Skytron 3600 Ultra Slide User manual

SURGICAL T ABLE
OPERAT ORS MANU AL
REV 6/06
3600B UltraSlide
Page 1

TABLE OF CONTENTS

Title Page
EQUIPMENT LABELS AND SPECIFICATIONS..................................................................................2
3600B UltraSlide General Purpose Surgical Table Specifications.......................................................3
SPECIAL USER ATTENTION..............................................................................................................4
SECTION I INTRODUCTION..............................................................................................................10
1-1. General.............................................................................................................................10
1-2. Power Requirements........................................................................................................10
1-3. Pendant Control Unit ........................................................................................................11
1-4. Floor Lock/Brake System .................................................................................................11
SECTION II OPERATION...................................................................................................................12
2-1. Electrical Power.................................................................................................................12
2-2. AC 120V Operation ...........................................................................................................12
2-3. Battery Operation ..............................................................................................................13
2-4. Automatic Shut-Off............................................................................................................13
2-5. Charging the Battery .........................................................................................................14
2-6. Positioning Functions ........................................................................................................15
a. Floor Lock/Brake system.............................................................................................15
b. Trendelenburg .............................................................................................................16
c. Lateral Tilt ....................................................................................................................16
d. Back Section................................................................................................................16
e. Elevation......................................................................................................................17
f. Top Slide .....................................................................................................................17
g. Leg Section..................................................................................................................18
h. Flex Positioning ...........................................................................................................18
i. Kidney Lift....................................................................................................................19
j. Return To Level ...........................................................................................................19
k. Beach Chair .................................................................................................................19
2-7. Emergency Back-up controls ............................................................................................20
2-8. Emergency Brake Release................................................................................................21
2-9. Head Section.....................................................................................................................22
2-10. Leg and Back Section Removal ........................................................................................22
2-11. Positioning.........................................................................................................................23
SECTION III MAINTENANCE.............................................................................................................26
3-1. Preventive Maintenance...................................................................................................26
3-2. Cleaning Recommendations ............................................................................................26
3-3. Service .............................................................................................................................27
REV 6/06
Page 2
TYPE B
EQUIPMENT

EQUIPMENT LABELS AND SPECIFICATIONS

INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED. INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING. IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY, THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED HOSPITAL GRADE OUTLET.
N
IPX4
V A
HZ
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS TEMPERATURE: 15˚ - 30˚ C (60˚ - 85˚ F) HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS. INDICATES SPECIAL USER ATTENTION.
POWERED BY AC VOLTAGE
POWERED BY BATTERY
BATTERY TYPE:
SEALED LEAD ACID 12V, VALVE REGULATED 16AH, 10HR (530W/10MIN)
15 AMP INTERNAL CIRCUIT BREAKER
15A
Electrical Specifications
Power requirements Current Leakage Power Cord
120 VAC, 60Hz, 450 Watts Less than 100 micro amps 15 feet w/hospital grade connector (removable)
12-1/2"
18-1/4" 25" 23"
14-1/2"
19-3/4"
5-3/4"
3"
82"
TOP VIEW
90˚
60˚
14-1/2"
40-3/4"
SIDE VIEW
8-3/4"
5-3/4"
19"
22"
42-1/2" MAX
24" MIN
END VIEW
ENTELA CERTIFIED
TO UL2601-1 CAN/CSA601.1, IEC 60601-2-46
SERIAL NUMBER TAG

