Smiths System 250 User manual

Service
System
250
Manual
Fluid
Warmer
:
Unpacking Filling
Electricat
Alarm
Performance
Maintenance
Troubleshooting
Limited Service Specifications...
Electrical
Parts Mechanical
and
with
Distilled
Safety
Power
Leakage
Testing
Testnd
Inoperable Flow
Restriction,
Warranty
Schematics
List,
Maintenance
Parts
Set-Up
Reguirement.
Current
.
List
..
Water
ii
System
and
.
Test.
250
|.V.
Disposable
User
Sets
Replaceable
Parts
6
Revised
REV
A
May
1993
215)
Purpose
The
Level 1 System
patients.
The
Minimum are Level 1 1.V.
board infusion
warming
Disposable
standard
delivery
delivered
safety
of
for
250
sets
Disposable
temperature
near
37°C.
Sets,
and
and
monitoring
normothermic
and
|.C.U.
is
designed
specifically
Set
allows
at
Blood
and
will
be
delivered
systems
fluids.
Emergency
for
designed
infusion
these
packed
make
Typical
Room
safe
and
for
of
6°C
flow rates
RBCs
near 37°C.
the
Level 1 System
uses
of
use.
rapid
various
blood
is
35°C.
from
ail
High
the
system
in-line
warming
infusion
up
to
At
the
auto
warming
of
rates
are
100
ml/min.
somewhat
transfusion
and
flow
250
the ideal
include
all
|.V.
fluids
available.
or
6°C
crystalloid
lower
flow rates
systems
capacity,
choice
routine
surgery
as
may
simple
for
they
are
generally
be
infused
all
procedures and
_
administered
up
to
150
used,
through
operation,
blood
to
mMmin.
fluids
the
and
on-
requiring
and
fluid
Messages
improper
results.
Messages
cause
improper
Messages
cause
improper
Concept
The
Level 1 System
makes used
The
rapidly
rapid System
Taking
very
available
in
the
System
circulated
water
250.
advantage
cold
blood.
that
are
headed
that
are
headed
results
that
are
headed
results,
to
all
Open
Heart
250
employs a Disposable
counter-flow
circulation,
of
The
Filter
and
250
other
the
by
by
damage
by
damage
is
the
areas
Suite
for
and
counter-flow
high
Holder
“NOTE:"
indicate
“CAUTION:"
to
the
“WARNING:"
to
the
equipment,
first
significant
of
the
on-pump
through
warming
of
an
capacity,
the
System
information
indicate
equipment.
indicate
advance
hospital
warming
Heat
Exchanger.
anodized
warming
the
250
or
information
information
injury
to
personnel,
in
routine
the
same
for
over
25
As
aluminum
all
contribute
Disposable
is
warmed
procedures
or
procedures
or
procedures
or
blood
fluid
warming
years.
with
extracorporeal
tube.
The
to
the
Sets
allow
to
help
maintain
that
if
not
that i not
`
that
serious
and
use
excellent
patient
fluid
technology
of a thermal
increased
peak
followed
followed
if
not
followed
harm.
warming
Perfusionists
warming,
conductor
high
warming
fluid
efficiency
correctly
correctly can
correctly
in
over
warmed
capacity
flow
rates,
at
low
can
cause
can
20
years.
have
water
It
safety
is
(aluminum),
of
the
even
with
flow
rates.
The
Hardware
and
easy
to
use.
the
system
safely
Safety
FILTER
The
released
delivered
WITH
Level 1 Disposable
from to
the
and
Disposables
By
following
set-up
GAS
VENT
fluids
as
patient
and
Sets
they
along
of
the
Level 1 System
the
1-2-3
instructions
operating
employ a unique
are
with
within
warmed. the
warmed
seconds.
In
many
fluids.
Gas
found
Vent
other
250
on
which
warming
are
integrated
the
unit,
vents
into
properly
micro-bubbles
systems,
the
design
trained
these
to
be
highly
users
can
confidently
of
gas
which
micro-bubbles
effective
are
always
are
actually
have
The
ability
Gas
Vent
with
luer-locks Fitter vent
replacement
For
D-50
For
D-100
with
of
in
gas
Set
Set
the
the
and
vent
order
order
Gas
Vent
I.V.
