Stryker Spirit Plus, Spirit Select User guide

Page 1
Please read carefully before using your new bed
TM
User Manual for Spirit™ Beds
Spirit Plus™ • Spirit Select™
1-70-001 – R
Page 2
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 2 of 159
Contact Information
Technical Service
Phone:
(519) 963-4010
Toll Free:
1 (866) 516-5446
Fax:
(519) 963-4013
Mailing Address
CHG Hospital Beds Inc. 1020 Adelaide Street South London, Ontario Canada N6E 1R6
Spirit™ Service Call Information
Your facility’s phone number:
Your facility's fax number:
Bed(s) Model and Type:
Bed(s) Serial Number
Date of Purchase:
List of defective part(s):
List of deficiencies or type of problem:
To ensure prompt and accurate attention to your concerns please have the following information available before you call:
Page 3
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 3 of 159
Spirit Plus™ Bed – Quick Reference Guide
To Operate the Bed:
Use the footboard staff control, rail controls, or the pendant to raise or lower the head section, the knee section, or the entire bed.
To Operate Electronic CPR:
PRESS and HOLD the CPR button to flatten both the head and knee sections simultaneously.
To Operate Trendelenburg:
PRESS and HOLD the TREND (Trendelenburg) or rev.TREND (Reverse Trendelenburg) button on the
footboard control to achieve either function. To level the bed, PRESS and hold the opposite function until the bed automatically stops in the horizontal position.
To Operate ‘Chair’:
PRESS and HOLD the CHAIR button. Green LED illuminates when full CHAIR position is achieved. PRESS and HOLD CHAIR button again to level bed. If Green LED is not illuminated PRESS and HOLD
the CPR button to flatten both the head and knee sections and PRESS and HOLD the TREND button until the bed automatically stops in the horizontal position.
To Operate ‘Auto Contour’:
Press the CONTOUR button so that the green light is on. Knee section will raise automatically when head section is raised. (NOTE: KNEE lock-out will override Auto Contour).
Lock-Outs:
The side rail controls and pendant can be locked or unlocked (to restrict patient use). Lock-outs for each bed function are located on the footboard staff control underneath each function. When the LOCK button LED is illuminated, patient control of that particular function with the side rail controls or pendant is restricted. If the LOCK button LED is not illuminated, then patient control of that function is permitted.
Master Lock-Out:
To restrict all side rail, pendant,
and
footboard controls, press all three LOCK buttons simultaneously (the LOCK button LED’s will sequentially flash). When Master Lock-Out has been activated, the Electronic CPR function and Nurse Call remains operable.
To Operate Plastic Siderails:
To lower the rails, gently depress the green PRESS button, then rotate the rail down. Head rails rotate toward the headboard, foot rails rotate toward the footboard. The rail will automatically lock in the UP position.
To Move and Lock the Bed:
Use the Central “Lock & Steer” system:
BRAKE: All four wheels are locked in “BRAKE” mode STEER: Foot-end wheels are locked in “STEER” mode, head-end wheels swivel freely NEUTRAL: All four wheels swivel freely in “NEUTRAL” mode
Power Status Indication:
LED is solid GREEN:
Bed is plugged into an AC power outlet and the battery is fully charged
LED is flashing GREEN:
Bed is plugged into an AC power outlet and the battery is charging
LED is solid RED:
Bed is running on battery power
LED is flashing RED:
Bed is running on battery power but has used up the battery power. Plug bed into an AC power outlet and allow battery to charge for 24 hr
LED is alternating flashing GREEN and RED:
Bed is plugged into an AC power outlet but there is a problem with the battery or another component. Service bed immediately
Page 4
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 4 of 159
F A D
C
B
E
Page 5
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 5 of 159
Page 6
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 6 of 159
A Message from CHG Hospital Beds
Steve Walker Vice President of Operations CHG Hospital Beds
Kyle Sobko Vice President of Sales & Marketing CHG Hospital Beds
To Our Valued Customers,
Thank you for selecting CHG Hospital Beds as your hospital bed provider and congratulations on your purchase
of a Spirit™ bed. You are a part of the growing majority of healthcare providers that recognize the benefits of
a low hospital bed.
CHG Hospital Beds is committed to meeting the needs of our customers through quality products that are innovative, user-friendly, easy to maintain, and most importantly, focused on patient safety. We are sure that
your hospital will enjoy the benefits and safety features of the Spirit™ bed for many years to come.
The Spirit™ bed is manufactured in London, Ontario, Canada. Each and every Spirit™ bed is inspected and tested by our team of highly trained and dedicated quality technicians to ensure that all Spirit™ beds meet our
strict quality standards before entering your hospital.
Our sales team is focused on providing exceptional sales support before, during, and after the delivery of your
Spirit™ bed. If at any time you have questions or concerns regarding your Spirit™ bed, we strongly
encourage you to call us at 1-866-516-5446. Our hands-on company philosophy enables us to respond to our customers quickly and effectively.
We recommend that you carefully read this Spirit™ Bed Owner’s Manual prior to operating your bed.
Becoming acquainted with the safe operation and maintenance procedures outlined in this manual will ensure proper bed use and can extend the service life of the bed.
Sincerely,
Page 7
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 7 of 159
Spirit™ Bed User Instruction Manual Content
Spirit Plus™ Bed – Quick Reference Guide ............................................................................................. 3
Spirit Select™ Bed – Quick Reference Guide .......................................................................................... 4
A Message from CHG Hospital Beds ....................................................................................................... 6
Section 1: ...................................................................................................................................... 11
Specifications & Precautions ......................................................................................................... 11
1.1 Full Bed Warranty ....................................................................................................................... 12
1.2 Intended Use .............................................................................................................................. 13
1.3 Standard Conventions Used in this Manual .................................................................................... 13
1.3.1 Patient Left & Patient Right Determination .......................................................................... 13
1.4 Symbols Used on the Spirit™ Bed ................................................................................................. 14
1.5 Additional Symbols Used on the Spirit Select™ Bed ........................................................................ 15
1.6 Technical Specifications ............................................................................................................... 16
1.7 Important Electrical Precautions ................................................................................................... 18
1.7.1 Battery Back-up ............................................................................................................... 20
1.7.2 Grounding ....................................................................................................................... 22
1.8 Important Mechanical Precautions ................................................................................................ 22
1.9 Mattress Specifications ................................................................................................................ 23
1.10 General Precautions .................................................................................................................... 24
1.11 Standard & Optional Features ...................................................................................................... 27
1.12 Siderail Assemblies ...................................................................................................................... 28
1.13 Optional Accessories.................................................................................................................... 28
1.14 Mass of Detachable Components .................................................................................................. 29
1.15 Applied Parts .............................................................................................................................. 29
1.16 Storage & Transport .................................................................................................................... 29
1.17 Ultra-Low Feature ....................................................................................................................... 30
Section 2: ...................................................................................................................................... 31
Set-Up Instructions ....................................................................................................................... 31
2.1 Unpacking Instructions ................................................................................................................ 32
2.1.1 Delivery ........................................................................................................................... 32
2.1.2 Lowering the Bed ............................................................................................................. 33
2.2 Set-Up of the Spirit™ Bed ............................................................................................................ 34
Section 3: ...................................................................................................................................... 39
Bed Operation ............................................................................................................................... 39
3.1 Central “Lock & Steer” System ..................................................................................................... 40
3.1.1 Bed Mobilization & Stabilization ......................................................................................... 41
3.2 Bed Control ................................................................................................................................ 42
3.2.1 Patient Handheld Control (Pendant) ................................................................................... 42
3.3 Bed Positioning ........................................................................................................................... 43
3.3.1 Bed Elevation – HI-LO Operation ....................................................................................... 43
3.3.2 Head Deck (Back Rest) Elevation - Head Actuator Operation ................................................ 43
3.3.3 Knee Deck Elevation - Foot Actuator Operation ................................................................... 44
3.3.4 Auto Contour Mode .......................................................................................................... 44
3.3.5 Chair Positioning .............................................................................................................. 45
3.4 Emergency & Staff Functions ....................................................................................................... 45
3.4.1 Electronic CPR (Cardiopulmonary Resuscitation) Function .................................................... 45
3.4.2 Optional Manual CPR (Cardiopulmonary Resuscitation) Release ............................................ 46
3.4.3 Trendelenburg Operation .................................................................................................. 49
Page
Page 8
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 8 of 159
3.4.4 Reverse Trendelenburg Operation ...................................................................................... 49
3.5 Patient Lock-Out Functions .......................................................................................................... 50
3.5.1 Regular Patient Lock-Outs ................................................................................................. 50
3.5.2 Master Lock-Out ............................................................................................................... 50
3.5.3 Security Lock-Out ............................................................................................................. 51
3.6 Other Bed Functions & Features ................................................................................................... 54
3.6.1 Nurse Call ........................................................................................................................ 54
3.6.2 Standard 37 Pin Connector ................................................................................................ 54
3.6.3 Spirit Plus™ Power Status Indicator ................................................................................... 55
3.6.4 Spirit Select™ Power Status Indicator ................................................................................. 56
3.7 Siderail Operation ....................................................................................................................... 57
3.7.1 Intended Application/Use of Spirit™ Bed Siderail Assemblies ................................................ 57
3.7.2 Plastic Regular Height Siderails .......................................................................................... 58
3.7.3 Plastic High Siderails ......................................................................................................... 59
3.7.4 Integrated Features of Plastic Regular Height and High Siderails .......................................... 61
3.7.5 Other Optional Siderails .................................................................................................... 62
3.7.6 Optional Electronic Foot Siderail Release ............................................................................ 62
3.8 Optional Underbed Light on Spirit Plus™ Beds ............................................................................... 63
3.9 Optional Underbed Obstruction Sensors ........................................................................................ 64
3.10 Optional “Easy Bed” Extension System .......................................................................................... 66
3.11 Optional Auxiliary AC Power Outlet ............................................................................................... 68
3.12 Optional Pillow Speaker Ready ..................................................................................................... 69
Section 4: ...................................................................................................................................... 71
Spirit Select™ - Unique Functions & Features ............................................................................... 71
4.1 LCD Display ................................................................................................................................ 72
4.2 Patient Weigh Scale .................................................................................................................... 74
4.3 Patient Weight Scale OIML BEDS ONLY ......................................................................................... 76
4.5.1 Calibration OIML BEDS ONLY ............................................................................................ 78
4.5.2 Scale Accuracy Verification ................................................................................................ 78
4.4 Level All Function ........................................................................................................................ 79
4.5 Bed Exit Feature ......................................................................................................................... 79
4.5.1 Bed Exit Alarm: Auto Reset ............................................................................................... 83
4.5.2 Bed Exit Alarm: Auto Re-Arm ............................................................................................ 84
4.5.3 Bed Exit Alarm: Auto Arm ................................................................................................. 85
4.5.4 Bed Exit Light .................................................................................................................. 86
4.5.5 Bed Exit Alarm History ...................................................................................................... 86
4.6 Watchdog™ Bed Monitoring System ............................................................................................. 87
4.6.1 Always On™ Monitoring .................................................................................................... 87
4.6.2 Bed Position/Status Monitoring .......................................................................................... 88
4.6.3 Nurse Reminder™ ............................................................................................................ 89
4.6.4 Available Light Configurations ............................................................................................ 91
Menu Navigation ......................................................................................................................... 92
4.7 Scale Menu ................................................................................................................................. 94
4.7.1 Selecting Patient Weight Display Units ............................................................................... 94
4.7.2 Changing Equipment ........................................................................................................ 94
4.7.3 Adjusting Patient Weight ................................................................................................... 95
4.7.4 View Weight Change ........................................................................................................ 95
4.7.5 Weight Change Alarm ....................................................................................................... 96
4.7.6 Reset All .......................................................................................................................... 97
4.8 Service Menu .............................................................................................................................. 98
4.8.1 Locking LCD Display Units ................................................................................................. 98
Page 9
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 9 of 159
4.8.2 Priority & Nurse Call Settings ............................................................................................. 98
4.8.3 Obstruction Sensor Settings .............................................................................................. 99
4.8.4 Head Angle Settings ........................................................................................................ 100
4.8.5 Language Settings ........................................................................................................... 101
4.8.6 Control Box Versions ....................................................................................................... 101
4.8.7 Configuration List/Code .................................................................................................... 101
4.9 Maintenance Menu ..................................................................................................................... 102
4.9.1 Inputting Access Code ..................................................................................................... 102
4.9.2 Disabling Manual CPR Release Items ................................................................................. 103
4.9.3 Advanced Configurations .................................................................................................. 103
4.9.4 Diagnostic Menu.............................................................................................................. 103
4.10 Calibration ................................................................................................................................. 104
4.10.1 Angle Calibration ............................................................................................................. 104
4.10.2 Scale Calibration ............................................................................................................. 106
4.11 Exit All Menus ............................................................................................................................ 110
Section 5: .................................................................................................................................... 111
Care & Maintenance .................................................................................................................... 111
5.1 Regular Maintenance & Cleaning ................................................................................................. 112
5.1.1 Cleaning & Disinfection ..................................................................................................... 112
5.1.2 Semi Annual Inspections ................................................................................................... 113
5.1.3 Yearly Inspections & Maintenance ...................................................................................... 113
5.1.4 Specific Maintenance Requirements for Spirit Select™ Beds .................................................. 115
5.2 Controls Servicing ...................................................................................................................... 115
5.2.1 Soft Reset of Controls ...................................................................................................... 115
5.2.2 Hard Reset of Controls ..................................................................................................... 116
5.2.3 Updating Firmware on the CB10 Control Box ..................................................................... 116
5.2.4 Updating Firmware on the CB11 Control Box ..................................................................... 117
5.2.5 Obtaining Updated Firmware ............................................................................................ 118
5.3 Mattress Deck Removal .............................................................................................................. 119
5.3.1 Mattress Deck Removal - Foot Section............................................................................... 119
5.3.2 Accessing the CB10/CB11 Control Box ............................................................................... 119
5.3.3 Mattress Deck Removal - Knee Section .............................................................................. 119
5.3.4 Mattress Deck Removal - Head Section (Back Rest) ................................................................... 119
5.3.5 Mattress Deck Removal - Seat Section ................................................................................ 120
5.4 Before Calling The Factory .......................................................................................................... 121
5.5 Colour Information ..................................................................................................................... 121
5.6 Troubleshooting Guide ............................................................................................................... 122
5.7 Troubleshooting Guide for Unique Features of Spirit Select™ Bed ..................................................... 125
5.8 CB10/CB11 Warnings ................................................................................................................. 126
5.9 Schematics & Technical .............................................................................................................. 127
5.9.1 Spirit Wiring Schematic .................................................................................................... 127
5.9.2 Spirit Select™ with OIML Scale Wiring Schematic ................................................................ 129
5.9.3 Electrical Connections – Overview ...................................................................................... 131
5.9.4 Spirit Plus™ CB10Lite Control Box – Overview .................................................................... 132
5.9.5 Spirit Select™ CB10Full Control Box – Overview ................................................................. 133
5.9.6 Spirit Select™ CB11 Control Box – Overview ...................................................................... 134
5.10 Service Parts ............................................................................................................................. 135
Page 10
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 10 of 159
Page 11
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 11 of 159
Section 1:
Specifications & Precautions
Page 12
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 12 of 159
1.1 Full Bed Warranty
CHG Hospital Beds warrants to the original purchaser of its products that the electrical components are guaranteed to be free from manufacturing defect for one (1) year from the date of purchase.
CHG Hospital Beds warrants to the original purchaser of its products that the product or any part thereof, in the judgment of CHG Hospital Beds, will not be defective in material or workmanship for one (1) year from date of original purchase.
CHG Hospital Beds specifically warrants to the original purchaser of its products that the reinforced fibreglass sections of the mattress deck will not be defective in material or workmanship for ten (10) years from date of original purchase.
2nd year
frame parts only
list price less 60%
3rd year
frame parts only
list price less 40%
4th year
frame parts only
list price less 20%
5th year
frame parts only
list price less 10%
Any alteration made to and/or performed on the bed and/or one of its components voids this warranty.
CHG Hospital Beds’ pro-rated warranty covers manufacturing defect, or failure due to workmanship or materials.
Such defects will be repaired or replaced (at the company’s option) free of charge for parts only, provided the defective product is shipped to the factory or an authorized depot at the customer’s own expense. Shipping charges for return of the product to the customer will also be at the customer’s own expense.
This warranty does not include any labour or shipping charges incurred in replacement part installation or
repair of any such part. CHG Hospital Beds’ sole obligation and your exclusive remedy under this warranty
shall be limited to such repair and/or replacement.
After one year, for the next four years, frame parts only will be covered by the warranty according to the following five (5) year prorated fee schedule:
Parts, in the judgment of CHG Hospital Beds, found to be damaged through normal wear and tear, impact, abuse, overloading, fire, lightening, water damage, power surges, brown-outs, improper supply voltage or use of other manufacturer’s head and/or footboards and/or accessories will not be covered. Replacement parts will be charged out at the full list price.
A Return Goods Authorization number (RGA) must be obtained before returning any products or parts to the factory. We regret that no products will be accepted at the factory without an RGA number. No other warranty expressed or implied by CHG Hospital Beds or its agents shall be considered valid.
CHG Hospital Beds Inc. 1020 Adelaide Street South London, Ontario Canada N6E 1R6
Phone: (519) 963-4010 Fax: (519) 963-4013 Toll Free: (866) 516-5446
Page 13
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 13 of 159
1.2 Intended Use
Application Environment
Use of
Spirit™ Bed
No.
Description
1
Intensive/critical care provided in a hospital where 24 hour medical supervision and constant monitoring is required and the provision of life support system/equipment used in medical procedures is essential to maintain or improve the vital functions of the patient.
Suitable
2
Acute care provided in a hospital or other medical facility where medical supervision and monitoring is required and medical equipment used in medical procedures is often provided to help maintain or improve the condition of the patient.
Suitable
3
Long-term care in a medical area with medical supervision is required and monitoring is provided if necessary and medical equipment used in medical procedures may be provided to help maintain or improve the condition of the patient.
Suitable
4
Care provided in a domestic area where medical equipment is used to alleviate or compensate for an injury, disability, or disease.
Not Suitable
5
Outpatient (ambulatory) care which is provided in a hospital or other medical facility, under medical supervision where medical equipment is provided for the need of persons with illness, injury, or disability for treatment, diagnosis, or monitoring.
Suitable
Environmental Conditions for Normal Use
Ambient Temperature Range
Ambient Humidity Range
(Non Condensing)
Atmospheric Pressure Range
DANGER
Information appearing under the DANGER caption concerns the protection of patient, staff and others from the immediate and imminent hazards that, if not avoided, will result in immediate, serious personal injury or loss of life in addition to equipment damage.
WARNING
Information appearing under the WARNING caption concerns the protection of patient, staff, and others from potential hazards that can result in personal injury or loss of life in addition to equipment damage.
CAUTION
Information appearing under the CAUTION caption concerns the protection of patient, staff and others from potential hazards that can result in minor personal injury or equipment damage.
NOTE:
Information appearing in a NOTE caption provides additional information which is helpful in understanding the item being explained.
The Spirit™ bed is intended for low to moderate acuity patients in the medical and/or surgical area of the hospital. The Spirit™ bed is also intended for use as a general-purpose variable height hospital bed for general care, post-operative and general medicine wards. The Spirit™ bed is intended for used in Application Environments 1-3, 5 as described in IEC 60601-2-52, Clauses 201.3.201 - 205.
1.3 Standard Conventions Used in this Manual
This manual includes information essential to the safety of the patient, staff, and equipment during the normal operation of the Spirit™ bed. Before operating the Spirit™ bed be sure you have read and understood the contents of this manual. It is important that you use this equipment in accordance with the procedures outlined in this manual. As you read through this manual be alert to the four signal words.
1.3.1 Patient Left & Patient Right Determination
CHG Hospital Beds’ determination of the “Patient Left” and the “Patient Right” side of the bed is made from the patient’s point of view while positioned normally on the bed facing up.
