3-3. Service .......................................................................................................................................13
REV 3/06
Although current at time of publication, SKYTRON's policy of continuous development makes this manual
subject to change without notice.
Page 2
EQUIPMENT LABELS AND SPECIFICATIONS
10A
TYPE B
EQUIPMENT
N
FUSE TYPE 10 AMP, FAST ACTING (FA)
INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED.
INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING.
IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY,
THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED
HOSPITAL GRADE OUTLET.
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS
TEMPERATURE: 15˚ - 30˚ C (60˚ -85˚ F)
HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
IPX4
V
A
HZ
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS.
INDICATES SPECIAL USER ATTENTION.
POWERED BY AC VOLTAGE
19.7"
3000 Series Imaging Table Specifications
TOP VIEW
65"
82.5"
Page 3
39" MAX
28" MIN
5.75"
8.25"
40.5"19"
SIDE VIEWEND VIEW
Electrical Specifications
Power requirements
Current Leakage
Power Cord
15 feet w/hospital grade connector
ENTELA CERTIFIED
TO UL2601-1,
CAN/CSA601.1, IEC 60601-2-46
120 VAC, 60Hz, 450 Watts
Less than 100 micro amps
Page 4
Prior to use, all personnel that may operate this
table must be instructed in the correct operational procedures. This table is designed for
use by trained and qualified personnel for human medical purposes only.
Initial use should not begin until after the users
have been instructed by the manufacturer's
representative.
A routine instructional program must be implemented by the facility for proper usage instructions for all personnel that may operate this table.
Certain accessories can be damaged when changing the position of the table top. Always look first to
see if a desired movement is going to interfere with
any accessories in use.
SPECIAL USER ATTENTION
Do not use worn or damaged accessories, they
represent an injury hazard.
Remove possible obstacles before lowering or
tilting the operating table
The operator has the ultimate responsibility of
preventing damage to the table and surrounding
equipment or possible injury to the patient or staff.
In general, common sense will dictate when there
is a potential hazard.
During table set-up and patient positioning, the
following items should be observed.
The table top is constructed of a radioluscent
composite material. Use caution when using accessories to protect the table top from damage
including nicks or scratches to maintain clear imaging capability.
The 500 pound lift and articulate capacity is
based on the patient weight load being evenly
distributed on the table top. Refer to figures 3-1,
3-2 and 3-3 for proper patient positioning.
The following precautions should be reviewed
by all personnel prior to operating the table.
WARNING
Indicates a possibility of personal injury.
Do not place objects on the base of the table, a
danger of damage exists during positioning.
Use caution when articulating the table top, pinch
hazards exist.
WARNING
Risk of electrical shock. Make sure
power cord is disconnected prior to
accessing fuses.
NOTE
Activating any function button will activate the brake system. Using the
TABLE UP function to set the brakes
provides a visual assurance that the
brakes are locked without altering the
table position, except when emergency
brake is released.
CAUTION
Indicates a possibility of damage to
equipment.
NOTE
Indicates important facts or helpful hints.
WARNING
DO NOT use the table in the presence
of FLAMMABLE GASES (Flammable
gases are not commonly used in the
U.S.).
WARNING
DO NOT unlock brakes when a patient
is on the table. An uneven patient
weight load may cause instability.
WARNING
Page 5
CAUTION
MAXIMUM END LOAD 180 POUNDS.
During transfer, DO NOT seat the pa-
tient at the distal end of the table. Table
stability may be compromised.
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg and lateral tilt positioning.
NOTE
•The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any hydraulic
function.
•If the Emergency Brake Release Valve
has been operated, the BRAKE UNLOCK button on the pendant control will
have to be pressed before brakes will
lock again.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
Lithotomy positioning is NOT ADVISED
for the 3000 series tables.
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor
WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table
or an alternate approved replacement.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
Page 6
SECTION I INTRODUCTION
Figure 1-1. Elite 3000
1-1. General
SKYTRON’s Elite 3000 Series Imaging Tables are
electro-hydraulically operated. See figure 1-1.
The electro-hydraulic positioning functions operated by the hand-held, pendant control unit are:
Trendelenburg, lateral tilt, elevation, return to level,
and the floor lock/brake system.
A manual control is provided for emergency brake
release.
1-2. Power Requirements
The Elite 3000 Surgical Table requires a 120VAC,
60 Hz electrical power supply. The table is equipped
with a 15 foot long power cord with a standard three
prong, hospital grade plug. The electrical protection fuses (2) are located behind a cover plate in the
electrical enclosure on the front edge of the base.
