Product provides terminal for connection of a potential equalization conductor. The potential
equalization conductor provides direct connection between the product and potential equalization
busbar of the electrical installation.
Protective earth ground
Protection from dripping water from above the device
Defibrillation proof type BF applied part
8001-109-002 Rev D.0EN
CCAAUUTTIIOONN -- Federal law (USA) restricts this device to sale by or on the order of a physician.
CCAAUUTTIIOONN -- Always use sterile distilled water or distilled water that has been passed through a
filter less than or equal to 0.22 microns with this product.
87VL Medical
Electrical Equipment
In accordance with European Directive 2012/19/EU on Waste Electrical and Electronic
Equipment, this symbol indicates that the product must not be disposed of as unsorted municipal
waste, but should be collected separately. Contact your local distributor for disposal information.
Medical Equipment Classified by Underwriters Laboratories Inc. with Respect to Electric Shock,
Fire, Mechanical and Other Specified Hazards Only in Accordance with IEC 60601-1:20 05 (3rd
edition), ANSI/AAMI ES60601-1 (2005, 3rd edition), CAN/CSA C22.2 No. 60601-1:20 08, IEC
80601-2-35:2009, CAN/CSA C22.2 NO 80601-2-35:12, ISO 80601-2-56:2009, CAN/CSA C22.2
NO 80601-2-56:12, IEC 60601-1-8:2007, CAN/CSA C22.2 NO 60601-1-8-08, IEC 60601-110:2008, CAN/CSA C22.2 NO 60601-1-10-09, IEC 60601-1-6, CAN/CSA-C22.2 No. 60601-16:11
Summary of safety precautions ................................................................................................................................5
Contact information .................................................................................................................................................6
Serial number location.............................................................................................................................................7
Expected service life ...............................................................................................................................................7
Theory of operation .................................................................................................................................................7
System modes........................................................................................................................................................7
System overview.....................................................................................................................................................8
Physiological Closed-Loop Control System (PCLCS) responses.................................................................................8
Specifications International ......................................................................................................................................9
Service icons ........................................................................................................................................................ 10
Service .................................................................................................................................................................... 12
Powering off the product ........................................................................................................................................ 12
Service log ........................................................................................................................................................... 12
System state ........................................................................................................................................................13
Protecting against Electrostatic Discharge (ESD) ....................................................................................................14
Water flow diagram ............................................................................................................................................... 15
Removing and replacing the reservoir..................................................................................................................... 15
Connecting and disconnecting thermal transfer devices ........................................................................................... 16
Disconnecting the insulated hoses .........................................................................................................................17
Back cover removal...............................................................................................................................................17
Front cover removal .............................................................................................................................................. 18
Side cover removal (left or right).............................................................................................................................18
Back cover replacement ........................................................................................................................................ 18
Battery removal and replacement ........................................................................................................................... 19
User interface replacement ....................................................................................................................................20
Front wheel replacement (left or right)..................................................................................................................... 20
Rear swivel lock caster replacement (left or right) .................................................................................................... 21
Power cord replacement........................................................................................................................................22
Valved panel mount female coupling ......................................................................................................................23
Drain valve body replacement ................................................................................................................................25
Condenser air filter replacement............................................................................................................................. 27
Refrigeration temperature probe replacement ......................................................................................................... 32
Thermal unit replacement ...................................................................................................................................... 33
Temperature probe replacement ............................................................................................................................34
Fan assembly, thermal unit replacement.................................................................................................................36
Main control board (MCB) replacement...................................................................................................................42
Temperature or Monitor probe jack replacement...................................................................................................... 44
System flow sensor replacement ............................................................................................................................49
Storing the controller ............................................................................................................................................. 57
Storing the power cord and hoses .......................................................................................................................... 57
Draining water from the controller and hoses........................................................................................................... 57
Draining the thermal transfer devices...................................................................................................................... 58
Draining water from the reservoir............................................................................................................................ 59
Test equipment .......................................................................................................................................................60
Functional test data sheet ...................................................................................................................................... 62
LCD functionality, testing the visual and audible alarms ........................................................................................... 63
High thermal cutout test ....................................................................................................................................65
Probe port resistance, flow and water temperature verification..................................................................................65
Quick Reference Replacement Parts .....................................................................................................................69
Remove from use code entry .................................................................................................................................75
Remove from use codes ........................................................................................................................................ 76
Main Control Board (MCB) assembly .....................................................................................................................83
Replaceable fuse ratings and type.......................................................................................................................... 88
Final assembly ........................................................................................................................................................ 89
Base assembly ..................................................................................................................................................... 107
Control assembly .................................................................................................................................................. 116
Power supply assembly ........................................................................................................................................ 120
Back cover assembly............................................................................................................................................143
EMC Information ................................................................................................................................................... 173
The words WWAARRNNIINNGG, CCAAUUTTIIOONN, and NNOOTTEE carry special meanings and should be carefully reviewed.