Sets
is
easily
should
fitters
are
replacement
replacement
to
quickly
contains a 170-micron
changed
be
changed
available
fitter
without
as
F-10
filter
F-10
vent
air
every
follows:
patient
three
also
screen
permits
hours
rapid
filter
to
trap
disconnection
to
maintain
priming
of
the
macro-aggregates.
or
repriming
efficient
system
Disposable
This
of
the
entire
performance.
Set.
The
filter
is
disposable
Fitter
Fitter
with
connected
set.
he
with
gas
TEMPERATURE
The
System
temperature
this
circuit, a second
temperature
Fluid
present
temperatures.
ELECTRICAL
A
special
through
low
the
power
CONTROL
250
employs a safe
control
reaches
in
the
circuit
“watchdog”
about
heat
SAFETY
leakage
cord
limits
41°C.
exchanger
immersion
providing
circulating
the
circulating
circuit
will
while
heater
an
is
extra
Unpacking
The
Level 1 System
Contents
water
water
visually
the
unit
used
measure
heating
to a 40°C
and
audibly
is
operating
in
the
Level 1 System
of
patient
and
Set-Up
250
system,
maximum.
alarm
is
and
is
packaged
inherently
and
never
exposed
250.
operator
and
tn
the
stop
This
safety.
partially
Quantity
free
of
unlikely
the
circulating
to
any
heater
assembied.
"hot
spots".
event
of a malfunction
water
damaging
is
electrically
The
primary
pump
or
dangerous
grounded
if
of
the
Figure
UNIT
PACKAGING
Power
Rolling
ACCESSORY
Service Operators Casters
ГУ IV. Fill
Set
Level 1 Video
Hitch
1
Вад Pole
Plug
Up
Pin
Unit
Base
Manual
Manual
Напдег
Knob
Card
1
1
PACK
a
da
WARNING
This
unit
prior
to
1,
With
wheels.
2.
Slide
base
as
3.
insert
must
placing
the
Rolling
the
Power
shown
the
PIN
be
assembled
the
device
base
Unit
in
Figure
through
into
on
Pole
1.
the
and
service.
its
side,
down
holes
tested
push
over
in
the
by
Level 1 technical
the
four
the
vertical
bottom
of
casters
square
the
pole
FULLY
base
and
personnel
into
their
tube
with
vertical
or
sockets.
Power
base
an
authorized
Unit
tube,
Place
facing
locking
distributor
the
base
the
longer
them
together.
of
Level
upright
side
of
on
1
its
the
Filling
Set-Up
CAUTION
For
WATER
water
operating
DO
creates
A.
B. port
in
Figure
way
port
proper
in
the
NOT
an
Unscrew
Fil
with
exits
3.
to
be
and
fill
with
system
should
unit
efficiency.
FILL
air
and
1.3
at
the
This
sure
the
Distilled
operation
be
used
will
build
the
unit
lock
preventing
remove
liters
of
bottom
vent
port
the
reservoir
reservoir
Water
and
in
the
up
mineral
with a DISPOSABLE
proper
the
Fillport
Distilled
of
the
also
is
until
water
component
water
Plug.
Water.
Power
serves
full
is
begins
life,
reservoir.
scale
and
SET
filling.
Note: A water
Module
as
to
at
an
overflow
place a cup
to
exit
only
DISTILLED
The
use
seriously
IN
PLACE
reservoir
the
location
tube.
under
the
port.
of
reduce
as
this
vent
shown
An
easy
this
vent
tap
its
Fillport
Plug
Figure
2
C.
Replace
CAUTION
The
Distilled
manual.
sure
there
The
are
the
Fillport
Water
System
no
in
250
large
Plug
the
reservoir
draws a maximum
power
consumers
should
on
be
of
7.5
that
changed
Amps.
circuit.
every It
should
30
days.
be
plugged
See
the
Maintenance
into a 10
to
20
section
Amp
of
circuit,
being
this
Electrical
Safety
Power Requirement
Most
of
the
current
600
Watt
immersion
turned
temperatures
full
proportional proportionally water
Leakage
on
and
ON
condition.
nears
controller
the
Current
drawn
the
digital
below
Beyond
shorter
40°C
heater.
