Spacelabs 91369 User manual

4 (1)

91369 Service Manual

070-1156-00 Rev. E

©2007 Spacelabs Medical, Inc.

All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Spacelabs Medical. Products of Spacelabs Medical are covered by U.S. and foreign patents and/or pending patents. Printed in U.S.A. Specifications and price change privileges are reserved.

Spacelabs Medical considers itself responsible for the effects on safety, reliability and performance of the equipment only if:

assembly operations, re-adjustments, modifications or repairs are carried out by persons authorized by Spacelabs Medical, and

the electrical installation of the relevant room complies with the requirements of the standard in force, and

the equipment is used in accordance with the operations manual.

Spacelabs Medical will make available, on request, such circuit diagrams, component part lists, descriptions, calibration instructions or other information which will assist appropriately qualified technical personnel to repair those parts of the equipment which are classified by Spacelabs Medical as field repairable.

Spacelabs Medical is committed to providing comprehensive customer support beginning with your initial inquiry through purchase, training, and service for the life of your Spacelabs Medical equipment.

CORPORATE OFFICES

U.S.A.

Spacelabs Medical, Inc.

5150 220th Ave SE

Issaquah, WA 98029

Telephone: 425-657-7200

Telephone: 800-522-7025

Fax: 425-657-7212

Authorized EC Representative UNITED KINGDOM

Spacelabs Healthcare, Ltd.

Beech House, Chiltern Court

Asheridge Road, Chesham

Buckinghamshire HP5 2PX

Telephone: 44 (0) 1494 784422

Fax: 44 (0) 1494 794414

BirthNet, Clinical Browser, Data Shuttle, Flexport, Intesys, Mermaid, MOM, Multiview, PCIS, PCMS, PrintMaster, Quicknet, Sensorwatch, TRU-CAP, TRU-CUFF, TruLink, Ultralite, Ultraview, Ultraview Care Network, Ultraview Clinical Messenger, Ultraview Digital Telemetry, Ultraview SL, Uni-Pouch, UCW, Varitrend and WinDNA are trademarks of Spacelabs Medical, Inc.

Other brands and product names are trademarks of their respective owners.

Caution:

Rx Only

!

US Federal law restricts the devices documented herein to sale by, or on the order of, a physician.

Before use, carefully read the instructions, including all warnings and cautions.

Table of Contents

Contents

Page

Introduction

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

Physical Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

Regulatory Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

Monitor Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

Setup

Unpacking the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Assembling the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7

SDLC Bus Termination. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8

Network Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11

Power-ON Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12

Configuring the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13

Theory

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

Major System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

Printed Circuit Board Assemblies (PCBAs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

Interconnect and Connector PCBAs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15

Bezel Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16

Boot Sequence Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18

Normal Operation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19

Parameter Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21

CPU PCBA Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21

CPU PCBA Jumpers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22

I/O PCBA Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22

Interconnect PCBA Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23

Maintenance

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

Mechanical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

Electrical Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4

Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

Assembly/Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24

Troubleshooting

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1

System Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2

Boot Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3

Power-ON Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

Extended Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

Diagnostic Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7

91369 Service Manual

i

Table of Contents

Error Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10

Diagnostics Failure Messages and Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11

System Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13

Parts

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

Assembly Drawings and Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4

Directory of Keys

BIOMED Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

Glossary

Appendix A — Electromagnetic Compatibility

Electromagnetic Emissions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2

Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3

Appendix B — Symbols 1

91369 Service Manual

ii

Introduction

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

Physical Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Regulatory Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Monitor Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Overview

Spacelabs Medical’s products are designed and manufactured under good manufacturing practices and in compliance with all applicable regulatory requirements. To ensure proper operation in accordance with these guidelines, this product must be maintained by trained technicians, using Spacelabs Medical authorized replacement parts.

Warnings, cautions, and notes are used throughout this manual. They are identified by the formats shown below. Be sure to read all warnings, cautions, and notes included in each section of this manual.

Warning:

Alerts the user to potentially serious outcomes (death, injury, or serious adverse events) to the patient or user.

Caution:

Alerts the user to actions to be taken to avoid non-serious injury to the patient or user, or to adverse effects to the device.

Note:

Failure to observe notifications may result in unexpected outcomes.

The 91369 monitor is a lightweight, portable monitor designed for use as a compact bedside monitor or as a battery-operated transport monitor. The monitor features a five-wire, resistive touchscreen and can be operated on either AC mains or battery power.

The single-high module slot on the right side of the monitor accepts all single-high modules, including the 90496 Ultraview® Command module and 91496 Ultraview SLCommand module, to permit the monitoring of parameters such as electrocardiography (ECG), pulse oximetry (SpO2), temperature, and invasive pressure.