WWAARRNNIINNGG -- Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also
describe potential serious adverse reactions and safety hazards.
CCAAUUTTIIOONN -- Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or damage to the product or other property. This includes special care necessary for the safe
and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or
misuse.
NNoottee -- Provides special information to make maintenance easier or important instructions clearer.
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WWAARRNNIINNGG
• Always use Stryker accessories. Only IEC 60601-1 equipment shall be hooked to the patient temperature ports. Failure
to comply with these instructions may invalidate any or all warranties and may negatively affect the products EMC
performance. This also protects the product from cardiac defibrillation.
• Avoid reduction in water flow. Do not connect two or more thermal transfer devices in a series on a single port.
CCAAUUTTIIOONN
• Federal law (USA) restricts this device to sale by or on the order of a physician.
• Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
• Do not place your fingers in between the reservoir and the sides of the controller, to avoid the risk of pinching your
fingers.
• Avoid the use of materials of good thermal conductivity, such as water, gel, or similar substances, with the AAllttrriixx system
not powered on. This can decrease the temperature of the body of a patient.
• Do not apply thermal transfer devices to patients with ischemic limbs. This may result in harm to the patient.
• Do not use this product if the patient has a transdermal medication (patch) as this can result in increased drug delivery.
• Do not use three or more adult Mul-T-Blanket products at the same time to avoid the risk of water overflow when you
power off the controller.
• Always pre-fill the thermal transfer devices with sterile distilled water before you apply it to the patients. This is to reduce
the risk of pressure ulcers.
• Always clamp the hoses when disconnecting the thermal transfer devices.
• Always wear gloves when replacing the battery in case of battery leak.
• Do not put the unit on its side. You will not be able to put the product back into service for 24 hours.
• Do not store the product with water in the device.
• Always store the product within the specified environmental condition values.
• Do not hang items on the controller handle to avoid the risk of tipping the product.
• Always store the power cord, cables, and hoses before you transport the product to reduce the risk of trip hazard.
NNoottee -- Disinfection of the Altrix internal water system was validated using
8001-109-002 Rev D.05EN
M. mucogenicum.
IInnttrroodduuccttiioonn
This manual assists you with the maintenance of your Stryker product. Read this manual before maintaining this product.
Set methods and procedures to educate and train your staff on the safe maintenance of this product.
CCAAUUTTIIOONN -- Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
NNoottee -- Stryker continually seeks advancements in product design and quality. This manual contains the most current
product information available at the time of release. There may be minor discrepancies between your product and this
manual. If you have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770.
PPrroodduucctt ddeessccrriippttiioonn
The Stryker model 8001AAllttrriixx® Precision Temperature Management System can supply water to an individual or multiple
thermal transfer devices simultaneously with each of these circuits monitored separately. Three operating modes are
available to ease patient care: Automatic, Manual, and Monitor. The controller uses the patient temperature probe to
provide closed loop feedback for automatic patient temperature management and monitoring. The controller alarms visual
and audible indications for when safety parameters are exceeded or it detects system function or performance irregularities.
The AAllttrriixx system is able to provide a patient temperature output reference signal to be connected to a non-specific third
party device or system.
The controller regulates water temperatures between 4.0° C (39.2° F) and 40.0° C (104.0° F) and circulates the heated or
cooled water via hose sets through the thermal transfer devices. A graphical display provides the user an interface for
selecting desired water or patient temperature settings, operating modes, help menus, and other key parameters. Visual
indicators are displayed to inform the user of system status or when the user must confirm a setting selection. The system’s
water temperature and flow outputs can be monitored with 400 series compatible devices to optimize system operation.
The AAllttrriixx system includes the following components:
•controller
•reusable hose sets
•thermal transfer devices (blankets, vests, and leg wraps)
•patient temperature probes
•reusable adapter cables
•reusable patient temperature output cable
NNoottee -- The blankets, vests, leg wraps, and patient temperature probes are type BF applied parts.
CCoonnttaacctt iinnffoorrmmaattiioonn
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
To view your operations or maintenance manual online, see https://techweb.stryker.com/.
Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support.
Include the serial number in all written communication.
