3-3. Service ..................................................................................................................................... 27
REV 6/06
Although current at time of publication, SKYTRON's policy of continuous development makes this manual
subject to change without notice.
Page 2
TYPE B
EQUIPMENT
EQUIPMENT LABELS AND SPECIFICATIONS
INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED.
INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING.
IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY,
THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED
HOSPITAL GRADE OUTLET.
N
IPX4
V
A
HZ
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS
TEMPERATURE: 15˚ - 30˚ C (60˚ - 85˚ F)
HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS.
INDICATES SPECIAL USER ATTENTION.
POWERED BY AC VOLTAGE
POWERED BY BATTERY
BATTERY TYPE:
SEALED
LEAD ACID 12V, VALVE REGULATED
16AH, 10HR (530W/10MIN)
15 AMP INTERNAL CIRCUIT BREAKER
15A
6700B Hercules General Purpose Surgical Table Specifications
14-1/2"
Page 3
60˚
90˚
12-1/2"
19-1/4"23"24-1/4"
14-1/2"
TOP VIEW
5-3/4"
76"
3"
19-3/4"
21-3/4"
44" MAX
23" MIN
8-1/2"
36"
SIDE VIEW
ENTELA CERTIFIED
TO UL2601-1
CAN/CSA601.1, IEC 60601-2-46
5-3/4"
Electrical Specifications
Power requirements
Current Leakage
Power Cord
19"
END VIEW
120 VAC, 60Hz, 450 Watts
Less than 100 micro amps
15 feet w/hospital grade
connector (removable)
SERIAL NUMBER TAG
Page 4
SPECIAL USER ATTENTION
Prior to use, all personnel that may operate this
table must be instructed in the correct operational procedures. This table is designed for
use by trained and qualified personnel for human medical purposes only.
Initial use should not begin until after the users
have been instructed by the manufacturer's
representative.
A routine instructional program must be implemented by the facility for proper usage instructions for all personnel that may operate this
table.
The maximum lifting capacity of the 6700B
Hercules table is 1,200 pounds and the maximum articulation weight capacity is 1,000
pounds. When lifting or articulating large patients, pay close attention to the patient position as well as the positioning guidelines and
limitations listed in the operation instructions.
In general, common sense will dictate when there
is a potential hazard.
The following precautions should be reviewed
by all personnel prior to operating the table.
WARNING
Indicates a possibility of personal injury.
CAUTION
Indicates a possibility of damage to
equipment.
NOTE
The extreme positioning capabilities of the 6700B
Hercules Table requires special attention for possible interference points when using multiple function positioning. As with the operation of any
surgical table, a certain amount of care should be
exercised to position the patient safely. Although
the thick pads and sheets substantially protect the
patient, pinch points, located at the joints of the top
section should always be considered. BE SURE
THAT THE ARMS, HANDS AND FINGERS OF
THE PATIENT AND THOSE OF THE OPERATING ROOM PERSONNEL ARE CLEAR OF ALL
MOVING PARTS BEFORE MOVING THE
TABLE. Proper restraints should always be used
for patient safety.
Certain accessories such as the Uro-Drain Tray,
Armboards and X-Ray top can be damaged when
changing the position of the table top sections.
Always look first to see if a desired movement is
going to interfere with any accessories in use.
The operator has the ultimate responsibility of
preventing damage to the table and surrounding
equipment or possible injury to the patient or staff.
Indicates important facts or helpful hints.
Do not use worn or damaged accessories, they
represent an injury hazard.
Remove possible obstacles before lowering or
tilting the operating table
Do not place objects on the base of the table, a
danger of damage exists during positioning.
Use caution when articulating the table top, pinch
hazards exist.
The operator must ensure proper positioning is
maintained to prevent compromizing respiration,
nerve pathways or circulation.
SPECIAL USER ATTENTION
Page 5
NOTE
Activating any function button will activate the brake system. Using the TABLE
UP function to set the brakes provides
a visual assurance that the brakes are
locked without altering the table position, except when emergency brake is
released.
WARNING
Possible explosion hazard exists if table
is used in the presence of FLAMMABLE
ANESTHETICS.
NOTE
An equalization grounding terminal is
located under the main power panel.
This is provided as an alternate pathway to reduce the risk of static shock
hazards. Always follow recommended
grounding procedures to ensure patient
and staff safety.
NOTE
Turning the Main Power Switch ON will
change the table operation to 120 VAC
power.
