Skytron 6700 Hercules User manual

SURGICAL T ABLE
OPERATORS MANUAL
REV 6/06
6700B HERCULES
Page 1

TABLE OF CONTENTS

Title Page
EQUIPMENT LABELS AND SPECIFICATIONS ................................................................................. 2
6700B Hercules General Purpose Surgical Table Specifications.........................................................3
SPECIAL USER ATTENTION ............................................................................................................. 4
SECTION I INTRODUCTION ............................................................................................................. 10
1-1. General..................................................................................................................................... 10
1-2. Power Requirements................................................................................................................ 10
1-3. Pendant Control Unit ................................................................................................................ 11
1-4. Floor Lock/Brake System ......................................................................................................... 11
SECTION II OPERATION ...................................................................................................................12
2-1. Electrical Power ....................................................................................................................... 12
2-2. AC 120V Operation .................................................................................................................. 12
2-3. Battery Operation ..................................................................................................................... 13
2-4. Automatic Shut-Off ................................................................................................................... 13
2-5. Charging the Battery ................................................................................................................14
2-6. Positioning Functions ............................................................................................................... 15
a. Floor Lock/Brake system ..................................................................................................... 15
b. Trendelenburg ...................................................................................................................... 16
c. Lateral Tilt ............................................................................................................................ 16
d. Back Section ........................................................................................................................ 17
e. Elevation .............................................................................................................................. 17
f. Leg Section .......................................................................................................................... 18
g. Flex Positioning.................................................................................................................... 18
h. Kidney Lift ............................................................................................................................ 19
i. Return To Level ................................................................................................................... 19
j. Beach Chair ......................................................................................................................... 19
2-7. Emergency Back-up controls................................................................................................... 20
2-8. Emergency Brake Release...................................................................................................... 20
2-9. Head Section............................................................................................................................ 21
2-10. Leg and Back Section Removal ...............................................................................................22
2-11.Table Top Rotation................................................................................................................... 22
2-12. Positioning................................................................................................................................ 23
SECTION III MAINTENANCE ............................................................................................................. 26
3-1. Preventive Maintenance........................................................................................................... 26
3-2. Cleaning Recommendations ....................................................................................................26
3-3. Service ..................................................................................................................................... 27
REV 6/06
Page 2
TYPE B
EQUIPMENT

EQUIPMENT LABELS AND SPECIFICATIONS

INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED. INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING. IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY, THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED HOSPITAL GRADE OUTLET.
N
IPX4
V A
HZ
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS TEMPERATURE: 15˚ - 30˚ C (60˚ - 85˚ F) HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS. INDICATES SPECIAL USER ATTENTION.
POWERED BY AC VOLTAGE
POWERED BY BATTERY
BATTERY TYPE:
SEALED LEAD ACID 12V, VALVE REGULATED 16AH, 10HR (530W/10MIN)
15 AMP INTERNAL CIRCUIT BREAKER
15A
6700B Hercules General Purpose Surgical Table Specifications
14-1/2"
Page 3
60˚
90˚
12-1/2"
19-1/4" 23" 24-1/4"
14-1/2"
TOP VIEW
5-3/4"
76"
3"
19-3/4"
21-3/4"
44" MAX
23" MIN
8-1/2"
36"
SIDE VIEW
ENTELA CERTIFIED
TO UL2601-1 CAN/CSA601.1, IEC 60601-2-46
5-3/4"
Electrical Specifications
Power requirements Current Leakage Power Cord
19"
END VIEW
120 VAC, 60Hz, 450 Watts Less than 100 micro amps 15 feet w/hospital grade connector (removable)
SERIAL NUMBER TAG
Page 4