EN68001-109-002 Rev D.0
SSeerriiaall nnuummbbeerr llooccaattiioonn
A
EExxppeecctteedd sseerrvviiccee lliiffee
The AAllttrriixx controller has a five year expected service life under normal use, conditions, and with appropriate periodic
maintenance. See the maintenance manual for preventive maintenance and service information.
TThheeoorryy ooff ooppeerraattiioonn
The temperature management system circulates temperature controlled water via centrifugal pump through a configured
combination of hoses connected to single or multiple thermal transfer devices which can be applied around, placed over or
under a patient. Using a heater to increase water temperature and a refrigeration system to decrease water temperature,
delivery and removal of thermal energy to and from the patient occurs at the contact point of the patient skin and thermal
transfer device to raise or lower patient temperature. A patient temperature probe can also be used to provide closed-loop
feedback patient temperature control capability or to monitor patient temperature.
SSyysstteemm mmooddeess
The AAllttrriixx system has eight modes of operation. This section contains general descriptions of each of the modes.
MMooddeeDDeessccrriippttiioonn
SleepInitial mode when the product is plugged into a Mains AC. System is active but not yet
available for any therapy or user input other than to turn the product on into Standby Mode
or Maintenance Mode.
PreparationThe system checks different subsystem statuses to make sure that it is ready to start
therapy. This mode begins once Sleep Mode initiates, continues, and concludes before the
system can move from Sleep to any other Mode.
Stand-byShows that the product is ready for therapy and presents the operator with the choice of
starting one of the active therapies.
MaintenanceAllows the maintenance technician the ability to perform the following functions:
•Review and clear the latest RFU code
•Review the controller software versions
•Test visual and audible indicators
•Review the service log
•Monitor the state of the system sensors
Remove from use (RFU)The product enters this mode when there is a problem with the system providing adequate
therapy. The operator will not be able to perform therapy and may be instructed to remove
the product from use for qualified service personnel to evaluate.
8001-109-002 Rev D.07EN
TThheerraappyy mmooddeeDDeessccrriippttiioonn
ManualIn this mode, the product will provide thermally treated water to the patient through the
thermal transfer device as it is targeting operator input water temperature control setting. In
this mode, the operator determines what the water temperature should be and as a result,
a patient temperature probe is not necessary to have. No rate selection is necessary as the
water controller will always use the maximum cooling/warming capacity of the product to
achieve the final Water Temperature Control Setting. During Manual Mode the AAllttrriixx
controller will utilize the pump to regulate the water flow to a prescribed target flow in order
to obtain maximum heat transfer through the thermal transfer device(s). The water
controller will also utilize the heater cartridge and refrigeration product to regulate the
Water Temperature at the supply manifold to the Water Temperature Control Setting
Automatic
MonitorThe patient's temperature is monitored through a patient temperature probe. No thermally
In this mode, the product will provide thermally treated water to the patient through thermal
transfer device(s) as it is targeting a user input patient temperature control setting. In this
mode, the CS determines what the water temperature should be according to the therapy
rate chosen and what it takes to reach the desired patient temperature control setting. A
patient temperature probe is necessary to have in this mode.
The AAllttrriixx water controller provides an Automatic Mode to allow the caregiver to monitor
and control the core body Patient Temperature. The AAllttrriixx water controller requires the
caregiver to select the Final Patient Target Temperature and to select the warming or
cooling rate.
In Automatic Mode, the AAllttrriixx water controller will utilize the pump to regulate the water
flow to obtain maximum heat transfer through the thermal transfer device(s). The water
controller will also utilize the heater cartridge and refrigeration product to regulate the
Primary Patient Temperature to the selected Final Patient Target Temperature by
manipulating the water temperature of the supply ports.
The AAllttrriixx water controller offers 3 cooling rates and 4 warming rates.
treated water is being provided to the patient through thermal transfer devices.
SSyysstteemm oovveerrvviieeww
The AAllttrriixx system includes the main controller, fluid management controller and a basic user interface controller. The
system gathers input from the Human machine interface (HMI) system to determine the correct mode for the system.
The control system receives input from the following systems.