91369 Service Manual

1-1

Introduction

 

Figure 1-1: 91369 monitor

 

 

 

 

Physical Dimensions

 

Assembled weight

10.0 pounds (4.6 kg)

Dimensions

8.3 (H) × 11.7 (W) × 6.2 (D) inches

(21.1 × 29.7 × 15.8 cm)

 

Electrical Specifications

Designed for continuous operation. Requires outlet with ground (Protective Earth) conductor. Designated Class I by applicable electrical safety standards.

AC Line Requirements

AC input voltage range

100 to 240 VAC

AC input current

1.0 A

AC input frequency range

50 – 60 Hz

91369 Service Manual

1-2

 

Introduction

 

 

 

Environmental Requirements

Operating temperature

0° to 50° C

Humidity (operating)

10% to 95% relative humidity, non-condensing

Regulatory Approvals

C

®

0123

US

 

CSA certified. Meets IEC60601-1, UL60601-1, and CSA C22.2 No. 601.1 for electrical safety. CE marked in accordance with the Medical Device Directive 93/42/EEC.

Monitor Options

The following options are available:

 

Table 1: 91369 Monitor Options

 

 

Option

Definition

 

 

D

Perioperative

 

 

J

Dual-channel internal recorder (Polish,Czech, Portuguese

language support only)

 

 

 

N

Vital Signs Calculations

 

 

O

Drug Dose Calculations

 

 

Q

Data Shuttle®

R

Patient Data Logger (PDL)

 

 

U

Dual-channel internal recorder (no Polish language support)

 

 

Z

Wireless networking

 

 

04

Four waveform zones

 

 

06

Six waveform zones

 

 

91369 Service Manual

1-3

Introduction

Display

The video input of the display conforms with the Video Electronics Standards Association (VESA) display resolution of 1024 × 768 pixels. The monitor does not support an external touchscreen.

 

Vertical

Horizontal

 

 

 

Rate

64 Hz

51.584 Hz

 

 

 

Front porch

58 μs

0.350 μs

 

 

 

Sync width

116 μs

1.985 μs

 

 

 

Back porch

562 μs

2.101 μs

 

 

 

Blank

737 μs

4.435 μs

 

 

 

Video clock rate

68.5 MHz

51.584 KHz

 

 

 

91369 Service Manual

1-4

Setup

Contents

Unpacking the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

Assembling the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7

Network Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

Power-ON Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

Configuring the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

Unpacking the Monitor

The 91369 monitor, one or two batteries, external AC power supply, and any optional accessories are all packaged and shipped in a single box. Keep at least one shipping box and its packing materials for reshipping, if the monitor should ever require factory service.

Caution:

Observe precautions for handling electrostatic-sensitive devices!

Note:

Never touch electrostatic-sensitive electronic components without following proper anti-static procedures, including the use of an ESD wrist band and mat. An electrostatic discharge from your fingers can permanently damage electronic components and cause latent failures.

All static-sensitive electronic components are packaged in static-shielding bags. Retain the bag for repackaging the component should you need to store it or return it to Spacelabs Medical for any reason.

The monitor is typically shipped as follows:

Top Assembly — Contains the main enclosure with installed CPU, power supply, and I/O PCBAs.

Accessories — Contains the external DC power supply, U.S. power cord, international power cords (if applicable), and any cable assemblies ordered.

Before installing the monitor:

Note:

When removing items from the shipping containers, ensure that you remove ALL components from each container.

1Unpack the received equipment.

2Unpack the mounting hardware.

3Conduct an equipment audit.

91369 Service Manual

2-1

Setup

Upon receiving the equipment, complete a detailed inventory to verify that the equipment you received matches your order. This inventory must include serial numbers, model numbers, and all options and cables received. Carefully inspect these items for shipping damage. If any damage is evident, immediately notify the freight company and Spacelabs Medical.

Assembling the Monitor

Power and Battery Status

The three LEDs on the monitor indicate whether the monitor is connected to the AC mains power and the status of any installed batteries. Battery status conditions are indicated as described in the following sections.

 

 

 

 

Battery

 

status

 

 

 

 

 

Power ON/OFF

Power LED

 

switch

 

 

LEDs

Figure 2-1: Battery status information

Power LED

The power LED is located immediately to the right of the power ON/OFF switch. This LED is lit whenever the monitor is connected to AC mains power via its power supply, and is not lit if the monitor is not connected to the AC mains power.

Battery Status LEDs

These LEDs are only active while the unit is connected to AC mains power.

Unlit LED

A battery LED that is neither solid ON nor flashing indicates a battery is not present.

Solid Green LED

A solid green battery LED indicates that the battery is fully charged. Only a charging cycle or a faulty battery will cause the green LED to flash, and these conditions only occur when a battery is installed in the monitor.

91369 Service Manual

2-2

Setup

Flashing Green LED — Battery Charging

A flashing green battery LED indicates an installed battery is being charged and the monitor is not completely ready to be used in transport mode. This LED flashes in a constant pattern with no delays with the monitor powered ON or OFF. The flashing is different than the battery fault detection flash.