NOTE
When the red light starts to blink (indicating low power in battery) the table
will operate for approximately 5 continuous minutes, typically long enough
to use the table for the rest of the day.
NOTE
The charging system operates ONLY
when the table is in AC120V operation
mode.
NOTE
The table can be operated on 120VAC
power while the battery is being recharged.
NOTE
The table will operate correctly on battery power with the power cord connected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF
at the pendant control. It can not be
turned Off using the main power switch.
NOTE
If the table is stored for a period greater
than 6 months, the batteries should be
removed and stored in a dry, clean condition at a storage temperature of 68° F
(20° C). Batteries should be recharged
every 6 months of product storage.
Page 6
SPECIAL USER ATTENTION
WARNING
•Do Not unlock brakes when patient is
on the table. An uneven patient weight
load may cause instability.
•If circumstances demand table brakes
to be unlocked, the patient must be
centered and evenly distributed on the
table top (i.e. supine or prone position)
with the table lowered to its lowest height
position. The maximum patient weight
should not exceed 500 pounds. Table
top rotation must be in normal orientation, that is, the back section over the
long end of the table. Patient's head
must be on the head section. Head
section must be attached in its normal
orientation to the table's back section.
•Prior to unlocking brakes, check for
obstructions on the floor that might prevent the table from moving smoothly to
new location. Relock the brakes immediately once the final position is reached
and before commencing surgery. Table
brakes should remain locked at all times
if patient weight exceeds 500 pounds.
NOTE
With an evenly distributed patient weight
load, all table positioning functions will
operate smoothly and quietly with a
patient weight of up to 1,000 pounds.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg positioning.
NOTE
To prevent table damage, a safety interlock limits lateral tilt positioning to 20° if
Trendelenburg positioning exceeds 20°.
An audible alarm will sound.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
lateral tilt positioning.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the back section from going above horizontal if the
kidney lift is not all the way down. If the
kidney lift is raised, an audible alarm will
sound when the BACK UP button is
pressed.
NOTE
NOTE
To prevent table damage, a safety interlock limits Trendelenburg positioning to
20° if lateral tilt positioning exceeds 20°.
An audible alarm will sound.
If back section is below horizontal,
BACK UP function is limited to 800
pound patient weight.
SPECIAL USER ATTENTION
WARNING
The Leg section may hit the table base
or the floor if both the leg and elevation
systems are placed in their full down
position.
NOTE
When REFLEX button is activated, if
Kidney Bridge is up, the back section
will not go above horizontal.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the kidney lift
from going up if the back section is
raised above horizontal. An audible
alarm will sound.
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
Page 7
NOTE
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
NOTE
The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any function.
•If the Emergency Brake Release Valve
has been operated, the BRAKE UNLOCK button on the pendant control will
have to be pressed before brakes will
lock again.
NOTE
To make the Back Section easier to
handle, remove the Head Section and
X-ray Top prior to removing the Back
Section.
WARNING
CAUTION
The Back Section/Kidney Lift safety
interlock is not operational when the
emergency back-up control switches
are used.
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake system timer. The switch must be held until
the brakes are completely locked, approximately 10 seconds.
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
Page 8
SPECIAL USER ATTENTION
NOTE
Normal table top position is with the
head (and back) section over the power
cord end of the base.
WARNING
Always lock the table top in position
after rotation. DO NOT rotate the top
with an unevenly distributed patient
weight load as instability may result.
WARNING
•Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes
are set before transferring the patient.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table
or an alternate approved replacement.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
•Exercise caution with the table top
rotated 90° to the base since an improperly distributed patient load may cause
the table to be tipped over. A table
support rod is required for 90° positioning. See Specialty Positioning.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
SPECIAL USER ATTENTION
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
CAUTION
Page 9
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
Page 10
SECTION I INTRODUCTION
HEAD
SECTION
SIDE
RAIL
HEAD SECTION
LOCKING KNOB
SERVICE
ACCESS COVER
REMOVABLE
BACK SECTION
PENDANT CONTROL
STORAGE BRACKET
TOP ROTATION
POWER
CORD
HANDLE
PENDANT
CONTROL
SEAT
SECTION
REMOVABLE
LEG SECTION
LEG SECTION
RELEASE LEVER
(RECESSED)
FLOOR/LOCK
BRAKE (4)
MAIN POWER
SWITCH
Figure 1-1. 6700B HERCULES
1-1. General
SKYTRON’s 6700B Hercules Surgical Table is an
electro-hydraulically operated, general purpose surgical table. See figure 1-1.