SPECIAL USER ATTENTION

Prior to use, all personnel that may operate this table must be instructed in the correct opera­tional procedures. This table is designed for use by trained and qualified personnel for hu­man medical purposes only.
Initial use should not begin until after the users have been instructed by the manufacturer's representative.
A routine instructional program must be imple­mented by the facility for proper usage instruc­tions for all personnel that may operate this table.
The maximum lifting capacity of the 6700B Hercules table is 1,200 pounds and the maxi­mum articulation weight capacity is 1,000 pounds. When lifting or articulating large pa­tients, pay close attention to the patient posi­tion as well as the positioning guidelines and limitations listed in the operation instructions.
In general, common sense will dictate when there is a potential hazard.
The following precautions should be reviewed by all personnel prior to operating the table.
WARNING
Indicates a possibility of personal injury.
CAUTION
Indicates a possibility of damage to equipment.
NOTE
The extreme positioning capabilities of the 6700B Hercules Table requires special attention for pos­sible interference points when using multiple func­tion positioning. As with the operation of any surgical table, a certain amount of care should be exercised to position the patient safely. Although the thick pads and sheets substantially protect the patient, pinch points, located at the joints of the top section should always be considered. BE SURE THAT THE ARMS, HANDS AND FINGERS OF THE PATIENT AND THOSE OF THE OPERAT­ING ROOM PERSONNEL ARE CLEAR OF ALL MOVING PARTS BEFORE MOVING THE TABLE. Proper restraints should always be used for patient safety.
Certain accessories such as the Uro-Drain Tray, Armboards and X-Ray top can be damaged when changing the position of the table top sections. Always look first to see if a desired movement is going to interfere with any accessories in use.
The operator has the ultimate responsibility of preventing damage to the table and surrounding equipment or possible injury to the patient or staff.
Indicates important facts or helpful hints.
Do not use worn or damaged accessories, they represent an injury hazard.
Remove possible obstacles before lowering or tilting the operating table
Do not place objects on the base of the table, a danger of damage exists during positioning.
Use caution when articulating the table top, pinch hazards exist.
The operator must ensure proper positioning is maintained to prevent compromizing respiration, nerve pathways or circulation.
SPECIAL USER ATTENTION
Page 5
NOTE
Activating any function button will acti­vate the brake system. Using the TABLE UP function to set the brakes provides a visual assurance that the brakes are locked without altering the table posi­tion, except when emergency brake is released.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
NOTE
An equalization grounding terminal is located under the main power panel. This is provided as an alternate path­way to reduce the risk of static shock hazards. Always follow recommended grounding procedures to ensure patient and staff safety.
NOTE
Turning the Main Power Switch ON will change the table operation to 120 VAC power.
NOTE
When the red light starts to blink (indi­cating low power in battery) the table will operate for approximately 5 con­tinuous minutes, typically long enough to use the table for the rest of the day.
NOTE
The charging system operates ONLY when the table is in AC120V operation mode.
NOTE
The table can be operated on 120VAC power while the battery is being re­charged.
NOTE
The table will operate correctly on bat­tery power with the power cord con­nected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF at the pendant control. It can not be turned Off using the main power switch.
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean con­dition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
Page 6
SPECIAL USER ATTENTION
WARNING
•Do Not unlock brakes when patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top rotation must be in normal orienta­tion, that is, the back section over the long end of the table. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for obstructions on the floor that might pre­vent the table from moving smoothly to new location. Relock the brakes imme­diately once the final position is reached and before commencing surgery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
NOTE
With an evenly distributed patient weight load, all table positioning functions will operate smoothly and quietly with a patient weight of up to 1,000 pounds.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme Trendelenburg positioning.
NOTE
To prevent table damage, a safety inter­lock limits lateral tilt positioning to 20° if Trendelenburg positioning exceeds 20°. An audible alarm will sound.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme lateral tilt positioning.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the back sec­tion from going above horizontal if the kidney lift is not all the way down. If the kidney lift is raised, an audible alarm will sound when the BACK UP button is pressed.
NOTE
NOTE
To prevent table damage, a safety inter­lock limits Trendelenburg positioning to 20° if lateral tilt positioning exceeds 20°. An audible alarm will sound.
If back section is below horizontal, BACK UP function is limited to 800 pound patient weight.
SPECIAL USER ATTENTION
WARNING
The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in their full down position.
NOTE
When REFLEX button is activated, if Kidney Bridge is up, the back section will not go above horizontal.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the kidney lift from going up if the back section is raised above horizontal. An audible alarm will sound.
NOTE
Elevation and brake system functions are not affected by the return to level function.
Page 7
NOTE
The emergency back-up control switches will function when the table is operating on 120VAC power, battery power, or turned off.
NOTE
The Emergency Brake Release Valve must be closed and tightened (clock­wise) before activating any function.
•If the Emergency Brake Release Valve has been operated, the BRAKE UN­LOCK button on the pendant control will have to be pressed before brakes will lock again.
NOTE
To make the Back Section easier to handle, remove the Head Section and X-ray Top prior to removing the Back Section.
WARNING
CAUTION
The Back Section/Kidney Lift safety interlock is not operational when the emergency back-up control switches are used.
CAUTION
The EMERGENCY BRAKE LOCK switch does not activate the brake sys­tem timer. The switch must be held until the brakes are completely locked, ap­proximately 10 seconds.
Ensure that the Leg and Back sections are properly engaged and secured to pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled for proper orientation. The Leg section cannot be installed on the Back section pins.
Page 8
SPECIAL USER ATTENTION
NOTE
Normal table top position is with the head (and back) section over the power cord end of the base.
WARNING
Always lock the table top in position after rotation. DO NOT rotate the top with an unevenly distributed patient weight load as instability may result.
WARNING
•Make sure the TOP ROTATION LOCK HANDLE is tightened and the brakes are set before transferring the patient.
WARNING
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replacement.
WARNING
Certain accessories may limit weight capacities. Check with your SKYTRON representative.
•Exercise caution with the table top rotated 90° to the base since an improp­erly distributed patient load may cause the table to be tipped over. A table support rod is required for 90° position­ing. See Specialty Positioning.
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and car­diac defibrillator monitors.
NOTE
Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedure.
WARNING
Always follow OSHA blood-borne patho­gens standards for protective clothing, including gloves, masks and eye pro­tection when cleaning the surgical table.
SPECIAL USER ATTENTION
CAUTION
Thoroughly read and follow the manufacturer's directions for all clean­ing fluids. DO NOT use cleaners con­taining phenolics.
CAUTION
When using spray cleaners DO NOT spray fluids directly into electrical re­ceptacles or micro switches.
CAUTION
Page 9
Before replacing pads on the table, make sure the pads and all mating surfaces are completely dry. Moisture trapped between the pads and mating surfaces may cause distortion of table tops.
Page 10

