Spacelabs Ultraview 1500, Ultraview 1600 User manual

®
Ultraview 1500 and
Ultraview 1600 Monitors
90363, 90364
Service Manual
070-0750-01 Rev. G
© 2004 Spacelabs Medical, Inc.
All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Spacelabs Medical. Products of Spacelabs Medical are covered by U.S. and foreign patents and/or pending patents. Printed in U.S.A. Specifications and price change privileges are reserved.
Spacelabs Medical considers itself responsible for the effects on safety, reliability and performance of the equipment only if:
assembly operations, re-adjustments, modifications or repairs are carried out by persons authorized by Spacelabs Medical, and
the electrical installation of the relevant room complies with the requirements of the standard in force, and
the equipment is used in accordance with the operations manual.
Spacelabs Medical will make available, on request, such circuit diagrams, component part lists, descriptions, calibration instructions or other information which will assist appropriately qualified technical personnel to repair those parts of the equipment which are classified by Spacelabs Medical as field repairable.
Spacelabs Medical is committed to providing comprehensive customer support beginning with your initial inquiry through purchase, training, and service for the life of your Spacelabs Medical equipment.
CORPORATE OFFICES
U.S.A.
Spacelabs Medical, Inc. 5150 220th Ave SE Issaquah, WA 98029 Telephone: 425-657-7200 Telephone: 800-522-7025 Fax: 425-657-7212
Authorized EC Representative UNITED KINGDOM
Spacelabs Medical Ltd. Basepoint Business Centre Metcalf Way, Crawley West Sussex RH11 7XX Telephone: 44 (0) 845 6017224 Fax: 44 (0) 845 6017225
BirthNet, Data Shuttle, Flexport, Intesys Clinical Suite, Maternal Obstetrical Monitor, MOM, Mermaid, Multiview, PCIS, PCMS, PrintMaster, Quicknet, Sensorwatch, TRU-CAP, TRU-CUFF, TRU-LINK, UCW, Ultralite, Ultraview, Ultraview Clinical Messenger, Ultraview SL, Uni-Pouch, Universal Flexport, Varitrend and WinDNA are trademarks of Spacelabs Medical, Inc.
Other brands and product names are trademarks of their respective owners.
CAUTION:
US Federal law restricts the devices documented herein to sale by, or on the order
Rx
Only
of, a physician.

Table of Contents

Chapter Page
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Equipment Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Setup
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Pre-Installation Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Network Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
Biomed Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
Theory
90363 Ultraview 1500 System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
90364 Ultraview 1600 System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
CPU PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
I/O PCBA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Interconnect PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20
Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Cleaning/Disinfecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Preventive Maintenance (PM) Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
90363 Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
90364 Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-20
Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Suggested Tools and Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Monitor Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
System Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Boot Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Diagnostic Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Parts
Field-Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Assembly Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Appendix A — Glossary
Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Appendix B — Electromagnetic Compatibility
Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
Frequency Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-3
Appendix C — Symbols
i

Introduction

Contents
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Equipment Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Overview

Spacelabs Medical’s products are designed and manufactured under good manufacturing practices and in compliance with all applicable regulatory requirements. To ensure proper operation in accordance with these guidelines, this product must be maintained by trained technicians using Spacelabs Medical authorized replacement parts.
This manual covers both the 90363 Ultraview® 1500 and the 90364 Ultraview 1600 monitors. The 90363 Ultraview 1500 monitor is a PCIS™ and Ultraview SL™-compatible bedside mainframe. It features a large 15-inch color display and an infrared touch screen.
The 90364 Ultraview 1600 monitor combines the module mounting capability of the 90491 module housing, with the processing and display capabilities of the 90363 Ultraview 1500. The right-hand side of the chassis is used to house the CPU and I/O PCBAs, while the left-hand side provides two bays for module plug-in. The Ultraview 1600 is capable of driving an external speaker (either amplified or unamplified) or may stand alone by use of the internal speaker. The result is a product that provides the same Ultraview Care Network capabilities as the Ultraview 1500, but adds plug-in support for two modules in a single, compact unit with a separate display. The Ultraview 1600 is primarily intended to function as a compact, two-module bedside unit.
1-1
Ultraview 1500/1600 Monitors — Service Manual
Refer to Theory on page 3-1 of this manual for a detailed description of each monitor and its components.
wall track
90491
module housing
monitor
Figure 1-1: Ultraview 1500 with wall mount
90363
1-2
Figure 1-2: Ultraview 1500 with pedestal mount
Introduction
Figure 1-3: 90364 Ultraview 1600 with external power supply