Display measurement accuracy±0.2° C (4.0° - 40.0° C)
Display / resolution setting0.1° C
Default setting104.0° F40.0° C
PPaattiieenntt tteemmppeerraattuurree
Control setting range89.6° - 100.4° F32.0° - 38.0° C
Control accuracy±0.1° C (32° - 38°C)
Measurement accuracy±0.3° C (25.0° - 45.0° C)
±0.4° C (0° C - 24.9° C, 45.1° C - 50° C)
Display / resolution setting0.1° C
Display range32.0° - 122° F0.0° - 50° C
Default setting98.6° F37.0° C
8001-109-002 Rev D.09EN
CCoonnttrroolllleerr
89.9°F
(32.2°C)
59°F
(15°C)
140 °F
(60 °C)
-40°F
(-40°C)
140 °F
(60 °C)
-20°F
(-29°C)
75 %
30 %
95 %
10 %
95 %
25 %
1060 hPa
700 hPa
1060 hPa
500 hPa
Heater capacity, maximum500 watts
Circulating fluidSterile distilled water or water that has been passed through a filter less than or
equal to 0.22 microns with this product
Battery9V Lithium
Alarm tone range75 - 85 dBA per standard IEC 60601-1-8
Water flow rate in each hose portTypical 1.2 lpm
Refrigerant typeR134a
Power cord length14 to 15 feet4.2 - 4.5 meters
Clinical thermometerDirect mode
Equipment ClassClass I
Rated for continuous operation
NNoottee -- The controller takes approximately 9 minutes to heat from 23.0±2° C (73.4° F) to 37.0° C (98.6° F) when not
connected to a patient. Time will vary when connected to a patient.
Stryker reserves the right to change specifications without notice.
For more information about thermal transfer devices, cables, or probes, see the manufacturer’s instructions for use.
LCD functionality, testing the visual and audible alarms
(page 63). Complete steps 1 though 4.
FFiigguurree 55 –– SSeerrvviiccee lloogg
NNoottee -- In the upper right hand corner of the service log screen (Figure 5), xx is the current page number The yy will
display the total number of service log pages.
SSyysstteemm ssttaattee
1. To gain access, see
2. The system state. (Figure 6 sample screen)
CCoooollStarts the cooling sub-function of the System State maintenance function
HHeeaattStarts the heating sub-function of the System State maintenance function
SSttooppStops the warming or cooling sub-function of the System State maintenance function
LCD functionality, testing the visual and audible alarms
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
NNoottee -- Always ship back circuit boards to Stryker in the same antistatic bags that the new boards were originally shipped in.
The electronic circuits in the product are completely protected from static electricity damage when factory assembled.
Always use adequate static protection when servicing the electronic systems of the product. All service personnel must use
static protection whenever they are touching wires.
The removable reservoir enables you to fill or drain the reservoir away from the controller without interrupting therapy. You
will need to have the reservoir installed before starting a therapy.
CCAAUUTTIIOONN -- Do not place your fingers in between the reservoir and the sides of the controller, to avoid the risk of pinching
your fingers.
To remove the reservoir, pull forward at an angle, and lift out the reservoir (Figure 8).
Read the operations manual for the individual thermal transfer devices for warnings, cautions, and safe operating
instructions before use.
WWAARRNNIINNGG
• Always use Stryker accessories. Only IEC 60601-1 equipment shall be hooked to the patient temperature ports. Failure
to comply with these instructions may invalidate any or all warranties and may negatively affect the products EMC
performance. This also protects the product from cardiac defibrillation.
• Avoid reduction in water flow. Do not connect two or more thermal transfer devices in a series on a single port.
CCAAUUTTIIOONN
• Avoid the use of materials of good thermal conductivity, such as water, gel, or similar substances, with the AAllttrriixx system
not powered on. This can decrease the temperature of the body of a patient.
• Do not apply thermal transfer devices to patients with ischemic limbs. This may result in harm to the patient.
• Do not use this product if the patient has a transdermal medication (patch) as this can result in increased drug delivery.
• Do not use three or more adult Mul-T-Blanket products at the same time to avoid the risk of water overflow when you
power off the controller.
• Always pre-fill the thermal transfer devices with sterile distilled water before you apply it to the patients. This is to reduce
the risk of pressure ulcers.
• Always clamp the hoses when disconnecting the thermal transfer devices.
To connect or disconnect the CClliikk--TTiittee® connectors (Figure 9) to the insulated hoses.
EN168001-109-002 Rev D.0
FFiigguurree 99 –– CClliikk--TTiittee
A
To connect or disconnect the Colder style (Figure 10) to the insulated hoses.
To connect the AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt WWrraapp™™ to the AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt HHoossee™™ see
Figure 12.
Press the button (A) on the hose to disconnect from the wrap.
The AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt WWrraappss have an internal valve and do not need a clamp to stop the water flow.
Draining the thermal transfer devices
FFiigguurree 1111 –– HHoossee ccllaammppss
(page 58).
NNoottee -- The term “thermal transfer devices” is used throughout this manual and is interchangeable with blankets and wraps,
unless indicated otherwise.
