Skytron Stellar ST23 User manual

STELLAR SERIES • SURGICAL LIGHTS
FOCUSABLE LIGHTING SYSTEM WITH
EXTENDED DEPTH OF FIELD CONTROL
NON-STERILE
POSITIONING HANDLES
ST2319WC
360° RANGE OF MOVEMENT
(Pitch, Roll, Rotation)
110° VERTICAL
TRAVEL
SINGLE YOKE
BALANCE MECHANISM
LIGHT WEIGHT, EXTREMELY
DURABLE POLYMER LIGHTHEAD
POSITIONING/FOCUS
INTRODUCTION
SKYTRON’s Stellar Series Surgical Lighting provides advanced variable and fixed focus optics within separate lightheads. Exceptionally durable, lightweight polymer lightheads create the perfect balance of performance, maneuverability and dependability. The Vertical Segmented Reflector optics provide shadow controlled illumination and a generous field size. Typical applications include General, ASC ORs, Labor & Delivery, Plastics, OB/GYN and Emergency rooms.
TECHNICAL DATA
Electrical: 120V, 60 Hz, 1 phase Power Consumption: 400watts (1,360 BTU/hr) Amperage: Approximately 4.25 amps, fuse protected. Intensity: 226,800 lux (21,000 fc) Color Temperature: 4,000K Beam Temperature: Less than 25,000 microwatts per
cm2 as recommended by the I.E.S. handbook.
Current Leakage: Less than 30 microamps Bulb Type: Halogen 24V/50W. Fixture Weight: 151 lbs. Moment Load: 580 ft. lbs.
STERILE
WALL CONTROL WITH
EXCLUSIVE SKYTRON SOF-START
BULB PROTECTION
ACCESSORIES
Wall Mounted Intensity Control provides continuous
intensity adjustment according to surgeon's requirements and can be set for high or low line voltage to assure maximum performance. Required, priced separately.
ARCHITECTURAL SPECIFICATIONS
Surgical lighting system shall consist of a dual radial arm mounted 23" and 19" lightheads. Each lighthead to be equipped with a sterilizable center positioning handle and non-sterile positioning handles. Total system output is to be 226,800 lux (21,000 foot-candles) at a color temperature of 4,000K. System is to be designed Halogen fuve-bulb and Halogen three-bulb to operate at
18.5V to 20.5V, 50W. Bulb life is protected by SKYTRON's Sof-Start System. Wall Control provides intensity adjustment and ON/OFF control. The system shall be SKYTRON Model ST2319WC.
ST2319WC
ELEVATION VIEW
PLAN VIEW
45-1/2"
1155mm
2-3/4"
70mm
360°
360°
360°
fMinimum ceiling height: 8'-7". Consult your SKYTRON representative for lower ceiling heights and special applications.
110°
80"
2030mm
360°
360°
52-1/2"
1335mm
35-1/2"
900mm
38"
965mm
110°
360°
360°
3/4" Diameter
Conduit
Wall
Control
360°
360°
35-1/4"
895mm
360°
43"
1090mm
35-1/4"
895mm
360°
360°
78.34"
1990mm
35-1/4"
895mm
91"
2310mm
360°
REFER TO APPROPRIATE MOUNTING STRUCTURE GUIDELINE
• STANDARD MOUNTING STRUCTURE GUIDELINE
• SPECIFIC SEISMIC CALCULATIONS IF APPLICABLE
WALL CONTROL
3-1/2"
10"
1"
7-7/8"
5-7/8"
3-1/2"
4"
BACK
VIEW
7-5/8"
TOP VIEW
1"
1-1/16" Dia.
(2 places)
11-3/4" 10-1/4"
RECESS
MOUNT
FLANGE
12-7/16"
14"
6-7/8"
6mm Dia.
(4 places)
4mm Dia.
(4 places)
5000 36th Street SE • Grand Rapids, MI 49512
1.800.SKYTRON • 616.957.0500 • FAX: 616.957.5053
4/05