3600B UltraSlide General Purpose Surgical Table Specifications

Page 3
Page 4

SPECIAL USER ATTENTION

Prior to use, all personnel that may operate this table must be instructed in the correct opera­tional procedures. This table is designed for use by trained and qualified personnel for hu­man medical purposes only.
Initial use should not begin until after the users have been instructed by the manufacturer's representative.
A routine instructional program must be imple­mented by the facility for proper usage instruc­tions for all personnel that may operate this table.
The maximum lifting capacity of the 3600B Ultraslide table is 1,000 pounds and the maxi­mum articulation weight capacity is 800 pounds. When lifting or articulating large patients, pay close attention to the patient position as well as the positioning guidelines and limitations listed in the operation instructions.
The extreme positioning capabilities of the 3600B UltraSlide Table requires special attention for pos­sible interference points when using multiple func­tion positioning. As with the operation of any surgical table, a certain amount of care should be exercised to position the patient safely. Although the thick pads and sheets substantially protect the patient, pinch points, located at the joints of the top section should always be considered. BE SURE THAT THE ARMS, HANDS AND FINGERS OF THE PATIENT AND THOSE OF THE OPERAT­ING ROOM PERSONNEL ARE CLEAR OF ALL MOVING PARTS BEFORE MOVING THE TABLE. Proper restraints should always be used for patient safety.
In general, common sense will dictate when there is a potential hazard.
The following precautions should be reviewed by all personnel prior to operating the table.
WARNING
Indicates a possibility of personal injury.
CAUTION
Indicates a possibility of damage to equipment.
NOTE
Indicates important facts or helpful hints.
Do not use worn or damaged accessories, they represent an injury hazard.
Remove possible obstacles before lowering or tilting the operating table
Certain accessories such as the Uro-Drain Tray, Armboards and X-Ray top can be damaged when changing the position of the table top sections. Always look first to see if a desired movement is going to interfere with any accessories in use.
The operator has the ultimate responsibility of preventing damage to the table and surrounding equipment or possible injury to the patient or staff.
The operator must ensure proper positioning is maintained to prevent compromising respiration, nerve pathways or circulation.
Do not place objects on the base of the table, a danger of damage exists during positioning.
Use caution when articulating the table top, pinch hazards exist.
SPECIAL USER ATTENTION
NOTE
Activating any function button will acti­vate the brake system. Using the TABLE UP function to set the brakes provides a visual assurance that the brakes are locked without altering the table posi­tion, except when emergency brake is released.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
Page 5
NOTE
Turning the Main Power Switch ON will change the table operation to 120 VAC power.
NOTE
When the red light starts to blink (indi­cating low power in battery) the table will operate for approximately 5 con­tinuous minutes, typically long enough to use the table for the rest of the day.
NOTE
NOTE
An equalization terminal is located un­der the main power panel. This is pro­vided as an alternate pathway to re­duce the risk of static shock hazards. Always follow recommended ground­ing procedures to ensure patient and staff safety.
NOTE
The table will operate correctly on bat­tery power with the power cord con­nected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF at the pendant control. It cannot be turned OFF using the main power switch.
The charging system operates ONLY when the table is in AC120V operation mode.
NOTE
The table can be operated on 120VAC power while the battery is being re­charged.
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean con­dition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
Page 6
SPECIAL USER ATTENTION
WARNING
•DO NOT unlock brakes when a patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top slide must be centered (indi­cated by a green LED light on the pendant control) prior to unlocking brakes. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for obstructions on the floor that might prevent the table from moving smoothly to new location. Relock the brakes im­mediately once the final position is reached and before commencing sur­gery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme lateral tilt positioning.
NOTE
If the table top is slid toward the foot end, the back section will not go below horizontal. An audible alarm will sound.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the back sec­tion from going more than 45° above horizontal if the kidney lift is not all the way down. An audible alarm will sound.
NOTE
If the leg section is positioned more than 45° below horizontal, the top will not slide toward the head end. An audible alarm will sound.
NOTE
With an evenly distributed patient weight load, all table positioning functions will operate smoothly and quietly with a patient weight of up to 800 pounds.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme Trendelenburg positioning.
NOTE
If the leg section is below horizontal, slide toward head is limited to 7-1/2"
NOTE
If the back section is positioned below horizontal, the top will not slide toward the foot end. An audible alarm will sound.
SPECIAL USER ATTENTION
Page 7
CAUTION
The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in their full down position.
NOTE
If the top is slid toward the head end less than 7-1/2", the leg section will only go down 45°. An audible alarm will sound. If the top is slid toward the head end more than 7-1/2", the leg section will not go below horizontal.
NOTE
If the top is slid toward the foot end, the FLEX function will not operate. An audible alarm will sound.
CAUTION
The safety interlock system is not op­erational when the emergency back­up control switches are used.
CAUTION
The EMERGENCY BRAKE LOCK switch does not activate the brake sys­tem timer. The switch must be held until the brakes are completely locked, approximately 10 seconds.
NOTE
The emergency back-up control switches will function when the table is operating on 120VAC power, battery power, or turned off.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the kidney lift from going up if the back section is 45° above horizontal. An audible alarm will sound .
NOTE
Elevation and brake system functions are not affected by the return to level function.
NOTE
The table top must be centered or slid toward the foot end for the Beach Chair function to operate.
NOTE
The Emergency Brake Release Valve must be closed and tightened (clock­wise) before activating any function.
•If the Emergency Brake Release Valve has been operated, the UNLOCK button on the pendant control will have to be pressed before brakes will lock again.
NOTE
To make the Back Section easier to handle, remove the Head Section, pad and X-ray top prior to removing the Back Section. Remove the pad and X­ray top prior to removing the Leg Sec­tion.
Page 8
SPECIAL USER ATTENTION
WARNING
Ensure that the Leg and Back sections are properly engaged and secured to pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled for proper orientation. The Leg section cannot be installed on the Back section pins.
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and car­diac defibrillator monitors.
WARNING
Certain accessories may limit weight capacities. Check with your SKYTRON representative.
NOTE
Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedure.
WARNING
Always follow OSHA blood-borne patho­gens standards for protective clothing, including gloves, masks and eye pro­tection when cleaning the surgical table.
WARNING
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replace­ment.
CAUTION
Thoroughly read and follow the manufacturer's directions for all clean­ing fluids. DO NOT use cleaners con­taining phenolics.
CAUTION
When using spray cleaners DO NOT spray fluids directly into electrical re­ceptacles or micro switches.
CAUTION
Before replacing pads on the table, make sure the pads and all mating surfaces are completely dry. Moisture trapped between the pads and mating surfaces may cause distortion of table tops.
Page 10

