Super Systems 20PQ Calibration Manual

0831106 Rev.0
Calibration procedure
MKP PKP MKC PKC
Calibration procedure for MKP - PKP
1 GENERAL GUIDELINES FOR CALIBRATION
For an accurate calibration it is necessary to proceed as follows: a) - The instrument under calibration should be mounted in its case in
order to keep the internal temperature stable.
b) - The ambient temperature should be stable.
Avoid any drift due to air-conditioning or others. c) - The relative humidity should not exceed 70%. d) - The instrument must be in ON condition from 20 minutes at least. e) - Operate, possibly, in an environment with no electromagnetic
disturbances. f) - During calibration, connect to the instrument one input at a time. g) - Before to execute each calibration, be sure that the specific
hardware setting has been made (see "Preliminary hardware
setting" paragraph). For this calibration procedure it is necessary to use calibrators with
the following accuracy and resolution:
ACCURACY
1) For current input:
2) For voltage input : + 0.005% output + 0.001% range + 5 µV
3) For TC input: + 0.005% output + 0.001% range + 5 µV
4) For RTD input:
5) For cold junction compensation: better than 0.1 °C
RESOLUTION
1) For current input: 0.5 µA
2) For voltage input: 100 µV
3) For TC input: 1 µV
4) For RTD input: 10 m
5) For cold junction compensation: better than 0.1 °C
2 PRELIMINARY HARDWARE SETTINGS
+ 0.025% output + 0.0025% range + 0.01 µA
+ 0.02 % + 0.0025 /decade.
D
C
D
Fig. 2
2.2 MAIN INPUT SELECTION
Set J103 (see fig. 3) according to the desired input type as shown in the following table.
J103 INPUT TYPE
T/C, RTD,CJ 60 mV 5 V 10 V 20 mA 1-2 open open close open open 3-4 open open close close open 5-6 open open open open close 7-8 open open open open close 5-7 close close open close open 6-8 close close open open open
1 3 5 7
J102
2 4 6 8
1 3 5 7
J103
2 4 6 8
CPU card
2.1 HOW TO REMOVE THE INSTRUMENT FROM ITS CASE
1) Switch off the instrument.
2) Push gently the lock A on the right.
3)While the lock A is maintained out, slide out the right side of the
instrument (see fig. 1).
B
A
B
Fig. 1
4) Push gently the lock C on the left.
5)While the lock C is maintained out, slide out the instrument
(see fig. 2).
J205
ON DIP
V301
1 2 3 4
Fig. 3
2.3 AUXILIARY INPUT SELECTION (option)
Set J102 (see fig. 3) according to the desired input type as shown in the following table.
J102 INPUT TYPE
5 V 10 V 20 mA 1-2 close open open 3-4 close close open 5-6 open open close 7-8 open open close 5-7 open close open 6-8 open open open
2
Calibration procedure for MKP - PKP
2.4 IN CT / FEEDBACK SELECTION
This instrument can use the "IN CT" input or the "Feedback" input; the two inputs are not contemporarily. The current transformer input allows you to measure and display the current running in a load driven by a time proportional control output during the ON and OFF periods of the output cycle time. By this feature it is also available the "Out failure detection" function (see page 66 in the USER MANUAL). The feedback input is used when the servomotor close loop or the servomotor open loop with valve position indication outputs is required. To select the desired input type, set V301 (see fig. 3) as detailed in the following table:
Input V301.1 V301.2 V301.3 V301.4 IN CT ON OFF ON ON Feedback OFF ON OFF ON
2.5 CALIBRATION MODE SELECTION
To start the calibration procedure, the DIP SWITCH V101, mounted on CPU card, must be set as follows: V101.1 = Not care condition V101.2 = OFF V101.3 = Not care condition V101.4 = ON NOTE: during calibration procedure the serial communication interface will be disabled.
ON DIP
1 2 3 4
V101
3 CALIBRATION PROCEDURE
3.1 FOREWORD
Calibration parameters are divided in groups. Each group is comprised of two parameters (initial and full scale values) plus a specific calibration check.
Follows a complete list of the "calibrations groups".
