Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Smart-Trak® 2 Series 100
Mass Flow Meters and Controllers
INSTRUCTION MANUAL
IM-100-ST2, Revision: D
January 2012
1
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
GLOBAL SUPPORT LOCATIONS: WE ARE HERE TO HELP!
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940
Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
www.sierrainstruments.com
EUROPE HEADQUARTERS
Bijlmansweid 2 1934RE Egmond aan den Hoef
The Netherlands
Phone +31 72 5071400 Fax +31 72 5071401
ASIA HEADQUARTERS
Rm. 618, Tomson Centre, Bldg A, 188 Zhang Yang Road
Pu Dong New District, Shanghai, P.R. China
Phone: + 8621 5879 8521 Fax: +8621 5879 8586
IMPORTANT CUSTOMER NOTICE: OXYGEN SERVICE
Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the use of
Sierra Instruments standard mass flow meters or controllers for oxygen gas. You are responsible for
determining if this mass flow meter or controller is appropriate for your oxygen application. You are
responsible for cleaning the mass flow meter or controll er to the degree required for your oxygen flow
application.
No part of this publication may be copied or distributed, transmitted, transcribed, st ored in a retrieval system,
or translated into any human or com puter language, in any form or by any means, electronic, mechanical,
manual, or otherwise, or disclosed to third parties without the expre ss written permission of Sierra Instruments.
The information contained in this manual is subject to change without notice.
TRADEMARKS
®
Smart-Trak
product and company names listed in this manual are trademarks or trade names of their respective
manufacturers.
100 Series and Dial-A-Gas™ is a Registered Trademark of Sierra Instruments, Inc. Other
2
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
• Appendix F: Special Instructions for Installation and Operation of the
Micro-Trak ultra-low flow instruments.
Throughout this manual, we use the word instrument as a generic term to
represent all models of Sierra Instruments’ Smart-Trak 2 Series 100 mass flow
meters and controllers.
SAFETY INFORMATION
Caution and warning statements are used throughout this book to draw your
attention to important information.
Warning!
This statement appears with information that
is important to protect people and equipment
from damage. Pay very close attention to all
warnings that apply to your application.
Caution!
This statement appears with information that is
important for protecting your equipment and
performance. Read and follow all cautions that
apply to your application.
6
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
RECEIPT OF YOUR INSTRUMENT
When receiving the instrument, carefully check the outside packing carton for
damage that may have incurred during shipment. If the carton is damaged, notify
the local carrier and submit a report to the factory or distributor. Remove the
packing slip and check that all ordered components are present and match your
specifications (as ordered). Make sure any spare parts or accessories are not
discarded with the packing material. Do not return any equipment to the factory
without first contacting one of Sierra’s Technical Support Centers:
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
DEFINITIONS USED IN THIS MANUAL
The following terms are used frequently in this manual. They are presented here
with their definitions for your information.
Setpoint—The command or control signal supplied to a flow controller is called its
setpoint. The controller will maintain the flow at this value.
Full scale—The highest flow that an instrument will meter within its specified
accuracy. It is often possible for an instrument to measure a flow beyond its full
scale value, but the accuracy of this measurement may be outside of published
specifications.
Purge—The Smart-Trak 2 Mass Flow Controller is supplied with the ability to open
the valve far beyond the full scale position to allow them to be cleaned. This is
usually accomplished by blowing clean, dry nitrogen through the instrument. When
the valve is opened to this cleaning position, it is said to be in the Purge mode.
LFE—Laminar Flow Element (LFE) or bypass generates pressure drop forcing a
small fraction of the total flow to pass through the sensor capillary tube.
8
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
THE SMART-TRAK 2 FLOW SENSING PRINCIPLE
The operating principle of the Smart-Trak 2 instruments is based on heat transfer
and the first law of thermodynamics. During operation process gas enters the
instrument’s flow body and divides into two flow paths, one through the sensor
tube, and the other through the laminar flow bypass. The laminar flow bypass (often
called LFE which stands for “laminar flow element”) generates a pressure drop, P1–
P2, forcing a small fraction of the total flow to pass through the sensor tube (m1).
Figure 1-1. Flow Paths through the Instrument
Two resistance temperature detector (RTD) coils around the sensor tube direct a
constant amount of heat (H) into the gas stream. During operation, the gas mass
flow carries heat from the upstream coil to the downstream coil. The resulting
temperature difference (∆T) is measured by the Smart-Trak 2 microprocessor.
From this, Smart-Trak 2 calculates the output signal. Since the molecules of the gas
carry away the heat, the output signal is linearly proportional to gas mass flow.
Figure 1-2. Flow Measuring Principle
9
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Figure 1-3. Sensor Temperature Distribution
Figures 1-2 and 1-3 show the mass flow through the sensor tube as inversely
proportional to the temperature difference of the coils. The coils are legs of a bridge
circuit with an output voltage in direct proportion to the difference in the coils’
resistance; the result is the temperature difference (∆T). Two other parameters, heat
input (H) and coefficient of specific heat (Cp) are both constant. Through careful
design and attention to these parameters, this output signal is made linear over the
transducer’s normal operating range (Figure 1-4). As a result, the measured flow
through the sensor tube is directly proportional to the gas flow in the main body.
Figure 1-4. Linear Range of the Transducer’s Output Signal
In the Smart-Trak 2 mass flow controllers, the gas which flows through the
monitoring section is precisely regulated by the built-in electromagnetic valve. The
normally closed valve is similar to an on/off solenoid valve, except that the current
to the valve coil, and hence the magnetic field, is modulated so that the
ferromagnetic valve armature, or valve plug, assumes the exact height above the
valve’s orifice required to maintain the valve’s command flow (set point). The
result is excellent resolution.
10
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Before You Begin Installation
Warning!
Injury can result if line
pressure exceeds the
maximum rating of
500 psig (34 barg).
CHAPTER 2 INSTALLATION
Before installing the instrument, ensure that the installation site conforms to the
specific operating parameters recorded on the instrument’s Data Label. The Data
Label is mounted on the back of the instrument electronics enclosure (see sample
Data Labels in Figure: 2-1). This is critical because each instrument is configured
for a specific application range. Please review the gas or gases, the mounting
orientation, the maximum flow range(s), the inlet and outlet pressure(s), and the
operating temperature(s). The line pressure should not exceed 500 psig (34 barg).
The temperature should not exceed 122°F (50°C). The minimum operating
temperature is 32°F (0°C) and ambient temperature is 0-50°C. If your application
exceeds any of these parameters, contact your Sierra Sales Agent before
installation. You may also contact one of Sierra’s Technical Support Centers.
FACTORY USA: TOLL FREE: 800-866-0200 or PHONE: 831-373-0200 or FAX:
831-373-4402 or EMAIL: service@sierrainstruments.com
Figure 2-1: Examples of Smart-Trak 2 Data Labels
11
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Pre-Installation Check List
1. Double-check to be sure that the O-ring material used in your
instrument is compatible with the gas to be measured. The O-ring
material used in your Smart-Trak 2 can be found in the Data Label. See
Appendix A for a table of elastomer compatibility with a wide variety of
gases.
2. Sierra strongly recommends you install an in-line filter upstream of the
instrument. Recommended filter size: 10 micron. A 10 micron filter is
available from Sierra as an accessory. See Appendix B or contact your local
Sierra distributor.
3. Do not locate the instrument in areas subject to sudden temperature
changes, excessive moisture or near equipment radiating significant
amounts of heat. Be sure to allow adequate space for cable connectors and
wiring.
4. For controllers, use a properly sized pressure regulator. Make sure the
pressure regulator is not too small or too big. There can be no restrictions
(such as valves, tubing or pipe internal diameters, reducers, etc.) upstream or
downstream of the controller with a dimension that is less than the valve
orifice diameter. To determine orifice diameter, consult the calibration
certificate included with your instrument. If restricted, controller will not
reach full scale.