SECTION I INTRODUCTION

SIDE RAIL
HEAD SECTION
LOCKING KNOB
ACCESS COVER
HEAD
SECTION
PENDANT CONTROL STORAGE BRACKET
SERVICE
POWER
CORD
REMOVABLE
BACK SECTION
PENDANT CONTROL
SEAT
SECTION
REMOVABLE
LEG SECTION
LEG SECTION
RELEASE LEVER
(RECESSED)
FLOOR/LOCK
BRAKE (4)
MAIN POWER
SWITCH
Figure 1-1. 3600B UltraSlide
1-1. General
SKYTRON’s 3600B UltraSlide Surgical Table is an electro-hydraulically operated, general purpose sur­gical table. See figure 1-1.
The electro-hydraulic positioning functions oper­ated by the hand-held, push button, pendant con­trol unit are: Trendelenburg, lateral tilt, back sec­tion, elevation, leg section, top slide, flex/reflex, kidney lift, return to level, beach chair and the floor lock/brake system.
Manual controls are provided for head section positioning, emergency brake release, back sec­tion removal and leg section removal.
EMERGENCY
BRAKE RELEASE
main power ON/OFF switch is located on the electrical panel on the front edge of the table base. See figure 1-2.
The battery charging indicator and foot control connector are also located on the electrical panel.
1-2. Power Requirements
The 3600B UltraSlide Surgical Table requires a 120VAC, 60 Hz electrical power supply. The table is equipped with a removable 15 foot long power cord with a three prong, hospital grade plug. The
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
Figure 1-2. Electrical Panel
FOOT CONTROL
CONNECTOR
1-3. Pendant Control Unit
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TABLE UP
(BRAKE LOCK)
LATERAL
TILT LEFT
BACK UP
LEG UP
SLIDE HEAD
FLEX
KIDNEY LIFT UP
RETURN
TO LEVEL
BATTERY
LATERAL TILT RIGHT
BACK DOWN TABLE DOWN
LEG DOWN
SLIDE FOOT
REFLEX KIDNEY
LIFT DOWN BEACH
CHAIR BRAKE
UNLOCK
REVERSE
TRENDELENBURG
TRENDELENBURG
Page 11
The hand-held pendant control unit (figure 1-3) has a non-slip rubber cover which assures a posi­tive grip during use. A spring clip hanger is located on the back of the control for storage. When the Pendant Control is not in use, it should be stored on a convenient side or end rail. A bracket is located under the table top next to the pendant control connector for storage of the Pendant Control when the table is not in use and during cleaning. See figure 1-4.
POWER
SIDE RAIL
CLIP
INDICATOR
FUNCTION
BUTTONS
The function push buttons are identified with ab­breviated descriptions for all functions. See figure 1-5. When illuminated the Trendelenburg and table up buttons are red, the remaining buttons are all green.
PENDANT CONTROL
STORAGE BRACKET
Figure 1-3. Pendant Control Unit
BACK
SECTION
Figure 1-4. Pendant Control Storage Bracket
SEAT
SECTION
Figure 1-5. Function Buttons
1-4. Floor Lock/Brake System
The floor lock/brake system consists of four self­leveling, hydraulic brake cylinders which raise and support the table base off from the casters. Press the TABLE UP button on the pendant control to set the table’s brakes. An electronic timer will activate the brake system until the brakes are completely set, approximately 8-10 seconds.
NOTE
Activating any function button will acti­vate the brake system. Using the TABLE UP function to set the brakes provides a visual assurance that the brakes are locked without altering the table posi­tion, except when emergency brake is released.
Page 12