Note:

The green LED stops flashing and stays ON when the charging cycle is complete.

Intermittent Flashing Green LED — Battery Fault Detected

An intermittent flashing green LED indicates that this battery will not hold a charge, or is taking too long to charge. The intermittent signal is a repeating pattern of a solid green LED for one second and a flashing LED for one second. An error message is also added to the error log for review by your system administrator.

To determine whether a battery is faulty, power the monitor ON using the front-panel switch and observe the message displayed along the bottom of the monitor screen. Replace a faulty battery with the same battery type.

Installing/Replacing Batteries

NiMH batteries are used in the monitor. Refer to Figure 2-2 to install one or two batteries.

Figure 2-2: Monitor battery installation

91369 Service Manual

2-3

Setup

While the monitor is operational, a single battery can be exchanged under any of the following conditions without a loss of patient data:

The monitor is being powered by the external power supply.

The monitor is operating on two batteries, and one charged battery remains connected at all times during the exchange.

Warning:

Batteries exposed to short circuit, high temperature, or fire may leak, vent, or explode.

Caution:

Follow the manufacturer’s recommended handling procedure. Collect and transport batteries in a manner that prevents short circuit, compacting, mutilation, or any other abuse that would compromise the physical integrity.

Connections

Refer to Figure 2-3 for available connections on the monitor’s rear panel. Refer to Figure 2-4 for available connections on the monitor’s side panel.

Rear Panel

Equipotential ground post

External power supply input

SDLC port

Alarm relay output

Video output

Serial port

 

 

Figure 2-3: Rear panel connections

91369 Service Manual

2-4

Setup

Table 1: Rear Panel Cables

Rear Panel

Description

Part Number

Connection

 

 

 

 

 

 

Cable, Serial I/O (RS-232)

As required

 

 

 

 

Cable, Video, DB15HD Male to DB15HD Male,

012-0593-00

1.8 m (6 feet)

 

 

 

 

 

 

Cable, SDLC

As required

 

 

 

 

Cable, Monitor to Module Housing, 0.61 m (2 feet)

012-0532-02

 

 

 

 

Cable, Monitor to Module Housing, 1.22 m (4 feet)

012-0532-04

 

 

 

 

Cable, Monitor to Module Housing, 2.44 m (8 feet)

012-0532-08

 

 

 

 

Cable, Monitor to Module Housing, 3.05 m (10 feet)

012-0532-10

 

 

 

Caution:

For continued electromagnetic interference (EMI) radiation compliance, use only cables that have been tested and approved by Spacelabs Medical. Refer to Table 2 on page 6-3 for all cable part numbers.

91369 Service Manual

2-5

Setup

Side Panel

 

USB ports

 

10/100BaseT network

 

connection

Figure 2-4: Side panel connections

Note:

The USB ports are to be used for Spacelabs-approved devices only (Symbol/Metrologic barcode scanner and Microsoft USB keyboard/mouse devices).

Table 2: Side Panel Cables

Side Panel

Description

Part Number

Connection

 

 

 

 

 

 

Cable, Ethernet, 10/100BaseT, 0.94 m (3 feet)

175-0951-00

 

 

 

 

Cable, Ethernet, 10/100BaseT, 1.8 m (6 feet)

175-0951-01

 

 

 

 

Cable, Ethernet, 10/100BaseT, 3.7 m (12 feet)

175-0951-02

 

 

 

 

Cable, Ethernet, 10/100BaseT, 6.1 m (20 feet)

175-0951-03

 

 

 

91369 Service Manual

2-6

Setup

Cabling

Maximum Cable Lengths

The following cables are limited to the indicated maximum length:

SDLC Cable — 12.2 m (40 feet) maximum (total length from the monitor to the last device on the bus). For longer SDLC cable runs, contact a Spacelabs Medical Field Service Engineer.

Video Cable — 30.5 m (100 feet) maximum (total length from the monitor to the last display).

Ethernet cable (10/100BaseT) — 100 m (328 feet) maximum.

SDLC External Devices

External devices (for example, Flexport® system interfaces) can be connected to the SDLC bus. (In this context, the term “external” means connected to the SDLC bus by cable via an external connector. This is in contrast to modules, which are connected by inserting them into a module housing.)

If no supplementary module housings are present (in addition to the module slot integral to the monitor itself), then external devices are connected directly to the SDLC connector of the monitor.

If one or more supplementary module housings are present, Flexport devices are connected to connector J2 on one of the 90499 or 90491 supplementary module housings, or J3 on model 90485. Refer to the Module Housings and Power Supplies Service Manual (P/N 070-0680-xx).

If multiple module housings are present, external devices must be connected to the last module housing in the daisy-chain; that is, the housing electrically farthest from the monitor on the SDLC bus. Even though multiple connectors may be available, only the SDLC connector on the most distal module housing can be used for connecting external devices. Do not use more than a single Flexport connector, regardless of how many module housings are present.

