Smeg DRY50V User Manual

Contents
1. INSTRUCTIONS FOR SAFE AND PROPER USE_______________ 13
2. INSTALLATION OF THE STERILIZER _______________________ 15
3. DESCRIPTION OF CONTROLS ____________________________ 17
4. USE OF THE STERILIZER ________________________________ 19
5. CLEANING AND MAINTENANCE ___________________________ 20
6. EXTRAORDINARY MAINTENANCE _________________________ 22
IDENTIFICATION: HOT-AIR STERILIZER
MODEL: DRY STERIL 60V
YEAR OF CONSTRUCTION: .....................
INSTRUCTIONS FOR THE INSTALLER: these are intended for the qualified technician who must install the appliance, set it functioning and carry out an inspection test.
INSTRUCTIONS FOR THE USER: these contain user advice, description of the commands and the correct procedures for cleaning and maintenance of the appliance.
12
Introduction
1. INSTRUCTIONS FOR SAFE AND PROPER USE
THIS MANUAL IS AN INTEGRAL PART OF THE APPLIANCE AND THEREFORE MUST BE KEPT IN ITS ENTIRETY AND IN AN ACCESSIBLE PLACE FOR THE WHOLE WORKING LIFE OF THE STERILIZER. WE ADVISE READING THIS MANUAL AND ALL THE INSTRUCTIONS THEREIN BEFORE USING THE STERILIZER. INSTALLATION MUST BE CARRIED OUT BY QUALIFIED ENGINEERS IN ACCORDANCE WITH THE REGULATIONS IN FORCE. THIS STERILIZER IS INTENDED FOR COMMERCIAL USES AND CONFORMS TO CURRENT REGULATIONS IN FORCE. THIS APPLIANCE MUST ONLY BE USED BY SPECIFICALLY AUTHORISED STAFF TRAINED IN ITS CORRECT USE. THE APPLIANCE MUST ONLY BE USED FOR ITS INTENDED PURPOSE: THE STERILIZATION OF SURGICAL INSTRUMENTS IN A MEDICAL AND PARA-MEDICAL CONTEXT. THE APPLIANCE MUST NOT BE USED FOR TREATING CONTACT LENSES UNDER RULE 15 - ANNEX 9 - SUBSECTION 1. ALL OTHER USES ARE CONSIDERED IMPROPER.
THE MANUFACTURER DECLINES ALL RESPONSIBILITY FOR IMPROPER USE.
DO NOT LEAVE THE PACKING IN THE HOME ENVIRONMENT. SEPARATE THE VARIOUS WASTE MATERIALS AND TAKE THEM TO THE NEAREST SPECIAL GARBAGE COLLECTION CENTRE. THE STERILIZER IS SUPPLIED IN SUITABLE PACKAGING WHICH REQUIRES THE NORMAL PRECAUTIONS FOR TRANSPORTATION OF FRAGILE MATERIALS, AS INDICATED ON THE PACKAGING ITSELF. IF THE APPLIANCE IS RETURNED, USE THE ORIGINAL PACKAGING.
IT IS OBLIGATORY FOR THE ELECTRICAL SYSTEM TO BE GROUNDED ACCORDING TO THE METHODS REQUIRED BY SAFETY RULES.
WHEN LINKING UP TO MAINS BY PLUG AND SOCKET, MAKE SURE THAT BOTH ARE COMPATIBLE AND CONNECT BY MEANS OF A POWER CABLE COMPLYING WITH APPLICABLE REGULATIONS.
NEVER UNPLUG BY PULLING ON THE CABLE.
DO NOT OBSTRUCT VENTILATION OPENINGS AND HEAT DISPERSAL SLITS.
IMMEDIATELY AFTER INSTALLATION CARRY OUT A BRIEF INSPECTION TEST OF THE STERILIZER, FOLLOWING THE INSTRUCTIONS BELOW. SHOULD THE APPLIANCE NOT FUNCTION, DISCONNECT IT FROM THE SUPPLY AND CALL THE NEAREST TECHNICAL ASSISTANCE CENTRE. NEVER ATTEMPT TO REPAIR THE APPLIANCE.
13
ALWAYS CHECK THAT THE CONTROL KNOBS ARE IN THE 0 (OFF) POSITION WHEN YOU FINISH USING THE STERILIZER.
NEVER PUT INFLAMMABLE OBJECTS IN THE STERILIZER: THEY COULD BE ACCIDENTALLY LIGHTED AND CAUSE FIRES.
THE I.D. PLATE WITH TECHNICAL DATA, REGISTRATION NUMBER AND BRAND NAME IS LOCATED IN AN EXPOSED POSITION AT THE REAR OF THE STERILIZER.
THE PLATE MUST NOT BE REMOVED.
THE APPLIANCE IS INTENDED FOR COMMERCIAL USE BY APPROPRIATELY TRAINED PERSONNEL. CHILDREN SHOULD ALWAYS BE
KEPT OUT OF ITS REACH AND NEVER BE ALLOWED TO PLAY WITH IT.
THIS APPLIANCE IS MARKED ACCORDING TO THE EUROPEAN DIRECTIVE 2002/96/EC ON WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT (WEEE). THIS GUIDELINE IS THE FRAME OF A EUROPEAN-WIDE VALIDITY OF RETURN AND RECYCLING ON WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT.
BEFORE THE APPLIANCE IS PUT INTO OPERATION, ALL THE LABELS AND PROTECTIVE FILMS APPLIED INSIDE OR OUTSIDE MUST BE REMOVED.
Introduction
The manufacturer declines all responsibility for damage to persons
or things caused by non-observance of the above prescriptions or by interference with any part of the appliance or by the use of non-original spares.
THE WARRANTY COVERS PRODUCTION DEFECTS AND IS SUBJECT TO THE CONDITIONS OF THE CERTIFICATE WHICH ACCOMPANIES THE APPLIANCE.
14
Instructions for the installer
2. INSTALLATION OF THE STERILIZER
The sterilizer must be set up on a perfectly flat surface and left standing free. Make sure to allow enough room at the back for easy access to the terminal strip for connection of the power cable to mains. Installation must be carried out by a qualified technician in compliance with applicable regulations in force.
2.1 Electrical connection
Make sure that the power line voltage matches the specifications indicated on the rating plate located to the rear of the sterilizer. This rating plate must never be removed. The sterilizer is supplied complete with a H07RN-F three-core supply cable for operation from a 230V power supply.
The standard cable is of the three-core 3 x 1.5 mm2 H07RN-F type. If necessary the electrical supply cable should be replaced with an oil-resistant type (min. H05RN-F 3 x 1.5 mm2). The (yellow/green) earth cable on the end of the cable to be connected to the sterilizer should be at least 20 mm longer.
Make sure that plug and wall socket are of the same type and are in conformity with regulations in force. Also check that the power supply line is provided with adequate earthing. Avoid using adapters or shunters.
In case of permanent connection, ensure that power cable is equipped with an easily accessible omnipolar switch fitted in proximity to the apparatus and with a gap between contacts equal to or greater than 3 mm.
15