See figure 1-2. The main power ON/OFF switch is
located on the electrical enclosure.
WARNING
Risk of electrical shock. Make sure
power cord is disconnected prior to
accessing fuses.
Page 7
*
PRESS TO
LOCK BRAKES
RETURN
ELEVATION
LATERAL TILT
TRENDELENBURG
POWER
BRAKE
UNLOCK
*
PRESS TO
LOCK BRAKES
RETURN
ELEVATION
LATERAL TILT
TRENDELENBURG
POWER
BRAKE
UNLOCK
ELITE / 3000
CAUTION
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
MOVING PARTS.
*
PRESS TO
LOCK BRAKES
RETURN
ELEVATION
LATERAL TIL T
TRENDELENBURG
POWER
BRAKE
UNLOCK
TRENDELENBURG
LA TERAL TIL T
POWER
INDICATOR
TABLE UP
(BRAKE LOCK)
RETURN
TO LEVEL
TABLE DOWN
BRAKE UNLOCK
ELECTRICAL
ENCLOSURE
POWER
CORD
MAIN POWER
COVER
SWITCH
PLATE
(2) FUSES
10 Amp FA
Figure 1-2. Power Switch and Fuse Location
1-3. Pendant Control Unit
The hand-held pendant control unit (figure 1-3) has
a non-slip rubber cover which assures a positive
grip during use. A spring clip hanger located on the
back of the control allows it to be stored on the table
bracket.
SPRING
CLIP
HANGER
ELITE / 3000
ELITE / 3000
POWER
INDICATOR
1-4. Floor Lock/Brake System
The floor lock/brake system consists of four selfleveling, hydraulic brake cylinders which raise and
support the table base off from the casters. Press
the ELEVATION-UP button on the pendant control
to set the table’s brakes. An electronic timer in the
brake circuit will activate the brake system until the
brakes are completely set, approximately 8-10
seconds.
NOTE
Activating any function button will activate the brake system. Using the
TABLE UP function to set the brakes
provides a visual assurance that the
brakes are locked without altering the
table position, except when emergency
brake is released.
To unlock the brakes, press the BRAKE UNLOCK
button and release. The brakes will retract automatically in approximately 7-8 seconds.
FUNCTION
BUTTONS
CAUTION
CAUTION
HANG PENDANT CONTROL
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
ON SIDE RAIL WHEN NOT IN
MOVING PARTS.
MOVING PARTS.
USE. KEEP CORD CLEAR OF
USE. KEEP CORD CLEAR OF
Figure 1-3. Pendant Control Unit
The function buttons are identified with symbols
and descriptions for all functions. See figure 1-4.
The Trendelenburg and table-up buttons are red,
the remaining buttons are all green.
Figure 1-4. Function Push Buttons
Page 8
ELITE / 3000
ELEVATION
LATERAL TILT
TRENDELENBURG
POWER
ELEVATION
UP
SECTION II OPERATION
2-1. Electrical Power
a. Check to be sure the power cord is plugged
into 120VAC outlet.
WARNING
Prior to operating the table, observe all
table caution labels and review the
SPECIAL USER ATTENTION section
in the front of this manual.
WARNING
DO NOT use the table in the presence
of FLAMMABLE GASES (Flammable
gases are not commonly used in the
U.S.)
b. Depress Main Power ON/OFF switch on the
electrical enclosure. See figure 2-1. The green
Power ON indicator light on the pendant control unit
should now be illuminated.
2-2. Positioning Functions
The hand-held pendant control unit (figure 2-2)
activates the following table functions:
TRENDELENBURG
LA TERAL TIL T
TABLE UP
(BRAKE LOCK)
RETURN
TO LEVEL
ELITE / 3000
POWER
TRENDELENBURG
LATERAL TILT
ELEVATION
BRAKE
RETURN
UNLOCK
PRESS TO
*
LOCK BRAKES
CAUTION
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
MOVING PARTS.
POWER
INDICATOR
TABLE DOWN
BRAKE UNLOCK
Figure 2-2. Pendant Control Unit
POWER
CORD
MAIN
POWER
SWITCH
Figure 2-1. Main Power Switch
a. Floor Lock/Brake System. To activate the
brakes without affecting table positioning, press
the ELEVATION-UP button. See figure 2-3. The
elevation cylinder will not function until the brakes
are completely extended.
Figure 2-3. Brake System Activation
Press the BRAKE-UNLOCK button on the pendant
20˚
20˚
ELITE / 3000
CAUTION
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
MOVING PARTS.