If multiple Flexport interfaces are to be installed, they must be daisy-chained using the T-cable supplied with those devices. Up to three Flexport interfaces may be connected in this way.

Warning:

Unreliable system operation will occur if the SDLC bus is not correctly terminated or the maximum cable length is exceeded. Flexport interfaces must be attached to the most distal module housing on the SDLC bus.

SDLC Cable Interconnection

To ensure electromagnetic interference (EMI) compliance, the appropriate Spacelabs Medical 9-pin connector must be used. Refer to the Module Housings and Power Supplies Service Manual

(P/N 070-0680-xx).

91369 Service Manual

2-7

Setup

SDLC Bus Termination

The SDLC bus must be properly terminated for correct operation. If no external devices (for example, Flexports or multigas analyzers) are connected, proper termination of the SDLC bus is accomplished automatically. If external devices are connected, the switch on the module housing farthest from the monitor must be set to the terminated () position. All others must be set to the non-terminated () position. The SDLC clock and data signals are switched by the terminator switches and are not present “downstream” of any switch set to the position.

Non-terminated Terminated

Figure 2-5: Terminator switch settings

Because bus termination is handled by setting the switches appropriately, an external terminator is only required when external devices are connected.

If external devices are connected, an external cable terminator is required to terminate the SDLC bus. This must be installed at the end of the SDLC bus (following the last external device). In this case, all module housings must have their switches in the position.

Note:

Flexports require a powered Flexport cable (P/N 012-0555-00) when used with the 90491/90499 module housing or 91369 monitor. SDLC data is only passed along to the external device(s) when the terminator switch (SW2) is in the position.

Alarm Relay

Alarm output signals are available at the Nurse Alert () connector instantaneously when an alarm occurs. Table 3 describes the connector pinouts for remote alarms. Figure 2-6, Figure 2-7, and Figure 2-8 illustrate the circuits for each alarm function.

External Alarm Pinout

Alarm connector pinouts are as follows:

Table 3: Connector Pinouts

Pin

Alarm Circuit

Meaning

 

 

 

1

 

Common

 

 

 

2

Alarm 0 (high priority)

Normally Closed

 

 

 

3

 

Normally Open

 

 

 

4

 

GND

 

 

 

91369 Service Manual

2-8

Setup

Table 3: Connector Pinouts (continued)

Pin

Alarm Circuit

Meaning

 

 

 

5

 

Normally Closed

 

 

 

6

Alarm 1 (medium priority)

Normally Open

 

 

 

7

 

Common

 

 

 

8

 

GND

 

 

 

9

 

+12 V, 140 mA

 

 

 

10

 

GND

 

 

 

11

 

GND

 

 

 

12

 

Normally Open

 

 

 

13

Alarm 2 (low priority)

Common

 

 

 

14

 

Normally Closed

 

 

 

Warning:

For operational safety and reliability, the following relay contact ratings MUST NOT BE EXCEEDED:

Current = 250 ma

Voltage = 28 V AC/DC

 

+5 V

 

 

 

common 1

Relay maximum ratings:

 

NC

2

28 V AC/DC

 

 

0.25 A

 

NO

 

 

3

 

 

relay

 

 

4

GND

ALMON

 

 

8

GND

 

 

 

 

10

GND

 

 

11

GND

+12 V

9

140 mA

 

Figure 2-6: Alarm 0 (high priority) relay schematic

91369 Service Manual

2-9

Setup

 

+5 V

 

 

 

common 7

Relay maximum ratings:

 

NC

5

28 V AC/DC

 

 

0.25 A

 

NO

 

 

6

 

 

relay

 

 

4

GND

ALMON

 

 

8

GND

 

 

 

 

10

GND

 

 

11

GND

 

+12 V

9

140 mA

 

 

Figure 2-7: Alarm 1 (medium priority) relay schematic

 

+5 V

 

 

 

common 13

Relay maximum ratings:

 

NC

14

28 V AC/DC

 

 

0.25 A

 

NO

 

 

12

 

 

relay

 

 

4

GND

ALMON

 

 

8

GND

 

 

 

 

10

GND

 

 

11

GND

 

+12 V

9

140 mA

 

 

Figure 2-8: Alarm 2 (low priority) relay schematic

91369 Service Manual

2-10

Setup

Network Installation

A typical network consists of bedside and central monitors and an optional clinical information system (Figure 2-9).

94263

Flat-panel display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

94260-19

 

 

 

 

 

 

 

 

M

 

 

 

 

 

 

 

 

 

 

 

o

 

 

 

 

 

Flat-panel display

 

 

 

 

 

 

 

 

 

 

 

 

 

91387-27

 

 

 

d

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bedside #n

 

 

 

u

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

l

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

e

 

 

 

 

 

 

 

 

 

 

 

 

 

91387-38

 

 

 

 

 

 

 

 

s

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Central #n

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Clinical

 

 

 

 

 

 

 

 

Ethernet Network

 

 

 

Information

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

System

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

91369 Bedside #n

Module

Figure 2-9: Typical network configuration

Warning:

Ensure that the Ethernet wall plate and the shield of the Ethernet connecting cable are bonded to the hospital grounding system.