5. Output Signals: The Smart-Trak 2 has two analog outputs that are linearly
proportional to the gas mass flow rate, one voltage and one current. Choose
from 0-20 mA or 4-20 mA for current plus one voltage signal: 0-5 VDC or
0-10 VDC or 1-5 VDC. The output signals specified at time of order will be
indicated on the data label. You may change among the current and the
voltage output signal at your discretion after receipt of the instrument using
the Pilot Module or the Smart-Trak 2 Software (see Chapters 4 & 5).
Changing the output signals has no influence on the instrument’s accuracy.
6. The CAT-5 connector on the side of the Smart-Trak 2 is NOT an Ethernet
connector. It is for use with the optional Remote Pilot Module. Do not plug
an Ethernet cable here as damage may result. Keep this connector covered if
possible whenever it is not in use.
7. The instrument has specific power supply requirements. See the table
later in this chapter for a complete listing of power requirements.
Installing the Instrument—Plumbing
Smart-Trak 2 instruments are supplied with compression, VCO®, VCR®, or female
NPT process connections. To ensure a successful installation, inlet and outlet tubing
12
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
should be clean prior to plumbing the instrument into the system. The shipping caps
covering the inlet/outlet fittings should not be removed until immediately before
installation.
Follow the installation instructions that are applicable to your instrument’s process
connection. Ensure that the tubing is free from burrs or sharp rims that may result
from cutting.
CAUTION: Before use, all plumbing should be checked carefully for leaks,
especially at the connecting fittings. All instruments are leak-tested prior to
shipping. It is not a requirement to leak test your instrument. Do not use liquid
leak detectors such as Snoop
Trak 2. Instead, monitor pressure decay.
®
to search for leaks inside or outside the Smart-
Compression Fittings
1. Position the instrument with the
flow direction arrow pointing in the
direction of flow.
2. Verify the position of the front and
back ferrule. Insert the tubing into
the fitting. Be sure that the tubing rests firmly on the shoulder of the fitting
and that the nut is finger-tight. Scribe the nut at the six o’clock position.
3. While holding the fitting body steady with a backup wrench, tighten the nut
1-1/4 turns, watching the scribe mark make one complete revolution and
continue to the nine o’clock position. For 1/16-inch, 1/8-inch and 3/16-inch
(2, 3 and 4 mm) sizes, tighten only 3/4 turns from finger-tight. Do not over-tighten!
4. If you use flexible tubing (Example: Polyflow) use an “Insert” (see
www.swagelok.com)
5. Check the system’s entire flow path thoroughly for leaks. Do not use liquid
leak detectors. Instead, monitor pressure decay. Exposing the instrument to
leak detector fluid may cause damage.
VCO Fittings
1. Position the instrument with the flow direction arrow pointing in the
direction of flow.
2. Tighten the nut finger-tight, and then 1/8 turn tighter with a wrench. Do not
over-tighten!
13
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
3.Check the system’s entire flow path thoroughly for leaks. Do not use liquid
leak detectors. Instead, monitor pressure decay. Exposing the instrument to
leak detector fluid may cause damage.
VCR Fittings
1. Position the instrument with the flow direction arrow pointing the direction
of flow.
2. Install new gaskets that are compatible with the gas to be used.
3. Tighten the nut finger-tight, and then 1/8 turn tighter with a wrench. Do not
over-tighten!
4. Check the system’s entire flow path thoroughly for leaks. Do not use liquid
leak detectors. Instead, monitor pressure decay. Exposing the instrument to
leak detector fluid may cause damage.
1/4InchFemaleNPT
1. Position the instrument with the flow direction arrow pointing the direction
of flow.
2. Apply high quality Teflon tape to the male NPT fitting. Alternatively, use a
high quality paste pipe thread sealant suitable for the application and gas,
and apply this compound to the inlet and outlet fittings. Avoid getting the
tape or the thread sealant onto the first two threads to keep it out of your
process gas.
3. Tighten each fitting by hand. Then, tighten no more than one (1) turn. Do
not over-tighten.
4. Check the system’s entire flow path thoroughly for leaks. Do not use liquid
leak detectors. Instead, monitor pressure decay. Exposing the instrument to
leak detector fluid may cause damage.
14
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Installing your Instrument—Mechanical Mounting
Mounting your Instrument
The base plate or bottom of the instrument has 4 mounting holes. Two are SAE
thread and two are metric thread. For location and dimensions, please see Appendix
E.
Your Smart-Trak 2 instrument is made from premium 316 stainless steel. As a result,
it may require substantial mounting brackets to properly support its weight. Exercise
caution when installing to avoid damage or injury.
Mounting the Optional Remote Pilot Module
If you have the optional Remote Pilot Module control unit, you have several
mounting options.
1. Wall or Panel mounting—your Remote Pilot Module may be mounted to a
flat surface using the supplied plate and 2 special “shoulder” screws.
Simply screw the shoulder screws into the plate. Then, attach the plate to
the wall by driving 2 screws (not included) through the 2 large central holes.
MAKE SURE THE UP ARROW IMPRINTED ON THE MOUNTING
BRACKET POINTS UPWARD. Your Remote Pilot Module will slip onto
the shoulder screws. Push against the plate and then down. To remove,
push up and pull. Attach the CAT-5 cable to the socket on the bottom of the
Remote Pilot Module.
2. Rack mounting—to mount your Remote Pilot Module to a standard 19”
laboratory rack, first purchase a “blank” panel from your industrial supplier.
Decide where the Remote Pilot Module will be attached, then follow the
procedure listed above for wall mounting. When you are ready, insert the
CAT-5 cable into the socket in the Remote Pilot Module.
3. Desktop mounting—the Remote Pilot Module will sit on a desk in a
fashion that makes it convenient to view and operate. Insert the CAT-5
cable into the socket in the back of the Module. Use the cable as a
“kickstand.” Gently bend the cable to recline the Remote Pilot Module to
an angle suitable for easy viewing. Alternately, you can lay the Remote
Pilot Module on its back and insert the CAT-5 cable into the socket in the
bottom.
15
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Installing your Instrument—Electrical Connections
All electrical connections for your Smart-Trak 2 instrument are made on the left
(inlet) side panel. See Figure 2-2: Smart-Trak 2 Connections below for the location
of all connections. Note that the CAT-5 connector is not an ethernet connector.
CAT 5
HD DB 15
Figure 2-2: Smart-Trak 2 connections
16
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Smart-Trak 2 is provided with a high density 15-pin D Connector called an “HD
DB-15” located on the side of the enclosure and either one of our optional preassembled communications cables or an empty mating connector. Power must be
supplied to the HD DB-15 connector. Other features may be accessed there as well.
The pin numbers and assignments for the HD DB-15 connector are shown in Figure
2-3: HD DB-15 Connector Pin Configuration (on the instrument). The
corresponding colors of the optional communication cable wires and the functions
of each are listed in Figure 2-4: Wiring Definitions for Optional Communication
Cable. The connections for input power, analog output signal and analog input
signal (controllers only) are all made at the HD DB-15 connector. There is a
second copy of these 2 figures in Appendix D for your convenience
17
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Note: Pins 1, 3, and 10 are connected together inside the
instrument. Do not tie these grounds together outside the
instrument. Must have one connection per analog ground.
Recommended use listed.
18
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Instrument Power:
The Smart-Trak 2 requires a 15-24 VDC power supply. If you are using the power
supply supplied by Sierra, connect it to the 15-pin HD DB-15 connector on the side of
the instrument. If you are supplying your own power source, it must be a regulated
15-24 VDC with ripple not to exceed 100 mV peak-to-peak. It must be capable of
producing the current specified for the appropriate voltage shown in Figure 2-5: Power
Supply Requirements. Apply power as follows: positive (+) to the green (pin 6) and
negative (-) to the yellow (pin 5) wires. The instrument is polarity sensitive. If you
reverse this wiring, the instrument will not be damaged, but it will not function.