SECTION I INTRODUCTION

HEAD
SECTION
SIDE RAIL
HEAD SECTION
LOCKING KNOB
SERVICE
ACCESS COVER
REMOVABLE
BACK SECTION
PENDANT CONTROL STORAGE BRACKET
TOP ROTATION
POWER
CORD
HANDLE
PENDANT CONTROL
SEAT
SECTION
REMOVABLE
LEG SECTION
LEG SECTION
RELEASE LEVER
(RECESSED)
FLOOR/LOCK
BRAKE (4)
MAIN POWER
SWITCH
Figure 1-1. 6700B HERCULES
1-1. General
SKYTRON’s 6700B Hercules Surgical Table is an electro-hydraulically operated, general purpose sur­gical table. See figure 1-1.
The electro-hydraulic positioning functions oper­ated by the hand-held, push button, pendant control unit are: Trendelenburg, lateral tilt, back section, elevation, leg section, flex/reflex, kidney lift, return to level, beach chair and the floor lock/brake sys­tem.
Manual controls are provided for head section posi­tioning, table top rotation, emergency brake re­lease, back section removal and leg section re­moval.
EMERGENCY
BRAKE RELEASE
cord with a three prong, hospital grade plug. The main power ON/OFF switch is located on the electrical panel on the front edge of the table base. See figure 1-2.
The battery charging indicator and foot control connector are also located on the electrical panel.
1-2. Power Requirements
The 6700B Hercules Surgical Table requires a 120VAC, 60 Hz electrical power supply. The table is equipped with a removable 15 foot long power
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
Figure 1-2. Electrical Panel
FOOT CONTROL
CONNECTOR
1-3. Pendant Control Unit
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TABLE UP
(BRAKE LOCK)
LATERAL
TILT LEFT
BACK UP
LEG UP
FLEX
KIDNEY LIFT UP
RETURN
TO LEVEL
LATERAL TILT RIGHT
BACK DOWN TABLE DOWN
LEG DOWN REFLEX KIDNEY
LIFT DOWN BEACH
CHAIR BRAKE
UNLOCK
REVERSE
TRENDELENBURG
TRENDELENBURG
Page 11
The hand-held pendant control unit (figure 1-3) has a non-slip rubber cover which assures a positive grip during use. A spring clip hanger is located on the back of the control for storage. When the Pendant Control is not in use, it should be stored on a convenient side or end rail. A bracket is located under the table top next to the pendant control connector for storage of the Pendant Control when the table is not in use and during cleaning. See figure 1-4.
POWER
SIDE RAIL
CLIP
INDICATOR
FUNCTION
BUTTONS
The function push buttons are identified with abbre­viated descriptions for all functions. See figure 1-5. When illuminated the Trendelenburg and table up buttons are red, the remaining buttons are all green.
PENDANT CONTROL STORAGE BRACKET
Figure 1-4. Pendant Control Storage Bracket
Figure 1-3. Pendant Control Unit
BACK
SECTION
TOP ROTATION
HANDLE
SEAT
SECTION
Figure 1-5. Function Buttons
1-4. Floor Lock/Brake System
The floor lock/brake system consists of four self­leveling, hydraulic brake cylinders which raise and support the table base off from the casters. Press the TABLE UP button on the pendant control to set the table’s brakes. An electronic timer will activate the brake system until the brakes are completely set, approximately 8-10 seconds.
NOTE
Activating any function button will acti­vate the brake system. Using the TABLE UP function to set the brakes provides a visual assurance that the brakes are locked without altering the table posi­tion, except when emergency brake is released.
Page 12
2-1. Electrical Power