Ethernet Network Connection

Caution:

Only qualified personnel should attempt to connect a monitor to an Ethernet LAN the first time.

Do not connect the monitor to an Ethernet local area network (LAN) prior to configuring the following settings. The monitor must be properly configured for LAN access before you operate the monitor. If you fail to correctly configure the monitor, you may interrupt other units also using the LAN.

Note:

Detailed installation instructions for the physical Ethernet LAN are beyond the scope of this document.

91369 Service Manual

2-11

Setup

To connect a monitor onto an existing Spacelabs Medical Ethernet LAN, complete the following steps:

1Install the monitor on a suitable table or shelf, ensuring that the air flow to the side air intake vents is unobstructed, or use a Spacelabs Medical mounting option.

2Ensure that the monitor is not connected to the LAN.

3Plug the power cord attached to the monitor’s DC power supply into a standard hospital-grade AC power supply.

4Power ON the monitor.

5Enter a unique MONITOR ID, BED NAME, and SUBNET for the monitor. Refer to Network Setup on page 2- 13 for more information.

6Attach the 10/100BaseT LAN transceiver cable into the RJ45 connector on the left side of the monitor ( in Figure 2-4 on page 2-6).

7Connect the other end of the Ethernet cable from the monitor to the nearest port.

8Configure the monitor’s other network settings as necessary to ensure proper communication on the network. Refer to Network Setup on page 2-13.

Ethernet Network Disconnection

To remove a monitor from the LAN, disconnect the network cable from the 10/100BaseT network connection ( in Figure 2-4 on page 2-6).

Power-ON Test

Each time the monitor is powered ON:

Diagnostic information displays for approximately 10 seconds.

The embedded alarm light cycles red and yellow. Some models may cycle red, yellow, and cyan.

Monitor keys display on the right side of the screen.

The monitor is now ready for normal operation.

External Devices

If an external SDLC device, such as a Flexport interface, is to be installed, the 9-pin SDLC connector on the rear of the monitor or the module housing must be used. If multiple SDLC ports on module housings are available, only the SDLC port on the module housing farthest from the monitor can be used for external devices. Set the termination switch to non-terminated () for all module housings and then terminate the external device.

91369 Service Manual

2-12

Setup

Module Tests

To verify that the monitor functions correctly with parameter modules:

1Insert an ECG module without the patient cables connected. Verify that the ECG parameter key is displayed.

2Connect a patient simulator to the ECG input with a 5-lead patient cable, and set the simulator to a known rate.

Verify that the heart rate and lead being monitored are displayed to the right of the ECG parameter key.

Verify that the ECG waveform is displayed.

3Disconnect the patient cable. After 10 seconds, verify that the LEADS OFF message appears, the parameter key flashes, and an alarm tone sounds.

4Reconnect the patient cable and verify that the LEADS OFF message clears and the alarm stops.

5Connect a patient simulator to the invasive pressure inputs.

6Zero the pressures and verify that the numerics and waveforms are accurate.

7Verify that the key tone sounds each time a key is selected.

Configuring the Monitor

The Biomed Level menu displays when the biomed password (default is biomed) is entered into the Privileged Access window. Refer to Directory of Keys on page 7-1 for the Biomed Level menu structure.

Network Setup

Note:

The NETWORK SETUP key only displays on monitors that are configured for network operation.

Touch NETWORK SETUP to display the Monitor Setup - Network Configuration window. This window contains an on-screen keyboard and three tabs: TCP/IP, Monitor, and Printers. Proper network operation requires that each device on the network have a unique network address, monitor ID, and monitor name. If the wireless option has been installed on the monitor, three more tabs will be present: WLAN, Security, and

Advanced.

Editing Tab Fields

The fields within a tab on the Monitor Setup - Network Configuration window can be edited by selecting the field and entering new information using the on-screen or optional external keyboard.

When editing, adding, or deleting, press ENTER or TAB to cycle to the next input cell. Any changed or added items are stored in the monitor’s non-volatile memory when SAVE is selected. The description of each tab indicates when that change takes effect (for example, immediately or after a monitor reset occurs).

To edit text within a tab:

1Select an item from the list. -OR-

91369 Service Manual

2-13

Setup

2Select an input cell’s text and type any combination of letters, characters, or spaces.

To add an item to a list:

• Select the input cell and type the new information.

To delete an item from a list:

1Select the item.

2Touch Del.

3Enter at least one space (an error message is displayed if no spaces are entered).

TCP/IP Tab

Figure 2-10: TCP/IP tab

The TCP/IP tab enables you to define the monitor’s attributes for networking.