*
PRESS TO
LOCK BRAKES
RETURN
ELEVATION
LATERAL TILT
TRENDELENBURG
POWER
BRAKE
UNLOCK
LA TERAL TIL T
ELITE / 3000
CAUTION
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
MOVING PARTS.
*
PRESS TO
LOCK BRAKES
RETURN
ELEVATION
LATERAL TILT
TRENDELENBURG
POWER
BRAKE
UNLOCK
TABLE UP
TABLE DOWN
28"
39"
control to release the four self-leveling brake feet in
order to move the table. See figure 2-3. The brake
delay circuit automatically retracts the brake system with just one press of the BRAKE-UNLOCK
button. It takes approximately 7-8 seconds to
totally release the system.
WARNING
DO NOT unlock brakes wh
en a patient is on the table. An uneven
patient weight load may cause instability.
b. Trendelenburg. To achieve Trendelenburg
positioning, press the appropriate
TRENDELENBURG button (figure 2-4). Trendelenburg positioning of up to 20° may be obtained.
Page 9
c. Lateral Tilt. To achieve lateral tilt right
positioning, press the appropriate LATERAL-TILT
button (figure 2-5). Tilt of up to 20° may be obtained.
Figure 2-5. Lateral Tilt Positioning
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg and lateral tilt positioning.
ELITE / 3000
POWER
TRENDELENBURG
LATERAL TILT
ELEVATION
BRAKE
UNLOCK
PRESS TO
*
LOCK BRAKES
CAUTION
MOVING PARTS.
TRENDELENBURG
RETURN
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
d. Elevation. To raise table top, press the UP
button (figure 2-6.). The table will lift a patient weight
of 500 pounds up to a maximum height of 39". To
lower the table top, press the DOWN button. The
table top will go down to a minimum height of 28"
(minus pad).
20˚
20˚
Figure 2-4. Trendelenburg Positioning
Figure 2-6. Elevation Function
Page 10
e. Return To Level. To return the table top to a level
position, press the RETURN button (figure 2-7).
NOTE
ELITE / 3000
POWER
TRENDELENBURG
LATERAL TILT
ELEVATION
RETURN
TO LEVEL
BRAKE
RETURN
UNLOCK
PRESS TO
*
LOCK BRAKES
CAUTION
HANG PENDANT CONTROL
ON SIDE RAIL WHEN NOT IN
USE. KEEP CORD CLEAR OF
MOVING PARTS.
Figure 2-7. Return To Level
2-3. Emergency Brake Release
In case of a power failure or an electrical problem
within the table, the emergency brake release
system can be used to move the table. The control
lever for this function is located on the side of the
table base and is identified by an EMERGENCY
BRAKE RELEASE label. Turn the lever counterclockwise to release the brakes. See figure 2-8.
•The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any hydraulic
function.
•If the Emergency Brake Release Valve
has been operated, the BRAKE UNLOCK button on the pendant control will
have to be pressed before brakes will
lock again.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
BRAKE (4)
EMERGENCY
BRAKE RELEASE
Figure 2-8. Emergency Brake Release
WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table
or an alternate approved replacement.
2-4. Positioning
During table set-up and patient positioning, the
following items should be observed.
The table top is constructed of a radioluscent
composite material. Use caution when using accessories to protect the table top from damage
including nicks or scratches to maintain clear imaging capability.
The 500 pound lift and articulate capacity is
based on the patient weight load being evenly
distributed on the table top and the patient's head
over the elevation column. The weight capacity
with the patient's head away from the elevation
column is 400 pounds. Refer to figures 3-1,
3-2 and 3-3 for proper patient positioning.
RESTRAINT
STRAP
6" ACCESSORY
Page 11
SINGLE
CARBON FIBER
ARMBOARD (2)
ELEVATION
COLUMN
RAIL CLAMP
CAUTION
MAXIMUM END LOAD 180 POUNDS.
During transfer, DO NOT seat the pa-
tient at the distal end of the table. Table
stability may be compromised.
Lithotomy positioning is NOT ADVISED
for the 3000 series tables.
STANDARD
ARMBOARD (2)
RESTRAINT
STRAP
500 POUND CAPACITY
Figure 3-2.
DUAL
CARBON FIBER
ARMBOARD
RESTRAINT
STRAP
6" ACCESSORY
RAIL CLAMP
Figure 3-1.
ELEVATION
COLUMN
400 POUND CAPACITY
6" ACCESSORY
RAIL CLAMP
400 POUND CAPACITY
Figure 3-3.
ELEVATION
COLUMN
Page 12
SECTION III MAINTENANCE
3-1. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the serviceability and proper operation of your SKYTRON
Surgical Table.
a. During normal cleaning, a general visual examination should be made checking for leaks,
loose bolts or parts and cracked or chipped paint.