The electro-hydraulic positioning functions operated by the hand-held, push button, pendant control
unit are: Trendelenburg, lateral tilt, back section,
elevation, leg section, flex/reflex, kidney lift, return
to level, beach chair and the floor lock/brake system.
Manual controls are provided for head section positioning, table top rotation, emergency brake release, back section removal and leg section removal.
EMERGENCY
BRAKE RELEASE
cord with a three prong, hospital grade plug. The
main power ON/OFF switch is located on the
electrical panel on the front edge of the table base.
See figure 1-2.
The battery charging indicator and foot control
connector are also located on the electrical panel.
1-2. Power Requirements
The 6700B Hercules Surgical Table requires a
120VAC, 60 Hz electrical power supply. The table
is equipped with a removable 15 foot long power
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
Figure 1-2. Electrical Panel
FOOT CONTROL
CONNECTOR
1-3. Pendant Control Unit
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TABLE UP
(BRAKE LOCK)
LATERAL
TILT LEFT
BACK UP
LEG UP
FLEX
KIDNEY
LIFT UP
RETURN
TO LEVEL
LATERAL
TILT RIGHT
BACK DOWN
TABLE DOWN
LEG DOWN
REFLEX
KIDNEY
LIFT DOWN
BEACH
CHAIR
BRAKE
UNLOCK
REVERSE
TRENDELENBURG
TRENDELENBURG
Page 11
The hand-held pendant control unit (figure 1-3) has
a non-slip rubber cover which assures a positive
grip during use. A spring clip hanger is located on
the back of the control for storage. When the
Pendant Control is not in use, it should be stored on
a convenient side or end rail. A bracket is located
under the table top next to the pendant control
connector for storage of the Pendant Control when
the table is not in use and during cleaning. See
figure 1-4.
POWER
SIDE RAIL
CLIP
INDICATOR
FUNCTION
BUTTONS
The function push buttons are identified with abbreviated descriptions for all functions. See figure
1-5. When illuminated the Trendelenburg and table
up buttons are red, the remaining buttons are all
green.
PENDANT CONTROL
STORAGE BRACKET
Figure 1-4. Pendant Control Storage Bracket
Figure 1-3. Pendant Control Unit
BACK
SECTION
TOP ROTATION
HANDLE
SEAT
SECTION
Figure 1-5. Function Buttons
1-4. Floor Lock/Brake System
The floor lock/brake system consists of four selfleveling, hydraulic brake cylinders which raise and
support the table base off from the casters. Press
the TABLE UP button on the pendant control to set
the table’s brakes. An electronic timer will activate
the brake system until the brakes are completely
set, approximately 8-10 seconds.
NOTE
Activating any function button will activate the brake system. Using the TABLE
UP function to set the brakes provides
a visual assurance that the brakes are
locked without altering the table position, except when emergency brake is
released.
Page 12
2-1. Electrical Power
SECTION II OPERATION
The 6700B table will operate on either 120 VAC or
battery power.
WARNING
Prior to operating the table, observe
all table caution labels and review the
SPECIAL USER ATTENTION section in the front of this manual.
WARNING
Possible explosion hazard exists if table
is used in the presence of FLAMMABLE
ANESTHETICS.
NOTE
b. Activate the POWER SWITCH located on the
electrical panel. The switch will illuminate.
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
TWIST TO
LOCK
FOOT CONTROL
CONNECTOR
EQUALIZATION
GROUNDING TERMINAL
Figure 2-1. Electrical Panel
An equalization terminal is located under the main power panel. This is provided as an alternate pathway to reduce the risk of static shock hazards.
Always follow recommended grounding procedures to ensure pattient and
staff safety.
2-2. AC 120V Operation
Use the following procedures to operate the table
on 120 VAC power.
a. Make sure the Power cord is securely attached to the table. To install the power cord, align
the cord connector with the base connector, insert
the cord and twist clockwise to lock the cord into
the connector. See figure 2-1. Plug the cord into a
properly grounded, Hospital Grade, 120 VAC outlet. Make sure the power cord is routed to the outlet
to prevent it from being in the way of operating
personnel.
The pendant control buttons and the green AC
120V, POWER indicator light located in the upper
right corner of the pendant control will illuminate.
See figure 2-2.