Accessories

Monitor Options
Option Definition
-F
-N
-O
-P
-Q
-R
-X
-04
-05
-06
Table 1: Ultraview 1500 and Ultraview 1600 Options
Ethernet Network Communications
Local Vital Signs Calculations Package
Local Drug Dose Calculations
Interactive and Advanced Network Functions
Data Shuttle
Patient Data Logger (PDL)
200V/240V Operation (Ultraview 1500 only)
Four Waveform Zones
Five Waveform Zones
Six Waveform Zones
1-3
Ultraview 1500/1600 Monitors — Service Manual
Example:
A 90363-XQ04 (where “X” represents the language designator) is an Ultraview 1500 equipped with:
• Q = Data Shuttle option
• 04 = Four waveform zones
Accessory devices supported by the Ultraview 1500/1600 are as follows:
Modules
90428, 90432, and 90410. The use of modules requires a module housing(s) and DC power supply(ies).
The 90364 Ultraview 1600 supports all existing Ultraview modules except 90425, 90428, 90432, and
90410. The unit provides plug-in support for either two single-height modules or one double-height module. The Ultraview 1600 supports additional modules by means of external module housings.
Flexport® System Interfaces
Spacelabs Medical Flexport interfaces. Refer to Parts on page 6-1 for installation requirements.
Repeater Displays
monitor. The repeater displays serve only to duplicate the information shown on the Ultraview 1500/1600 monitor and does not accept user input. Repeater displays may be located up to 100 feet from the Ultraview 1500/1600 monitor.
Touchscreen — The Ultraview 1600 supports the ELO Touch brand touchscreens. The Ultraview 1500 uses an integral infrared touchscreen.
Mouse
Keyboard
Nurse Alert
device. This device provides an external indication that one of the parameters being monitored is in an alarm condition.
The 90363 Ultraview 1500 supports all existing Ultraview modules except 90425,
The Ultraview 1500/1600 supports the use of all existing
Up to two repeater displays may be connected to the Ultraview 1500/1600
An Ultraview 1500/1600 supports the use of a PS/2-style mouse.
An Ultraview 1500/1600 supports the use of a PS/2-style keyboard.
The Ultraview 1500/1600 supports the use of a Spacelabs Medical nurse alert

Equipment Illustrations

90363 Ultraview 1500
The stand-alone unit consists of a color display with touchscreen and an optional mouse and keyboard.
touchscreen
Figure 1-4: Front view of 90363
1-4
Introduction
2
1
2
1
printer port (unavailable)
nurse call (unavailable)
nurse alert
speaker
microphone (unavailable)
mouse port
keyboard
1
serial out 1
2
serial out 2
WARNING:
For safety, the power cord retainer hardware must always be used.
SDLC
video out 1
video out 2
SDLC
SDLC
network connection
ON/OFF switch
power cord
ground lug
antenna cable (unavailable)
Figure 1-5: Rear view of the 90363
1-5
Ultraview 1500/1600 Monitors — Service Manual
90364 Ultraview 1600
SDLC
1
2
Nurse call (unavailable), J17 Microphone (unavailable), J13
1
Serial port 1, J15 Nurse alert, J14
Video port 1, J7
2
Serial port 2, J9 (Touchscreen)
NiMH battery
Equipotential ground
1-6
Audio output, J10
SDLC
SDLC/IO, J16
Mouse, J11
Video port 2, J8
Ethernet (10BaseT), J18
External keyboard, J12
SDLC terminator, SW2
SDLC/power input, P1
SDLC/power output, J2
Unit power switch, SW1
High level output, J3
IRTS interface for 91416-B, J19
Figure 1-6: Rear view of the 90364

Setup

Contents
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Pre-Installation Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Biomed Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Inspection

Prior to installation of your 90363/90364 patient monitor, conduct an equipment audit.
Upon receipt of the equipment, a detailed inventory must be taken to verify that the equipment received matches your order. This inventory should include serial numbers, model numbers with options, and cables. Carefully inspect these items for shipping damage. If damage is apparent, notify the freight company and Spacelabs Medical immediately.
When removing items from the shipping container, make sure you remove ALL
!
The monitors are typically shipped as follows:
Ultraview 1500 Monitor (90363) — Contains the assembled monitor, I/O and CPU printed circuit board assemblies (PCBAs), power cord (U.S.), and AC cord retainer.
Ultraview 1600 Monitor (90364) — Contains the main assembly with installed I/O and CPU printed circuit board assemblies (PCBAs), power cord (U.S.), and AC cord retainer.
External Module Housing — The 90491 four-bay housing or the 90499 two-bay housing, DC power supply, SDLC cables, and power cord.
Accessories — Contains the international power cords (if applicable) and cable assemblies ordered.
Mounting Hardware — All ordered mounting hardware is supplied with appropriate instruction sheets.
components from each container.
2-1
Ultraview 1500/1600 Monitors — Service Manual

