The Spectrum and Spectrum with Master Drug Library are intended to be used for the controlled
administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood
products, antibiotics, nutritional fluids and mixtures of required patient therapy. The intended
routes of administration consist of the following clinically acceptable routes: intravenous, arterial,
subcutaneous, intrathecal, epidural or irrigation of fluid space. The Spectrum is intended to be used
in conjunction with legally marked intravenous administration sets and medications provided by the
user.
The Spectrum and Spectrum with Master Drug Library are suitable for many user facility applications such as, but not limited to, hospitals, outpatient care areas, homecare and ambulatory care services.
The Spectrum and Spectrum with Master Drug Library are intended to reduce operator interaction
through automated programming, thereby helping to reduce errors associated with complex device
programming. Parameter programming requires trained healthcare professional confirmation of limits and drug therapy to physician's directive.
Related Documents
The following documents also pertain to the Spectrum and Spectrum with Master Drug Library:
■Operator’s Manual P/N 41018
■MDL User Manual P/N 41020
1Introduction and Safety
SIGMA Manual Number 41019
Listings & Certifications
StandardVersionDescription
UL 60601-12003UL Standard for Safety Medical Electrical
Equipment, Part 1: General Requirements for
Safety First Edition
Revision AA
CAN/CSA C22.2
NO 601.1-M90
IEC 60601-2-241998–2002Medical Electrical Equipment – Part 2 –24:
EN 60601-11990Medical Electrical Equipment Part 1: General
FCC Section 15C
ID: MCQ-50M879
FCC Section 15C
ID: MCQ-50M1538
See the SIGMA Spectrum Operator’s Manual (P/N 41018) for specific flow rate accuracy (trumpet
curve) information and electromagnetic compatibility compliance levels.
1990Medical Electrical Equipment – Part 1: General
Requirements for Safety General Instruction No
1: with UL60601-1 Supplement I
Particular Requirements for Safety of Infusion
Pumps and Controllers First Edition
Requirements for Safety Includes Amendments +
A1(93) + A2(95)
N/AFCC approved wireless transmitter/receiver
module
(wireless battery module only – SIGMA P/N
35083)
N/AFCC approved wireless transmitter/receiver
module
(wireless battery module only – SIGMA P/N
35162)
NRTL – Intertek Testing Services NA Inc. Reference Control Number 3063464
Contacting SIGMA Technical Support
Contact SIGMA, LLC for all service information at:
Telephone:1-800-356-3454
Email: techsupport@sigmapumps.com
2
SIGMA Manual Number 41019
Revision AA
Conventions
WARNING:Indicates a potentially hazardous situation which, if not avoided, could
CAUTION:Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate personal injury or property damage. This word is used to also
alert against unsafe practices.
NOTE: Additional information to clarify or strengthen wording.
Summary of Warnings and Cautions
result in death or serious injury.
WARNING:
- THIS EQUIPMENT IS NOT SUITABLE FOR USE IN THE
PRESENCE OF A FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR OR OXYGEN OR NITROUS OXIDE.
- LITHIUM BATTERY (SIGMA P/N 55075)
(LOCATION: BT1 ON PCBA P/N 60087)
DANGER OF EXPLOSION IF THE BATTERY IS INCORRECTLY
REPLACED. REPLACE ONLY WITH THE SAME OR EQUIVALENT
TYPE RECOMMENDED BY SIGMA. REFERENCE THIS MANUAL
FOR DISPOSAL INSTRUCTIONS.
- DO NOT SHORT CIRCUIT THE BATTERY TERMINALS,
DISASSEMBLE OR MODIFY BATTERY PACKS.
- DO NOT DISPOSED OF BATTERIES OR BATTERY PACKS IN
FIRE.
- ANY PUMP THAT HAS BEEN DROPPED OR IS OTHERWISE
VISIBLY DAMAGED MUST NOT BE USED. Discontinue use
immediately and refer to a qualified Service Technician for inspection
and repair prior to further use.