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
Figure 2-2. Pendant Control
c. The table is now ready for 120VAC operation.
Page 13
2-3. Battery Operation
a. Make sure the Battery Indicator and Main
Power Switch indicator lights, on the electrical
panel, are OFF. See figure 2-2. If the indicator
lights are ON, turn AC120V operation OFF with the
main power switch.
NOTE
The table will operate correctly on battery power with the power cord connected to a wall outlet or disconnected.
b. Press the BATT button on the hand-held
pendant control. The pendant control buttons and
the red BATTERY indicator light, located in the
upper right corner of the pendant control, will illuminate.
2-4. Automatic Shut-Off
a. To prevent unnecessary discharge of the
battery, a timer is built into the battery circuit. This
timer will automatically shut the battery power OFF
after 1½ hours of table inactivity.
b. To turn the table ON again, press the BATT
button on the pendant control, the pendant control
buttons and the red indicator light will illuminate.
NOTE
Turning the Main Power Switch ON will
change the table operation to 120 VAC
power.
c. The table is now ready for BATTERY opera-
tion.
d. To extend the battery charge life, turn the
BATTERY power OFF with the pendant control
when the table is not going to be used.
NOTE
Battery Operation must be turned OFF
at the pendant control. It cannot be
turned OFF using the main power
switch.
Page 14
2-5. Charging the Battery
Batteries should be charged:
• When the table is placed into initial
service
• As indicated by Battery Indicator
• Every week under normal service
conditions
a. Battery Indicator The Battery Indicator
consists of ten lighted bars, 3 red, 4 yellow and 3
green. See figure 2-3. Each bar represents a
percentage of the battery charge condition. When
all ten bars are illuminated, the batteries are fully
charged. The following list shows the battery
charge level as indicated by the lighted bars;
b. If the battery needs to be charged when
operating the table on battery power, the red indicator light on the pendant control will begin to blink.
NOTE
When the red light starts to blink (indicating low power in battery) the table
will operate for approximately 5 continuous minutes, typically long enough
to use the table for the rest of the day.
During charging, the bars will light in sequence to
the respective charge level, turn off and light in
sequence again.
The charging system operates ONLY
when the table is in AC120V operation
mode.
c. To recharge the battery, make sure the power
cord is connected, plugged into a 120VAC wall
outlet and the main POWER SWITCH - ON.
NOTE
The table can be operated on 120VAC
power while the battery is being recharged.
d. A full battery charge will last approximately 2
weeks under normal operating conditions. However, it is recommended to charge the batteries at
the end of each week to establish a normal routine
protocol. Lead acid batteries last longer if they are
not permitted to fully discharge. The table features
(2) 12 volt, sealed, lead acid batteries which require
no manual maintenance. Lead acid gel batteries,
under a proper charging program, feature an approximate normal life of 4 years.
BATTERY
INDICATOR
Figure 2-3. Battery Indicator
NOTE
If the table is stored for a period greater
than 6 months, the batteries should be
removed and stored in a dry, clean condition at a storage temperature of 68° F
(20° C). Batteries should be recharged
every 6 months of product storage.
2-6. Positioning Functions
The hand-held pendant control (figure 2-4) activates the following table functions:
TRENDELENBURG
REVERSE
LATERAL
TILT LEFT
BACK UP
TABLE UP
(BRAKE LOCK)
LEG UP
FLEX
KIDNEY
LIFT UP
RETURN
TO LEVEL
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TRENDELENBURG
LATERAL
TILT RIGHT
BACK DOWN
TABLE DOWN
LEG DOWN
REFLEX
KIDNEY
LIFT DOWN
BEACH
CHAIR
BRAKE
UNLOCK
Figure 2-4. Pendant Control Function Buttons
a. Floor Lock/Brake System. To activate the
brakes without affecting table positioning, press
the TABLE UP button. See figure 2-5. The elevation cylinder will not function until the brakes are
completely extended.
TABLE UP
(BRAKE LOCK)
Page 15
Press the BRAKE UNLOCK button on the pendant
control to release the four self-leveling brake feet in
order to move the table. See figure 2-5. The brake
delay circuit automatically retracts the brake system. It takes approximately 7-8 seconds to totally
release the system.
WARNING
•Do Not unlock brakes when patient is
on the table. An uneven patient weight
load may cause instability.
