Super Systems 7EK 31080 Calibration Manual

Page 1
0831106 Rev.0
Page 2
INSTRUMENT CALIBRATION PROCEDURES
1 GENERAL GUIDELINES FOR CALIBRATION
For an accurate calibration it is necessary to proceed as follows: a) - The instrument under calibration should be mounted in its case in
order to keep the internal temperature stable.
Avoid any drift due to air-conditioning or others. c) - The relative humidity should not exceed 70%. d) - The instrument must be in ON condition from 20 minutes at least. e) - Operate, possibly, in an environment with no electromagnetic
disturbances. f) - During calibration, connect to the instrument one input at a time. g) - Before to execute each calibration, be sure that the specific
hardware setting has been made (see "Preliminary hardware
setting" paragraph).
For this calibration procedure it is necessary to use calibrators with the following accuracy and resolution:
ACCURACY
1) For current input:
2) For voltage input : + 0.005% output + 0.001% range + 5 µV
3) For TC input: + 0.005% output + 0.001% range + 5 µV
4) For RTD input:
5) For cold junction compensation: better than 0.1 °C
RESOLUTION
1) For current input: 0.5 µA
2) For voltage input: 100 µV
3) For TC input: 1 µV
4) For RTD input: 10 m
5) For cold junction compensation: better than 0.1 °C
2 PRELIMINARY HARDWARE SETTINGS
INPUT TYPE SELECTION
1) Remove the instrument from its case.
2) Set J1according to the desired input type.
3) Set the dip switch V1to the open condition (see fig. 1).
4) Re-insert the instrument in its case.
+ 0.025% output + 0.0025% range + 0.01 µA
+ 0.02 % + 0.0025 /decade.
3 CALIBRATION PROCEDURE
3.1 FOREWORD
Calibration parameters are divided in groups. Each group is comprised of two parameters (initial and full scale values) plus a specific calibration check. Follows a complete list of the "calibration groups" is reported.
A) TC input B) Cold junction C) RTD input D) Linear mA input E) Linear 5 V input
F) Linear 10 V input G) Current transformer input calibration H) Default calibration data loading
3.2 HOW TO PROCEED
1) Switch on the instrument the display will show "CONF".
2) Push the the lower display is blank.
3) Push the FUNC pushbutton to display the first calibration group on the lower display. Depress FUNC pushbutton more times until the desired calibration group is reached.
GENERAL NOTES ABOUT CALIBRATION PROCEDURE:
1) During calibration procedure, when the initial or full scale value of a group is selected and the upper display shows "OFF", pushing the FUNC pushbutton the instrument will jump to the next parameter for checking without to modify the previous calibration setting. In this way it is possible to recalibrate only the desired input or output. It is also possibe to make a check of one or more calibration group without to remake the specific calibration. Pushing the MAN pushbutton it is possible to scroll back through the calibration parameters without to memorize the setting.
2) During calibration procedure the serial communication interface and the time out will be disabled.
A) TC INPUT CALIBRATION
pushbutton, the upper display will show "CAL" while
1JtupnI
Type 1-2 3-4 5-6 7-8 9-10
TC-RTD open close open open open
60 mV open close open open open
5 V close open close open open
10 V open open close open open
20 mA open open open close close
NOTE: the not used jumper can be placed on pin 7-9
2 4 6 8 10
1 3 5 7 9
V1
Fig. 1
J1
A.1) "tL" - INITIAL SCALE VALUE
The lower display will show "tL" while the upper display will show "OFF"
1) Ma
ke the specific hardware setting as described at paragraph 2.
2) Connect the instrument under test to the calibrator as shown in
Fig. 2.
_
3
+
1
Fig.2
3) Set calibrator to 0.000 mV.
4) Push or pushbutton, the upper display will change to "On".
5) After few seconds, start the calibration by pushing the FUNC
pushbutton. At the end of this calibration routine, the instrument will go to the next step.
2
Page 3
A.2) "tH" - FINAL SCALE VALUE
The lower display will show "tH" while "OFF" is shown on the upper display.
1) Set the calibrator to 60.000 mV (see Fig. 2).
2) Push or pushbutton, the upper display will change to "On".
3) After few seconds, start calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
..
A.3) "t
." - TC INPUT CHECK
..
The upper display show "t." followed by the measured value in counts as shown in the following figure:
Check symbol
The "tH" calibration is correct if the indication is equal to "t. 3 0000"
1) Check the zero calibration, by setting the calibrator to 0.000 mV,
2) Check the half scale linearity by setting the calibrator to 30.000
+ 10 counts.
the read-out must be equal to "t. 0 0000"
mV. The read-out must be "t. 1 5000" NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
Instrument readout (in counts) =
3) Push FUNC pushbutton, the instrument will go to the next calibration group.
