Smiths Cadd-Legacy 1 PCA Plus 6500 User manual

T
E
C
H
N
L
A
U
N
A
M
L
C
I
y
S
P
M
U
P
N
C
A
A
M
B
D
U
L
D
A
T
O
-
R
L
Y
e
I
N
g
F
U
a
I
S
c
O
44
CADD-Legacy™ 1 Pump
CADD-Legacy
CADD-Legacy
PCA Pump
PLUS Pump
Deltec
For detailed instructions, specifications, warnings, warranties, and additional information on operat­ing CADD additional comments or questions concerning the operation of CADD
®
pumps, please refer to the Operator’s Manual supplied with the product. If you have
®
pumps, please call this num­ber: 800-426-2448. Our staff is available to help you twenty-four hours a day with the programming and operation of CADD® pump infusion systems.
The issue date of this Technical Manual is included on the back cover for the user’s information. In the event one year has elapsed between the issue date and product use, the user should contact SIMS Deltec, Inc. to see if a later revision of this manual is available.
Issue Date: January 2000
41
Table of Contents
1. Introduction ............................................. 1
Limited Warranty .......................................... 1
Exposing CADD® Pumps to Radiation, Ultra­sound, or MRI or use near ECG equipment ... 1
2. CADD-Legacy™ Pump
Delivery Modes .............................................. 2
Specifications (Nominal) ................................ 5
Compatible Medication Cassette™ Reservoirs and CADD
Remote Dose Cord ......................................... 7
®
Administration Sets ................... 7
3. Batteries ................................................... 8
Battery Compatibility ..................................... 8
DURACELL® Alkaline Battery Life............... 8
4. Construction .......................................... 11
5. Theory of Operation .............................. 12
Keyboard Circuitry ...................................... 12
Data Memory in Real Time Clock RAM ...... 12
EEPROM .................................................... 12
Battery Backed RAM ................................... 12
Time Base Circuitry ..................................... 12
LCD Circuitry .............................................. 12
LED Indicator ............................................. 13
Flash PROM Technology ............................. 13
Audible Alarm Circuitry ............................... 13
Watchdog Timer Circuit .............................. 13
Motor Drive/Motor Watchdog Circuit ......... 13
Power Circuitry ............................................ 14
Voltage Reference Circuit ............................. 14
Pumping Mechanism .................................... 14
Pumping Characteristics ............................... 15
Air Detector ................................................. 15
Downstream Occlusion Sensor .................... 16
Upstream Occlusion Sensor ......................... 16
Cassette Attachment Detection..................... 16
6. Safety Features and Fault Detection ....... 17
Hardware Safety Features ............................ 17
Software Safety Features .............................. 19
7. Hardware and Software Fault Detection 20
Overview ..................................................... 20
Order of Error Code Events ......................... 20
8. Cleaning and Inspection Procedures ....... 21
Inspection Recommendation ........................ 21
Cleaning ....................................................... 21
Battery Contact Cleaning ............................. 21
Visual Inspection .......................................... 22
Mechanical Inspection .................................. 22
9. Testing Procedures ................................. 23
Testing Recommendation ............................. 23
Changing to Lock Level 0 (LL0) .................. 23
CADD-Legacy™ PCA pump ................... 23
Air Detector Test ...................................... 25
Upstream Occlusion Sensor Test ............... 25
Occlusion Pressure Range Tests ................ 25
Accuracy Testing ....................................... 27
CADD-Legacy™ 1 pump ........................ 29
Air Detector Test ...................................... 30
Upstream Occlusion Sensor Test ............... 30
Occlusion Pressure Range Tests ................ 31
Accuracy Testing ....................................... 32
CADD-Legacy™ PLUS pump .................. 34
Air Detector Test ...................................... 35
Upstream Occlusion Sensor Test ............... 36
Occlusion Pressure Range Tests ................ 36
Accuracy Testing ....................................... 37
Cleaning and Functional Testing
Checklist ................................................ 40
42
1 Introduction
This Technical Manual is intended to provide an understanding of the mechanical and electri­cal operation of the CADD-Legacy
CADD-Legacy
1, and CADD-Legacy™ PLUS Computerized Ambulatory Drug Delivery pumps to persons familiar with these devices. The CADD-Legacy and CADD-Legacy
PCA, CADD-Legacy™ 1,
PLUS pump Operator’s
Manuals should be used in conjunction with this publication for complete information.
This manual also outlines cleaning and func­tional testing procedures that can be performed on the CADD-Legacy and CADD-Legacy
PCA, CADD-Legacy™ 1,
PLUS pumps.
WARNING:
This Technical Manual must be used by Bio­medical Technicians only. Do not permit patients to have access to this manual. Do not disclose to the patient the pump’s security codes or any other infomation that would allow the patient complete access to all programming and operating functions. Improper programming could result in death or serious injury to the patient.
IMPORTANT NOTICE:
CADD-Legacy CADD-Legacy
PCA, CADD-Legacy™ 1, and
PLUS pump operations and safety features are based on a microcomputer design. Inadequate servicing or tampering with the safety features of the pumps may seriously affect performance and safety.
For that reason, ALL SERVICING AND
REPAIR OF THE CADD-Legacy MUST BE PERFORMED BY DELTEC OR ITS AUTHORIZED AGENTS.
The manufacturer’s warranty agreement shall become null and void if the pump is not used in accordance with the Operator’s Manual and Instructions for Use for the pump accessories; or, the pump is serviced by persons other than Deltec or those authorized by Deltec.
Limited Warranty
The limited warranty associated with the CADD-Legacy CADD-Legacy the product literature supplied with the product when originally purchased, which is incorpo­rated herein by reference. DELTEC SPECIFI-
PCA, CADD-Legacy™ 1, and
PLUS pumps can be found in
PUMPS
PCA,
CALLY DISCLAIMS ANY OTHER WAR­RANTY, WHETHER EXPRESS, IMPLIED OR STATUTORY, INCLUDING, WITHOUT LIMITATION, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR USE. Deltec further disclaims responsibility for the suitability of the system for a particular medical treatment or for any medical complica­tions resulting from the use of the system. The manufacturer shall not be responsible for any incidental damages or consequential damages to property, loss of profits, or loss of use caused by any defect or malfunction of the system.
If you wish to receive additional information about the extent of the warranty on these prod­ucts, please contact your Deltec representative or call Customer Service at 1-800-426-2448.
All recommendations, information, and litera­ture supplied by Deltec with respect to the CADD
®
product line are believed to be accurate and reliable, but do not constitute warranties. No agent, representative, or employee of Deltec has authority to bind Deltec to any representa­tion or warranty, expressed or implied.
Exposure to Radiation, Ultrasound or Magnetic Resonance Imaging (MRI), or use near ECG equipment
CAUTION:
Do not expose the pump to therapeutic levels of
ionizing radiation as permanent damage to the pump’s electronic circuitry may occur. The best procedure to follow is to remove the pump from the patient during therapeutic radiation sessions. If the pump must remain in the vicinity during a therapy session, it should be shielded, and its ability to function properly should be confirmed following treatment.
• Do not expose the pump directly to ultrasound,
as permanent damage to the pump’s electronic circuitry may occur.
• Do not use the pump in the vicinity of magnetic
resonance imaging (MRI) equipment as mag­netic fields may adversely affect the operation of the pump. Remove the pump from the pa­tient during MRI procedures and keep it at a safe distance from magnetic energy.
• Do not use the pump near ECG equipment as
the pump may interfere with the operation of the equipment. Monitor ECG equipment care­fully when using this pump.
1
2 CADD-Legacy™ Pump
Delivery Modes
The CADD-Legacy™ ambulatory drug delivery pump provides measured drug therapy to patients in hospital or outpatient settings. The CADD-Legacy nous, intra-arterial, subcutaneous, intraperi­toneal, epidural space, or subarachnoid space infusion.
Epidural administration is limited to use with indwelling catheters for short term delivery of anesthetics and short or long term delivery of analgesics. Subarachnoid administration is limited to use with indwelling catheters for short-term delivery of analgesics.
(used here as a loading dose)
Dosage
pump is indicated for intrave-
Clinician Bolus
Demand Doses
Continuous Rate
The CADD-Legacy
PCA pump may be programmed to deliver medication in one of three ways: 1) continuous rate only, 2) patient­activated dose only and 3) continuous rate and patient-activated dose. (See figure 1.)
The CADD-Legacy
PLUS pump may be programmed to deliver in one of two modes: (1) Continuous, (2) Intermittent . (See figures 2 and 3.)
The CADD-Legacy
1 pump operates in continuous mode. (See figure 2.)
Figure 4 shows a diagram of the CADD­Legacy
pump.
PCA Delivery Profile
The PCA (patient-controlled analgesia) delivery mode is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patient­controlled analgesia.
Time
Figure 1. PCA mode delivery profile.
Delivery
Rate
CADD-Legacy™ PLUS
(ML/HR)
or
CADD-Legacy™ 1
(ML/24 HR)
Time
Continuous Delivery
Figure 2. Continuous mode delivery profile.
Dose Cycle
Dose
Volume
Dose
Starts
in
Dose
Duration
Time
Intermittent Delivery
Dose
Duration
Continuous Mode Delivery Profile
The Continuous delivery mode allows the infusion of drug at a constant, programmed rate.
Intermittent Mode Delivery Profile
The Intermittent delivery mode allows the infusion of a specific volume of drug at regular programmed intervals.
Figure 3. Intermittent mode delivery profile.
2
Power Jack
Display
Threaded Mounting Hole
Power Jack symbol
Accessory
Jack
AC Indicator
Light
Air Detector
®
Cassette Lock
Keypad
Dose Key on CADD-Legacy™ PCA
Front View
Figure 4. Front and Rear views of the CADD-Legacy™ Pump. Features are identical on all CADD-Legacy™ pumps except as illustrated for the CADD-Legacy™ PCA pump.
Cassette
CADD-Legacy™ PCA
Cassette Lock
Rear View
Accessory Jack Symbol
Battery Compartment
3
PCA Delivery Mode Scroll Ranges
Units
ML
MG
MCG
Starting
0.10
10% of Concentration
10% of Concentration
Increment
All values:
Values between 0.01 and 0.5: Values between 0.50 and 100.0: Values between 100.0 and 1000.0: Values greater than 1000.0:
Values between 0.1 and 100: Values between 100 and 1000: Values greater than 1000:
Table 1. PCA delivery mode: continuous rate scroll ranges.
Milligrams
Concentration Demand Dose Clinician Bolus
mg/ml increment max. increment max.
0.1 0.01 0.99 0.01 2
0.2 0.02 1.98 0.02 4
0.3 0.03 2.97 0.03 6
0.4 0.04 3.96 0.04 8
0.5 0.05 4.95 0.05 10 1 0.05 9.9 0.05 20 2 0.10 19.8 0.10 40 3 0.15 29.7 0.15 60 4 0.20 39.6 0.20 80
5 0.25 49.5 0.25 100 10 0.50 99.0 0.50 200 15 0.75 148.5 0.75 300 20 1.00 198.0 1.00 400 25 1.25 247.5 1.25 500 30 1.50 297.0 1.50 600 35 1.75 346.5 1.75 700 40 2.00 396.0 2.00 800 45 2.25 445.5 2.25 900 50 2.50 495.0 2.50 1000 55 2.75 544.5 2.75 1100 60 3.00 594.0 3.00 1200 65 3.25 643.5 3.25 1300 70 3.50 693.0 3.50 1400 75 3.75 742.5 3.75 1500 80 4.00 792.0 4.00 1600 85 4.25 841.5 4.25 1700 90 4.50 891.0 4.50 1800 95 4.75 940.5 4.75 1900
100 5.00 990.0 5.00 2000
Table 2. Demand dose, clinician bolus scroll ranges, milligrams
4
Maximum
0.10
0.01
0.10
1.00
10.0
0.10
1.00
10.00
50.00
Concentration x 50
Concentration x 50
Micrograms
Concentration Demand Dose Clinician Bolus
mcg/ml increment max. increment max.