Any necessary repairs should be made.
b. Semi-annually the following checks and services should be performed:
1. Check all hydraulic fittings, mini-valves
and slave cylinders for proper operation
and any signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3. Pressure check (with a gauge) the pressure relief valve.
4. Check all mechanical adjustments and
adjust as necessary.
5. Check hydraulic fluid level.
3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
Remove major contaminants from the table with
disposable materials following appropriate biohazard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
6. Lubricate the elevation slider assembly.
7. Inspect table top for nicks, scratches or
dents.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
3-3. Service
Page 13
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
Repeat cleaning procedure for all table surfaces
including the top, sides, elevation column, base and
all accessories.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
When the cleaning procedure is complete, replace
all pads and accessories as applicable.
Remove pendant control from table bracket and
apply cleaning solution to the pendant control and
cord.
Use a clean cloth dampened with water to remove
cleaning solution.
Table maintenance can be performed by trained
maintenance personnel using SKYTRON authorized replacement parts and service techniques.
Service instructions and parts are available from
SKYTRON.
To obtain service instructions, replacement parts,
factory service or preventive maintenance contracts, contact the SKYTRON representative listed
below.
Or contact:
SKYTRON
5000 - 36th Street S.E.
Grand Rapids, MI 49512
1-800-SKYTRON (1-800-759-8766)
FAX: (616) 957-5053
Use another clean damp cloth to remove any remaining residue.
Install pendant control on bracket for storage when
cleaning procedure is complete.
5000 36th Street S.E., Grand Rapids, MI 49512
1-800-SKYTRON or 1-616-957-0500 • FAX 1-616-957-5053
ELITE SERIES SURGICAL TABLES
MAINTENANCE MANUAL
MODEL ELITE 3000 and 3001
Page 37
Table of Contents
SECTION I HYDRAULIC SYSTEM......................................................................................................... 1
WRENCHES 1.5mm-8mm
ADJUSTABLE CRESCENT WRENCH
DIGITAL VOLTMETER, TRUE RMS
METRIC OPEN END WRENCHES 7mm-18mm
LEVEL (CARPENTERS)
BASIC RECOMMENDED MAINTENANCE PROCEDURES
The basic items noted below should be inspected at a minimal interval period of 6 months (dependant on
usage). For optimal usage, safety and longevity of the product, have it serviced only by an authorized
Skytron representative with authentic Skytron replacement parts.
• Check brake pads for wear and inspect brake cylinders for proper operation.
• Check Carbon Fiber Top for Damage
Only facility-authorized SKYTRON trained, maintenance personnel should troubleshoot the
SKYTRON Surgical Table. Trouble shooting by unauthorized personnel could result in
personal injury or equipment damage.
How to contact us:
Skytron
5000 36th St. SE, Grand Rapids, MI 49512
PH: 1-800-759-8766 (SKY-TRON)
FAX: 616-957-5053
Page 33
TYPE B
EQUIPMENT
EQUIPMENT LABELS AND SPECIFICATIONS
INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED.
INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING.
IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY,
THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED
HOSPITAL GRADE OUTLET.
N
IPX4
V
A
HZ
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS
TEMPERATURE: 15˚ - 30˚ C (60˚ -85˚ F)
HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS.
INDICATES SPECIAL USER ATTENTION.
Page 34
POWERED BY AC VOLTAGE
ENTELA CERTIFIED
TO UL2601-1
CAN/CSA601.1, IEC 60601-2-46
3001 Series General Purpose Surgical Table Specifications
TOP VIEW
30"
13.8"
42" MAX
32" MIN
40.5"
82.5"
8.25"
SIDE VIEW
65"
19.7"
20˚
21.875"
20˚20˚
20˚
19"
END VIEW
3000 Series General Purpose Surgical Table Specifications
19.7"
39" MAX
28" MIN
TOP VIEW
82.5"
40.5"19"
SIDE VIEW
65"
20˚
20˚20˚
20˚
8.25"
END VIEW
Page 35
SECTION VI ELECTRICAL SYSTEM ADJUSTMENTS
6-1. Relay Box Adjustments
The Relay Box contains variable resistors for adjusting the operating timers for the BRAKE SET
and BRAKE UNLOCK functions. These timers are
set at the factory and usually never need adjustment. If an adjustment is necessary, remove the
relay box cover and use the following procedures.
