Thank you for choosing the DFC Force Controller.
This Quick Reference Guide is an overview of the basic
functions available with your instrument. For detailed
instructions on instrument setup and operation, please refer
to the electronic user manual. You may download a copy
of the user manual at http://www.Starrett.com/u?dfc-um.
INTRODUCTION
The DFC may be used as a classic digital force gage. It may
also be used as a digital controller when connected to the
FMM Digital Force Tester.
The DFC, as a controller, is the only gage available that
lets you setup an automated test, including the operating
parameters of the force tester. Configure speed and
distance limits through the DFC.
STARRETT DFC ADVANCED FORCE GAGE
SPECIFICATIONDFC
Accuracy, Full Scale0.1%
Data Sampling (Hz)25,000
Display Resolution10,000:1
Safe Overload, Full Scale200%
Maximum Tare10%
COMMUNICATIONS
®
Bluetooth
USB 2.0Yes
RS-232Yes
Digital I/O2 channels
Memory, saved results max.99
OPERATING MODE
Machine Control
Real TimeYes
Peak CompressionYes
Peak TensionYes
Load LimitYes
Break LimitYes
Load AverageYes
Load-Time AverageYes
Cyclic Count, Max.99,999
Cyclic Duration, Max.27 Hrs.
Hold Duration, Max.27 Hrs.
Coefficient of FrictionYes
POWER, ENVIRONMENTAL
Battery TypeLithium Ion
Battery Life, @ 20% brightness>30 hours
Charge Time, using 110/240V Mains <3 hours
Display (OLED)320 x 240
Operating Temperature
Thread, for adapters (Metric)M6, M10
Instrument Weight (approx.)
CE COMPLIANCE
EN61010-1 Safety for Electrical Equipment Yes
EN61000-6-3 EMC Generic Emissions Yes
EN61000-6-1 EMC Generic Immunity Yes
NOTES
1. Machine control is exclusive to the DFC. When
connected to the FMM digital force tester,
configuration of force gage and tester is
performed through the gage.
1
Yes
Yes
40°F to 110°F
(4°C to 43°C)
3 lbs
(1.36 kgs)
6 PKG08812-QRGDFC
Page 7
ENGLISH
0.91 (23.0)
7.50 (190.5)
DFC - ADVANCED FORCE CONTROLLER
MODEL
DFC-21012329000.0130.33M6 x 1-6HSPK-FG-A
DFC-52525802,2000.0070.18M6 x 1-6HSPK-FG-A
DFC-10505101605,0000.0060.15M6 x 1-6HSPK-FG-S
DFC-2010010203209,0000.0080.20M6 x 1-6HSPK-FG-S
DFC-50250255080023,000 0.0150.39M6 x 1-6HSPK-FG-S
DFC-100 500501101,60045,000 0.0240.60M6 x 1-6HSPK-FG-S
DFC-200 1,000100225--0.0210.54M6 x 1-6HSPK-FG-M
DFC-500 2,500250550--0.0280.70M10 x 1.5-5H SPK-FG-L
NOTES
Load measurement accuracy is ±0.1% of load cell capacity. Display resolution is 10,000:1.
The DFC keypad is multi-functional. There are four keys.
ON/OFF/MENU KEY
This key is used to power the gage on/off. Press to
power the gage on. Press and hold for 3 seconds to
power the gage off.
This key is also used to access the Setup Menu. When
the gage is powered ON, press to access the Setup Menu.
This key is also used to navigate UP when in the
Setup Menu. p
ZERO KEY
This key is used to zero the displayed values. Press
to zero load and if connected to the FMM digital force
tester, zero distance.
This key is also used to navigate DOWN when in the
Setup Menu. q
SOFTKEY 1
This is the left arrow key. It is used to move out of a
setup when in the Setup Menu. t
This key may also be mapped to a specific function.
The Setup Menu has a Key setup function where you
may assign how Softkey 1 performs. For example, you
can assign a SAVE function to the key. When pressed,
the measured values are "saved" to memory.