1 0.05 9.9 0.05 20 2 0.10 19.8 0.10 40 3 0.15 29.7 0.15 60 4 0.20 39.6 0.20 80
5 0.25 49.5 0.25 100 10 0.50 99.0 0.50 200 15 0.75 148.5 0.75 300 20 1.00 198.0 1.00 400 25 1.25 247.5 1.25 500 30 1.50 297.0 1.50 600 35 1.75 346.5 1.75 700 40 2.00 396.0 2.00 800 45 2.25 445.5 2.25 900 50 2.50 495.0 2.50 1000 55 2.75 544.5 2.75 1100 60 3.00 594.0 3.00 1200 65 3.25 643.5 3.25 1300 70 3.50 693.0 3.50 1400 75 3.75 742.5 3.75 1500 80 4.00 792.0 4.00 1600 85 4.25 841.5 4.25 1700 90 4.50 891.0 4.50 1800 95 4.75 940.5 4.75 1900
100 5.00 990.0 5.00 2000 200 10.00 1980.0 10.00 4000 300 15.00 2970.0 15.00 6000 400 20.00 3960.0 20.00 8000 500 25.00 4950.0 25.00 10000
Table 3. Demand dose, clinician bolus scroll ranges, micrograms
Milliliters
Demand Dose Clinician Bolus
increment max. increment max.
0.05 9.9 0.05 20
Table 4. Demand dose, clinician bolus scroll ranges, milliliters
Specifications (Nominal)
High Pressure Alarm
26 (±14) psi, 1.79 (± 0.97) bar
General Pump Specifications
Resolution
Medication Cassette Administration Set, 0.050 ml/pump stroke nominal
Size
4.1 cm x 9.5 cm x 11.2 cm (1.6 in. x
3.8 in. x 4.4 in.) excluding cassette or other accessories
Weight
391 g (13.8 oz.) including 2 AA batteries, empty 100-ml Medication Cassette Reservoir, and air detector, excluding other accessories
Pump Alarms
Low battery power; depleted battery power; battery dislodged; pump stopped; pump fault; low reservoir volume; high delivery pressure; air in line; disposable not attached when run attempted; motor locked; upstream occlusion; reservoir volume empty; program incomplete; remote dose cord removed; key stuck; disposable detached, power removed, value not saved.
Bolus Volume at Occlusion Alarm Pressure
0.050 ml resolution sets/reservoirs: < 0.15 ml
Power Sources
Two AA alkaline batteries such as DURA-
®
CELL
or EVEREADY Energizer®; AC
adapter. An internal battery powers the clock.
When it is depleted, it cannot reliably maintain the clock time. This battery must be replaced by SIMS Deltec, Inc. The internal battery has an expected life of 5 years.
System Operating Temperature*
+2°C to 40°C (35°F to 104°F)
System Storage Temperature*
-20°C to 60°C (-4°F to 140°F)
System Delivery Accuracy*
± 6% (nominal)
Reservoir or CADD
Air Detector Alarm
Single bubble
®
Low sensitivity = greater than 0.250 ml High sensitivity = greater than 0.100 ml
Multi-bubble = 1.0 ml nominal
Delivery Mode Specifications
CADD-Legacy™ PCA pump
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Units
Milliliters (ml), milligrams (mg), micro­grams (mcg) Default: milligrams
Concentration
Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 10, 15, ... 95, 100 (Default: 100 mg/ml) Mcg/ml: 1, 2, 3, 4, 5, 10, 15, ...95, 100, 200, 300, 400, 500 (Default: 500 mcg/ml)
Continuous Rate
0 to 50 ml/hr (or the mg or mcg equivalent) (See Table 1 for scroll ranges)
Demand Dose
0 to 9.9 ml in 0.05 ml increments (or the mg or mcg equivalent) (See Tables 2 and 3 for scroll ranges) Delivery rate (Continuous Rate + Demand Dose): 125 ml/hr nominal
Dose Lockout
5 minutes to 24 hours in the following increments: 1 minute for values between 5 and 20 minutes 5 minutes between 20 minutes and 24 hours Default: 24 hours
Doses Per Hour
1 to 12 doses in 1 dose increments (will also be limited by the Demand Dose Lockout value)
Default: 1 dose/hr
*System is defined as a CADD-Legacy™ pump with an attached Medication Cassette™ Reservoir and CADD® Extension Set with integral anti-siphon valve, or an attached CADD® Administration Set with integral or add-on anti-siphon valve.
5
Doses Given
0 to 999
5 minutes from 00:10 to 24:00 Default: 30 minutes
Doses Attempted
0 to 999
Given
0 to 99999.95 in 0.05 unit increments or 0 to 99999.99 in 0.01 unit increments
(increments converted to current units based on concentration)
Clinician Bolus
0.05 ml to 20.00 ml (or mg or mcg equiva­lent) (See Tables 1, 2 and 3 for scroll ranges) Delivery rate (Continuous Rate + Clinician Bolus): 125 ml/hr nominal
CADD-Legacy
1 pump
Continuous Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Continuous Rate
1 to 3000 ml/24 hr in increments of 1 ml/24hr Default: 0 ml/24hr
Given
0 to 99999.95 in 0.05 ml increments
Dose Cycle
10 minutes to 96 hours in 5 minute incre­ments
Default: 4 hours
KVO Rate
0 to 125.0 ml/hr in increments of 0.1 ml/hr Default: 0 ml/hr
Dose Starts in
Immediate or 1 minute to 96 hours in the following increments:
00:01 from 00:00 to 00:10 00:05 from 00:10 to 96:00 Default: Immediate
Continuous Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Continuous Rate
0.1 ml/hr to 125.0 ml/hr in increments of
0.1 ml/hr Default: 0.0 ml/hr
Given
0 to 99999.95 in 0.05 ml increments
Biomed Functions Specifications
CADD-Legacy
PLUS pump
Intermittent Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Dose Volume
0.1 to 1000.0 ml in increments of 0.1 Default: 0.0 ml
Dose Duration
1 minute to 24 hours in the following increments:
1 minute from 00:01 to 00:10
6
Air Detector Status:
Off On- low
On- high Default: On-high
Upstream Occlusion Status:
Off On Default: On
Delivery Mode (CADD-Legacy
Continuous Intermittent Default: Intermittent
PLUS only):
Compatible Medication Cassette
Reservoirs and CADD® Administration Sets
• 50-ml or 100-ml Medication Cassette Reservoir, used with the CADD® Extension Set with Anti-siphon Valve.
• CADD
®
Administration Set with integral anti-siphon valve, with or without bag spike (allows use of flexible plastic bag or sterile vial with injector)
• CADD
®
Administration Set with add-on anti-siphon valve and bag spike (allows for gravity priming before attaching the add on anti-siphon valve)
Remote Dose Cord
Deltec provides a Remote Dose Cord for the CADD-Legacy sion of the Single Pole Double Throw (SPDT) switch which operates in the same manner as the key. When the Remote Dose Cord is attached to the pump, the patient may press either the Remote Dose button or the a Demand Dose. The clinician may also use either the Remote Dose button or the to deliver a clinician-activated bolus. For easy access, the Remote Dose Cord may be fastened to the patient’s clothing or bedsheet with the attached clip.
There is an alarm/function present in the CADD-Legacy Dose Cord is removed, the display shows a message “Remote Dose Removed”. The pump sounds an audible alarm until the pressed to acknowledge the alarm.
PCA pump which is an exten-
Í key. The push button is a
Í key to receive
Í key
PCA pump. If the Remote
key is
Í
NOTE:
To detach the Remote Dose Cord from the pump, grasp the Remote Dose Cord connector and pull back using a straight, steady motion.
7
3 Batteries
Battery Compatibility
Recommended Batteries
Two AA alkaline batteries are recommended for use in the CADD-Legacy
pumps. Carbon­zinc, mercury, nickel-cadmium, nickel-metal­hydride, or zinc-air AA batteries should not be used.
Battery Life
The CADD-Legacy
pumps have been de­signed to provide optimal battery life. The expected battery life in the CADD-Legacy
pumps depends on the following factors:
• Programmed delivery rate
• Operating temperatures
• Battery type and brand
• Battery age
DURACELL® Alkaline Battery Life
Recommended storage conditions are 10°C to 25°C (50°F to 77°F) with no more than 65% relative humidity noncondensing.
The following table may be used to predict typical alkaline battery life at different delivery rates when alkaline batteries are used in the
CADD-Legacy
pump. As expected, battery life decreases as the delivery rate increases. This table is based on laboratory tests using fresh DURACELL CADD-Legacy
®
alkaline batteries in
pumps while the pumps were operating at room temperature.
Actual battery life may be significantly shorter depending on the operating temperature and the storage conditions of the battery.
Battery life is shortened significantly at very low operating temperatures. For example, at 0°C (32°F), an alkaline battery will yield approximately 30% of its normal capacity.
Alkaline batteries do not need to be stored in a refrigerator. After four years of storage at 21°C (70°F), an alkaline battery retains approxi­mately 86% of its original capacity. Battery life will be shorter if the battery is stored above room temperature. An alkaline battery stored at 43°C (110°F) will be down to approximately 80% of its capacity within one year.
Continuous Delivery Battery Life with Alkaline Batteries
Rate Life Volume
0.4 ml/hr 338 hrs 135 ml 4 ml/hr 178 hrs 712 ml
10 ml/hr 112 hrs 1120 ml 15 ml/hr 96 hrs 1440 ml 30 ml/hr 53 hrs 1590 ml 75 ml/hr 18 hrs 1350 ml
125 ml/hr 15 hrs 1875 ml
Table 5. Two AA Alkaline-type batteries used with the CADD-Legacy™ pumps.
8
Intermittent Delivery Battery Life with Alkaline Batteries
Dose Volume Duration Dose Cycle KVO Life Volume
23.5 ml 1:00 hr 5:00 hr 0.2 ml/hr 193 hr 915 ml 61 ml 1:00 hr 6:00 hr 0.2 ml/hr 120 hr 1224 ml
125 ml 1:00 hr 6:00 hr 0.2 ml/hr 65 hr 1356 ml
Table 6. Two AA Alkaline-type batteries used with the CADD-Legacy™ pumps.
135
120
105
90 75
60
45
Rate (ml/hr)
30
15
0
10
20 30
40 50
Hours
Figure 5. Operating time to low battery alarm using alkaline batteries.
18
16
14
12
10
8
6
Rate (ml/hr)
4 2
0
50
100
150
200
Hours
60
250
70
80
300
90
100
350
Figure 6. Operating time to low battery alarm using alkaline batteries.
9
Continuous Delivery Battery Life with Lithium Batteries
Rate Life Volume
0.4 ml/hr 413 hrs 165 ml 4 ml/hr 307 hrs 1228 ml
10 ml/hr 190 hrs 1900 ml 15 ml/hr 163 hrs 2445 ml 30 ml/hr 90 hrs 2700 ml 75 ml/hr 33 hrs 2475 ml
125 ml/hr 22 hrs 2750 ml
Table 7. Two AA Lithium-type batteries used with the CADD-Legacy™ pumps.
Intermittent Delivery Battery Life with Lithium Batteries
Dose Volume Duration Dose Cycle KVO Life Volume
23.5 ml 1:00 hr 5:00 hr 0.2 ml/hr 300 hrs 1458 ml 61 ml 1:00 hr 6:00 hr 0.2 ml/hr 185 hrs 1911 ml
125 ml 1:00 hr 6:00 hr 0.2 ml/hr 125 hrs 2625 ml
Table 8. Two AA Lithium-type batteries used with the CADD-Legacy™ pumps.
135
120
105
90
75
60
45
Rate (ml/hr)
30
15
0
Figure 7. Dual-stroke operating time on lithium batteries.
18
16
14
12
10
8
6
Rate (ml/hr)
4 2
0
50
Figure 8. Single-stroke operating time on lithium batteries.
20
100
40 60
150
80 100
200
Hours
250
Hours
120
300
140
350
160
400
180
200
450
10
4 Construction
The pump’s housing is made of a special high impact plastic. It is composed of two sections: the rear and front housing. The pump housing is sealed to ensure that the pump is water resistant. The battery compartment is not water resistant.
NOTE:
The CADD-Legacy
ambulatory infusion pump is water resistant, but not waterproof. The pump is “Splashproof” and is characterized by the IEC code of IPX4.
The battery compartment is accessed through a removable door on the rear housing. Within the battery compartment is space for the batteries and the four battery contacts.
On CADD-Legacy Cassette
Reservoir or the administration set is
pumps the Medication
attached to the bottom of the pump by inserting the two hooks on the cassette into the mating hinge pins on the pump. The pump and the reservoir or the administration set are then placed in an upright position on a firm, flat surface. The reservoir or the administration set must be secured in place by inserting a coin (or key if using the CADD-Legacy
PCA pump) in the slot on the pump’s locking button, pushing the button in and turning the button one­quarter turn counter-clockwise.
