Skytron 6701 hercules Operator's Manual

SURGICAL TABLE
OPERATORS MANUAL
TEC-D-0010 REV2 6/11
6701 HERCULES
Page 1
Title Page
EQUIPMENT LABELS ..........................................................................................................................2
6701 Hercules General Purpose Surgical Table Specifications ............................................................3
SPECIAL USER ATTENTION ..............................................................................................................4
SECTION I INTRODUCTION ...............................................................................................................10
1-1. General .................................................................................................................................10
1-2. Power Requirements ............................................................................................................10
1-3. Pendant Control Unit ...........................................................................................................11
1-4. Floor Lock/Brake System ..................................................................................................... 11
SECTION II OPERATION .....................................................................................................................12
2-1. Electrical Power ....................................................................................................................12
2-2. AC 120V Operation .............................................................................................................. 12
2-3. Battery Operation ................................................................................................................. 13
2-4. Automatic Shut-Off ...............................................................................................................13
2-5. Charging the Battery ............................................................................................................14
2-6. Positioning Functions .......................................................................................................... 15
a. Floor Lock/Brake system..................................................................................................15
b. Trendelenburg ..................................................................................................................16
c. Lateral Tilt ........................................................................................................................ 16
d. Back Section ....................................................................................................................17
e. Elevation .......................................................................................................................... 17
f. Leg Section ......................................................................................................................18
g. Flex Positioning ................................................................................................................18
h. Kidney Lift ........................................................................................................................ 19
i. Return To Level ................................................................................................................ 19
j. Beach Chair ..................................................................................................................... 19
2-7. Emergency Back-up controls ...............................................................................................20
2-8. Emergency Brake Release ...................................................................................................20
2-9. Head Section ........................................................................................................................ 21
2-10. Leg and Back Section Removal ...........................................................................................22
2-11. Table Top Rotation. ............................................................................................................... 22
2-12. Positioning ............................................................................................................................23
SECTION III MAINTENANCE ..............................................................................................................26
3-1. Routine Inspections .............................................................................................................. 26
3-2. Preventive Maintenance ......................................................................................................26
3-3. Cleaning Recommendations ............................................................................................... 27
3-4. Service .................................................................................................................................28
Although current at time of publication, SKYTRON's policy of continuous development makes this manual subject to change without notice.
6701 OPERATOR • REV2
Page 2
DO NOT SIT ON END OF LEG SECTION(S) AS LOADS IN EXCESS OF 140 LBS, MAY CAUSE INSTABILITY THAT COULD CAUSE THE TABLE TO BE TIPPED OVER.
WARNING
WARNING
USE HEAD SECTION AS FOOT EXTENSION ONLY - WHEN REVERSING PATIENT ON TABLE REFER TO OPERATOR MANUAL.
6701 EQUIPMENT LABELS
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
18
17
16
15
14
13
12
11
10
9
87
6
5
4
3
2
1
TOP
ROTATION
HAND
LE
ALWAYS LOCK THE TABLE TOP IN POSITION AFTER A FULL
ROTATION OF
180°
DO NOT ROTATE THE TOP WITH AN
UNDISTRIBUTED
PATIENT LOAD AS
INSTABILITY MAY
RESULT,
EXERCISE CAUTION
WITH THE TABLE TOP ROTATED
90°
TO THE
BASE SINCE
AN
IMPROPERLY
DISTRIBUTED
PATIENT LOAD MAY CAUSE THE TABLE TO BE TIPPED
OVER.
TABLE TOP
---------- WARNING ----------
Table Capacity: Lift
12
00 lbs.
Articulate 10
00 lbs.
See Operators
Manual for Limitations.
POSSIBL
E E
XPLOSION
HAZARD
IF USED IN THE PRESENCE OF
FLAMMABLE ANESTHETICS.
DANGER
THIS PRODUCT COMPLIES WITH RADIA­TION PERFORMANCE STANDARD 21 CFR AT THE TIME OF MANUFACTURE
Manufactured:
Model No.
DANGER - E
XPLOS
ION
H
AZARD.