SECTION II OPERATION

2-1. Electrical Power
The 3600B table will operate on either 120 VAC or battery power.
WARNING
Prior to operating the table, observe all table caution labels and review the SPECIAL USER ATTENTION section in the front of this manual.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR TWIST TO LOCK
FOOT CONTROL
EQUALIZATION
GROUNDING TERMINAL
Figure 2-1. Electrical Panel
CONNECTOR
NOTE
An equalization terminal is located un­der the main power panel. This is pro­vided as an alternate pathway to re­duce the risk of static shock hazards. Always follow recommended ground­ing procedures to ensure patient and staff safety.
2-2. AC 120V Operation
Use the following procedures to operate the table on 120 VAC power.
a. Make sure the Power cord is securely at­tached to the table. To install the power cord, align the cord connector with the base connector, insert the cord and twist clockwise to lock the cord into the connector. See figure 2-1. Plug the cord into a properly grounded, Hospital Grade, 120 VAC out­let. Make sure the power cord is routed to the outlet to prevent it from being in the way of operating personnel.
The pendant control buttons and the green AC 120V, POWER indicator light located in the upper right corner of the pendant control will illuminate. See figure 2-2.
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
Figure 2-2. Pendant Control
b. Activate the POWER SWITCH located on the
electrical panel. The switch will illuminate.
c. The table is now ready for 120VAC operation.
Page 13
2-3. Battery Operation
a. Make sure the Main Power Switch indicator
light, on the electrical panel, are OFF. See figure 2-2. If the indicator light is ON, turn AC120V operation OFF with the main power switch.
NOTE
The table will operate correctly on bat­tery power with the power cord con­nected to a wall outlet or disconnected.
b. Press the BATT button on the hand-held pendant control. The pendant control buttons, the red BATTERY indicator light, located in the upper right corner of the pendant control and the Battery Indicator on the electrical panel will illuminate.
2-4. Automatic Shut-Off
a. To prevent unnecessary discharge of the
battery, a timer is built into the battery circuit. This timer will automatically shut the battery power OFF after 1½ hours of table inactivity.
b. To turn the table ON again, press the BATT button on the pendant control, the pendant control buttons and the red indicator light will illuminate.
NOTE
Turning the Main Power Switch ON will change the table operation to 120 VAC power.
c. The table is now ready for BATTERY opera-
tion.
d. To extend the battery charge life, turn the BATTERY power OFF with the pendant control when the table is not going to be used.
NOTE
Battery Operation must be turned OFF at the pendant control. It cannot be turned OFF using the main power switch.
Page 14 2-5. Charging the Battery
Batteries should be charged:
• When the table is placed into initial service
• As indicated by Battery Indicator
• Every week under normal service conditions
a. Battery Indicator The Battery Indicator consists of ten lighted bars, 3 red, 4 yellow and 3 green. See figure 2-3. Each bar represents a percentage of the battery charge condition. When all ten bars are illuminated, the batteries are fully charged. The following list shows the battery charge level as indicated by the lighted bars;
3 green 100% -Fully charged
2 green 89%
1 green 78%
4 yellow 67%
3 yellow 56%
2 yellow 45% -Needs Charging (BATT
indicator on pendant will flash) 1 yellow 34% -Needs Charging 3 red 23% -Needs Charging
(poor performance) 2 red 12% -Needs Charging
(intermittent performance) 1 red 1% -Needs Charging
(inoperable)
During charging, the bars will light in sequence to the respective charge level, turn off and light in sequence again.
NOTE
When the red light starts to blink (indi­cating low power in battery) the table will operate for approximately 5 con­tinuous minutes, typically long enough to use the table for the rest of the day.
NOTE
The charging system operates ONLY when the table is in AC120V operation mode.
c. To recharge the battery, make sure the power cord is connected, plugged into a 120VAC wall outlet and the main POWER SWITCH - ON.
NOTE
The table can be operated on 120VAC power while the battery is being re­charged.
d. A full battery charge will last approximately 2 weeks under normal operating conditions. How­ever, it is recommended to charge the batteries at the end of each week to establish a normal routine protocol. Lead acid batteries last longer if they are not permitted to fully discharge. The table features (2) 12 volt, sealed, lead acid batteries which re­quire no manual maintenance. Lead acid gel bat­teries, under a proper charging program, feature an approximate normal life of 4 years.