NOTE:
The Medication Cassette
Reservoir and the administration set are intended for single use only.
The keyboard, located on the front housing, is composed of eight membrane switches (nine membrane switches on the CADD-Legacy
PCA pump) and is sealed against moisture. All of the keys contain domes to provide a tactile feel when the key is pressed. The keyboard keys are sensed by the pump’s microprocessor.
The Liquid Crystal Display (LCD), also located on the front housing, shows the pump status and programmed settings. The dot matrix display consists of 16 character columns with 2 rows of characters, and is selected by the pump’s microprocessor according to status conditions and keyboard entries.
The microprocessor and other circuitry which control the pump are located on a printed circuit board. The board contains the Central Processing Unit (CPU), motor driver circuitry, and other circuitry. The circuitry is designed to reduce susceptibility to interference from electromagnetic fields and to dissipate electro­static discharge.
The LCD controller is mounted on the LCD using chip on glass technology.
The pumping mechanism subassembly contains the motor, gear train, camshaft, valves, expul­sor, sensing disk, infrared light source, infrared detector, occlusion sensors, disposable sensor, and cassette locking button. Via the motor driver circuitry, the pump’s microprocessor controls motor rotation.
Two external port connectors are utilized for remote dose and external power input. The accessory jack is used for attachment of the Remote Dose Cord (CADD-Legacy
PCA pump only) and interface cable. The Remote Dose Cord enables the patient to use either of two options to begin a Demand Dose when using the PCA delivery mode: (1) the Remote Dose button; or (2) the
Í key.
The second port allows connection to an AC adapter.
The keyboard is connected to the printed circuit board via a flex circuit tail. Discrete wires connect the pumping mechanism, motor, and sensors to the printed circuit board.
The accessory jack in conjunction with the interface cable allows download of events using the CADD-DIPLOMAT
software.
11
5 Theory of Operation
Keyboard Circuitry
The CADD-Legacy™ pumps are controlled by a microprocessor. The actions of the microproces­sor are controlled by a program, which is contained in the memory.
Commands are issued to the microprocessor from the user via the eight keys on the keyboard (nine keys and Remote Dose Cord on CADD-Legacy keyboard are arranged in a 3x3 matrix which feeds into the keyboard encoder. A key closure applies a ground to the associated input of the keyboard encoder. Key debounce circuitry resident in the keyboard encoder provides a clean output signal to the microprocessor for the duration of the key closure. The micropro­cessor reads keyboard status by accessing special memory locations in the keyboard encoder.
The Remote Dose Cord button (CADD-Legacy PCA pump only) consists of an SPDT switch with one switch output going to the micropro­cessor and the other going to the keyboard encoder. The switch has a common input line and two output signal lines. The two signal lines are complementary such that one line is always logic high and the other is always low. When the Remote Dose Cord button is pressed, both signal lines change to the alternate logic state. This redundancy prevents a single line failure from starting a dose delivery.
The
Å button allows the pump to be placed in a
very low power mode by turning off all sensors and LCD, but some battery energy is still used by the electronics. To maximize battery life, remove the batteries when pump is not in use.
PCA pump). The keys on the
Data Memory in Real Time Clock RAM
Many settings of the pump’s delivery and record-keeping parameters are stored by the microprocessor in a Battery backed RAM in the Real Time Clock. Data to and from the memory is presented serially. Whenever the microproces­sor uses data from the Real Time Clock, the data is checked for validity.
EEPROM
Data describing the current delivery protocol is stored in an EEPROM included in the micro­processor. Whenever this data is used, it is checked for validity.
Battery Backed RAM
Additional settings of the pump’s delivery and record keeping parameters are stored in a battery backed Random Access Memory (RAM). Battery backup is provided by a printed circuit board-mounted lithium battery. This battery provides a minimum of five years of memory retention during normal pump usage. Whenever the microprocessor uses data from the RAM, the data is checked for validity.
Time Base Circuitry
An accurate 3.6864 MHz timebase is provided by a quartz crystal. The 3.6864 MHz signal is connected to the microprocessor, where it is frequency-divided to access the program memory at a cycle rate of 921 kHz.
In addition, an accurate 32.768 kHz timebase is provided by a second quartz crystal. The
32.768 kHz signal is used for the real time clock.
LCD Circuitry
The CADD-Legacy™ pumps feature a 2 line by 16 character Liquid Crystal Display (LCD). The characters on this dot matrix display are formed by a matrix of 5 by 7 dots. It is reflec­tive only, with a black on silver appearance, with no backlight.
The display includes a controller chip mounted directly on the glass capable of interfacing with 4 and 8 bit systems to display 92 kinds of characters, numerals, symbols, and special characters under control of a built in character generator ROM. A RAM is also included to make other special characters possible.
12
LED Indicator
A green Light Emitting Diode (LED) is pro­vided under the pump’s front panel overlay to provide pump power status to the user. When this LED is lit, it indicates that an AC adapter is being used to power the pump.
Flash PROM Technology
Program memory for the pump is stored in Flash Programmable Read Only Memory (Flash PROM). This type of memory allows modification of the contents without physically removing the device from the circuit board. Under certain circumstances the program can also be downloaded. Several layers of redun­dancy in the programming system prevent accidental erasing or modification of the PROM.
Audible Alarm Circuitry
Audible alarm circuitry consists of two piezo electric disks and an independent oscillator. The disks flex or bend in resonance with the output of the oscillator. The piezo disks are mounted to the pump housing to enhance sound level.
The microprocessor controls the audible alarm by selecting the alarm control line for more than 0.5 seconds. The oscillator which drives the piezo disks is capable of providing two driving frequencies. The low frequency is in the range of 700 to 1500 Hz and the high fre­quency is in the range of 1600 to 2500 Hz. When the microprocessor selects the audible alarm, the alarm enters a warble mode where it oscillates between the low and high frequency sound at a rate of 0.8 and 2 Hz.
Low battery voltage detection and watchdog timer circuitry also have the ability to enable the audible alarm via the microprocessor.
The audible alarm circuitry is backed up by a super capacitor. The super capacitor provides energy for the alarm in the instance where all power is lost while pump is in the RUN mode. There is enough energy in the super capacitor to drive the audible alarm for 3 minutes when the pump has been in the RUN mode for 2 minutes or longer.
Watchdog Timer Circuit
Watchdog timer circuitry is provided to moni­tor the status of the microprocessor. If the microprocessor fails to function properly, the watchdog circuit issues a reset signal which disables the motor and enables the audible alarm. To assure proper function, the micro­processor must strobe the watchdog circuit at least once every second in order to prevent the watchdog from performing its reset function.
The reset output from the watchdog circuit is a pulse output. This acts to “jump start” the microprocessor. This unique feature allows the microprocessor to test the watchdog circuit on every power-up. By setting a flag in memory and not strobing the watchdog, the micropro­cessor can force a watchdog time-out. After being reset, the microprocessor checks the status flag to see if this was a time-out test. If so, the microprocessor verifies the watchdog’s ability to disable the motor and then continues normal power-up activities.
If a reset occurs when the microprocessor is not expecting it, the microprocessor traps the event, sounds the audible alarm and displays an error message on the LCD.
Motor Drive/Motor Watchdog Circuit
The motor drive circuitry is composed of a series of power FET transistors, passive com­ponents, and two voltage comparators. Built into the motor drive circuitry is an RC timer which times how long the motor runs each time it is turned on. If the motor runs for more than an average of 4 seconds, the circuit will time out and disable the motor.
A unique feature of this circuit is that control lines to and from the microprocessor circuit allow the microprocessor to perform a com­plete functional test of the motor drive circuit without running the motor. The microproces­sor performs this test function every several minutes to assure its continued functionality. An input from the watchdog circuit prevents motor operation if the watchdog timer expires.
Rotation of the motor is sensed by the micro­processor via an infrared-sensitive photo detector. An infrared light source is mounted
13
so that its light beam illuminates the infrared detector. An opaque flag is mounted concentri­cally to the camshaft and rotates with it be­tween the infrared light source and detector.
When the flag interrupts the light beam, the output of the detector is sensed by the micro­processor via an input port bit. Power to the infrared LED light source is controlled by the motor drive circuit and is off when the motor is not running to conserve battery life.
In the microprocessor software, multiple checks are made on motion of the camshaft. When the motor is commanded to start, the infrared sensor must show that half a revolu­tion has occurred within four seconds and that the motor has stopped when half a rotation was completed. In addition, no camshaft rotation can take place when the motor has not been commanded to run.
Power Circuitry
Power for the pump is normally supplied by two AA alkaline batteries, two AA lithium batteries, or an AC adapter. These types of batteries have a fairly low internal resistance over their discharge range, which will keep power supply noise low. Other types of batter­ies, such as carbon-zinc, exhibit high internal resistance, especially near depletion. A voltage drop across the internal resistance occurs when current is drawn by the motor during pump activations. This current is demanded in short pulses when the motor is first turned on and
generates large spikes in the battery voltage. This noise can cause the low battery detection circuit to shut down the pump.
The power from the two AA batteries is boosted to +5VDC. This 5V is used to power the motor and a 3.3V linear regulator. The linear regulator provides power to all the other circuitry including the microprocessor.
Voltage Reference Circuit
A voltage reference circuit provides a constant DC voltage to the microprocessor Analog to Digital Converter (ADC). By reading this input and comparing the value to a predetermined range, the microprocessor can validate the accuracy of the 3.3-volt power supply. Varia­tions in the 3.3-volt supply left undetected can result in inaccuracy in the low battery alarm set points and variations in other calculated values. (Also refer to Voltage Detector Circuit description on page 18.)
Pumping Mechanism
The pumping mechanism is linear peristaltic with two active valves and an expulsor. Pump­ing occurs when the expulsor presses on the reservoir pump tubing in sequence with the inlet and outlet valves. At rest, the outlet valve is pressing down fully on the tubing and the expulsor and inlet valve are retracted. (See Figure 9.)
Motor
Pump Housing
Expulsor
Occlusion Sensor
Cassette Hinge
Pump Tubing
Figure 9. A simulated pumping mechanism in a CADD-Legacy™ pump.
14
Camshaft
Lock Cassette
Upstream Sensor
Pressure Plate Inlet Valve
Outlet Valve
When the microprocessor commands the mecha­nism to pump, the camshaft begins to rotate, thus controlling the following pump cycle:
1. The inlet valve closes.
2. In synchrony with the expulsor moving down to compress the tubing, the outlet valve opens, expelling 0.050 ml of fluid toward the patient.
3. The outlet valve closes.
4. The inlet valve opens as the expulsor is retracted, causing fluid from the reservoir to again fill the pump tubing segment.
5. The camshaft rotation stops after half a revolution and the cycle is completed.
The microprocessor uses its timer circuits to accurately time the 18 seconds (in this ex­ample) between mechanism activations. The timebase accuracy is ultimately determined by the 3.6864 MHz quartz crystal oscillator.
Air Detector
The air detector is designed to detect air in the outlet tubing fluid path. The air detector can be set to On-high sensitivity, On-low sensitivity, or Off by accessing Biomed Functions. When the On-high sensitivity setting is selected the pump will detect a single bubble greater than 0.100 ml. When the On-low sensitivity setting is selected the pump will detect a single bubble greater than
0.250 ml.
Pumping Characteristics
To deliver the amount of drug specified by the parameter settings, the pump’s microprocessor causes the pump mechanism to deliver fluid “pulses” timed according to the desired rate. At rates of 15 ml/hr or less the microprocessor delivers a single pulse to the motor circuit causing a half revolution of the camshaft and fluid delivery in 0.05 ml increments. At rates greater than 15 ml/hr the microprocessor delivers two back to back pulses to the motor circuit causing a full revolution of the camshaft and fluid delivery in 0.1 ml increments. Thus, to deliver 20 ml/hr, for example, the micropro­cessor solves these equations:
Mechanism activations per hr
= 20 ml per hr/0.1 ml per activation
= 20/0.1
= 200
Time (seconds) between activations
= 3600 sec per hr/number of activations per hr
= 3600/200
=18
NOTE:
At rates 15 ml/hr the pump delivers 0.05 ml per stroke. This allows a more contin­uous delivery at low rates.
Multi-bubble Sensing
The air detector is also designed to sense if an accumulation of more than 1 ml of air has passed through the outlet tubing path in the last 15 minutes. This feature is active anytime the air detector is on.
The air detector is compatible with all of the Medication Cassette
Reservoirs and CADD
®
Administration Sets indicated for use with the CADD-Legacy
pump, and all pump accessories. It is powered directly from the pump and no additional power is required.