See figure 6-1.
a. Brake Release Timer
The Brake Release Timer is set for about 7 seconds and is controlled by the variable resistor VR1
on the relay box circuit board. Turn the adjuster
clockwise to increase the operating time. Counterclockwise to decrease the operating time.
b. Brake Set Timer
The Brake Set Timer is set for about 7 seconds
and is controlled by the variable resistor VR2 on
the relay box circuit board. Turn the adjuster clockwise to increase the operating time. Counterclockwise to decrease the operating time.
VR-1VR-2VR-3VR-4
CN-4
CN-2B
CN-3
CN-2ACN-1
Figure 6-1. Relay Box Adjustments
Page 29
SECTION I HYDRAULIC SYSTEM
1-1. General
Electro- Hydraulic System
The hydraulic system (with the exception of the
hydraulic cylinders and hoses) is contained within
the base of the table. The hydraulic valves and
pump are electrically controlled by the use of a
hand- held push button pendant control. The power
requirements for the table are 120 VAC, 5 amp,
60 Hz.
The table contains the following components. Refer to the block diagram ( figure 1- 1) for relationship.
a. Oil Reservoir - Main oil supply. Approximately
two quarts.
LA TERAL TIL T
LATERAL SLIDE
67
45
b. Motor/Pump Assembly - A positive displace-
ment gear type pump provides the necessary oil
pressure and volume.
c. Pressure Relief Valve - Provides an alternate
oil path when the hydraulic cylinders reach the end
of their stroke.
d. Electro/Hydraulic Mini-Valv e Assemblies -These
direct the fluid to the appropriate hydraulic cylinders.
e. Hydraulic Lines, Fittings, Connections - They
provide a path for the hydraulic oil.
f. Hydraulic Cylinders - They convert the hydraulic
fluid pressure and volume into mechanical motion.
LONGITUDINAL
SLIDE
889
ELEVATION
1
TREND
2
3
9
2
1
4
3
ELEVTREND
6
5
2
1
8
PLUMBING
7
TERMINAL
8
9
LONG
SLIDE
12
7
3
4
TILT
9
6
5
LAT
SLIDE
12
PRESSURE
RELIEF V AL VE
BRAKES
EMERGENCY
BRAKE RELEASE
DR
MINI-VALVES
MOTOR/PUMP
ASSEMBLY
Figure 1-1. Hydraulic Block Diagram - Model 3001
M
Page 1
1-2. Component Operation
a. Motor/Pump Operation
The motor/pump assembly is a gear type pump
that provides the oil pressure and volume for the
entire hydraulic system. The pump has an inlet
side and an outlet side. The inlet side is connected
to the reservoir which provides the oil supply . The
reservoir has a very fine mesh screen strainer
which prevents foreign material from entering the
oil system.
The output line of the pump is connected to the
main oil galley which is internal and common to all
the hydraulic mini-valves and pressure relief valve.
Also, common to the hydraulic mini-valves and
pressure relief valve is an oil galley that internally
connects to the oil reservoir to provide a return
path for the hydraulic oil. See figure 1-2.
Turning the adjustment nut clockwise increases
the amount of oil pressure required to open the
valve, and turning it counterclockwise decreases
the amount of oil pressure. (See adjustment section for specification.)
PRESSURE RELIEF
ADJUSTMENT NUT
PRESSURE
GAUGE
SPRING LOADED
PLUNGER
MAIN RETURN
OIL GALLEY
(INTERNAL)
MINI-VALVES
MOTOR/PUMP
ASSEMBLY
PUMP
OUTLET
PUMP INLET
MAIN PRESSURE
OIL GALLEY
(INTERNAL)
PRESSURE
RELIEF V AL VE
OIL
RESERVOIR
Figure 1-2.
b. Pressure Relief Valve
This device provides an alternate oil path when
the hydraulic cylinders reach the end of their stroke
and the pump continues to run. If this path were
not provided, the pump motor would stall because
the oil cannot be compressed. The pressure relief
valve is directly connected to the mini-valve bodies and shares both the common internal main
pressure oil galley, and the return oil galley, that
internally connect to the reservoir. See figure 1-3.
Figure 1-3. Pressure Relief Valve Not
Functioning
PRESSURE RELIEF
ADJUSTMENT NUT
PRESSURE
GAUGE
SPRING LOADED
PLUNGER
The main component of the valve is an adjustab le
spring loaded plunger that is pushed off from its
seat by the oil pressure. The oil then flows back
into the reservoir. See figure 1-4.
Page 2
Figure 1-4. Pressure Relief Valve Functioning
c. Mini-Valves
The operation of the mini-valves is identical for all
table functions except the ele v ation and br ak e circuits. These two hydraulic circuits use a 3-way
(single check valve) type mini-v alve. All other functions use a 4-way (dual check valve) type minivalve.