BATTERY
INDICATOR
Figure 2-3. Battery Indicator
b. If the battery needs to be charged when
operating the table on battery power, the red indi­cator light on the pendant control will begin to blink.
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean con­dition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
2-6. Positioning Functions
Page 15
The hand-held pendant control (figure 2-4) acti­vates the following table functions:
TRENDELENBURG
REVERSE
LATERAL
TILT LEFT
BACK UP
TABLE UP
(BRAKE LOCK)
SLIDE HEAD
LEG UP
FLEX
KIDNEY LIFT UP
RETURN
TO LEVEL
BATTERY
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TRENDELENBURG LATERAL
TILT RIGHT BACK DOWN
TABLE DOWN SLIDE FOOT LEG DOWN
REFLEX KIDNEY
LIFT DOWN BEACH
CHAIR BRAKE
UNLOCK
Figure 2-4. Pendant Control Function Buttons
a. Floor Lock/Brake System. To activate the
brakes without affecting table positioning, press the TABLE UP button. See figure 2-5. The elevation cylinder will not function until the brakes are completely extended.
Press the BRAKE UNLOCK button on the pendant control to release the four self-leveling brake feet in order to move the table. See figure 2-5. The brake delay circuit automatically retracts the brake sys­tem. It takes approximately 7-8 seconds to totally release the system.
WARNING
•DO NOT unlock brakes when a patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top slide must be centered (indi­cated by a green LED light on the pendant control) prior to unlocking brakes. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for obstructions on the floor that might prevent the table from moving smoothly to new location. Relock the brakes im­mediately once the final position is reached and before commencing sur­gery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
TABLE UP
(BRAKE LOCK)
BRAKE UNLOCK
Figure 2-5. Brake System Activation
NOTE
With an evenly distributed patient weight load, all table positioning functions will operate smoothly and quietly with a patient weight of up to 800 pounds.
Page 16
b. Trendelenburg. To place the table in a
Trendelenburg (head down) position, press the TREND button (figure 2-6). To place the table in a reverse Trendelenburg (head up) position, press the REV TREND button. Trendelenburg position­ing of up to 30° may be obtained.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme Trendelenburg positioning.
WARNING
To maximize patient safety, utilize proper restraint methods during extreme lateral tilt positioning.
LATERAL
TILT LEFT
LATERAL TILT RIGHT
TRENDELENBURG
REVERSE
30˚
30˚
TRENDELENBURG
Figure 2-6. Trendelenburg Positioning
c. Lateral Tilt. To achieve lateral tilt right (as
viewed from the head end of the table), press the TILT RIGHT button (figure 2-7). To achieve lateral tilt left, press the TILT LEFT button. Tilt of up to 30° may be obtained.
30˚ 30˚
Figure 2-7. Lateral Tilt Positioning
d. Back Section. To raise the back section,
press the BACK UP button (figure 2-8). The back section will raise up to 90° above horizontal. To lower the back section, press the BACK DOWN button. The back section will go down to 40° below horizontal.
NOTE
If the table top is slid toward the foot end, the back section will not go below horizontal. An audible alarm will sound.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the back sec­tion from going more than 45° above horizontal if the kidney lift is not all the way down. An audible alarm will sound.
BACK UP
9-1/2"
SLIDE HEAD
SLIDE FOOT
13-1/2"
90˚
40˚
BACK DOWN
Figure 2-8. Back Section Positioning
Page 17
f. Top Slide. To move the table top toward the
head end, press the SLIDE HEAD button. From center position, the top will slide up to 9-1/2". See figure 2-10.
NOTE
If the leg section is positioned more than 45° below horizontal, the top will not slide toward the head end. An audible alarm will sound.
NOTE
If the leg section is below horizontal, slide toward head is limited to 7-1/2"
e. Elevation. To raise table top, press the TABLE UP button (figure 2-9). The table will lift a patient weight of 1,000 pounds up to a maximum height of 42-1/2" (46-1/2" with X-Ray top and 2" pad). To lower the table top, press the TABLE DOWN button. The table top will go down to a minimum height of 24" (minus pad).
TABLE
UP
42-1/2"
24"
TABLE DOWN
