J -100 to 1000 °C
-150 to 1830 °F
2) TC type K, -100 to 1370 °C
-150 to 2500 °F
3) TC type T -200 to 400 °C
-330 to 750 °F
4) TC type E -100 to 800 °C
-150 to 1470 °F
5) TC type N -100 to 1400 °C
-150 to 2550 °F
6) TC type S -50 to 1760 °C
-60 to 3200 °F
7) TC type R -50 to 1760 °C
-60 to 3200 °F
8) TC type B 0 to 1820 °C 32 to 3300 °F
9) TC type L [DIN43710-1977]
-100 to 900 °C
-150 to 1650 °F
10) TC type U [DIN43710-1977]
-200 to 600 °C
-330 to 1110 °F
11) TC type G [W - W, 26% Re]
0 to 2300 °C 0 to 4170 °F
12) TC type D [W, 3% Re - W, 25% Re]
0 to 2300 °C 0 to 4170 °F
13) TC type C 0 to 2300 °C
0 to 4170 °F
14) TC type Platinel II
-100 to 1400 °C
-150 to 2550 °F
15) RTD PT100 -200 to 850 °C
-330 to 1560 °F
16) Linear 0 to 20 mA
17) Linear 0 to 60 mV
18) Linear 0 to 5 V
19) Linear 0 to 10 V
7sl-5-00.p65 9/17/02, 2:43 PM27
27
Source Impedance
100 maximum for TC/mV input Less than 20 per wire for RTD input
Input Impedance
Greater than 1 M for TC/mV input Greater than 400 K for 5V to 10V input range Less than 5 for mA input
Cold Junction
Automatic compensation from 0 to 50 °C
Digital Input Input from dry contact (voltage free) to
restart process (optional)
Control OUT 1
Relay (form C) 3 Amp @ 250 Vac resistive load (with varistor on contacts) (Relay operates in failsafe mode)
Limit Type High limit (for heating process)
Low limit (for cooling process) High/Low limit (for special process)
Acknowledgment Mode
To restart the process, two conditions must apply a) The process returns within setpoint b) The operator has acknowledged the process (from keyboard, digital input or serial link) The “a” and “b” sequence may or may not matter depending on C2 (see C2)
Power-Up Mode
Automatic or Manual restart. When in automatic restart the control output is guar­anteed OFF for 1 second.
Setpoint Threshold
Span limits
Threshold hysteresis
From 0.1 to 10.0% of span value or 1 LSD (whichever is greater)
OUT 2 Relay (form A) 2 Amp @ 250 Vac resistive
load. OUT 2 is used as alarm output (optional)
Alarm Function
Process alarm Deviation alarm Band alarm
Type of Alarm
High/Low (Outside/Inside if band alarm) Direct/Reverse Automatic/Manual reset Standby sequence/No standby sequence
Alarm Threshold
Span limits for process alarm From 0 to 500 digits for band alarm From -500 to 500 digits for deviation alarm
Alarm Hysteresis
From 0.1 to 10.0% of span value or 1 LSD (whichever is greater)
Serial Interface
RS-485 Opto-isolated (optional)
Protocol Type
Modbus/Jbus (RTU mode)
Device Address
From 1 to 255. NOTE: The interface can support up to 31 devices
Baud Rate 600 up to 19200 baud Format 1 start bit
8 bit with/without parity 1 stop bit
Parity Even/Odd Watch dog Hw/Sw is provided for automatic restart Protection Internal dip switch for calibration and con-
figuration parameter protection.
7sl-5-00.p65 9/17/02, 2:43 PM28
28
References UL 94 Tests for flammability of plastic materials
for pans in devices and appliances
CEI 70-1 (IEC 529)
Degrees of protection provided by enclo­sures (IP Code)
NEMA 250-1991
Enclosures for equipment (1000 Volts maximum)
DIN 43700 Measurements and control instruments
for panel mounting. Nominal front and cut-out dimensions.