SOFTKEY 2
This is the right arrow key. It is used to move in to a
setup when in the Setup Menu. u
Like the Softkey 1, it also may be mapped to a specific
function. For example, you can assign a MODE
function to the key. Pressing the key can change the
operating modes of the gage. You can switch from
Real Time to Peak Tension by pressing Softkey 2.
DISPLAY LAYOUT
The force gage features a high-resolution OLED color
display with adjustable backlight. The backlight may
be adjusted from a setting in the Main Menu.
The DFC features a primary and secondary
display window.
DISPLAY TYPES
PRIMARY DISPLAYSECONDARY DISPLAY
Real TimePeak Measurement
Peak TensionReal Time Measurement
Peak CompressionReal Time Measurement
Average
Average (Functional Capacity) Peak Measurement Time Duration
Static COFKinetic COF
Kinetic COFStatic COF
Peak Measurement Preload to
Start Average
8 PKG08812-QRGDFC
Page 9
ENGLISH
lbf
7
4
7
lbf
lbf
7
Real time load being
placed on load cell
Compression Direction
Shown: Real Time view showing Peak Compression
Peak/Max Compression
0.00
16.89
16.84
Real Time Mode Indicator
Real Time value
lbf
5.66
5.71
Peak Tension Mode Indicator
Shown: Peak Tension view with Tolerance result
Peak/Max Tension
22.29
Tensile Direction
Shown: Peak Tension view
Peak/Max Compression
Compression
Direction
21.10
Peak Compression Mode Indicator
Shown: Peak Compression view
Real Time value
19.24
Peak Tension Mode Indicator
Real Time value
21.06
PKG08812-QRGDFC 9
Page 10
ENGLISH
Battery Status
Shows remaining battery life.
Primary Result
Shows load direction using a symbol for
Tension or Compression.
Shows functions mapped to softkeys
1 and 2. Menu function is always
center key.
Softkey 1
Associated with above Function
Use to navigate LEFT and move out
of setup option.
Zero key
Use to zero the display
User to tare the force gage
Use to navigate DOWN
Softkey 2
Associated with above Function
Use to navigate LEFT and move out of setup option.
Shows functions mapped to softkeys 1 and 2.
Menu function is always center key.
Shows Load measured by load cell
Shows Distance measured by encoder
on the FMM Test Frame.
Unit of Measure
Shows unit of measure for displayed results.
Shown: Display in Classic mode
Secondary Display
Load Bargraph
Shows active load on sensor.
Shows direction of load.
Softkey 2
Associated with above Function
Use to navigate RIGHT and INTO setup option
On/Off/Menu key
Use to power gage ON/OFF
Use to access the SETUP MENU
Use to navigate UP
Softkey 1
Associated with above Function
Use to navigate RIGHT and INTO setup option
Shows setpoint target that completes
the test method.
Shows Load unit of measure.
Shows Distance unit of measure.
Load Bargraph
Shows active load on sensor
Shows direction of load
10 PKG08812-QRGDFC
Shown: Display in Controller mode used with FMM Test Frame
Page 11
ENGLISH
9
5
5
5
Average LoadAverage Load
lbf
3.65
N
7.15
4
Shown: Functional Capacity result
Start Load
8.71
1.00
Average Mode IndicatorShown: Average view
Countdown TimerAverage Load
7.15
8.1
Average Mode Indicator
Kinetic COF
Static COF
33.45
CONTROLLER DISPLAY FORMATS
As a Controller, the DFC will display both load and
distance information and results. The gage will serve
is the universal interface for the "system" and will be
used to setup the force gage and the tester.
The DFC force gage will display load information, as
measured by the load cell.
The DFC force gage will display distance information,
as measured by the FMM digital force tester.
Start/Stop Test SoftkeyLoad Limit Target
100.00N
Active Load
L
D
Shown: Load Limit Test results
Active Load
L8.80
D
16.55
16.64
Distance Limit Target Start/Stop Test Softkey
8.78
N
mm
50.00mm
N
mm
Active Distance
Active Distance
Shown: COF result
µ
µ
k
s
33.450
17.055
5
Average Mode Indicator
Shown: Distance Limit Test results
Break Limit Target
Active Load
L27.45
D
Shown: Break Limit Test results
28.28
PKG08812-QRGDFC 11
Start/Stop Test Softkey
50%
N
Active Distance
mm
Page 12
ENGLISH
4
USING TOLERANCE LIMITS
Use tolerances to setup "pass" and "fail"
measurements. You may specify a Limit 1 and a
Limit 2 to create a tolerance band. Measured results
that equal or fall within the range created by the two
limits are considered "pass" results. If the measured
result falls outside the band created by the two
Limits, the result is considered a "fail" result. A "fail"
results displays in RED.