To move the table top toward the foot end, press the SLIDE FOOT button. From center position, the top will slide up to 13-1/2". Slide function will stop and RETURN CTR Indicator will illuminate when table is centered.
NOTE
If the back section is positioned below horizontal, the top will not slide toward the foot end. An audible alarm will sound.
Figure 2-9. Elevation Function
Figure 2-10. Top Slide
Page 18
g. Leg Section. To lower the leg section, press
the LEG DOWN button (figure 2-11). The leg section will go down to 100° below horizontal. To raise the leg section, press the LEG UP button. The leg section will go up to 15° above horizontal.
CAUTION
The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in their full down position.
NOTE
If the top is slid toward the head end less than 7-1/2", the leg section will only go down 45°. An audible alarm will sound. If the top is slid toward the head end more than 7-1/2", the leg section will not go below horizontal.
h. Flex Positioning. To place the table top in a flex position from horizontal, press the FLEX button (figure 2-12). To return the table top to a horizontal position or into a reflex position, press the LEVEL or REFLEX button.
NOTE
If the top is slid toward the foot end, the FLEX function will not operate. An audible alarm will sound.
FLEX
REFLEX
LEG UP
LEG DOWN
15˚
100˚
Figure 2-11. Leg Section Positioning
Figure 2-12. Flex/Reflex Positioning
i. Kidney Lift. To raise the built-in kidney lift,
press the KIDNEY UP button (figure 2-13). Up to 5-3/4" of lift can be achieved. Press the KIDNEY DOWN button to lower the kidney lift.
NOTE
To prevent damage to the kidney lift, a safety interlock prevents the kidney lift from going up if the back section is 45° above horizontal. An audible alarm will sound .
KIDNEY UP
5-3/4"
KIDNEY DOWN
Page 19
RETURN
TO LEVEL
Figure 2-14. Return To Level
k. Beach Chair. To place the top in the beach
chair position from a level position, press the BEACH CHAIR button (figure 2-15). The back section will raise, the leg section will lower and the Trendelenburg positioning will function simulta­neously. The functions will stop when Trendelen­burg reaches it limit.
Figure 2-13. Kidney Lift Positioning
j. Return To Level. To return the table top to a
level position, press the LEVEL button (figure 2-14).
NOTE
Elevation and brake system functions are not affected by the return to level function.
NOTE
The table top must be centered or slid toward the foot end for the Beach Chair function to operate.
BEACH CHAIR
Figure 2-15. Beach Chair Positioning
Page 20
2-7. Emergency Back-up Controls
a. The emergency back-up control switches are
located under the access door on the service access cover in the table base. See figure 2-16.
REV
TREND
TILT
BACKUPTABLEUPLEGUPBRAKE
RIGHT
LOCK
FUNCTION CONTROL
ACCESS DOOR
Figure 2-16. Emergency Controls Location
b. In the event of either a power failure or a
problem with the hand-held pendant control, the table can be operated using the emergency back­up switches. Simply push the desired emergency switch in the appropriate direction to operate the table functions. See figure 2-17.
TREND TILT
LEFT
BACK
DOWN
TABLE DOWN
LEG
DOWN
KIDNEY
DOWN
Figure 2-17. Emergency Back-Up Controls
CAUTION
The EMERGENCY BRAKE LOCK switch does not activate the brake sys­tem timer. The switch must be held until the brakes are completely locked, approximately 10 seconds.
NOTE
The emergency back-up control switches will function when the table is operating on 120VAC power, battery power, or turned off.
CAUTION
The safety interlock system is not op­erational when the emergency back­up control switches are used.
c. Switches are provided for Trendelenburg, lateral tilt, back section, elevation, leg section, kidney down and brake lock. These switches are spring-loaded so they return to the neutral or center position when released.
2-8. Emergency Brake Release.
SERVICE ACCESS
COVER
POWER
CORD
EMERGENCY
BRAKE RELEASE
BRAKE (4)
In case of a power failure or an electrical problem within the table, the emergency brake release system can be used to move the table. The control knob for this function is located on the side of the table base and is identified by an EMERGENCY BRAKE RELEASE label. Turn the knob counter­clockwise to release the brakes. See figure 2-18.
WARNING
•DO NOT unlock brakes when a patient is on the table. An uneven patient weight load may cause instability.