Page 14
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 14 of 159
1.4 Symbols Used on the Spirit™ Bed
PROTECTIVE EARTH GROUND
OR
ELECTRONIC CARDIOPULMONARY RESUSCITATION (CPR) FUNCTION
PRESS AND HOLD TO LOWER ALL SECTIONS OF MATTRESS DECK FLAT TO BED FRAME
OR
CHAIR FUNCTION
PRESS AND HOLD TO BRING PATIENT TO SITTING POSITION
POTENTIAL EQUALIZATION (EQUIPOTENTIAL POINT)
OR
TRENDELENBURG FUNCTION
PRESS AND HOLD TO MOVE THE BED INTO TRENDELENBURG POSITION
OR
CONTOUR FUNCTION
TOGGLE TO INITIATE AUTOMATIC ACTION OF KNEE-FOOT SECTION WHEN ACTION IS INITIATED TO HEAD SECTION OF MATTRESS DECK
SAFE WORKING LOAD
THE MAXIMUM ALLOWABLE LOAD THAT CAN BE PLACE ON THE BED
OR
REVERSE TRENDELENBURG FUNCTION
PRESS AND HOLD TO MOVE THE BED INTO REVERSE TRENDELENBURG POSITION
OR
FUNCTION LOCK-OUT
ACTIVATE TOGGLE SWITCH TO RESTRICT FUNCTION OF PATIENT AND PENDANT
MAXIMUM PATIENT WEIGHT
THE MAXIMUM ALLOWABLE WEIGHT OF PATIENT THAT CAN BE PLACE ON THE BED
INITIATE ACTION TO HEAD SECTION OF MATTRESS DECK
OR
UP CONTROL
ELEVATES THE CORRESPONDING SECTION OF THE MATTRESS DECK
READ OWNER’S MANUAL
BEFORE USE
INITIATE ACTION TO ENTIRE MATTRESS DECK (HI-LO)
OR
DOWN CONTROL
LOWERS THE CORRESPONDING SECTION OF THE MATTRESS DECK
SYMBOL TO DRAW USER ATTENTION TO AN IDENTIFIED DANGER, WARNING, OR CAUTION
INITIATE ACTION TO KNEE­FOOT SECTION OF MATTRESS DECK
ELECTRONIC FOOT SIDERAIL RELEASE
CLASS 1
EQUIPMENT CLASSIFICATION IN TERMS OF PROTECTION AGAINST ELECTRICAL SHOCK
ALTERNATING CURRENT (AC) POWER
POWER STATUS INDICATOR (SPIRIT™ & SPIRIT PLUS™)
IP54
INGRESS PROTECTION CODE (IP RATING)
TYPE B EQUIPMENT
DUTY CYCLE
2 MINUTES ON 18 MINUTES OFF
CATALOGUE NUMBER
INDICATES THE
MANUFACTURER’S
CATALOGUE NUMBER SO THAT THE MEDICAL DEVICE CAN BE IDENTIFIED
SERIAL NUMBER
INDICATES THE MANUFACTURER’S SERIAL NUMBER SO THAT A SPECIFIC MEDICAL DEVICE CAN BE IDENTIFIED
CONSULT INSTRUCTIONS FOR USE
INDICATES THE NEED FOR THE USER TO CONSULT THE INSTRUCTIONS FOR USE
CAUTION
INDICATES THE NEED FOR THE USER TO CONSULT THE INSTRUCTIONS FOR USE FOR IMPORTANT INFORMATION
MANUFACTURER
INDICATES THE MEDICAL DEVICE MANUFACTURER
WEEE SYMBOL
WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT DIRECTIVE
Page 15
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 15 of 159
SCALE ZERO BUTTON
PRESS AND HOLD TO ZERO SCALE
PRESS AT ANY TIME IN ANY
MENU TO “ESCAPE HOME”
TO HOME SCREEN
SET/RESET BUTTON
PRESS TO TURN ON/OFF AND SELECT DESIRED BED EXIT SENSITIVITY LEVEL
LEVEL ALL FUNCTION
PRESS AND HOLD WILL FULLY FLATTEN ALL SECTIONS OF MATTRESS DECK AGAINST BED FRAME AND RETURN BED FRAME TO HORIZONTAL POSITION
DISPLAY ON/ENTER BUTTON
PRESS TO ACTIVATE OR “WAKE UP” HOME SCREENS
PRESS TO MAKE SELECTIONS WITHIN MENUS AND MENU OPTIONS
CHIME BUTTON
PRESS TO SCROLL THROUGH CHIME VOLUME AND NURSE/PRIOIRITY CALL SETTINGS
LIGHT KEY
MENU SCROLL ARROW BUTTONS
PRESS TO SCROLL UP/DOWN THROUGH HOME SCREENS, MENUS, MENU OPTIONS, AND AVAILIABLE SELECTIONS WITHIN MENU OPTIONS
BED EXIT ALARM SENSITIVITY LEVELS
FUNCTION FEATURE
POWER STATUS INDICATOR
BED EXIT ALARM VOLUME AND NURSE/PRIORITY CALL ACTIVIATION SETTINGS
WATCHDOG FEATURE
NOTE: THESE FEATURES MAY NOT BE PRESENT ON ALL SPIRIT SELECT BED CONFIGURATIONS
1.5 Additional Symbols Used on the Spirit Select™ Bed
Page 16
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 16 of 159
1.6 Technical Specifications
Certifications
Certified to: CAN/CSA C22.2 No. 60601-1:2008 & 60601-2-52:2011, UL 60601-1:2003, IEC 60601-2-52/C1:2010 & 60601-2-38/A1:1999, ANSI/AAMI ES 60601-1:2005/A2:2010 Conforms to: IEC 60601-1:2005/C2:2007
Articulation Range
Spirit Plus™ & Select™
Head Section Angle
0° to 60°
Head Section Angle (Beds with High Siderails and/or Manual CPR Release)
0° to 60°
Knee Section Angle
0° to 25°
Foot Section Angle
0° to 10°
Foot Section Angle in Vascular Position - Knee Fully Flat (Beds with Optional Vascular Bail)
15°
Foot Section Angle in Vascular Position - With Knee Articulation (Beds with Optional Vascular Bail)
0° to -4°
Trendelenburg
0° to 14°
Reverse Trendelenburg
0° to -14°
Mattress Deck Height - Lowest Elevation
Ø5” (125 mm) Casters
10 ¾”
273 mm
Mattress Deck Height - Highest Elevation
34”
863 mm
Minimum Deck Height Required to Accommodate Patient Lift Clearance
48” x 5” (1219 mm x 127 mm)
24”
610 mm
Underbed Clearance - Lowest Elevation
1¾”
45 mm
Mattress Deck Height - Lowest Elevation
Ø6” (150 mm) Casters
11¼”
279 mm
Mattress Deck Height - Highest Elevation
35"
889 mm
Minimum Deck Height Required to Accommodate Patient Lift Clearance
48” x 5” (1219 mm x 127 mm)
21½”
546 mm
Underbed Clearance - Lowest Elevation
2¾”
70 mm
Electrical Specifications
120V Models
230V Models
Mode of Operation - Intermittent Operation
Duty: 10% (2 Minutes ON, 18 Minutes OFF)
Mains Input Voltage
120 Volts AC
230 Volts AC
Mains Input Current
4.0 Amps
2.1 Amps
Mains Input Current (Beds with Auxiliary Mains AC Outlet)
9.0 Amps
5.1 Amps
Mains Input Current Frequency
60 Hertz
50 Hertz
Motor (Actuator) Voltage - Safety Extra Low Voltage (SELV)
24 Volts DC
24 Volts DC
Controls Voltage - Safety Extra Low Voltage (SELV)
5-12 Volts DC
5-12 Volts DC
Power Consumption - at Maximum Load (Safe Working Load)
480 Watts
480 Watts
Power Consumption - at Idle
24 Watts
24 Watts
Auxiliary Mains AC Outlet - Maximum Output Power
120 Volts, 5 A (600 Watts)
230 Volts, 3 A (690 Watts)
Page 17
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 17 of 159
Control Locations
Spirit Plus™
Spirit Select™
Footboard Staff Control
Standard Control
Enhanced Control
Siderail Controls (Patient and Staff)
Standard
Standard
Six Function Handheld Pendant
Optional
Optional
Safety Features
Protective Earth Ground
Class I (Equipotential Point)
Electric Shock Protection
Type B
Enclosure Protection
IPX4
Dual Foot Pedal – Lock & Steer Feature
Standard
On-Board Battery Back-up
Standard
Safety & Convenience Features on Footboard Staff Control
PRESS and Hold Electronic CPR Function
Standard
Trendelenburg/ Reverse Trendelenburg Capability
Standard
Chair Positioning
Standard
Auto Contour Mode
Standard - Toggle Button with LED
Security
Patient Lock-Outs - Bed Elevation, Head Section, and Knee/Foot Section (Patient Controls only)
Standard – Individual Toggle Button with LED’s
Master Lock-Out - All Patient and Staff Control Locations (including Footboard Staff Control)
Standard – Requires deactivation of Master Lock-out to operate bed
Security Lock-Out - All Patient and Staff Control Locations (including Footboard Staff Control)
Standard – Requires facility-set user passcode to operate bed
Dimensions & Weight
Spirit Plus™
Spirit Select™
Length (Overall Bed Length)
91¼”
2318 mm
91¼”
2318 mm
Length - Between Head & Footboards
81”
2057 mm
81”
2057 mm
Length - Mattress Deck (in 80” Mattress Position)
79¾”
2026 mm
79¾”
2026 mm
Length - Mattress Deck (in 84” Mattress Position)
83¾”
2127 mm
83¾”
2127 mm
Width - Between Outer Edges of Siderails
40”
1016 mm
40”
1016 mm
Width - Mattress Deck
35”
889 mm
35”
889 mm
Load Capacity (Safe Working Load)
500 lbs
227 kg
500 lbs
227 kg
Weight of Bed (Empty)
520 lbs
235 kg
525 lbs
238 kg
Weight of Bed (Empty with High Siderails, Manual CPR Release, Easy Bed Extension System)
565 lbs
256 kg
Sound Pressure Level (Measured 1m from Device)
Bed Operation - at Maximum Load (Safe Working Load)
< 54 dBA
CB10/CB11 Confirmation/Notification “Chirp”
< 62 dBA
Bed Exit Alarm Volume Settings
Low Setting
Medium Setting
High Setting
54 - 62 dBA
68 - 87 dBA
80 - 101 dBA
*All dimensions are nominal and have approximate manufacturing tolerances of ± ½” (12.7 mm) for length/height and ± 2° for angular dimensions
Page 18
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 18 of 159
1.7 Important Electrical Precautions
AC Power Outlet Requirements
120 VAC models
Plug the three prong power supply cord ONLY into a properly grounded 120 VAC, 60 Hz power outlet that is rated to at least 10 A.
230 VAC models
Plug the three prong power supply cord ONLY into a properly grounded 230 VAC, 50 Hz power outlet that is rated to at least 5 A.
DANGER
Possible ELECTROCUTION Hazard
DO NOT expose the AC power supply cord or other electrical components to water.
DO NOT allow the power supply cord, AC power outlet, auxiliary AC outlet socket (if
equipped), actuators, control boxes or pendant (if equipped) to become wet or submerged during normal bed operation or when performing cleaning.
DO NOT submerge the bed frame or electrical parts. DO NOT allow liquids to enter electrical components. If a liquid is spilled in or around the bed, unplug the bed before cleaning. Clean up the spill and allow the bed and/or the area around the bed to dry thoroughly before using the controls again.
DO NOT open assemblies such as actuators, control boxes, battery, and pendant (if equipped). These parts are not serviceable. Only service technicians, specifically trained to service CHG Hospital Beds Spirit™ beds, should attempt to service bed and/or replace electrical components.
DO NOT use the auxiliary AC outlet socket (if equipped) if the socket receptacles appear to be damaged and/or if the plug (of the device that you are connecting the bed to) does not remain connected when inserted in the outlet socket receptacles.
DANGER
Possible EXPLOSION Hazard
DO NOT use near explosive gases. DO NOT operate this bed in the presence of a flammable anaesthetic mixture with air or with nitrous oxide.
DO NOT use this bed in Active Treatment “Ether Environments”.
DANGER
Possible FIRE Hazard exists when connecting (plugging in) the power supply cord to an AC power outlet if oxygen administering equipment is in use.
Use either a patient nasal mask or half-length tent, ensuring the tent NEVER extends below the mattress deck, when administering oxygen.
DANGER
Possible FIRE Hazard exists if the pendant is not disabled when oxygen administering equipment is in use. If bed is equipped with a pendant, disconnect the pendant from the port at the side of the bed and insert pendant into its holster or attach pendent to bed linen using the included linen clip when not in use.
WARNING
DO NOT operate the bed if any actuator has malfunctioned or has been damaged in any manner.
DO NOT open assemblies such as actuators, control boxes, battery, and pendant (if equipped). These parts are not serviceable.
Page 19
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 19 of 159
WARNING
Possible ELECTRIC SHOCK Hazard exists when servicing bed and/or replacing electrical components.
ALWAYS disconnect bed from the AC power outlet and ensure that the capacitor inside the CB09 control box has been fully discharged before servicing bed and/or replacing any electrical components. After the bed has been disconnected (unplugged) from the AC power outlet, fully discharge the CB09 control box by sequentially depressing the UP and DOWN buttons on one of the siderail controls or on the pendant (depending upon how bed is equipped) at least 20 times.
Only service technicians, specifically trained to service CHG Hospital Beds Spirit™ beds,
should attempt to service bed and/or replace electrical components.
WARNING
All Spirit™ beds are equipped with a power supply cord storage device located at the head end of the bed. ALWAYS keep the power supply cord clear of any moving bed parts or mechanisms. The power supply cord should ALWAYS be stowed whenever moving, transporting, or storing the bed.
DO NOT allow the power supply cord and/or pendant cord to become entangled around other objects. A pinched power supply cord and/or pendant cord can become damaged and could be dangerous. Be aware of the power supply cord and/or pendant cord location especially when moving the bed.
DO NOT roll the bed over the power supply cord or pendant cord. DO NOT operate the bed if the power supply cord and/or pendant cord is damaged. ALWAYS keep power supply cord and/or pendant cord clear of moving equipment. The power supply cord is a replaceable component by service personnel. Contact CHG Hospital Beds’ Customer Service at (519) 963-4010 or toll free 1-866-516-5446 for further assistance.
DO NOT position the bed so that it is difficult to disconnect the power supply cord should the need ever arise. DO NOT use an extension cord.
WARNING
Depending upon the features your Spirit™ bed has, it will be equipped with either a CB10
or CB11 control box. Some versions of CB10 control box are NOT interchangeable and MUST NOT be installed on some beds. Bed may not operate as intended if incorrect version of the CB10 control box is used.
NEVER install a CB10 control box on your Spirit™ bed if it originally came equipped with a
CB11 control box and vice versa. Both control boxes have different wiring terminations on the circuit board.
Some versions of CB09 control boxes are NOT interchangeable. If your Spirit™ bed is not equipped with a CB10 or CB11 control box but contains additional electrical components (such as a DJB Junction Box), then this bed uses an older version of the CB09 control box.
ALWAYS confirm replacement part numbers and/or control box circuit board revisions are the same prior to installation. Contact CHG Hospital Beds’ Customer Service at (519) 963- 4010 or toll free 1-866-516-5446 for further assistance.
CAUTION
Connecting electrical equipment to auxiliary mains AC outlet effectively leads to creating a medical equipment system and can result in a reduced level of safety. When connecting devices, the system shall be evaluated in end use application to ensure compliance to medical device standard 60601-1.
Page 20
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 20 of 159
CAUTION
A safety feature of this product includes protection against overheating caused by excessive or extended periods of operation. Depending on the duration, this includes multiple or repeated adjustments or the use of multiple functions at once.
To ensure trouble free operation, ALWAYS allow a slight pause between multiple adjustments. DO NOT exceed the maximum continuous mode of operation. Refer to page 16 for complete technical specifications.
If thermal protection activation should occur, the bed will not respond to staff commands from any control location and the CB09 control box will need to be service and/or replaced.
WARNING
Always keep the footboard installed and locked when a patient is left unattended. When removing the footboard, if the mating spring-return connector cover on the bed is damaged or does not return to properly to cover the terminals, reinstall and lock the footboard then have the bed serviced. Never touch an exposed terminal and the patient simultaneously.
Battery Location
All Spirit™ beds are equipped with a rechargeable, back-up battery. The battery is located under the knee section of the mattress deck mounted to the bed frame cross member of the patient left side of the bed on all Spirit™ beds. Refer to pages 127 and 129 respectively for battery location.
Battery Charging and Operation Characteristics
The battery is continuously charged when connected to the CB10 or CB11 control box and the bed is plugged into an AC power outlet. When the bed is unplugged from the AC power outlet, the bed is powered by the battery. Refer to pages 34 to ensure proper battery cable connection.
The footboard staff control on Spirit™ beds has an integrated power status indicator. When the ATTN light is flashing, the battery requires inspection. Refer to pages 55 and 56 for power status indication.
The battery is maintenance free, however battery life is not indefinite, and thus semi-annual inspection is required to verify operation. Refer to page 113 for details. In the event the battery is deemed to be faulty, does not hold a charge, will not operate the functions of the bed while the bed is disconnected (unplugged) from the AC power outlet, or if the battery case or mounting hardware has been damaged in any way, replace the battery.
WARNING
After connecting a battery to the CB10 or CB11 control box, the bed MUST be plugged into an AC power outlet to energize the battery circuit and enable bed operation under battery power. This MUST be performed regardless of the battery charge condition.
WARNING
Emergency bed functions cannot be guaranteed under abnormal operating conditions.
ALWAYS be aware of the power status indicator and take appropriate action when the ATTN light is flashing. ALWAYS perform periodic battery inspection.
1.7.1 Battery Back-up
Page 21
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 21 of 159
WARNING
Potential ELECTROCUTION, EXPLOSION, and/or FIRE HAZARD
These potential hazards could occur resulting in injury or loss of life in addition to equipment damage can occur if battery is replaced with components other than those approved for use on the Spirit™ bed by CHG Hospital Beds.
WARNING
EXPLOSION or FIRE HAZARD
DO NOT expose the battery to heat. Keep away from flames and sparks
WARNING
DO NOT connect battery to CB09 control box if your bed is also equipped with a CB10 or
CB11 control box. The battery may overheat causing it to swell, leak acid, and/or explode.
If your bed has a CB10 or CB11 control box, the DC jack in the CB09 control box is no longer configured as a charging circuit for the battery. The DC jack has been modified to provide constant 24VDC power for the CB10 control box.
Connect battery only to CB10 or CB11 control box. If you have to replace your CB09 control box call CHG Hospital Beds’ Customer Service to ensure that you have the correct version of this component at (519) 963-4010 or Toll Free 1-866-516-5446.
WARNING
If a sulphuric odour (smells like a burnt match) is detected, or if any residual leakage is found on or around the area of the battery, ALWAYS replace the battery immediately.
Battery is NOT a serviceable item. DO NOT attempt to open the battery. In the event that the battery enclosure has been compromised, avoid contact with internal components. Internal components are primarily lead oxide and electrolyte (sulphuric acid).
In the event a person is exposed to sulphuric acid, flush contacted area with large amounts of water for at least 15 minutes. Remove contaminated clothing and seek medical attention if necessary. An eye wash station and emergency shower should be readily available. If swallowed, give large amounts of water. DO NOT induce vomiting. Seek medical attention immediately.
Lead-acid batteries can be HAZARDOUS to your health. Short-term exposure – Sulphuric acid may cause irritation of eyes, nose and throat. Prolonged contact may cause severe burns. Long term exposure – repeated contact causes irritation and skin burns. Repeated exposure to mist may cause erosion of teeth, chronic eye irritation and/or chronic inflammation of the nose, throat, and bronchial tubes.
If a sulphuric acid spill is found under the bed, dilute the spill cautiously with five to six volumes of water and gradually neutralize with sodium bicarbonate, soda ash or lime. When exposure level is not known, wear NIOSH approved positive pressure self-contained breathing apparatus. Reference North American Emergency Response Guidebook #154 (or equivalent current documentation). Be sure to wear acid resistant gloves, safety glasses, and acid resistant clothing when cleaning a sulphuric acid spill.
CAUTION
ALWAYS unplug the bed from the AC power outlet before connecting or disconnecting the battery.
Page 22
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 22 of 159
CAUTION
ALWAYS dispose of faulty batteries or batteries that have reached the end of their service life according to local laws and regulations. Lead acid batteries are completely recyclable.
DANGER
Improper connection of the equipment-grounding conductor can result in electrocution. Check with a qualified electrician or service person if you are doubtful that the electrical outlet is properly grounded.
DO NOT modify the three prong plug provided. If it will not fit into the AC power outlet, have a proper AC power outlet installed by a qualified electrician.
Grounding reliability can only be achieved when bed is connected to an equivalent receptacle marked “HOSPITAL ONLY” or “HOSPITAL GRADE”.
WARNING
Safe Working Load of Bed
227 kg (500 lbs)
DO NOT overload the bed. The total combined weight of patients,
visitors, mattress, additional equipment/accessories, and bedding MUST
NOT exceed 227 kg (500 lbs)
WARNING
Maximum Patient Weight
162 kg (357 lbs)
Patient weight is based on a typical bed setup (20 kg mattress + 45 kg
additional accessories)
WARNING
Safe Working Load of Siderails
80 kg (176 lbs)
DO NOT exceed a 80 kg (176 lbs) load on any siderail
WARNING
DO NOT use headboards or footboards from other manufacturers on any Spirit™ bed.
Spirit™ beds are specifically designed and manufactured for use in conjunction with CHG
Hospital Beds accessories. Accessories designed by other manufacturers have not been
tested by CHG Hospital Beds and are not recommended for use on Spirit™ beds.
On Spirit™ beds, ALWAYS ensure that the headboard and footboard brackets are properly
attached to the frame before use.
1.7.2 Grounding
This electric bed must be grounded. In the event of a malfunction or breakdown, grounding provides a path of least resistance for electric current, thereby reducing the risk of electric shock.
This product is equipped with a cord having an equipment-grounding conductor and a grounded plug. The plug must be inserted into an appropriate AC power outlet that is properly installed and grounded in accordance with all local electrical codes and ordinances.
1.8 Important Mechanical Precautions
Page 23
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 23 of 159
WARNING
Possible PINCHING HAZARD exists when bed is configured with manual CPR release. Powered bed mechanisms can cause SERIOUS INJURY.
ONLY trained healthcare practitioners shall operate the manual CPR release in the event of a medical emergency. ONLY operate the Electronic CPR button and/or the manual CPR release with persons clear of mechanisms.
A mattress is not included with the Spirit™ bed. A CHG Hospital Beds mattress is recommended.
Any mattress that is used on the Spirit™ bed must
suitably fit on the mattress deck and fit snugly within the confines of the 4 corner mattress keepers having a width of 35” (889mm).
The length and width of the mattress MUST suitably fit on the mattress deck in either the 80” (2032 mm) or 84” (2134 mm) configuration.