EN 50081-2 Electromagnetic compatibility - Generic
emission standard - Part 2. Industrial environment.
EN 55011 Limits and methods of measurement of
radio disturbance characteristics of in­dustrial, scientific and medical (ISM) radio frequency equipment.
EN 50082-2 Electromagnetic compatibility - Part 2 -
Industrial environment
ENV50140 Electromagnetic compatibility - Basic im-
munity standard ­Radiated radio - frequency electro-mag­netic field- Immunity test.
IEC 1000-4-2 Electromagnetic compatibility (EMC)
Part 4: Testing and measurement tech­nique. Section 2: Electrostatic discharge immunity test.
EN-61000-4-8Electromagnetic compatibility (EMC)
Part 4: Testing and measurement tech­niques Section 8: Power frequency magnetic field immunity test.
IEC 1000-4-4 Electromagnetic compatibility. Part 4:
Testing and measurement techniques ­Section 4: Electrical fast transient/burst immunity test.
ENV50141 Electromagnetic compatibility - Basic im-
munity standard - Conducted disturbances induced by radio-frequency fields - Immu­nity test.
DIN 43710-1977
Thermocouples - Reference table.
IEC 584-1 Thermocouples - Part 1 - Reference table. DIN 43760 Thermometer Reference table.
7sl-5-00.p65 9/17/02, 2:43 PM29
29
CALIBRATION PROCEDURE
Calibration parameters are logically divided into groups of two parameters each – initial scale value and final scale value. A calibration check is provided after entering the values of each group. It is also possible to perform a calibration check without making an entry: press the FUNC button twice when “OFF” is displayed. The instru­ment goes directly to the group check.
Before beginning calibration, be sure the internal DIP switch V101 is open.
WARNING: Perform the calibration procedure according to J106 jumper positions as shown in Figure 1. Other­wise, the stored calibration values may be lost.
epyTtupnI
2-14-36-58-7
601J
,DTR,C/T
Vm06-21,Vm06-0
nepo
esolc
nepo nepo
)noitarugifnocdradnats(
V5-1,V5-0
nepo esolc nepo nepo
V01-2,V01-0 nepo nepo esolc nepo
Am02-4,Am02-0
nepo nepo nepo esolc
K081=291RCTTCtupnI
nepO=101HCdesolC=021HS
Figure 1. J106 Settings
General Guidelines
a) The instrument should be mounted in its case in order
to keep the internal temperature constant.
b) Ambient temperature should be stable. Avoid drift due
to air conditioning or other mechanical devices. c) Relative humidity should not exceed 70%. d) Minimum warm up time of at least 20 minutes. e) Operate as much as possible in a noise free
environment.
f) during calibration, connect one input at a time to the
rear terminal block.
g) Use calibrators with the following:
Accuracy TC Input: +0.005% output
+0.001% range +5 mV
RTD Input: +0.02%
+0.0025 decade CJ Compensation: better than 0.1°C Current Transformer: 0.1 mA AC rms Resolution TC Input: 1 mV RTD Input: 10 m CJ Compensation: better than 0.1°C Current Transformer: 0.1 mA AC rms
Calibration Parameters
Following is a complete list of calibration symbols: CodeParameter tL TC Input Initial Scale Value tH TC Input Final Scale Value t. TC Input Check rJ Cold Junction Compensation rJ. Cold Junction Compensation Check PL RTD Input Initial Scale Value PH RTD Input Final Scale Value P. RTD Input Check AL Current Input Initial Scale Value AH Current Input Final Scale Value A. Current Input Check nL 5 Volt Input Initial Scale Value nH 5 Volt Input Final Scale Value n. 5 Volt Input Check UL 10 Volt Input Initial Scale Value UH 10 Volt Input Final Scale Value U. 10 Volt Input Check
7sl-6-00.p65 9/17/02, 2:44 PM30
30
Procedure
Switch on the instrument; the upper display will show “COnf”. Press the button; the upper display will show “CAL”.