SECTION II OPERATION

The 6700B table will operate on either 120 VAC or battery power.
WARNING
Prior to operating the table, observe all table caution labels and review the SPECIAL USER ATTENTION sec­tion in the front of this manual.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
NOTE
b. Activate the POWER SWITCH located on the
electrical panel. The switch will illuminate.
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR TWIST TO LOCK
FOOT CONTROL
CONNECTOR
EQUALIZATION
GROUNDING TERMINAL
Figure 2-1. Electrical Panel
An equalization terminal is located un­der the main power panel. This is pro­vided as an alternate pathway to re­duce the risk of static shock hazards. Always follow recommended ground­ing procedures to ensure pattient and staff safety.
2-2. AC 120V Operation
Use the following procedures to operate the table on 120 VAC power.
a. Make sure the Power cord is securely at­tached to the table. To install the power cord, align the cord connector with the base connector, insert the cord and twist clockwise to lock the cord into the connector. See figure 2-1. Plug the cord into a properly grounded, Hospital Grade, 120 VAC out­let. Make sure the power cord is routed to the outlet to prevent it from being in the way of operating personnel.
The pendant control buttons and the green AC 120V, POWER indicator light located in the upper right corner of the pendant control will illuminate. See figure 2-2.
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
Figure 2-2. Pendant Control
c. The table is now ready for 120VAC operation.
Page 13
2-3. Battery Operation
a. Make sure the Battery Indicator and Main
Power Switch indicator lights, on the electrical panel, are OFF. See figure 2-2. If the indicator lights are ON, turn AC120V operation OFF with the main power switch.
NOTE
The table will operate correctly on bat­tery power with the power cord con­nected to a wall outlet or disconnected.
b. Press the BATT button on the hand-held pendant control. The pendant control buttons and the red BATTERY indicator light, located in the upper right corner of the pendant control, will illumi­nate.
2-4. Automatic Shut-Off
a. To prevent unnecessary discharge of the
battery, a timer is built into the battery circuit. This timer will automatically shut the battery power OFF after 1½ hours of table inactivity.
b. To turn the table ON again, press the BATT button on the pendant control, the pendant control buttons and the red indicator light will illuminate.
NOTE
Turning the Main Power Switch ON will change the table operation to 120 VAC power.
c. The table is now ready for BATTERY opera-
tion.
d. To extend the battery charge life, turn the BATTERY power OFF with the pendant control when the table is not going to be used.
NOTE
Battery Operation must be turned OFF at the pendant control. It cannot be turned OFF using the main power switch.
Page 14
2-5. Charging the Battery
Batteries should be charged:
• When the table is placed into initial service
• As indicated by Battery Indicator
• Every week under normal service conditions
a. Battery Indicator The Battery Indicator consists of ten lighted bars, 3 red, 4 yellow and 3 green. See figure 2-3. Each bar represents a percentage of the battery charge condition. When all ten bars are illuminated, the batteries are fully charged. The following list shows the battery charge level as indicated by the lighted bars;
b. If the battery needs to be charged when operating the table on battery power, the red indica­tor light on the pendant control will begin to blink.
NOTE
When the red light starts to blink (indi­cating low power in battery) the table will operate for approximately 5 con­tinuous minutes, typically long enough to use the table for the rest of the day.
NOTE
3 green 100% -Fully charged 2 green 89% 1 green 78% 4 yellow 67% 3 yellow 56% 2 yellow 45% -Needs Charging (BATT
indicator on pendant will flash) 1 yellow 34% -Needs Charging 3 red 23% -Needs Charging
(poor performance) 2 red 12% -Needs Charging
(intermittent performance) 1 red 1% -Needs Charging
(inoperable)
During charging, the bars will light in sequence to the respective charge level, turn off and light in sequence again.
The charging system operates ONLY when the table is in AC120V operation mode.
c. To recharge the battery, make sure the power cord is connected, plugged into a 120VAC wall outlet and the main POWER SWITCH - ON.
NOTE
The table can be operated on 120VAC power while the battery is being re­charged.