DO NOT use the bed without a mattress having a thickness of at least 6" (152 mm) but not more than 9-1/2” (241 mm).
LENGTH:
OR
80 inch (2032mm)
84 inch (2134 mm)
WIDTH:
35 inch (889 mm)
THICKNESS:
6 inch (152 mm)
WEIGHT:
80 inch (with Bolsters)
24 lbs (10.9 kg)
84 inch (with Bolsters)
28 lbs (12.7 kg)
WARNING
The mattress MUST entirely rest upon the mattress deck. The mattress MUST fit snugly within the 4 corner mattress keepers.
Incompatible mattresses can create hazards. DO NOT use this bed without a special mattress specifically designed to bend and conform to the shape of the bed. DO NOT use water filled or gel filled mattress on this bed.
WARNING
Possible patient ENTRAPMENT Hazard or FALL RISK if using non-specified mattress.
Patient entrapment may result in injury or death. Use only a mattress of recommended specifications with this bed. CHG Hospital Beds will not be responsible for any injury to patient and/or staff and/or damage to bed that may result with use of non-specified mattress.
1.9 Mattress Specifications
Page 24
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 24 of 159
WARNING
CHG Hospital Beds recommends that the customer perform a thorough patient assessment to determine if the bed system and mattress selection is appropriate for the patient on the basis of their clinical needs, fall risk, and mental capacity. CHG Hospital Beds is of the belief that to ensure maximum patient safety, there is simply no substitute for frequent patient monitoring by qualified healthcare practitioners. CHG Hospital Beds also recommends that the customer conduct and document a patient entrapment risk assessment for every mattress combination
that is intended for use on the Spirit™ bed in both the flat (horizontal) and articulated positions
to identify and address any exposure to areas of potential patient entrapment and/or fall risk as part of a comprehensive and proactive bed safety program.
WARNING
A potential increase in FALL RISK may result when using the Spirit™ bed with a mattress having a thickness greater than 6” (152mm). All patients should be monitored closely and CHG Hospital Beds recommends that when a patient is to be left unattended the bed should ALWAYS be lowered to the lowest position.
WARNING
Close supervision by trained healthcare practitioners is ALWAYS necessary when this product is used by or near children and/or people with disabilities.
DO NOT let any person climb/crawl underneath the bed, between the bed legs and/or the raised bed frame components at any time.
WARNING
DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such
as owner’s manuals, service manuals, or instruction sheets supplied with this product or
optional equipment.
If you are unable to understand the warnings, cautions or instructions contact a healthcare
professional, your Spirit™ bed dealer, or a service technician, specifically trained to service CHG Hospital Beds Spirit™ beds, before attempting to install and/or use this equipment,
otherwise injury to the patient and/or staff and equipment damage may occur.
WARNING
DO NOT drop the bed. DO NOT allow patients to fall onto and/or jump on the bed.
These types of activity can cause impact loads that can permanently damage the actuators and/or other bed components resulting in an inoperable bed.
In the case of an inoperable bed due to a damaged actuator, replace the actuator immediately. If left unattended, a damaged actuator could result in injury to the patient and/or staff and additional equipment damage.
WARNING
Check all parts for shipping damage and test bed to confirm proper operation before putting the bed into active service.
DO NOT use bed if any component damage is discovered or a service technician suspects that damage has occurred. A damaged bed may expose staff and/or patients to unforeseen safety
hazards. Contact your Spirit™ bed dealer or a service technician, specifically trained to service CHG Hospital Beds Spirit™ beds, for further instruction.
Before the bed is returned to active service after any adjustment, repair, and/or service have been performed, ALWAYS ensure that all attaching hardware is tightened securely.
1.10 General Precautions
Page 25
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 25 of 159
WARNING
NEVER allow patients to use trapeze or traction units as a total individual weight support.
Traction units are to be only used for immobilizing a patient in various, therapeutic, traction set ups and/or positions that have been clinically prescribed by a trained healthcare practitioner. Trapeze units (lifting poles) are to be only used to assist patient when repositioning and/or transferring into or out of the bed.
WARNING
Unauthorized modifications to the equipment can result in HAZARDS. DO NOT modify the Spirit™ bed and/or any accessories without written authorization from CHG Hospital Beds.
Use only authorized CHG Hospital Beds replacement parts and/or accessories otherwise the warranty is void. CHG Hospital Beds will not be responsible for any injury to patient and/or staff and/or damage to bed that may result.
WARNING
ALWAYS keep all moving parts, including the main frame of the bed, the bed legs, the mattress deck, and all actuator shafts free of obstructions (i.e.: window sills, radiators, bed side cabinets, under bed tables/trays, chair rails, consoles, blankets/bed linens, heating blankets/pads, tubing, wiring, etc., and other types of products using electric cords which may get tangled around the bed, siderails or legs) during bed operation. NEVER store anything under the bed.
ALWAYS take the necessary precautions to avoid squeezing/shearing of routed cables from other equipment in the moving parts of the bed.
CAUTION
DO NOT stand on the bed or concentrate weight on any particular sections of the mattress deck. Patient body weight should be evenly distributed over the surface of the bed. DO NOT lay, sit or lean in such a way that the patient’s entire body weight is placed only on elevated head or foot sections of the bed. This includes situation when assisting the patient to reposition and/or transfer into or out of the bed.
Ensure all hinges of the mattress deck sections are properly aligned before raising head or knee sections. All four sections of the mattress deck can be detached for thorough cleaning, sanitization, and maintenance. Refer to page 119 for instructions.
CAUTION
To ensure the accuracy of the scale and alarms, keep the area around and/or under the bed open and free of obstructions and/or equipment.
CAUTION
A patient hoist can be used with this bed however due to the ultra-low height, it has limited underbed clearance therefore the bed must be partially raised to allow the patient hoist to roll freely under it. Failure to raise the bed may result in damage to the bed when patient hoists are used.
CAUTION
The Footboard is NOT designed for use as a patient therapy device. NEVER use the Footboard as a resistance device. Any undue forces, bending or flexing of the footboard may permanently damage the footboard and any associated electrical connections.
Page 26
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 26 of 159
CAUTION
When bed is configured with manual CPR release, ONLY trained healthcare practitioners shall operate the manual CPR release in the event of a medical emergency. ONLY operate the CPR button and/or manual CPR release with persons clear of mechanisms. PRESS and HOLD the CPR button to lower the head (back rest) and knee sections to a flat position.
To interrupt the CPR function, simply release the CPR button. Refer to device manual for additional safety precautions.
To avoid unintended CPR release keep area around CPR handle clear of obstructions.
CAUTION
Potential PINCHING HAZARD exists between articulating deck sections and bed frame. Moving parts and/or powered bed mechanisms can cause SERIOUS INJURY. Only operate bed with persons clear of all mechanisms. ALWAYS keep hands clear of moving bed parts.
Ensure that the backrest remains free from all obstructions. Obstructions could impair normal bed operation and/or cause component damage and/or injury. If an obstruction is detected elevate the backrest to clear obstruction before attempting to remove any obstruction.
CAUTION
The Spirit™ bed has been supplied from an environmentally aware manufacturer that
complies with the Waste of Electrical and Electronic Equipment (WEEE) directive
2002/96/CE. The Spirit™ bed may contain substances that could be harmful to the
environment if disposed of in places (landfills, etc.) that are not appropriate according to legislation. Please be environmentally responsible and recycle this bed through your recycling facility when the bed has reached the end of its service life.
Page 27
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 27 of 159
1.11 Standard & Optional Features
Feature
Spirit Plus™
Spirit Select™
5” (125 mm) Casters
Standard
Standard
6” (150 mm) Casters
Optional
Optional
4 Corner Mattress Keepers
Standard
Standard
6 Patient Restraint Loops
Standard
Standard
2 Drainage Bag Holders
Standard
Standard
Wood Laminate Head/Footboards
Optional
Optional
Patient Handheld Control (Pendant)
Optional
Optional
Pillow Speaker Ready
-
Optional
Moulded Plastic Head/Footboards
Standard
Standard
Moulded Plastic Siderails c/w Integrated Staff & Patient Controls, and Nurse Call
Standard
Standard
Enhanced Footboard Staff Controls with Integrated Patient Weight Scale Display & Bed Exit Controls
-
Standard
Integrated Patient Weight Scale & Bed Exit System
-
Standard
Manual CPR Release
-
Optional
Bed Exit Alarm Pad & External Alarm Unit
Optional
-
Underbed Obstruction Sensing
Optional
Optional
Underbed Lighting
Optional
Standard
Auxiliary AC Power Outlet Socket
Optional
Optional
4” (102mm) “Easy Bed” Extension System
Optional
Optional
2 Position Vascular Foot Bail
Optional
Optional
Moulded Plastic High Siderails c/w Integrated Staff, Patient Controls, and Nurse Call Feature
Optional
Optional
Mental Health Package c/w Tamperproof Fasteners, Deck Section Connectors, Non-Removable Head and Footboards, Removable Brake Pedal, and Shortened Power Supply Cord
Optional*
Optional*
WARNING
*A Spirit™ bed outfitted with the optional Mental Health package is a customer requested bed configuration that is intended to perform a specialized function and has several unique differences from regular production beds. The head and footboards and deck sections may not be removable without tools. The head and footboards and rail panels may not have any openings. The brake pedal may be removable. The power cord may also be shorter.
* A Spirit™ bed outfitted with the Mental Health package is NOT intended for normal unit use.
Page 28
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 28 of 159
Your Spirit™ bed may be equipped with one of the following types of siderail assemblies:
Siderail Types
Spirit Plus™
Spirit Select™
Regular Height Plastic Siderails
Standard
Standard
High Plastic Siderails
Optional
Optional
High “Filled-in” Plastic Siderails
Optional
Optional
Long High Head Plastic Siderails
Optional
Optional
Long High Head “Filled-in” Plastic Siderails
Optional
Optional
Accessory
Mass*
Spirit Plus™
Spirit Select™
Bed Exit Alarm Pad & External Alarm Unit
n/a**
A2700-PLUS
-
4” (102 mm) Bolt-on Bed Extension Kit
n/a**
A2500
A2500
4” (102 mm) “Easy Bed” Extension Kit
22 lbs (9.9 kg)
A2508
A2508
4” Bolster
2 lbs (0.9 kg)
A2501
A2501
4” Bolster (For “Easy Bed” System)
6 lbs (2.7 kg)
A2503
A2503
2 Position Vascular Foot Bail
n/a**
A2140
A2140
Padded Floor Mat
10 lbs (4.5 kg)
A3321
A3321
Padded Head/Footboard Covers
2 lbs (0.9 kg)
A3201/A3202
A3201/A3203
Padded Regular Height Rail Covers
1 lbs (0.5 kg)
A3210
A3210
Padded High Rail Covers
1 lbs (0.5 kg)
A3215
A3215
CPR Board & Bracket
6 lbs (2.7 kg)
A2101/A2100
A2101/A2100
Mattress Pump & Accessory Hanger
3 lbs (1.4 kg)
A2601
A2601
Removable Universal Trapeze Adapter
13 lbs (5.9 kg)
A2404
A2404
Fixed IV Pole
3 lbs (1.4 kg)
A2320
A2320
Collapsible IV Pole and Brackets
6 lbs (2.7 kg)
A2315
A2315
Horizontal Oxygen Tank Holder
7 lbs (3.2 kg)
A2211
A2211
Vertical Oxygen Tank Holder
2 lbs (0.9 kg)
A2210
A2210
1.12 Siderail Assemblies
1.13 Optional Accessories
*All masses are nominal and have approximate manufacturing tolerances of ± ¼ lbs (± .11 kg) **Accessory, once installed onto and/or fastened to bed is no longer considered detachable
Page 29
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 29 of 159
Detachable Component
Mass*
Headboard
Footboard
Spirit Plus™ Wood Laminate Head/Footboard
21 lbs (9.5 kg)
21 lbs (9.5 kg)
Spirit Plus™ Moulded Plastic Head/Footboard
6 lbs (2.7 kg)
6 lbs (2.7 kg)
Spirit Select™ Wood Laminate Head/Footboard
21 lbs (9.5 kg)
22 lbs (10.0 kg)
Spirit Select™ Moulded Plastic Head/Footboard
6 lbs (2.7 kg)
7 lbs (3.2 kg)
CAUTION
NEVER attempt to install a headboard at the foot end of the bed and vice versa. ALWAYS ensure that the footboard is installed in the proper orientation. The footboard staff control MUST ALWAYS face away from the patient. NEVER drop a footboard. A damaged footboard staff control and/or footboard staff control connector could result in the inability to operate the bed from this control point.
CAUTION
On Spirit™ beds configured with wood laminate head/footboards, the footboard cable MUST be disconnected before the footboard can be removed from the bed. Failure to disconnect the footboard cable prior to footboard removal may result in component damage and/or the inability to operate the bed from the footboard staff control.
CHG Hospital Beds considers that the following are the Applied Parts of the Spirit™ bed: any head/footboard assemblies, head/footboard bracket assemblies, any siderail assemblies, all sections of the mattress support platform (deck) and associated mattress keepers, any CHG Hospital Beds’ supplied pendant and/or pillow speaker (if equipped), any CHG Hospital Beds supplied (Prevention Surface) mattress (if equipped).
Environmental Conditions for Storage and Transport
Ambient Temperature Range
Ambient Humidity Range
(Non Condensing)
Atmospheric Pressure Range
1.14 Mass of Detachable Components
*All masses are nominal and have approximate manufacturing tolerances of ± ¼ lbs (± .11 kg)
1.15 Applied Parts
1.16 Storage & Transport
Page 30
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 30 of 159
During extended periods of disconnect from an AC power outlet, the battery is susceptible to permanent
damage caused by being deep discharged. All Spirit™ beds are equipped with a disconnect feature which
automatically disengages the battery from the battery circuit to prevent damage from being deep discharged. There is no need to disconnect the battery when a Spirit™ bed is going to be put into storage, however, the CB10 or CB11 control box will continue to draw power until the battery voltage has hit the 18V threshold which triggers the automatic circuit disconnection.
IMPORTANT: A battery voltage of 18V is insufficient to operate the bed. Reconnect bed to an AC power outlet as soon as possible to and allow battery to charge for a period of 24hr to ensure that the bed will reliably operate on battery power when needed.
IMPORTANT: To conserve battery life during storage and transport, disconnect the battery from the CB10 or
CB11 control box. On Spirit™ beds equipped with CB10 control boxes, even if a battery was fully charged
when it was unplugged from the control box, the bed will not operate on battery power if the battery is simply reconnected to the control box. The bed must be connected to an AC power outlet to energize the battery circuit. This will enable bed operation on battery power. On Spirit™ beds equipped with CB11 control boxes, the bed will become operational when the battery is reconnected.
CAUTION
DO NOT re-use any original packaging material to transport the bed.
ALWAYS transport the bed by rolling the bed on its casters.
ALWAYS ensure the Central “Lock & Steer” system is in the “BRAKE” position before
attempting to lift/lower bed. ALWAYS ensure that the Central “Lock & Steer” system is in the “BRAKE” position after the bed is loaded onto transportation vehicle. Use additional
strapping or tie downs as necessary to ensure the bed does not move while in transport.
DO NOT use the siderails as a means of mechanically restraining the bed during
transportation and/or storage. DO NOT use the siderails as lifting points for the bed during transportation and/or storage.
NOTE:
CHG Hospital Beds recommends that following any storage and/or transportation in extreme temperature conditions that you allow the bed to acclimatize for a period of at least 24 hours before the bed is operated/powered up and/or put into service.
CAUTION
When beds are configured with no side rails: Keep the bed in the lowest position with the wheels locked and patient lock-outs on as prescribed by the care plan when occupied, adjusting the level for activities such as administering care or for patient transfers in/out of bed. Transport patients as prescribed by the care plan.
1.17 Ultra-Low Feature
The Spirit™ product line is a family of ultra-low beds, designed to reduce patient injury due to falls out of bed. Therefore, in an effort to reduce patient injury, CHG Hospital Beds recommends that you ALWAYS lower the bed completely to the lowest position before leaving a patient unattended.
When configuring a bed with no rails:
Regardless of the purpose for which bed rails are being used or considered, a decision to utilize or remove those in current use should occur within the framework of an individual patient assessment. For Further information refer to the FDA’s website at http:/www.fda.gov/cdrh/beds/ for Clinical Guidance for the Assessment and Implementation of Bed Rails in Hospital, Long Term Care Facilities, and Home Care Settings.
Available floor mats, mattresses with side bolsters, bed exit alarm, and nurse call can be used in conjunction with the low bed based on the patient assessment and care plan.
Page 31
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 31 of 159
Section 2:
Set-Up Instructions
Page 32
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 32 of 159
2.1 Unpacking Instructions
CAUTION
Equipment DAMAGE may result from improper plastic tie wrap removal.
DO NOT cut any trimmed plastic tie wraps. Trimmed plastic tie wraps are permanent features on the Spirit™ bed.
If your Spirit™ bed is shipped standing on-end, use three people to carefully lower it to the floor. Refer to page 33 for safe lowering instructions.
DO NOT attempt to lower the bed without assistance.
Once the bed is lowered, you may begin the set-up of your Spirit™ bed.
If your Spirit™ bed is shipped on a pallet or in a crate, roll the bed onto the floor and begin the set-up procedure.
For the purpose of clarity, the packaging has been removed from this bed.
WARNING
BED IS TOP HEAVY!!!
Bed may TIP OVER when positioned upright during storage and/or transport
WARNING
BED IS HEAVY!!!
Improper handling may result in serious injury to personnel and/or damage to bed.
Use proper lifting techniques when lifting or lowering bed.
DO NOT attempt to lower bed without assistance. Refer to page 33 for safe lowering instructions.
Upon delivery of the bed, remove all cardboard, Styrofoam, crates, or other packaging material. Follow these instructions to unpack and set up the bed.
2.1.1 Delivery
Page 33
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 33 of 159
2.1.2 Lowering the Bed
1. Beds in typical position when being
transported or stored.
IMPORTANT: BED IS HEAVY!!! Never attempt to lower the bed without assistance. CHG Hospital Beds recommends that the bed is most safely lowered with 3 people as shown in the following steps.
2. Use the corner of the shipping stand
(bolted to lower end of the bed) and the siderails (must be strapped to the bed) to pivot the bed into a position that will allow you to lower the bed.
IMPORTANT: Maintain your hold on the bed. The bed will fall if not held upright at this point.
3. Ensure the brake is engaged before
attempting to lower/lift the bed. Attempt to rotate a caster to confirm that the brake has been engaged.
IMPORTANT: Prior to lowering the bed, ensure that you have adequate room. At least 8 feet is required to lower the bed from the upright transportation/storage position.
4. Firmly grasp the bed frame at the foot
end of the bed (circled). Begin lowering the bed by gently pulling the bed from the upright position and slowing walking backwards.
5. Continue lowering the bed by slowing
walking backwards. It may be necessary to reposition your grip to allow further lowering of the bed.
IMPORTANT: Communicate with the other people before adjusting your grip!
6. At this point, maintain your grip, lock
your arms, and lower the bed to the floor by bending your knees.
IMPORTANT: Do not drop bed!
WARNING
DO NOT drop the bed when lowering during storage and/or transport. If the bed is dropped, permanent damage to bed components may occur. This may result in an inoperable bed or a bed with severely impaired operation.
DO NOT use a bed that has been dropped. Non obvious damage may have occurred that may expose staff and/or patients to unforeseen safety hazards.
The Central “Lock & Steer” system MUST be in the “BRAKE” position prior to attempting to lower/lift the bed. If the brake has not been engaged, the bed may roll away from personnel while attempting to lower/lift the bed. This may result in a very rapid descent and/or unexpected bed motion that may cause severe INJURY to personnel lowering/lifting bed and/or permanent DAMAGE to bed.
Before lowering/lifting bed, ALWAYS attempt to roll/rotate a caster to confirm that the brake is engaged.
DO NOT use the siderails as lifting/lowering points for the bed.
DO NOT use the siderails as a means of mechanically restraining the bed during transportation and/or storage.
Page 34
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 34 of 159
2.2 Set-Up of the Spirit™ Bed
1. Cut packaging straps holding the headboard and footboard box and other packaging materials to the bed. Cut all
untrimmed plastic tie wraps. Discard all packaging materials.
DO NOT cut any trimmed plastic tie wraps. Trimmed plastic tie wraps are permanent features on all Spirit™ beds.
1. Using a 3/4 SAE socket, 9”
extension, and ratchet remove the 4 bolts holding the shipping bed stand onto the headboard bracket.
2. Lift shipping bed stand away from
headboard bracket. Discard shipping bed stand and fasteners.
3. Remove the plastic tie wrap
around the main power supply cord bundle.
Exercise caution to ensure that you do not cut into or otherwise damage the of the power supply cord.
4. Connect the bed to mains A/C
power either directly into a properly grounded wall outlet or into a suitable extension cord connected to a properly grounded wall outlet.
5. Cut plastic tie wrap holding head
rail assembly.
6. Rotate both head rail assemblies
and lock/latch in the UP position. Pull out all packaging tucked in between the rail arms, panel, and bracket. Discard all packaging materials.
Page 35
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 35 of 159
7. Reach under bed and cut
packaging strap wrapped around head-end leg assembly and the frame.
8. If your bed is equipped with the
optional obstruction sensors, pull out protective shipping cover installed over the lower leg obstruction sensor. Discard protective shipping cover.
9. Cut packaging strap wrapped
around foot rail assemblies and the frame.
10. Reach under bed and cut
packaging strap wrapped around foot-end leg assembly and the frame.