Theory of Operation
The air detector consists of sensor electronics and two ultrasonic transducers positioned on oppo­site sides of the fluid path. One transducer acts as an acoustic transmitter and the other as an acoustic receiver. Air detection occurs when air in the fluid path causes a reduction in the signal level to the receiver. When the signal is inter­rupted for a preset length of time, the sensing circuitry sends a signal to the microprocessor indicating air in the fluid path. To maximize the reliability of the system and to reduce false alarms, the transmitted signal is swept over a frequency range. This accommodates varying resonance frequencies of the transducer and reduces sensitivity to tubing tolerances and other mechanical variations.
15
Downstream Occlusion Sensor
The downstream occlusion sensor is designed to detect excessive pressure in the outlet tubing.
If the fluid path to the patient becomes blocked, the pump tubing will expand as pumping occurs. When there has been an amount of inflation corresponding to 179.3 kPa ±
96.5 kPa, 1.79 (± 0.97) bar, or 26 (±14) psi, the occlusion sensor trips, whereupon the microprocessor stops the pump mechanism and issues visual and audible alarms. Thus the maximum pressure which can be developed is 276 kPa (2.76 bar, 40 psi).
Construction
The downstream occlusion sensor consists of a membrane switch located on the bottom of the pump next to the outlet valve. The switch is fastened to the housing with an adhesive to ensure that the overall assembly is water resistant.
Theory of Operation
The membrane switch is in contact with the outlet tubing when a cassette is installed. Tubing expansion caused by a downstream occlusion results in closure of the membrane switch. Switch closure sends a logic low to the microprocessor indicating a downstream occlusion.
Upstream Occlusion Sensor
The upstream occlusion sensor detects an occlusion in the inlet tubing which would prevent or restrict the flow of fluid to the pump.
Construction
The upstream occlusion sensor consists of a strain gauge sensor located on the bottom of the pump next to the inlet valve. The sensor is fastened to the housing with an adhesive to ensure that the overall assembly is water resistant.
Theory of Operation
When a cassette is installed on the pump, the inlet tubing is in contact with the sensor. In order to conserve battery power, the upstream occlusion sensor circuit is only activated while the motor circuitry is enabled. Pressure on the sensor is read just prior to the motor starting and after the end of the motor stroke. The microprocessor uses an average of the pressure exerted by the unoccluded tubing to establish a baseline pressure. If the tubing pressure at the end of a motor stroke is below the baseline pressure, the upstream tubing is occluded.
Cassette Attachment Detection
The pump uses the upstream occlusion sensor and cassette present sensor to verify the pres­ence of a cassette. If an infusion is started by pressing or if a cassette is improperly seated, the pump will initiate a visual and audible alarm.
Theory of Operation
During manufacture of the pump, upstream occlusion sensor readings are recorded for no cassette installed and typical cassette installed. These readings are used to calculate threshold levels for cassette detection. When a cassette is first attached to the pump, the new sensor reading must be above the calculated threshold level.
when there is no cassette installed
16
Additional readings are taken periodically while the pump is in use. If the sensor readings drop below the threshold when the motor is off, or the cassette present sensor circuit does not sense the presence of a cassette, the cassette is considered removed.
6 Safety Features and
Fault Detection
Hardware Safety Features
Key hardware safety features include a watch­dog timer circuit, motor drive and motor watchdog circuits, cassette present sensor circuit, and a voltage detector circuit. Each safety circuit performs a unique function to insure the overall safety of the device. (See Figure 10.)
Watchdog Timer Circuit
The microprocessor must send an appropriate signal to the watchdog circuit at least once per second. If the microprocessor does not, the watchdog circuit will time out and shut down the pump controller.
Watchdog timer circuitry is provided to monitor the status of the microprocessor and disable the motor and enable the audible alarm if the microprocessor fails to function properly. The microprocessor must strobe the watchdog circuit at least once every second in order to prevent the watchdog from performing its reset function. The reset output from watchdog circuit is a pulse output. This acts to “jump start” the microprocessor. This unique feature allows the microprocessor to test the watchdog circuit on every power-up. By setting a flag in memory and not strobing the watchdog, the
microprocessor can force a watchdog time-out. After being reset, the microprocessor checks the status flag to see if this was a time-out test. If so, the microprocessor verifies the watchdog’s ability to disable the motor and then continues normal power-up activities. If the reset occurred when the microprocessor was not expecting it, the microprocessor traps the event, sounds the audible alarm and dis­plays an error message on the LCD.
Motor Drive/Motor Watchdog Circuit
Motor drive circuitry is composed of a series of power FET transistors, passive components, and two voltage comparators. Built into the motor drive circuitry is an RC timer which times how long the motor runs each time it is turned on. If the motor runs for more than an average of 4 seconds, the circuit will time out and disable the motor. A unique feature of this circuit is that control lines from the micropro­cessor can perform a complete functional test of the motor drive circuit without running the motor. The microprocessor performs this test function every several minutes to assure its continued functionality. An input from the watchdog circuit prevents motor operation if the watchdog timer expires.
PROGRAM
MEMORY
DATA
LCD DISPLAY VOLTAGE
CPU/IO
▼ ▼ ▼
▼ ▼
MEMORY
KEYBOARD
Figure 10. CADD-Legacy™ pump hardware block diagram.
AUDIBLE
ALARM
MONITOR
MOTOR
DRIVE
WATCHDOG
REAL-TIME
CLOCK
MOTOR
WATCHDOG
POWER
INPUT
SENSORS
17
Voltage Trip Point
2.4 V
Voltage
Trip Point
Source
Battery
Motor Status CADD-Legacy
Running/not running
No alarm
Pump Status
< 2.4 V
< 1.8 V
< 4.75 V
< 1.0 V
Battery
Battery
5 Volt supply
motor voltage
Battery
Not running
Running
Running
Running/not running
Table 9. CADD-Legacy™ pump low battery conditions.
Cassette Present Sensor Circuit
The cassette present sensor system consists of a switch on the pump mechanism that interfaces to the attached cassette and associated cir­cuitry. This switch senses the presence of a cassette.
When a cassette is latched to the pump, the cassette presses against the switch in the pump mechanism. Electronic circuitry on the circuit board detects this and reports the information to the microprocessor. This system acts as a safety feature to detect a damaged or detached cassette. If, during operation, the microproces­sor detects the switch open, the pump will enable audible and visual alarms and stop delivery.
Redundancy with the upstream occlusion sensor prevents single fault failures from causing over or under delivery of fluid. Addi­tional circuitry allows these sensors to be turned on and off by the microprocessor to conserve battery power.
Voltage Detector Circuit
Low voltage detection is performed by part of the watchdog circuit and by the microproces­sor via software. Three low voltage levels are detected. The first two levels (Low Battery and Battery Depleted) are detected by software and the third by hardware.
The first level to be reached is the Low Battery Warning threshold which occurs when the battery voltage decays to a nominal value of
2.4 volts when motor is off or 1.8 volts when motor is active. An Analog to Digital Con-
Audible alarm (3 beeps every 5 minutes); Low Bat message appears
Audible alarm (3 beeps every 5 minutes); Low Bat message appears
Battery Depleted message appears
Hardware reset occurs; Pump continues to indicate depleted battery condition
††
verter (ADC) built into the microprocessor allows the microprocessor, via software, to monitor the battery voltage and motor voltage.
At the Low Battery Warning threshold, the microprocessor enables a periodic series of beeps and displays a “Low Bat” warning message on the LCD.
The second level is Battery Depleted Warning threshold. As the voltage operating the motor reaches a nominal value of 4.75 volts, the software disables delivery, places a “Battery Depleted” message on the LCD, and enables a continuous two tone audible alarm.
The third level is a hardware reset which is reached when the battery voltage decays to a nominal value of 1.0 volt. At this point a hardware reset circuit is triggered which places the microprocessor in reset. This prevents ambiguous microprocessor operation as the battery voltage continues to decay. The hard­ware reset continues until the battery is com­pletely discharged or until it is removed. A hardware reset can only be cleared by replac­ing the old batteries with two fresh ones.
The pump emits 3 beeps every 5 minutes, and the message “Low Bat” appears on the pump’s display, indicating that the battery power is low, but the pump is operable.
††
The pump emits a continuous, variable-tone alarm, and the message “Battery Depleted” appears on the display, the battery power is too low to operate the pump and pump operation has stopped.
18
Software Safety Features
Data Handling Software Safety Features
Hardware-related Software Safety Features
Program Memory Check
At power up and at regular intervals thereafter, the program memory is tested by calculating a Cyclic Redundancy Code (CRC) on the program and then comparing it with the CRC stored with the program. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
RAM Memory Check
At power up, the random access memory is checked. A particular bit pattern is written to and read from each address in the RAM. If the read data is different from the written data, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Motor Circuit Check
At power up and at regular intervals thereafter, the motor circuit is checked to ensure that no power is being applied to the motor unless the motor is actually on. If the software detects power being applied to the motor at any other time, it will sound a continuous two-tone audible alarm and will no longer attempt to deliver medication. During every pump activa­tion, the software checks to see whether the motor completes one activation. If the motor fails to turn, or fails to complete a cycle, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Keypad Encoder Check
Key presses are routed to the microprocessor via a keypad encoder. Every time the software receives data from the keypad encoder, it is checked. If the data is not a valid key press, the software disregards it.
The keypad contains a redundant switch in the
key, ‹ key, and Í key (CADD-Legacy
PCA). The redundant switch in each of these keys is routed to the microprocessor via an I/O chip. The microprocessor must see a valid signal simultaneously from the redundant switch and the normal switch (routed through the keypad encoder) before it will start infusing.
Data Stored in RAM
Before use, data associated with delivery and stored in RAM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Data Stored in EEPROM
Before use, data associated with delivery and stored in EEPROM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Data Stored in NOVRAM
Before use, data associated with delivery and stored in NOVRAM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Data Used in Calculations
Calculations on data used in some way to control the delivery of drug are performed redundantly. The two calculated values are then compared. If the two values do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all drug delivery.
Timer Data Registers
The data in the timer Real Time Clock is checked at regular intervals. If the data is not reasonable, the software will turn on a con­tinuous two-tone audible alarm and stop all drug delivery.
19
7 Hardware and Software
Fault Detection
Overview
If the CADD-Legacy™ pump displays an error code, a hardware or software fault has been detected by the microprocessor, and the pump should be returned for servicing.
When hardware or software faults are detected by the microprocessor, pump operation stops and a continuous two-tone audible alarm will be activated. An error message will be dis­played on the LCD. On the next power up, the error code will again be displayed. If the error detected was a data fault, the pump will be in Lock Level 2, and all other programmed functions will have default values. (See the pump’s Operator’s Manual for specific defaults.)
Order of Error Code Events
1. There is a continuous two-tone audible alarm and the display will read
Error Detected
Error Code Description Range
1010-1040 Software Application Errors
1110-1160 Software Control Errors
1210-1270 CRC Errors
1310-1340 Real Time Clock Errors
1410-1450 Standard Delivery Calcula-
tion Errors
1510-1530 Air Detector System Errors
1610-1670 CPU Test Errors
1710-1720 Miscellaneous Hardware
Errors
1810-1872 Motor Errors
Table 10. CADD-Legacy™ pump error codes.
E(XXXX)
NOTE:
“XXXX” is a 4-digit code.
2. To silence the error code alarm, remove the batteries.
3. At each subsequent power up the pump will display the initial power up screen and then the following screen:
LEC XXXX
Thus there is always a display of the “Last Error Code (LEC)” detected by the micro­processor.
20
8 Cleaning and Inspection
Procedures
Inspection Recommendation
Deltec recommends annual functional inspec­tion on the CADD-Legacy PLUS, and CADD-Legacy™ PCA pumps. The following inspection procedures should be performed annually to verify function and accuracy.
NOTE:
Persons performing the following tests and procedures should be familiar with the Deltec CADD-Legacy the Operator’s Manual supplied with the pump before proceeding.
WARNING:
CADD-Legacy
pumps are sealed units. A broken or damaged seal will, therefore, be considered conclusive evidence that the pump has been misused and/or altered, which voids any and all warranties. All service and repair of CADD-Legacy must be performed by Deltec or its author­ized agents.