- USE ONLY IV ADMINISTRATION SETS THAT ARE LISTED ON
SIGMA COMPATIBLE SET LISTING (see Operator’s Manual for
listing). Use of any other sets may result in malfunction or inaccurate
delivery.
- DO NOT USE IF ROOM OR PUMP TEMPERATURE IS BELOW
15°C (60°F) OR ABOVE 32°C (90°F).
3Introduction and Safety
CAUTION:
SIGMA Manual Number 41019
Revision AA
- Federal (USA) law restricts this device to sale by or on the order of a
physician or other licensed practitioner.
- Operate only on the AC line voltage for which the AC power adaptor is
marked.
- The AC power adaptor should be plugged in during storage to
maintain proper battery charge level.
- Disconnect from AC power source and remove the battery pack prior
to servicing this pump.
- Routine cleaning and periodic maintenance is needed to assure that the
pump remains safe and functional. Refer to “Service and
Maintenance” on page 26, and “Preventive Maintenance” on page 31.
- Clean up solution spills as soon as possible using a damp cloth or
sponge. Dry thoroughly before use. Keep the upstream and
downstream sensors’ tube channels clean and dry at all times during
pump operation. Do not use hard or sharp objects to clean sensor tube
channel areas. Failure to comply with this requirement may cause the
pump to fail. Refer to “Service and Maintenance” on page 26.
- Static sensitive electronics used in SIGMA infusion pumps may be
damaged by electrostatic discharge. Service technicians must follow
proper ESD procedures when working on pumps. Refer to MIL-STD1686 for protective handling procedures.
- Do not use a pen or any other sharp object to actuate the keyboard
switches. Replace torn or punctured membranes immediately, to assure
solutions won't internally damage the membrane switch and prevent
proper operation of the SIGMA Spectrum. (Replacement should only
be performed by authorized personnel).
- Use only filter sets that are specifically listed in the Operator's Manual
for use with Spectrum pumps.
NOTE: For a list of serviceable parts, contact SIGMA.
4
SIGMA Manual Number 41019
Revision AA
DESI GN AN D THE OR Y OF OPERAT ION
Case
SIGMA Spectrum consists of two case halves: front and rear case assemblies.
Front Case Assembly
The front case assembly is comprised of the following primary components:
■Front case
■Display
■Two circuit boards
The circuit boards of this assembly are fastened with four hex standoffs and four Phillips machine
screws.
Rear Case Assembly
The rear case assembly consists of the following primary components:
■Rear case
■Back flex circuit
■Power connector
■Speaker
The back flex circuit is adhered to the rear case. The power connector is assembled to the rear case.
Battery Pack
The SIGMA Spectrum employs a removable lithium ion 7.2V battery pack. This assembly consists of
the battery cells, contact pins and the two plastic case halves. The battery pack uses both cell and
battery pack protection to guard against short-circuit, over voltage and over current conditions. The
battery terminals are slightly recessed to minimize the possibility that the terminals may be shorted
on a conductive surface. The battery pack is labeled to identify it and provide handling / disposal
information as necessary for user safety.
There are three battery types:
■P/N 35724 - standard battery
■P/N 35083 - 802.11b wireless module
■P/N 35162 - 802.11g wireless module.
The standard battery (P/N 35724) and 802.11b wireless module (P/N 35083) are compatible with all
SIGMA Spectrum pumps.
The 802.11g wireless module (P/N 35162) is only compatible with wireless G compatible pumps.
Wireless G compatible pumps are identified by the letter G preceding the serial number on the pump
(see Figure 1.). Pumps that do NOT have a letter G preceding the serial number (see Figure 2.) are
not compatible with the 802.11g wireless module (P/N 35162).