11. If your bed is equipped with the
optional obstruction sensors, pull out protective shipping cover installed over the lower leg obstruction sensor. Discard protective shipping cover.
12. Press the UP button on either
siderail staff controls and elevate bed to a comfortable working height.
13. Rotate both foot rail assemblies
and lock/latch in the UP position. Pull out all packaging tucked in between the rail arms, panel, and bracket. Discard all packaging materials.
14. On Spirit Select beds there are two shipping pins located behind the foot
rail assemblies.
IMPORTANT: Both of these shipping pins MUST be removed for proper scale function.
15. Removed both shipping pins by pulling straight out. Discard both shipping
pins.
Page 36
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 36 of 159
16. Unpack the headboard and
footboard from their separate packaging box.
17. Lay the headboard and footboard
on top of their separate packaging box.
There should also be a small bag of 4 black plastic barbed clips in the box. Other items may be shipped inside this box. Ensure box is empty then discard box.
18. Place the headboard into the
headboard bracket assembly.
Electrical connector is not live. It is used only as an indexing feature to ensure correct headboard installation orientation.
19. Place the footboard into the
footboard bracket assembly.
Electrical connection is made when the footboard is inserted into bracket assembly; footboard becomes inactive when removed.
20. Install both footboard lanyard
pins into the holes in the footboard bracket.
21. Install 4 black plastic barbed clips
(from step 17) into the 4 holes on the headboard bracket that the shipping stand was attached to.
22. Remove manual from cardboard
coving mattress deck.
IMPORTANT: DO NOT discard manual. Thoroughly read manual before using bed. Store manual in a readily accessible location for future reference.
23. Removed cardboard covering
mattress deck. Discard cardboard.
24. Lift the knee/foot sections of the
mattress deck (as shown above) to gain access the CB10 or CB11 control box.
Page 37
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 37 of 159
25. Continue to lift the knee/foot
hinged connection and gently lay the deck sections back as shown above. The CB10 or CB11 control box is now accessible.
26. Connect the battery cable to the
side DC jack of the CB10 or CB11 control box.
IMPORTANT: This will ensure the on-board battery remains fully charged and will be ready for use in the event that mains AC power is disrupted.
27. If your bed is equipped with the
optional pendant, unscrew cap on desired pendant port (either side of bed). Align arrows and insert pendant connector into pendant port. Slide up and screw on threaded retaining collar attached to the pendant.
The bed MUST be initialized before it can be used.
28. Using the BED DOWN button fully
lower the bed using any control location (footboard staff control, siderail staff control, rail patient, or pendant).
IMPORTANT: Ensure that the power cord is free of moving bed parts.
29. Once the bed is fully lowered and
all bed motion has automatically stopped CONTINUE to HOLD the BED DOWN button until the two tone audible indicator sounds. Continue to Hold bed down for an additional 20 seconds to complete the Soft Reset.
This process will synchronize the hi-lo actuators so that the bed will perform properly
30. Your new Spirit™ bed is now
ready for use.
Page 38
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 38 of 159
Page 39
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 39 of 159
Section 3:
Bed Operation
Page 40
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 40 of 159
3.1 Central “Lock & Steer” System
Mode
Caster Functionality
Pedal Position
“BRAKE”
Mode used to stabilize the bed from shifting.
This mode prevents the bed from moving forwards, backwards or sideways.
Casters DO NOT swivel or roll
“NEUTRAL”
Mode used only to manoeuvre the bed in a tight area.
This mode allows the bed to move forwards, backwards or sideways.
All casters swivel and roll
“STEER”
Mode used when attempting to steer the bed in a desired direction.
All caster wheels can still rotate, enabling the bed to move forwards or backwards.
Head end casters swivel
Foot end casters DO NOT swivel
All casters roll
WARNING
Unintended bed movement may occur if bed is left in either of the two mobilized positions; “STEER” or “NEUTRAL”.
NEVER leave the bed unattended in either the “STEER” or “NEUTRAL” positions.
ALWAYS engage the “BRAKE” when leaving a patient unattended.
DO NOT attempt to move the bed until the “BRAKE” has been released.
WARNING
When transferring into or out of the bed, ALWAYS ensure that the “BRAKE” is engaged (casters are locked). Inspect the caster locks for correct locking action before actual use.
Even with the “BRAKE” properly engaged (caster properly locked), some flooring surfaces
such as tile or wood will allow the bed to move under some conditions. Bed use on surfaces such as these MUST be evaluated by the healthcare facility and deemed safe before the bed is put into active service.
Page 41
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 41 of 159
3.1.1 Bed Mobilization & Stabilization
Bed Mobilization
The bed is mobile when the Central “Lock & Steer” pedal is in either the “NEUTRAL” or “STEER” position. Use either of these two pedal positions depending on the situation, when bed mobility is needed.
Enable “Steer”
Fully depress the right side of the Central
“Lock & Steer” pedal at either end of the
bed. Pedal actuation mechanism should make an audible engagement when switching between modes.
IMPORTANT: Depending
upon the orientation of the castors, it may be necessary to roll the bed sideways, in a back and forth motion, at the foot end of the bed until the steering casters become
engaged in the “STEER”
mode.
Side-to-Side Motion
Put the bed into “Neutral”
Depress or lift the Central “Lock & Steer” pedal with your foot until the pedal is level. Pedal actuation mechanism should make an audible engagement when switching between modes.
The bed can be put in “NEUTRAL” regardless of caster orientation.
Bed Stabilization
The bed is stable when the Central “Lock & Steer” pedal is in the “BRAKE” position. Use this pedal position whenever the bed is left unattended or when the bed needs to remain stable.
Apply the “BRAKE”
Fully depress the left side of the Central “Lock & Steer” pedal at either end of the bed. Pedal actuation mechanism should make an audible engagement when switching between modes.
The “BRAKE” can be applied regardless of caster orientation.
WARNING
DO NOT to move the bed until the siderail assemblies have been fully raised/closed and locked/latched in the UP position Refer to page 57 for siderail operation instructions.
NOTE:
When the “NEUTRAL” mode is activated properly, the bed should move freely without any unusual noises. If any clicking noises are heard when in the “NEUTRAL” position, stop and ensure that the Central “Lock & Steer” pedal is level. Adjust, if necessary.
Page 42
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 42 of 159
3.2 Bed Control
Depending on the Spirit™ bed you purchased, you will have two or three locations for controlling bed functions.
Control Location
Spirit Plus™
Spirit Select™
Patient Handheld Control (Pendant)
Optional
Optional
Footboard Staff Control
Standard
Enhanced
Patient/Staff Siderail Controls
Standard
Standard
All Spirit™ beds can be configured to connect an optional patient, handheld, 6 function control (pendant).
The pendant offers patient control of basic bed operations:
Head UP/DOWN, Knee/Foot UP/DOWN
Bed UP/DOWN
Pendant Storage
The pendant can be affixed to the bed using a holder. The pendant slips into the holder for easy location and convenient patient access.
On Spirit™ beds equipped with high siderails:
Install the pendant holder into the opening in either foot rail panel. To permanently affix the holder to the bed, screws the holder to the foot rail panel with the fasteners provided.
Optionally, the pendant can be secured to the bed linens with the linen clip provided. Choose a clip location so that the pendant is within easy reach by the patient.
Pendant Connection
The pendant can be connected on either side of the bed (the connection port is just beneath the seat section of the mattress deck on both sides of the bed).
To relocate the pendant to the opposite side of the bed, first, unscrew the threaded retaining cap then carefully disconnect the jack. Reconnect the pendant control cable to the existing port on the other side of the bed.
Pendant Port Alignment and Care
Ensure that the indexing arrows are aligned to enable connection. Insert pendant cable into pendant port until O-ring seats into pendant port and then thread on retaining cap to retain this connection.
To ensure that the pendant port remains free of dirt and debris, ALWAYS cap any pendant port not in use. The threaded cap is simply screwed onto the unused end of the pendant T-cable.
If you relocate the pendant to the other side of the bed, ensure that you cap the unused pendant port.
3.2.1 Patient Handheld Control (Pendant)
Page 43
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 43 of 159
3.3 Bed Positioning
To raise the bed, press the BED UP arrow on the footboard staff control, the rail staff controls, or on the pendant.
Release the button when the desired elevation has been achieved.
To lower the bed, press the BED DOWN arrow on the footboard staff control, the rail staff controls, or on the pendant.
Release the button when the desired elevation has been achieved.
To raise the head section of the bed, press the HEAD UP arrow on the footboard staff control, the rail staff/patient controls, or on the pendant.
Release the button when the desired elevation has been achieved.
To lower the head section of the bed, press the HEAD DOWN arrow on the footboard staff control, the rail staff/patient controls, or on the pendant.
Release the button when the desired elevation has been achieved.
3.3.1 Bed Elevation – HI-LO Operation
3.3.2 Head Deck (Back Rest) Elevation - Head Actuator Operation
Page 44
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 44 of 159
3.3.3 Knee Deck Elevation - Foot Actuator Operation
To raise the knee section of the bed, press the KNEE UP arrow on the footboard staff control, the rail staff/patient controls, or on the pendant.
Release the button when the desired elevation has been achieved.
To lower the knee section of the bed, press the KNEE DOWN arrow on the footboard staff control, the rail staff/patient controls, or on the pendant.
Release the button when the desired elevation has been achieved.
When the head section of the mattress deck (back rest) is elevated, there is often the tendency for the patient to slide towards the foot end of the bed. The Auto Contour Mode helps to prevent this motion. The Auto Contour mode automatically raises/lowers the knee section to, correspondingly, whenever the head section is raised or lowered.
To activate Auto Contour mode, simply press the CONTOUR button. The green LED in the top right corner of the button (Spirit Plus™) or the entire
CONTOUR button (Spirit Select™) will
illuminate when the Contour mode has been activated.
To deactivate Auto Contour mode, simply press the CONTOUR button again. If the CONTOUR button is not illuminated, Auto Contour mode has been deactivated.
NOTE:
Patient lock-out will override Auto Contour if the knee elevation lock-out has been activated.
3.3.4 Auto Contour Mode
Page 45
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 45 of 159
3.3.5 Chair Positioning
The Chair position allows patient to be placed in an upright seating position.
To Achieve Chair Positioning
Simply press-and-hold the CHAIR button.
The bed will automatically articulate into the chair position. The green LED in the top right corner of the CHAIR button (Spirit
Plus™ beds) will illuminate once the proper
reverse Trendelenburg angle has been achieved. The CHAIR button does not illuminate on Spirit Select™.
Release the CHAIR button anytime when the desired chair position has been achieved or the bed will stop automatically once the full chair position has been achieved.
To Return the Bed to a Horizontal Position
On Spirit Plus™ beds:
Simply press-and-hold the CHAIR button until the bed is flat and level. The bed will stop automatically.
IMPORTANT: This will only work if the bed is in the full chair position (the green LED in the top right corner of
the CHAIR button is illuminated). Either continue to press-and-hold the CHAIR button until the full chair position is achieved, then release and press-and-hold the CHAIR button until the bed is flat and level or press-and- hold the TREND button until the bed is level and press-and-hold the CPR button until the bed is flat.
On Spirit Select™ beds:
Simply press-and-hold the LEVEL ALL button until the bed is flat and level. The bed will stop automatically.
CAUTION
The reverse Trendelenburg position is integral to Chair positioning configuration. The
Spirit™ bed may shift during reverse Trendelenburg activation. Attempt Chair positioning only after the Central “Lock & Steer” system has been put in the “BRAKE” position.
Activation of the Electronic CPR function allows press-and-hold flattening of the mattress deck allowing staff to administer CPR to the patient.
Activation of the Electronic CPR Function on the Footboard Staff Control
On Spirit Plus™ beds:
Press-and-hold the CPR button on the
footboard staff control. The head and knee sections will automatically lower to flat position.
To interrupt the CPR function, simply release the CPR button. Normal bed operation can be resumed at any time. The Electronic CPR function does not need be deactivated or reset.
3.4 Emergency & Staff Functions
3.4.1 Electronic CPR (Cardiopulmonary Resuscitation) Function
Page 46
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 46 of 159
On Spirit Select™ beds:
Press-and-hold the CPR button on the
footboard staff control. The head and knee sections will lower to flat position.
IMPORTANT: The LCD display will show a notification when the CPR button on the footboard staff control is pressed to remind staff of this functional difference.
To interrupt the Electronic CPR function, simply release the CPR button. Normal bed operation can be resumed at any time. The Electronic CPR function does not need be deactivated or reset.
CAUTION
Improper use of the Electronic CPR function may cause patient injury. Once activated, Electronic CPR function will lower head and knee sections to flat position as
long as the CPR button is pressed. To interrupt Electronic CPR function, simply release the CPR button.
CHG Hospital Beds offers an optional manual CPR release feature on Spirit Select™ beds.
In the event of a medical emergency, activation of the manual CPR release handle will mechanically flatten the head section (back rest) of the mattress deck allowing staff to administer CPR to the patient. The knee section of the mattress deck will also automatically flatten.
The manual CPR release will work when mains AC power has been disrupted in the event of either a power failure or when the bed is not connected to mains AC power (unplugged from the wall outlet).
If your Spirit Select™ bed has been equipped with the optional manual CPR release it will be outfitted with two CPR pull handles, one mounted on each side of the bed in the location shown.
PR E S S & HO L D C P R TO COM P LET E
PRESS & HOLD CPR TO COMPLETE
3.4.2 Optional Manual CPR (Cardiopulmonary Resuscitation) Release
Page 47
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 47 of 159
Activating the Manual CPR Release Feature
To flatten the head section (back rest), Pull Up and Hold either of the manual CPR release handles.
Continue to hold the CPR release handle until the head section (back rest) has been completely flattened.
CAUTION
When the bed is configured with manual CPR release, ONLY trained healthcare practitioners shall operate the manual CPR release in the event of a medical emergency. ONLY operate the CPR button and/or manual CPR release with all persons clear of mechanisms. PRESS and HOLD the CPR button to lower the head (back rest) and knee sections to a flat position.
To interrupt the CPR function, simply release the CPR button. To avoid unintended CPR release keep area around CPR handle clear of obstructions.
Notifications of the Manual CPR Release Feature
Immediately after either one of the CPR release handles has been pulled, the LCD display will display the following message.
IMPORTANT: The manual CPR release feature MUST be reset whenever the manual CPR release
handle is pulled.
If the head section (back rest) is fully flattened, the bed will automatically reset the manual CPR release feature without requiring additional action from staff.
The LCD display will display the following message to notify staff that the automatic CPR reset procedure is in progress.
The LCD display will display the following message and an audible “chirp” will sound to indicate that the automatic CPR reset procedure has been completed.
MA N UAL CP R PU L L D E TEC T E D
MA N UAL CP R HAS B E E N R E S ET
WA I T F O R C P R TO COM P LET E
Page 48
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 48 of 159
Resetting the Manual CPR Release Feature
IMPORTANT: The manual CPR release feature MUST be reset whenever the manual CPR release
handle is pulled.
However, in the event that the manual CPR release has been interrupted before the head section (back rest) has been fully flattened, staff will need to follow the prompts on the LCD display to complete the CPR reset procedure.
To Reset the Manual CPR Release Feature:
Pull Up and Hold either of the manual CPR release
handles
OR
Press-and-hold the CPR button
Additionally, in the event that any button is pressed on any control location before the head section (back rest) has been fully flattened, the automatic reset procedure will also be interrupted.
The LCD display will continue to display the following message to instruct staff to continue to press-and- hold the CPR button until the CPR reset procedure has been completed.
The LCD display will display the following message and an audible “chirp” will sound to indicate that the manual CPR reset procedure has been completed.
PU L L & HO L D CP R KE Y TO RE S ET
MA N UAL CP R HAS B E E N R E S ET
PU L L & HO L D CP R KE Y TO RE S ET
PU L L & HO L D CP R HA N DL E OR . . .
MA N UAL CP R IN T ERR U PTE D . . .
Page 49
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 49 of 159
3.4.3 Trendelenburg Operation
Press-and-hold the TREND button. Release the button
once the desired bed angle between 0° and 14° has been achieved.
To return the bed to the horizontal position, press-and- hold the rev.TREND button. The bed will automatically stop when the mattress deck has returned to the horizontal position.
The Trendelenburg angle is shown on the bed frame angle gauge integrated on the staff side of both foot siderails.
On Spirit Select™ beds, the LCD display on the enhanced
footboard staff control will show the current Trendelenburg angle. The entire TREND button will also illuminate once the bed has achieved a Trendelenburg angle of greater than 2°.
Press-and-hold the rev.TREND button. Release the button once the desired bed angle between 0° and -14° has been achieved.
To return the bed to the horizontal position, press-and- hold the TREND button. The bed will automatically stop when the mattress deck has returned to the horizontal position.
The reverse Trendelenburg angle is shown on the bed frame angle gauge integrated on the staff side of both foot siderails.
On Spirit Select™ beds, the LCD display on the enhanced
footboard staff control will show the current reverse Trendelenburg angle. The entire rev.TREND button will also illuminate once the bed has achieved a Reverse Trendelenburg angle greater than 2°
CAUTION
Patient discomfort may result from normal operation of the Trendelenburg and reverse Trendelenburg function.
Trendelenburg and reverse Trendelenburg modes should only be used on the advice of a medical practitioner.
WARNING
The Spirit™ bed may shift during Trendelenburg or reverse Trendelenburg activation.
Initiate Trendelenburg or reverse Trendelenburg only after the Central “Lock & Steer” system has been put in the “BRAKE” position.
NOTE:
ACCURACY: The Trendelenburg and reverse Trendelenburg angles shown on the LCD
display of the enhanced footboard staff control on Spirit Select™ bed has an approximate accuracy of ±2°.
3.4.4 Reverse Trendelenburg Operation
Page 50
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 50 of 159
3.5 Patient Lock-Out Functions
Patient Lock-Outs restrict the patient from initiating head, knee, and bed motion from the siderail controls or the pendant (if equipped). The degree of restriction depends on the Lock-Out option selected.
Activating Patient Lock-Out
Staff may choose to restrict patient access to one, two, or all three bed functions. To restrict a bed function, press the Patient Lock-Out button under that particular function. The yellow LED in the top right corner of the button (Spirit Plus™) or the entire Lock-Out button (Spirit Select™) will illuminate when patient control of a particular function has been Locked-Out.
IMPORTANT: Patient Lock-Outs restrict bed operation from both of the siderail controls and pendant (if equipped).
Deactivating Patient Lock-Out
To deactivate any/all Patient Lock-Out(s), simply press the Patient Lock-Out button(s) again. When a Patient Lock-Out button is no longer illuminated, patient control of that particular function has been restored.
NOTE:
When Patient Lock-Outs have been activated, the footboard staff control remains fully active.
To completely restrict access to all bed functions (except the Electronic CPR function) from all control locations (staff and patient controls), use the Master Lock-Out function.
Activating Master Lock-Out
To activate the Master Lock-Out, press-and-hold all three Patient Lock-Out
buttons at the same time. All three yellow LED’s in the top right corner of the three Patient Lock-Out buttons (Spirit Plus™) or all three entire Patient Lock-Out buttons (Spirit Select™) will flash sequentially to indicate that Master Lock-Out has been
activated.
IMPORTANT: Master Lock-Out restricts bed operation from all control locations (both siderail controls and the footboard staff control).
Deactivating Master Lock-Out
To deactivate the Master Lock-Out function, simply repeat this process. Press-and­hold all three Patient Lock-Out buttons at the same time. When all three Patient
Lock-Out buttons are no longer flashing sequentially, Master Lock-Out has been deactivated and both staff and patient control from all control locations has been restored.
NOTE:
When Master Lock-Out has been activated, the Electronic CPR function, Nurse Call, and the
underbed light remain operable. On Spirit Select™ beds, the bed exit feature also remain
operable.
WARNING
When a patient is left unattended, the bed should ALWAYS be lowered to its lowest position and the bed elevation controls should be locked-out in order to reduce the risk of patient injury.
3.5.1 Regular Patient Lock-Outs
3.5.2 Master Lock-Out
Page 51
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 51 of 159
3.5.3 Security Lock-Out
The Security Lock-Out feature allows for a high-security lock-out of ALL bed functions (including the Electronic CPR function). A
“User Passcode”
is required to perform any/all bed functions. The Security
Lock-Out feature must be activated or deactivated as required by the facility.
IMPORTANT: The manual CPR release function on bed (if equipped) remains operational.
WARNING
IMPORTANT: When the Security Lock-Out feature has been activated, ALL bed operation including the Electronic CPR function from any control location (footboard staff
control, both siderail controls, and pendant) is restricted.
Activating the Security Lock-Out Feature
To deactivate the Security Lock-Out feature, a User Passcode must first be created. The User Passcode must be between 1 and 6 digits long. The digits are interpreted from a virtual numeric keypad superimposed over the footboard staff control buttons. Imagine the numbers 1 2 3, 4 5 6, 7 8 9, just as they would appear on a telephone. This is demonstrated below.
Spirit Plus™ keypad
Spirit Select™ keypad
Guidelines for Choosing a User Passcode
The number of digits chosen for the User Passcode will determine the level of security provided. A single digit User Passcode is easy for staff to remember and unlock but offers virtually no security. A six digit User Passcode offers very high security but may be a nuisance to staff.
It is recommended that simple “lines” such as 1-2-3 or 1-4-7 are not chosen since they are relatively
easy to guess.
It is recommended that the code contains at least one of the Patient Lock-Out buttons (7, 8, or 9)
since these buttons are not generally used during normal bed operation therefore adding an enhanced level of security.