DO
NOT USE IN
T
HE
PRESENCE OF
FLA
MMAB
LE
ANAE
STHETICS
DANGER - RISQUE
D'E
XPL
OSION. NE
PAS
EMPLOYER EN
PRE
SENCE
D'A
NESTHESIQUES
INFLAMMABLES
Grounding
reliability can only be a
chiev
ed w
hen
the equip
ment is connected t
o an
equivalent
receptacle mar
ked "Hospit
al Only"
or "Hosp
ital Grade"
TWIST TO LOCK OR RELEASE PLUG
6701 EQUIPMENT LABELS
6701 OPERATOR • REV2
Electrical Specifications
Power requirements Current Leakage Power Cord
120 VAC, 60Hz, 450 Watts Less than 100 micro amps 15 feet w/hospital grade connector (removable)
12-1/2"
19-1/4" 23" 24-1/4"
14-1/2"
19-3/4"
6"
3"
79"
TOP VIEW
90°
60°
14-1/2"
37"
SIDE VIEW
8-1/2"
5-3/4"
19"
22"
41" MAX
23" MIN
END VIEW
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS TEMPERATURE: 15° - 30° C (60° -85° F) HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED. INTERNALLY POWERED EQUIPMENT
ETL LISTED
CONFORMS TO
UL STD 60601-1
CERTIFIED TO
CAN/CSA STD C22.2 NO.601.1
Page 3
6701 Hercules General Purpose Surgical Table Specifications
6701 OPERATOR • REV2
Page 4
SPECIAL USER ATTENTION
Prior to use, all personnel that may operate this table must be instructed in the correct operational procedures. This table is designed for use by trained and qualified personnel for human medical purposes only.
Initial use should not begin until after the users have been instructed by the manufacturer's representative.
In general, common sense will dictate when there is a potential hazard.
The following precautions should be reviewed by all personnel prior to operating the table.
WARNING
A routine instructional program must be implemented by the facility for proper usage instructions for all personnel that may operate this table.
The maximum lifting capacity of the 6701 Hercules table is 1,200 pounds and the maximum articulation weight capacity is 1,000 pounds. When lifting or articulating large patients, pay close attention to the patient position as well as the positioning guidelines and limitations listed in the operation instructions.
The extreme positioning capabilities of the 6701 Hercules Table requires special attention for possible interference points when using multiple function positioning. As with the operation of any surgical table, a certain amount of care should be exercised to position the patient safely. Although the thick pads and sheets substantially protect the patient, pinch points, located at the joints of the top section should always be considered. BE SURE THAT THE ARMS, HANDS AND FINGERS OF THE PATIENT AND THOSE OF THE OPERATING ROOM PERSONNEL ARE CLEAR OF ALL MOVING PARTS BEFORE MOVING THE TABLE. Proper restraints should always be used for patient safety.
Indicates a possibility of personal injury.
CAUTION
Indicates a possibility of damage to equipment.
NOTE
Indicates important facts or helpful hints.
This equipment is intended for use by healthcare professionals only. This equipment may cause radio interference or may disrupt the operation of nearby equipment. It may be necesssary to take mitigation measures, such as re-orienting or relocating the table or shielding the location.
Certain accessories such as the Uro-Drain Tray, Armboards and X-Ray top can be damaged when changing the position of the table top sections. Always look first to see if a desired movement is going to interfere with any accessories in use.
The operator has the ultimate responsibility of preventing damage to the table and surrounding equipment or possible injury to the patient or staff.
The operator must ensure proper positioning is maintained to prevent compromizing respiration, nerve pathways or circulation.
Do not use worn or damaged accessories, they represent an injury hazard.
Remove possible obstacles before lowering or tilting the operating table
Do not place objects on the base of the table, a danger of damage exists during positioning.
Use caution when articulating the table top, pinch hazards exist.
6701 OPERATOR • REV2
SPECIAL USER ATTENTION
Page 5
NOTE
Activating any function button will activate the brake system. Using the TABLE UP function to set the brakes provides a visual assurance that the brakes are locked without altering the table position, except when emergency brake is released.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
NOTE
An equalization grounding terminal is located under the main power panel. This is provided as an alternate pathway to reduce the risk of static shock hazards. Always follow recommended grounding procedures to ensure patient and staff safety.