5Design and Theory of Operation
SIGMA Manual Number 41019
Revision AA
Figure 1. Serial Number for Pump Compatible with the 802.11b/g Wireless Battery Module (P/N 35162)
Figure 2. Serial Number for Pump NOT Compatible with the 802.11b/g Wireless Battery Module (P/N
35162)
6
SIGMA Manual Number 41019
Revision AA
Pole Clamp
The pole clamp assembly (SIGMA P/N 35712) is attached to the rear of the pump. A thumbscrew is
used to lock the pole clamp assembly to the SIGMA Spectrum pump. The pole clamp assembly is
comprised of an anodized aluminum C-clamp and a slide mount receptacle bracket. The bracket is
attached to the C-clamp with two 1/4-20 x 5/8” Phillips machine screws.
Double Rotating Pole Clamp Assembly (Accessory)
The Double Rotating Pole Clamp Assembly (SIGMA P/N 35743) is an accessory to the Spectrum
infusion pump. The double rotating feature of the Double Rotating Pole Clamp enables the pump to
be offset on the left or right side of the single (standard) IV pole. The clamp can also be clamped to
hospital bedside rail diameters ranging from 0.75” to 1.25”. Refer to the cleaning section of this
manual for compatible cleaners. Refer to the installation manual (SIGMA P/N 41092) on how to
install the clamp on the pump. The dimension of the clamp is less than 7.0” long X2.5” wide. The
weight of the clamp is less than 1.2 lbs.
CAUTION:Always operate the pump in vertical direction with the key hole at the top of the
pump.
Tubing Channel/Pump Door
The tubing channel consists of
■Proximal air/occlusion sensor
■Pump valves and fingers
■Downstream tubing guide
These components are designed to facilitate in the proper loading of the IV tube.
Set Slide Clamp Sensor
A reflective slide clamp sensor is employed to determine whether a closed slide clamp is present in
the keyhole. If a slide clamp is present in the keyhole, the sensor LED light source illuminates and
detects the presence of the clamp. The sensor component alerts the user that a closed slide clamp is
present in the keyhole when the user presses the RUN/STOP key.
Processing Architecture
The SIGMA Spectrum employs three computational processors.
■Two of the three processors, “Sharp” (Processor Board – 60087) and “PIC” (I/O Board –
60091), form a diverse checked redundancy scheme and perform all of the safety critical
computations. The scheme is diverse because they are different processor types manufac-
7Design and Theory of Operation
SIGMA Manual Number 41019
Revision AA
tured by separate, independent manufacturers. Likewise, these processors use different software compilers and math libraries, which yields additional independence. This processor
pair uses a master/slave checked redundancy scheme where safety critical computations are
computed by the master processor and checked by the slave processor. If either processor’s
check calculation does not match its own computation or if a fault condition is detected, the
processor can independently fail the system to a safe state, which includes an audible alarm.
■The third processor is used to charge the battery (Backflex – 35149), interface to the bar-
code scanner, and pass asynchronous data through the rear case connector.
Pumping Mechanism
Physical Method
The SIGMA Spectrum uses a linear peristaltic motion to pump the fluid. That is, the tubing is repeatedly squeezed to create fluid pressure and thereby flow of fluid.
Operation
There are two phases of the pumping cycle:
■Fill phase
■Pumping phase
The fill phase allows the tubing to be filled from the proximal or “upstream” end of the pump. The
second phase of the cycle is the pumping phase. The pump consists of two valves, one of which
occludes the administration set tubing at all times. The proximal valve occludes the tubing during
the pumping phase of the cycle. Similarly, the distal valve occludes the tubing during the fill phase of
the cycle. There are two “fingers” whose function is to squeeze fluid towards the distal end of the
pump, thereby creating fluid pressure and fluid flow during the pumping phase of the cycle. During
the fill phase of the cycle, the fingers are in a position that allows maximum volume of fluid to fill
the tubing proximal to the distal valve.
Pumping Mechanism Assembly
This assembly is comprised of a cast aluminum chassis and doorplate assembly. The motor, cams,
pumping fingers, upstream sensor and downstream sensor are also part of this assembly. All components are fastened with standard, Phillips and socket head machine screws.