•If circumstances demand table brakes
to be unlocked, the patient must be
centered and evenly distributed on the
table top (i.e. supine or prone position)
with the table lowered to its lowest height
position. The maximum patient weight
should not exceed 500 pounds. Table
top rotation must be in normal orientation, that is, the back section over the
long end of the table. Patient's head
must be on the head section. Head
section must be attached in its normal
orientation to the table's back section.
•Prior to unlocking brakes, check for
obstructions on the floor that might prevent the table from moving smoothly to
new location. Relock the brakes immediately once the final position is reached
and before commencing surgery. Table
brakes should remain locked at all times
if patient weight exceeds 500 pounds.
BRAKE
UNLOCK
Figure 2-5. Brake System Activation
NOTE
With an evenly distributed patient weight
load, all table positioning functions will
operate smoothly and quietly with a
patient weight of up to 1,000 pounds.
Page 16
b. Trendelenburg. To place the table in a
Trendelenburg (head down) position, press the
TREND button (figure 2-6). To place the table in a
reverse Trendelenburg (head up) position, press
the REV TREND button. Trendelenburg positioning of up to 30° may be obtained if lateral tilt
positioning is less than 20°.
NOTE
To prevent table damage, a safety interlock limits Trendelenburg positioning to
20° if lateral tilt positioning exceeds 20°.
An audible alarm will sound.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg positioning.
c. Lateral Tilt. To achieve lateral tilt right (as
viewed from the head end of the table), press the TILT
RIGHT button (figure 2-7). To achieve lateral tilt left,
press the TILT LEFT button. Tilt of up to 30° may be
obtained if Trendelenburg positioning is less than
20°.
NOTE
To prevent table damage, a safety interlock limits lateral tilt positioning to 20° if
Trendelenburg positioning exceeds 20°.
An audible alarm will sound.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
lateral tilt positioning.
TRENDELENBURG
REVERSE
30˚
30˚
TRENDELENBURG
Figure 2-6. Trendelenburg Positioning
LATERAL
TILT LEFT
30˚30˚
LATERAL
TILT RIGHT
Figure 2-7. Lateral Tilt Positioning
Page 17
d. Back Section. To raise the back section,
press the BACK UP button (figure 2-8). The back
section will raise up to 90° above horizontal. To
lower the back section, press the BACK DOWN
button. The back section will go down to 40° below
horizontal.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the back section from going above horizontal if the
kidney lift is not all the way down. If the
kidney lift is raised, an audible alarm will
sound when the BACK UP button is
pressed.
NOTE
If back section is below horizontal,
BACK UP function is limited to 800
pound patient weight.
e. Elevation. To raise table top, press the
TABLE UP button (figure 2-9). The table will lift a
patient weight of 1,200 pounds up to a maximum
height of 44" (48" with X-Ray top and 2" pad). To
lower the table top, press the TABLE DOWN
button. The table top will go down to a minimum
height of 23".
TABLE
UP
44"
23"
TABLE
DOWN
BACK UP
90˚
40˚
BACK DOWN
Figure 2-8. Back Section Positioning
Figure 2-9. Elevation Function
Page 18
f. Leg Section. To lower the leg section, press
the LEG DOWN button (figure 2-10). The leg
section will go down to 100° below horizontal. To
raise the leg section, press the LEG UP button.
The leg section will go up to 20° above horizontal.
WARNING
The Leg section may hit the table base
or the floor if both the leg and elevation
systems are placed in their full down
position.
LEG UP
LEG DOWN
g. Flex Positioning. To place the table top in a
flex position from horizontal, press the FLEX button
(figure 2-11). To return the table top to a horizontal
position or into a reflex position, press the LEVEL
or REFLEX button.
NOTE
When REFLEX button is activated, if
Kidney Bridge is up, the back section
will not go above horizontal.
FLEX
REFLEX
20˚
100˚
Figure 2-10. Leg Section Positioning
Figure 2-11. Flex/Reflex Positioning
Page 19
h. Kidney Lift. To raise the built-in kidney lift,
press the KIDNEY UP button (figure 2-12). Up to
5.75 inches of lift can be achieved. Press the
KIDNEY DOWN button to lower the kidney lift.
KIDNEY UP
6"
KIDNEY DOWN
Figure 2-12. Kidney Lift Positioning
NOTE
j. Beach Chair. To place the top in the beach
chair position from a level position, press the
BEACH CHAIR button (figure 2-14). The back
section will raise, the leg section will lower and the
Trendelenburg positioning will function simultaneously. The functions will stop when Trendelenburg reaches it limit.