Codes that contain repeating digits (i.e.: 1-3-1-7) can enhance security since they are considerably
less likely to be guessed.
Codes that contain patterns of digits (i.e.: all four corners) can help staff remember the User
Passcode but offer slightly less security than random codes.
1 2 3 4 5 6 7 8 9 4 5 6 1 2 3 7 8
9
Page 52
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 52 of 159
Creating/Changing User Passcode
A User Passcode must first be created before the Security Lock-Out feature can be activated.
To Create the User Passcode:
1. Press the Bed Lock-Out, Knee Lock-Out, and Contour buttons simultaneously. These three Lock-Out
buttons will then begin to flash to indicate that the bed is now ready to receive the Security Access Code.
2. Enter the Security Access Code.
Security Access Code: “1881 825 153”
IMPORTANT: This Security Access Code is factory set and is not customizable
3. Once the Security Access Code has been successfully entered, an audible “chirp” will sound three
times and the Contour button will also begin to flash to indicate that the bed is now ready to receive the customized User Passcode.
4. Enter a 1 to 6 digit User Passcode of your choosing. Time-out will occur if more than 30 seconds
elapses between User Passcode digit input.
5. When the User Passcode has been entered, press the Contour button to complete the process. An
audible “chirp” will sound to indicate that the User Passcode has been set and that the Security Lock­Out feature has now been activated.
NOTE: DO NOT allow patients to learn either the Security Access Code or the User Passcode or the Security Lock-Out feature could become compromised!
To Change the User Passcode:
In the event that the User Passcode has been lost or compromised, a new User Passcode can be created. Simply repeat the above steps to change the User Passcode. The User Passcode can be changed as often as required to maintain effective control of the Security Lock-Out feature
If, for some reason, you wish to abort the User Passcode creation/change procedure, simply press the Bed Lock-Out, Knee Lock-Out, and Contour buttons simultaneously to exit User Passcode set up mode. An audible “chirp” will sound twice and the bed will return to its previous state (Security Lock-Out activated or deactivated). The bed will also automatically return to its previous state after 30 seconds of keypad inactivity.
Locking/Unlocking the Bed for use by Staff
To enable any/all bed functions, the bed must be unlocked.
To unlock the Security Lock-Out:
1. Enter the 1 to 6 digit User Passcode.
2. Once the User Passcode has been successfully entered, an audible “chirp” will sound and the three
Patient Lock-Out buttons will begin to flash to indicate that the bed is now unlocked. All bed functions are now available from all control locations (the footboard staff control and both siderail controls).
IMPORTANT: There is no feedback during this step. If you make a mistake, or if nothing happens after the User Passcode has been entered, simply re-enter the User Passcode starting at the first digit again There is
no need to “cancel” or abort the procedure if you make a mistake while entering the User Passcode as the bed is “looking” only for the complete sequence of digits that comprise the User Passcode.
To re-lock the bed:
When you are finished adjusting the bed positions, press any one of the flashing Patient Lock-Out buttons to re-lock the bed. An audible “chirp” will sound twice to indicate the bed has been re-locked.
OR
The bed will automatically re-lock after 30 seconds of keypad inactivity.
Page 53
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 53 of 159
Notification of the Security Lock-Out Feature
Once the Security Lock-Out feature has been activated the LCD display will display the following message whenever any button is pressed.
IMPORTANT: ALL bed operation including the Electronic CPR function from any control location
(footboard staff control, both siderail controls, and pendant) is restricted.
Deactivating the Security Lock-Out Feature
To restore normal bed operation (i.e.: full time control of bed functions from any control location), the Security Lock-Out feature must be deactivated.
To deactivate the Security Lock-Out feature:
1. Press the Bed Lock-Out, Knee Lock-Out, and Contour buttons simultaneously. These three buttons
will then begin to flash to indicate that the bed is now ready to receive the Security Access Code.
2. Enter the Security Access Code (NOT the User Passcode).
3. Once the Security Access Code has been successfully entered, an audible “chirp” will sound three
times and the Contour button will begin to flash.
4. Press the Contour button. An audible “chirp” will sound twice to indicate that the Security Lock-Out
feature has now been deactivated.
WARNING
When the Security Lock-Out feature has been activated, the Electronic CPR function will NOT work. The Security Lock-Out feature is only intended to be used by qualified medical practitioners on a case-by-case basis based upon clinically assessed needs of the patient as requiring a hospital bed having restricted motion. CHG Hospital Beds strongly recommends that Security Lock-Out NEVER be activated when the bed is in normal ward use.
NOTE:
After a hard reset, all bed settings will be automatically restored. The Security Lock-Out feature will be deactivated after a hard reset.
SEC U RI T Y L O CK O UT AC T I V A T ED
Page 54
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 54 of 159
3.6 Other Bed Functions & Features
All Spirit™ beds are equipped with one Nurse Call port to accommodate Nurse Call control.
The Nurse Call port is located on the underside of the headboard mounting bracket on the patient right side of the bed.
To operate the Nurse Call feature, press the red NURSE button on the rail patient control (inside of each head rail control). This will send a Nurse Call signal to the nurses’ station when appropriately connected to the communication system of your healthcare facility.
Operation of the Nurse Call feature requires a conventional connection cable
(not supplied) to be plugged into the ¼” Phono jack on the headboard
mounting bracket and into the designated outlet in the patient station.
Ensure that the attached “dummy plug” is inserted to an inactive Nurse Call port.
The “dummy plug” has an integral lanyard that is screw connected to the headboard mounting bracket to ensure that the “dummy plug” is not lost
when the Nurse Call port is being used. All Spirit™ beds are equipped with a standard 37 pin connector to accommodate connection to the
communication system typical of most healthcare facilities.
The 37 pin connector is located on the underside of the headboard mounting bracket directly beside the Nurse Call port on the patient right side of the bed.
The “pin outs” for the 37 pin connector have been configured to work with the communication set-up typical of most 37 pin enabled healthcare facilities. If the configuration of the communication system used in your healthcare facility differs from conventional set-up, successful interconnection can only be guaranteed through the use of interconnection cables that have been specifically configured for your application.
If your healthcare facility requires a particular interconnection cable configuration, CHG Hospital Beds offers customized interconnection cables. Contact your sales representative for details.
3.6.1 Nurse Call
3.6.2 Standard 37 Pin Connector
Page 55
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 55 of 159
3.6.3 Spirit Plus™ Power Status Indicator
The footboard staff control on Spirit Plus™ beds is equipped with an integrated power
status indicator. This cluster of 3 discrete icons indicates the real-time status of power available to operate the bed. The icons will illuminate respective of the source of power, the operation of the battery charging circuit, and the level of battery power.
The battery is continuously charged when connected to the CB10 control box and the bed is plugged into an AC power outlet. If the battery voltage drops below 18VDC, battery power will be automatically disrupted to prevent further discharge which could impact the battery’s ability to fully recharge. When the bed is plugged back into an AC power outlet, battery power will be automatically restored.
Indicator LED is Solid GREEN
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10 control box and is fully charged. Bed is ready to operate on battery power if required. This is the optimum operating condition for the bed.
Indicator LED is Flashing GREEN
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10 control box but is currently recharging.
Indicator LED is Alternating Flashing GREEN and RED
IMPORTANT: Bed may not reliably operate on battery power if required.
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10 control box but not holding proper charge. Service bed and/or replace battery immediately.
OR
The battery is not connected to the CB10 control box. Connect battery to CB10 and allow battery to charge for 24hr.
Indicator LED is Solid RED
Bed is disconnected from an AC power outlet and is operating on battery power. The battery is connected to the CB10 control box and has not yet exhausted the available battery power. Bed is OK to operate on battery power but it should be connected to an AC power outlet as soon as is possible to recharge the battery.
IMPORTANT: Exercise caution when operating the bed on battery power. Ensure that only vital and necessary bed functions are performed to extend battery power as long as possible and ensure emergency functions remain available when needed.
Indicator LED is Flashing RED
Bed is disconnected from an AC power outlet and has been operating on battery power. Although the battery is connected to the CB10 control box, it has exhausted its battery power. Reconnect bed to an AC power outlet as soon as possible and allow battery to charge for 24hr.
IMPORTANT: At this point, reliable bed operation cannot be guaranteed although some bed functionality may still be available. Exercise caution operating the bed as the bed will cease operation without warning.
No Indicator LED illumination
IMPORTANT: Bed may NOT operate.
The bed has been disconnected from an AC power outlet and the battery is disconnected from the CB10 control box. Reconnect bed to an AC power outlet and allow battery to charge for 24hr.
OR
The bed has a faulty connection or component (CB09 control box, CB10 control box, battery, and/or connection cable). Confirm all connections and begin troubleshooting potentially faulty components.
OR
The bed has been disconnected from an AC power outlet and the battery has been automatically disengaged from the battery circuit because the voltage has dropped below 18VDC. Reconnect bed to an AC power outlet and allow battery to charge for 24hr.
NOTE:
A new back-up battery, when fully charged, has sufficient power to deliver the equivalent of approximately 5 CPR cycles with a patient weight of 250 lbs. Battery power discharge rate depends upon factors such as patient weight, age of battery, ambient temperature and humidity, and the number of charge/discharge cycles the battery has been subjected to. For example, the heavier the patient, the faster battery power will be exhausted during battery powered bed operation. Conversely, the lighter the patient, battery power may offer a longer duration of battery powered bed operation.
Page 56
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 56 of 159
3.6.4 Spirit Select™ Power Status Indicator
The enhanced footboard staff control on Spirit Select™ beds is equipped with an integrated power status indicator. This cluster of 3 discrete icons indicates the real­time status of power available to operate the bed. The icons will illuminate respective of the source of power, the operation of the battery charging circuit, and the level of battery power.
The battery is continuously charged when connected to the CB10/CB11 control box and the bed is plugged into an AC power outlet. If the battery voltage drops below 18VDC, battery power will be automatically disrupted to prevent further discharge which could impact the battery’s ability to fully recharge. When the bed is plugged back into an AC power outlet, battery power will be automatically restored.
AC Light is Solid GREEN
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10/CB11 control box and is fully charged. Bed is ready to operate on battery power if required. This is the optimum operating condition for the bed.
AC Light is Solid GREEN + DC Light is Flashing AMBER
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10/CB11 control box but is currently recharging.
AC Light is Solid GREEN + ATTN Light is Flashing RED
IMPORTANT: Bed may not reliably operate on battery power if required.
Bed is connected to an AC power outlet and operating on AC power. The battery is connected to the CB10/CB11 control box but not holding proper charge. Service bed and/or replace battery immediately.
OR
The battery is not connected to the CB10/CB11 control box. Connect battery to CB10/CB11 and allow battery to charge for 24hr.
DC Light is Solid AMBER
Bed is disconnected from an AC power outlet and is operating on battery power. The battery is connected to the CB10/CB11 control box and has not yet exhausted the available battery power. Bed is OK to operate on battery power but it should be connected to an AC power outlet as soon as possible to recharge the battery.
IMPORTANT: Exercise caution when operating the bed on battery power. Ensure that only vital and necessary bed functions are performed to extend battery power as long as possible and ensure emergency functions remain available when needed.
DC Light is Solid AMBER + ATTN Light is Flashing RED
Bed is disconnected from an AC power outlet and has been operating on battery power. Although the battery is connected to the CB10/CB11 control box, it has exhausted its battery power. Reconnect bed to an AC power outlet as soon as possible and allow battery to charge for 24hr. IMPORTANT: At this point, reliable bed operation cannot be guaranteed although some bed functionality may still be available. Exercise caution operating the bed as the bed will cease operation without warning.
NO Light illumination
IMPORTANT: Bed may NOT operate.
The bed has been disconnected from an AC power outlet and the battery is disconnected from the CB10 control box. Reconnect bed to an AC power outlet and allow battery to charge for 24hr.
OR
The bed has a faulty connection or component (CB09 control box, CB10/CB11 control box, battery, and/or connection cable). Confirm all connections and begin troubleshooting potentially faulty components.
OR
The bed has been disconnected from an AC power outlet and the battery has been automatically disengaged from the battery circuit because the voltage has dropped below 18VDC. Reconnect bed to an AC power outlet and allow battery to charge for 24hr.
NOTE:
A new back-up battery, when fully charged, has sufficient power to deliver the equivalent of approximately 5 CPR cycles with a patient weight of 250 lbs. Battery power discharge rate depends upon factors such as patient weight, age of battery, ambient temperature and humidity, and the number of charge/discharge cycles the battery has been subjected to. For example, the heavier the patient, the faster battery power will be exhausted during battery powered bed operation. Conversely, the lighter the patient, battery power may offer a longer duration of battery powered bed operation.
Page 57
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 57 of 159
3.7 Siderail Operation
WARNING
Safe Working Load of Siderails
80 kg (176 lbs)
DO NOT exceed a 80 kg (176 lbs) load on any siderail
WARNING
Possible PATIENT ENTRAPMENT if use of siderails that are damaged or modified in any way.
NEVER use siderails from other manufacturers and/or with dimensions different that the original components and/or assemblies that came equipped with the bed. Variations in siderail design, (width, height, shape, profiles, opening, locking/latching mechanisms, etc.) could cause/contribute to patient entrapment and/or could potentially increase the probability/possibility for patient entrapment.
CAUTION
DO NOT use the siderails as a patient lifting device/apparatus. Siderails can be deformed or broken if excessive side loading/pressure is exerted. The siderails are used for the purpose of preventing a patient from inadvertently rolling out of bed.
DO NOT use the siderails as part of a patient restraint mechanism. Siderails are NOT intended for nor may be used for restraint purposes. If a patient is capable of inflicting self-injury or injury to staff, a physician or other suitably trained healthcare practitioner should be consulted for alternate means of safe patient restraint. The Spirit™ bed is equipped with 6 patient restraint loops for this purpose.
CAUTION
When operating the bed, ALWAYS ensure that the individuals using the bed is positioned properly on the mattress within the confines of the bed.
DO NOT let any patient extremities protrude over the side and/or between the siderails when operating the bed.
CAUTION
DO NOT use the siderails as lifting points for the bed during transportation and/or storage.
DO NOT use the siderails as a means of mechanically restraining the bed during transportation
and/or storage.
All Spirit™ bed siderail assemblies fully comply with FDA and Health Canada patient entrapment reduction guidelines and have successfully passed rail entrapment testing using test procedures outlined by the Hospital Bed Safety Workgroup. Spirit™ bed siderail assemblies were qualified using the Cone and Cylinder Tool as specified by FDA and Health Canada patient entrapment reduction guidelines. Entrapment Test Kits are readily available so on-going compliance can be monitored by the healthcare provider. For further information, refer to the FDA’s website at http://www.fda.gov/cdrh/beds/.
3.7.1 Intended Application/Use of Spirit™ Bed Siderail Assemblies
When fully raised/closed and locked/latched in the UP position, the siderails define a residence area for patients to rest upon the bed. This provides positive patient location/support helping to protect against the potential for patients to fall from the bed as the siderails give the patient a visual and physical indicator of the edge of the bed. When fully lowered/open in the DOWN position, the siderails permit patient ingress to and egress from the bed. The integral handles common to all siderails facilitate patient mobility by providing a gripping region during ingress to and egress from the bed.
Page 58
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 58 of 159
CAUTION
DO NOT use a bed with damaged or non-latching rails. If a bed is found with damaged or non­latching rails, remove bed from patient use and service bed immediately.
All Spirit™ bed comes with moulded plastic siderails as standard equipment.
Operation of the Plastic Siderails
It is important that caregivers know how to operate the plastic siderails safely. The plastic siderails lock/latch in the UP position and enables patient ingress, egress, and transfer in the DOWN position.
In the UP position, the Plastic siderails provide positive patient location and support helping to protect the patient from the potential of falling from the bed.
IMPORTANT: An audible “click” should be heard when each siderail assembly has been completely rotated into the UP position as the locking/latching mechanism engages.
The DOWN position, the plastic siderails fully rotate out of the way to provide unimpeded and unassisted patient ingress to and egress from the bed and also enables patient transfer by staff.
IMPORTANT: The siderails will not lock/latch in the DOWN position.
Raising the Foot Siderails
Gently rotate the foot siderail to the UP position. The foot siderail will first arc toward the foot end of the bed then back towards the head end of the bed as it rotates to the UP position.
IMPORTANT: An audible “click” should be
heard when the foot siderail has been completely raised to the UP position as the locking/latching mechanism engages.
3.7.2 Plastic Regular Height Siderails
Page 59
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 59 of 159
Lowering the Foot Siderails
Depress the green PRESS button then gently PRESS down on the foot siderail. The foot siderail will first arc towards the foot end of the bed then back towards the head end of the bed as it rotates to the DOWN position.
IMPORTANT: The foot siderails will not lock/latch in the DOWN position.
Raising the Head Siderails
Gently rotate the head siderail to the UP position. The head siderail will first arc toward the head end of the bed then back towards the foot end of the bed as it rotates to the UP position.
IMPORTANT: An audible “click” should be
heard when the head siderail has been completely raised to the UP position as the locking/latching mechanism engages.
Lowering the Head Siderails
Depress the green PRESS button then gently PRESS down on the head siderail. The head siderail will first arc towards the head end of the bed then back towards the foot end of the bed as it rotates to the DOWN position.
IMPORTANT: The head siderails will not lock/latch in the DOWN position.
Your Spirit™ bed may be equipped with the optional moulded plastic high siderails*.
*Plastic high siderails come standard with CHG beige colour matched inlays. Plastic high siderails shown have optional coloured inlays.
Operation of the Plastic High Siderails
It is important that caregivers know how to operate the plastic high siderails safely. The high siderails lock in the UP position and enables patient ingress, egress, and transfer in the DOWN position.
In the UP position, the high siderails provide positive patient location and support helping to protect the patient from the potential of falling from the bed.
IMPORTANT: An audible “click” should be heard when each high siderail assembly has been completely rotated into the UP position as the locking/latching mechanism engages.
3.7.3 Plastic High Siderails
Page 60
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 60 of 159
The DOWN position, the high siderails fully rotate out of the way to provide unimpeded and unassisted patient ingress to and egress from the bed and also enables patient transfer by staff.
IMPORTANT: The high siderails will not lock/latch in the DOWN position.
Raising the High Foot Siderails
Gently rotate the high foot siderail to the UP position. The high foot siderail will first arc toward the foot end of the bed then back towards the head end of the bed as it rotates to the UP position.
IMPORTANT: An audible “click” should
be heard when the high foot siderail has been completely raised to the UP position as the locking/latching mechanism engages.
Lowering the High Foot Siderails
Depress the green PRESS button then gently PRESS down on the high foot siderail. The high foot siderail will first arc towards the foot end of the bed then back towards the head end of the bed as it rotates to the DOWN position.
IMPORTANT: The high foot siderails will not lock/latch in the DOWN position.
Raising the High Head Siderails
Gently rotate the high head siderail to the UP position. The high head siderail will first arc toward the head end of the bed then back towards the foot end of the bed as it rotates to the UP position.
IMPORTANT: An audible “click” should be
heard when the high head siderail has been completely raised to the UP position as the locking/latching mechanism engages.
Page 61
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 61 of 159
Lowering the High Head Siderails
Depress the green PRESS button then gently PRESS down on the high head siderail. The high head siderail will first arc towards the head end of the bed then back towards the foot end of the bed as it rotates to the DOWN position.
IMPORTANT: The high head siderails will not lock/latch in the DOWN position.
CAUTION
To ensure the accuracy of the scale and alarms, keep the area around and/or under the bed open and free of obstructions and/or equipment.
To ensure optimal scale accuracy, CHG Hospital Beds recommends that the siderails be fully raised/closed and locked/latched in the UP position before reading patient weight.
Staff Siderail Controls
The head siderails incorporate integrated staff control that offers staff control of basic bed operations:
Head UP/DOWN, Knee/Foot UP/DOWN Bed UP/DOWN
IMPORTANT: Bed operation from this siderail control is disabled when the patient lock-outs are activated.
Patient Siderail Controls
The head siderails incorporate integrated patient control that offers patient control of basic bed operations:
Head UP/DOWN, Knee/Foot UP/DOWN Nurse Call feature
IMPORTANT: Bed operation from this siderail control is disabled when the patient lock-outs are activated however the Nurse Call feature always remains active.
Angle Gauges
All four siderails have integrated angle gauges:
The angle gauge in the head siderail panel
displays the angle of the head section of the mattress deck.
The angle gauge in the foot siderail panel
displays the Trendelenburg or reverse Trendelenburg angle.
NOTE: Angle Gauges are for Reference ONLY
Head Angle Gauge
Trend or rev Trend
Angle Gauge
3.7.4 Integrated Features of Plastic Regular Height and High Siderails
Page 62
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 62 of 159
3.7.5 Other Optional Siderails
Your Spirit™ bed may be equipped other optional moulded plastic siderails as shown below.
IMPORTANT: These other optional siderails operate exactly the same as described above.
Plastic High Head Siderails (“Filled-in” version with beige inlays shown)
Plastic Long High Head Siderails (optional wood head and footboards with beige inlays shown)
To accommodate the needs of some patients to have a greater level of mobility and independence, CHG Hospital Beds offers an optional Electronic Foot Siderail Release feature. This feature allows the patient to unlock/unlatch the foot siderails while in the bed by pressing a button on the head siderail control. This will permit patient ingress to and egress from the bed without the aid from staff.
If your Spirit Select™ bed has been equipped with the the optional electronic foot siderail release it will be outfitted with enhanced staff/patient head siderail controls as shown below.