Motor
The motor is connected to the I/O Board (60091) through a 10-conductor, 1 mm pitch flex circuit.
Motor Drive
The motor drive circuitry will only allow the motor to run in a clockwise direction, which creates
fluid flow from the proximal end of the pump to the distal end of the pump. Both the PIC and Sharp
processors have the ability to independently remove power from the motor through independent
hardware switches, thereby halting the pump in the event that single fault condition is detected.
There are two mechanisms for detecting rotation of the motor. The first rotation sensing mechanism
is the two-channel encoder, which can sense motor angular velocity as well as rotation direction. The
second rotation sensing mechanism is the once-per-revolution hall effect sensor. These two sensing
8
SIGMA Manual Number 41019
Revision AA
mechanisms eliminate the possibility of incorrect sensing motor velocity after a detectable single
fault condition.
Motor and Gear Drive
SIGMA Spectrum employs a 16 mm DC motor, encoder and planetary gearbox. The output shaft of
the gearbox is coupled to the pump drive camshaft by two identical gears, thereby creating a 1:1
drive ratio between the camshaft and output of the gearbox. Motor speed is controlled by a closedloop pulse width modulation (PWM) drive system that derives its power directly from the battery or
9VDC source when the AC power adapter is connected.
Fluid Temperature
Physical properties of administration set tubing are subject to the effects of fluid temperature. In general, colder temperatures can cause negative flow rate errors in an uncompensated system. Similarly,
hotter temperatures can cause positive flow rate errors in an uncompensated system. The SIGMA
Spectrum employs a temperature compensation system to maintain flow rate accuracy over the operating temperature range of the device. Approximate fluid temperature is measured by two negative
temperature coefficient thermistors, one proximal to the pumping mechanism and one distal to the
pumping mechanism. The proximal sensor is housed by the upstream sensor assembly. The distal
sensor is located in the lower tubing guide, just below the distal occlusion sensor. Both sensors are
calibrated at the factory, and are used to check each other for single fault failure conditions. The output of both thermistors are filtered and fed to A/D converters for processing by Spectrum processors
for compensating pump motor speed to maintain flow rate accuracy.
Upstream Sensors
Optical Tube Sensor
The optical tube sensor consists of a light emitting diode (LED) emitting infrared light and a phototransistor. The LED and phototransistor are located at the bottom of the ultrasonic sensor housings,
facing opposite to each other. This position is marked as load point 2 and is indicated in red color
next to the “direction of flow” label inside the pump door. The optical tube sensor detects that an IV
set is loaded at load point 2 to assist in proper loading of the IV set into the pump. It must recognize
the presence or absence of the IV tube. The sensor is activated only when the door is open. If the IV
tube is properly placed in the space between the LED and phototransistor, the infrared light is diffracted and the phototransistor outputs a high voltage. If there is no IV tube present, the phototransistor receives continuous infrared light and a low voltage is output.
The software monitors the output voltage and displays a green bar with a check mark to indicate a
properly loaded tube or a red bar with loading instructions when there is no tube or if the tube is
improperly placed in the sensor. This sensor is calibrated through a biomed option.
Proximal Occlusion Detection
The proximal occlusion sensor consists of two pairs of ultrasonic transmitters and receivers. A
“pusher” connected to the door assembly assures that a constant force is exerted to push the administration set tubing against the sensor housing. The transmitted ultrasonic signal is a constant wave
9Design and Theory of Operation
SIGMA Manual Number 41019
Revision AA
(CW) signal centered at a nominal 293 KHz. This signal is generated by exciting a piezoelectric crystal near its natural resonant frequency. The ultrasonic signal propagates from the sensor housing laterally through the administration set tubing. The ultrasonic receiver is a piezoelectric crystal
matched in frequency to the transmitter. The differential output of the receiver is amplified, envelope
detected, and filtered before being presented to a 12-bit A/D converter where it is converted to a 12bit digital word. Three samples are taken for each pump revolution. The digital words are processed
by the SIGMA Spectrum processor and fed into a detection algorithm. The detection algorithm compares successive samples and looks for a trend of decreasing signal amplitude caused by a reduction
in pressure against the sensor housing. If the trend continues through a predetermined threshold over
a predetermined unit of time, a proximal occlusion is declared and a visual and audible alarm is presented to the user.