1, CADD-Legacy
pump. Please read
pumps
Cleaning
Use any of the following solutions to clean the pump and accessories:
• Soap solution
• Benzalkonium chloride concentrate (0.13%)
• Glutaral concentrate, USP (2%)
• 10 percent solution of household bleach (one part household bleach to nine parts water)
• Alcohol, USP (93%)
• Isopropyl Alcohol, USP (99%)
• PDI - Super Sani-Cloth
• Mada Medical - MadaCide
®
1. Dampen a soft, lint-free cloth with cleaning solution. Apply the solution to exterior surface of the pump or accessory. Do not
allow the solution to soak into the pump or accessory.
2. Wipe the entire surface dry with another soft, lint-free cloth. Allow the pump to dry completely before use.
CAUTION:
Do not immerse the pump in cleaning fluid or water. Do not allow solution to soak into the pump, accumulate on the keypad, or enter the battery compart-ment. Moisture build-up inside the pump may damage the pump.
Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners, as damage to the pump may occur.
Battery Contact Cleaning
Pump battery contact cleaning can be per­formed easily using a clean cotton swab wetted with Isopropyl Alcohol or by using a pre­moistened Alcohol swab. Use a minimum of 70% concentration by volume Isopropyl Alcohol. Do not use Alcohol formulation that
has other additives besides Alcohol and water.
• Using a cotton swab wetted with Alcohol or the pre-moistened Alcohol swab, rub with medium pressure over the entire contact surface a minimum of ten back and forth cycles (twenty total wipes over the contact).
• Select a fresh surface of the swab and repeat the cleaning process on the second battery contact. Dispose of the swab when finished.
• Using a second Alcohol wetted swab, rub over each contact surface again a minimum of four back and forth cycles (eight total wipes over the contact). Allow the contacts to dry for a few minutes.
• 70% Chlorohexine
21
Visual Inspection
Mechanical Inspection
• Visually inspect the pump for any damage to the LCD, occlusion sensor seals, valves and expulsor, cassette hinge area, cassette lock, cassette sensor, keypad, indicator light, power jack, accessory jack, air detector, and housing. If any damage is noted, the pump should be returned for service.
• Check the battery door for proper operation. It should not be broken or damaged. The battery door mating tabs on the pump housing should not be broken or damaged.
• Examine the battery compartment for damage. If the battery contacts appear corroded, clean them as instructed on page
20. If the battery contacts appear to be bent or pushed in, straightening may be possible with a small screwdriver or other suitable tool. Care must be taken not to damage the pump housing or to incur further damage to the contacts.
• Press each key on the keypad. Each key should have a distinctive dome feeling. The keys should not feel flat.
• Attach the battery door. The battery door should fit snugly in place when it is closed on the pump.
• Attach either a 50-ml or 100-ml Medication Cassette
Reservoir or a CADD® Adminis­tration Set to the pump. Using a coin (key for the CADD-Legacy
PCA pump), turn the lock 1/4 turn counter-clockwise. Check for smooth operation and a definite feel when the lock pulls the cassette firmly against the bottom of the pump. The slot on the cassette lock should be aligned with the “LOCKED” indicator on the side of the pump.
• Gently twist and pull on the cassette to make sure it is firmly attached.
22
9 Testing Procedures
Testing Recommendation
Deltec recommends annual functional testing on the CADD-Legacy testing procedures should be performed annu­ally to verify function and accuracy.
NOTE:
To perform the following functional tests the pump must be in Lock Level 0.
pumps. The following
Changing to Lock Level 0 (LL0)
Before programming the pump, make sure the lock level is 0. LL0 allows the operator to access all programming and operating func­tions.
1. Make sure the pump is stopped. Press The current lock level will appear. (If the lock level is already LL0, press
2. Press
3. Press
4. Press
5. Press
´ or Î until “LL0” appears. Œ again or ¤. “Code 0” will
appear.
´ or Î until the Lock Level Code
“63” appears for CADD-Legacy “64” for CADD-Legacy CADD-Legacy
PLUS.
Œ or ¤ to set the new lock level.
to exit.)
1 or “65” for
PCA,
Œ.
CADD-Legacy™ PCA Pump
Power-up Check
• Insert batteries or press Å and observe the
LCD during power up. The first screen will display the serial number, model number, and software number with revision level. The second screen will display 32 character blocks. (If “LEC” and four digits appear prior to the pump displaying the 32 charac­ter blocks, the pump has experienced an electrical or mechanical fault and should be returned for service.) If no error message is immediately shown, the pump has powered up normally. The pump will then sequen­tially display all of the programmed values and beep at each screen. After all screens are
displayed, successful power up is indicated with six audible beeps and the “STOPPED” screen displayed. Continue with the Lock check.
To Access Biomed Functions Loop
1. Press Œ. The current lock level will
appear.
2. Press
3. Press
Œ or ¤. “CODE 0” will appear. ´ or Î until the Biomed function
code “163” appears (Lock Code +100). Then press
Œ or ¤.
Air Detector ON/OFF
1. Press until “Air Detector” appears.
2. Use
3. Press
´ or Î to select “Off.”
¤ to enter the change.
Lock Check
• Attach a 50- or 100-ml Medication Cas-
sette
Reservoir or a CADD® Administra­tion Set to the pump. The mark on the Cassette Lock button should be aligned with the “Locked” symbol.
Cassette Sensor Check
• Unlock the cassette by inserting a key into the lock and turning clockwise.
• The pump should issue an audible alarm and the display should read “No Disposable Clamp Tubing”.
• Press
or to silence the alarm. Press
and hold
The following three checks (LCD, Motor and Gear Train, and Reservoir Volume Empty Alarm Check) should be performed in the sequence shown.
Å to turn the pump off.
LCD Check
• With the pump turned off, press Å. The second screen that the pump displays will consist of 32 blocks of characters. Examine the LCD to verify that there are no missing dark pixels in the character blocks.
23
Testing Procedures - CADD-Legacy™ PCA pump continued
Motor and Gear Train Check
• Program the Reservoir Volume to 2.0 ml.
• Attach either a 50- or 100-ml Medication Cassette tion Set to the pump. Lock the cassette.
• Press and hold dashes appear. Release
motor for excessive noise or grinding sounds. Count the number of pump activa­tions. The pump should prime ten double activations and then stop. Press return to main menu.
Reservoir or CADD® Administra-
until three series of
. Press and hold
. While priming the pump, listen to the
to
Reservoir Volume Empty Alarm Check
• Program the Reservoir Volume to 1.0 ml. Press played on the LCD. Press
1.0 ml is displayed. Then press
• Press and hold dashes appear. Release
activations and then stop. The pump will alarm and display “Reservoir Volume Empty.” Press
until Reservoir Volume is dis-
´ or
until three series of
. Press and hold
. The pump should prime ten double
.
Î
¤
until
.
Starting/Stopping the Pump
• Program the pump with the following values: Reservoir Volume: 1.0 ml Units: milliliters Continuous Rate: 50 ml/hr Demand Dose: 0.00 ml Given: 0.00 (Press
• Program the Air Detector Off.
• Press and hold followed by three sets of dashes, each ac­companied by a beep. A review of the programmed parameters then appears. The main screen should appear with “RUN” in the display.
• To stop the device, press and hold “Stopping” appears followed by three sets of dashes that disappear one at a time, each accompanied by a beep. The main screen should appear with “STOPPED” in the display.
⁄.
)
¤
“Starting” appears
.
Activation Timing Check
• Reprogram the Reservoir Volume to 1.0 and clear the Given screen.
• Press and hold disappear from the display. The pump should sequentially display all of the pro­grammed values. Start a timer at the first motor activation.
• Count the activations. One activation should occur every six seconds. Approximately sixty-six seconds and ten activations later, the Reservoir Volume alarm should occur. The display should show “Reservoir Volume Empty.”
until three dashes
DOSE Key Check (CADD-Legacy™ PCA pump only)
• Check the ming the pump with the following values:
Reservoir Volume: 10.0 ml Units: Milliliters Continuous Rate: 0.0 ml/hr Demand Dose: 1.00 ml Dose Lockout: 00 hrs 5 min Doses Per Hour: 12 Doses Given: 0 doses (Press
key operation by program-
Í
¤ to clear)
Doses Attempted: 0 doses (Press
¤ to clear)
Given: 0.00 ml (Press
¤ to clear)
• Press and hold sequentially display all of the programmed values.
• After “RUN” appears on the display, press
and note the time. The pump should
Í
beep twice and begin to deliver. Count the number of pump activations. The pump should make ten double activations. After ten double activations, the display should show a Reservoir Volume of 9.0 ml. Press
two more times within the next 5
Í
minutes. The pump should not deliver.
Remote Dose Cord Check (CADD-Legacy™ PCA pump only)
• Wait 5 minutes after the dose given above; then, instead of pressing button on the Remote Dose Cord. The pump should make ten double activations. After ten double activations, the display should show a Reservoir Volume of 8.0 ml. Press
. The pump should
Í
, press the
24
Testing Procedures - CADD-Legacy™ PCA pump continued
the Remote Dose Cord button two more times within the next 5 minutes. The pump should not deliver.
Doses Given and Doses Attempted Check (CADD-Legacy™ PCA pump only)
• Stop the pump by pressing and holding Use screen. The screen should show 2. Use to advance to the Doses Attempted screen. The display should show 6. (If the above steps have not been followed exactly, differ­ent values may appear.)
to advance to the Doses Given
GIVEN Mode Check
• Press to advance to the Given screen. The display should now show
2.00 ml. (If the above steps have not been followed exactly, a different value may appear.)
• Press the show 0.00 ml.
¤ key. The display should now
Air Detector Test
This test will verify the function of the air detector. To perform this test, the air detector must be turned on. The previous program from the
Í key check can be used to perform this
test.
• Attach an empty Medication Cassette Reservoir or CADD® Administration Set to the pump.
• Start the pump.
• Deliver a demand dose. (NOTE: Five min­utes must have passed since the delivery of the last demand dose.)
• The pump should deliver the dose without
.
an air detection alarm.
Upstream Occlusion Sensor Test
• Verify that the Upstream Occlusion Sensor is turned on. (See page 23, To Access Biomed Functions Loop.)
• Obtain a CADD bag spike and anti-siphon valve. Also obtain a clamp (slide clamp or hemostat).
• Insert the CADD into an appropriate, standard IV bag filled with water. Attach the cassette to the pump. Prime the entire fluid path.
• Program the pump to deliver a continuous maximum rate. Press and hold the pump.
• Clamp the tubing halfway between the IV bag and the pump. The pump should alarm within three activations after clamping the tubing.
®
Administration Set with
®
Administration Set spike
to start
Occlusion Pressure Range Tests
Occlusion Pressure Range Test I
• Secure it using the lock button.
• Thread the tubing through the air detector groove.
• Start the pump.
• The pump should respond with a continuous two-tone alarm and the display should read:
“Air In Line Detected”
• Press
• Now attach a Medication Cassette
• Secure it using the lock button.
• Thread the tubing through the air detector
or to silence the alarm, and
remove the Medication Cassette or CADD
voir containing fluid, or a primed CADD Administration Set to the pump. Make certain there is no air in the fluid path.
groove.
®
Administration Set.
Reservoir
Reser-
®
Description
Pressure is generated by activating the pump­ing mechanism with an attached filled, clamped Medication Cassette pump is started and a Demand Dose is given until the high pressure alarm sounds.
Equipment needed
50- or 100-ml Medication Cassette containing distilled water
Procedure
1. Insert two AA batteries and wait for the pump to power up.
2. Attach a Medication Cassette containing water to the pump. Lock the cassette.
Reservoir. The
Reservoir
Reservoir
25
Testing Procedures - CADD-Legacy™ PCA pump continued
3. Prime the Medication Cassette™ Reservoir tubing. The tubing should be filled with fluid to the end of the luer lock connector. The system must be free from air bubbles for this test.
4. Close the slide clamp on the distal end of the tubing near the female luer of the
Medication Cassette
Reservoir.
5. Program the pump to the following param­eters:
Reservoir Volume: 10.0 ml Units: Milliliters Continuous Rate: 0.00 ml/hr Demand Dose: 1.00 ml Dose Lockout: 00 hrs 5 min Doses Per Hour: 12 Dose Given: 0 (Press
¤ to
clear)
Dose Attempts: 0 (Press
¤ to
clear)
Milligrams Given: 0.00 ml (Press
¤
to clear)
6. Start the pump. When the pump is running, activate a Demand Dose, noting when the high pressure alarm is activated.
Equipment needed
• Pressure gauge, 40 psi ± 1 psi (2.76 bar ± 0.07 bar)
• Pressure vessel, partially filled with water
• Pressure regulator, 40 psi (2.76 bar ± 0.07 bar)
• 50 or 100 ml Medication Cassette
Reservoir
containing water
CAUTION:
At the completion of the test, the pressure must be reduced to zero before detaching the cassette from the pump; otherwise, the cassette may rupture. Safety glasses should be worn while conducting or observing this test.