Instrument Grounding:
The Smart-Trak 2 has very high levels of RFI and EMI shielding built into the metal
electronics cover (meets or exceeds the CE Standard EN 61326-1; 2006). To maintain
the integrity of this CE rating, it is critical that a path be provided for any residual
internal noise to exit the instrument or it may register on the outputs. Grounding
provides this path.
To properly ground your instrument, secure the chassis to solid earth ground using the
mounting holes on the bottom of the flow body. If the instrument will be used without
permanent mounting (on a laboratory bench, for instance) then, using the provided
cable, connect the shield wire (no insulation) to earth ground in your facility. If you
purchased a Sierra power supply, a ground wire is provided for your convenience.
Figure 2-5: Power Supply Requirements
Instrument Type Recommended Input
Voltage
Minimum Current
Required (mA)
M100L Meter 15-24 VDC (+ 10%) 230
M100M Meter
M100H Meter
15-24 VDC (+
15-24 VDC (+
10%)
10%)
230
230
C100L Controller 24 + 10% VDC 500
C100M Controller 24 + 10% VDC 800
C100H Controller 24 + 10% VDC 1260
NOTE: The Compod™ adds an additional 100mA to the ST2 current ratings.
CAUTION: This instrument is not a loop-powered device! Do NOT apply power
to the 4-20 mA output or input connections.
19
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Analog Output Signals:
•Output Signal—Voltage: Measure the voltage output signal across the red (pin 2)
wire and any of the analog grounds: brown (pin 1), orange (pin 3) or gray (pin 10).
The minimum load is 1000 Ohms. We recommend pin 1. DO NOT USE THE SAME
ANALOG GROUND FOR CURRENT OUTPUT, SETPOINT OR RS232.
•Output Signal—Current: Measure the current output signal, 4-20 mA or 0-20 mA,
across the red/white stripe (pin 14) wire and any of the analog grounds: pin 1, 3, or
10. The maximum load is 500 Ohms. We recommend pin 1. DO NOT USE THE
SAME ANALOG GROUND FOR VOLTAGE OUTPUT, SETPOINT OR RS232.
For Mass Flow Controllers, the following analog features are also
available at the HD DB-15 connector:
Setpoint: To transmit an analog setpoint, supply the voltage or current signal (check
the data label and/or setting) across the blue (pin 8) wire and pin 10.
Valve Close: To force the valve closed, connect the black (pin 12) wire to pin 10
Purge: To force the valve to its maximum open position which we call “Purge,”
connect the pink (pin 4) wire to the white (pin 11) wire. Note that this will allow
much greater flow than the rated full-scale value.
For Digital Communication Using Your Personal Computer:
You can communicate with your instrument using the Smart-Trak 2 Software
package and
your PC running the Windows operating system. Simply connect the light green (pin 7)
wire, the brown/white stripe (pin 13) wire and one of the analog grounds (pin 1,3, or 10)
to a standard DB-9 connector according to Figure 2-6: Digital Communication.
Figure 2-6: Digital Communication
RS-232 Transmit (pin 7) to DB-9 pin #2
RS-232 Receive (pin 13) to DB-9 pin #3
Analog ground (pin 3) to DB-9 pin #5
20
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
NOTE: Transit and Receive may need to be reversed, depending on which type of
device or cable is connected. (No damage will result—attempt communication after
reversal.
With the connections in Figure 2-6 in place, plug the DB-9 connector into an appropriate
serial port on your PC.
To minimize the potential for RF interference, it is recommended to shield these wires.
Use a metal DB-9 connector and connect one end of the shield to the DB-9 shell and the
other end to the outer shell of the Smart-Trak 2 HD DB-15 connector.
CAUTION: The CAT-5 connector on the side of the Smart-Trak 2 is NOT an
Ethernet connector. It is for use with the optional Remote Pilot Module or CRN
cable. Do not plug an Ethernet cable here as damage may result.
21
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CHAPTER 3: ANALOG OPERATION
Your Smart-Trak 2 instrument may be operated in three different ways:
THREE CONTROL OPTIONS
A.Analog Input/Output Operation (This Chapter): Using analog input/output signals
at the 15-pin mini-D connector.
B. Digital Operation with Pilot Module (Chapter 4): Using the optional Pilot Module.
C. Digital Operation with RS-232 and Smart-Trak 2 Software (Chapter 5): Using the
RS-232 link, the supplied Smart-Trak 2 Software package, and a PC-style
computer running Windows operating system.
This chapter will discuss the first of these—Analog Operation. Please see subsequent
chapters for other options.
Regardless of control options, the standard output for all Smart-Trak 2 instruments are
two linear analog output signals corresponding to 0% to 100% of the mass flow fullscale range. Please note that one of these output signals is a current signal of either 4-20
or 0-20 mA. The other is a voltage signal of 0-5 VDC, 0-10 VDC or 1-5 VDC. You can
choose any combination of these signals at any time in your facility.
For mass flow controllers, one input signal of 4-20 mA, 0-20 mA, 0-5 VDC, 0-10 VDC
or 1-5 VDC (selectable by user) may be chosen to set the gas mass flow rate to any
desired value within the range of the device. This input signal must be a direct linear
representation of 0% to 100% of the desired gas mass flow full-scale value. For the
location of these signals on the HD DB-15 connector, refer to Figure 2-4 or Appendix D.
Analog Operation, Mass Flow Meter
After your instrument is installed and the system has undergone a complete leak check
as discussed in detail in Chapter 2, you are ready to supply power.
Power Your Instrument: Provide adequate power per Figure 2-5. Apply power using
Sierra’s power supply or your own power source. The green LED at the top of the inlet
side will light to confirm power. If your instrument has a Pilot Module, it will begin its
start-up cycle. See Chapter 4 for details on Pilot Module operation. NOTE:
recommended you connect power to the Smart-Trak 2, and then power your supply
(plug into wall or switch on). The opposite may cause the unit to take longer to power
on. Let the instrument warm up for at least 15 minutes for optimal performance.
Your Smart-Trak 2 instrument is now ready for use!
It is highly
22
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Analog Operation, Mass Flow Controller
After your instrument is installed and the system has undergone a complete leak
check as discussed in detail in Chapter 2, follow these steps:
1. The valve will remain closed until power is supplied. See Chapter 2 for
wiring instructions. Remember that the valve in the Smart-Trak 2 is not a
positive shut-off device. When power is applied, the flow control valve will
operate per any instructions it receives. When the Smart-Trak 2 is delivered, the
valve will be in the Automatic (Normal) state and the Pilot Module or analog
signal will provide the correct zero setpoint reference for the instrument. As a
result, the valve will be closed. However, upon subsequent power-ups, the
valve will return to the state it was in the last time the instrument was operated.
CAUTION: If you do not know the value of the setpoint or the valve state
given to the Smart-Trak 2 when it was last operated, you must assume that
the valve will open when power is applied. Take necessary precautions.
You may use the Pilot Module or the Smart-Trak 2 Software to check the
setpoint or the valve state currently on your instrument. See Chapter 4 or
Chapter 5 for information on Setpoint and Valve State.
2. Power Your Instrument: Provide adequate power per Figure 2-5. Apply power
using Sierra’s power supply or your own power source. The green LED at the
top of the inlet side will light to confirm power. If your instrument has a Pilot
Module, it will begin its start-up cycle. See Chapter 4 for details on Pilot
Module operation. NOTE: It is highly recommended you connect power to the
Smart-Trak 2, and then power your supply (plug into wall or switch on). The
opposite may cause the unit to take longer to power on. Let the instrument warm
up for at least 15 minutes for optimal performance.