Pre-Installation Testing

Before mounting the equipment, it is recommended that you first verify its operation as follows:
1. Power ON the unit and let the power-ON diagnostics run.
2. Look in the upper-left corner of the display for any diagnostic failures or other system errors. The monitor will start up normally if no unacceptable diagnostic errors occur. Refer to Troubleshooting on page 5-1 for more information on the diagnostics features if the monitor does not start up normally.
3. Insert at least one parameter module for testing purposes.
4. Refer to the Preventive Maintenance (PM) Procedures on page 4-2 of this manual and complete those procedures.

Mounting

The monitor can be mounted in one of three ways:
Wall mount — The unit is attached to an adjustable mounting arm that projects out from a wall
track. This mounting scheme is suitable for a bedside configuration.
Pedestal mount — The unit is securely attached to a flat surface, such as a shelf or an operating room cart.
Roll stand — The unit can be easily rolled from one location to another.
The 90364 can also be operated unmounted in a free-standing, table-top configuration, or it can be console-mounted.
For installation requirements of the module housing(s), refer to the Module Housings and Power Supplies Service Manual (P/N 070-0680-xx, located on CD-ROM 084-0700-xx).
90363 Installation Specifications
Assembled weight of monitor: 36 lbs (16.4 kg)
Dimensions of monitor with pedestal at 0° tilt: 14 (H) × 14.5 (W) × 17.5 (D) in. (35.6 × 36.8 × 44.5 cm)
Temperature (operating): +10° to 40° C (50° to 104° F)
Humidity (operating): 10 to 95% relative humidity, non-condensing
90363 total power dissipation: 180 watts; 680 BTU/hour
AC input voltage range: 100 to 120; 220 to 230 VAC
AC input current: 2.5 A @100 to 120 VAC
AC input frequency range: 50 to 60 Hz
Power supply output voltage: +18 VDC
Power supply output current: 4.25 A (maximum)
2-2
Setup
90364 Installation Specifications
Assembled weight of monitor: 9.7 lbs (4.4 kg)
Assembled weight DC power supply: 1.8 lbs (0.8 kg)
Dimensions of monitor:
Dimensions of DC power supply:
AC input voltage range: 100 to 120 VAC, 200 to 230 VAC
AC input current: 1.5 A (maximum)
AC input frequency range: 50/60 Hz
Power supply output voltage: +18 VDC
Power supply output current: 4.25 A (maximum)
Temperature (operating): +10° to 40° C (50° to 104° F)
Humidity (operating): 10 to 95% relative humidity, non-condensing
90364 power dissipation: 100 watts, 378 BTU/hour
Base Tension Adjustment
10.25 (H) × 6.0 (W) × 9.5 (D) in. (26.0 × 15.2 × 24.1 cm)
2.1 (H) × 4.0 (W) × 6.5 (D) in. (5.3 × 10.2 × 16.5 cm)
The 90363 monitor’s base can be adjusted to increase or decrease the tension that allows tilting of the monitor screen. Tension is factory-adjusted for optimum performance and further adjustment is usually not necessary. If adjustment is necessary, turn the base screw clockwise (refer to Figure 2-1) to increase base tension; turn it counterclockwise to decrease tension. Set the tension loosely enough to allow normal viewer adjustments, yet tightly enough to hold the adjusted position.
base tension adjusting screw
Figure 2-1: 90363 base tension adjustment
If necessary, base adjustment must be done prior to installing the monitor onto a
!
mount.
Not all monitors will have the base tension adjustment described here.
2-3
Ultraview 1500/1600 Monitors — Service Manual
Wall Mounting
Specific wall requirements for the installation of wall mounts are not covered in
!
this manual and are the responsibility of each hospital. Refer to the Module Housings and Power Supplies Service Manual, (P/N 070-0680-xx, located on CD­ROM 084-0700-xx) for detailed instructions.
90363 or 91415-A
AC mains
10’ patient cable
Minimum of 60” from floor
patient bed
Route all power cables inside the wall track.
Figure 2-2: Central station configuration for the Ultraview 1500
The wall mounting hardware consists of:
Wall track — The wall track is mounted vertically on the wall with the mounting arm installed in
the track. The mounting arm can be slid up and down in the wall track to the desired height. Wall tracks are available in multiple lengths.