1) ñ.I.tc = Main input calibration, TC input
2) ñ.I.CJ = Main input calibration, cold junction
3) ñ.I.rt = Main input calibration, RTD input
4) ñ.I.ñA = Main input calibration, mA input
5) ñ.I.5 = Main input calibration, 5 V input
6) ñ.I.10 = Main input calibration, 10 V input
7) A.I.ñA = Auxiliary input calibration, mA input
8) A.I.5 = Auxiliary input calibration, 5 V input
9) A.I.10 = Auxiliary input calibration, 10 V input
10) In.Ct = Current transformer input calibration
11) FEEd = Feedback input
12) 05.ñA = Out 5 calibration
13) 06.ñA = Out 6 calibration
14) CAL = Default calibration data loading NOTE: Calibration groups from group 7 to group 13 will be shown only when the specific hardware is fitted.
GENERAL NOTE ABOUT CALIBRATION PROCEDURE: During calibration procedure, when the initial or full scale value of a group is selected and the middle display shows "OFF", pushing the FUNC pushbutton the instrument will jump to the next parameter or check without to modify the previous calibration setting. In this way it is possible to recalibrate only the desired input or output. It is also possibe to make a check of one or more calibration group without to remake the specific calibration.
Fig. 4 When it is desired to exit from calibration mode proceed as follows:
1) switch off the instrument;
2) remove the instrument from its case;
3) select the desired operative mode by setting the DIP switch V101 as described in "Operative mode and hardware lock" paragraph reported at page 23 of the USER MANUAL.
3.2 CALIBRATION PROCEDURE
HOW TO PROCEED
Switch on the instrument, the upper display will show CAL while the lower display will show the firmware version. Push the FUNC pushbutton to display the first calibration group on the upper display. Depress FUNC pushbutton more times until the desired calibration group is reached.
1) "ñ.I.tc" - MAIN INPUT CALIBRATION - TC INPUT
The upper display will show "ñ.I.tc".
1.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 5.
_
3
+
1
Fig.5 c) The upper display will show "ñ.I.tc", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 mV. e)Push pushbutton, the middle display will change to "On". f) After few seconds, start calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
3
Calibration procedure for MKP - PKP
1.2) "Hr" - FINAL SCALE VALUE
The upper display will show "ñ.I.tc", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 60.000 mV (see Fig. 5). b) Push pushbutton, the middle display will change to "On". c) After few seconds, start calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
1.3) "
∀." - TC INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts as shown in the following figure:
Selected calibration group Check simbol
Fig. 6 The "ñ.I.tc" "Hr" calibration is correct if the indication is equal to "
0000"
a)Check the zero calibration, by setting the calibrator to 0.000 mV, b)Check the half scale linearity by setting the calibrator to 30.000 mV.
+ 10 counts. the read-out must be equal to " The read-out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
∀∀
∀." followed by the measured
∀∀
Measured or generated value (in counts)
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
+ 10 counts.
input value
60 (mV)
30000
∀∀
∀. 3
∀∀
∀.∀.
2.2) "
∀." - COLD JUNCTION COMPENSATION CHECK
∀.∀.
The middle display will show " The lower display will show the measured cold junction temperature (in C° and tenths of °C). a)Make sure that the cold junction temperature measured by the
instrument is equal to the value measured by the measuring device (MEMOCAL).
b)Push FUNC pushbutton, the instrument will go to the next
calibration group.
3) "ñ.I.rt" - MAIN INPUT CALIBRATION - RTD INPUT
The upper display will show "ñ.I.rt".
NOTE: make sure that "ñ.I.tc" "Lr", "ñ.I.tc" "Hr" and "ñ.I.CJ" param-
eters are correctly calibrated before to calibrate "ñ.I.rt" parameter.
3.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Made a short circuit between terminals 1, 3 and 4 as shown in
Fig. 8.
1
3
4
Fig.8
∀∀
∀.".
∀∀
2) "ñ.I.CJ" - MAIN INPUT CALIBRATION - COLD JUNCTION
The upper display will show "ñ.I.CJ"
NOTE: make sure that "ñ.I.tc" "Lr" and "ñ.I.tc" "Hr" parameters are
correctly calibrated before to calibrate "ñ.I.CJ" parameter.
2.1) Lr - ACTUAL VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Measure the temperature close to terminals 1 and 3 using a
calibrator, for instance, the MEMOCAL (see Fig. 7).
1
+
RTD
_
3
Fig. 7 c) Wait a few minutes to allow the temperature stabilisation of the
entire system (sensor, calibrator and instrument).
d)The middle displays will show "OFF".