Using the and pushbuttons (buttons), it is possible to select between ON and OFF. To go to the next parameter without modifying the calibration, press the FUNC button when the display shows “OFF”. To enter a calibration value, press the FUNC button when the dis­play shows “ON”.
Press the FUNC button to show the first calibration code on the lower display. Repeatedly press the FUNC button until the desired calibration (parameter) code appears.
NOTE: by pushing the SMRT button it is possible to go back to a previous parameter without memorizing the new calibration.
Entering Calibration Values
Following is a detailed, sequential procedure for entering and checking values for each calibration parameter:
tL TC Input Initial Scale Value
a) Connect calibrator and instrument as shown in
Figure 2.
10
9
Figure 2. Calibrator Connection
b) The upper display shows “OFF”, the lower displays
shows “tL”. c) Set calibrator to 0.000 mV. d) Push the button; the display changes to “ON”. e) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the instrument will go to the next parameter.
tH TC Input Final Scale Value
a) Set the calibrator to 60.000 mV. b) Push the
and “tH”.
c) After a few seconds, start calibration by pressing the
FUNC button. When this calibration is complete, the instrument will go to the next parameter.
t. TC Input Check
The display (Figure 3) shows “t.” followed by a number showing the measured value in counts:
button; the displays will show “ON”
t.300
00
Figure 3. TC Calibration Check Display
a) Check the calibration (linear) by setting:
0.000 mV – the readout must be equal to “t.000 00” ± 10 counts;
60.000 mV – the readout must be equal to “t.300 00” ± 10 counts;
30.000 mV – the readout must be equal to “t.150 00” ± 10 counts.
b) Push the FUNC button to go to the next parameter.
rJ Cold Junction Compensation
NOTE: Make sure tL and tH are correctly calibrated before attempting rJ calibration. a) Measure the temperature close to terminals 9 and 10
using an appropriate instrument – for example, MEMOCAL. See Figure 4.
31
7sl-6-00.p65 9/17/02, 2:44 PM31
10
Measuring Device
9
Figure 4. Measure Temperature Near Terminals
b) Wait a few minutes to allow temperature stabilization
of the entire system (compensation cable, sensor,
calibrator and instrument). c) The displays show “rJ” and “OFF”. Using the or
button, make the readout value equal to the
temperature measured by the measuring device in
tenths of °C. d) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
rJ. Cold Junction Compensation Check
The displays show “rJ.” and the cold junction temperature in tenths of °C. Make sure the display readout is equal to the value read on the measuring device. Then, push the FUNC button to go to the next parameter.
PL RTD Input Initial Scale Value
a) Connect a resistor box and the instrument as shown
in Figure 5.
8
9
b) Set 0.00 on the resistor box. c) Push the button; the displays show “ON” and “PL”.
d) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the instrument will go to the next parameter.
PH RTD Input Final Scale Value
a) Set resistor box to 375.00 Ω. b) Push the button; the displays will show “ON” and
“PH”.
c) After a few seconds, start calibration by pressing the
FUNC button. When this calibration is complete, the instrument will go to the next parameter.
P. RTD Check
The display (Figure 6) shows “P.” followed by a number showing the measured value in counts:
P.300
00
Figure 6. RTD Calibration Check Display
a) Check the calibration (linear) by setting:
0.00 – the readout must be equal to “P.000 00” ± 10 counts;
125.000 – the readout must be equal to “P.101 90” ± 10 counts;
375.00 – the readout must be equal to “P.300 00” ± 10 counts.
b) Push the FUNC button to go to the next parameter.
10
Figure 5. Resistor Box Connection
7sl-6-00.p65 9/17/02, 2:44 PM32
32
AL Current Input Initial Scale Value
a) Connect calibrator and instrument as shown in
Figure 7.
nL 5 VOLT Input Initial Scale Value
a) Connect calibrator and instrument as shown in
Figure 8.