Wall track end caps — The wall track end caps are installed on the top and bottom of the wall track to give the track a finished appearance and to prevent the equipment from sliding off the end of the mounting track.
Mounting arm — The mounting arm projects horizontally out from the wall track and holds the display assembly and module housings. Two types of arms are available: one for a monitor-only installation; the other for installation of the monitor and module housing(s).
Mounting screws — Mounting screws are provided to install the end caps and to secure the mounting arm at the desired vertical position.
Ethernet connector
90491 or 90364
power supply
Mounting Considerations
Power supply should be placed no closer than three inches (7.62 cm) to the remote housings to allow sufficient air flow. Install a minimum of four AC outlets.
The monitor, monitor wall plate, and AC outlet must be located on the same side
!
2-4
of the patient bed. The power cord has a standard length of 10 ft. (3.05 m).
Pedestal Mounting
90363 Monitor
To attach the pedestal mount to the display assembly refer to Figure 2-3.
1. Verify that the pedestal’s six rubber feet and four retaining studs with retaining rings have not been dislodged during shipping.
2. Carefully place the display on its side. A cushion should be used under the display to avoid scratching the display case.
3. Rotate the display’s round base (if necessary) so that its tick mark points toward the front of the unit.
4. Line up the four retaining studs on the pedestal with the four holes in the display’s round base.
5. Push the pedestal onto the round base and secure it by turning each of the four retaining studs 90 degrees clockwise with a straight blade screwdriver. The studs will lock in place and can no longer be turned.
6. Carefully return the display to its normal upright position.
Setup
tick mark
Figure 2-3: Pedestal mount
rubber feet (6)
retaining studs (4)
2-5
Ultraview 1500/1600 Monitors — Service Manual
Console Mounting
90364 Monitor
The physical layout and other console characteristics depend on the requirements of the hospital or other institution. It is the responsibility of the institution to define its specific console needs and to design and construct the console according to its specifications.
For this reason, Spacelabs Medical does not supply a pre-fabricated console assembly. Instead, this section provides the information necessary to create a custom console design. In designing the console, the following considerations must be kept in mind.
• Access
The rear of the console must be designed to allow for easy access. The dimensions of the rear access door must allow the 90364 and display monitor to be easily installed and removed.
• Cooling
The 90364 does not contain a fan.
!
The console cabinet must be designed to allow sufficient air flow for natural convection. If the console cabinet is enclosed, one of two cooling methods can be used:
Central air conditioning — The console can be connected directly to an air conditioning cooling duct. The maximum ambient air temperature should be limited to approximately 32° C, and vented from the console at the highest exit.
Forced air fan — Small, quiet muffin fans rated at 100 cubic feet per minute (CFM) or equivalent (to be provided by the institution) can be installed. In this case, provide one fan for every central station or two system printers/module recorders that are installed in the console.
If a network printer is to be installed in the console, the power dissipation and
!
dimensional requirements will be different. Refer to the printer specifications for further information.
2-6
Setup
An example console design is shown in Figure 2-4. This figure shows the minimum dimensions required to satisfy the above requirements.
exhaust fan with louvers
removable back
air inlet
26.5”
(67.3 cm)
cable access
hole
10×
DC power supply
optional DC power supply mount using wall mount
optional air inlet
16.5” (41.91 cm)
30”
(76.2 cm)
46.5”
(118.20 cm)
AC power
!
Figure 2-4: Minimum cavity sizes for console setup
When mounting the Ultraview 1600 as shown in Figure 2-4, this cabinet specification may not allow for appropriate clearance for displays larger than 15 inches (38.1 cm).
2-7
Ultraview 1500/1600 Monitors — Service Manual