IP Address — Enables you to specify the monitor’s IP address. The default is 164.90.254.10. Either a static IP address must be specified, or DHCP networking must be enabled.

Subnet Mask — Enables you to identify which parts of the IP address are to be used for TCP/IP subnet determination. The TCP/IP network’s subnet mask is not related to the Spacelabs Medical network’s subnet name. The standard and factory-default subnet mask is 255.255.255.0. Either a subnet mask must be specified, or DHCP networking must be enabled.

Gateway Address — Enables you to specify the IP address of the TCP/IP gateway (bridge or router) though which communication to other devices should flow. The default is blank.

DHCP — (Dynamic Host Configuration Protocol) Used to configure and enable DHCP network configuration. When DHCP is enabled, IP Address, Subnet Mask, and Gateway Address are automatically filled in. To use this service, a DHCP server must be available on the network to respond to DHCP requests.

91369 Service Manual

2-14

Setup

Note:

A DHCP lease is a TCP/IP configuration given out from the DHCP server that is valid for a period defined by the DHCP server or forever (no expiration).

Monitors configured for DHCP operation request a lease from the DHCP server when they boot up or when their existing lease expires during operation. If the DHCP server is not present, the monitor checks the expiration time of the last DHCP lease obtained. If the lease is still valid, the monitor continues to use those values and operates normally. If the lease has expired, the monitor disables TCP/IP networking and displays a NETWORK SIGNAL LOST message to indicate that it is unable to communicate over the network. The monitor continues to request a DHCP lease until it receives one.

If the monitor’s configured DECNET node ID is a duplicate on the network, the DHCP server can be confused. This may result in a duplicate or invalid DHCP lease and may prevent full network communication.

The subnet mask must correctly correspond to the network size and type during operation. Monitors may not be able to fully communicate with each other if the DHCP server fails to set the network mask properly.

The DNS server address is in standard TCP/IP address form, while the DNS default domain is a string of ASCII characters. A DHCP server may also provide this information.

Editing this tab is performed as described in Editing Tab Fields on page 2-13. Tabbing order is IP Address >>

Subnet Mask >> Gateway Address >> DHCP >> DNS Server Address >> DNS Default Domain >> IP Address. Changes made to this tab’s settings take effect after a successful save and monitor reset.

Secondary Display — The Secondary Display field is only available if option -D, Perioperative, has been activated.

Select ENABLE on the TCP/IP tab to activate the primary monitor’s secondary display. When you enable the secondary display, the IP Address and Secondary Hostname fields become available and must contain a valid entry for the secondary monitor to operate. The IP Address and Secondary Hostname fields are exclusive to each other.

Enter a valid IP Address when your network is using static IP addresses. -OR-

Enter a valid Secondary DNS hostname when your network is using DHCP for network configuration. Changes made to the Secondary Display settings take effect after the changes have been saved.

91369 Service Manual

2-15

Setup

Monitor Tab

Figure 2-11: Monitor tab

The Monitor tab enables you to determine what monitor names, monitor (node) IDs, and subnet names are currently in use. The tab also enables you to enter settings for the monitoring network.

In Figure 2-11, the monitor’s current settings are displayed in each input cell. The scroll list below each input cell displays the remaining items that have been detected on the network. Items in the Subnet Name and Monitor Name lists are displayed in alphanumeric order. Items in the Monitor ID list are displayed in numeric order.

Brackets (< >) surround strings that consist solely of spaces. The separation within the brackets indicates the number of spaces within that string.

Editing is performed as described in Editing Tab Fields on page 2-13. The tabbing order is Subnet Name >>

Monitor Name >> Monitor ID >> DECNET/IP >> Network >> TTL >> Subnet Name. Network and TTL are skipped if unavailable. Changes made to this tab’s settings take effect after a successful save and monitor reset.

Subnet Name

The subnet name contains up to five characters (default is five blanks). Items in this scroll list are selectable.

Monitor Name and Monitor ID

The Monitor Name is the name given to each bedside and central monitor (does not apply to telemetry bed names) to help the users identify monitors on the network. The Monitor Name contains five characters (default is SL001).

The Monitor ID is the numeric ID assigned to a monitor. Each device on the network must have a unique Monitor ID. This can be any number from 1 to 1023, depending on the Network Size selected.

91369 Service Manual

2-16

Setup

To prevent duplication of currently used monitor names and IDs, items in these lists are not selectable. The error checking procedure performed when SAVE is selected also specifically checks for duplications.

Note:

Items in the Monitor Name and Monitor ID lists only display when the monitor is connected to the network.

When entering a monitor name or ID, do not use a space between characters.

DECNET/IP

You can configure the monitor to operate using either Spacelabs DECNET or TCP/IP network protocols. If you are communicating with 903xx Spacelabs Medical monitors, you must select DECNET.