Procedure
1. Insert two AA batteries and wait for the pump
to power up.
2. Attach a Medication Cassette
Reservoir to the
pump. Lock the cassette.
NOTE:
The pressure from the source must be zero when the cassette is attached.
7. The pump should alarm when the pump delivers between 1 and 2 activations.
Occlusion Pressure Range Test II
Description
An adjustable metered pressure source is connected to the Medication Cassette voir tubing. The pressure is slowly increased until the high pressure alarm sounds.
1998-11-22 D. Zurn «Lgc Occlusion Setup»
40
PSI
Pressure GaugeRegulator
Reser-
3. Assemble the apparatus as shown in Figure 11.
4. Connect the Medication Cassette
Reservoir
outlet tube to the metered pressure source.
NOTE:
Do not use a CADD
®
Extension Set with
Anti-siphon Valve.
5. Program the pump for a continuous rate of 50 ml/hr. Press
⁄.
6. Slowly increase the back pressure, noting when the high pressure alarm is activated.
NOTE:
The pressure may be increased rapidly to 8 psi (0.55 bar), after which the pressure should be increased at 3 psi/min (0.21 bar/min) or less until the alarm sounds.
7. The high pressure alarm should sound within
1.79 (±0.97) bar, or 26 (±14) psi.
Figure 11. Occlusion test set-up.
26
Testing Procedures - CADD-Legacy™ PCA pump continued
Accuracy Testing
Gravimetric Accuracy Testing
Description
A Medication Cassette filled with water and weighed. The cassette is then attached to the pump and the pump is set to deliver a certain amount of water. The cassette is then removed and weighed again. The amount of water delivered is compared to the amount that the pump should have deliv­ered.
Nominal system accuracy is given in the techni­cal specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medication Cassette Reservoir will be nominal with a 90% confi­dence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed
• 50- or 100-ml Medication Cassette voir
• 50 or 60 ml syringe
• CADD
®
Extension Set with Anti-siphon Valve
Reservoir is partially
Reser-
3. Weigh the entire Medication Cassette
Reservoir/CADD® Extension Set assembly and record the weight. This is the pre- delivery weight. (This weight includes the empty Medication Cassette CADD
®
Extension Set, and weight of the
Reservoir,
water.)
4. Attach the Medication Cassette
Reservoir to the pump. Program the reservoir volume to 20 ml. Now press
¤. This value is the
intended delivery volume. Remove the slide clamp.
5. With the pump in Lock Level 0, program a continuous rate of 50 ml/hr. Start the pump and deliver 20 ml.
6. Again, secure the slide clamp as close as possible to end of the CADD
®
Extension Set luer lock connector. Remove the Medication Cassette weigh the entire Medication Cassette
Reservoir from the pump and
Reservoir/CADD® Extension Set assembly. This is the post-delivery weight.
7. Calculate the difference in weight between the pre-delivery weight and the post-delivery weight. This is the weight of the amount delivered. (1 ml of water at 20° weighs 1 gram.)
• A balance accurate to 0.1 g
• 40 ml of room temperature water
Procedure
1. Fill the 50 or 60 ml syringe with 40 ml of water. Transfer the water into a Medication Cassette
Reservoir.
8. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy volume.
9. Divide the inaccuracy volume by the in­tended delivery volume and multiply by 100. This is the accuracy error percentage. (See Table 11.)
2. Remove any air from the Medication Cas-
sette
Reservoir by aspirating the air with
the syringe. Attach the CADD
®
Extension Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock connector.
10.If the accuracy error percentage is greater than ± 6%, repeat the test with a new Medi­cation Cassette
Reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.
Pre- Post- Weight of Intended Inaccuracy Accuracy Accuracy
Delivery Delivery Amount Delivery Volume Error Error Percentage
Weight Weight Delivered Volume
61.1 g 41.6 g 19.5 g = 19.5 ml 20 ml -0.5 ml -0.5 ml ÷ 20.0 ml = -0.025 x 100 = -2.5%
-0.025
Table 11. Gravimetric percentage calculation
27
Testing Procedures - CADD-Legacy™ PCA pump continued
Example:
Pre-delivery Weight: 61.1 g Post-delivery Weight: – 41.6 g
Weight of Amount Delivered: 19.5 g
= 19.5 ml
Volume of Amount Delivered: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: -0.025
x 100.00
Accuracy Error Percentage = -2.5%
Volumetric Accuracy Testing
Description:
A predetermined amount of water is delivered into a collection device such as a burette or graduated cylinder. The amount of water delivered is compared to the amount that the pump should have delivered.
Nominal system accuracy is given in the technical specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medica­tion Cassette a 90% confidence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed:
• 50- or 100- ml Medication Cassette voir
• 50- or 60- ml syringe
• CADD Valv e
• A fluid collection device such as a burette or a class A 25 ml capacity graduated cylinder
Reservoir will be nominal with
®
Extension Set with Anti-siphon
Reser-
• 40 ml of room temperature water
NOTE:
The test procedure calls for the use of a Medication Cassette CADD
®
Extension Set with Anti-siphon
Valve. An IV bag and CADD
Reservoir and a
®
Admini­stration Set with integral or add-on anti­siphon valve can be substituted for the Medication Cassette
®
CADD
Extension Set with Anti-siphon
Reservoir and
Valve.
Procedure:
1. Fill the 50- or 60-ml syringe with 40 ml of water. Transfer the water into a Medication Cassette
Reservoir.
2. Remove any air from the Medication Cassette with the syringe. Attach the CADD
Reservoir by aspirating the air
®
Exten­sion Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock connec-
tor.
3. Insert the end of the CADD
®
Extension Set
into the fluid collection device.
4. Attach the Medication Cassette
Reservoir to the pump. Program the Reservoir Volume to 20 ml. This is the intended delivery volume. Remove all clamps.
5. Program a continuous rate of 50 ml/hr. Start the pump and deliver 20 ml.
6. When delivery is complete, record the volume of fluid delivered. This is the actual delivery.
7. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy volume.
8. Divide the inaccuracy volume by the in­tended delivery volume and multiply by
100. This is the accuracy error percentage. (See Table 12.)
Intended Actual Inaccuracy Accuracy Accuracy Delivery Delivery Volume Error Error Percentage
Volume Volume
20 ml 19.5 ml -0.5 ml -0.5 ml ÷ 20.0 ml = -0.025 -0.025 x 100 = -2.5%
Table 12. Volumetric percentage calculation
28
9. If the accuracy error percentage is greater than ± 6%, repeat the test with a new Medication Cassette pump fails a second time, call Deltec’s Customer Service Department.
Example:
Actual Delivery Volume: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: -0.025
Accuracy Error Percentage: = -2.5%
Reservoir. If the
x 100.00
CADD-Legacy™ 1 Pump
Power-up Check
• Insert batteries or press Å and observe the
LCD during power up. The first screen will display the serial number, model number, and software number with revision level. The second screen will display 32 character blocks. (If “LEC” and four digits appear prior to the pump displaying the 32 charac­ter blocks, the pump has experienced an electrical or mechanical fault and should be returned for service.) If no error message is immediately shown, the pump has powered up normally. The pump will then sequen­tially display all of the programmed values and beep at each screen. After all screens are displayed, successful power up is indicated with six audible beeps and the “STOPPED” screen displayed. Continue with the Lock check.
To Access Biomed Functions Loop
1. Press appear.
2. Press
3. Press code “164” appears (Lock Code +100). Then press
Air Detector ON/OFF
1. Press
2. Use
3. Press
. The current lock level will
Œ
or
Œ
or Î until the Biomed function
´
Œ
until “Air Detector” appears.
or Î to select “Off.”
´
to enter the change.
¤
. “CODE 0” will appear.
¤
or
¤
.
Lock Check
• Attach a 50- or 100-ml Medication Cas-
sette
Reservoir or a CADD® Administra­tion Set to the pump. The slot on the Cas­sette Lock button should be aligned with the “Locked” symbol.
Cassette Sensor Check
• Unlock the cassette by inserting a coin into the latch and turning clockwise.
• The pump should issue an audible alarm and the display should read “No Disposable Clamp Tubing”.
• Press
or to silence the alarm. Press
and hold
The following three checks (LCD, Motor and Gear Train, and Reservoir Volume Empty Alarm Check) should be performed in the sequence shown.
Å to turn the pump off.
LCD Check
• With the pump turned off, press Å. The second screen that the pump displays will consist of 32 blocks of characters. Examine the LCD to verify that there are no missing dark pixels in the character blocks.
Motor and Gear Train Check
• Program the Lock Level to LL0. Program the Reservoir Volume to 2.0 ml.
• Attach either a 50- or 100-ml Medication Cassette tion Set to the pump. Lock the cassette.
• Press and hold dashes appear. Release
motor for excessive noise or grinding sounds. Count the number of pump activa­tions. The pump should prime ten double activations and then stop. Press return to the main menu.
Reservoir or CADD® Administra-
until three series of
. Press and hold
. While priming the pump, listen to the
to
Reservoir Volume Empty Alarm Check
• Program the Reservoir Volume to 1.0 ml.
• Press and hold dashes appear. Release
. The pump should prime ten double
activations and then stop. The pump will alarm and display “Reservoir Volume Empty.” Press
until three series of
.
. Press and hold
29
Testing Procedures - CADD-Legacy™ 1 pump continued
Starting/Stopping the Pump
• Program the pump with the following values:
Reservoir Volume: 1.0 ml Rate: 3000 ml/24hrs
¤
.
Given: 0.0 ml (press
to clear)
• Program the Air Detector Off.
• Press and hold followed by three sets of dashes, each ac­companied by a beep. A review of the programmed parameters then appears. The main screen should appear with “RUN” in the display.
• To stop the device, press and hold “Stopping” appears followed by three sets of dashes that disappear one at a time, each accompanied by a beep. The main screen should appear with “STOPPED” in the display.
“Starting” appears
.
Activation Timing Check
• Reprogram the Reservoir Volume to 1.0 and clear the Given screen.
• Press and hold disappear from the display. The pump should sequentially display all of the pro­grammed values. Start a timer at the first motor activation.
• Count the activations. One activation should occur every three seconds. Approximately twenty-seven seconds and ten activations later, the Reservoir Volume alarm should occur. The display should show “Reservoir Volume Empty.”
until three dashes
GIVEN Mode Check
• Press to advance to the Given screen. The display should now show 1.00 ml. (If the above steps have not been followed exactly, a different value may appear.)
• Press the show 0.00 ml.
¤ key. The display should now
• Attach an empty Medication Cassette Reservoir or CADD® Administration Set to the pump.
• Secure it using the lock button.
• Thread the tubing through the air detector groove.
• Start the pump.
• The pump should respond with a continuous two-tone alarm and the display should read:
“Air In Line Detected”
• Press
• Now attach a Medication Cassette
• Secure it using the lock button.
• Thread the tubing through the air detector
• Program the Reservoir Volume to 1.0 ml.
• The pump should deliver without an air
or to silence the alarm, and
remove the Medication Cassette or CADD
voir containing fluid or a primed CADD Administration Set to the pump. Make certain there is no air in the fluid path.
groove.
Start the pump.
detection alarm.
®
Administration Set.
Reservoir
Reser-
®
Upstream Occlusion Sensor Test
• Verify that the Upstream Occlusion Sensor is turned on. (See page 29, To Access Biomed Functions Loop.)
• Obtain a CADD bag spike and anti-siphon valve. Also obtain a clamp (slide clamp or hemostat).
• Insert the CADD into an appropriate, standard IV bag filled with water. Attach the cassette to the pump. Prime the entire fluid path.
• Program the pump to deliver a continuous maximum rate. Press and hold the pump.
®
Administration Set with
®
Administration Set spike
to start
Air Detector Test
This test will verify the function of the air detector. To perform this test, the air detector must be turned on. The previous program from the Activation Timing Check can be used to perform this test.
30
• Clamp the tubing halfway between the IV bag and the pump. The pump should alarm within three activations after clamping the tubing.