•If circumstances demand table brakes to be unlocked, the patient must be centered and evenly distributed on the table top (i.e. supine or prone position) with the table lowered to its lowest height position. The maximum patient weight should not exceed 500 pounds. Table top slide must be centered (indi­cated by a green LED light on the pendant control) prior to unlocking brakes. Patient's head must be on the head section. Head section must be attached in its normal orientation to the table's back section.
•Prior to unlocking brakes, check for obstructions on the floor that might prevent the table from moving smoothly to new location. Relock the brakes im­mediately once the final position is reached and before commencing sur­gery. Table brakes should remain locked at all times if patient weight exceeds 500 pounds.
Page 21
Figure 2-18. Emergency Brake Release
NOTE
The Emergency Brake Release Valve must be closed and tightened (clock­wise) before activating any function.
•If the Emergency Brake Release Valve has been operated, the UNLOCK button on the pendant control will have to be pressed before brakes will lock again.
Page 22
2-9. Head Section
a. A quick release positioning bar located under
and to the front of the head section (figure 2-19) is used to raise or lower the head section. Pull the release bar toward the head end to allow the section to pivot up or down. Positioning from 60° above horizontal to 90° below horizontal in 15° increments is available. Release the bar to lock the head section in position.
HEAD SECTION RELEASE BAR
Figure 2-19. Head Section Adjustment
b. By loosening two locking knobs beneath the
back section, an additional 2" of longitudinal ad­justment can be achieved. If desired, the head section may be removed by loosening the locking knobs and pulling it straight out of the back section.
3600B UltraSlide Table has the capability of at­taching the head section to the leg section
for use as a foot extension ONLY. Do Not reverse the patient on the table without first consulting with SKYTRON.
FOOT/LEG SECTION
LOCKING KNOB
HEAD SECTION
LOCKING KNOB
Figure 2-20. Repositioning Head Section
(for use as a Foot Extension)
2-10. Leg and Back Section Removal.
a. The Leg Section and the Back Section on the
3600B UltraSlide table are both removable. See figure 2-21 (Leg Section shown).
b. To remove either section, level the table top, simultaneously depress both release levers and pull the section out. Press the LEG DOWN or BACK DOWN button on the pendant control to position the attachment pins down and out of the way.
NOTE
To make the Back Section easier to handle, remove the Head Section, pad and X-ray top prior to removing the Back Section. Remove the pad and X­ray top prior to removing the Leg Sec­tion.
Two locking knobs are located on the inside of the leg section for securing the head section. See figure 2-20.
Page 23
FOOT/LEG SECTION
LOCKING LEVERS
Figure 2-21. Leg Section Removal
c. To install the Leg Section, press and hold the
LEG UP button until the attachment pins com­pletely stop. Install the section on the pins. Level the table top and pull out on the section to make sure the release levers are completely locked.
d. To install the Back Section if the attachment pins are not aligned, make sure the table top is centered or toward the head end, press and hold the BACK DOWN button until the attachment pins completely stop (40° down). Press REV TREND to bring the pins up to a level position and install the section on the pins. Level the table top and pull out on the section to make sure the release levers are completely locked.
WARNING
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and car­diac defibrillator monitors.
WARNING
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replace­ment.
2-11. Positioning
The use of certain optional accessories available from SKYTRON further extend the positioning capabilities of the 3600B UltraSlide Table. Refer to the following "Positioning Guidelines" or contact your SKYTRON representative for further details.
Ensure that the Leg and Back sections are properly engaged and secured to pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled for proper orientation. The Leg section cannot be installed on the Back section pins.
WARNING
Certain accessories may limit weight capacities. Check with your SKYTRON representative.
Page 24
UltraSlide 3600B Bariatric Recommended Patient Positioning
Abdominal Laparoscopic Gastric Bypass
Cysto/GYN
Upper Body Imaging
ENT
UltraSlide 3600B Positioning Guidelines
Lower Body Imaging
Lap Nissen
Urology
Page 25
Ophthalmic/ENT
Neuro (neck) Neuro (lumbar)
Nephrectomy
Shoulder Arthroscopy
Orthopedic (with Traction Accessories)
Lumbar
Endovascular (with optional 48” Carbon Fiber Leg Extension)
Page 26