d. A full battery charge will last approximately 2 weeks under normal operating conditions. How­ever, it is recommended to charge the batteries at the end of each week to establish a normal routine protocol. Lead acid batteries last longer if they are not permitted to fully discharge. The table features (2) 12 volt, sealed, lead acid batteries which require no manual maintenance. Lead acid gel batteries, under a proper charging program, feature an ap­proximate normal life of 4 years.
BATTERY
INDICATOR
Figure 2-3. Battery Indicator
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean con­dition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
2-6. Positioning Functions
The hand-held pendant control (figure 2-4) acti­vates the following table functions:
TRENDELENBURG
REVERSE
LATERAL
TILT LEFT
BACK UP
TABLE UP
(BRAKE LOCK)
LEG UP
FLEX
KIDNEY LIFT UP
RETURN
TO LEVEL
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TRENDELENBURG LATERAL
TILT RIGHT BACK DOWN
TABLE DOWN LEG DOWN
REFLEX KIDNEY
LIFT DOWN BEACH
CHAIR BRAKE
UNLOCK
Figure 2-4. Pendant Control Function Buttons
a. Floor Lock/Brake System. To activate the
brakes without affecting table positioning, press the TABLE UP button. See figure 2-5. The eleva­tion cylinder will not function until the brakes are completely extended.
TABLE UP
(BRAKE LOCK)
Page 15
Press the BRAKE UNLOCK button on the pendant control to release the four self-leveling brake feet in order to move the table. See figure 2-5. The brake delay circuit automatically retracts the brake sys­tem. It takes approximately 7-8 seconds to totally release the system.
WARNING
•Do Not unlock brakes when patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top rotation must be in normal orienta­tion, that is, the back section over the long end of the table. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for obstructions on the floor that might pre­vent the table from moving smoothly to new location. Relock the brakes imme­diately once the final position is reached and before commencing surgery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
BRAKE UNLOCK
Figure 2-5. Brake System Activation
NOTE
With an evenly distributed patient weight load, all table positioning functions will operate smoothly and quietly with a patient weight of up to 1,000 pounds.
Page 16
b. Trendelenburg. To place the table in a
Trendelenburg (head down) position, press the TREND button (figure 2-6). To place the table in a reverse Trendelenburg (head up) position, press the REV TREND button. Trendelenburg position­ing of up to 30° may be obtained if lateral tilt positioning is less than 20°.
NOTE
To prevent table damage, a safety inter­lock limits Trendelenburg positioning to 20° if lateral tilt positioning exceeds 20°. An audible alarm will sound.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme Trendelenburg positioning.
c. Lateral Tilt. To achieve lateral tilt right (as viewed from the head end of the table), press the TILT RIGHT button (figure 2-7). To achieve lateral tilt left, press the TILT LEFT button. Tilt of up to 30° may be obtained if Trendelenburg positioning is less than 20°.
NOTE
To prevent table damage, a safety inter­lock limits lateral tilt positioning to 20° if Trendelenburg positioning exceeds 20°. An audible alarm will sound.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme lateral tilt positioning.
TRENDELENBURG
REVERSE
30˚
30˚
TRENDELENBURG
Figure 2-6. Trendelenburg Positioning
LATERAL
TILT LEFT
30˚ 30˚
LATERAL TILT RIGHT
Figure 2-7. Lateral Tilt Positioning
Page 17
d. Back Section. To raise the back section,
press the BACK UP button (figure 2-8). The back section will raise up to 90° above horizontal. To lower the back section, press the BACK DOWN button. The back section will go down to 40° below horizontal.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the back sec­tion from going above horizontal if the kidney lift is not all the way down. If the kidney lift is raised, an audible alarm will sound when the BACK UP button is pressed.
NOTE
If back section is below horizontal, BACK UP function is limited to 800 pound patient weight.
e. Elevation. To raise table top, press the TABLE UP button (figure 2-9). The table will lift a patient weight of 1,200 pounds up to a maximum height of 44" (48" with X-Ray top and 2" pad). To lower the table top, press the TABLE DOWN button. The table top will go down to a minimum height of 23".
TABLE
UP
44"
23"
TABLE DOWN
BACK UP
90˚
40˚
BACK DOWN