30°C,
cycles
target
Test
NOTE This
device
device
the
Parts
WARNING DO
NOT
WITHOUT A DISPOSABLE
and
is
equipped
perform
List
section
DEFEAT
leakage
of
THE
΄
by
the
System
When
display
the
shows
600
approximately
the
heater
ON
times temperature.
with
disposable
current
this
manual.
DISPOSABLE
HEAT
250
the
device
rapidly
Watt
heater
30°C,
ON/OFF
as
the
circulating
sensing
testing.
EXCHANGER
Non-Sterile
SENSING
is
for
its
is
first
rising
is in
a
the
with
interlocks. A Level 1 Heat
Test
Sets
are
INTERLOCKS
IN
PLACE.
Exchanger
available
OR TRY
Figure
from
Level 1 for
TO
Air
intake
3
is
required
this
OPERATIVE
Vent
Port
Slots
to
operate
purpose.
THE
the
See
UNIT
Power in
be
warms
Although less
Units
repair.
On
order
performed
up
than
exceeding
Alarm
NOTE
Alarm avatlable
testing
Leakage
to
be
quickly,
the
specifications
50
microamps.
sure
on
100
Current
the
immersion
units
test
readings
microamps
which
Testing
requires a Level 1 Heat
from
Level 1 for
this
Tests
have
should
claim
purpose.
should
heater
less
electrical
circuit
been
be
than
Exchanger
See
be
allowed
taken
100
leakage
the
performed
is
in a full
to
stand
within 1 minute
microamps
should
to
be
Parts
List
with
the
ON
condition,
until
electrical
be
removed
in
place
in
this
immersion
Power
close
to
of
startup.
leakage,
from
on
the
manual.
heater
On
room
temperature.
units
service
System
circuit
Leakage
leaving
and
250.
Non-sterile
in
the
full
Current
As
the
the
factory
returned
ON
condition.
Tests
should
System
are
usually
to
Level 1 for
Test
Sets
250
are
Add
Water
The
System
water
level
condition
Alarm
250
is
too
the
circulating
is
equipped
low, a RED
pump
with a float
L.E.D.
should
on
not
switch
the
Display
be
running.
which
senses
Panel
the
water
illuminates
level
in
the
and a sonic
on-board
alarm
reservoir,
sounds.
In
When the
the
alarm
With a Test
tube
on
the
Replace
Check
Two
and
running,
1.
2.
Over
any
Disposables
interlocks
the
Power
each
Top
Heat
Heat
Exchanger
Temperature
With a Test
the
water
Fil/Drain
DISTILLED
this
manual.)
illuminate
above,
normal
and
turn
operation.
Heat
Exchanger
back
of
the
Power
water
drained
Alarm
sense a property
Switch
interlock
Exchanger
is
may
Socket - Pull
Guide - Bend
Alarm
Heat
Exchanger
tube
on
WATER
the
Turn the
Power
which
the
Sonic
Switch
Power
Alarm
in
place,
Unit
with
DISTILLED
installed
ON,
the
RED
be
independently
tube
in
place and
the
back
has
been
Switch
should
OFF
and
the
ADD
and
tilting
Heat
indicator
Release
portion
the
of
the
heated
ON.
sound.
allow
WATER
it
down
WATER.
Exchanger.
tested:
knob
of
Heat
Power
Power
to
45°C
The
Over
The
the
unit
ALARM
to
See
will
illuminate
and
slide
Exchanger
Switch
Unit
and
to
48°C.
temperature
water
circulating
to
cool
may
allow
some
the
Filling
If a properly
and a pulsed
up.
out
OFF,
drain
tilting
it
(See
FILLING
pump
to
room
be
tested
of
the
water
with
Distilled
installed
of
Guide
down
RED
groove.
the
water
to a container.