Installation

Table 1: Common Cables and Adapters
Cable/
Adapter
A 012-0182-00 Cable, Shielded RS-232, Display, Alarm
B 012-0391-xx Cable, Display Assembly to Module Housing
C 175-0951-xx Cable, Assembly, Ethernet, 10BaseT
E 012-0395-00 Cable, Serial Data, RS-232
Part Number Description
90363 Ultraview 1500
The cables and adapters available for making interconnections to the 90363 and other components are listed in Table 1 and Table 2. The use of these cables is illustrated schematically in Figure 2-5: 90363 cabling diagram.
CAUTION:
Spacelabs Medical has tested and approved the cables listed in Table 1 and Table 2 for use with the 90363 monitor. For continued EMI radiation compliance, use only these cables.
Table 2: 90363 Cables and Adapters
Cable/
Adapter
F 012-0584-00 Cable, Display Assembly, male DB15 to RGB, 6 feet (1.8 m)
Refer to Parts on page 6-1 of this document for additional cable part numbers.
2-8
Part Number Description
A
Setup
nurse alert
90363 monitor
SDLC
B, G
P1
DC power supply
90491/90499
module housing
1
Video output is a repeater display and is not configurable.
2
Serial output may also be used for external touchscreen input in some external display
applications.
Figure 2-5: 90363 cabling diagram
C
F
E
ferrite beads
Ethernet
1
Video
serial I/O
2
90364 Ultraview 1600
The cables and adapters available for making interconnections to the 90364 and other components are listed in schematically in Figure 2-6.
Adapter
Table 1 on page 2-8 and Table 3 below. The use of these cables is illustrated
CAUTION:
Spacelabs Medical has tested and approved the cables listed in Table 3 for use with the 90364. For continued EMI radiation compliance, use only these cables.
Table 3: 90364 Cables and Adapters
Cable/
G 012-0601-00 Cable, SDLC with power interface, 1 ft. (0.305 m)
H 012-0593-00 Cable, video
J 012-0555-00 Cable, powered Flexport
K 012-0595-00 Cable, audio
L 012-0620-00 Cable, touchscreen/video sync.
Part
Number
Description
2-9
Ultraview 1500/1600 Monitors — Service Manual
Refer to Parts on page 6-1 for additional cable part numbers.
*
*
91416-B display
Generic
display with touch and audio
J7
J9
H
E
K
serial I/O
L
G
J16 P1
J7
J9 J10 J19
90364
J2
Figure 2-6: 90364 cabling diagram
J14
J15
J18
DC power
supply
A
E
C
G
serial I/O
Ethernet
P1 J2
90499
module
housing
J
nurse
alert
terminator
Flexport
† The J9 COM port must be used to interface a serial touchscreen. Using J15 to interface the
touchscreen is not supported.
* Only one display with touchscreen can be attached to the 90364. A second display can only be
driven for a video touchscreen; any audio is not available.
Ethernet LAN Cables
Pre-molded cables — When connecting a monitor to the wall plate or in the construction of a
central station, use cables with both ends pre-molded. These cables have ferrite beads molded into each end for EMI compliance and are available in multiple lengths.
Bulk cables — If desired, cables can be constructed from bulk 500-foot (152.4 m) spools. This
type of cable construction is preferable in cases where conduit lengths are unknown.
SDLC Bus Connections
The 90364 has two integral module bays providing plug-in support for either two single-height modules or one double-height module. If additional modules are required, these are installed in external module housings and connected to the 90364 by the SDLC interface. The 90363 requires an external module housing for parameter module support. Flexport devices are also connected by the SDLC interface. External module housings or Flexport devices require that the SDLC bus be configured correctly.
The 90363/90364 support the 90491 four-bay external module housing, or the 90485/90499 two-bay external module housings.
Multiple module housings may be used. They are interfaced to the 90363/90364 by daisy-chaining them together to create an extended SDLC bus.
2-10
Maximum Cable Lengths
The following cables are limited to the indicated maximum length:
Ethernet Cable (10Base5) — 10Base5 cables have a maximum length of 164 feet (50 m)
including the total cable length from the monitor to the wall plate and/or transceiver, and a 35-foot (10 m) propagation delay for any Multiport present.
Ethernet Cable (10BaseT) — 10BaseT cables have a maximum length of 328 feet (100 m).
SDLC Cable — SDLC cables have a maximum length of 40 feet (12.2 m), including the total
SDLC cable length from the monitor to last device on the SDLC bus.
Video display cable — Video display cables have a maximum length of 100 feet (30.5 m), including the total length of cable from the monitor to the last repeater display.
91416-B Touchscreen Cable — 91416-B touchscreen cables have a maximum length of 6 feet (1.83 m).
Setup
2-11
Ultraview 1500/1600 Monitors — Service Manual