Figure 2-8. Back Section Positioning
Figure 2-9. Elevation Function
Page 18
f. Leg Section. To lower the leg section, press
the LEG DOWN button (figure 2-10). The leg section will go down to 100° below horizontal. To raise the leg section, press the LEG UP button. The leg section will go up to 20° above horizontal.
WARNING
The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in their full down position.
LEG UP
LEG DOWN
g. Flex Positioning. To place the table top in a
flex position from horizontal, press the FLEX button (figure 2-11). To return the table top to a horizontal position or into a reflex position, press the LEVEL or REFLEX button.
NOTE
When REFLEX button is activated, if Kidney Bridge is up, the back section will not go above horizontal.
FLEX
REFLEX
20˚
100˚
Figure 2-10. Leg Section Positioning
Figure 2-11. Flex/Reflex Positioning
Page 19
h. Kidney Lift. To raise the built-in kidney lift,
press the KIDNEY UP button (figure 2-12). Up to
5.75 inches of lift can be achieved. Press the KIDNEY DOWN button to lower the kidney lift.
KIDNEY UP
6"
KIDNEY DOWN
Figure 2-12. Kidney Lift Positioning
NOTE
j. Beach Chair. To place the top in the beach
chair position from a level position, press the BEACH CHAIR button (figure 2-14). The back section will raise, the leg section will lower and the Trendelenburg positioning will function simulta­neously. The functions will stop when Trendelen­burg reaches it limit.
BEACH CHAIR
To prevent damage to the kidney lift, a safety interlock prevents the kidney lift from going up if the back section is raised above horizontal. An audible alarm will sound .
i. Return To Level. To return the table top to a level
position, press the LEVEL button (figure 2-13).
NOTE
Elevation and brake system functions are not affected by the return to level function.
Figure 2-14. Beach Chair Positioning
RETURN
TO LEVEL
Figure 2-13. Return To Level
Page 20
2-7. Emergency Back-up Controls
a. The emergency back-up control switches are
located under the access door on the service access cover in the table base. See figure 2-15.
FUNCTION CONTROL
ACCESS DOOR
Figure 2-15. Emergency Controls Location
b. In the event of either a power failure or a
problem with the hand-held pendant control, the table can be operated using the emergency back­up switches. Simply push the desired emergency switch in the appropriate direction to operate the table functions. See figure 2-16.
NOTE
The emergency back-up control switches will function when the table is operating on 120VAC power, battery power, or turned off.
c. Switches are provided for Trendelenburg, lateral tilt, back section, elevation, leg section, kidney down and brake lock. These switches are spring-loaded so they return to the neutral or center position when released.
2-8. Emergency Brake Release.
In case of a power failure or an electrical problem within the table, the emergency brake release system can be used to move the table. The control knob for this function is located on the side of the table base and is identified by an EMERGENCY BRAKE RELEASE label. Turn the knob counter­clockwise to release the brakes. See figure 2-17.
CAUTION
The Back Section/Kidney Lift safety interlock is not operational when the emergency back-up control switches are used.
REV
TILT
TREND
TREND TILT
BACKUPTABLEUPLEGUPBRAKE
RIGHT
BACK
DOWN
TABLE
DOWN
LEFT
LEG
DOWN
LOCK
KIDNEY
DOWN
Figure 2-16. Emergency Back-Up Controls
CAUTION
The EMERGENCY BRAKE LOCK switch does not activate the brake sys­tem timer. The switch must be held until the brakes are completely locked, ap­proximately 10 seconds.
WARNING
•Do Not unlock brakes when patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top rotation must be in normal orienta­tion, that is, the back section over the long end of the table. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
HEAD SECTION
FOOT/LEG SECTION
LOCKING KNOB
LOCKING KNOB
•Prior to unlocking brakes, check for obstructions on the floor that might pre­vent the table from moving smoothly to new location. Relock the brakes imme­diately once the final position is reached and before commencing surgery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
Page 21
HEAD SECTION RELEASE BAR
SERVICE ACCESS
COVER
POWER
CORD
EMERGENCY
BRAKE RELEASE
BRAKE (4)
Figure 2-17. Emergency Brake Release
NOTE
The Emergency Brake Release Valve must be closed and tightened (clock­wise) before activating any function.
•If the Emergency Brake Release Valve has been operated, the UNLOCK button on the pendant control will have to be pressed before brakes will lock again.