IMPORTANT: The mechanical operation of beds equipped with the Electronic Foot Siderail Release feature is the same as standard production high siderails but simply have an additional button on the head rail controls which permits the foot siderail to be electronically unlocked/unlatched.
Enhanced Siderail Controls
The head siderails incorporate integrated controls which offer both staff and patient control of basic bed operations:
Head UP/DOWN, Bed UP/DOWN Knee/Foot UP/DOWN Nurse Call feature Electronic Foot Siderail Release
IMPORTANT: Similarly to the standard siderail controls, bed operation from this from this siderail control is disabled when the patient lock-outs are activated however the Nurse Call feature always remains active.
IMPORTANT: The Electronic Foot Siderail Release feature is disabled when the bed UP/DOWN lock-out is activated.
3.7.6 Optional Electronic Foot Siderail Release
Page 63
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 63 of 159
Electronic Foot Siderail Release Functionality
To avoid accidental release of the foot siderail, the release button must be pressed 3 times to activate the electronic release.
To Electronically Release the Foot Siderails:
Press the “FOOT RAIL/PRESS 3X” button 3 times within a period of 5 seconds. An audible chirp
will sound after each button press as confirmation to the patient. An audible “click” should be heard when the foot siderail(s) electronically unlocks/unlatches.
IMPORTANT: If a period of greater than 5 seconds elapses before the button is pressed 3 times, the foot rail will not release. An audible two tone chirp will sound to indicate a failed release attempt.
In the event that the foot siderails fail to electronically unlock/unlatch, if, say, bed linens prevent foot rail rotation, the patient may be inclined to make additional attempts to release the foot rail. To prevent premature component failure, only 3 attempts to activate this feature will be permitted in a row. Followed by a rest period of 1 minute to prevent the solenoid from overheating.
If your bed has been equipped with this optional feature, the factory default setting for Electronic Foot Siderail Release is ENABLED (both PL and PR sides). This setting can be changed by following the instructions below.
WARNING
NEVER attempt to release the foot siderail when the patient is supporting their weight on the foot siderails. NEVER hang or drape anything over foot siderails.
CHG Hospital Beds recommends that the customer perform a thorough patient assessment to determine if enabling the electronic foot siderail release is appropriate for the patient on the basis of their clinical needs, fall risk, and mental capacity.
To Enable/Disable the Electronic Foot Siderail Release Feature
From the SERVICE Menu:
1. Press the UP/DOWN arrow buttons to scroll to the FOOT RAIL RELEASE screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
LEFT ENABLED (patient left foot siderail release enabled) RIGHT ENABLED (patient right foot siderail release enabled) BOTH ENABLED (both patient left and patient right foot siderail release enabled) RAIL RELEASE DISABLED
3. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Electronic Foot Siderail Release setting has been changed. Press the 0.0/ZERO button to “escape” to the home screen.
Spirit Plus™ beds can be equipped with integrated underbed lighting as an option however the underbed light cannot be deactivated using the footboard staff control. If the underbed light needs to be deactivated, both of the underbed lights need to be disconnected from the CB10Lite control box. Refer to page 132 to identify the two connectors for the underbed lights:
To disable the underbed lights, disconnect both of the light connectors from the LT1 and LT2 headers on
the CB10Lite circuit board.
The ONLY setting for the underbed light: LOW brightness ALWAYS ON. IMPORTANT: There are no MEDIUM or HIGH brightness setting or visual indicator on Spirit Plus™ beds.
NOTE:
Underbed lighting WILL NOT function (illuminate) when the bed is operating under battery power. This is to reserve battery power for emergency functions.
3.8 Optional Underbed Light on Spirit Plus™ Beds
Page 64
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 64 of 159
Underbed & Night Light on Spirit Select™ Beds
The Spirit Select™ is equipped with integrated underbed lighting. See Section
4.6.4 Available Light
Configurations
for more information
If your Spirit™ bed has been equipped with the optional underbed obstruction sensors, the bed will be outfitted with 6 obstruction sensors as shown below.
IMPORTANT: The operation of the obstruction sensors is identical for all beds. While all beds have the audible component of this feature, only Spirit Select™ beds have visual indication and message display.
Top Cover Obstruction Sensor
Caster Cover Obstruction Sensor
Bottom Cover Obstruction Sensor
Contact with an Obstruction While the Bed is in Downward Motion
When an obstruction is detected during any downward motion of the bed (using any of the following: BED DOWN, TREND, rev.TREND, LEVEL ALL, or CHAIR position), the bed will immediately stop all downward
motion and the bed platform will automatically elevate up a safety distance of 1”. This will provide clearance
for removal of the obstruction.
On Spirit Select™ beds, once an obstruction has been detected, the
obstruction indicator
(BED UP button) will begin to flash and an audible “chirp” will sound. A warning message will also be displayed. The
obstruction indicator
will continue to flash to remind staff that an obstruction was
detected during the last attempt to initiate downward motion of the bed platform.
On Spirit Plus™ beds an audible “chirp” will sound.
Any further attempts to initiate downward motion of the bed platform will not be permitted if an obstruction is still being detected and the bed platform will emit an audible “chirp” after each attempt.
Contact with an Obstruction While Bed is Stationary (i.e.: Bed not in Motion)
When an obstruction is detected prior to any bed movement, all downward motion of the bed (using any of the following: BED DOWN, TREND, rev.TREND, LEVEL ALL, or CHAIR position) will be automatically disabled.
On Spirit Select™ beds, the
obstruction indicator
will flash until the obstruction is removed. A warning message will also be continuously displayed and an audible “chirp” will sound when any downward motion of the bed platform is attempted. Press the ENTER button to acknowledge and clear the warning message.
On Spirit Plus™ beds an audible “chirp” will sound.
Any further attempts to initiate downward motion of the bed platform will not be permitted if an obstruction is still being detected and the bed platform will emit an audible “chirp” after each attempt.
3.9 Optional Underbed Obstruction Sensors
Page 65
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 65 of 159
When an Obstruction is not Detected (i.e.: Obstruction Removed)
When an obstruction is no longer detected:
On Spirit Select™ beds, the
obstruction indicator
will stop flashing and normal bed operation will be restored.
On Spirit Plus™ beds, normal bed operation will be restored.
Visual and Audio Indicators
Visual Obstruction Indicator:
(Spirit Select™ only)
In addition to operating BED UP functionality, the BED UP serves as an
obstruction indicator
by illuminating & flashing whenever an obstruction has been detected. The “CHECK UNDER BED FOR OBSTRUCTION” warning message will also be displayed.
Audio Obstruction Indicator:
(all Spirit™ beds)
When an obstruction is contacted during bed motion or if an obstruction is detected prior to any bed lowering motion and a downward motion of the bed platform (using any of the following: BED DOWN, TREND, rev.TREND, LEVEL ALL, or CHAIR position) is attempted an audible “chirp” will sound.
On Spirit Select™ beds, the factor default setting for obstruction sensors is ENABLED. In the rare event of component failure and/or malfunction, the obstruction sensor system can be disabled in the Service Menu to restore bed functionality until the bed can be properly serviced. Refer to page 99 for instructions.
On Spirit Plus™ beds the factor default setting for obstruction sensors is CONNECTED and OPERATING. This cannot be changed using the footboard staff control. The obstruction sensors can only
be enabled/disabled by connecting/disconnecting the obstruction sensor circuits from the CB10Lite circuit board. Refer to page 132 to identify the two connectors for the obstruction sensor circuits.
CHECK UNDER BED FOR OBSTRUCTION!
Page 66
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 66 of 159
3.10 Optional “Easy Bed” Extension System
To accommodate the physical needs of taller patients, while maintaining the easy of frequent bed movement about space confined healthcare facilities, CHG Hospital Beds offers an optional “Easy Bed” extension system that allows for easy and rapid extension/retraction of the mattress deck by 4 inches (102 mm) from 80”
(2032 mm) to 84” (2134 mm). No tools are required to perform bed “Easy Bed” extension/retraction
operations. If your Spirit™ bed has been equipped with the optional “Easy Bed Extension” system, the bed will be
outfitted with two clamping knobs and an extendable foot deck section of the mattress deck as shown below.
Clamping Knobs
(viewed from foot end of bed)
Extendable Foot Deck Section
Bed Extension Procedure
1. Elevate bed to comfortable working height.
Unclamp the footboard bracket slide mechanism by turning both knobs counter-clockwise until they stop turning freely.
2. Grasp footboard bracket by the corners as shown
above. Gently and smoothly pull away from the head of the bed until the footboard bracket slide mechanism contacts its mechanical stops.
Knob clamped
Knob unclamped
Page 67
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 67 of 159
3. Clamp the footboard bracket slide mechanism by
turning both retention knobs clockwise until hand tight.
IMPORTANT: Ensure that the clamping knob
fastener extends through the retention hole located in the footboard bracket channel. Typical both sides.
4. Unlock foot deck extension pan by grasping the
chrome handle and sliding the locking mechanism towards the patient left side of the bed to the unlocked position (green, unlocked icon)
5. Grasp the chrome handle as shown at right. Gently
and smoothly pull away from the head end of the bed until the foot deck extension pan contacts its mechanical stops.
6. Lock foot deck extension pan by grasping the
chrome handle and sliding the locking mechanism towards the patient right side of the bed to the locked position (red, locked icon).
WARNING
Possible PATIENT ENTRAPMENT Hazard could occur if the mattress is not securely located and retained on the mattress deck by the four corner mattress keepers.
NEVER perform a bed extension/retraction with a patient on the bed.
NEVER place a patient on the bed if the mattress is not properly located and retained on
the mattress deck by the four corner mattress keepers.
CAUTION
Possible UNEXPECTED MOTION could occur if the slide lock is not in the “lock” position and the mattress deck is articulated from the horizontal position.
When the bed is in either the extended or retracted position, the clamping knob fasteners MUST be fully engaged and the slide lock MUST be in the “lock” position prior to placing a patient on the bed and/or allowing a patient to return to the bed.
NOTE:
Bed retraction procedure is the same as the above except that the footboard bracket and the foot deck extension pan are pushed towards the head end of the bed until contact is made with their respective mechanical stops.
Knob fully clamped (knob fastener fully engaged)
Knob not fully clamped (knob fastener not fully engaged)
Page 68
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 68 of 159
3.11 Optional Auxiliary AC Power Outlet
NEMA 5-15R Outlet Socket configuration shown
All Spirit™ beds can be equipped with an optional auxiliary AC power outlet mounted in the patient right side of the footboard bracket.
IMPORTANT: The auxiliary AC power outlet has the following electrical ratings:
MAXIMUM Output Power Rating: 120 Volts AC, 600 Watts
OR 230 Volts AC, 690 Watts
DO NOT EXCEED MAXIMUM OUTPUT POWER RATING
Auxiliary AC Power Outlet Cover
The auxiliary AC power outlet has a cover that is integrated into the footboard bracket. The attachment screw is retained with a nylon nut so that it will not be misplaced when the auxiliary outlet is in use.
IMPORTANT: Always install the cover over the auxiliary AC power outlet when it is not in use.
Circuit Breaker: Rating 5 Amps (120V Beds) OR 3 Amps (230V Beds)
The auxiliary AC power outlet socket is protected by a circuit breaker. The circuit breaker is located in the footboard bracket. The circuit breaker is the resettable type. Simply depress the button to re-engage the circuit breaker.
IMPORTANT: The circuit breaker is a safety device that is designed to disengage an electrical circuit when the load exceeds the maximum rating of the circuit. An overloaded circuit may cause irreparable damage to the AC power wiring in the bed. If the circuit breaker trips, it is important to understand what has happened to cause this event. The device(s) the bed is connected to may be drawing too much power and exceeding the circuit breaker rating.
CAUTION
The auxiliary AC power outlet socket is rated for a MAXIMUM 120 Volts AC, 600 Watts OR 230V
Volts AC, 690 Watts output power. DO NOT exceed the maximum output power rating.
DO NOT use the auxiliary AC power outlet socket if the socket receptacles appear to be damaged
and/or if the plug (of the device that you are connecting to the bed) does not remain connected when inserted in the outlet socket receptacles.
Only replace this outlet socket with a NEMA 5-15R duplex outlet socket that is marked as hospital grade. Contact CHG Hospital Beds’ Customer Service for a replacement outlet socket.
WARNING
A circuit breaker is NOT designed to act as a ground fault circuit interrupter (GFCI) and DOES NOT offer the same level of protection. The circuit breaker is a safety device that is designed to only disengage an electrical circuit when the load exceeds the maximum rating. If the circuit breaker trips,
discontinue bed use immediately and have a service technician, specifically trained to service Spirit™
beds, inspect the wiring. Circuit Breaker Rating: 5 Amps (120V beds) or 3 Amps (230V beds)
CAUTION
Connecting electrical equipment to auxiliary mains AC outlet effectively leads to creating a medical equipment system and can result in a reduced level of safety. When connecting devices, the system shall be evaluated in end use application to ensure compliance to medical device standard 60601-1 and 60601-1-1.
Page 69
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 69 of 159
3.12 Optional Pillow Speaker Ready
All Spirit™ Select beds can be configured to connect an optional pillow speaker.
The pillow speaker offers patients control of the following basic operations (but are not limited to):
Entertainment Controls
o TV Channels o TV Volume o Etc.
Nurse Call Button Room Light Read Light
Pillow Speaker Storage
The pillow speaker can be affixed to the bed using a holder. The pillow speaker slips into the holder for easy location and convenient patient access.
The pillow speaker holder can be installed into any of the high rail openings. To permanently affix the holder to the bed, screws the holder to the foot or head rail panel with the fasteners provided.
Locate the optimal position for the holster within the rail openings. Ensure the pillow speaker can freely slip in and out of the holster with minimal effort. To permanently affix the holder to the bed, fasten the holder to the foot or head rail panel with the hardware provided.
CAUTION
If connecting a Pillow Speaker for the first time refer to document 3-80-097 for installation instructions. Contact CHG Hospital Beds’ Customer Service at (519) 963-4010 or toll free 1­866-516-5446 to obtain instructions.
Pillow Speaker Connection
The pillow speaker can be connected on either side of the bed (the connection port is between the foot and head rails on the lower frame).
To relocate the pillow speaker to the opposite side of the bed, first, carefully disengage the coiled cable from the pillow speaker bracket. Disconnect the pillow speaker from the DB15 port. Reconnect the pillow speaker cable to the existing port on the other side of the bed. Route
the coiled portion of the pillow speaker cable through the pillow speaker bracket as depicted in the label located directly above the pillow speaker port.
Page 70
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 70 of 159
Pillow Speaker Cable Routing
CHG recommends Pillow Speaker cabling to ALWAYS be safely routed through the gap between the Head and Foot rails. Routing the cable in this fashion will minimize the creation of a trip hazard and rail entanglement.
Check rail movement to ensure Pillow Speaker cable does not become entangled in rails
Pillow Speaker Port Care
ALWAYS ensure that the coiled portion of the pillow speaker cable if
correctly and securely routed thought the pillow speaker bracket to provide strain relief for the connection port.
To ensure that the pillow speaker port remains free of dirt and debris, ALWAYS cap any pillow speaker port not in use. The cap is simply snap fit onto the unused end of the pillow speaker T-cable.
If you relocate the pillow speaker to the other side of the bed, ensure that you cap the unused pillow speaker port.
CAUTION
Ensure cables are routed properly to prevent damage to Pillow Speaker and rail entanglement. Using the label located directly above the connection point, route cable into pillow speaker bracket as shown. Failure to route cable properly may result in serious injury.
CAUTION
ALWAYS ensure protective caps are installed when a pillow speaker is not connected to the pillow speaker port to prevent contamination and to provide protection from Electrostatic Discharge (ESD)
WARNING
Only use Pillow Speakers provide by CHG Hospital Beds. The use of a third party pillow speaker may cause adverse effects, and may permanently damage parts of the nurse call system. Contact CHG
Hospital Beds’ Customer Service at (519) 963-4010 or toll free 1-866-516-5446 for additional information
CAUTION
If upon 1st connection of the Pillow Speaker Pendant to the CHG Spirit™ Bed the TV system begins to exhibit abnormal operation (i.e. channels switching automatically), TV polarity may need to be adjusted. Contact CHG Hospital Beds’ Customer Service at (519) 963-4010 or toll free 1-866-516-5446 for additional instruction on how to resolve this issue.
WARNING
Pillow Speakers contain magnets and may interfere with sensitive devices.
CAUTION
Not for standalone use or direct connection to wall. To be used only when connected to a CHG Spirit Hospital Bed.
Page 71
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 71 of 159
Section 4:
Spirit Select™ - Unique Functions & Features
Page 72
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 72 of 159
4.1 LCD Display
The enhanced footboard staff control on Spirit Select™ beds is equipped with an integral LCD display.
Home Screens
LB units and 30° HEAD and 0° TREND angle for this example
When the LCD display is activated, the home screen will default display the patient weight in either kilograms (KG) or pounds (LB) depending upon which unit of measure has been set for patient weight display.
Pressing the UP/DOWN arrow buttons will scroll through the home screens:
PATIENT WEIGHT HEAD angle and TREND/rev.TREND angle WARNINGS (if any)
IMPORTANT: No warning screens will be displayed if the bed has not detected any problems or no safety feature has been disabled.
Between the HEAD angle and TREND/rev.TREND angle home screen and WARNINGS home screen (if any) the LCD display the following message to help staff navigate into the menus.
After 20 seconds of keypad inactivity has elapsed, the LCD display will go into “sleep” mode (become inactive).
The LCD display is activated or “wakes up” when the ON/ENTER, TREND, or rev.TREND buttons are pressed.
20 SECONDS OF KEY PAD INACTIVITY
On/Enter Button
“ON” Functionality of this Button
Press this button to activate or “wake up” the LCD display.
“ENTER” Functionality of this Button
Press this button (as enter button) when prompted by the LCD display to make desired selections within menu options.
UP/DOWN Arrow Buttons
Press the UP/DOWN arrow buttons to scroll up or down in the different menus and to scroll through the available options within the menus when making desired selections.
TO ACC E SS M EN U S PR E S S E N TE R 
WA R NIN G #X DES C RI P TI O N !
HEA D TR E N D 30° 0°
PA T I EN T W E I GH T
0. 0 LB
Page 73
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 73 of 159
Warning Screens
The bed has the ability to detect warning conditions and display them to the user.
Warnings will be displayed whenever the bed is in an alarm condition or detects a fault or problem (i.e.: failed component) or whenever a safety feature has been disabled (turned OFF in the associated menu).
To view active warnings, return to the home screen and use the UP/DOWN arrow buttons to scroll through the list. If there are no warnings, only the regular home screens will appear.
When a warning is detected an audible “chirp” will
sound and the warning indicator icons in the lower right hand corner of the LCD display will sequentially flash between the exclamation mark (!) and the UP/DOWN arrow characters (). This indicator is meant to notify staff that the bed has detected a warning condition and that attention is required.
IMPORTANT: This indicator will continue to flash until the warning condition has been appropriately addressed.
SEQUENTIALLY FLASH
Each warning screen is numbered and contains a description of the cause of the warning. This is to aid in troubleshooting (if required). Refer to page 126 for a list of warnings.
Escape Function
Press the 0.0/ZERO button to return back to or
“escape” to the home screen. This function can be
performed at any time in any menu option screen.
IMPORTANT: No settings will be changed if no selections were made prior to pressing the 0.0/ZERO button.
Automatic Return Home Feature
To help staff with menu navigation, an automatic “return home” feature has been included.
How it works:
After 1 minute of keypad inactivity, the LCD display will automatically enter “sleep” mode. No
unselected change to any setting in any menu option screen will occur (i.e.: if a setting change attempt has been abandoned).
Pressing the ON/ENTER button after the LCD display enters “sleep” mode will “wake up” the LCD
display at the home screen.
The automatic “return home” feature will work from any menu option screen.
IMPORTANT: No settings will be changed if no selections were made prior to LCD display going to “sleep”.
NOTE:
The LCD display WILL NOT function or become illuminated when bed is operating under battery power. This is to conserve battery power for emergency functions.

!
WA R NIN G #X DES C RI P TI O N !
Page 74
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 74 of 159
4.2 Patient Weigh Scale
All Spirit Select™ beds are equipped with an integral scale to facilitate the weighing of patients while in the bed. The scale is always activated and will display the patient’s weight in either kilograms (KG) or pounds (LB)
depending upon which unit of measure has been set for patient weight display. The factory default for patient weight display is imperial pounds (LB). This setting can be changed in the Maintenance Menu.
The Spirit Select™ scale is very sensitive. The patient can be weighed in any bed position but patient weight will be determined and displayed most accurately when the bed is in the flat position with nothing touching the bed and the bed is not contacting the wall, patient station, or any other equipment.
ACCURACY: The Spirit Select™ scale system has an accuracy of ±3 lbs (1.4 kg) for patients weighing 50 to 110 lbs (22.7 to 49.9 kg) and ±3% of the total patient weight for patients weighing 110 to 500 lbs (49.9 to
226.8 kg). The specified scale accuracy pertains to Mattress Deck Heights from 17 in (431.8 mm) to 34 in (863.6 mm) in the horizontal position.
WARNING
The scale output is NOT intended to be used to determine diagnosis or treatment.
CAUTION
To ensure the accuracy of the scale and alarms, keep the area around and/or under the bed open and free of obstructions and/or equipment.
To ensure optimal scale accuracy, CHG Hospital Beds recommends that the siderails be fully raised/closed and locked/latched in the UP position before reading patient weight.
Zeroing the Scale (Establish Tare Value)
LB units used for this example
1. Press-and-hold the 0.0/ZERO button for 3
seconds. During this period, a message will be displayed to remind staff that the patient must not be in the bed while attempting to zero the scale.