NOTE
NOTE
Turning the Main Power Switch ON will change the table operation to 120 VAC power.
NOTE
When the red light starts to blink (indicating low power in battery) the table will operate for approximately 5 continuous minutes, typically long enough to use the table for the rest of the day.
NOTE
The charging system operates ONLY when the table is in AC120V operation mode.
NOTE
The table can be operated on 120VAC power while the battery is being recharged.
The table will operate correctly on battery power with the power cord connected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF at the pendant control. It can not be turned Off using the main power switch.
6701 OPERATOR • REV2
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean condition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
Page 6
SPECIAL USER ATTENTION
WARNING
•Do Not unlock brakes when patient is
on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes
to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top rotation must be in normal orientation, that is, the back section over the long end of the table. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for
obstructions on the floor that might prevent the table from moving smoothly to new location. Relock the brakes immediately once the final position is reached and before commencing surgery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
NOTE
With an evenly distributed patient weight load, all table positioning functions will operate smoothly and quietly with a patient weight of up to 1,000 pounds.
NOTE
WARNING
To maximize patient safety, utilize proper restraint methods during extreme Trendelenburg positioning.
NOTE
To prevent table damage, a safety interlock limits lateral tilt positioning to 20° if Trendelenburg positioning exceeds 20°. An audible alarm will sound.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme lateral tilt positioning.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the back section from going above horizontal if the kidney lift is not all the way down. If the kidney lift is raised, an audible alarm will sound when the BACK UP button is pressed.
NOTE
If back section is below horizontal, BACK UP function is limited to 800 pound patient weight.
To prevent table damage, a safety interlock limits Trendelenburg positioning to 20° if lateral tilt positioning exceeds 20°. An audible alarm will sound.
6701 OPERATOR • REV2
SPECIAL USER ATTENTION
Page 7
WARNING
The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in their full down position.
NOTE
When REFLEX button is activated, if Kidney Bridge is up, the back section will not go above horizontal.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the kidney lift from going up if the back section is raised above horizontal. An audible alarm will sound.
NOTE
The e mergency back-up co ntro l switches will function when the table is operating on 120VAC power, battery power, or turned off.
NOTE
•The Emergen cy Bra ke Release
Valve must be closed and tightened (counter-clockwise) before activating any hydraulic function.
•If the Emergency Brake Release
Valve has been operated, the BRAKE UNLOCK button on the pendant control will have to be pressed before brakes will lock again.
NOTE
NOTE
Elevation and brake system functions are not affected by the return to level function.
CAUTION
The Back Section/Kidney Lift safety interlock is not operational when the emergency back-up control switches are used.
CAUTION
The EMERGENCY BRAKE LOCK switch does not activate the brake system timer. The switch must be held until the brakes are completely locked, approximately 10 seconds.
To make the Back Section easier to handle, remove the Head Section and X-ray Top prior to removing the Back Section.
WARNING
Ensure that the Leg and Back sections are properly engaged and secured to pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled for proper orientation. The Leg section cannot be installed on the Back section pins.
6701 OPERATOR • REV2
Page 8
SPECIAL USER ATTENTION
NOTE
Normal table top position is with the head (and back) section over the power cord end of the base.
WARNING
Always lock the table top in position after rotation. DO NOT rotate the top with an unevenly distributed patient weight load as instability may result.
WARNING
•Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes are set before transferring the patient.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replacement.
WARNING
Certain accessories may limit weight capacities. Check with your SKYTRON representative.
WARNING
SKYTRON Products are guaranteed for proper performance with the use of genuine SKYTRON accessories.
•Exercise caution with the table top
rotated 90° to the base since an improperly distributed patient load may cause the table to be tipped over. A table support rod is required for 90° positioning. See Specialty Positioning.
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and cardiac defibrillator monitors.
WARNING
When an antistatic pathway is required, the table has to be used on an antistatic floor.
Accessories and products not furnished by SKYTRON have not been tested for proper performance and safety. Such applications or use are at the discretion of the user to ensure patient and staff safety.
CAUTION
Do not use the table if any of the inspection points fail.
WARNING
Personal injury to patient or staff may result from a lack of proper maintenance of this equipment.
6701 OPERATOR • REV2
Loading...
+ 22 hidden pages