ONE-YEAR MAINTENANCE PROTOCOL
for HOTLINE®Blood and Fluid Warmer and Level 1®Fast Flow Fluid Warmer
Benefits for you:
Maintain a microbe-free environment
We use a disinfecting procedure, unlike hotplate warmers which use superficial cleaning that does not disinfect or sterilize.
Maintenance time is reduced for the HOTLINE®Warmer from twice a year to once
®
year and for the Level 1
Fast Flow Fluid Warmer from 12 times a year to once
a
a year
Using hydrogen peroxide mixing bottles will make for easier mixing, pouring and storage of the
gen peroxide/distilled water solution. To find out more about the mixing bottles please
0.3% h
ydro
contact your local sales representative or Customer Service at 800-258-5361.
Yearly maintenance costs are reduced
The HO
TLINE only once a year, unlike hotplate warmers which may require cleaning maintenance costs up to 12 times a year.
If you would like additional information regarding the new one-year maintenance protocol for the HOTLINE Blood and Fluid Warmer and/or the Level 1®Fast Flow Fluid Warmer, please do not hesitate to contact us at the number belo
w .
Technical Support 1-800-553-8351, prompt 5 TECHSERV@smiths-medical.com
®
Warmer and the Level 1®Fast Flow Fluid Warmer require cleaning maintenance costs
®
*Data on f
ile
ONE-YEAR RESERVOIR MAINTENANCE PROTOCOL
The one-year maintenance solution is a 0.3% hydrogen peroxide/distilled water solution. The below protocol should be followed annually for the Level 1
®
Fast Flow Fluid Warmer (H-1200,
H-1025, H-1000, H-1100), as well as for the initial set up in your facility.
Level 1®Fast Flow Fluid Warmer
To prepare 0.3% hydrogen peroxide/distilled water solution for the reservoir, mix 140 mL of 3% hydrogen peroxide solution and 1,260 mL of distilled water.
1. Empty reservoir of current solution.
®
2. Place a container under the drain valve of the Level 1
3. Drain the re-circulating solution by turning the drain valve clockwise 90 degrees.
4. When all of the re-circulating solution has drained from the unit, close the drain valve.
5. Refill the unit with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water solution.
6. Insert a D-series disposable set (D-50/DI-50, D-60HL/DI-60HL, D-100/DI-100,
®
D-300/DI-300) into the Level 1
Fast Flow Fluid Warmer. If you use a Level 1®Fast Flow Fluid Warmer for urological and/or gynecological procedures, insert an IR-series disposable set (IR-500, IR-600/IRI-600/IRI-600B, IR-700, IR-40).
7. Turn the unit on and run for a 30-minute disinfection period.
Fast Flow Fluid Warmer.
8. Switch the unit off, discard the disposable according to established hospital procedures and empty the reservoir of the re-circulating solution by turning the drain valve clockwise 90 degrees. When all of the re-circulating solution has been drained from the unit, close the drain valve.
9. Refill the reservoir with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water
®
solution. The Level 1
Fast Flow Fluid Warmer is now ready for use.
During the 12-month period, any solution lost from the reservoir through the disposable must be topped off with the 0.3% hydrogen peroxide/distilled water solution.