Figure 2-18. Head Section Adjustment
b. By loosening two locking knobs beneath the
back section, an additional 1.5" of longitudinal ad­justment can be achieved. If desired, the head section may be removed by loosening the locking knobs and pulling it straight out of the back section.
The 6700B Hercules Table has the capability of attaching the head section to the leg section for use as a foot extension ONLY. Do Not reverse the patient on the table without first consulting with SKYTRON.
Two locking knobs are located on the inside of the leg section for securing the head section. See figure 2-19.
2-9. Head Section
a. A quick release positioning bar located under
and to the front of the head section (figure 2-18) is used to raise or lower the head section. Pull the release bar toward the head end to allow the section to pivot up or down. Positioning from 60° above horizontal to 90° below horizontal in 15° increments is available. Release the bar to lock the head section in position.
Figure 2-19. Repositioning Head Section
(for use as a Foot Extension)
Page 22
2-10. Leg and Back Section Removal.
a. The Leg Section and the Back Section on the
6700B Hercules tables are both removable. See figure 2-20 (Leg Section shown).
b. To remove either section, level the table top, simultaneously depress both release levers and pull the section out. Press the LEG DOWN or BACK DOWN button on the pendant control to position the attachment pins down and out of the way.
NOTE
To make the Back Section easier to handle, remove the Head Section and X-ray Top prior to removing the Back Section.
WARNING
Ensure that the Leg and Back sections are properly engaged and secured to pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled for proper orientation. The Leg section cannot be installed on the Back section pins.
2-11. Table Top Rotation.
NOTE
Normal table top position is with the head (and back) section over the power cord end of the base.
FOOT/LEG SECTION
LOCKING LEVERS
Figure 2-20. Leg Section Removal
c. To install the Leg Section, press and hold the
LEG UP button until the attachment pins com­pletely stop. Install the section on the pins. Level the table top and pull out on the section to make sure the release levers are completely locked.
d. To install the Back Section if the attachment pins are not aligned, press and hold the BACK DOWN button until the attachment pins completely stop (40° down). Press REV TREND to bring the pins up to a level position.
a. The table top can be horizontally rotated 210° without having to rotate the entire table. Refer to figure 2-21. To rotate the top, turn the TOP ROTA­TION LOCK HANDLE counterclockwise to re­lease, grasp the table by the head end and rotate the top counterclockwise. Lock the top in position by tightening the TOP ROTATION LOCK HANDLE clockwise.
WARNING
Always lock the table top in position after rotation. DO NOT rotate the top with an unevenly distributed patient weight load as instability may result.
TOP ROTATION
HANDLE
Figure 2-21. Top Rotation
b. The use of the optional support rod allows the
table top to be rotated 90° from the base. See figure 2-22.
WARNING
•Make sure the TOP ROTATION LOCK HANDLE is tightened and the brakes are set before transferring the patient.
Page 23
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and car­diac defibrillator monitors.
•Exercise caution with the table top rotated 90° to the base since an improp­erly distributed patient load may cause the table to be tipped over. A table support rod is required for 90° position­ing. See figure 2-22.
SUPPORT
ROD
Figure 2-22. 90 Degree Top Rotation
WARNING
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replacement.
2-12. Positioning
The use of certain optional accessories available from SKYTRON further extend the positioning ca­pabilities of the 6700B Hercules Tables. Refer to the following "Positioning Guidelines" or contact your SKYTRON representative for further details.
WARNING
Certain accessories may limit weight capacities. Check with your SKYTRON representative.
Page 24
6700B HERCULES General Purpose Patient Positioning Guidelines
Upper Body Imaging
Lap Nissen
Lower Body Imaging
Urology
Ophthalmic/ENT
Neuro (neck)
Nephrectomy
Neuro (lumbar)
Page 25
Shoulder Arthroscopy
Hip Revision
Neuro (head)
Cardiovascular
6700B HERCULES Bariatric Recommended Patient Positioning Guidelines
Abdominal
Cysto/GYN
Gall Bladder
Cardiovascular
Page 26