3. Using a T27 Torx, remove and save the two screws at the top of the display assembly screen that secures the display
assembly screen to the product.
4. Push forward on the display assembly screen to unseat it from the top bracket.
5. Using wire cutters, cut the cable tie that secures the communication cable to the display assembly.
6. Unscrew the two thumbscrews.
7. Unplug communication cable and remove display assembly.
Protecting against Electrostatic Discharge (ESD)
Back cover removal
(page 17).
(page 14).
NNoottee -- Complete the following procedures with the display assembly removed from the product. Face display assembly
down onto an ESD controlled workbench.
8. Using T20 torx remove and save the eight screws(C) (
9. Using a 3/8” socket remove and save the hex nut (A) that secures the display bracket, and remove the bracket.
10.Using a T20 torx remove and save the two screws (D) that secure the Controller Communications Board (CCB) (R) to
the User Interface.
11.Pull up on CCB to unplug it from the User Interface, set aside for reinstallation.
12.Remove the User Interface from the Bezel and discard.
NNoottee -- Do not dispose of as unsorted municipal waste. See your local distributor for return or collection systems
available in your country.
13.Reverse steps to reinstall.
14.Perform
sheet.
15.Verify proper operation before you return the product to service.
LCD functionality, testing the visual and audible alarms
Display assembly
(page 63). Check complete on the functional test data
CCAAUUTTIIOONN -- Do not put the unit on its side. You will not be able to put the product back into service for 24 hours.
TToooollss rreeqquuiirreedd::
•9/16” Combination wrench
•2” x 4” Board
PPrroocceedduurree::
1. Remove the back cover. See
2. Slightly tip the product and place a 2” x 4” board under the front corner of the frame on the side you are replacing the
rear swivel caster.
3. Using a 9/16” combination wrench, loosen and remove the nut that secures the rear swivel lock caster to the frame.
Remove and discard the rear swivel lock caster (Figure 14).
7. Using a T27 torx, remove and save the four screws (C) on the back of the control assembly. Pull the cover towards the
back of the product, set aside (Figure 31).
1. Remove the Manifold assembly from the product. See Manifold assembly replacement (Removal only).
2. Once the manifold assembly is removed, using a T25 torx remove the two screws (B) securing the outer manifold
bracket to the MCB support bracket (Figure 40).
6. Unscrew the two thumbscrews that secure the display cable to the control assembly.
7. Unscrew the two thumbscrews that secure the pump cable to the control assembly.
8. Disconnect the two lines of tubing to the control assembly, utilize the quick disconnects on the patient left side of the
product.
9. Using a 7/16” socket, remove and save the two lock nuts on the back of the control assembly.
10.Using a 727 torx, remove and save the two screws that secure the control assembly to the side panels.
11.Lift up and pull out the control assembly, removing the control assembly from the front of the product.
Side cover removal (left or right)
Draining water from the controller and hoses
Back cover removal
Front cover removal
(page 17).
(page 18).
(page 18).
(page 57).
NNoottee
• Guide the control assembly tubing while you remove it from the unit.
• Complete the following procedures with the control assembly removed from the product.
12.Using a T27 torx, remove and save the four screws that secure the control box cover.
13.Using a T27 torx, remove and save the screw that secures the power supply bracket to the control assembly bracket.
14.Disconnect the ten connectors from Main Control Board (MCB) labeled E, F, G, H, J, M, N, L, AG, and AH.
15.Using a T27 torx, remove and save the two screws (B) (Figure 43) that secure the MCB bracket to control assembly
bracket.
8001-109-002 Rev D.051EN
Torque item B
to 25 ± 5 in-lb
Torque item D
to 25 ± 5 in-lb
FFiigguurree 4433 –– MMCCBB bbrraacckkeett
16.Using 7/16” socket loosen the two nuts(D) that secure the MCB bracket to the control assembly bracket.
17.Using a 7/16” socket remove and save the two nuts that secure the MCB bracket to the power supply bracket. Move the
power supply bracket out of your way.
18.Lift up on the underside of MCB bracket. Move the MCB bracket towards the back end of the assembly out of your way.
19.Using a 5/32 Allen, remove and save the four screws (C) that secures the return manifold assembly (N) to the supply
manifold assembly (M). (Figure 44)
EN528001-109-002 Rev D.0
Torque item C
to 10 ± 2 in-lb
Torque item C
to 10 ± 2 in-lb
FFiigguurree 4444 –– MMaanniiffoollddss
20.Separate the supply manifold assembly (M) from the return and bypass manifolds (P).
NNoottee -- Verify all 3 Manifold Flow Sensors remain seated in the Return Manifold.