WITH
L.E.D.
should
temperature.
by
releasing
to
drain
Water
Heat
Exchanger
alarm
will
out
of
the
DISTILLED
on
the
not
be
Restart
the
out
section
sound.
reservoir
Fill
WATER
Display
running.
the
water
Fil/Drain
into a container.
of
this
manual.
is
not
in
place
With
the
unit
by
releasing
with
1.3
liters
of
section
Panel
Confirming
unit
and
of
should
the
confirm
NOTE
The
sonic
alarm
Performance
NOTE
Performance Disposable
THREE
1.
COLD
Store
in
piace,
Rapidly
minutes
simpie
the
rising
on
Test
START
System
note
an
and
all
RED
Testing
testing
the
numbers
efficiently
requires a Level 1 0-50
Sets
are
tests
allow
TEST
250
unit
time
and
will
operating
L.E.D.s
available
periodic
in a room
turn
checking
the
appear
unit.
may
from
Level 1 for
where
Power
on
be
the
tested
Disposable
of
the
the
room
Switch
Water
by
pressing
Set
this
purpose.
overall
operating
temperature
ON.
The
Temperature
the
to
See
Green
alarm
test
be
in
place
the
Parts
efficiency
is
approximately
System
display.
37°C
button
of
on
List
the
21°C
on
the
System
section
System
Operational
will
appear
the
back
250.
in
this
250.
(70°F).
With a D-50
indicator
in
approximately
of
the
cabinet.
Non-sterile
manual.
will
illuminate.
Set
3
2.
SYSTEM
RECOVERY
TEST
Chill a 1
allow
the
place
the
Pole
so
at
250
mUmin.
Note
the
display.
not
read
3.
STEADY
Thermal
For
use
NOTE
Use
of
this
accurate
liter
bag
of
unit
to
reach
Male
Luer
the
fluid
bag
(This
time
and
OPEN
The
fluid
bag
below
37°C.
STATE
Calibration
with:
within
Level 1 Fluid
unit
requires a digital
0.1°C.
Normal
in a container
may
RUN
Well
at
least
outlet
be
the
should
TEST:
(TCW)
Saline
39.5°C.
is
4"
confirmed
clamp
empty
Warmers
thermometer
to
10°C
to
Remove
on
the
floor.
above
the
level
with a graduated
below
the
in
about 4 minutes,
'.
Model
15°C.
the
Close
of
bag,
P/N
H-500
H-500 H-250 H-250
NIST
With a D-50
cap from
the
Bag
the
Heat
Exchanger
beaker
allowing
80-03-002
traceable
and
INT
INT
the
the
and
Test
the
Spike
and
fluid
digital
Set
in
place,
Male
Luer
clamps,
inlet.
stopwatch.)
to
flow
freely.
display
at
and
With
in
the
spike
an
turn
the
Power
end
of
the
the
fluid
all
clamps
Watch
the
efficiently
Switch
Patient
bag.
open,
fluid
Water
Temperature
operating
Set
ON
Line
the
will
unit
and
and
LV.
flow
will
Required
Short
a.
Install
(sockets 1 &
“For
b.
Turn thermometer Compare
Unit
correctly
Examples
Thermometer
Level 1 Unit
probe
checklist
TCW
H:500 & H-500
the
temperatures.
display
calibrated
size:
to
verify
in
machine
2).
machine
probe
will
be
39.750
WITHIN
0.099"
0.50" - 1.50"
displayed
INT,
on
and
fully
into
within
unit.
THIS
in
normal
also
install a test
allow
well
0.3°C
of
RANGE
OD
maximum
long
circulating
heat
to
warm
and
allow
the
thermometer
39.4°C - 40.0°C = OK
water
exchanger
filter
at
up
for
15
display
temperature:
position
position
to
display
3.
minutes.
stabilize.
on
Insert
a
7
pe
|
3
1
Thermometer Levelt
NOTE
IF
SERVICE.
THE
Unit
SYSTEM
Maintenance
40.0°C WITHIN
250
THIS
DOES
RANGE
NOT
39.7°C - 40.3°C = OK
MEET
1,
2,
or
3,
IT
SHOULD
BE
RETURNED
TO
THE
FACTORY
FOR
CLEAN
Clean
and a soft
discolor
CHANGE
Release
container.
water.