SECTION III MAINTENANCE

3-1. Preventive Maintenance
The following preventive maintenance checks and services are recommended to ensure the service­ability and proper operation of your SKYTRON Surgical Table, and should only be performed by qualified SKYTRON trained personnel.
a. During normal cleaning, a general visual ex­amination should be made checking for leaks, loose bolts or parts, and cracked, chipped, or missing paint. Any necessary repairs should be made. Refer to the 6700B Maintenance Manual.
b. Semi-annually the following checks and ser­vices should be performed:
1. Check all hydraulic fittings, mini-valves and slave cylinders for proper operation and any signs of leaks.
2. Check the hydraulic speed controls and adjust if necessary.
3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedure.
The following procedures should be followed when cleaning the surgical table between cases.
Place table top in level position prior to starting cleaning procedure.
WARNING
Always follow OSHA blood-borne patho­gens standards for protective clothing, including gloves, masks and eye pro­tection when cleaning the surgical table.
3. Pressure check (with a gauge) the pres­sure relief valve.
4. Check all mechanical adjustments and ad­just as necessary.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of foot leg release levers, (release knob early models). Lubricate as necessary.
Remove major contaminants from the table with disposable materials following appropriate biohaz­ard waste disposal procedures.
Remove all table pads and place them on a flat surface for cleaning.
CAUTION
Thoroughly read and follow the manufacturer's directions for all clean­ing fluids. DO NOT use cleaners con­taining phenolics.
Apply cleaning fluid liberally to top and sides of each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to remove all moisture.
Repeat the steps to clean the bottom of the each pad.
CAUTION
When using spray cleaners DO NOT spray fluids directly into electrical re­ceptacles or micro switches.
Repeat cleaning procedure for all table surfaces including the top, sides, elevation column, base and all accessories.
CAUTION
Before replacing pads on the table, make sure the pads and all mating surfaces are completely dry. Moisture trapped between the pads and mating surfaces may cause distortion of table tops.
When the cleaning procedure is complete, replace all pads and accessories as applicable.
Page 27
3-3. Service
Table maintenance can be performed by trained maintenance personnel using SKYTRON autho­rized replacement parts and service techniques. Service instructions and parts are available from SKYTRON.
Preventive Maintenance contracts are available through your local SKYTRON representative.
To obtain service instructions, replacement parts, factory service or preventive maintenance con­tracts, contact the SKYTRON representative listed below.
Or contact:
SKYTRON 5000 36th Street S.E. Grand Rapids, MI 49512 1-800-SKYTRON (1-800-759-8766) Fax. 1-616-957-5053
Remove pendant control from table side rail and apply cleaning solution to the pendant control and cord.
Use a clean cloth dampened with water to remove cleaning solution.
Use another clean damp cloth to remove any remaining residue.
Install pendant control on side rail for storage when cleaning procedure is complete.
5/05

INTRODUCTION

This manual contains the exploded views and replacement parts lists for the service­able components of the SKYTRON Model 6700B Surgical Table.
Each serviceable part in these exploded views is identified by a reference number. Use this number to locate necessary part information in the parts list adjacent to the exploded view.
Always use the complete SKYTRON part number and description when ordering replacement parts.
Always use the complete table serial number (S.N.) when ordering replacement parts.
Special Tools and Maintenance Items listed on page 38.
Abbreviations
As Required .......................................... A/R
Optional................................................... opt
Serial Number ....................................... S.N.
Not Shown.............................................NS
SERIAL NUMBER TAG
5/05
Although current at the time of publication, SKYTRON's policy of continuous development makes this manual subject to change without notice.
Loading...
+ 97 hidden pages