3. Adjust the controller setpoint to the desired flow rate by supplying an
appropriate signal (mA or VDC). The effective control range of the unit is 2%
to 100% of the calibrated full scale flow range. Automatic shut-off occurs at
1.9% of the factory full scale calibrated range unless specifically modified at
time of order. Smart-Trak 2 will immediately begin accurately monitoring and
controlling the gas mass flow rate. Let the instrument warm up for at least 15
minutes for optimal performance.
Your Smart-Trak 2 instrument is now ready for use!
23
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Smart-Trak 2 Features
Setpoint Adjustment
The setpoint (command) input signal you supply to Smart-Trak 2 must be a direct
linear representation of 0% to 100% of the mass flow full-scale value. Apply the
setpoint signal from pin 8 to any of the analog grounds (see Chapter 2 for wiring
details). A setpoint value of 0 VDC or mA (or 1 VDC or 4 mA) will regulate the
flow to 0% and a setpoint value of 5.00 VDC (or 10 VDC or 20 mA) will adjust the
flow to 100% of the instrument’s full scale range.
When the setpoint (command) signal is applied, the flow controller will reach the
setpoint value within two seconds to within ±2% of the selected flow rate.
CAUTION: DO NOT LEAVE A SETPOINT APPLIED FOR AN
EXTENDED PERIOD OF TIME TO A CONTROLLER WHEN THE
GAS SUPPLY IS SHUT OFF OR BLOCKED. Damage may result and the
instrument will become hot to the touch. Instead, see below for use of the
“Valve Close” feature which allows you to disable the valve while
maintaining the setpoint signal. This may be set by the Pilot Module, the
Smart-Trak 2 Software, or an external analog signal.
Changing the Output or Setpoint Signals
To modify the analog output or setpoint signals (from 4-20mA to 0-10Vdc, for
example), you must use the Pilot Module or the Smart-Trak 2 Software. The data
label will indicate the form these signals had when the instrument was last
calibrated. We strongly recommend that you adapt the data label if the
configuration is changed for future reference. See Chapter 4 or 5 for the necessary
procedure.
Over-Range Condition
If the mass flow rate exceeds the full-scale range listed on the Smart-Trak 2 data
label (see samples on page 2-1), the output signal will measure above full-scale.
However, the device has not been calibrated for flows in excess of the calibrated
full scale value and the value will be both non-linear and inaccurate if an over-range
condition exists. Please be aware that the analog outputs can exceed full scale by as
much as 20%, or more.
Once the over-range condition has been removed, it may take up to 30 seconds for
the Smart-Trak 2 to recover and resume normal operation. An over-range condition
will not
harm the instrument.
24
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Manual Valve Override—Valve Close
Manual valve override is provided for all Sierra mass flow controllers. This feature
includes both a valve close command and a valve maximum open command (called
purge). When the valve is directed to close or to purge, it will no longer respond to
a setpoint command.
FOR VALVE CLOSE: connect pin 12 to analog ground
Remember that the valve in the Smart-Trak 2 is not a positive shut-off device.
The Controller will return to normal automatic operation about 4 seconds
after pin 12 is left floating.
Manual Valve Override—Valve Purge Function
The purge function opens the controller valve completely for the purpose of quickly
flushing unwanted gas from the flow path. When the valve is opened for purging, it
allows flows far in excess of the rated full scale of the controller.
FOR VALVE PURGE: connect pin 4 to pin 11.
CAUTION: PURGE MODE ALLOWS FAR MORE GAS TO FLOW THROUGH
THE CONTROLLER! BEFORE USING VALVE PURGE OPERATION, INSURE
PROPER DOWNSTREAM CAPACITY AND VENTILATION.
IMPORTANT NOTES ABOUT PURGING
Purging Non-Reactive Gases:
Purge your Smart-Trak 2 with clean, dry nitrogen for a minimum of two hours.
Purging Reactive Gases:
One of the following methods may be used:
Cycle purge. This is done by alternately evacuating and purging the
instrument for 2 to 4 hours with clean, dry nitrogen.
Purge the instrument with clean, dry nitrogen for 18 to 24 hours.
Evacuate the instrument for 18 to 24 hours.
25
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
IMPORTANT SAFETY NOTES ABOUT PURGING
WARNING: When toxic or corrosive gases are used,
purge unit thoroughly with inert dry gas before
disconnecting from the gas line to prevent personnel
from being injured when coming in contact with the
instrument.
WARNING: If an instrument used with a toxic or
corrosive gas is returned to the factory, a Material
Safety Data Sheet (MSDS) must be enclosed &
attached to the outside of the box to alert Sierra
personnel of the potential hazard. Also, make sure the
inlet & outlet are securely sealed.
26
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CHAPTER 4: Digital Operation with Pilot Module
Your Smart-Trak 2 instrument may be operated in three different ways:
THREE CONTROL OPTIONS
A. Analog Input/Output Operation (Chapter 3): Using analog
input/output signals at the HD DB-15 connector.
B. Digital Operation with Pilot Module (This Chapter): Using the
optional Pilot Module.
C. Digital Operation with RS-232 and Smart-Trak 2 Software
(Chapter 5): Using the RS-232 Smart-Trak 2 Software package and a
computer running the Windows operating system.
This chapter will discuss the second of these—Digital Operation with the optional
Pilot Module. Please see alternate chapters for other options.
Although you have chosen to use the optional Pilot Module, please note that all the
Analog control functions are still available on your instrument. Consult Chapter 3
for details on Analog operation. Also, computer control using the RS-232
communication is available. See Chapter 5 for details on operation with a computer.
CAUTION—If RS-232 digital communication is to be used in conjunction with the
Pilot Module, the HD DB15 connector must be properly wired with a three wire
serial DB9 cable to your computer. Often, this is done with the same HD DB15 that
supplies power to your instrument. You can run both RS-232 communication and
Pilot Module communications in parallel, but the unit will only respond to one set of
commands at a time. DO NOT attempt to control the unit simultaneously with both
the Pilot Module and the computer, this can lock up the unit.
Introduction to Pilot Module Features and Capabilities
The optional Pilot Module functions as both display and a control unit for your
Smart-Trak 2 instrument. The standard Pilot Module is available mounted directly
on the face of your instrument or as a handheld / remote mountable control interface
attached to the Smart-Trak 2 via a detachable cable.
27
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If your instrument has a standard Pilot Module mounted locally on the face of the
unit, no additional set-up is required. See picture below.
On the other hand, if your instrument has a Remote Pilot Module, attach one end of
the included Category 5 (CAT 5, also called RJ-45) connecting cable into the jack
at the top of the instrument’s left side, immediately above the HD DB-15 connector.
Next, place the other end into one of the two matching jacks on the Pilot Module.
For your convenience, Sierra has provided two jacks—one on the back and one on
the bottom of the Remote Pilot Module. You may use whichever jack is most
convenient for your application as they both have identical functions.
The Pilot Module includes a large LCD graphic display screen and six buttons. The
LCD will show a variety of information and the buttons can be used to view and
modify this information. The convenient buttons are:
Left arrow
Right arrow
Up arrow
Down arrow
Enter button
Escape button
28
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
These are shown in the photo below:
Up, Down
Left, Right
Pilot Module Operation, Mass Flow Meter
Caution!
The Smart-Trak 2 is
not a loop-powered
device. Do not
apply power to the
4-20 mA outputs.