GREASE
Place a small Rings
sockets
CHANGE
Note: the socket
Ring Caps
EXTERIOR
the
entire
some
DISTILLED
the
See
O-RING
in
the
to
O-RING
An
O-Ring
Parts
List
and
sockets.
and
button
cloth
Replace
FILLING
grease
remove
System
after
of
the
water
amount
bottom
Kit
jn
this
Coat
head
--
EVERY
every
plastics
WATER
Fil/Drain
the
tube
WITH
SEALS
of
and
top
them.
SEALS
containing 2 O-Rings, 1 1/8"
manual.
the
rectanguiar
the
new
screws.
USE
250
with a spray
use.
Cold
on
the
--
tube
by
pushing
DISTILLED
--
EVERY
silicone
heat
--
EVERY
Using a 1/8"
O-Rings
sterilizing
device.
EVERY
behind
firmiy
WATER
30
DAYS
grease
exchanger
on a cotton
12
MONTHS
Socket
with
of
Do
30
the
hex
Caps.
silicone
warm
solutions
not
DAYS
Power
up
into
section
sockets
hex
wrench,
After
grease
soapy
use
Unit
the
swab
of
wrench
unscrew
removing
water
may
be
cleaning
and
tilt
fitting
of
this
or
the
the
Pole
and
and
press
solution
used,
agents
it
down
until
it
bottoms.
manual.
end
of a smali
Assembly.
silicone
the
four
and
discarding
them
or a commercial
but
strong
containing abrasives.
to
allow
It
grease
stainless
into
the
Refill
finger
is
not
is button
the
their
chemical
non-abrasive
concentrations
water
to
with
1.3
and
apply
necessary
available
head
old
O-Rings,
sockets.
drain
into
liters
of
all
around
to
disassemble
from
Level
screws
clean
Replace
cleaner
may
a
distilled
the
1.
See
of
each
the
O-
the
Socket
O-
the
GENERAL
INSPECTION
--
EVERY
USE
Visually
service
Disposable
This
bottom Pole
check
which
unit
is
of
mount ( figure
the
shows
Set
cooled
the
cabinet
condition
physical
does
not
by
4)
of
damage
install
convection.
(refer
to
are
not
blocked.
Air
Vent
Figure
4
the
device.
easily.
Be
Figure
or
one
sure
3)
and
Remove
in
which
the
air
air
any
the
intake
vent
below
unit
Test
slots
from
on
the
the
Ι.Ν.
Troubleshooting
Inoperable System
PROBLEM
NO
POWER
NO
GREEN
OPERATIONAL
and
CHECK ALARM
ADD WHEN
DISPOSABLES
WATER
UNIT.IS
250
SYSTEM
ALARM
MOVED
LIGHT
CHECK
Unit
is
PLUGGED
Power
Minimum
Note:
from Switch will
Heat properly
Pole
property
Switch
If
power
the
is
GLOW.
Exchanger
installed?
interlocks
engaged
Exchanger?
Add
about
water
to
the
Fill
Port.
ON?
10
AMP
is
wall
and
turned
(2)
150
mt
reservoir
IN?
circuit?
reaching
the
Power
ON,
the
switch
by
the
Heat
distilled
through
the
unit
the
ADD WHEN
ON
OVER ALARM
WATER
UNIT
ALARM
IS
TEMPERATURE
TURNED
Fill
with
distilled
alarm
stops.
Manual
filling
Turn
and
service.
or
for
complete
instructions.
off
the
remove
It
must
replaced
by
Technologies,
Note:
Room
above
unusual
off
40°C,
before
situation,
returning
water
until
Refer
to
Service
draining
Power
the
Switch
Unit
be
repaired
Level
from
1
Inc.
temperature
may
cause
turn
the
it
to
service.
above
the
Power
and
40°C
H-250
(104°F)
to
shut
Switch
down
OFF
or
filling
and
and
allow
with
alarm.
the
distilled
In
unit
this
to
water
cool
Loading...
+ 18 hidden pages