21.Discard supply manifold.
22.Reverse the steps to reinstall.
NNoottee -- Verify O-ring (E) are seated properly between the Supply and Bypass manifolds during reinstall.
23.Check for leaks.
24.Perform all test procedures. Complete the
25.Verify proper operation before you return the product to service.
6. Unscrew the two thumbscrews that secure the display cable to the control assembly.
7. Unscrew the two thumbscrews that secure the pump cable to the control assembly.
8001-109-002 Rev D.053EN
Side cover removal (left or right)
Draining water from the controller and hoses
Back cover removal
Front cover removal
(page 17).
(page 18).
(page 18).
(page 57).
8. Disconnect the two lines of tubing to the control assembly, utilize the quick disconnects on the patient left side of the
Torque item B
to 25 ± 5 in-lb
Torque item D
to 25 ± 5 in-lb
product.
9. Using a 7/16” socket, remove and save the two lock nuts on the back of the control assembly.
10.Using a 727 torx, remove and save the two screws that secure the control assembly to the side panels.
11.Lift up and pull out the control assembly, removing the control assembly from the front of the product.
NNoottee
• Guide the control assembly tubing while you remove it from the unit.
• Complete the following procedures with the control assembly removed from the product.
12.Using a T27 torx, remove and save the four screws that secure the control box cover.
13.Using a T27 torx, remove and save the screw that secures the power supply bracket to the control assembly bracket.
14.Disconnect the ten connectors from Main Control Board (MCB) labeled E, F, G, H, J, M, N, L, AG, and AH.
15.Using a T27 torx, remove and save the two screws (B) (Figure 45) that secure the MCB bracket to control assembly
bracket.
FFiigguurree 4455 –– MMCCBB bbrraacckkeett
16.Using 7/16” socket loosen the two nuts (D) that secure the MCB bracket to the control assembly bracket.
17.Using a 7/16” socket remove and save the two nuts that secure the MCB bracket to the power supply bracket. Move the
power supply bracket out of your way.
18.Lift up on the underside of MCB bracket. Move the MCB bracket towards the back end of the assembly out of your way.
19.Using a crescent wrench, remove and save the six hose connectors (G) and O-rings (E) (Figure 46).
20.Disconnect all three Manifold Flow Sensors labeled Port 1, 2, and 3.
21.Remove and discard the manifold block.
NNoottee -- Although the system was drained of water residual water may be present use CAUTION when removing to make
sure that water does not come in contact with electrical boards.
22.Using a 5/32” Allen remove and save the six screws (C) that secure the Return Manifold to the Bypass and Supply
Manifolds (Figure 47).
FFiigguurree 4477 –– MMaanniiffoollddss
23.Separate the return manifold assembly (N) from the supply and bypass manifolds (M and P).
24.Reverse the steps to reinstall.
8001-109-002 Rev D.055EN
NNoottee -- Verify O-ring (E) is seated properly between Return and Bypass manifolds during reinstall.
25.Check for leaks.
26.Perform all test procedures. Complete the
27.Verify proper operation before you return the product to service.
Functional test data sheet
(page 62).
EN568001-109-002 Rev D.0
SSttoorraaggee
SSttoorriinngg tthhee ccoonnttrroolllleerr
Storage is equal to or greater than 7 days without use.
CCAAUUTTIIOONN
• Do not store the product with water in the device.
• Always store the product within the specified environmental condition values.
To store the controller refer to the operations manual instructions.
Read the manufacturer’s operations manual for the individual thermal transfer devices (blankets and wraps) for warnings,
cautions, and safe operating instructions before use. Make sure that you drain the hoses before you put them in storage.
1. Unplug the product.
2. Remove the thermal transfer device from the patient.
3. Open the clamps on the hoses and thermal transfer devices, if applicable. See Figure 11.
4. Raise the thermal transfer devices attached to the hose above the ports on the controller. Gravity helps to drain the
water into the controller.
5. Allow most of the water to drain back into the controller. (Approximately 10 minutes).
6. See
7. See
Connecting and disconnecting thermal transfer devices
Disconnecting the insulated hoses
(page 17).
(page 16).
EN588001-109-002 Rev D.0
8. See
9. Discard the disposable thermal transfer devices based on your local waste management protocol.
Storing the power cord and hoses
a. Discard the disposable thermal transfer devices based on your local waste management protocol, if applicable.
•Safety Analyzer (Ground Resistance and Current leakage test)
•One gallon sterile distilled water
EN608001-109-002 Rev D.0
PPrreevveennttiivvee mmaaiinntteennaannccee
Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified service technician.