Testing Procedures - CADD-Legacy™ 1 pump continued
Occlusion Pressure Range Tests
Occlusion Pressure Range Test I
Description
Pressure is generated by activating the pump­ing mechanism with an attached filled, clamped Medication Cassette pump is started and a Demand Dose is given until the high pressure alarm sounds.
Equipment needed
50- or 100-ml Medication Cassette containing distilled water.
Procedure
1. Insert two AA batteries and wait for the pump to power up.
2. Attach a Medication Cassette containing water to the pump. Lock the cassette.
3. Prime the Medication Cassette tubing. The tubing should be filled with fluid to the end of the luer lock connector. The system must be free from air bubbles for this test.
Reservoir. The
Reservoir
Reservoir
Reservoir
7. The pump should alarm when the pump delivers between 1 and 2 activations.
Occlusion Pressure Range Test II
Description
An adjustable metered pressure source is connected to the Medication Cassette voir tubing. The pressure is slowly increased until the high pressure alarm sounds.
Equipment needed
• Pressure gauge, 40 psi ± 1 psi (2.76 bar ±
0.07 bar).
• Pressure vessel, partially filled with water.
• Pressure regulator, 40 psi (2.76 bar ± 0.07
bar).
• 50 or 100 ml Medication Cassette
voir containing water.
CAUTION:
At the completion of the test, the pressure must be reduced to zero before detaching the cassette from the pump; otherwise, the cassette may rupture. Safety glasses should be worn while conducting or observing this test.
Reser-
Reser-
4. Close the slide clamp on the distal end of the tubing near the female luer of the Medication Cassette
Reservoir.
5. Program the pump to the following param­eters:
Reservoir Volume: 10.0 ml Continuous Rate: 3000.0 ml/24hr
Given: 0.0 ml (Press
¤ to
clear)
6. Start the pump. When the pump is running, note when the high pressure alarm is acti­vated.
1998-11-22 D. Zurn «Lgc Occlusion Setup»
40
PSI
Pressure GaugeRegulator
Procedure
1. Insert two AA batteries and wait for the pump to power up.
2. Attach a Medication Cassette
Reservoir to
the pump. Lock the cassette.
NOTE:
The pressure from the source must be zero when the cassette is attached.
3. Assemble the apparatus as shown in Figure
12.
4. Connect the Medication Cassette
Reser­voir outlet tube to the metered pressure source.
NOTE:
Do not use a CADD
®
Extension Set with
Anti-siphon Valve.
5. Program the pump for a continuous rate of 3000 ml/24hr. Press
⁄.
Figure 12 . Occlusion test set-up.
6. Slowly increase the back pressure, noting when the high pressure alarm is activated.
31
Testing Procedures - CADD-Legacy™ 1 pump continued
NOTE:
The pressure may be increased rapidly to 8 psi (0.55 bar), after which the pressure should be increased at 3 psi/min (0.21 bar/ min) or less until the alarm sounds.
7. The high pressure alarm should sound within 1.79 (±0.97) bar, or 26 (±14) psi.
Accuracy Testing
Gravimetric Accuracy Testing
Description
A Medication Cassette filled with water and weighed. The cassette is then attached to the pump and the pump is set to deliver a certain amount of water. The cassette is then removed and weighed again. The amount of water delivered is compared to the amount that the pump should have deliv­ered.
Nominal system accuracy is given in the technical specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medica­tion Cassette
Reservoir will be nominal with a 90% confidence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed
• 50- or 100-ml Medication Cassette
voir
• 50 or 60 ml syringe
• CADD
®
Extension Set with Anti-siphon
Valv e
• A balance accurate to 0.1 g
• 40 ml of room temperature water
Reservoir is partially
Reser-
Procedure
1. Fill the 50 or 60 ml syringe with 40 ml of water. Transfer the water into a Medication Cassette
Reservoir.
2. Remove any air from the Medication Cassette with the syringe. Attach the CADD
Reservoir by aspirating the air
®
Exten­sion Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock connec-
tor.
3. Weigh the entire Medication Cassette
Reservoir/CADD® Extension Set assembly and record the weight. This is the pre- delivery weight. (This weight includes the empty Medication Cassette CADD
®
Extension Set, and weight of the
Reservoir,
water.)
4. Attach the Medication Cassette
Reservoir to the pump. Program the reservoir volume to 20 ml. Now press
¤. This value is the
intended delivery volume. Remove the slide clamp.
5. With the pump in Lock Level 0, program a continuous rate of 3000 ml/24hr. Start the pump and deliver 20 ml.
6. Again, secure the slide clamp as close as possible to end of the CADD
®
Extension Set luer lock connector. Remove the Medication Cassette weigh the entire Medication Cassette
Reservoir from the pump and
Reservoir/CADD® Extension Set assembly. This is the post-delivery weight.
7. Calculate the difference in weight between the pre-delivery weight and the post-deliv­ery weight. This is the weight of the amount delivered. (One ml of water at 20°C weighs 1 gram.)
8. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy
volume.
Pre- Post- Weight of Intended Inaccuracy Accuracy Accuracy
Delivery Delivery Amount Delivery Volume Error Error Percentage
Weight Weight Delivered Volume
61.1 g 41.6 g 19.5 g = 19.5 ml 20 ml -0.5 ml -0.5 ml ÷ 20.0 ml = -0.025 x 100 = -2.5%
-0.025
Table 13. Gravimetric percentage calculation
32
Testing Procedures - CADD-Legacy™ 1 pump continued
9.Divide the inaccuracy volume by the in­tended delivery volume and multiply by
100. This is the accuracy error percentage. (See Table 13.)
10.If the accuracy error percentage is greater than ± 6%, repeat the test with a new
Medication Cassette
Reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.
Example:
Pre-delivery Weight: 61.1 g Post-delivery Weight: – 41.6 g
Weight of Amount Delivered: 19.5 g
= 19.5 ml
Volume of Amount Delivered: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: -0.025
x 100.00
Accuracy Error Percentage: = -2.5%
Volumetric Accuracy Testing
Description:
A predetermined amount of water is delivered into a collection device such as a burette or graduated cylinder. The amount of water delivered is compared to the amount that the pump should have delivered.
Nominal system accuracy is given in the technical specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medica­tion Cassette a 90% confidence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed:
• 50- or 100- ml Medication Cassette voir
• 50- or 60- ml syringe
• CADD Valv e
Reservoir will be nominal with
®
Extension Set with Anti-siphon
Reser-
• A fluid collection device such as a burette or a class A 25 ml capacity graduated cylinder
• 40 ml of room temperature water
NOTE:
The test procedure calls for the use of a Medication Cassette CADD
®
Extension Set with Anti-siphon
Valve. An IV bag and CADD
Reservoir and a
®
Admini­stration Set with integral or add-on anti­siphon valve can be substituted for the
Medication Cassette CADD
®
Extension Set with Anti-siphon
Reservoir and
Valve.
Procedure:
1. Fill the 50- or 60-ml syringe with 40 ml of water. Transfer the water into a Medication Cassette
Reservoir.
2. Remove any air from the Medication Cassette with the syringe. Attach the CADD
Reservoir by aspirating the air
®
Exten­sion Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock
connector.
3. Insert the end of the CADD
®
Extension Set
into the fluid collection device.
4. Attach the Medication Cassette
Reservoir to the pump. Program the Reservoir Volume to 20 ml. This is the intended delivery volume. Remove all clamps.
5. Program a continuous rate of 3000 ml/24hr. Start the pump and deliver 20 ml.
6. When delivery is complete, record the volume of fluid delivered. This is the actual delivery.
7. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy volume.
8. Divide the inaccuracy volume by the in­tended delivery volume and multiply by
100. This is the accuracy error percentage. (See Table 14.)
9. If the accuracy error percentage is greater than ± 6%, repeat the test with a new Medication Cassette
Reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.
33
Example:
Air Detector ON/OFF
Actual Delivery Volume: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: = -0.025
x 100.00
Accuracy Error Percentage: = -2.5%
CADD-Legacy™ PLUS Pump
Power-up Check
• Insert batteries or press Å and observe the
LCD during power up. The first screen will display the serial number, model number, and software number with revision level. The second screen will display 32 character blocks. (If “LEC” and four digits appear prior to the pump displaying the 32 charac­ter blocks, the pump has experienced an electrical or mechanical fault and should be returned for service.) If no error message is immediately shown, the pump has powered up normally. The pump will then sequen­tially display all of the programmed values and beep at each screen. After all screens are displayed, successful power up is indicated with six audible beeps and the “STOPPED” screen displayed. Continue with the Lock check.
To Access Biomed Functions Loop
1. Press appear.
2. Press
3. Press code “165” appears (Lock Code +100). Then press
. The current lock level will
Œ
or
Œ
or Î until the Biomed function
´
Œ
. “CODE 0” will appear.
¤
or
¤
.
1. Press until “Air Detector” appears.
2. Use
3. Press
´ or Î to select “Off.”
¤ to enter the change.
Changing Delivery Modes
1. Press until “Delivery Mode” appears.
2. Press
3. Press
´ or Î to select “Continuous”. ¤ to enter the change.
Lock Check
• Attach a 50- or 100-ml Medication Cas-
sette
Reservoir or a CADD® Administra­tion Set to the pump. The slot on the Cas­sette Lock button should be aligned with the “Locked” symbol.
Cassette Sensor Check
• Unlock the cassette by inserting a coin into the lock and turning clockwise.
• The pump should issue an audible alarm and the display should read “No Disposable Clamp Tubing”.
• Press
or to silence the alarm.
Press and hold
The following three checks (LCD, Motor and Gear Train, and Reservoir Volume Empty Alarm Check) should be performed in the sequence shown.
Å to turn the pump off.
LCD Check
• With the pump turned off, press Å. The second screen that the pump displays will consist of 32 blocks of characters. Examine the LCD to verify that there are no missing dark pixels in the character blocks.
Motor and Gear Train Check
• Program the Reservoir Volume to 2.0.
• Program Biomed to Continuous Mode.
Intended Actual Inaccuracy Accuracy Accuracy
Delivery Delivery Volume Error Error Percentage
Volume Volume
20 ml 19.5 ml -0.5 ml -0.5 ml ÷ 20.0 ml =-0.025 -0.025 x 100 = -2.5%
Table 14. Volumetric percentage calculation
34
Testing Procedures - CADD-Legacy™ PLUS pump continued
• Attach either a 50- or 100-ml Medication Cassette tion Set to the pump. Lock the cassette.
• Press and hold dashes appear. Release
motor for excessive noise or grinding sounds. Count the number of pump activa­tions. The pump should prime ten double activations and then stop. Press return to the main menu.
Reservoir or CADD® Administra-
until three series of
. Press and hold
. While priming the pump, listen to the
to
Reservoir Volume Empty Alarm Check
• Program the Reservoir Volume to 1.0 ml.
• Press and hold dashes appear. Release
. The pump should prime ten double
activations and then stop. The pump will alarm and display “Reservoir Volume Empty.” Press
until three series of
.
. Press and hold
Starting/Stopping the Pump
• Program the pump with the following values:
Reservoir Volume: 1.0 ml Rate: 125 ml/hr Given: 0.0 ml (press
to clear)
• Program the Air Detector Off.
• Press and hold followed by three sets of dashes, each ac­companied by a beep. A review of the programmed parameters then appears. The main screen should appear with “RUN” in the display.
• To stop the device, press and hold “Stopping” appears followed by three sets of dashes that disappear one at a time, each accompanied by a beep. The main screen should appear with “STOPPED” in the display.
“Starting” appears
⁄.
¤
.
Activation Timing Check
• Reprogram the Reservoir Volume to 1.0 and clear the Given Screen.
• Press and hold disappear from the display. The pump should sequentially display all of the pro­grammed values. Start a timer at the first motor activation.
until three dashes
• Count the activations. One activation should occur every three seconds. Approximately twenty-seven seconds and ten activations later, the Reservoir Volume alarm should occur. The display should show “Reservoir Volume Empty.”
GIVEN Mode Check
• Press to advance to the Given screen. The display should now show 1.00 ml. (If the above steps have not been followed exactly, a different value may appear.)
• Press the show 0.00 ml.
¤ key. The display should now
Air Detector Test
This test will verify the function of the air detector. To perform this test, the air detector must be turned on. The previous program from the Activation Timing Check can be used to perform this test.
• Attach an empty Medication Cassette Reservoir or CADD® Administration Set to the pump.
• Secure it using the lock button.
• Thread the tubing through the air detector groove.
• Start the pump.