Network Configuration

Figure 2-7 illustrates a typical network system composed of bedside monitors, central monitors, and a clinical information system.
router
Unix Workstation equip #192 equip # 100 equip # 200
164.90.254.192
subnet 254 (standard Spacelabs Medical network)
164.90.254.0
164.90.254.10
monitor #10 monitor #11
Unix Workstation equip # 192 equip #100
164.90.250.192
MFS UCN Chartmaster
164.90.254.100
MFS Chartmaster
164.90.250.100 164.90.250.101
Workstation
164.90.254.11
NCD X-terminal
equip # 101
164.90.254.200
Ethernet
Figure 2-7: Example of Spacelabs Medical network
For the example in Figure 2-7, the Network Setup entries are as follows:
Monitor #10
• Monitor ID: 10
• IP Address: 164.90.254.10
Monitor #11
• Monitor ID: 11
• IP Address: 164.90.254.11
2-12
Ethernet
Setup
Ethernet LAN Installation
Detailed installation instructions for an Ethernet local area network (LAN) are
!
To install a 90363/90364 monitor onto an existing Spacelabs Medical Ethernet LAN, complete the following (refer to page 2-8, and Table 3 on page 2-9).
1. Power OFF the monitor.
2. Attach one end of a 10BaseT Ethernet cable (C) to of the 90363/90364 monitor.
3. Power ON the monitor and refer to System Configuration on page 2-14 to set up the monitor for network operation.
beyond the scope of this document.
CAUTION:
Only qualified personnel should attempt to install a monitor onto an Ethernet LAN.
Figure 2-5 on page 2-9, Figure 2-6 on page 2-10, Table 1 on page 2-8, Table 2 on
Do not connect the monitor to an Ethernet LAN prior to configuring the network
!
4. Connect the other end of the Ethernet cable (C) from the 90363/90364 to the nearest
network port.
parameters. If it is incorrectly configured for LAN access, other units on the network may be interrupted.
WARNING:
When 10Base5 hardware is used for the LAN, make sure the Ethernet wall plate and the shield of the Ethernet connecting cable are bonded to the hospital grounding system.
2-13
Ultraview 1500/1600 Monitors — Service Manual