Assuming no gas is flowing, after another 5-10 seconds the display will read:
Escape
Enter
After your instrument is installed and the system has undergone a complete leak check
as discussed in detail in Chapter 2, follow these steps:
1.Power Up Your Instrument: Apply power to your instrument. See Chapter 2,
Figure 2-5: Power Supply Requirements. When power is first applied, the Pilot
Module will display:
Version
2.04X
Read Parameters
Mass Flow
0.000sl/m
29
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Air
NOTE: If gas is flowing the Pilot Module will immediately begin to accurately
display the gas mass flow rate on the LCD panel. If you have chosen alternate units
or another gas, the display will show the selected units instead of the above.
2. Open the gas supply: Smart-Trak 2 is now ready to monitor the gas mass flow
rate. Let the instrument warm up for at least 15 minutes for optimal
performance.
Your Smart-Trak 2 instrument is now ready for use!
Pilot Module Operation, Mass Flow Controllers
After your instrument is installed and the system has undergone a complete leak
check as discussed in detail in Chapter 2, follow these steps:
1.The valve will remain closed until power is supplied. See Chapter 2
for wiring instructions.
CAUTION: Remember that the valve in the Smart-Trak 2 is not a guaranteed
positive shut-off device. For dangerous applications, Sierra recommends use of
an external shut-off safety valve.
When power is applied, the flow control valve will operate per the instructions it
receives from the Pilot Module. When the Smart-Trak 2 is delivered, the valve will
be in the Automatic (Normal) state and the Pilot Module will provide the correct
zero setpoint reference. As a result, the valve will be closed. However, the valve
will return to the state it was in the last time the instrument was operated.
WARNING: If you do not know the setpoint or the valve state of the Mass
Flow Controller before it was shut down, you must assume that the valve will
open when power is applied. TAKE NECESSARY PRECAUTIONS.
2. Power Up Your Instrument: Apply power to your instrument using
Sierra’s power supply or your own input power source. See Chapter 2,
Figure 2-5: Power Supply Requirements. When power is first applied,
the Pilot Module will display:
Version
2.04X
Read Parameters
If no gas is flowing, after another 5-10 seconds the display will read:
30
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Mass Flow
0.000sl/m
Air
NOTE: If gas is flowing and the Pilot Module has a setpoint greater than zero, it
will immediately begin to accurately display the gas mass flow rate on the LCD
panel. If you have chosen alternate units or another gas, the display will show the
selected units instead of those above.
3. Open the gas supply. Smart-Trak 2 is now ready to monitor and
control the gas mass flow rate. The display will show 0.000 until it is
given a setpoint. Let the instrument warm up for at least 15 minutes for
optimal performance.
Your Smart-Trak 2 instrument is now ready for use!
CAUTION: DO NOT LEAVE A SETPOINT APPLIED FOR AN EXTENDED
PERIOD OF TIME TO A CONTROLLER WHEN THE GAS SUPPLY IS
SHUT OFF OR BLOCKED. Damage may result and the instrument will become
hot to the touch. Instead, see below for use of the “Valve Close” feature which
allows you to disable the valve while maintaining the setpoint signal. This may be
set by the Pilot Module, the Smart-Trak 2 Software, or an external analog signal.
Using the Pilot Module Menus & User Interface
The features of the Pilot Module can be considered in three groups:
1 Upper Level Screens: display information (no password is required to view this
information). These include:
Mass flow rate
Gas (10 options pre-programmed)
Engineering units (mass per unit time)
Current Setpoint with units
Source of Setpoint (analog or digital and type)
Valve operation mode (normal, valve close or purge)
Current meter full scale value with units (user selectable)
31
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
2 Lower Level Screens: permit changes to instrument operation. They are password
protected. These include:
Setpoint value
Engineering units
Gas
Valve operation
Source of the setpoint signal
Form of the output signals
Full scale of the instrument
Password
Zero
Span
3 Maintenance Features:
Re-boot the Smart-Trak 2 microprocessor. Press the LEFT ARROW,
DOWN ARROW, ENTER, AND ESCAPE keys at the same time.
Move the decimal point. Navigate to the “change the setpoint” menu. Press
ENTER to make one of the digits flash. While it is flashing, press ESCAPE at the same
time you press LEFT ARROW or RIGHT ARROW.
Return all parameters to the factory default values. Navigate to the
“Change the Setpoint” menu. Press LEFT ARROW, UP ARROW, and ENTER at the
same time. Choose “yes” to restore factory default settings.
If you are trying to navigate using the pilot but the message appears,
“Must be in Pilot Mode,” use the LEFT ARROW from this screen to navigate to the
“Change Setpoint Source” and change to “Pilot / RS-232.” Escape and try again.
32
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Map of the Pilot Module Interface
The Pilot Module user interface is presented below in a graphical format. Once you
have some familiarity with the user interface, you may find you want to make a
copy of this and keep it with the instrument for reference. You can find a larger
version of this Flow Chart in Appendix C.
33
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
The seven Upper Level Screens display a variety of information. You are able to
move between the screens by pressing the left or right arrows. No password is
required for the Upper Level Screens.
Mass Flow Screen
When the instrument is powered up, or whenever the escape button is pressed, it
always returns to the Mass Flow Screen. This screen displays the mass flow rate,
the engineering units and your gas choice. It looks something like this:
Mass Flow
0.000sl/m
Air
Setpoint Screen
Pushing the right arrow takes you to the Setpoint Screen. The Setpoint Screen
displays the current setpoint given to the controller, the engineering units and the
source of the setpoint signal.
The source of the setpoint can be:
• Pilot Module/RS-232
• 4-20 mA
• 1-5 Vdc
• 0-5 Vdc
• 0-10 Vdc
For operation with the Pilot Module, the display will look something like this:
Setpoint
10.00
Pilot Module/RS-232
If this screen does not show “Pilot Module/RS-232” at the bottom, you will not be
able to give the controller a setpoint command from the Pilot Module because the
instrument is waiting for an analog setpoint. See section below titled “Change
Setpoint Source Screen” to change the source of the setpoint signal. Alternately,
you may supply an analog setpoint to the HD DB-15 connector (see Chapter 3).
sl/m
34
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Valve Position Screen (Mass Flow Controllers only)
Pushing the right arrow again takes you to the Valve Position Screen, if you have a
Mass Flow Controller. This screen will display the current state of the Smart-Trak 2
valve.
The state of the valve can be:
1. Closed (Remember that the Smart-Trak 2 is not a positive shut-off device).
2. Purge--Maximum Open (recommended 120% of the calibrated full scale
value, but can be much more and can be dangerous)
3. Automatic (the normal position, where the controller responds to a setpoint
signal).
WARNING: The flow rate in Purge is much greater than the
calibrated full scale value and as a result can be dangerous.
For normal operation of the flow controller, this screen should display:
Valve
Automatic
Normal
If this is visible, the instrument will automatically control flow as soon as a setpoint
is given to it. If this screen displays Closed or Purge, the instrument has been
placed into an override position and it will not respond to any setpoint signal. The
valve state may be changed using the “Change Valve Operation” as described on
page 4-16.
Full Scale Screen
Pressing the right arrow again takes you to the Full Scale Screen. This screen
displays the current
displays the gas. Note that this is not necessarily the factory calibrated full scale
value. The screen will display:
full scale value of the instrument with engineering units. It also
Full Scale
10.00
To change the full scale value, see the section below titled “Change Full Scale
Screen.” Pressing the right arrow again takes you back to the Mass Flow Screen.
sl/m
Nitrogen
35
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Lower Level Screens (Changing Parameters)
Getting to the Lower Level Screens:
Your instrument is password protected so that unauthorized personnel will be
unable to change the operating parameters of Smart-Trak 2. To enter the Lower
Level Screens at any time you must first supply the correct password.
Password Screen: By pressing the “enter” key from any of the Upper Level
Screens you will come to the Password Screen. (If you do not know if you are in an
Upper Level Screen, press escape and you always automatically go to the Mass
Flow Screen in the Upper Level) The display will show:
Enter Password
0000
The first digit will blink. At this point, you must enter the correct password to gain
access to the Lower Level Screens.