Air Sensor
The air detector uses the same sensor hardware as described in the Proximal Occlusion Detection
section. The air detection algorithm uses an “air” threshold that is set by a calibration routine at the
factory. If the signal amplitudes received by both sets of ultrasonic receivers are below the air threshold, an air alarm is declared and a visual and audible alarm is presented to the user. The SIGMA
Spectrum air sensor is designed to detect air bubbles greater than 1 inch of tubing (125–150 µl
depending on tubing calibration).
Downstream Sensors
Set Loading
The downstream sensor consists of a channel piezoelectric bridge force transducer and is located
between loading points 3 and 4 inside the pump door. This sensor has two basic functions. The first
is to acknowledge an IV administration set as being installed at load points 3 and 4. Acknowledgement is identified on the display with a third green bar and check mark. The second function is to
detect distal occlusions.
Distal Occlusion Detection
The distal occlusion sensor uses a piezoelectric bridge force transducer. A pusher connected to the
door assembly assures that a constant force is exerted to push the administration set tubing against
the sensor housing. The detection algorithm compares successive samples and looks for a trend of
increasing/decreasing signal amplitude caused by a increase/reduction in pressure against the sensor
housing. Four samples are taken for each pump revolution. The differential output of the force transducer is amplified, envelope detected, and filtered before being presented to a 12-bit A/D converter
where it is converted to a 12-bit digital word. The digital words are processed by the SIGMA Spectrum processor and fed into a detection algorithm. The detection algorithm compares the received
signal levels to thresholds set by the user selected occlusion alarm trip points. If the received signal
exceeds the trip points, a distal occlusion alarm is declared and a visual and audible alarm is presented to the user.
10
SIGMA Manual Number 41019
Revision AA
Audible Alarm Device and System
The audible alarm device is a 1-inch moving coil speaker. The speaker is driven by two separate
audio paths, one by the Sharp and one by the PIC. Either the Sharp or PIC can independently sound
an audio alarm tone if an alarm occurs or a fault is detected.
Electronic Components
AC Power Adaptor / Charger
The SIGMA Spectrum infusion pump uses an AC power adaptor capable of providing 9 VDC. The
power adaptor connects to Spectrum via a positive locking connector, which will reduce the possibility of accidental disconnection. When the AC power adaptor is plugged into an AC receptacle, the
lithium ion battery pack will be charged. Normal charger operation occurs as follows: If the battery
requires charging, the Spectrum battery charger will initially maintain the battery at a constant current of approximately 250 mA (400 mA if pump is non-wireless or wireless is turned off). After the
battery voltage reaches a nominal charge of 8.3 volts, the charger maintains this voltage until the
charging current supplied to the battery is less than 60 mA. The battery charge cycle is completed
and will terminate. There are three mechanisms employed to prevent overcharging of the battery:
■The charging microcontroller’s ability to monitor and control voltage and current applied to
the battery.
■A current clamp circuit is employed to prevent charge currents that would cause charge cir-
cuit overheating.
■Battery pack protection circuits within the battery pack itself prevent overheating and over
charging.
The adaptor side of the AC power adaptor is equipped with a Protector (P/N 45742). The Protector
is a plastic enclosure which snaps onto the adaptor. The purpose of the Protector is to protect the
adaptor from damage during use in a hospital environment. The AC power adaptor with the Protector can be used on wall outlet and pole mount power strip receptacles. The Protector comes installed
with the pump. The Protector cannot be removed from the adaptor once it is installed. The Protector
can be ordered separately as an accessory for installing on an old AC power adaptor. A separate
instruction bulletin on how to install the Protector is sent along with the Protector.