PressqMove to TOLERANCE
PressuSelect ENABLE
q
Press
Press
Press
PresstMove out of Limit 1
Press
Press
PresstMove out of LIMIT 2
PresstMove out of TOLERANCE
Move down to LIMITS
q
Move down to LIMIT 1
u
Select Limit 1 value using qp
q
Move down to LIMIT 1
u
Select Limit 2 value using qp
Real Time value
lbf
5.66
5.71
Peak Tension Mode Indicator
Shown: Peak Tension view with Tolerance result
SAVING RESULTS
The DFC feature an internal memory for saving
results for the purpose of calculating and displaying
statistics. You may save up to 99 individual results
in memory.
To save results to memory, you must configure
a softkey with the SAVE function. Press the SAVE
softkey to save the displayed results to memory.
Results saved to memory MUST be of the same type.
You cannot have mixed type results. For example, you
cannot save tension results with COF results. Results
MUST be of the same type.
Results obtained using the TESTS feature are saved in
pairs. For each test, there is a Load and Distance result
(L-D). You cannot save results from different types of
tests. For example, you cannot save results from a
Load Limit test with results from a Break Limit test.
Each load-distance pair of results must have come
from the same test type: Load, Distance or Break.
RESULT TYPEMEMORY TYPE
MODES
Real TimeReal Time
Tension PeakTension Peak
Compression PeakCompression Peak
AverageAverage
Static COFStatic COF
Kinetic COFKinetic COF
TESTS
LOAD LIMIT TEST
Distance @ LimitDistance @ Limit
Load LimitLoad Limit
DISTANCE LIMIT TEST
Load @ LimitLoad @ Limit
Distance LimitDistance Limit
BREAK LIMIT TEST
Load Max (Peak Load)Load Max (Peak Load)
Distance @ Load MaxDistance @ Load Max
12 PKG08812-QRGDFC
Page 13
ENGLISH
Management of the DFC memory is important to
ensure correct statistical analysis. Always clear old
results that are from a different test method.
CLEAR RESULTS FROM MEMORY
Results in memory may be cleared individually
or collectively. To clear an individual result from
memory, go to the STATS view. Select the results you
want to clear. Select the "X" (delete) key.
To clear all results from memory, go to the Memory
setup and select CLEAR. This will erase all results for
the gage's memory.
VIEW STATISTICS
You must configure a softkey with STATS to view
statistics. When results are saved to memory, you
press the STATS softkey to view results in memory
and the statistics for those results.
EXPORT RESULTS
You must configure a softkey with SEND to export
to an external device via Bluetooth, RS232 or USB.
Press the SEND softkey to export the displayed result.
GAGE SETUP
MODES
p
u
for Menu
for Modes
t to exit setup
Press
Press
Select your Mode, press
UNITS
p
q
for Menu
for Units
t to exit setup
Press
Press
Select your Units, press
MEMORY
p
Press
for Menu
Pressqfor Memory
u
Press
Press
Press
Select your Units, press
*You must configure a softkey to display memory and statistics.
to Display results from memory
u
to Clear results from memory
u
to Export results from memory
t to exit setup
TESTS
p
Press
Press
Press
Press
Press
*Refer to the User Manual for detailed instructions on using
the TESTS feature in your DFC.