10
9
Figure 7. Calibrator Connection
b) The upper display shows “OFF”, the lower displays
shows “AL”. c) Set calibrator to 0.000 mA (even if the minimum range
value is 4mA). d) Push the button; the display changes to “ON”. e) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
AH Current Input Final Scale Value
a) Set the calibrator to 20 mA b) Push the button; the displays will show “ON”
and “AH”. c) After a few seconds, start calibration by pressing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
A. Current Input Check
The display shows “A.” followed by a number showing the measured value in counts.
a) Check the calibration (linear) by setting:
0.000 mA – the readout must be equal to
“A.000 00” ± 10 counts;
20.000 mA – the readout must be equal to
“A.300 00” ± 10 counts;
10.000 mA – the readout must be equal to
“A.150 00” ± 10 counts. b) Push the FUNC button to go to the next parameter.
10
9
Figure 8. Calibrator Connection
b) The upper display shows “OFF”, the lower displays
shows “nL”.
c) Set calibrator to 0.000 V (even if the minimum range
value is 1V). d) Push the button; the display changes to “ON”. e) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
nH 5 Volt Input Final Scale Value
a) Set the calibrator to 5.000 V. b) Push the button; the displays will show “ON”
and “nH”. c) After a few seconds, start calibration by pressing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
n. 5 Volt Input Check
The display shows “n.” followed by a number showing the measured value in counts.
a) Check the calibration by setting:
0.000 V – the readout must be equal to
“n.000 00” ± 10 counts;
5.000 V – the readout must be equal to
“n.300 00” ± 10 counts;
2.500 V – the readout must be equal to
“n.150 00” ± 10 counts. b) Push the FUNC button to go to the next parameter.
33
7sl-6-00.p65 9/17/02, 2:44 PM33
UL 10 VOLT Input Initial Scale Value
a) Connect calibrator and instrument as shown in
Figure 9.
10
9
Figure 9. Calibrator Connection
b) The upper display shows “OFF”, the lower displays
shows “UL”.
c) Set calibrator to 0.000 V (even if the minimum range
value is 2 V). d) Push the button; the display changes to “ON”. e) After a few seconds, start calibration by pushing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
UH 10 Volt Input Final Scale Value
a) Set the calibrator to 10.000 V. b) Push the button; the displays will show “ON”
and “10UH”. c) After a few seconds, start calibration by pressing the
FUNC button. When this calibration is complete, the
instrument will go to the next parameter.
U. 10 Volt Input Check
The display shows “10UH.” followed by a number show­ing the measured value in counts.
a) Check the calibration by setting:
0.000 V – the readout must be equal to “U.000 00” ± 10 counts;
10.000 V – the readout must be equal to “U.300 00” ± 10 counts;
5.000 V – the readout must be equal to “U.150 00” ± 10 counts.
b) Push the FUNC button.
7sl-6-00.p65 9/17/02, 2:44 PM34
34
MAINTENANCE
1. Remove power from the power supply terminals and
from relay output terminals.
2. Remove the instrument from case.
3. Using a vacuum cleaner or a compressed air jet (max.
3kg/cm2) remove dust and dirt which may be present
on the louvers and on the internal circuits, being
careful to not damage the electronic components.
4. Clean external plastic or rubber parts only with a cloth
moistened with ethyl alcohol (pure or denatured)
[C2H5OH]; or isopropyl alcohol (pure or denatured)
[(CH3)2CHOH]; or water [H2O]
5. Verify that there are no loose terminals.
6. Before re-inserting the instrument in its case, be sure
it is perfectly dry.
7. Re-insert the instrument and turn it ON.
7sl-z-00.p65 9/17/02, 2:46 PM35
35
170.IU0.LHL.S00 02
Cincinnati, OH 800-666-4330 http://www.supersystems.com
Manual Part No. 31302 7SL
7sl-z-00.p65 9/17/02, 2:46 PM36
Copyright © 2001 Super Systems Inc.
Loading...