• The pump should respond with a continuous two-tone alarm and the display should read:
“Air in Line Detected”
• Press
• Now attach a Medication Cassette
• Secure it using the lock button.
• Thread the tubing through the air detector
• Program the Reservoir Volume to 1.0 ml.
• The pump should deliver without an air
or to silence the alarm, and
remove the Medication Cassette or CADD
voir containing fluid, or a primed CADD Administration Set to the pump. Make certain there is no air in the fluid path.
groove.
Start the pump.
detection alarm.
®
Administration Set.
Reservoir
Reser-
®
35
Testing Procedures - CADD-Legacy™ PLUS pump continued
Upstream Occlusion Sensor Test
• Verify that the Upstream Occlusion Sensor is turned on. (See page 34, To Access Biomed Functions Loop.)
• Obtain a CADD
®
Administration Set with bag spike and anti-siphon valve. Also obtain a clamp (slide clamp or hemostat).
• Insert the CADD
®
Administration Set spike into an appropriate, standard IV bag filled with water. Attach the cassette to the pump. Prime the entire fluid path.
• Program the pump to deliver a continuous maximum rate. Press and hold
to start
the pump.
• Clamp the tubing halfway between the IV bag and the pump. The pump should alarm within three activations after clamping the tubing.
Occlusion Pressure Range Tests
Occlusion Pressure Range Test I
Description
Pressure is generated by activating the pump­ing mechanism with an attached filled, clamped Medication Cassette pump is started until the high pressure alarm sounds.
Equipment needed
50- or 100-ml Medication Cassette containing distilled water
Procedure
1. Insert two AA batteries and wait for the
pump to power up.
2. Attach a Medication Cassette
containing water to the pump. Lock the cassette.
3. Prime the Medication Cassette
tubing. The tubing should be filled with fluid to the end of the luer lock connector. The system must be free from air bubbles for this test.
4. Close the slide clamp on the distal end of
the tubing near the female luer of the Medication Cassette
36
Reservoir. The
Reservoir.
Reservoir
Reservoir
Reservoir
5. Program the pump to the following param­eters:
Reservoir Volume: 10.0 ml Continuous Rate: 125.0 ml/hr Given: 0.0 ml (Press
¤ to
clear)
6. Start the pump. When the pump is running, note when the high pressure alarm is acti­vated.
7. The pump should alarm when the pump delivers between 1 and 2 activations.
Occlusion Pressure Range Test II
Description
An adjustable metered pressure source is connected to the Medication Cassette voir tubing. The pressure is slowly increased until the high pressure alarm sounds.
Equipment needed
• Pressure gauge, 40 psi ± 1 psi (2.76 bar ±
0.07 bar)
• Pressure vessel, partially filled with water
• Pressure regulator, 40 psi (2.76 bar ± 0.07
bar)
• 50- or 100-ml Medication Cassette
voir containing water
CAUTION:
At the completion of the test, the pressure must be reduced to zero before detaching the cassette from the pump; otherwise, the cassette may rupture. Safety glasses should be worn while conducting or observing this test.
Procedure
1. Insert two AA batteries and wait for the pump to power up.
2. Attach a Medication Cassette the pump. Lock the cassette.
NOTE:
The pressure from the source must be zero when the cassette is attached.
3. Assemble the apparatus as shown in Figure
13.
4. Connect the Medication Cassette voir outlet tube to the metered pressure source.
Reser-
Reser-
Reservoir to
Reser-
Testing Procedures - CADD-Legacy™ PLUS pump continued
NOTE:
Do not use a CADD
®
Extension Set with
Anti-siphon Valve.
5. Program the pump for a continuous rate of 125 ml/hr. Press
⁄.
6. Slowly increase the back pressure, noting when the high pressure alarm is activated.
NOTE:
The pressure may be increased rapidly to 8 psi (0.55 bar), after which the pressure should be increased at 3 psi/min (0.21 bar/ min) or less until the alarm sounds.
7. The high pressure alarm should sound within
1.79 (±0.97) bar, or 26 (±14) psi.
Accuracy Testing
Gravimetric Accuracy Testing
Description
A Medication Cassette filled with water and weighed. The cassette is then attached to the pump and the pump is set to deliver a certain amount of water. The cas­sette is then removed and weighed again. The amount of water delivered is compared to the amount that the pump should have delivered.
Nominal system accuracy is given in the techni­cal specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medication Casssette
Reservoir will be nominal with a 90% confidence level. The nominal test condi­tions are as follows: degassed water at 25 ± 5°C without back pressure.
Equipment needed
• 50- or 100-ml Medication Cassette
• 50 or 60 ml syringe
1998-11-22 D. Zurn «Lgc Occlusion Setup»
40
PSI
Reservoir is partially
Pressure GaugeRegulator
Reservoir
• CADD
®
Extension Set with Anti-siphon
Valve
• A balance accurate to 0.1 g
• 40 ml of room temperature water
Procedure
1. Fill the 50- or 60-ml syringe with 40 ml of water. Transfer the water into a Medication Cassette
Reservoir.
2. Remove any air from the Medication Cassette with the syringe. Attach the CADD
Reservoir by aspirating the air
®
Exten­sion Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock
connector.
3. Weigh the entire Medication Cassette
Reservoir/CADD® Extension Set assembly and record the weight. This is the pre- delivery weight. (This weight includes the empty Medication Cassette CADD
®
Extension Set, and weight of the
Reservoir,
water.)
4. Attach the Medication Cassette
Reservoir to the pump. Program the reservoir volume to 20 ml. Now press
¤. This value is the
intended delivery volume. Remove the slide clamp.
5. With the pump in Lock Level 0, program a continuous rate of 125 ml/hr. Start the pump and deliver 20 ml.
6. Again, secure the slide clamp as close as possible to end of the CADD
®
Extension Set luer lock connector. Remove the Medication Cassette weigh the entire Medication Cassette
Reservoir from the pump and
Reservoir/CADD® Extension Set assembly. This is the post-delivery weight.
7. Calculate the difference in weight between the pre-delivery weight and the post-deliv­ery weight. This is the weight of the amount delivered. (One ml of water at 20°C weighs 1 gram.)
8. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy volume.
Figure 13. Occlusion test set-up.
37
Testing Procedures - CADD-Legacy™ PLUS pump continued
9. Divide the inaccuracy volume by the in­tended delivery volume and multiply by
100. This is the accuracy error percentage. (See Table 15.)
10.If the accuracy error percentage is greater than ± 6%, repeat the test with a new
Medication Cassette
Reservoir. If the pump fails a second time, call Deltec’s Customer Service Department.
Example:
Predelivery Weight: 61.1 g Postdelivery Weight: – 41.6 g
Weight of Amount Delivered: 19.5 g
= 19.5 ml
Volume of Amount Delivered: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: = -0.025
x 100.00
Accuracy Error Percentage: = -2.5%
Equipment needed:
• 50- or 100- ml Medication Cassette
Reser-
voir
• 50- or 60- ml syringe
• CADD
®
Extension Set with Anti-siphon
Valv e
• A fluid collection device such as a burette or a class A 25 ml capacity graduated cylinder
• 40 ml of room temperature water
NOTE:
The test procedure calls for the use of a Medication Cassette CADD
®
Extension Set with Anti-siphon
Reservoir and a
Valve. An IV bag and CADD Admini­stration Set with integral or add-on anti­siphon valve can be substituted for the Medication Cassette CADD
®
Extension Set with Anti-siphon
Reservoir and
Valve.
Procedure:
1. Fill the 50- or 60-ml syringe with 40 ml of
water. Transfer the water into a Medication Cassette
Reservoir.
Volumetric Accuracy Testing
Description:
A predetermined amount of water is delivered into a collection device such as a burette or graduated cylinder. The amount of water delivered is compared to the amount that the pump should have delivered.
Nominal system accuracy is given in the technical specifications section for the pump. That is, under the test conditions described below, the accuracy of the pump and Medica­tion Cassette
Reservoir will be nominal with
2. Remove any air from the Medication
Cassette with the syringe. Attach the CADD
Reservoir by aspirating the air
®
sion Set with Anti-siphon Valve. Prime the tubing so it is filled with fluid to the end of the CADD
®
Extension Set luer lock connec-
tor.
3. Insert the end of the CADD
®
Extension Set
into the fluid collection device.
4. Attach the Medication Cassette
Reservoir to the pump. Program the Reservoir Volume to 20 ml. This is the intended delivery volume. Remove all clamps.
a 90% confidence level. The nominal test conditions are as follows: degassed water at 25 ± 5°C without back pressure.
Pre- Post- Weight of Intended Inaccuracy Accuracy Accuracy
Delivery Delivery Amount Delivery Volume Error Error Percentage
Weight Weight Delivered Volume
.
61.1 g 41.6 g 19.5 g = 19.5 ml 20 ml -0.5 ml -0.5 ml ÷ 20.0 ml = -0.025 x 100 = -2.5%
.
-0.025
Exten-
Table 15. Gravimetric percentage calculation
38
Testing Procedures - CADD-Legacy™ PLUS pump continued
5. Program a continuous rate of 125 ml/hr. Start the pump and deliver 20 ml.
6. When delivery is complete, record the volume of fluid delivered. This is the actual delivery.
7. Find the difference between the volume of the amount delivered and the intended delivery volume. This is the inaccuracy volume.
8. Divide the inaccuracy volume by the intended delivery volume and multiply by
100. This is the accuracy error percentage. (See table 16.)
9. If the accuracy error percentage is greater than ± 6%, repeat the test with a new Medication Cassette pump fails a second time, call Deltec’s Customer Service Department.
Reservoir. If the
Example:
Actual Delivery Volume: 19.5 ml Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = -0.5 ml
Inaccuracy Volume: -0.5 ml Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: = -0.025
x 100.00
Accuracy Error Percentage: = -2.5%
Intended Actual Inaccuracy Accuracy Accuracy
Delivery Delivery Volume Error Error Percentage
Volume Volume
20 ml 19.5 ml -0.5 ml -0.5 ml ÷ 20.0 ml =-0.025 -0.025 x 100 = -2.5%
Table 16. Volumetric percentage calculation
39
CADD-Legacy™ Pump Cleaning and Functional Testing Checklist
The following checklist is provided as a guide only to assist in establishing documentation of cleaning and func­tional testing for the CADD-Legacy
pump. If service is required, fill out this sheet and return it with the device.
Serial # _________________ Reference Number _________________________ Date _________________________
(Refer to the Technical Manual procedures.)
I. Cleaning Completed 䡺 Yes 䡺 No
II. Visual Inspection
LCD Cassette Sensor Accessory JackOcclusion Sensor Seals Keypad Air DetectorValves and Expulsor Indicator Light Pump HousingCassette Hinge Area Power Jack Battery DoorCassette Lock Battery Compartment
III. Mechanical Inspection
Keypad Cassette LockBattery Door
IV. Functional Inspection
Power-up Reservoir Volume Empty Alarm Dose Given/Attempted (PCA)Cassette Lock Starting/Stopping MG Given/GivenCassette Sensor Activation Timing Air DetectorLCD Dose Key (PCA)Motor/Gear Train Remote Dose Cord (PCA)
V. Downstream Occlusion Pressure Range Tests Test 1: Activations Before Alarm ______________
Test 2: High Pressure Alarm At _______________ psi
VI. Accuracy Testing
Volumetric Accuracy Test
Intended Actual Inaccuracy Accuracy Accuracy Delivery Delivery Volume Error Error
Volume Volume Percentage
ml ml ml %
Gravimetric Accuracy Test
Pre-Delivery Post-Delivery Amount Intended Inaccuracy Accuracy Accuracy
Weights Weight Delivered Delivery Volume Error Error
g g ml ml ml %
40
Volume
p
CADD, CADD-Legacy, Medication Cassette and Medication Cassette reservoir design are SIMS trademarks. The products described are covered by one or more of the following U.S. Patent Nos. 4,559,038; 4,565,542; 4,650,469; 5,364,242; 5,531,697; 5,538,399; 5,540,561; 5,564,915; 5,567,119; 5,567,136; 5,647,854; 5,695,473; Japanese Patent No. 2034590; European Patent No. 0182502; other patent(s) pending; foreign patents(s) pending. DURACELL® is a registered trademark of DURACELL, Inc. Energizer® is a registered trademark of EVEREADY Battery Company, Inc. SaniCloth® is a registered trademark of PDI. ©2000 SIMS Deltec, Inc. All rights reserved. Printed in U.S.A. 1/00 19876
T
A
43
Loading...