System Configuration

The Biomed menu enables a Spacelabs Medical Field Service Engineer to set user defaults for monitor operation.
To access the Biomed menu:
1. Touch MONITOR SETUP.
2. Touch PRIVILEGED ACCESS to display the Keyboard menu (refer to Keyboard Menu on
page 2-17).
3. Enter the correct password (case-sensitive) to display the Biomed Monitor Setup menu.
Based on your software version, your screen may differ from the following.
!
MONITOR SETUP - Select parameter
NETWORK
SETUP
RECORDER
SETUP
PRESELECTED
RECORDINGS
SERIAL PORTS SETUP
ALARM USER
ACCESS
UNITS OF MEASURE
MORE
TIME/ RESET
DATE
NETWORK SETUP - Restart monitor after entering changes
SUBNET
RECORDER SETUP - Select parameter
NAMES
(SELECTION BACKUP) or
PRIMARY
PRESELECTED RECORDINGS - Select configuration to change
PRESELECTED A PRESELECTED B
Select option to change - x: current type and bed configuration (where x = A or B)
SELECT
TYPE
NODE NODE
ID NAME DIRECTORY
SELECTRECORDER
RECORDERS
BACKUP
SELECT
BED
TCP/IP SETUP
SELECTION
AUTO
BED
BACKUP
(SELECTION = AUTO)
RECORDER A
TONE
CONFIGURATON
RECORDER B
SUBNET ACCESS
ON
SERVICE
FUNCTIONS
OFF
MONITOR
Select beds/subnets - x: current type configuration (where x = A or B)
ALL THIS
SUBNETS
Select recording type - x: current bed configuration (where x = A or B)
ALL
BEDSIDE PARAM(S)
SUBNET
THIS
MONITOR
2-14
CONFIGURED PARAMETERS
MONITOR SETUP - Select parameter
NETWORK
SETUP
Select serial port to assign owner to:
1
BIOMED LEVEL - Serial Port Configuration
SETTINGS
BIOMED LEVEL - Serial port settings: Changes committed on exit
DATA BITS
RECORDER
SETUP
2
2
ASSIGNMENT
Restart monitor after selecting owner for serial port
DIAGNOSTICS
8
PAR ITY
NONE
PRESELECTED
RECORDINGS
DATA
LOGGER
STOP BITS
1
ECHO NONE
SERIAL PORTS
DISABLE
USER
ACCESS
BAUD RATE
ALARM SETUP MEASURE
CR/LF XON/XOFF NONE
UNITS OF
9600
Setup
MORE
ALARM SETUP - Select parameter
REMOTE ACCESS
ON OFF
ALARM RELAY SETUP - Select parameter
RELAY TIMEOUT
0 SEC 10 SEC
ALARM LEVEL - Select minimum alarm priority level to trigger relay activation
HIGH MEDIUM LOW
USER ACCESS - Enable user access to functions
PATI E NT TY PE
ON OFF
Restart monitor after selecting units of measure
mmHg
kPa
ALARM SUSPEND
ON OFF
FLASHING
STEADY ON
PARAMETER CONFIG
ON OFF
inches
cm
TREND SUSPEND
ON OFF
ALARM
LEVEL
RECORDING DURATION
ON OFF
lb
kg
MIN VOLUME
ON OFF
ALARM
RELAY
MORE - Select parameter
QRS/SPO2 TONE ENABLE
ALWAYS DURING ALARM SINGLE TRIPLE
MORE
ALARM WATCH TONE
2-15
Ultraview 1500/1600 Monitors — Service Manual
MONITOR SETUP - Select parameter
SETUP
MONITOR SETUP - Select parameter
TIME/ TONE DATE CONFIGURATION
TONE CONFIGURATION - Restart Monitor after selecting tone configuration
ISO STANDARD ALARM TONES
RECORDER
SETUP
CONFIGURABLE
PRESELECTED
RECORDINGS
SERVICE
FUNCTIONS
ALARM TONES
use arrow keys to adjust the selected alarm priority’s period
HIGH
15s
ALARM TONES
MEDIUM
30s
PORTS SETUP MEASURE
RESET
RESET
MONITOR
MONITOR
CONTINUOUS
ALARM USER
LOW
30s
ACCESS
UNITS OFNETWORK SERIAL
FAC TORY
DEFAULTS
MORE
BIOMED LEVEL - service functions
DISPLAY
ERROR LOG
ERROR LOG MESSAGES
PREVIOUS
PAG E
BIOMED LEVEL - calibrate monitor subsystems
CRT
POSITION
MONITOR
CALIBRATION
NEXT PAG E
SYNC ON GREEN
ON OFF
MOST
RECENT
SYSTEM
Figure 2-8: Biomed Monitor Setup menu
INFO
CHANGE
PASSWORD
TOUCHSCREEN
CALIBRATION
BIOMED LEVEL - touchscreen margin settings: select a TS margin setting
DEFAULT
SYSGEN
CLEAR
LOG
TOUCHSCREEN MARGIN
A B D E
RESET
MONITOR
PRINT
ADJUSTMENT
LOG
90363 only
C
2-16
Setup
Keyboard Menu
The Keyboard menu appears whenever a menu is selected that requires entering alphanumeric data. (A-Z, 0-9):
RESTART CLEAR
! 1
CAPS
SHIFT
<--- --->
When the keyboard first appears, the cursor is positioned at the first character in the input line and moves through the input line as you press the keys. Input changes are displayed on the prompt line but are not displayed on the monitor until you touch ENTER.
When you select a character key, the cursor is located at the last character in the name. The cursor will remain in place but will type over the old character, replacing it with the new character.
Touch BACKSPACE to remove the character under the cursor and move the remaining characters one position to the left.
Use the key to move the cursor one position to the left. If the cursor is at the first character in the
name, no change occurs.
@
2
Q W E TR Y U I PO
A S D F G H J K L
# 3
Z X C BV N M
$ 4
%
5
Figure 2-9: Keyboard menu
^ 6
&
7
*
8
(
9
) 0
<
,
BACK-
SPACE
_
-
: ;
>
.
ENTER
?
/
Use the key to move the cursor one position to the right. If the cursor is at the last character in
the line, no change occurs.
Touch RESTART to redisplay the last name recognized by the system. Any changes that have not been stored in memory (before touching ENTER) are deleted.
Touch CLEAR to delete the displayed characters; this enables you to enter new ones.
2-17
Ultraview 1500/1600 Monitors — Service Manual