BEACH
CHAIR
To prevent damage to the kidney lift, a
safety interlock prevents the kidney lift
from going up if the back section is
raised above horizontal. An audible
alarm will sound .
i. Return To Level. To return the table top to a level
position, press the LEVEL button (figure 2-13).
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
Figure 2-14. Beach Chair Positioning
RETURN
TO LEVEL
Figure 2-13. Return To Level
Page 20
2-7. Emergency Back-up Controls
a. The emergency back-up control switches are
located under the access door on the service
access cover in the table base. See figure 2-15.
FUNCTION CONTROL
ACCESS DOOR
Figure 2-15. Emergency Controls Location
b. In the event of either a power failure or a
problem with the hand-held pendant control, the
table can be operated using the emergency backup switches. Simply push the desired emergency
switch in the appropriate direction to operate the
table functions. See figure 2-16.
NOTE
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
c. Switches are provided for Trendelenburg,
lateral tilt, back section, elevation, leg section,
kidney down and brake lock. These switches are
spring-loaded so they return to the neutral or center
position when released.
2-8. Emergency Brake Release.
In case of a power failure or an electrical problem
within the table, the emergency brake release
system can be used to move the table. The control
knob for this function is located on the side of the
table base and is identified by an EMERGENCY
BRAKE RELEASE label. Turn the knob counterclockwise to release the brakes. See figure 2-17.
CAUTION
The Back Section/Kidney Lift safety
interlock is not operational when the
emergency back-up control switches
are used.
REV
TILT
TREND
TRENDTILT
BACKUPTABLEUPLEGUPBRAKE
RIGHT
BACK
DOWN
TABLE
DOWN
LEFT
LEG
DOWN
LOCK
KIDNEY
DOWN
Figure 2-16. Emergency Back-Up Controls
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake system timer. The switch must be held until
the brakes are completely locked, approximately 10 seconds.
WARNING
•Do Not unlock brakes when patient is
on the table. An uneven patient weight
load may cause instability.
•If circumstances demand table brakes
to be unlocked, the patient must be
centered and evenly distributed on the
table top (i.e. supine or prone position)
with the table lowered to its lowest height
position. The maximum patient weight
should not exceed 500 pounds. Table
top rotation must be in normal orientation, that is, the back section over the
long end of the table. Patient's head
must be on the head section. Head
section must be attached in its normal
orientation to the table's back section.
HEAD
SECTION
FOOT/LEG
SECTION
LOCKING
KNOB
LOCKING
KNOB
•Prior to unlocking brakes, check for
obstructions on the floor that might prevent the table from moving smoothly to
new location. Relock the brakes immediately once the final position is reached
and before commencing surgery. Table
brakes should remain locked at all times
if patient weight exceeds 500 pounds.
Page 21
HEAD SECTION
RELEASE BAR
SERVICE ACCESS
COVER
POWER
CORD
EMERGENCY
BRAKE RELEASE
BRAKE (4)
Figure 2-17. Emergency Brake Release
NOTE
The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any function.
•If the Emergency Brake Release Valve
has been operated, the UNLOCK button
on the pendant control will have to be
pressed before brakes will lock again.
Figure 2-18. Head Section Adjustment
b. By loosening two locking knobs beneath the
back section, an additional 1.5" of longitudinal adjustment can be achieved. If desired, the head
section may be removed by loosening the locking
knobs and pulling it straight out of the back section.
The 6700B Hercules Table has the capability of
attaching the head section to the leg section for use
as a foot extension ONLY. Do Not reverse the
patient on the table without first consulting with
SKYTRON.
Two locking knobs are located on the inside of the
leg section for securing the head section. See
figure 2-19.
2-9. Head Section
a. A quick release positioning bar located under
and to the front of the head section (figure 2-18) is
used to raise or lower the head section. Pull the
release bar toward the head end to allow the
section to pivot up or down. Positioning from 60°
above horizontal to 90° below horizontal in 15°
increments is available. Release the bar to lock the
head section in position.
Figure 2-19. Repositioning Head Section
(for use as a Foot Extension)
Page 22
2-10. Leg and Back Section Removal.
a. The Leg Section and the Back Section on the
6700B Hercules tables are both removable. See
figure 2-20 (Leg Section shown).
b. To remove either section, level the table top,
simultaneously depress both release levers and
pull the section out. Press the LEG DOWN or
BACK DOWN button on the pendant control to
position the attachment pins down and out of the
way.