At the first pressure of or push-button, the middle display starts to show a read-out value.
e)Using or push-button, set a read-out value equal to the
temperature measured by the measuring device (in C° and tenths of °C).
f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument will go to the next step.
c) The upper display will show "ñ.I.rt", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Push pushbutton, the middle display will change to "On". e)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
3.2) "Hr" - FINAL SCALE VALUE
The upper display will show "ñ.I.rt", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Connect the instrument under test to the calibrator as shown in
Fig. 9.
1
3
4
Fig.9 b)Set the calibrator to 375.00 .
c) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
3.3) "
∀." - RTD INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "ñ.I.rt" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.00 , the b)Check the linearity by setting the calibrator to 175.00 . The
+ 10 counts.
0000" read-out must be equal to " read-out must be "
∀∀
∀. 1 4213"
∀∀
∀∀
∀. 0 0000"
∀∀
∀∀
∀." followed by the measured
∀∀
+ 10 counts.
+ 10 counts.
∀∀
∀. 3
∀∀
4
Calibration procedure for MKP - PKP
30000
∀∀
∀. 2 2166"
∀∀
c) Set the calibrator to 275.00 . The read-out must be "
+ 10 counts.
d)Push FUNC pushbutton, the instrument will go to the next
calibration group.
4) "ñ.I.ñA" - MAIN INPUT CALIBRATION - mA INPUT
The upper display will show "ñ.I.ñA".
4.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 10.
_
3
+
1
Fig.10 c) The upper display will show "ñ.I.ñA", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 mA. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
4.2) "Hr" - FINAL SCALE VALUE
The upper display will show "ñ.I.ñA", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 20.000 mA. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
4.3) "
∀." - mA INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "ñ.I.ñA" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.000 mA, b)Check the linearity by setting the calibrator to 10.000 mA The
+ 10 counts.
0000" the read-out must be equal to " read-out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
+ 10 counts.
input value
20 (mA)
∀∀
∀. 3
∀∀
d)Set calibrator to 0.000 V. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument will go to the next step.
5.2) "Hr" - FINAL SCALE VALUE
The upper display will show "ñ.I.5", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 5.000 V. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
5.3) "
∀." - 5 V INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "ñ.I.5" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.0000 V, the b)Check the linearity by setting the calibrator to 2.500 V The read-
+ 10 counts.
0000" read-out must be equal to " out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
6) "ñ.I.10" - MAIN INPUT CALIBRATION - 10 V INPUT
The upper display will show "ñ.I.10".
6.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 5.
c) The upper display will show "ñ.I.10", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 V. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
6.2) "Hr" - FINAL SCALE VALUE
The upper display will show "ñ.I.10", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 10.000 V. b) Push pushbutton, the middle display will change to "On". c) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 3
∀∀
+ 10 counts.
input value
5 (V)
30000
5) "ñ.I.5" - MAIN INPUT CALIBRATION - 5 V INPUT
The upper display will show "ñ.I.5".
5.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 5.
c) The upper display will show "ñ.I.5", the lower display will show
"Lr" while "OFF" will appear on the middle display.
∀.∀.
6.3) "
∀." - 10 V INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "ñ.I.10" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.000 V, the b)Check the linearity by setting the calibrator to 5.000 V The read-
+ 10 counts.
0000" read-out must be equal to " out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and
∀∀
∀. 1 5000"
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
∀∀
∀." followed by the measured
∀∀
+ 10 counts.
5
∀∀
∀. 3
∀∀
the instrument read-out (in counts).
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
input value
10 (V)
30000
Calibration procedure for MKP - PKP
_
6
+
5
7) "A.I.ñA" - AUXILIARY INPUT CALIBRATION - mA INPUT
The upper display will show "A.I.ñA".
7.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 11.
_
6
+
5
Fig.11 c) The upper display will show "A.I.ñA", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 mA. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
7.2) "Hr" - FINAL SCALE VALUE
The upper display will show "A.I.ñA", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 20.000 mA. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
7.3) "
∀." - mA AUXILIARY INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "A.I.ñA" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.000 mA, b)Check the linearity by setting the calibrator to 10.000 mA The
+ 10 counts.