NNoottee -- Complete functional checks as indicated in the maintenance manual.
Inspect all of the following items:
All fasteners secure
Power cord and plug for fraying
Condition of covers and push handle for damage
Hose ports are operational
Ground chain attached
LCD is not cracked
LCD operational
Touch screen operational
Wheels for smooth operation
Rear caster wheels for free swivel action
Both rear wheels lock secure when applied
Front and rear wheels are not loose or wobbly
Battery backup, functional check
Alarm system - visual and audible, functional check
High temp cutout checks
Water temperature and flow verification, functional check
Probe resistance, functional check
Clear RFU codes
Ground impedance not more than 100mΩ (millohms)
Current leakage not more than 300 (microamps)
Integrity of all clamps and clamped joints located in the air elimination circuit.
Replace the following on an annual basis:
9 V battery
Condenser inlet filter
Air eliminator hose
Inlet filter
Product Serial Number:
Completed by:
Date:
8001-109-002 Rev D.061EN
FFuunnccttiioonnaall tteesstt ddaattaa sshheeeett
Serial #:
Functional checksCheck complete
LCD functionality, visual, and audible alarms
High temp cutout check
Probe resistance verificationRecord value
Probe A resistance1355 ohms37.0° C +/- 0.3° C (98.6° F +/- 0.5° F)
1471 ohms35.0° C +/- 0.3° C (95.0° F +/- 0.5° F)
1667 ohms32.0° C +/- 0.3° C (89.6° F +/- 0.5° F)
Probe B resistance1355 ohms37.0° C +/- 0.3° C (98.6° F +/- 0.5° F)
1471 ohms35.0° C +/- 0.3° C (95.0° F +/- 0.5° F)
1667 ohms32.0° C +/- 0.3° C (89.6° F +/- 0.5° F)
Flow and water temperature verificationRecord value
Before you place the product into service, make sure that the visual and audible alarms are functioning.
Stand-by
Back
Next or more
NNoottee -- CClliikk--TTiittee, colder style hoses with adaptor hose, or loop back hoses are required to run this procedure.
1. Make sure that the controller is in Stand-by.
2. At the same time, press and hold the SSttaanndd--bbyy button and the BBaacckk button.
3. Once Stryker is displayed on the screen, continue holding the SSttaanndd--bbyy button. Release the back button and press the
NNeexxtt oorr mmoorree button.
4. When you see PPrreeppaarraattiioonn iinn pprrooggrreessss...... , release the buttons.
5. Tap the nneexxtt oorr mmoorree button displayed on the screen under MMoorree.
6. Tap the VViissuuaall//AAuuddiibbllee icon.
7. Tap Confirm.
8. The controller will run through the LCD and alarm process.
a. Verify the alarm sounds.
b. Verify the green, yellow, and white indicators illuminate, mirroring the display. (See Figure 51, Figure 52, and Figure
53).
c. Verify the fluid controller light test illuminates, mirroring the display. (See Figure 54).
5. Verify the reservoir has at least a 2L of sterile distilled water.
6. Plug the product into a wall outlet.
7. Make sure that the controller is in SSttaanndd--bbyy.
8. At the same time, press and hold the SSttaanndd--bbyy button and the BBaacckk button.
9. Once Stryker is displayed on the screen, continue holding the SSttaanndd--bbyy button. Release the back button and press the
NNeexxtt oorr mmoorree button.
10.When you see PPrreeppaarraattiioonn iinn pprrooggrreessss...... , release the buttons.
11.Tap system state, press confirm.
12.Using RS-201 resistance box (or equivalent) set resistance to 1355 ohms.
13.Connect RS-201 resistance box with resistance cable attached to port probe A.
14.Verify that Patient A reads 98.6° F ± 0.5° F (37.0° C ±0.3° C).
15.Change the resistance on the box to 1471 ohms.
16.Verify that Patient A reads 95.0° F ±0.5° F (35.0° C ± 0.3° C).
17.Change the resistance on the box to 1667 ohms.
18.Verify that patient A reads 89.6° F ±0.5° F (32.0° C ± 0.3° C).
19.Connect RS-201 resistance box to port probe B.
20.Repeat steps 12-18 on port probe B, patient B.
21.Remove RS-201 and resistance cable.
22.Make sure that the flow has stabilized for more than five minutes.
23.Using the reference flow meter verify the displayed flow on the product reads within +/- 0.2 l/m and a minimum flow of
0.8 l/m for Port 1.
EN668001-109-002 Rev D.0
24.Verify the manifold water temperature reads within +/- 0.2° C (+/- 0.4° F) on the thermometer for Port 1.