9 If the instrument is being operated for the first time or if no password has ever
been set on the instrument: You can use the factory default password. The factory
default password is “0000.” To proceed to the Lower Level Screens by using the
factory default password, simply press the “enter” key a second time. If you want
rapid access to permit regular changes to your instrument and you do not desire a
password, this is the fastest way to enter the lower level.
9If You Have a Password: If you have already set a password, enter it now. To
enter the password, push the up arrow to increase the blinking digit or the down
arrow to decrease the blinking digit. To move to the next digit, press the left or
right arrow and repeat the process. When you have selected your four digit
password, press the enter key.
9If You Want to Set a New Password: If you want to change the password, you
must first get to the Lower Level Screens. Proceed by entering your known
password or if no password has ever been set on the instrument, us the factory
default password. Follow the instructions in the “Change Password Screen” section
found later in this chapter.
If the password you have entered is correct, you will enter the Lower Level at the
Change Setpoint Value Screen.
36
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If the password is not correct, the display will show:
Access Denied
Press any button
To continue
When you press any button, you will return to the Mass Flow Screen in the Upper Level.
Press the “enter” key to try again.
LOST PASSWORDS & GENERAL CUSTOMER SERVICE: If you lose your
password, it will be necessary to contact one of Sierra’s Technical Support Centers.
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Making Changes Using the Lower Level Screens:
The ten Lower Level Screens are at the heart of Dial-A-Gas, allowing you complete
control of your Smart-Trak 2 instrument. It is possible to make several changes on
different Lower Level Screens before exiting.
For example, you could change between one of the ten pre-programmed gases,
change the engineering units, and change the setpoint all in one visit to the Lower
Level Screens. The Smart-Trak 2 will make each adjustment as you complete it.
At any time, you may press the escape button to return to the Upper Level.
Change Setpoint Value Screen
This screen is the entry point to the Lower Level. As soon as a correct password is
entered, you will arrive here. If you are already in the Lower Level Screens, push
the right or left arrow until you reach the Change Setpoint Value screen. The
display will show:
Change Setpoint
00.00 sl/m
Air
This is the position where you can change the setpoint value of the mass flow
controller. To make a change to the displayed value, press the enter key. The first
number in the display will blink. Use the up and down arrows to change the value
of this digit or the left and right arrows to move to another digit. For example, if
you wish to enter a setpoint of 12.5 sl/m, push the “up” arrow once when the first
digit is blinking. You will see:
Change Setpoint
10.00 sl/m
Air
Next, press the “right” arrow so that the second digit blinks. Push the up arrow
twice. You will now see:
Change Setpoint
12.00 sl/m
Air
Press the right arrow again. The first digit after the decimal point will now blink.
Press the up arrow 5 times until you see:
38
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Caution!
The Smart-Trak 2 will not
allow you to enter a
setpoint greater than the
current full scale value
set on the instrument.
Change Setpoint
12.50 sl/m
Air
Now that you have made your selection, press the enter key. The display from our
example will show:
Change Setpoint
12.50 sl/m
Air
If you are finished or wish to observe the changes you have made on the LCD
panel, press the escape key to return to the Upper Level Mass Flow Screen.
If you prefer to make additional changes, use the left and right arrow keys to move
to other Lower Level Screens.
Note: If you enter a Setpoint that exceeds the full scale value (displayed in the “Full
Scale” screen in the Upper Level), the Smart-Trak 2 will automatically modify this
value to equal the current full scale value. For example, if the current full scale value
of your instrument is 10 slpm and you have entered a setpoint of 15 slpm, the SmartTrak 2 will modify your setpoint to 10 slpm when you implement the change. The
Setpoint Value screen will show 10 slpm, not 15 slpm.
Change Units Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Units screen. To get to this screenat any time, Press Escape—
Press Enter—type password and Press Enter. Then, press the Right or Left arrow
until you reach the Change Units Screen
The display will show:
.
Change Units
sl/m
39
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If you wish to change the engineering units, press the enter button at this point. The
“mass units” will begin to blink. Use the up or down arrows to select an alternate
unit. You can choose from the following mass units:
sl
NL
g
kg
lb
scc
Ncc
SCF
3
NM
3
SM
When you are satisfied, push the left or right arrow. You will now see the “time
unit” blink. Use the up or down arrows to select your choice of time units. You can
choose from the following time units:
m (minutes)
H (hours)
S (seconds)
When you are finished, press the enter button again. You will see:
Change Units
XXX/x
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens and to observe your change.
NOTE: If a small unit is chosen, the device might not be able to display the valve
and might show 9999. You might need to restore factory defaults. See Maintenance
Functions starting on page 47.
Change Gas Screen (Dial-A-Gas)
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Gas screen. To get to this screenat any time: Press Escape—Press
Enter—type password and Press Enter. Then, press the Right or Left arrow until
you reach this screen. The display will show:
Change Gas
Nitrogen
40
Caution!
The Smart-Trak 2
valve is not a positive
shut-off device.
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If you wish to change the gas used in the instrument, press enter. The name of the
gas will blink. Use the up and down arrows to make your selection. When you
reach the desired gas, press enter. You will see:
Change Gas
XXXX
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens and to observe your change.
Note: your instrument comes with 10 pre-programmed standard gases. These are
listed in the Specifications in Appendix B. Smart-Trak 2 may be ordered with
alternate gases programmed. If your device was so ordered, you may choose from
these 10 gases instead. You may see the 10 gases programmed in your instrument
by using this screen and simply scrolling up or down.
Change Valve Operation-Close, Purge
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Valve Operation screen. To get to this screenat any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen. The display will show:
Change Valve
Operation
Automatic
From this screen you may set the valve to open all the way (“Purge”), force the
valve to remain closed until further changes are made (“Valve Closed”) or set the
valve to control flow when it receives a setpoint from some source (“Automatic”).
To make a change to the valve operation, press the enter key. Use the up and down
arrows to make your selection. When you are satisfied, press the enter key again.
You will see:
Change Valve
Operation
XXXXX
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens.
41
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Note: The valve will move to the desired position immediately when you press the
enter key.
IMPORTANT SAFETY NOTES ABOUT PURGING
WARNING: When toxic or corrosive gases are used, purge
unit thoroughly with inert dry gas before disconnecting from
the gas line to prevent personnel from being injured when
coming in contact with the instrument. Chapter 3 discusses
how to purge your instrument. Always neutralize any toxic
gas trapped inside the instrument before removing it from
the gas line.
WARNING: The flow rate in Purge is much greater than the
calibrated full scale value and as a result can be dangerous.
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Setpoint Source screen. To get to this screenat any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen. The Change Setpoint Source screen allows
you to re-configure the location and type of the setpoint for the Smart-Trak 2
controller. If you intend to supply the setpoint command signal from the Pilot
Module or a computer using the RS-232 link, the display must show:
Change Setpoint
Source
Pilot/RS232
If, instead of using the Pilot Module or the RS-232 link, you prefer to supply an
analog setpoint signal to the Smart-Trak 2, press the enter button. “Pilot/RS232”
will begin to blink. Use the up and down arrows to make your selection from the
following choices:
0-5 VDC
0-10 VDC
1-5 VDC
42
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
4-20 mA
Pilot/RS232
When you are satisfied, press enter. You will see:
Change Setpoint
Source
X-XX XX
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens.
Caution: If you change the source of the setpoint to an analog
value, you will not be able to control your Smart-Trak 2 mass flow
controller via the Pilot Module or the RS-232 link.
Change Output Signals Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Output Signals screen. To get to this screenat any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen
Change Output
. The display will show:
Signals
0-5 VDC/4-20 mA
Here you can re-configure the analog output signals for the instrument. The SmartTrak 2 always outputs one current signal of 4-20mA but the voltage signal may be
selected using this screen. Use the up and down arrows to make your selection.