0000" the read-out must be equal to " read-out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
+ 10 counts.
input value
20 (mA)
30000
∀∀
∀. 3
∀∀
Fig.12 c) The upper display will show "A.I.5", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 V. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
8.2) "Hr" - FINAL SCALE VALUE
The upper display will show "A.I.5", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 5.000 V. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
8.3) "
∀." - 5 V AUXILIARY INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "A.I.5" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.0000 V, the b)Check the linearity by setting the calibrator to 2.500 V The read-
+ 10 counts.
0000" read-out must be equal to " out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
9) "A.I.10" - AUXILIARY INPUT CALIBRATION - 10 V INPUT
The upper display will show "A.I.10".
9.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 12.
c) The upper display will show "A.I.10", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 V. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 3
∀∀
+ 10 counts.
input value
5 (V)
30000
8) "A.I.5" - AUXILIARY INPUT CALIBRATION - 5 V INPUT
The upper display will show "A.I.5".
8.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 12.
9.2) "Hr" - FINAL SCALE VALUE
The upper display will show "A.I.10", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 10.000 V. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
6
∀.∀.
9.3) "
∀." - 10 V AUXILIARY INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "A.I.10" "Hr" calibration is correct if the indication is equal to "
0000"
a)Check the zero calibration, by setting the calibrator to 0.000 V, the b)Check the linearity by setting the calibrator to 5.000 V The read-
+ 10 counts. read-out must be equal to " out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 1 5000"
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
10) "In.Ct" - CURRENT TRANSFORMER INPUT CALIBRATION
The upper display will show "In.Ct".
10.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 13.
14
15
Fig.13
Generator
∀∀
∀." followed by the measured
∀∀
+ 10 counts.
input value
10 (V)
mA AC
30000
∀∀
∀. 3
∀∀
Calibration procedure for MKP - PKP
11) "FEEd" - FEEDBACK INPUT CALIBRATION
The upper display will show "FEEd".
11.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Two resistors (1 KΩ, 1/4 W, 1%) must be connected to the
instrument under test as shown in Fig. 14.
12
13
14
Fig.14 c) The upper display will show "FEEd", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Made a short circuit between terminal 13 and 14. e)Push push-button, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument
will go to the next step.
11.2) "Hr" - FINAL SCALE VALUE
c) The upper display will show "FEEd", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Remove the short circuit between terminal 13 and 14. Made a
short circuit between terminal 12 and 13. b) Push push-button, the middle display will change to "ON". d)After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument
will go to the next step.
1 K
1 K
c) The upper display will show "In.Ct", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.00 mA AC. e)Push pushbutton, the middle display will change to "On". f) After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
10.2) "Hr" - FINAL SCALE VALUE
The upper display will show "In.Ct", the lower display will show
"Hr" while "OFF" will appear on the middle display. a)Set the calibrator to 50.00 mA AC. b) Push pushbutton, the middle display will change to "On". d)After a few seconds, start the calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
10.3) "
∀." - CURRENT TRANSFORMER INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "In.Ct" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by setting the calibrator to 0.00 mA AC, b)Check the linearity by setting the calibrator to 25.00 mA AC. The
+ 10 counts.
1000" the read-out must be equal to " read-out must be "
NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal input and the instrument read-out (in counts).
∀∀
∀. 0 0500"
∀∀
Instrument readout (in counts) =
c) Push FUNC pushbutton, the instrument will go to the next
calibration group.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
+ 10 counts.
input value
50 (mA AC)
∀∀
∀. 0
∀∀
1000
∀.∀.
11.3) "
∀." - FEEDBACK INPUT CHECK
∀.∀.
The lower and the middle display show " value in counts (see fig 6).
The "FEEd" "Hr" calibration is correct if the indication is equal to " a)Check the zero calibration, by removing the short circuit between
b)Push FUNC push-button, the instrument will go to the next
12) "O5.ñA" - OUT 5 CALIBRATION.
12.1) "Lr" - INITIAL SCALE VALUE
+ _
+ 10 counts.
1000" terminal 12 and 13 and making a short circuit between terminal 13
and 14; the read-out must be equal to " calibration group.
The upper display will show "O5.ñA".
a)Connect the instrument under test to the calibrator as shown in
Fig. 15.
16 17
Fig. 15 b)The upper display will show "O5.ñA", the lower display will show
"Lr" while the middle display will show the actual zero offset in counts (a number from 0 to 5000).
c) Using or push-button, adjust the instrument output until
0.000 mA
d)After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument will go to the next step.