NOTE
To make the Back Section easier to
handle, remove the Head Section and
X-ray Top prior to removing the Back
Section.
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
2-11. Table Top Rotation.
NOTE
Normal table top position is with the
head (and back) section over the power
cord end of the base.
FOOT/LEG SECTION
LOCKING
LEVERS
Figure 2-20. Leg Section Removal
c. To install the Leg Section, press and hold the
LEG UP button until the attachment pins completely stop. Install the section on the pins. Level
the table top and pull out on the section to make
sure the release levers are completely locked.
d. To install the Back Section if the attachment
pins are not aligned, press and hold the BACK
DOWN button until the attachment pins completely
stop (40° down). Press REV TREND to bring the
pins up to a level position.
a. The table top can be horizontally rotated 210°
without having to rotate the entire table. Refer to
figure 2-21. To rotate the top, turn the TOP ROTATION LOCK HANDLE counterclockwise to release, grasp the table by the head end and rotate
the top counterclockwise. Lock the top in position
by tightening the TOP ROTATION LOCK HANDLE
clockwise.
WARNING
Always lock the table top in position
after rotation. DO NOT rotate the top
with an unevenly distributed patient
weight load as instability may result.
TOP ROTATION
HANDLE
Figure 2-21. Top Rotation
b. The use of the optional support rod allows the
table top to be rotated 90° from the base. See figure
2-22.
WARNING
•Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes
are set before transferring the patient.
Page 23
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
•Exercise caution with the table top
rotated 90° to the base since an improperly distributed patient load may cause
the table to be tipped over. A table
support rod is required for 90° positioning. See figure 2-22.
SUPPORT
ROD
Figure 2-22. 90 Degree Top Rotation
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table
or an alternate approved replacement.
2-12. Positioning
The use of certain optional accessories available
from SKYTRON further extend the positioning capabilities of the 6700B Hercules Tables. Refer to
the following "Positioning Guidelines" or contact
your SKYTRON representative for further details.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
Page 24
6700B HERCULES General Purpose Patient Positioning Guidelines
The following preventive maintenance checks and
services are recommended to ensure the serviceability and proper operation of your SKYTRON
Surgical Table, and should only be performed by
qualified SKYTRON trained personnel.
a. During normal cleaning, a general visual examination should be made checking for leaks,
loose bolts or parts, and cracked, chipped, or
missing paint. Any necessary repairs should be
made. Refer to the 6700B Maintenance Manual.
b. Semi-annually the following checks and services should be performed:
1. Check all hydraulic fittings, mini-valves and
slave cylinders for proper operation and
any signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
3. Pressure check (with a gauge) the pressure relief valve.
4. Check all mechanical adjustments and adjust as necessary.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of foot leg release levers,
(release knob early models). Lubricate as
necessary.
Remove major contaminants from the table with
disposable materials following appropriate biohazard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
Repeat cleaning procedure for all table surfaces
including the top, sides, elevation column, base
and all accessories.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
When the cleaning procedure is complete, replace
all pads and accessories as applicable.
Page 27
3-3. Service
Table maintenance can be performed by trained
maintenance personnel using SKYTRON authorized replacement parts and service techniques.
Service instructions and parts are available from
SKYTRON.
Preventive Maintenance contracts are available
through your local SKYTRON representative.
To obtain service instructions, replacement parts,
factory service or preventive maintenance contracts, contact the SKYTRON representative listed
below.
Or contact:
SKYTRON
5000 36th Street S.E.
Grand Rapids, MI 49512
1-800-SKYTRON (1-800-759-8766)
Fax. 1-616-957-5053
Remove pendant control from table side rail and
apply cleaning solution to the pendant control and
cord.
Use a clean cloth dampened with water to remove
cleaning solution.
Use another clean damp cloth to remove any
remaining residue.
Install pendant control on side rail for storage when
cleaning procedure is complete.
5/05
INTRODUCTION
This manual contains the exploded views and replacement parts lists for the serviceable components of the SKYTRON Model 6700B Surgical Table.
Each serviceable part in these exploded views is identified by a reference number.
Use this number to locate necessary part information in the parts list adjacent to the
exploded view.
Always use the complete SKYTRON part number and description when ordering
replacement parts.
Always use the complete table serial number (S.N.) when ordering replacement
parts.
Special Tools and Maintenance Items listed on page 38.
Abbreviations
As Required .......................................... A/R