You can choose between:
0-5VDC and 4-20mA
0-10VDC and 4-20mA
1-5VDC and 4-20mA
When you are finished making your selection, press enter. The screen will read:
43
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Change Output
Signals
X-XX XX/X-XX XX
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens.
Change Full Scale Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Full Scale screen. To get to this screenat any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen. The display will show:
Change Full Scale
XX.XX sl/m
Air
This screen allows you to re-range the outputs of your instrument. You may select
any full-scale value between 100% and 50% of the displayed maximum value (this
is the factory full-scale calibration value).
The new full-scale value that you select will re-define the analog outputs of the
instrument. The 20 mA signal and the corresponding voltage signal (5 VDC, 1-5
VDC or 10 VDC) will now represent this new full-scale value.
Caution: Changing the full-scale value of the instrument does not affect the
accuracy.
The accuracy is always ±1% of the original factory full-scale calibration value.
Caution: For any instrument, if a value greater than the factory full scale
calibration value is entered on this screen, the Smart-Trak 2 will modify the
requested value to equal the factory full scale calibration value.
If you choose to change the full-scale value, press the enter key. The first digit will
blink. Use the up and down arrows to adjust the value of the digit or the left and
right arrows to choose another digit. When you have completed your modification,
press the enter key. The display will show:
Change Full Scale
XX.XX sl/m
Air
44
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens.
Change Span Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Span screen. To get this screen at any time: Press Escape—Press
Enter—enter password and Press Enter. Then, press the Right or Left arrow until
you reach this screen.
From this screen you can change the instrument’s span distance by a percentage
factor ratio. For example, if your meter is reading 1% high of reading, change span
to 0.990 (99.0%) and the full scale value should be reduced by 1%. If your
controller is controlling 1% below of reading, the same change will bring it in. If
the meter is reading 1% low (controller controls 1% high), then change span to
1.010 (101.0%). The display will show:
Change Span
1.000
Air
To make a change, press the enter key. The first digit will begin to blink. Use the
up and down arrows to modify this digit or the left and right arrows to choose
another digit. When you are satisfied, press enter. The display will show:
Change Span
X.XXX
Air
Keep in mind that you can change the span for each individual gas, thus, a change
in the span to one gas will not affect the span of another gas. You can make
additional changes by using the left and right arrow keys to move to other Lower
Level Screens. You may also choose to press the escape key to return to the Upper
Level Screens.
Zero Meter Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Zero Meter screen. To get this screen at any time: Press Escape—Press
Enter—enter password and Press Enter. Then, press the Right or Left arrow until
you reach this screen.
45
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
From this screen you can change the zero flow bridge differential value to match
the zero flow conditions of your application. Mount (or place) the instrument where
it is intended to be used, minding orientation, tilt, etc.; The accuracy of this function
is dictated on how close to operating conditions the unit is when the meter is
zeroed. The display will show:
Zero Meter
Confirm zero flow
Then press enter
Once you push enter, the unit will monitor the sensor bridge differential value. The
display will show:
Zero Meter
Confirm zero flow
New Zero
Upon pressing the enter key again, the unit will record the bridge differential value
to the zero flow condition, leaving you with a freshly “zeroed” instrument! You can
make additional changes by using the left and right arrow keys to move to other
Lower Level Screens. You may also choose to press the escape key to return to the
Upper Level Screens.
Change Password Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Password screen. To get to this screenat any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen.
From this screen you can change the instrument password from the factory default
to any four-digit password of your choice. The display will show the factory default
password of four zero’s: “0000”. The display will show:
Change Password
0000
To make a change, press the enter key. The first digit will begin to blink. Use the
up and down arrows to modify this digit or the left and right arrows to choose
another digit. When you are satisfied, press enter. The display will show:
Change Password
0000
46
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
You can make additional changes by using the left and right arrow keys to move to
other Lower Level Screens. You may also choose to press the escape key to return
to the Upper Level Screens.
Caution: Once you change the Password, you will not be able to enter the
Lower Level without it. Be certain the new password is recorded.
Lost Passwords and General Customer Service
The “creator” password is 6363- this will override all other passwords. If this does
not work, it will be necessary to contact one of Sierra’s Technical Support Centers.
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CHAPTER 5: Digital Operation RS-232 & Smart-Trak 2
Software
Your Smart-Trak 2 instrument may be operated in three different ways:
THREE CONTROL OPTIONS
A.Analog Input/Output Operation (Chaper 3): Using analog input/output signals
at the 15-pin mini- D connector.
B.Digital Operation with Pilot Module (Chapter 4): Using the optional Pilot
Module.
C.Digital Operation with RS-232 and Smart-Trak 2 Software (This Chapter): Using
the RS-232 Smart-Trak 2 Software package and a PC-style computer running
the Windows operating system.
This chapter will discuss “C” above—Digital Operation with your computer via RS232 and Smart-Trak 2 Software. Although you have chosen to use the RS-232
option, please note that all the Analog control functions are still available on your
instrument. Consult Chapter 3 for details on Analog operation. Also, the Pilot
Module may be used so long as you make RS-232 communication with the supplied
CRN cable. See Chapter 4 for details on operation using the Pilot Module.
If you prefer to write your own software to communicate with the Smart-Trak 2 over
the RS-232 link, this is certainly possible. Sierra Instruments will provide the
Source Code including the Command Set upon request. Unfortunately, this is the
limit of software Technical Support we can extend.
Summary of the Smart-Trak 2 Features
Your Smart-Trak 2 instrument may be easily monitored and adjusted using the
supplied Smart-Trak 2 Software package. You should note that the Smart-Trak 2
Software was designed with all the same functions as the Pilot Module. The
software allows you to see all changes & parameters at a glance and will allow you
to make changes quickly and easily. For review, the features of the Smart-Trak 2
include:
1. Top Level Screens that display information In our software, these screens
appear as yellow boxes. They include:
Mass flow rate
48
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Gas (10 options pre-programmed, one must be AIR)
Engineering units (mass per unit time)
Current Setpoint with units
Source of Setpoint (analog or digital and type)
Valve operation mode (normal, valve shut or purge)
Current meter full scale value with units (user selectable)
2. Lower level Screens that permit changes to instrument operation In our
software, these screens appear as white boxes. They include:
Setpoint value
Engineering units
Gas
Valve operation
Source of the setpoint signal
Form of the output signals
Full scale of the instrument
Zero meter
Span meter
3. Additional Features include:
Re-boot the Smart-Trak 2 microprocessor
Change the Communication Port
Links to Sierra Instruments’ Web Site
Specifications
Power Up Your Instrument
CAUTION: The Smart-Trak 2 is not a loop-powered device. Do not apply
power to the 4-20mA outputs.
After your instrument is installed and the system has undergone a complete leak
check (discussed in detail in Chapter 2), apply power using Sierra’s power supply
or your own input power source. See Chapter 2, for power supply requirements.
The green LED at the top of the left side will light. If your instrument has a Pilot
Module, it will begin its start-up cycle. See Chapter 4 for details on Pilot Module
operation.
If you have a Mass Flow Controller, the valve will remain closed until power is
supplied. Remember that the valve in the Smart-Trak 2 controller is not a positive
shut-off device. When power is applied, the flow control valve will operate per any
instructions it receives. When the Smart-Trak 2 is delivered, the valve will be in the
Automatic (Normal) state and the Pilot Module will provide the correct zero
49
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
setpoint reference for the instrument. As a result, the valve will be closed.
However, the valve will return to the state it was in the last time the instrument was
operated.
CAUTION: If you do not know the position of the valve before it was shut
down, you must assume that the valve will open when power is applied. Take
necessary precautions.