for Menu
q
for Tests
u
to Enable Tests
q
to your Test Method
u
to enter specific Test Method options
TOLERANCE
p
Press
Press
Press
for Menu
q
for Tolerance
u
to Enable Tolerance
Pressuto setup Limit 1, select Limit 1 value pq
Pressuto setup Limit 2, select Limit 2 value pq
Presstto exit setup
KEYS
p
Press
Press
Press
for Menu
q
for Keys
q
to specify a Sound for key presses
Pressqto specify Softkey 1 function
q
Press
Press
*Certain gage functions require a softkey to be setup in order
to use the function- Memory, Statistics, Tests, Export (Send)
to specify Softkey 2 function
t
to exit setup
SETTINGS
p
Press
Press
for Menu
q
for Settings, select Settings Type
COMMUNICATIONS
Pressqfor Communications (Comms)
Press
Press
Press
Press
Press
Press
results: USB, RS-232, or BT (Bluetooth®)
q
to select RS-232
q
to select RS-232 Baud Rate
q
to setup direction sign (- default for Compression)
q
to transmit Units
q
to transmit Tolerance Limits
to specify the Data Channel method for exporting
u
Presstto exit Comms setup
DISPLAY
q
Press
Press
Press
Press
Press
Press
for Display
u
to setup Auto Off
u
to setup Backlight Brightness
to setup Flip orientation (required for
u
TESTS feature)
u
to setup Radix
t
to exit Display setup
FILTERS
q
Press
Press
Press
for Filters
u
to specify filter rate
t
to exit Filter setup
ABOUT
q
Press
Press
Press
*The About setup is read-only.
for About
to view the gage’s characteristics, serial
u
number, overload history and more.
t
to exit setup
PKG08812-QRGDFC 13
Page 14
ENGLISH
MEMORY
q
Press
u
Press
pq
Press
t
Press
LANGUAGE
Press
Press
Press
for Password
to Enable a Password
to select the Passcode
to exit setup
q
for Language
q
to your preferred display language
t
to exit setup
CHARGING BATTERY
Connect the USB cable supplied with your DFC to the
USB connector at the base of the gage. Connect the
opposite end of the cable to the USB connector on a
source device, i.e., personal computer, USB adapter.
If the DFC is connected to an FMM test frame
and power to the frame is ON, the DFC battery is
continuously charged.
Communicating with External Devices
The DFC can communicate using Bluetooth, USB 2.0
and RS-232. Select the Data Channel type to be used.
PROTOCOL WHERE TO USE
Bluetooth
USB 2.0
RS-232
RS-232USB 2.0
Gage Setup Menu Structure
MODESUNITSMEMORYTESTSTOLERANCEKEYSSETTINGS
Real TimeozfDisplayEnableEnableEnableComms
Peak TensionlbfClearLoad LimitLimit 1Softkey 1Data Channel
Peak Compression gfExportTypeLimit 2SendRS-232
AveragingkgfTargetSoundUnitsXmit Comp Time AvgNSpeedSaveXmit Units
COFUserDistance LimitModeXmit TOL
TypeStatsDisplay
TargetStart/StopAuto Off
SpeedReturn 0Backlight
Break LimitSoftkey 2Flip
TypeSendRadix
Minimum BreakUnitsFilter
% DropSaveAbout
Auto 0ModePassword
Auto ReturnStatsLanguage
Auto SaveStart/StopEnglish
Control
Print via Bluetooth printer
®
Export data to an external device
Charge DFC Battery
Upload firmware and new features to DFC
from a PC
Export data to a printer
Export data to a PC
Communicate with a serial printer
Communicate with a serial external computer
or hard drive
Controls the crosshead direction and velocity
the FMM from the DFC gage
Return 0Deutsch
Espanol
Portugues
Francais
Italiano
Chinese
Russian
Polski
Czech
14 PKG08812-QRGDFC
Page 15
NORTH AMERICA
WWW.STARRETT.COM
ATHOL, MA, USA, 01331-1915
EUROPE & ASIA
WWW.STARRETT.CO.UK
JEDBURGH, SCOTLAND, TD8 6LR
SOUTH & CENTRAL AMERICA
WWW.STARRETT.COM.BR
13306-900, ITU, SP, BRASIL
CNPJ 56.994.700/0001-01
CHINA
WWW.STARRETT.COM.CN
USER MANUALS AVAILABLE ONLINE
AT STARRETT.COM
PKG08812-QRGDFC 15
Page 16
Starrett.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.