Either type mini-valve is controlled by two pushing
type, electrically operated solenoids. The solenoids
push the spool valve (located in the lower portion
of the valve) one way or the other. This motion
opens the main supply galley (which has pump
pressure) allowing the oil to flow through the various parts of the mini-valve to the function. The
spool valve also opens an oil return circuit which
allows the oil to return to the oil reservoir.
The main components of the mini-valve and their
functions are listed below:
1. Spool Valve - Opens the main oil galley
(pump pressure) to either mini-valve outlet depending on which direction the spool valve is pushed.
Also it provides a return path for the oil returning
back into the reservoir.
2. Pilot Plunger - There are two plungers in a
four-way mini-valve (one in a 3-way mini-valve),
one under each check valv e. The purpose of the
pilot plungers is to mechanically open the return
check valve allowing the oil to return back into the
reservoir.
The speed controls are always located in the return oil circuit. This prevents uncontrolled movement of the piston in the slave cylinder due to one
side of the piston being loaded with hydraulic pressure and the other side having no load.
Also, by using this control method, it doesn’t matter
what size cylinder and piston is used because the
speed can be controlled by restricting the return oil.
If the pump puts out more volume to a certain slave
cylinder than the speed control is allowing to go back
to the reservoir, the pressure relief valve provides
an alternate path for the pump oil.
d. Mini-Valve in Neutral Position
(No fluid flow) See figure 1-5.
1. Spool V alv e Centered - This closes off both
oil pressure and oil return galleys.
2. Pilot Plungers Both Closed -The pilot plungers control the opening of the check valves. If they
are closed, the check valves must be closed.
3. Check Valves - Both check valves are
closed trapping the oil in the cylinder and oil lines.
4. Speed Adjustment - When the mini-valve
is in the neutral position, the speed adjustment
does not affect anything because there is not any
oil flow.
3. Check Valve - Two are provided in each
four-way mini-valve to seal the oil in the cylinders
and oil lines and prevent any movement of the
table. One check valv e is provided in a 3-wa y minivalve.
4. Speed Adjustments - There are two speed
adjustments in each mini-valve. They are needle
valve type controls which restrict the volume of oil
returning back into the reservoir, thereby controlling the speed of the table surface movement. A
3-way mini-valve has only one speed adjustment.
Figure 1-5. Mini-Valve in Neutral Position
Page 3
e. Mini-Valve Right Port Activated
(See figure 1-6)
Slave Cylinder Piston Moves to Left
Right Mini-Valve Port is Supply Line
Left Mini-Valve Port is Return Line
f. Mini-Valve Left Port Activated
(See figure 1-7.)
Slave Cylinder Piston Moves to Right
Left Mini-Valve Port is Supply Line
Right Mini-Valve Port is Return Line
INLETOUTLET
OUTLET
INLET
Figure 1-6. Mini-Valve Right Port Activated Figure 1-7. Mini-Valve Left Port Activated
1. Spool Valve - Pushed to the left by electric
solenoid. This opens the internal oil pressure galley allowing the fluid to go through the check valve
and on to the cylinder. Also , the spool valve opens
the oil return line providing an oil path through the
internal oil galley back to the reservoir.
2. Pilot Plunger V alv e - Left pilot plunger valv e
is pushed up by the incoming oil pressure mechanically opening the check valve located above it in
the return circuit. This action allows the oil from
the left side of the slave cylinder to go back into
the reservoir. The right pilot plunger valve is not
affected in this operation mode.
1. Spool Valve-Pushed to the right b y electric
solenoid. This opens the internal oil pressure galley allowing the fluid to go through the check valve
and on to the cylinder. Also , the spool valve opens
the oil return line providing an oil path through the
internal oil galley back to the reservoir.
2. Pilot Plunger V alv e - Right pilot plunger valv e
is pushed up by the incoming oil pressure mechanically opening the check valve located above it in
the return circuit. This action allows the oil from
the right side of the slave cylinder to go back into
the reservoir. The left pilot plunger valve is not
affected in this operation mode.
3. Check Valves - Both check valves are
opened in this operation mode. The right check
valve is pushed open by the oil pressure created
by the pump. The oil then continues to go through
the lines and pushes the slave cylinder piston to
the left. At the same time, the left check valve is
held open mechanically by the pilot plunger providing a return path for the oil through the minivalve back to the reservoir.
4. Speed Adjustment - The right speed control (output side) does not have any effect in this
operation mode because the oil is routed around
the speed adjustment through a by-pass valv e and
then to the output port. The left speed adjustment
controls the speed of the table function by restricting the amount of oil going back into the reservoir.