2. Release the 0.0/ZERO button immediately
when instructed.
3.
DO NOT TOUCH BED
during the zeroing
process.
4. Ensure that the home screen is displaying a
weight of 0.0 LB before allowing the patient to enter the bed.
IMPORTANT: DO NOT zero the scale with a patient in the bed. This will result in a zero patient weight being displayed. If the scale is accidentally zeroed with a patient on the bed, remove patient from bed, re-zero scale and return patient to bed.
RE L E AS E T O ZE R O S C ALE
EN S U R E PA T I EN T IS N OT IN B ED
PR E S S & HO L D TO Z ER O S C ALE
Page 75
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 75 of 159
If an accidental attempt to zero the scale is initiated with the patient in the bed, simply release the
0.0/ZERO button when the reminder message is displayed.
IMPORTANT: Make sure that all necessary equipment and linens are on the bed prior to zeroing the scale. Placing additional items on the bed or removing items from the bed after the zeroing process will result in an inaccurate patient weight being displayed.
SEVERAL SECOND DELAY
IMPORTANT: The scale cannot be zeroed if the bed has been placed in any Trendelenburg or Reverse Trendelenburg position. Ensure that the bed has been restored to fully horizontal and displaying 0° TREND or rev.TREND angle before attempting to zero the scale. Refer to page 49 for instructions.
IMPORTANT: The scale cannot be zeroed if the Bed Exit Alarm has been armed. Ensure that the Bed Exit Alarm has been disarmed before attempting to zero the scale. Refer to page 79 for instructions.
NOTE:
CHG Hospital Beds recommends that the bed is zeroed prior to use by each new patient and patient weight change is zeroed once the new patient enters the bed.
NOTE:
Zeroing the scale will override any adjustments made to patient weight.
ZE R OIN G SC A LE PL E A SE WA I T . . .
PA T I EN T W E I GH T XX X .X L B
PL E A SE LE V E L B E D BE F ORE Z ER O I NG
TU R N O F F B E D E X IT BE F ORE Z ER O I NG
Page 76
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 76 of 159
Some Spirit Select™ beds may be equipped with an OIML Certified integrated scale to facilitate the weighing of patients while in the bed. The scale is always activated and will display the patient’s weight in kilograms (kg). The patient can only be weighed in the weighing position (bed level and within height range) to ensure scale accuracy. The caregiver will be notified if the bed is outside of this weighing position.
ACCURACY: The Spirit Select™ with OIML Certified integrated scale system has a maximum permissible error in service of ±0.5 kg.
CAUTION
To ensure the accuracy of the scale and alarms, keep the area around and/or under the bed open and free of obstructions and/or equipment.
To ensure optimal scale accuracy, CHG Hospital Beds recommends that the siderails be fully raised/closed and locked/latched in the UP position before reading patient weight.
Zeroing the Scale
Upon receiving your new Spirit Select™ bed, the scale must be zeroed. This process should also be performed whenever a mattress is removed from the bed.
5. Ensure that there is no mattress, linens,
equipment or other weight on the bed
6. Press-and-hold the [ZERO0/TTARE]
button for 1 second. LCD will display TO ZERO SCALE PRESS ENTER
7. Press ENTER
8.
DO NOT TOUCH BED
during the zeroing
process.
9. Ensure that the home screen is displaying a
weight of 0 kg before installing or replacing a mattress.
TO ZERO SCALE PRESS ENTER
MEASURING… DO NOT TOUCH BED
COMPLETE
PATIENT WEIGHT
0 kg
4.3 Patient Weight Scale OIML BEDS ONLY
Page 77
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 77 of 159
Establish a Tare Value
1. Add a mattress, linens and/or standard
equipment to the bed
2. Press-and-hold the [ZERO0/TTARE]
button for 1 second. LCD will display TO TARE SCALE PRESS ENTER
3. Press ENTER
4.
DO NOT TOUCH BED
during the tare
process.
5. Ensure that the home screen is displaying a
weight of 0 kg N before allowing the patient to enter the bed.
IMPORTANT: DO NOT tare the scale with a patient in the bed. This will result in a zero patient weight being displayed. If a tare is performed accidentally with a patient on the bed, remove patient from bed, re-tare scale and return patient to bed.
If an accidental attempt to tare the scale is initiated with the patient in the bed, simply tap the [ZERO0/TTARE] button to exit the scale tare.
IMPORTANT: Make sure that all necessary equipment and linens are on the bed prior to a tare process being performed. Placing additional items on the bed or removing items from the bed after a tare process will result in an inaccurate patient weight being displayed.
IMPORTANT: the letter N represents NET weight and is used to signify that a tare has been properly established
IMPORTANT: The scale cannot be tare/zeroed if the bed has been placed in any Trendelenburg or Reverse Trendelenburg position. Ensure that the bed has been restored to fully horizontal and displaying 0° TREND or revTREND angle before attempting to tare/zero the scale. Refer to page 49 for instructions.
IMPORTANT: The scale cannot be tare/zeroed if the bed is below the weighing position. Using the BED UP key, restore the bed to the weighing position. When the bed has been returned to a position that satisfies the weighing position, the LCD screen will display TO TARE SCALE PRESS ENTER
TO TARE SCALE PRESS ENTER
MEASURING… DO NOT TOUCH BED
COMPLETE
PATIENT WEIGHT
0 kg N
BED NOT LEVEL
PRESS ENTER
BED TOO LOW
PRESS ENTER
Page 78
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 78 of 159
IMPORTANT: The scale cannot be tare/zeroed if the bed is above the weighing position. Using the BED DOWN key, restore the bed to the weighing position. When the bed has been returned to a position that satisfies the weighing position, the LCD screen will display TO TARE SCALE PRESS ENTER
IMPORTANT: if the bed exit alarm is armed, performing a zero/tare operation will automatically disarm it.
NOTE:
CHG Hospital Beds recommends that a tare value is established prior to use by each new patient and patient weight change is zeroed once the new patient enters the bed.
NOTE:
Establishing a tare value will override any adjustments made to patient weight.
Special Scale Messages
DESCRIPTION
DISPLAY
1. The scale is only rated to measure gross
weights 10.0 kg and 254.5 kg. The current gross weight is less than 10 kg.
2. The scale is only rated to measure gross
weights 10.0 kg and 254.5 kg. The current gross weight is MORE than 254.5 kg.
3. The current weight is within 0.125 kg of 0.0
kg.
4. The weight displayed is a NET WEIGHT
meaning that either a Tare or an equipment change has taken place
All OIML Certified beds come factory shipped with the scale in a calibrated state. Should a calibration need to be performed, contact
CHG Hospital Beds’ Customer Service: +01 (519) 963-4010 or Toll Free
1-866-516-5446
for further instructions.
Scale accuracy verification should be performed at an interval determined by the end user. Scale accuracy verification should only be performed by trained users
TO TARE SCALE PRESS ENTER
BED TOO HIGH
PRESS ENTER
PATIENT WEIGHT < 10 kg
PATIENT WEIGHT > 254.5 kg
PATIENT WEIGHT 0 kg
PATIENT WEIGHT 0 kg N
4.5.1 Calibration OIML BEDS ONLY
4.5.2 Scale Accuracy Verification
Page 79
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 79 of 159
4.4 Level All Function
All Spirit Select™ beds are equipped with a Level All Function.
Press-and-hold the LEVEL ALL button to simultaneously flatten both the head and foot sections of the mattress deck and to return the bed frame to a horizontal position (parallel with the floor).
Bed motion will stop automatically when both sections of the mattress deck have been fully flattened and the bed frame is horizontal.
The enhanced footboard staff control on Spirit Select™ beds is equipped with an integral bed exit feature to help monitor and report patient activity with audible and/or Nurse/Priority Call alarms.
IMPORTANT: This feature is not intended to replace patient monitoring by staff.
Three Zone Bed Exit
If equipped, The Bed Exit Alarm can be set to three different sensitivity levels (zones) corresponding to the level of monitoring a patient requires.
The Bed Exit Alarm can be set to produce an audible bed alarm (three different volume settings) or to send a Nurse/Priority Call alarm or it can be set to produce both audible and Nurse/Priority Call alarms.
Single Zone Bed Exit
If equipped, The Bed Exit Alarm can be set to a single sensitivity level. This single zone setting is intended as a Bed Exit Alarm.
The Bed Exit Alarm can be set to produce an audible bed alarm (three different volume settings) or to send a Nurse/Priority Call alarm or it can be set to produce both audible and Nurse/Priority Call alarms.
LEVELLING BED PLEASE WAIT...
4.5 Bed Exit Feature
Page 80
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 80 of 159
Bed Exit Alarm Arming
To arm the Bed Exit Alarm (Three Zone):
1. Press the RESET/SET button to scroll through the alarm zones.
Zone 1 – Alarm with ANY movement Zone 2 – Alarm only when patient is about to exit bed Zone 3 – Alarm only when patient exits bed
2. When the desired alarm zone has been selected, the RESET/SET
button and the selected alarm zone icon will remain illuminated to indicate that the bed exit alarm has been activated.
3. After a delay of several seconds an audible “chirp” will also sound to
indicate that the Bed Exit Alarm has been armed.
IMPORTANT: The Bed Exit Alarm can ONLY be armed if there is weight on the bed (20lbs [9.1kg] minimum). This is intended to prevent false arming of this feature that may otherwise become a nuisance to staff.
To arm the Bed Exit Alarm (Single Zone):
1. Press the RESET/SET button to turn on the alarm.
2. The RESET/SET button and the Bed Exit Alarm icon will remain
illuminated to indicate that the bed exit alarm has been activated.
3. After a delay of several seconds an audible “chirp” will also sound to
indicate that the Bed Exit Alarm has been armed.
IMPORTANT: The Bed Exit Alarm can ONLY be armed if there is weight on the bed (20lbs [9.1kg] minimum). This is intended to prevent false arming of this feature that may otherwise become a nuisance to staff.
Bed Exit Alarm Disarming
To disarm the Bed Exit Alarm (Three Zone):
Press the RESET/SET button to scroll through the three alarm zones. When the RESET/SET button and no alarm zone icons are illuminated
the Bed Exit Alarm has been disarmed.
IMPORTANT: The Bed Exit Alarm can ONLY be disarmed using the RESET/SET button.
The factory default sensitivity level setting is Zone 1. The factory default alarm chime setting is audible chime (high volume) AND Nurse/Priority Call alarm. To change these settings and/or to
change the chime tone, refer to the following sections.
To disarm the Bed Exit Alarm (Single Zone):
Press the RESET/SET button to scroll through until the Bed Exit Alarm
icon is no longer illuminated.
When the RESET/SET button and no alarm zone icons are illuminated
the Bed Exit Alarm has been disarmed.
IMPORTANT: The Bed Exit Alarm can ONLY be disarmed using the RESET/SET button.
Page 81
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 81 of 159
Bed Exit Alarm Volume Selection
To set the desired Bed Exit Alarm volume, the Bed Exit Alarm must first be armed. Press the RESET/SET button to arm the Bed Exit Alarm.
Press the CHIME button to begin scrolling through the alarm volume
settings. There are three alarm volume settings (low, medium, high) and a Nurse/Priority Call setting.
Selecting alarm volume:
Scroll through the alarm volume settings one time will allow staff to
select the desired alarm volume setting OR the Nurse/Priority Call alarm that is appropriate.
Selecting alarm volume AND Nurse/Priority Call:
Scroll through the alarm volume settings a second time will allow
staff to select the desired alarm volume setting AND the Nurse/Priority Call alarm that is appropriate.
When the desired Bed Exit Alarm volume is selected, the CHIME button and the selected alarm volume and/or Nurse/Priority Call icon(s) will remain illuminated.
IMPORTANT: The Nurse Call and Priority Call alarms annunciated by the bed can be separated to meet the alarm notification needs of your healthcare facility. This setting can be changed in the Maintenance Menu.
Bed Exit Alarm Chime Tone Selection
From the BED EXIT Menu:
2. Press the UP/DOWN arrow buttons to scroll to the ALARM CHIME screen.
3. Press the ON/ENTER button to enter this menu. The current Bed Exit Alarm chime tone will be shown.
4. Press UP/DOWN to scroll through available Bed Exit Alarm chime tones. Each chime tone will sound.
5. Press ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that Bed Exit Alarm chime tone has been changed.
Short Cut Key:
Press-and-hold the Chime button for 3 seconds to
jump directly to this selection in the Bed Exit Menu.
Alarm Silencing
Zone 3 used for this example
High Volume AND Nurse/Priority Call used for this example
Once a Bed Exit Alarm has been triggered, it may be silenced by pressing the RESET/SET button.
Page 82
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 82 of 159
Once the Bed Exit Alarm has been silenced, the RESET/SET or CHIME button icons will flash simultaneously as a visual indicator and reminder to staff that the alarm has been silenced and must be either reset or deactivated.
These button icons will flash until the Bed Exit Alarm has been disarmed (by pressing the RESET/SET button or, once the Bed Exit Alarm has been automatically re-armed if this feature is ON.
Locking Bed Exit Alarm Settings
You may find that your healthcare facility tends to use the same Bed Exit Alarm settings for patients, say, in the same ward and/or requiring the same level of monitoring. If this is the case for your healthcare facility, it may be desirable to lock the Bed Exit Alarm settings.
Locking the Bed Exit Alarm settings will enable one-touch arming of the Bed Exit Alarm feature without the need to:
Scroll through and then select desired zone. Scroll through and then select if alarm chime is audible and if it includes Nurse/Priority Call.
IMPORTANT: For added flexibility, both of the Bed Exit Alarm settings can be locked or the zone and alarm chime settings can be lock separately.
IMPORTANT: Once the Bed Exit Alarm settings have been locked, any attempt to change them by pressing either the RESET/SET and/or CHIME buttons will not be permitted.
To lock the Bed Exit Alarm settings, the Bed Exit Alarm must first be armed. Press the RESET/SET button to arm the Bed Exit Alarm.
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the LOCK ZONE CHIME screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press UP/DOWN arrow buttons to scroll through available selections:
OFF ZONE LOCK CHIME LOCK ZONE + CHIME LOCK
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm zone and/or chime tone lock setting has been changed. Press the 0.0/ZERO button to “escape” to the home screen.
Short Cut Key:
Press-and-hold the SET/RESET and CHIME button
for 3 seconds to jump directly to this selection in the Bed Exit Menu.
Page 83
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 83 of 159
When a patient attempts to exit or does exit the bed (depending upon the zone level selected), the Bed Exit Alarm will be triggered. If the patient returns to their original position on the bed, the Bed Exit Alarm can be set to automatically stop alarming or otherwise announcing the alarm.
Bed Exit Alarm: Auto Reset Functionality
The Bed Exit Auto Reset feature will allow the bed to automatically reset the Bed Exit Alarm without staff needing to press the RESET/SET button after the patient has returned to the bed.
The Bed Exit Alarm will resume patient monitoring operation with the previous or locked alarm
settings (zone and volume/Nurse/Priority Call levels).
An audible “chirp” will sound to indicate that the Bed Exit Alarm has been reset.
The factory default setting for the Bed Exit Auto Reset is ON. This setting can be changed by following the instructions below.
IMPORTANT: If the Auto Reset feature has been set to OFF (disabled) in the Bed Exit Menu, the Bed Exit Alarm will continue to sound regardless if the patient has returned to the original position or has returned to bed (depending upon the zone level selected) and has ceased all motion.
To Enable/Disable Auto Reset
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the AUTO RESET screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press the UP/DOWN arrow buttons to scroll through available selections:
AUTO RESET ON AUTO RESET OFF
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm Auto Reset setting has been changed. Press
the 0.0/ZERO button to “escape” to the home screen.
4.5.1 Bed Exit Alarm: Auto Reset
Page 84
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 84 of 159
4.5.2 Bed Exit Alarm: Auto Re-Arm
When a patient attempts to exit or does exit the bed (depending upon the zone level selected), the Bed Exit Alarm will be triggered. Staff may silence the Bed Exit Alarm by pressing the RESET/SET button. Once in this silenced mode, the staff needs to remember to reset the Bed Exit Alarm (by pressing the RESET/SET button) after settling the patient and/or when the patient has returned to the bed.
Bed Exit Alarm: Auto Re-Arm Functionality
The Bed Exit Auto Re-Arm feature will allow the bed to automatically re-arm the Bed Exit Alarm without staff needing to press the RESET/SET button after the patient has returned to the bed.
The Bed Exit Alarm will re-arm itself after a selected time delay period has elapsed. The Auto Re-Arm
feature has three factory set time delay periods to select from: 1, 3, or 9 minute delay.
The Bed Exit Alarm will resume patient monitoring operation with the previous or locked alarm
settings (zone and volume/Nurse/Priority Call levels).
An audible “chirp” will sound to indicate that the Bed Exit Alarm has been re-armed.
The factory default setting for Bed Exit Auto Re-Arm is ON with a time delay period of 1 minute. This setting can be changed by following the instructions below.
To Enable/Disable Auto Re-Arm
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the AUTO RE-ARM screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press the UP/DOWN arrow buttons to scroll through available selections:
1 MINUTE DELAY MINUTE DELAY 9 MINUTE DELAY OFF
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm Auto Re-Arm setting has been changed. Press the 0.0/ZERO button to “escape” to the home screen.
If the patient has not returned to the bed after 10 minutes has elapsed then a notification will be displayed to remind staff that the Bed Exit Alarm is still armed but has been silenced.
Press the RESET/SET button to cancel the
Bed Exit Alarm.
The LCD display will automatically return to the home screen.
NOTE:
If the Auto Re-Arm feature has been set to OFF (disabled) in the Bed Exit Menu, these screens will not appear.
PRESS SET/RESET TO CANCEL BED EXIT ALARM
PATIENT HAS NOT RETURNED TO BED
Page 85
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 85 of 159
4.5.3 Bed Exit Alarm: Auto Arm
You may find that your healthcare facility tends to set Bed Exit Alarm for most patients. If this is the case for your healthcare facility, it may be desirable to use the Bed Exit Alarm Auto Arm feature.
Bed Exit Alarm: Auto Arm Functionality
The Bed Exit Alarm Auto Arm feature automatically arms the Bed Exit Alarm once a stable weight of 50lbs (22.7kg) or more has been detected by the bed for a period greater than one minute. The Auto Arm feature will use the factory default or locked Bed Exit Alarm settings for Zone and Volume.
The factory default setting for Bed Exit Auto Arm is OFF. This setting can be changed following the instructions below.
To Enable/Disable Auto Arm
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the AUTO ARM screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press UP/DOWN arrow buttons to scroll through available selections:
AUTO ARM ON AUTO ARM OFF
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm Auto Arm setting has been changed. Press the
0.0/ZERO button to “escape” to the home screen.
Bed Exit Alarm: Auto Arm Pause Functionality Maximum delay time: 300 seconds
During patient care, you may wish to temporarily override the Bed Exit Auto Arm feature. For example, if it may be necessary to temporarily position a patient on the bed for a period greater than one minute which will cause the Bed Exit Alarm to automatically arm. As expected any patient motion after the Bed Exit Alarm has automatically armed may trigger a nuisance alarm (depending on the Zone selected).
During such patient care, it is possible to temporarily delay or “pause” the Auto Arm feature for 60, 90 or 120
seconds. A countdown timer is displayed. During the final 10 seconds of this countdown a series of warning “chirps” will sound. While the countdown is in progress, the Bed Exit Alarm will not Auto Arm.
When the countdown is nearing its end, Auto Arm Pause can be extended (up to 5 additional minutes, or 300 seconds, may be added). While the Auto Arm Pause countdown is in progress, the countdown can be set to zero at any time, re-establishing the normal Auto Arm feature.
To set an Auto Arm Pause, the Auto Arm feature must first be enabled. To enable Bed Exit Alarm Auto Arm, refer to the previous section. Maximum delay time: 300 seconds
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to the AUTO ARM PAUSE screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press the UP/DOWN arrow buttons to add/remove delay time in 30 second increments.
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound and
the countdown will begin immediately
During the final 10 seconds of the delay countdown, a chirp will sound every second as notification that the Auto Arm Pause is about to end and that the Bed Exit Alarm Auto Arm functionality is about to resume.
Press the 0.0/ZERO button to “escape” to the home screen.
Page 86
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 86 of 159
4.5.4 Bed Exit Light
The underbed light can also be set to operate as an automatic Bed Exit Light. The Bed Exit Light improves nighttime visibility around the bed and in the room providing an increased level of patient safety. When the Bed Exit Light is on, the underbed light automatically switches on just prior to the patient exiting the bed. The underbed light will remain illuminated until the patient returns to the bed, and automatically switches off after a short delay. If the patient does not immediately return to bed, the underbed light automatically shuts off after 15 minutes has elapsed.
The factory default setting for the Bed Exit Light is ON. This setting can be changed by following the steps below.
To Enable/Disable Bed Exit Light
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the BED EXIT LIGHT screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press UP/DOWN arrow buttons to scroll through available selections:
BED EXIT LIGHT ON BED EXIT LIGHT OFF
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Light setting has been changed. Press the 0.0/ZERO
button to “escape” to the home screen.
When the Bed Exit Alarm is armed, the bed automatically records the number of times the Bed Exit Alarm has been triggered. This data is stored as the Bed Exit Alarm History.
To View/Reset Bed Exit Alarm History
From the BED EXIT Menu:
1. Press the UP/DOWN arrow buttons to scroll to the ALARM HISTORY screen.
2. Press the ON/ENTER button to enter this menu option. The BED EXIT ALARM HISTORY will display
the number of times that the Bed Exit Alarms has been triggered.