The Protector is compatible with the cleaners mentioned in this Service Manual. Refer to “SIGMA
Spectrum Cleaning” on page 26 of this Service manual for cleaning methods and compatible cleaners
that can be used on the Protector.
CAUTION:IMPROPER REMOVAL MAY DAMAGE THE AC POWER ADAPTOR. Remove
the AC power adaptor by grabbing the handle of the Protector and pulling it back
off the receptacle. Do not pull the cord to unplug the Adaptor from the receptacle.
Improper twisting or pulling of the connector or cord may damage the AC power
adaptor.
11Design and Theory of Operation
SIGMA Manual Number 41019
Revision AA
Display
Spectrum uses a highly reflective thin film transistor (HRTFT) liquid crystal display containing:
■240 X 320 pixels (240 x270 are active)
■Color HRTFT - 16 out of a palette of 262,144 possible colors
■0.2235 mm X 0.2235 mm dot pitch
■Illumination by side mounted LEDs and diffuser.
The display employs front lighting using six white LEDs. Brightness of the LED front light is controlled by a constant current source implemented on the 60091 I/O Board. The display can be easily
viewed while mounted to an IV pole, with viewing distances up to 15 feet and a minimum viewing
angle of ±30 degrees from a line perpendicular to the display plane. The display has two flex circuit
connections: The first 50-conductor flex circuit carries digital display and control data as well as
power for the display. The second 5-conductor display flex circuit carries power for the back light.
Both of these flex circuits connect to 0.5 mm flex circuit connectors on the I/O Board.
I/O Board
This 10-layer, 100% surface-mount component circuit board distributes input and output (I/O) from
all sensors (except the ultrasonic air detector) through flex circuit connections that will be discussed
in detail below. The I/O board also receives the keypad, motor and display flex circuit signals. The I/
O board hosts A/D converters and discrete I/O to interface to the Spectrum sensors.
Processor Board
This 10-layer, 100% surface mount component circuit board distributes input and output (I/O) from
the ultrasonic air detector through a 40-pin flex circuit connection (detailed below). Additionally, the
processor board hosts the Sharp processor, which controls the HRTFT display, and performs most of
the primary computations for Spectrum. The processor board also hosts the real-time clock (RTC)
chip and lithium back-up battery for the RTC, Flash and static RAM memory.
Flex Circuits
■Ultrasonic Flex — This flex circuit provides connections to two pairs of ultrasonic piezoelec-
tric crystals (receive and transmit) as well as the secondary tube temperature thermistor. The
Ultrasonic Flex is connected to the processor board (60087) through a 0.5 mm pitch, 40conductor, right angle flex connector.
■Upper Aux Flex — This flex circuit provides connections to the upper door latch and link
discrete signals, once per revolution hall effect sensor, and slide clamp sensor signals. The
upper aux flex is connected to the I/O board (60091) through a 0.5 mm pitch, 30-conductor,
vertical flex connector.
■Lower Aux Flex — This flex circuit provides connections to the lower door latch and link
discrete signals. The lower aux flex is connected to the I/O board (60091) through a 0.5 mm
pitch, 10-conductor, vertical flex connector.
■Force Sensor Flex — This flex circuit provides connections to the distal force sensor and the
primary tube temperature thermistor. The force sensor flex is connected to the I/O board
(60091) through a 0.5 mm pitch, 10-conductor, vertical flex connector.
■Backflex — This flex is connected to the I/O board (60091) through a 0.5 mm pitch, 30-
conductor, vertical flex connector. The backflex provides the following functions:
12
SIGMA Manual Number 41019
Revision AA
■ System power supply control and fuse – A 2.5 Amp surface-mount primary power fuse
is mounted on this flex circuit. Additionally, four dual battery contacts are also on this
flex circuit.