Page 4
3. Check Valves - Both check valves are
opened in this operation mode. The left valve is
pushed open by the oil pressure created by the
pump. The oil then continues to go through the
lines and pushes the slave cylinder piston to the
right. At the same time, the right check valve is
held open mechanically by the pilot plunger providing a return path for the oil through the minivalve back to the reservoir.
4. Speed Adjustment - The left speed control
(output side) does not have any effect in this operation mode because the oil is routed around the
speed adjustment through a by-pass valve and
then to the output port. The right speed adjustment
controls the speed of the table function by restricting the amount of oil going back to the reservoir.
g. Hydraulic Cylinders (Slave Cylinders)
PISTON
PISTON
RAM
RAM
HYDRAULIC
PRESSURE
CYLINDER
HOUSING
SUPPORT
COLUMN
PIVOT
PIN
CONNECTING
ROD PIN
PISTON AND
CONNECTING RODS
There are several different types of hydraulic cylinders used in the table that activate the control
functions. With the exception of the elevation and
brake cylinders, all oper ate basically the same way .
The control functions are:
Reverse Trendelenburg/Trendelenburg
Lateral Tilt
Elevation
Brake Cylinders
Longitudinal Slide Function - 3001 Only
Lateral Slide Function - 3001 Only
1. Trendelenburg Cylinder - The double
action cylinders are closed at one end and have a
movable piston with hydraulic fluid on both sides.
Connected to this piston is a ram or shaft that exits
out of the other end of the cylinder. Through the use
of a ball and socket arrangement this ram is
connected to a movable table surface.
The movable surface can be moved one way or
the other by pumping hydraulic fluid into the cylinder on either side of the piston. Obviously, if oil is
pumped into one side of the cylinder, a return path
must be provided for the oil on the other side. See
figure 1-8.
TRENDELENBURG
CYLINDER
The pistons and connection rods are attached to a
non-movable surface. Therefore, when hydraulic
fluid is pumped into one side, the cylinder housing
itself moves around its pivot pin causing the table
top to tilt to one side. To tilt the table top in the
opposite direction, hydraulic fluid is pumped into
the other side of the cylinder assembly.
Figure 1-9. Lateral Tilt Cylinder Assembly
3. Longitudinal Slide Cylinder Assembly - 3001
Only - The longitudinal slide system utilizes two
double action cylinders working in tandem. The
two cylinder housings are connected together facing opposite directions. The ram of one cylinder is
connected to the table top frame and the other one
is attached to the elevation column. Hydraulic fluid
is pumped into the same end of both cylinders at
the same time. The two cylinder system allows
the 20" of top slide to be performed by a short
cylinder system. See figure 1-10.
with two pistons complete with connecting rods.
The connecting rods are attached to a stationary
pin in the housing on top of the main support column. The cylinder housing attaches to the table
frame and is attached to the support column with
der assembly consists of a dual cylinder housing
a pivot pin. See figure 1-9.
RAM
PISTON
O-RING
Figure 1-8. Trendelenburg Cylinder
2. Lateral Tilt Assembly - The lateral tilt cylin-
Figure 1-10. Longitudinal Slide Cylinders
Page 5
4. Later al Slide Function Cylinder - 3001 Only .
This cylinder is also a double action cylinder. It
differs from the Trendelenburg cylinder in that the
piston is connected to rams which exit from each
end of the cylinder tube. The ends of the rams
connect to each side of the table top . The cylinder
tube is connected to the elevation column. When
hydraulic fluid is pumped into one side of the cylinder, the entire tab le top assembly will slide laterally. See figure 1-11.
RAM
O-RING
PISTON
Figure 1-11. Lateral Slide Cylinder
5. Elevation Cylinder - This single action cylinder does not have hydraulic fluid on both sides
of the piston. It depends on the w eight of the tab le
top assembly to lower it.
The cylinder is set in the center of the elevation
main column. The piston and piston rod are elevated by the driven force of the oil pressure.
When lowering, the oil that is accumulated in the
cylinder is returned to the oil reservoir through the
mini-valve due to the table top weight.
A slider support assembly is used to support the
weight of the upper table section. A stainless steel
shroud covers the flexible hydraulic hoses and
slider. See figure 1-12.
Figure 1-12. Elevation Cylinder Assembly
6. Brake Cylinders - The brake cylinders are
single action type similar to the elevation cylinder.
The movable piston ram is connected to a brake
pad. See figure 1-13. Oil pumped into the top of
the cylinder pushes the piston down raising the
table base off its casters. An internal return spring
pushes the piston up to return the oil through the
mini-valve to the reservoir.