+0.005 mA is shown by the calibrator.
∀∀
∀." followed by the measured
∀∀
∀∀
∀. 0 0000"
∀∀
+ 10 counts.
∀∀
∀. 0
∀∀
7
Calibration procedure for MKP - PKP
12.2) "Hr" - FINAL SCALE VALUE
The upper display will show "O5.ñA", the lower display will show
"Hr" while the middle display will show the actual full scale offset
in counts (a number from 0 to 5000). a)Using or push-button, adjust the instrument output until
20.000 mA
b)After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument
will go to the next step.
∀.∀.
12.3) "
∀." - OUT 5 CHECK
∀.∀.
The upper display will show "O5.ñA", the lower display will show " while the middle display will show a read-out value in counts (a number from 0 to 8000). a)The instrument shows a value equal to 4000 count. The calibrator
measure must be equal to 10.000 mA NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal output and the instrument read-out (in counts).
OUT =
b)Using or push-button, set a value equal to 0 count. The
"O5.ñA" "Lr" calibration is correct if the calibrator measure an output equal to 0.000 mA
c) Using or push-button, set a value equal to 8000 count. The
"O5.ñA" "Hr" calibration is correct if the calibrator measure an output equal to 20.000 mA
d)Push FUNC push-button, the instrument will go to the next
calibration group.
+0.005 mA is shown by the calibrator.
+ 0.005 mA
displayed value
8000
20
+ 0.005 mA
+ 0.005 mA
∀∀
∀."
∀∀
∀.∀.
12.3) "
∀." - OUT 5 CHECK
∀.∀.
The upper display will show "O6.ñA", the lower display will show " while the middle display will show a read-out value in counts (a number from 0 to 8000). a)The instrument shows a value equal to 4000 count. The calibrator
measure must be equal to 10.000 mA NOTE: when it is desired to use a different check point, the following formula describes the ratio between the signal output and the instrument read-out (in counts).
OUT =
b)Using or push-button, set a value equal to 0 count. The
"O6.ñA" "Lr" calibration is correct if the calibrator measure an output equal to 0.000 mA
c) Using or push-button, set a value equal to 8000 count. The
"O6.ñA" "Hr" calibration is correct if the calibrator measure an output equal to 20.000 mA
d)Push FUNC push-button, the instrument will go to the next
calibration group.
12) "CAL" - DEFAULT CALIBRATION PARAMETER LOADING.
A complete and consistent set of calibration parameters is memorized in the instrument. These data are theoretical data and are used only to clear all calibration memory but after a default calibration data loading it is necessary to make all calibrations.
displayed value
8000
The upper display will show "CAL". The lower display will show "dFLt.". The middle display will show "OFF".
+ 0.005 mA
+ 0.005 mA
+ 0.005 mA
20
∀∀
∀."
∀∀
13) "O6.ñA" - OUT 6 CALIBRATION.
The upper display will show "O6.ñA".
13.1) "Lr" - INITIAL SCALE VALUE
a)Connect the instrument under test to the calibrator as shown in
Fig. 15.
18
+ _
19
Fig. 15 b)The upper display will show "O6.ñA", the lower display will show
"Lr" while the middle display will show the actual zero offset in
counts (a number from 0 to 5000). c) Using or push-button, adjust the instrument output until
0.000 mA
d)After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument
will go to the next step.
12.2) "Hr" - FINAL SCALE VALUE
The upper display will show "O6.ñA", the lower display will show
"Hr" while the middle display will show the actual full scale offset
in counts (a number from 0 to 5000). a)Using or push-buttons, adjust the instrument output until
20.000 mA
b)After a few seconds, start the calibration by pushing FUNC
push-button. At the end of this calibration routine, the instrument
will go to the next step.
+0.005 mA is shown by the calibrator.
+0.005 mA is shown by the calibrator.
When you desire to clear all calibration memory proceed as follows: a) Select the default calibration parameter loading b) Push push-button, the middle display will change to "On". c) Push the "FUNC". The central display will show
Then the upper display will show CAL while the lower display will show the firmware version.
The default calibration parameter loading procedure is ended.
WARNING: After default calibration data loading, it is necessary to remake all instrument calibrations.
8
Calibration procedure for MKP - PKP
9
Loading...