■ High-Speed UART Interface / MUX – A digital MUX allows either a smart battery
interface that includes an UART, or traditional “dumb” battery connections (battery
type ID resistor and battery thermistor).
■ Battery System Charge Control – An Atmel 8-bit microcontroller is hosted on the back-
flex. One function of this processor is to control battery charging.
■ Audio Output – The backflex provides connections to a 1-inch moving coil speaker that
is used during audio alarms.
■ Temperature Monitor – A temperature sensing thermistor is hosted on the backflex,
which provides internal case temperature to the Spectrum processors.
■ Advanced Battery Removal Detector – A magnetic reed relay is housed on the backflex,
and provides an early or advanced battery disconnection warning to the Spectrum processors.
Keypad
The keypad consists of 19 user-actuated keys. Keys 1–9 also have alpha designations A–Z (in groups
of 3 starting at the 1 key, with the exception that the 9 key has alpha designation “YZ”), a “0” key,
a decimal point, 4 soft keys, a RUN/STOP key, and an ON/OFF key. The keypad is constructed of
polyester dome keypad material with a circuit layer and adhesive layer. The keypad is edge sealed to
prevent fluid intrusion between the layers of the keypad. The keypad is connected to the I/O board
through a 10-conductor 1-mm pitch flex circuit. The following is a detailed description of the keys:
■ON /OFF: Powers up the pump.
■OK: This key accepts the information currently shown on the pump display and advances
the pump program to the next screen.
■SETUP: This key is used to access the Care Area Selection screen.
■RUN/STOP: This key signals the request to Start or Stop infusions.
■Soft keys: Four multi-function keys. Can be used in conjunction with the display to perform
any software programmable function including but not limited to Bolus operations, navigation arrows, help screens etc.
■Alphanumeric 1-9: These keys may be used to enter numeric sequences including real num-
bers with decimal point. Alphabetic strings may also be entered using the alphabetic groupings of characters found on these keys.
The infrared interface (Infrared Data Association) utilizes the IrDA protocol that can connect to personal computers and personal digital assistants (PDAs) that employ the IrDA standard. This digital
interface can transmit and receive data at rates of up to 115 thousand bits per second (115Kbs). The
IrDA interface is accessible through the SIGMA Spectrum keypad that is located on the front case
assembly. The IrDA port operates at distances of up to 3 meters assuming a direct line of sight within
a 15-degree conic section.
Power Connector
The power connector serves two functions:
■External power input
■Asynchronous data interface
13Design and Theory of Operation
SIGMA Manual Number 41019
Revision AA
Pump Specifications
Applications
■Large and small volume parenterals
■Neonatal
Administration Set
Compatible with standard gravity IV sets supplied by Baxter® and Hospira® and several other companies. (Visit www.sigmapumps.com for a full listing).
Performance
Downstream occlusion time to alarm for occlusion pressures adjusted to LOW (6 psi), MEDIUM (13
psi), and HIGH (19 psi)1:
RateLow (6 psi)Medium (13 psi)High (19 psi)
5 mL/hr<5 min.<10 min.<20 min
100 mL/hr<15 sec.<20 sec.<40 sec.
Max. Bolus0.25 mL0.50 mL0.80 mL
All safety related sensing systems are automatically checked for proper operation periodically during
setup and pump running conditions. Faults are identified by SYSTEM ERROR codes.
See the Specifications section of the Operator’s Manual for additional specifications.
Effect of Fluid Container Height
The performance of the infusion pump will be influenced by the forces of gravity on the fluid being
administered to the patient. When a fluid container is positioned above or below the patient’s administration site, pressure forces associated with the fluid’s head-height (distance measured from the center of the pumping mechanism to the top of the fluid in the source container) will cause deviations in
the nominal specification for device flow rate accuracy. The nominal head-height used for the flow
rate and the graphical presentation is 24” (61 cm). For example, the deviation from the nominal
flow rate accuracy from +24” to -20” (-51 cm) head-height has an influence of a possible -4% in
flow rate accuracy when the flow rate is set to 25 mL/hr.