25.Move the flow and temperature tool to Port 2 and repeat steps 22-24.
26.Move the flow and temperature tool to Port 3 and repeat steps 22-24.
27.Unplug the product and remove the test equipment.
These instructions are for the replacement of the condenser inlet filter, air eliminator hose assembly, and 9 volt battery
replacement as indicated on the preventive maintenance list.
NNoottee -- Order part number 8001-700-000 (PM kit) to receive the parts required for the annual replacements.
TToooollss rreeqquuiirreedd::
•T27 Torx
•T25 Torx
•7/16” Socket
•3/8” Drive ratchet
•Needle nose pliers
•Wire cutters
PPrroocceedduurree::
1. Apply the wheel locks.
2. Unplug the power cord.
3. Using a T27 Torx, remove and save the six screws that secures the back cover.
4. Remove and save the back cover by pulling outward and downward on the bottom of the back cover.
NNoottee -- The power cord will remain attached to the controller. Allow the power cord to slide through the back cover.
5. Remove and save the water reservoir.
6. Using a T27 Torx, remove and save the two screws inside the water catch tray that secure the bumper.
7. Remove and save the bumper by lifting up on the left corner of the tray while you push in on the front cover.
8. Using a T27 Torx, remove and save the top two screws inside the storage compartment that secures the front cover.
9. Using a T25 Torx, remove the three screws between the connection ports.
10.Using a 7/16” socket, remove and save the two nuts that secure the front cover to the product.
11.Unscrew the drain knob turning it counterclockwise, remove and save.
12.Pull outward on the bottom of the front cover so it lowers down. Remove and save the front cover.
13.Using a T25 Torx, remove and save the nine screws that secure the right side cover to the main frame.
NNoottee -- Removal of the front cover allows you to gain access to the screws securing the right side cover.
14.Remove the side cover, by rotating forward on the top and lifting upward. Now tip the side cover toward the back,
remove and save the cover.
15.Complete the following.
a.
Condenser inlet filter replacement
b.
9 volt battery replacement
c.
Air eliminator hose assembly replacement
16.Reverse the steps in this instruction to reinstall the covers.
17.Verify proper operation before you return the product to service.
1.1. If yes, hold Power Icon for at least 2 seconds.
1.1.1. Check for proper connection or damage
with the top ribbon cable on the back of
the User interface assembly. If ok, see
(User Interface Replacement in
Maintenance Manual)
1.2. If no, see
2. Verify proper display cable connection at display
assembly connector D and Control Box assembly
connector W.
3. Verify 24VDC to User Interface Board (UIB) LED is
green lit on FMB.
3.1. If present, go to step 4.
3.2. If not present, see
4. Verify 24VDC LED is present on Controller
Communications Board. (CCB)
4.1. If not present, replace Display cable assembly.
4.2. If present, go to step 5.
5. Verify CPU Stat is flashing on CCB.
6. Verify 3.3VDC and 15VDC test points on User Interface
Board.
•F7: Schurter Inc. 0034.3127 5x20 T-lag 10A 250V (8001-007-301)
EN888001-109-002 Rev D.0
FFiinnaall aasssseemmbbllyy
Torque item AE
to 8-12 in-lb
8001-103-010 Rev AB 100V (Reference only)
8001-123-010 Rev AB 120V (Reference only)
8001-223-010 Rev AB 220V (Reference only)
8001-233-010 Rev AB 230V (Reference only)
8001-109-002 Rev D.089EN
Torque item A
to 10 ± 2 in-lb
Torque item AC
to 10 ± 2 in-lb
Torque item C
to 10 ± 2 in-lb
Cable tie item J
pull force
33 ± 7 lb
EN908001-109-002 Rev D.0
Torque item C
to 10 ± 2 in-lb
Torque item B
to 10 ± 2 in-lb
Torque item C
to 10 ± 2 in-lb
Torque item E
to 25 ± 5 in-lb
8001-109-002 Rev D.091EN
Torque item B
to 10 ± 2 in-lb
Torque item AF
to 45 ± 5 in-lb
EN928001-109-002 Rev D.0
IItteemmNNuummbbeerrNNaammeeQQuuaannttiittyy
A0004-388-000Button head cap screw2
B0004-588-000Button head cap screw4
C0004-634-000Button head cap screw6
E0016-028-000Fiberlock hex nut2
F0034-022-000Cable clamp3
G1040-010-092Scrulok assembly2
H1040-010-008Cord ring, black2
J3000-300-1138" cable ties2
K8001-000-140