PISTON
OIL LINE
RETURN
SPRING
Page 6
BRAKE
PAD
Figure 1-13. Single Action Brake Cylinder
h. Elevation Cylinder Return Circuit
i. Brake System
A three-way (single check valve type) mini-valve
controls both the elevation and return circuits. The
elevation circuit operation within the mini-valve is
identical to the operation of the four-way valves
previously described (inlet pressure opens the
check valve allowing the oil to enter the cylinder).
In the return position, inlet pressure pushes the
pilot plunger up and opens the return check valve.
See figure 1-16. The open check valve allows a
path for the oil in the ele vation cylinder to return to
the reservoir. When the pilot plunger valve is
opened, the continuing pump pressure opens the
pressure relief valve which provides a return oil
path to the reservoir.
The mini-valve used in the elevation circuit contains only one check valv e (all four-way mini-v alves
use two check valv es). The check v alve is used to
trap the oil in the elevation cylinder thereby supporting the table top. When the top is being lowered the check valve is mechanically held open by
the pilot plunger through pump pressure.
The brake system consists of the following components: (figure 1-17)
1. Single action slave cylinders (4 each).
2. 3-way (single check valve type) mini-valve.
3. Manually controlled emergency brake re-
lease.
4. Plumbing terminal, flexible hoses, copper
lines and "O" rings.
5. Portions of the electrical system.
BRAKES
INLET
TO RESERVOIR
THROUGH
PRESSURE
RELIEF V AL VE
RETURN TO
RESERVOIR
Figure 1-16. Elevation Return Circuit
EMERGENCY
BRAKE RELEASE
PLUMBING
TERMINAL
MOTOR/PUMP
ASSEMBLY
PRESSURE
RELIEF V AL VE
DR
M
Figure 1-17. Brake System Block Diagram
Each corner of the cast-iron table base has a hydraulic brake cylinder. These single action cylinders are hydraulically connected in parallel to the
mini-valve and all f our are activ ated together. It is
normal for one corner of the table to raise before
the others due to the weight distribution of the table.
An electronic timer in the relay box is activated
when any function on the pendant control is
pushed momentarily . The pump/motor and brake
system mini-valve are activated and the brake cylinders are completely set. The electronic timer
runs for approx. 8-10 seconds.
Page 7
The brakes are released by pushing the BRAKE
UNLOCK button momentarily. An electronic timer
in the relay box activates the brake function hydraulic mini-valve and pump/motor.
1-3. Hydraulic Adjustments
a. Fluid Level.
When activated, the return hydraulic circuit operates similar to the elevation cylinder return circuit.
Return springs inside the single action brake cylinders retract the brake pads and provide the pressure to return the hydraulic oil back to the reservoir. The electronic timer operates the return circuit for approximately 8-10 seconds.
j. Emergency Brake Release
The emergency brake release is simply a manually operated bypass valve connected in parallel
to the brake cylinders and the oil reservoir. See
figure 1-18. When the valve is opened (turned
counter-clockwise) a return circuit for the brake
hydraulic fluid is opened. The return springs force
the pistons up pushing the hydraulic oil back into
the reservoir and retracting the brake pads.
The fluid level should be approximately 1/2" below
the filler hole or gasket surface. If additional fluid
is needed, remove the filler vent cap with a phillips
screwdriver and add fluid through this opening
using a funnel. See figure 1-19.
NOTE
The elevation cylinder should be completely down, the brakes released and
all the other control functions in their
neutral position when checking oil level.
FILLER VENT
FILLER
CAP
RESERVOIR
OPENING
EMERGENCY BRAKE
RELEASE LEVER
Figure 1-18.
NOTE
•The emergency brake release valve
must be tightened securely when not in
use.
•If the emergency brake release valve
has been operated, the BRAKE UNLOCK button on the pendant control
may have to be pressed before brakes
will lock again.
If the emergency brake release valve is open or
loose, the brakes will release slowly- depending
on how loose the valve is, this could take anywhere from a few minutes to several hours.
Figure 1-19.
The type of oil that should be used is Mobil DTE
#25 or equivalent. This is a very high quality hydraulic oil. The table requires approximately two
quarts of oil to operate properly. Exercise caution
when determining equivalence to avoid damage
to the hydraulic system.
b. Bleeding The Hydraulic System
To purge the air from the hydraulic system, operate each function back and forth at least two or
three times.
NOTE
Whenever a hydraulic line or component is replaced, bleed the air out of the
lines using the pump pressure before
making the final connection. Then operate the function until it stalls in both
directions.
Page 8
Loading...
+ 105 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.