Bolus Accuracy
The SIGMA Spectrum IV pump may have an optional bolus mode of operation. This feature allows
the user to perform a BOLUS SETUP action. To use this feature the pump must be programmed with
either a specific rate or a specific amount to be delivered in a certain amount of time.
2
1.With Occlusion 45 inches below the pump and alarm set to LOW, MEDIUM or HIGH.
2.Liquid container must be a vented or collapsible bag.
14
SIGMA Manual Number 41019
Revision AA
If the pump is currently operating in mL/hr delivery mode, the bolus rate value is entered in mL/hr
and the volume is entered in milliliter (mL). If the pump is operating in a non-mL/hr delivery mode
(for example mcg/kg/min), the bolus amount would be entered in mcg/kg however the ML/HR soft
key may be pressed in the setup screen to enter the bolus information in mL/hr format.
In either mode, the time is entered in minutes and seconds (min:sec). Limits are placed on the minimum and maximum amount of time for the bolus delivery. The limit constraints are contained
within the software of the Spectrum pump and are necessary to control the maximum or minimum
flow rate of the bolus infusion.
The accuracy of the bolus volume is dependent upon the resultant flow rate that is obtained from the
calculation of volume to be delivered in the time requested. For example, if the maximum bolus volume is 300 mL, the maximum flow rate is obtained with a bolus time of 18:02 (min:sec) or a flow
rate of approximately 999 mL/hr. Using this maximum bolus volume, and delivering the volume in
the shortest amount of time, the mean value of 302 mL ±5% may be expected. Whereas using a minimum bolus volume (.5 mL), and delivering the volume in a reasonably short amount of time (1 minute), the mean value of .52 mL ±16% may be expected.
Technical
■Type of pump - volumetric infusion pump.
■Operating principle - linear peristaltic
■Patient leakage: <15 micro amps, normal condition; <50 micro amps, single fault condition.
ium ion, or 7.2-volt battery pack consisting of (2) 3.7V/1.75 Ampere-hr rechargeable lithium ion.
■Alarms:
■ Air-In-Line
■ Downstream Occlusion
■ Upstream Occlusion
■ Door Open, Door Not Latched
■ Low and Depleted Battery
■ AC Applied / Removed
■ Battery Not Detected
■ Infusion Complete
■ Secondary Complete
■ Secondary Delivery
■ Primary Delivery
■ User Inactivity
■ Unload Set Caution
■ System Errors
■ Incomplete Titration Entry
■ Slide Clamp Closed
NOTE: For additional technical specifications, refer to the SIGMA Spectrum Operator’s
Manual.
15Design and Theory of Operation
SIGMA Manual Number 41019
CONFIGUR ATION OPTIO NS
The Spectrum pump has two sets of configuration options:
■User Options
■Biomed Options
NOTE:For a description of the User options, refer to the SIGMA Spectrum
Operator’s Manual.
BioMed Options
Biomed Options allow the Biomedical Technician access to certain pump functions and settings. This
section will describe the options available to the technician.
Revision AA
NOTE:Biomedical Technicians should access ONLY the options listed in this
section, unless otherwise Authorized.
To access the BioMed Options menu:
1.Press the OPTIONS soft key.
2.Select BioMed Options from the options list and press OK.
NOTE: The Bio Medical (BioMed) option menu requires an access password. This access
password may be obtained by authorized service personnel.
Contact SIGMA LLC Se rvi ce Department.
Teleph one : 1 -80 0-356-3454, or 585-7 98- 390 1
Or write:
SIGMA LLC
711 Park Avenue
Medina, N.Y. 1 410 3
Email Se rvi ce Coo rdi nator at:
techsupport @si gma pum ps. com
The BioMed Options menu has six categories:
■Sensor Configuration
■Calibration
■Pump Settings
■Pump Testing
■Update Software
■Network Configuration
16
Loading...
+ 47 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.