Network

The IP multicast group number of the monitor provides a filter to logically isolate one monitor from another on TCP/IP installations. Up to 32 network numbers are available (0 to 31) with 0 as the default.

Note:

Monitors must use the same network number to communicate.

This is unavailable if DECNET is selected.

TTL (Time to Live)

The allowed number of hops the IP packet can take across network devices. TTL values are 1 to 64, with 16 as the default.

Note:

TTL is unavailable if DECNET is selected.

Network Size

The network size allows configuration as:

64 — Monitor IDs from 1 to 64 are supported. No more than 64 total monitor devices can be on the network. Provides complete network compatibility with legacy Spacelabs Medical monitors.

250 — Monitor IDs from 1 to 250 are supported.

Note:

903xx monitors must have the Expanded Network option installed or they will not communicate correctly with devices with Monitor IDs above 64.

640 — Monitor IDs from 1 to 127 and from 512 to 1023 are supported with the following restrictions.

All model 903xx Spacelabs Medical monitors must use monitor IDs 1 to 127, inclusive.

All model 91xxx monitors must be configured with monitor IDs from 512 to 1023, inclusive.

1000 — Monitor IDs from 1 to 1023 are supported (compatible only with Spacelabs Medical 91xxx series monitors).

91369 Service Manual

2-17

Setup

Printers Tab

The Printers tab enables you to display controls for defining and storing printer names, choosing rules for printer selection, and selecting one or two network printers.

Figure 2-12: Printers tab

Editing is performed as described in Editing Tab Fields on page 2-13. The tabbing order is AUTO / PRIM/ BACKUP >> Printer A (or Primary) >> Printer B (or Backup) >> Printer Names (refer to Figure 2-13 on page 2-19). Changes made to this tab’s settings take effect after a successful save and monitor reset.

Printer Names

The Printer Names list displays up to eight, selectable printer names previously stored in the monitor, in the order in which they were stored. To display a new or changed name in the list, select that printer name from the list. Printer names contain up to five characters (default is five blanks).

Note:

Printer names are explicitly entered and may be duplicated. To clear a printer name from the list, select that name, touch Clear, touch Enter, and then touch SAVE.

A local (SDLC) printer can be either a bedside printer or network printer, depending on the printer name selected in this list. A local printer is configured as a network printer if the local monitor’s name is selected. Otherwise, a local printer functions as the bedside printer.

91369 Service Manual

2-18

Setup

AUTO / PRIM/BACKUP

The AUTO / PRIM/BACKUP key selects which set of printer selection rules the monitor uses for selecting network printers. It does not affect the monitor’s selection of whether a networked or non-networked printer is used. Any changes made to the printer selection mode using this key take effect immediately, regardless of monitor type. The default setting is AUTO, which selects the destination printer using the weight-based printer selection rules. (Refer to the Ultraview SL Operations Manual, P/N 070-1150-xx, located on CD-ROM

P/N 084-1101-xx for additional information).

When PRIM/BACKUP is selected, the monitor automatically selects the primary printer, unless that printer is unable to accept the recording. In that instance, the monitor then selects the backup printer. If the backup printer is also unable to accept the recording, the monitor displays an Unable to record message.

Printer Selection Fields

The default for the printer selection fields is blank (i.e., no printer selected). The labels above these fields vary, based on the current setting of the AUTO / PRIM/BACKUP key. If AUTO is selected, the labels are Printer A and Printer B. If PRIM/BACKUP is selected, the labels are Primary and Backup (Figure 2-13).

Figure 2-13: AUTO / PRIM/BACKUP selection differences

To define a printer name:

1Select a printer name from the Printer Names list.

2Select one of the two printer selection fields. The new value displays in the selected field.

WLAN Tab

This tab allows the operator to set basic wireless local area network (WLAN) related settings.

The WLAN tab only displays on monitors that are configured for wireless operation (option -Z). Monitors are configured for wireless operation by selecting the sysgen RADIO option. The RADIO ON/OFF keys and all other input cells and combo boxes illustrated in Figure 2-14 are not available for selection if the radio is not present.

Caution:

Do NOT power the radio ON until all network configuration has been completed. Powering the radio ON with the monitor set to the factory defaults could interfere with the wireless network.

91369 Service Manual

2-19

Spacelabs 91369 User manual

Setup

Figure 2-14: WLAN tab

The tabbing order is RADIO ON/OFF >> SSID >> Region >> Outbound Parameters >> Wireless DHCP >> WLAN IP Address >> WLAN Subnet Mask >> Gateway Address >> RADIO ON/OFF. WLAN IP Address, WLAN Subnet Mask, and Gateway Address are not available in the tabbing order if Wireless DHCP is selected. Changes made to this tab’s settings take effect after a successful save and monitor reset.

RADIO ON/OFF (default = OFF)

The RADIO ON/OFF setting defaults to OFF when a radio is present. If this key is toggled to ON, and SAVE is selected without a valid region defined [refer to Region (default = blank) on page 2-22], an error box displays (the radio cannot be switched to ON if the Region setting is invalid). Select OK to remove the error box. The RADIO ON/OFF key automatically toggles back to OFF.