Biomed Setup Menu

Network Setup Menu
The Network Setup menu enables you to define a subnet name, node ID, and bed name; view the active nodes on the network; and to select an appropriate subnet access level.
Touch NETWORK SETUP to display the following menu:
NETWORK SETUP - Restart monitor after entering changes
SUBNET ACCESS
ON
OFF
!
SUB NET
Additional keys may be present. Refer to the corresponding sections of this
NODE
ID
manual or to the Ultraview SL Operations Manual (P/N 070-1001-xx, located on CD-ROM 084-1101-xx) for configuration.
TCP/IP SETUP
BED
NAME
NODE
DIRECTORY
SUBNET Key
Touch SUBNET to display the Keyboard menu, enabling you to type in a subnet name (refer to
Keyboard Menu on page 2-17 for instructions on usage). This name can contain up to five characters, letters, or numbers.
To enter a subnet name:
1. Touch SUBNET.
2. Type in the subnet name.
3. When you have entered the name correctly, touch ENTER to store the name in memory.
4. Reset the monitor so that the network can recognize the new name.
NODE ID Key
Touch NODE ID to display the pop-up keypad.
Each monitor on the network must have a unique node ID. The default node ID for an uninitialized monitor is 1. You must change this ID to allow the monitor to function properly on the network. After changing the ID, you must reset the monitor to enable the system to recognize the new node ID.
To enter a node ID:
1. Touch NODE ID.
2. Enter a number between 1 and 64 (1 and 250 with monitors that support the expanded network).
3. Touch ENTER on the keypad to place it into memory.
2-18
Setup
TCP/IP SETUP Key
To enter a TCP/IP address:
1. Touch TCP/IP SETUP to display the Keyboard menu.
2. Type in the TCP/IP address.
3. Touch ENTER to initiate error checking and verify that the address is in the proper xxx.xx.xxx.xx format (e.g., 164.90.254.xx, where xx is monitor node ID).
4. When you have entered the name correctly, touch ENTER to store the name in memory.
5. Reset the monitor to have the network recognize the new TCP/IP address.
Entering an invalid IP address (e.g., 89.0.3.22x) sounds an error tone and displays the following message:
IP address 89.0.3.22x is out of range or invalid.
The TCP/IP settings are not currently used.
!
BED NAME Key
Touch BED NAME to display the Keyboard menu. Use the keys on the Keyboard menu to enter the bed name assigned to the monitor. This bed name is also the monitor’s node name on the network and is limited to five characters.
The message, Restart monitor after change in bed name: xxxxx, is displayed when you touch BED NAME. The xxxxx is the node name currently recognized by the network. The name you enter can contain from one to five alphanumeric characters but must be unique to the network. Reset the monitor after you enter the bed name to allow the network to recognize the new name.
NODE DIRECTORY Key
The NODE DIRECTORY key enables you to view which nodes are active on the network. Touching this key displays a set of node keys identifying the bed name (top) and the subnet where it belongs (bottom):
Nodes active on the network
BED01
WARD1
(The above names are provided only as examples.)
SUBNET ACCESS Key
BED02
WARD1
BED03
WARD1
BED04
WARD2
BED05
WARD2
CEN01
ICU
The SUBNET ACCESS key limits the monitor’s accessibility to beds physically located on other subnets. When SUBNET ACCESS is OFF, the monitor can access only those beds attached to its own subnet. When SUBNET ACCESS is ON, the monitor can also access beds located on other subnets.
To set SUBNET ACCESS to OFF:
1. Set SUBNET ACCESS to OFF.
2. Power the monitor OFF and ON again to ensure that all alarm watches, zone assignments, and remote views are cleared.
2-19
Ultraview 1500/1600 Monitors — Service Manual
3. Set the remote views (bedside monitors) as desired.
4. Select RECORDERS using only the recorders on the local subnet and set PRESELECTED RECORDINGS to access only the monitors in the local subnet.
5. Set alarm watches as desired.
Recorder Setup Menu
Touching the RECORDER SETUP key displays the following menu:
RECORDER SETUP - Select parameter
RECORDER
NAMES
SELECT
RECORDERS
RECORDER NAMES Key
This key enables you to assign names to each recorder on the system. Touching RECORDER NAMES displays the following information:
SELECTION
AUTO
BACKUP
Select recorder name to change
REC01
(The above names are provided only as examples.)
REC02 REC03 REC04 CENT1 ICU02 ER
Touch the RECORDER key you want to change, or touch the blank key to add another recorder. When touched, the Keyboard menu appears, enabling you to enter or change the name. Each name can contain from one to five characters.
SELECT RECORDERS Key
The SELECT RECORDERS key enables you to assign a priority to system recorders for each bed (where nnnnn is the bedname).
Touching this key displays the following menu:
when the SELECTION key is set to BACKUP
Select priority of recorder to be assigned to nnnnn
PRIMARY BACKUP
when the SELECTION key is set to AUTO
Select priority of recorder to be assigned to nnnnn
RECORDER BRECORDER A
1. Touch one of the above keys to display a list of available system recorders.
2. Touch the recorder name key assigned to the bed as either primary or secondary (depending upon your choice in step 1).
3. Touch Previous menu to return to one of the Select Priority menus.
2-20
Loading...
+ 94 hidden pages