B) COLD JUNCTION CALIBRATION
NOTE: make sure that "tL" and "tH" parameters are correctly
calibrated before to calibrate "rJ" parameter.
B.1) rJ - ACTUAL VALUE
The lower display will show "rJ" while the upper display will show "OFF"
1) Measure the temperature close to terminals 1 and 3 using a calibrator, for instance, the MEMOCAL (see Fig. 3).
Measured or generated value (in counts)
+ 10 counts.
+ 10 counts.
input value
60 (mV)
30000
B.2) "rJ" - COLD JUNCTION COMPENSATION CHECK
The upper display will show "rJ." while the lower display will show the measured cold junction temperature (in C° and tenths of °C).
1) Make sure that the cold junction temperature measured by the instrument is equal to the value measured by the measuring device (MEMOCAL).
2) Push FUNC pushbutton, the instrument will go to the next calibration group.
C) RTD INPUT CALIBRATION
NOTE: make sure that "tL", "tH" and "rJ" parameters are correctly
calibrated before to calibrate this parameter group.
C.1) "PL" - INITIAL SCALE VALUE
The lower display will show "PL" while the upper display will show "OFF"
ke the specific hardware setting as described at paragraph 2.
1) Ma
2) Make a short circuit among the 1, 3 and 4 terminals as shown in Fig. 4.
1
3
4
Fig.4
3) Push or pushbutton, the upper display will change to "On".
4) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
5) Remove the short circuit from the 1, 3 and 4 terminals.
C.2) "PH" - FINAL SCALE VALUE
The lower display will show "PH" while the upper display will show "OFF"
1) Connect the instrument under test to the calibrator as shown in Fig. 5.
1
+
1
RTD
_
3
Fig. 3
2) Wait for a few minutes to allow the temperature stabilisation of the entire system (sensor, calibrator and instrument). The first pressure of or push-button allows to show on the upper display a read-out value.
3) Using or push-button, set a read-out value equal to the temperature tenths of °C).
4) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
measured by the measuring device (in C° and
3
4
Fig.5
2) Set the calibrator to 375.00 .
3) Push or pushbutton, the upper display will change to "On".
4) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of t will go to the next step.
C.3) "P." - RTD INPUT CHECK
The upper and lower display show "P." followed by the measured value in counts.
Check symbol
his calibration routine, the instrument
Measured or generated value (in counts)
3
Page 4
The "PH" calibration is correct if the indication is equal to "P. 3 0000"
+ 10 counts.
1) Check the zero calibration, by making the short circuit as described at point C.1, the read-out must be equal to "P. 0 0000" + 10 counts.
2) Remove the short circuit.
3) Connect the calibrator as shown in Fig. 7.
4) Check the linearity by setting the calibrator to 125.00 . The read-out must be "P. 1 0190"
5) Set the calibrator to 250.00 . The read-out must be "P. 2 0189"
6) Push FUNC pushbutton, the instrument will go to the next
+ 10 counts.
calibration group.
+ 10 counts.
E) 5 V INPUT CALIBRATION
E.1) "5UL" - INITIAL SCALE VALUE
The lower display will show "5UL" while the upper display will show "OFF"
1) Make the specific hardware setting as described at paragraph 2.
2) Connect the instrument under test to the calibrator as shown in Fig. 7.
_
3
+
1
D) mA INPUT CALIBRATION
D.1) "nAL" - INITIAL SCALE VALUE
The lower display will show "nAL" while the upper display will show "OFF"
1) Make the specific hardware setting as described at paragraph 2.
2) Connect the instrument under test to the calibrator as shown in Fig. 6.
_
3
+
1
Fig. 6
3) Set calibrator to 0.000 mA.
4) Push or pushbutton, the upper display will change to "On".
5) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
D.2) "nAH" - FINAL SCALE VALUE
The lower display will show "nAH" while the upper display will show "OFF"
1) Set the calibrator to 20.000 mA.
2) Push or pushbutton, the upper display will change to "On".
3) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration rou will go to the next step.
D.3) "nA." - mA INPUT CHECK
The upper and lower show "nA." followed by the measured value in counts.
Check symbol
The "nAH" calibration is correct if the indication is equal to "nA. 3 0000"
+ 10 counts.
1) Check the zero calibration, by setting the calibrator to 0.000 mA, the read-out must be equal to "nA. 0 0000"
2) Check the linearity by setting the calibrator to 10.000 mA , the read-out must be "nA. 1 5000" NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
+ 10 counts.
Instrument readout (in counts) =
3) Push FUNC pushbutton, the instrument will go to the next calibration group.
tine, the instrument
Measured or generated value (in counts)
+ 10 counts.
input value
20 (mA)
30000
Fig.7
3) Set calibrator to 0.000 V.