The radio will function only if this key is set to ON. Keys, input cells, select boxes, and combo boxes are enabled without regard to the setting of this key. However, WLAN communication only occurs if ON is selected.

Note:

Any change in the wireless tabs (except for the setting of the OUTBOUND PARAMETERS input cell) requires a monitor reset before the changes take effect.

To activate the radio-only functions, select ON.

Wireless DHCP (default = OFF)

The Wireless DHCP tab is similar in function to the DHCP key on the TCP/IP tab, but for wireless networks. When Wireless DHCP is selected, the monitor asks the wireless DHCP server for the following information:

WLAN IP Address

WLAN Subnet Mask

Gateway Address

91369 Service Manual

2-20

Setup

The affected input cells then display the values provided by the wireless DHCP server, rather than the values from the monitor’s non-volatile memory. These cells then become unavailable for selection, to prevent these values from being changed, until Wireless DHCP is de-selected.

If Wireless DHCP is selected, as in Figure 2-14 on page 2-20, DNS Server Address and DNS Default Domain values display, and the WLAN IP Address, Wireless Subnet Mask, and Gateway Address input cells are unavailable for selection.

When Wireless DHCP is not selected, the DNS Server Address and DNS Default Domain values disappear, and the WLAN IP Address, Wireless Subnet Mask, and Gateway Address input cells are available.

Table 4: Wireless DHCP Input Cells

 

 

Is this cell

 

 

Can the wireless

unavailable if the

Can the cell be left

Cell name

DHCP server provide

wireless DHCP

blank?

 

this information?

server provides the

 

 

 

 

information?

 

 

 

 

 

WLAN IP Address

Yes

Yes

No

 

 

 

 

WLAN Subnet Mask

Yes

Yes

No

 

 

 

 

Gateway Address

Yes

Yes

Yes

 

 

 

 

Note:

Input cells that display wireless DHCP server-provided data may display values different than what is stored in the monitor’s non-volatile memory. DHCP server-provided values are not stored in non-volatile memory.

SSID (default = blank)

The SSID cell is used to define the WLAN Service Set Identifier (SSID) setting, which is the name used by acceptable access points (AP) for that WLAN. This field is up to 32 characters long, and it may contain any combination of case-sensitive characters. The default is blank.

Setting SSID and Security Mode

If enabled, WLAN card communication through the wireless network depends on the interaction between the WLAN card SSID setting (refer to SSID (default = blank)) and its security settings (refer to Security Tab on page 2-24).

91369 Service Manual

2-21

Setup

The WLAN card SSID and security settings must match the settings of the intended access point (AP). If the settings do not match, then wireless communication cannot occur.

Table 5: WLAN Card and Security States

 

Security

 

WEP

Monitor

Monitor

Can Wireless

SSID

Authentication

Keys

Communication

Mode

Associated?

Authenticated?

 

 

Match?

Occur?

 

 

 

 

 

 

 

No

WEP

Open

N/A

No

No

 

Match

Disabled

 

 

 

 

 

 

 

 

 

 

 

 

 

No

WEP

Open

N/A

No

No

No — SSID must

Match

Enabled

always match

 

 

 

 

 

 

 

 

 

 

 

No

WEP

Shared Key

N/A

No

No

 

Match

Enabled

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes, but only if

SSID

WEP

Open

N/A

Yes

Yes

Security Mode on

Matches

Disabled

AP matches (WEP

 

 

 

 

 

 

 

 

 

 

optional)

 

 

 

 

 

 

 

SSID

WEP

Open

Yes

Yes

Yes

Yes

Matches

Enabled

 

 

 

 

 

 

 

 

 

 

 

 

SSID

WEP

Open

No

Yes

No

No

Matches

Enabled

 

 

 

 

 

 

 

 

 

 

 

 

SSID

WEP

Shared Key

Yes

Yes

Yes

Yes

Matches

Enabled

 

 

 

 

 

 

 

 

 

 

 

 

SSID

WEP

Shared Key

No

No

No

No

Matches

Enabled

 

 

 

 

 

 

 

 

 

 

 

 

Region (default = blank)

This input cell displays the monitor’s current setting for the WLAN region. This cell is available for 802.11 radios that are configurable in the field. Region values and their corresponding geographical regions are listed in Table 6. The frequencies available in a particular country differ according to the regulations of that country. Consult your national and local regulations.

Table 6: Region Codes

Value

Center Frequency Range

Region

 

 

 

Blank

 

Disabled

 

 

 

1

2412 to 2472 MHz

Europe (ETSI), except France

and Spain

 

 

 

 

 

2

2412 to 2462 MHz

North America

 

 

 

91369 Service Manual

2-22

Loading...
+ 120 hidden pages