3. Press UP/DOWN arrow buttons to scroll through available selections:
DON’T RESET RESET NOW
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm History has been reset (if RESET NOW is selected). Press the 0.0/ZERO button to “escape” to the home screen.
4.5.5 Bed Exit Alarm History
Page 87
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 87 of 159
4.6 Watchdog™ Bed Monitoring System
If your Spirit Select™ bed has been equipped with the optional Watchdog Bed Monitoring System, the bed will be outfitted with a bumper light installed beneath the footboard in the center of the frame and a unique footboard staff control as shown below. The Watchdog Bed Monitoring System has the following modes of operation:
CAUTION
Always keep watchdog light free and clear of all obstructions. Ensure the watchdog light is always visible and not obstructed by linens, equipment, etc.
The Watchdog feature is used to monitor and indicate if a change has been made to the bed position or the status of important bed features. On bed power up, the bed defaults to “prepare bed for Watchdog arm” which allows the user 10 seconds (after last key press) to set the desired position of the bed. The Watchdog light will illuminate BLUE. The bed is now being monitored for bed position changes and important bed features.
A stationary bed is critical to patient safety. Therefore the status of the caster lock (brake) is continuously monitored. Whenever the caster lock (brake) is not engaged, the Watchdog lights will be illuminated Solid RED.
IMPORTANT: Continuous monitoring by this WATCHDOG mode is only possible when the bed is running on AC power (plugged into a wall outlet).
4.6.1 Always On™ Monitoring
Page 88
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 88 of 159
4.6.2 Bed Position/Status Monitoring
To adjust the position of the bed, simply disarm the Watchdog by pressing-and-holding the WATCHDOG key, then releasing the key when prompted. The LCD will display a countdown timer of 10 seconds. The timer will reset every time a change is made to the bed. Once all desired bed changes are made, simply allow the countdown timer to elapse. Watchdog Monitoring is now re-established to the new bed position.
To Adjust the Bed Position:
From the footboard staff control:
1. Press-and-hold the WATCHDOG button for
~4 seconds.
2. Release the WATCHDOG button when the
“PREPARE BED FOR WATCHDOG ARM 10s”
screen appears on the LCD display.
3. The Watchdog light will turn off and an
audible chirp will sound to indicate that the Bed is now ready to be repositioned. A countdown timer will be displayed on the LCD screen
4. Make all necessary changes to the bed as
required. The countdown timer will reset after each button press.
5. Once the desired changes have been made,
simply allow the countdown timer to elapse.
6. The Watchdog Light will illuminate BLUE after
the timer has elapsed.
PRESS-AND-HOLD FOR 4 SECONDS
Once the Bed Position/Status Monitoring has been auto-armed, any change to the bed (see following table) will be indicated by Watchdog lights turning yellow or red. The WATCHDOG button will also flash to indicate a change has occurred.
Right foot siderail change used for this example
To identify what has changed:
TAP (momentary press-and-release) the
WATCHDOG button. Changes will be displayed on the LCD screen.
Bed Position
Watchdog Lights Will Turn Yellow When…
Head Angle
Angle changes of approximately ± 5°
Knee Angle
Angle changes of approximately ± 3°
Bed Elevation
Height changes of approximately ± 3 inches
Bed Status
When…
Watchdog Lights Will Be…
Obstruction Sensors
An obstruction is present/detected
Flashing YELLOW
Patient Lock-Outs
Any change to settings has occurred
Solid YELLOW
Rail Locks
A change to any rail position has occurred
Solid YELLOW
RA I L R I GH T FO O T CH A N GE D 
WA T C HD O G AU T O -A R M ED
PR E P AR E B E D F O R WA T C HD O G A R M 1 0s
HO L D T O D I S ARM WA T C HD O G
Page 89
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 89 of 159
To View Bed Position/Status
At any time you can confirm Watchdog Bed Position/Status:
Tap (momentary press-and-release) the
WATCHDOG button. Confirmation screen will be displayed on the LCD screen.
To provide the best patient care possible, staff may wish to use the Nursing Reminder™ to identify if a particular patient/bed requires attention (example: patient requires time dependent repositioning or a bed needs a change of linens, etc.).
To Set the Nurse Reminder™:
Simply Press the REMINDER button. The LCD will display “NURSING REMINDER
OFF”
Use the MENU buttons to scroll through the
available time intervals
The Nursing Reminder can be set at 15
minute intervals from 1 minute to 3 hours maximum
When the desired time is display, Press the
ON|ENTER button to select
The Bumper lights will illuminate BLUE and an audible chirp will sound to indicate that the Nurse Reminder™ has been set.
NU R SIN G R E M IN D ER 1 h r 3 0 m i n 
NU R SIN G R E M IN D ER OFF 
BE D SE T T IN G S O K
4.6.3 Nurse Reminder™
Page 90
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 90 of 159
Nurse Call Trigger:
The Nursing Reminder can be set to trigger a nurse call (if connected) to a central nursing station. After the desired time is selected, the LCD display will ask the user if a nurse call is required.
Use the MENU buttons to select the desired
state
o ON o OFF
Press the ON|ENTER to confirm selection
Once the reminder countdown has expired, the Bumper lights will turn Solid PURPLE and the REMINDER button will flash.
IMPORTANT: A Nurse Call can only be sent if the bed is properly connected to a nurse call system.
Once the reminder countdown has expired, Nurse
Reminder™ will need to be cleared to turn off the
Bumper lights.
To Clear the Nurse Reminder™:
Press the REMINDER button. Set the Nurse Reminder to OFF and press the
ON|ENTER button
An audible chirp will sound to indicate that the Nurse Reminder™ has been cleared.
To Cancel Remaining Reminder Countdown
At any time you can cancel the Nurse Reminder™.
Press the REMINDER button. The “CANCEL?” screen will appear. The option “NO” will be displayed on screen Use the UP/DOWN arrow buttons to change
the option to “YES”
Press ON/ENTER button to select the desired
state.
An audible chirp will sound to indicate that the Nurse Reminder™ countdown has been cancelled.
NOTE:
While in the reminder countdown selection screen, use the UP/DOWN arrow buttons to set the reminder countdown to 0 minutes to clear the Nurse Reminder™.
CA N CE L ? E N T ER NO 
NU R SIN G R E M IN D ER OFF 
SEN D N U RS E CA L L ? YE S 
NU R SIN G R E M IN D ER CO M PLE T E
Page 91
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 91 of 159
4.6.4 Available Light Configurations
The Bumper, Watchdog and Underbed lights can be adjusted to enhance the room lighting. Light colour and intensity can be changed to various configurations.
SELECTION
SETTINGS
A. UNDERBED LIGHTS
OFF
LOW
MEDIUM
HIGH
B. BUMPER
LIGHTS
LT BLUE
LOW
WHITE
LOW
PURPLE
LOW
OFF
HIGH
HIGH
HIGH
C. WATCHDOG
LIGHTS
OFF
LOW
MEDIUM
HIGH
D. REMINDER
LIGHT
LOW
HIGH
E. BED EXIT
LIGHT
ON
OFF
To Adjust Light Configurations:
Press-and-hold the LIGHT button for 4
seconds. The “HOLD KEY FOR LIGHTS MENU” screen will appear.
Light Blue Nightlights used for this example
Release the LIGHT button when the “A.
UNDERBED LIGHT” screen appears on the LCD display.
Use the UP/DOWN arrow buttons to scroll
through the available Light Configurations.
Press the ON/ENTER button to ENTER the
desired selection.
Use the UP/DOWN arrow buttons to cycle
through the available lights settings.
The Bumpers, Watchdog or Underbed lights (depending on which selection menu) will change to match the current selection on the LCD display to allow the user to see the desired change before making the selection.
Press the ON|ENTER button to select the
desired light setting.
PRESS-AND-HOLD FOR 4 SECONDS
PRESS THE UP/DOWN ARROW BUTTONS
PRESS THE ON|ENTER BUTTON
NOTE:
The WATCHDOG feature is only operational when the bed is running on AC power (plugged into a wall outlet).
LT B LU E LO W
B. B UM P ER L I GH T S
A. UN D ER B ED LI G HT
HO L D K E Y F O R LI G HTS MEN U
Page 92
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 92 of 159
Menu Navigation
NOTE:
Depending upon the version of firmware that your Spirit Select™ bed is equipped with, the menus and menu options may have different titles and/or the menus may contain different and/or additional menu options.
Page 93
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 93 of 159
Menu Navigation Tip!
Press the 0.0/ZERO button at any time in any menu to “escape” back to the home screen.
Page 94
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 94 of 159
4.7 Scale Menu
A Scale Menu is available on Spirit Select™ beds. The Scale Menu includes menu options that are dirrectly related to scale operation and offer increased functionality of the bed.
The units that the patient weight is displayed in can be changed to suit the needs of your healthcare facility.
Selecting Patient Weight Display Units
From the Scale Menu:
1. Press the UP/DOWN arrow buttons to scroll to the SELECT LB/KG screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press UP/DOWN arrow buttons to scroll through available selections:
LB KG
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the LCD Display units have been selected. Press the 0.0/ZERO button to “escape” to the home screen.
During routine patient care, there may arise the need to add/remove equipment (medical devices, linens, etc) to/from the bed without disturbing the patient (ie: having to patient leave the bed). Any items added to or removed from the bed will impact the displayed patient weight.
Using the Change Equipment functions, staff can add required equipment to the bed or remove unneeded equipment from the bed at any time when the patient is on the bed.The displayed patient weight will not reflect this weight increase or decrease with the addition or removal of equipment to/from the bed when this procedure is performed.
Making an Equipment Change
From the Scale Menu:
1. Press the UP/DOWN arrow buttons to scroll to the CHANGE EQUIPMENT screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press the ON/ENTER button to begin the equipment change process. The bed will take a snapshot of
the current bed weight (patient, mattress, equipment, lines, etc.).
4. Perform desired equipment addition or removal.
5. Press the ON/ENTER button when done. The bed will take a second snapshot of the new bed weight.
The difference between the first and second weight snapshots will simply be ignored by the scale and will not be reflected in the displayed patient weight.
An audible chirp will sound to indicate that the equipment change has been successfully made. Press the
0.0/ZERO button to “escape” to the home screen.
HINT: Use this feature when adding additional pillows and/or linen to the bed.
4.7.1 Selecting Patient Weight Display Units
4.7.2 Changing Equipment
Page 95
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 95 of 159
4.7.3 Adjusting Patient Weight
Using the Adjust Patient Weight function, staff can adjust the displayed patient weight as necessary, at any time when the patient is on the bed, without disturbing the patient (ie: having to patient leave the bed).
The displayed patient weight will reflect a weight increase or decrease when this procedure is performed.
Making an Adjustment to Patient Weight
From the Scale Menu:
1. Press the UP/DOWN arrow buttons to scroll to the ADJUST WEIGHT screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press the ON/ENTER button to begin patient weight adjustment process.
4. Press the UP/DOWN Arrow buttons to make adjustments to patients displayed weight. Each time the
UP/DOWN Arrow buttons are pressed, an increase/decrease of 0.5 LB or 0.5 KG (depending upon which unit has been set for patient weight display) will register.
The UP Arrow button will increase the patient’s displayed weight. The DOWN Arrow button will decrease the patient’s displayed weight.
5. Press the ON/ENTER button to input the adjustment after the desired increase/decrease has been made.
An audible chirp will sound to indicate that the patient weight adjustment has been successfully made. Press the 0.0/ZERO button to “escape” to the home screen.
The patient display weight will now display the actual patient weight plus this inputted increase
OR
the
actual patient weight minus this inputted decrease.
IMPORTANT: To inform staff that the patient’s weight has been adjusted, “ADJ” will be displayed
next to the patient weight value on the home screen.
NOTE:
Zeroing the scale will override any adjustments made to patient weight.
During routine patient care, there may be the need to monitor changes in patient weight.
Using the View Weight Change function, staff can easily observe and record these changes. When the View
Weight Change function is set, the bed takes snapshots of the patient’s weight and stores this value as the
original patient weight. If the bed detects any change in patient weight, this changed amount will be displayed on the home screen.
Enable/Disable View Weight Change
From the Scale Menu:
1. Press the UP/DOWN arrow buttons to scroll to the VIEW WEIGHT CHANGE screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press the UP/DOWN arrow buttons to scroll between the available options:
OFF ON/RESET
An audible chirp will sound to indicate that the view weight change feature has been enabled. Press the
0.0/ZERO button to “escape” to the home screen.
Selecting ON/RESET will begin the VIEW WEIGHT CHANGE process. The scale will take a snapshot of the current bed weight (patient, mattress, equipment, lines, etc.) and use this value to compare patient weight changes.
PA T I EN T W E I GH T XX X .X L B A D J
4.7.4 View Weight Change
Page 96
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 96 of 159
Patient Weight GAIN used in this example
Any weight gain or loss is experienced by the patient will be displayed within parenthesis accompanied with either a “+” or “-” sign to indicate patient weight gain or loss respectively.
The home screen which will now sequentially flash
to display the original “snapshot” patient weigth
and the current, real time weight change that the patient has experienced.
SEQUENTIALLY FLASH
When the Weight Change Alarm function is set, the bed takes snapshots of the patient’s weight and stores this value. Weight change is a time related event and is assumed to occur gradually. If the difference in the
patient’s weight equals or exceeds the set amount after an elapsed period of an hour, a weight change
alarm is triggered.
If the difference is significantly greater than the set amount, the bed assumes that this occurrence is not related to patient weight change (example: a healthcare practitioner or family member is likely sitting on and/or something has been placed on the bed without using the “Change Equipment” feature – refer to previous section for instructions). In this case the bed simply ignores this change and a patient weight change alarm will not be triggered in this instance.
Patient Weight GAIN used in this example
When a weight change alarm is triggered, a warning message will be displayed and the bed will also send a priority call via the Nurse Call system (if connected). There is no audio component to this alarm.
Press the ON/ENTER button to cancel that the weight change alarm. The weight change alarm settings will also be cleared.
Setting a Weight Change Alarm
From the Scale Menu:
1. Press the UP/DOWN arrow buttons to scroll to the WEIGHT CHANGE ALARM screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press the UP/DOWN arrow buttons to set the WEIGHT CHANGE ALARM trigger weight GAIN/LOSS.
Each time the UP/DOWN Arrow buttons are pressed, an increase/decrease of 0.5 LB or 0.5 KG (depending upon which unit has been set for patient weight display) will register.
The UP Arrow button will increase the alarm trigger weight. The DOWN Arrow button will decrease the alarm trigger weight.
4. Press the ON/ENTER button when done. The scale will take a snapshot of the current patient weight
and store this value.
An audible chirp will sound to indicate that the WEIGHT CHANGE ALARM has been set. Press the 0.0/ZERO button to “escape” to the home screen.
If, after one hour elapses, the bed detects a weight change greater than the set alarm trigger weight, a weight change alarm will be triggered.
Press the ON/ENTER button to cancel the alarm. The Weight Change Alarm can be reset by following the steps above.
WEI G HT CH A N GE GA I N X X .X L B
PA T I EN T W E I GH T (+ X X. X LB )
PA T I EN T W E I GH T XX X .X L B
4.7.5 Weight Change Alarm
Page 97
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 97 of 159
4.7.6 Reset All
Reset All Settings – Impact
Performing a “Reset All”
WILL
impact the following
settings:
Performing a “Reset All”
WILL NOT
impact the
following settings:
Bed Exit History
– Cleared
Weight Change Alarm
– Cleared/Disabled
Weight Change View
– Disabled
Patient Adjust Weight
– Cleared
Equipment Change
– Cleared
Scale Calibration Angle Calibration Any Bed Exit Setting Obstruction Sensors Enabling/Disabling Nurse Call/Priority Call Selection Patient Weight Display Unit Selection LCD Display Language
Reset All Settings
From the SCALE Menu:
1. Press the UP/DOWN arrow buttons to scroll to the RESET ALL screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press UP/DOWN arrow buttons to scroll through available selections:
DON’T RESET RESET NOW
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Bed Exit Alarm History has been reset (if RESET NOW is selected). Press the 0.0/ZERO button to “escape” to the home screen.
Page 98
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 98 of 159
4.8 Service Menu
A Service Menu is available on Spirit Select™ beds. The Service Menu includes menu options that allow for the bed to be configured to best suit the needs of your healthcare facility. The Service Menu also includes menu options that are not typically accessed on a daily basis and/or are settings that are globally set for the healthcare facilty. Addtionally, in the rare event of component failure and/or malfunction, some bed features can be disabled in the Service Menu to restore bed functionality until the bed can be properly serviced.
The units that the patient weight is displayed in can be changed to suit the needs of your healthcare facility. This change can be easily made in the Scale Menu. See previous section for instructions.
Locking LCD Display Units
From the Maintenance Menu:
1. Press the UP/DOWN arrow buttons to scroll to the LOCK LB/KG screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press UP/DOWN arrow buttons to scroll through available selections:
LB KG
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the LCD Display units have been locked. Press the 0.0/ZERO button to “escape” to the home screen.
IMPORTANT: If the LCD Display units have been locked, the unit selection menu option WILL NOT appear in the Scale Menu.
The Nurse Call and Priority Call alarms annunciated by the bed can be separated to meet the alarm notification needs of your healthcare facility.
Setting Priority & Nurse Call
From the Maintenance Menu:
1. Press the UP/DOWN arrow buttons to scroll to the PRIORITY CALL screen.
2. Press the ON/ENTER button to enter this menu option.
3. Press UP/DOWN arrow buttons to scroll through available selections:
PRIORITY + NURSE CALL PRIORITY ONLY
4. Press the ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the Priority/Nurse Call settings have been changed. Press the
0.0/ZERO button to “escape” to the home screen.
NOTE:
Priority & Nurse Call settings can be further tailored. These adjustments can be done in the Maintenance Menu under the ADAVANCED CONFIGURATIONS menu option. Contact CHG Hospital Beds’ Customer Service at (519) 963-4010 or toll free 1-866-516-5446 for further assistance.
4.8.1 Locking LCD Display Units
4.8.2 Priority & Nurse Call Settings
Page 99
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 99 of 159
4.8.3 Obstruction Sensor Settings
In the rare event of component failure and/or malfunction, the obstruction sensor system can be disabled to restore bed functionality until the bed can be properly serviced
From the Maintenance Menu:
1. Press the UP/DOWN arrow buttons to scroll to the OBSTRUCTION SENSORS screen.
2. Press the ON/ENTER button to enter this menu option. The current selection will be shown.
3. Press UP/DOWN arrow buttons to scroll through available selections:
OBSTRUCTION SENSORS ENABLED OBSTRUCTION SENSORS DISABLED
4. Press ON/ENTER button to make desired selection.
An audible chirp will sound to indicate that the obstruction sensor operation has been changed. Press the
0.0/ZERO button to “escape” to the home screen. An OBSTRUCTION SENSOR DISABLED warning will now
be displayed.
On Spirit Plus™ beds the factor default setting for obstruction sensors is CONNECTED and OPERATING. This cannot be changed using the footboard staff control. The obstruction sensors can only
be enabled/disabled by connecting/disconnecting the obstruction sensor circuits from the CB10Lite circuit board. Refer to page 132 to identify the two connectors for the obstruction sensor circuits:
To enable/disable the obstruction sensors, connect/disconnect both of the obstruction sensor circuits
from the CB10Lite circuit board respectively.
Disabled Obstruction Sensor Warning
On Spirit Select™ beds, when the obstruction sensors have been disabled and a downward motion of the bed platform (using any of the following: BED DOWN, TREND, rev.TREND, LEVEL ALL, or CHAIR position) is attempted, a warning message will be displayed during bed operation to remind staff that the obstruction sensors have been disabled.
IMPORTANT: There is no warning or indicator on
Spirit™ and Spirit Plus™ beds. The bed will continue
to operate as though it was not equipped with the optional obstruction sensors.
WARNING
Disabling the obstruction sensors will disable a major safety device on the bed.
The obstruction sensors should only be disabled in the rare event that normal bed function has been impaired by the malfunction of a component in the obstruction sensing system.
WA R NI N G OB S TR UC T IO N SE N SO R S DI SA B LE D
WARNING OBSTRUCTION SENSORS DISABLED
Page 100
1-70-001 – R User Manual - Spirit Bed
© CHG Hospital Beds Inc.
Page 100 of 159
4.8.4 Head Angle Settings
Depending upon the preference of your healthcare facility, it is possible to set the origin for the head angle that is displayed on the second home page of the LCD display.
IMPORTANT: This setting impacts the displayed head angle when the bed is in a Trendelenburg, reversed Trendelenburg, or Chair position (ie: when the bed frame is not parallel with the floor).
When the bed frame is in the horizontal position, the bed frame is parallel with the floor. In this case, the head angle that is displayed would be the same if either setting was selected.
Displayed head angle when measured:
FROM FLOOR: 45° FROM FRAME: 45°
However in a Trendelenburg, reverse Trendelenburg, or Chair position the bed frame is no longer in the horizontal position and the bed frame is now at an angle to the floor. In this case, the head angle that is displayed would differ depending upon which setting was selected.
Bed in a 14° Trendelenburg angle used in this example
Displayed head angle when measured:
FROM FLOOR: 31° FROM FRAME: 45°
Displayed head angle when measured FROM FLOOR
Displayed head angle when measured FROM FRAME
HEA D TR E N D 31° 14°
HEA D TR E N D 45° 14°
HEA D TR E N D 45° 0°
HEA D TR E N D 30° 0°