4) Push or pushbutton, the upper display will change to "On".
5) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
E.2) "5UH" - FINAL SCALE VALUE
The lower display will show "5UH" while the upper display will show "OFF"
1) Set the calibrator to 5.000 V.
2) Push or pushbutton,
3) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
E.3) "5U." - 5 V INPUT CHECK
The upper and the lower displays show "5U." followed by the measured value in counts.
Check symbol
The "5UH" calibration is correct if the indication is equal to "5U. 3
+ 10 counts.
0000"
1) Check the zero calibration, by setting the calibrator to 0.0000 V, the read-out must be equal to "5U. 0 0000"
2) Check the linearity by setting the calibrator to 2.500 V The read­out must be "5U. 1 5000" NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
Instrument readout (in counts) =
3) Push FUNC pushbutton, the instrument will go to the next calibration group.
F) 10 V INPUT CALIBRATION
F.1) "10UL" - INITIAL SCALE VALUE
The lower display will show "10UL" while the upper display will show "OFF"
1) Make the specific hardware setting as described at paragraph 2.
2) Connect the instrument under test to the calibrator as shown in Fig. 8.
_
3
+
1
Fig.8
the upper display will change to "On".
Measured or generated value (in counts)
+ 10 counts.
+ 10 counts.
input value
5 (V)
30000
4
Page 5
3) Set calibrator to 0.000 V.
4) Push or pushbutton, the upper display will change to "On".
5) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
F.2) "10UH" - FINAL SCALE VALUE
The lower display will show "10UH" while the upper display will show "OFF"
1) Set the calibrator to 10.000 V.
2) Push or pushbutton, the upper display will change to "On".
3) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
F.3) "10U." - 10 V INPUT CHECK
The upper and the lower displays show "10U." followed by the measured value in counts.
Check symbol
The "10U." calibration is correct if the indication is equal to "10U.3
+ 10 counts.
0000"
1) Check the zero calibration, by setting the calibrator to 0.000 V, the read-out must be equal to "10U.0 0000"
2) Check the linearity by setting the calibrator to 5.000 V The read­out must be "10U.1 5000" NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
+ 10 counts.
Instrument readout (in counts) =
3) Push FUNC pushbutton, the instrument will go to the next calibration group.
G) CURRENT TRANSFORMER INPUT CALIBRATION
G.1) "ct.L" - INITIAL SCALE VALUE
The lower display will show "ct.L" while the upper display will show "OFF"
1) Made the specific hardware setting as described at paragraph 2.
2) Connect the instrument under test to the calibrator as shown in Fig. 9.
5
6
Measured or generated value (in counts)
+ 10 counts.
input value
10 (V)
mA AC
Generator
30000
G.3) "ct." - CURRENT TRANSFORMER INPUT CHECK
The upper and the lower displays show "Ct." followed by the measured value in counts.
Check symbol
The "ct.H" calibration is correct if the indication is equal to "ct. 0 1000"
+ 10 counts.
1) Check the zero calibration, by setting the calibrator to 0.00 mA AC, the read-out must be equal to "ct. 0 0000"
2) Check the linearity by setting the calibrator to 25.00 mA AC. The read-out must be "ct. 0 0500" NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
+ 10 counts.
Instrument readout (in counts) =
3) Push FUNC pushbutton, the instrument will go to the next calibration group.
H) DEFAULT CALIBRATION PARAMETER LOADING.
A complete and consistent set of calibration parameters is memorized in the instrument. These data are theoretical data and are used only to clear all calibration memory but after a default calibration data loading it is necessary to make all calibrations.
When is desired to clear all calibration memory proceed as follows: a) The upper display will show "CAL" while the lower display is blank. b) Push the push-button, the lower display will show the firmware
version.
c) Push and keep depressed the push-button again and push the
push-button, the lower display will show "dFLt". d) Push or pushbutton, the upper display will change to "On". e) Push the FUNC push-button, the upper display will show:
Then the upper display will show CAL again and the default calibra­tion parameter loading procedure is ended.
Measured or generated value (in counts)
+ 10 counts.
input value
50 (mA AC)
1000
WARNING: After default calibration data loading, it is necessary to remake all instrument calibrations.
Fig.9
3) Set calibrator to 0.00 mA AC.
4) Push or pushbutton, the upper display will change to "On".
5) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
G.2) "ct.H" - FINAL SCALE VALUE
The lower display will show "ct.H" while the upper display will show "OFF"
1) Set the calibrator to 50.00 mA AC.
2) Push or pushbutton, the upper display will change to "On".
3) After a few seconds, start the calibration by pushing FUNC pushbutton. At the end of this calibration routine, the instrument will go to the next step.
5
Loading...