SECTION III MAINTENANCE

3-1. Preventive Maintenance
The following preventive maintenance checks and services are recommended to ensure the service­ability and proper operation of your SKYTRON Surgical Table, and should only be performed by qualified SKYTRON trained personnel.
a. During normal cleaning, a general visual examination should be made checking for leaks, loose bolts or parts, and cracked, chipped, or missing paint. Any necessary repairs should be made.
b. Semi-annually the following checks and ser­vices should be performed:
1. Check all hydraulic fittings, mini-valves and slave cylinders for proper operation and any signs of leaks.
2. Check the hydraulic speed controls and adjust if necessary.
3. Pressure check (with a gauge) the pres­sure relief valve.
4. Check all mechanical adjustments and ad­just as necessary.
3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedure.
The following procedures should be followed when cleaning the surgical table between cases.
Place table top in level position prior to starting cleaning procedure.
WARNING
Always follow OSHA blood-borne patho­gens standards for protective clothing, including gloves, masks and eye pro­tection when cleaning the surgical table.
Remove major contaminants from the table with disposable materials following appropriate biohaz­ard waste disposal procedures.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of back and foot leg release levers, lubricate as necessary.
Remove all table pads and place them on a flat surface for cleaning.
CAUTION
Thoroughly read and follow the manufacturer's directions for all clean­ing fluids. DO NOT use cleaners con­taining phenolics.
Apply cleaning fluid liberally to top and sides of each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to remove all moisture.
Repeat the steps to clean the bottom of the each pad.
CAUTION
Page 27
3-3. Service
Table maintenance can be performed by trained maintenance personnel using SKYTRON autho­rized replacement parts and service techniques. Service instructions and parts are available from SKYTRON.
Preventive Maintenance contracts are available through your local SKYTRON representative.
When using spray cleaners DO NOT spray fluids directly into electrical re­ceptacles or micro switches.
Repeat cleaning procedure for all table surfaces including the top, sides, elevation column, base and all accessories.
CAUTION
Before replacing pads on the table, make sure the pads and all mating surfaces are completely dry. Moisture trapped between the pads and mating surfaces may cause distortion of table tops.
When the cleaning procedure is complete, replace all pads and accessories as applicable.
Remove pendant control from table side rail and apply cleaning solution to the pendant control and cord.
To obtain service instructions, replacement parts, factory service or preventive maintenance con­tracts, contact the SKYTRON representative listed below.
Or contact:
SKYTRON 5000 36th Street S.E. Grand Rapids, MI 49512 1-800-SKYTRON (1-800-759-8766) Fax. 1-616-957-5053
Use a clean cloth dampened with water to remove cleaning solution.
Use another clean damp cloth to remove any remaining residue.
Install pendant control on side rail for storage when cleaning procedure is complete.
5000 36th Street S.E., Grand Rapids, MI 49512
1-800-SKYTRON or 1-616-957-0500 • FAX 1-616-957-5053
SURGICAL T ABLE
PARTS MANUAL
10/05
3600B UltraSlide

INTRODUCTION

This manual contains the exploded views and replacement parts lists for the service-
able components of the SKYTRON Model 3600B UltraSlide Surgical Table.
Each serviceable part in these exploded views is identified by a reference number. Use this number to locate necessary par t information in the par ts list adjacent to the ex­ploded view .
Always use the complete SKYTR ON part number and description when ordering replacement parts.
Always use the complete table serial number (S.N.) when ordering replacement parts.
Special Tools and Maintenance Items listed on pa ge 46.
Page 1
SERIAL NUMBER TAG
Abbreviations
As Required .......................................... A/R
Optional................................................... opt
Serial Number ....................................... S.N.
Not Shown .............................................. N S
10/05
Although current at the time of publication, SKYTRON's policy of continuous development makes this manual subject to change without notice.
Loading...
+ 102 hidden pages