Skytron Nautilus User manual

NAUTILUS SERIES • SURGICAL LIGHTS
FULL 360° ROTATION
CAPABILITY FOR BOTH
LIGHTHEAD AND RADIAL
ARM ASSEMBLY
NON-STERILE
POSITIONING HANDLES
NT252525
110° VERTICAL
ADJUSTMENT
AUTOMATIC
VSRD ADVANCED
OPTICS AND REFLECTOR
DESIGN
RESERVE BULB
CHANGE OVER
WALL CONTROL
INTRODUCTION
SKYTRON’s Fixed Focus Nautilus Series with ADVANCED Vertical Segmented Reflector Design (VSRD) and swirl lens provides Superior Shadow Control and Extended Depth of Field. Typical applications include General, Orthopedic, Plastics, Micro Surgery, Cardiovascular, OB/GYN, Trauma and Urology.
TECHNICAL DATA
Electrical: 120v, 60 Hz, 1 phase Power Consumption: 510 watts (1,734 BTU/hr) Amperage: Approximately 3.9 amps, fuse
protected Intensity: 33,000 fc (360,000 lux)
11,000 fc per lighthead
Color Temperature: 4,000°K Beam Temperature: Less than 25,000 microwatts
per sq. centimeter as recommended by the I.E.S. handbook.
Current Leakage: Less than 100 microamps Bulb Type: Halogen 24v/170w Fixture Weight: 326 lbs. Moment Load: 1,560 ft.lbs.
HIGHLY VISIBLE
BULB CHANGE
INDICATOR
LIGHT
STERILE POSITIONING
AND PATTERN ADJUSTMENT
HANDLE
ACCESSORIES
Wall Mounted Intensity Control provides intensity
adjustment according to surgeon's requirements to assure maximum performance. Required, priced separately.
ARCHITECTURAL SPECIFICATIONS
Surgical lighting system shall consist of three radial arm mounted 25" diameter lightheads equipped with a sterilizable pattern adjustment and positioning handle as well as non-sterile positioning handles. Total system output is to be 33,000 foot­candles at a color temperature of 4,000°K. The system is to be designed with one primary bulb and one reserve bulb. An automatic bulb change-over system shall be provided to eliminate the need to change bulbs during surgery. Wall Control is to have an indicator/reset system to indicate reserve bulb operation. Bulbs are to be 24v, 170w Halogen type. System design is to provide cool, color­corrected light that is diffused evenly over the surgical site for optimum shadow reduction. System shall be SKYTRON Model NT252525.
NT252525
ELEVATION VIEW
53"
1350mm
2-3/4"
70mm
360°
360°
965mm
360°
38"
360°
45-1/2"
1155mm
360°
3/4" Diameter
Conduit
360°
PLAN VIEW
360°
360°
35-1/4"
895mm
91"
2310mm
360°
49"
58"
1250mm
1475mm
39"
34"
1000mm
865mm
360°
110°
Wall
Control
360°
35-1/4"
895mm
43"
1090mm
35-1/4"
895mm
360°
360°
360°
82-3/4"
2100mm
360°
110°
80"
2030mm
Minimum Ceiling Height: 9' 2". Consult your SKYTRON representative for lower ceiling heights and special applications.
REFER TO APPROPRIATE MOUNTING STRUCTURE GUIDELINE
• STANDARD MOUNTING STRUCTURE GUIDELINE
• SPECIFIC SEISMIC CALCULATIONS IF APPLICABLE
50-1/2"
1285mm
360°
98-1/2"
360°
2505mm
35-1/4"
895mm
WALL CONTROL
13-1/2"
1"
5"
6-3/8"
5"
4"
1"
10"
BACK
VIEW
11"
6mm Dia.
8-5/8"
(4 places)
5000 36th Street SE • Grand Rapids, MI 49512
1.800.SKYTRON • 616.957.0500 • FAX: 616.957.5053
TOP VIEW
1-1/16" Dia.
(2 places)
13-3/4"
12-3/16"
4mm Dia.
(4 places)
RECESS
MOUNT
FLANGE
15-3/4"
17"
4/05