Power Up Your Computer
Apply power to your computer per the manufacturer’s recommendations. The
Smart-Trak 2 Software is compatible with any computer running the following
Windows Operating Systems:
Windows 98, 2nd Edition
Windows XP
Windows XP Professional
Windows 2000
Loading the Smart-Trak 2 Software
If you are using your Smart-Trak 2 instrument or your computer for the first time, it
is necessary to install the Smart-Trak 2 Software into your computer. If this
software is already installed, skip this section. If you want to upgrade the SmartTrak 2 Software because you have a higher revision, continue below.
Each Smart-Trak 2 order is shipped with a CD-ROM containing the Smart-Trak 2
Software. Locate this disk. At this point, EXIT OUT OF ANY OPEN
APPLICATIONS BEING RUN ON YOUR COMPUTER.
PROCEDURE:
1. Insert the Smart-Trak 2 Software CD into your CD-ROM
2. Open “My Computer” on your desktop
3. Open the CD Named: “Smart-Trak 2” on your D-drive
4. Run “setup.exe”
5. Follow the instructions on screen
50
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CAUTION: It is recommended that you do not change the default installation
directory for this software. The default directory is:
Changing the installation directory may lead to malfunctions in the software.
C-drive:\ Program Files.
Connecting Smart-Trak 2 to Your Computer
If your computer has a serial port…
We suggest you use the supplied Sierra Instruments RS-232 communication cable
(part number CRN). This pre-manufactured cable has the correct DB9 connection
to mate with most computers and a CAT 5 connector which should be connected to
the jack on the side of your new instrument.
With your Smart-Trak 2 POWERED, plug CAT 5 connector into the receptacle on
the side of your Smart-Trak 2 instrument (See Figure 5-1: Smart-Trak 2
Connections). Next, plug the DB-9 connector to an appropriate serial port on your
computer. Note the serial port channel number, especially if there is more than one
serial port available. If your computer has only one serial port, it is often named
“Comm Port 1.” You will need to know the Comm Port number to communicate
with your Smart-Trak 2 instrument.
51
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CAUTION: The CAT-5 connector on the side of the Smart-Trak 2 is NOT an
Ethernet connector. Do not plug an Ethernet cable here as damage may result.
All electrical connections for your Smart-Trak 2 instrument are made on the left
(inlet) side panel.
Figure 5-1: Smart-Trak 2 Connections
CAT 5
HD DB 15
If you do not have the Pilot Module display, you can connect pins 7 (serial
transmit), 13 (serial receive) and 3 (analog ground) on the instrument’s 15-pin miniD Connector to an appropriate serial port using a standard DB-9 connector instead
of using the supplied CRN cable (see Chapter 2 for wiring instructions). If you use
the mini-D connector for your RS-232 communication, please shield the cable to
prevent EMI from disrupting communications.
52
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If your computer has no serial port, but has a USB port…
Perform the identical connections as if your computer has a serial port (see previous
paragraph), but do not plug the DB-9 connector into your computer. Instead,
purchase an appropriate Serial to USB converter. Plug the DB-9 connector into
your Serial to USB converter and then plug the USB connector into your computer.
Load the necessary software driver into your computer to operate the converter
(software and instructions should come with the converter).
Your computer will assign a Comm Port number to your USB adapter, but it will
probably not be “Comm Port 1.” Use the Control Panel feature of your computer to
identify which Comm Port number has been assigned to your USB adapter.
Open “My Computer” from your desktop.
Next, open “Control Panel.”
Then, open “System.”
Click on the tab labeled “Device Manager.”
Scan down the list until you see the word “Ports,” then right double-click to
examine the available Com Ports. Record the number(s). You will need to enter
this number or one of these numbers to communicate with your Smart-Trak 2
instrument.
If you plan to control more than one Smart-Trak 2 instrument
from your computer…
If your computer has a number of serial ports equal to the number of instruments
you wish to operate, simply plug each instrument into a separate serial port as
directed above in “If your Computer has a Serial Port…”
If you wish to operate more instruments than your computer has serial ports, you
will usually have to use USB ports instead. If the number of USB ports in your
computer equals the number of instruments, then purchase one serial to USB
converter per Smart-Trak 2 and follow the instructions above. If you have more
instruments than ports, you will need to acquire an edge port device (converts one
USB port into several serial ports) or a USB hub (converts one USB port into many
USB ports). Once you have attached the necessary device and loaded the proper
drivers, connect your Smart-Trak 2 instruments to the serial or USB ports as
outlined in the sections above. Open a separate software window for each meter
and you can monitor all of them at one time.
If connecting your computer to the Smart-Trak 2 creates any confusion, please
contact Sierra Instruments or your IT person for assistance.
53
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Running the Smart-Trak 2 Software
Locate the file named “Smart-Trak 2” and open it. You will see the following
screen: This is the main data screen. First, you must select the proper comport.
Select the comport pull down menu:
Use the pull-down menu to choose the port number that corresponds to the serial
port channel your Smart-Trak 2 is connected to (from 1 to 255). If you have only
one serial port, select “Comm Port 1.” When finished, click on the box marked
“OK.”
You will be returned to the the Smart-Trak 2 Master Screen. The screen is the same
for both meters and controllers. Controller functions will not be active for meters
and vice versa
54
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Establishing Communication
If the Screen appears with all the yellow and white boxes full, the Smart-Trak 2
instrument has established communication with your computer.
Proceed to the next section.
If the screen appears, but the yellow and white boxes are all empty, you have
selected the wrong “Com Port number.” See the section titled “Com Port” on page
5-16 for instructions on how to change the communication port number.
If the screen appears, but only the Mass Flow box shows a value, you are using the
correct Com Port, but your computer is not talking to the Smart-Trak 2. To begin
communication, move the mouse pointer to the top center screen and left click on
the words “Get Parameters.” then values will fill all the boxes on the screen.
Communication has been established. Your computer is now able to control your
Smart-Trak 2 instrument.
Your Smart-Trak 2 instrument is now ready for use!
55
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Using the Smart-Trak 2 Software
A. Upper Section of Software Window
Across the upper half of the Smart-Trak 2 Software window you will see 3 yellow
boxes. These are titled:
Flow
Setpoint (flow controllers only. Will show zero for meters)
Full Scale
These boxes display the current operating conditions of your Smart-Trak 2 instrument.
If you have the Pilot Module, these boxes on the upper half of the software screen are
identical to the Upper Level Screens on the Pilot Module discussed in Chapter 4. The
features of these boxes are described below.
Flow
The box displays the mass flow rate, the engineering units, the gas choice and the
analog output choice.
56
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Setpoint
If you have a Mass Flow Meter, this box is zero (as shown above). The Setpoint box
displays the current setpoint given to the flow controller, the engineering units, the
source of the setpoint signal, the current valve state (open, closed or purge) and the dac
values for the valve (for Sierra troubleshooting only).
CAUTION: If this box does not show Pilot Module/RS-232 on the first line,
you will not be able to give the controller a setpoint command from your
computer. This is because you Smart-Trak 2 is expecting an analog setpoint at
the 15-pin mini- D-connector. To supply a setpoint from your computer, see
“Change Setpoint Source” on page 5-13.
Note on Valve
If you have a Mass Flow Meter, this information can be bypassed. For normal
operation of the flow controller, this line should display
visible, the instrument will automatically control flow as soon as a setpoint is given to
it.
Automatic. If this is
57
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If this box does not display “Automatic,” it will not be possible to control gas flow.
Your valve is locked open (Purge) or closed (Closed). See the section titled “Change
Valve Operation” below for further instructions
Full Scale
This screen displays the current full-scale value of the instrument with engineering
units. It also displays the com port selected and a counter for the com port (for Sierra
Troubleshooting only).
B. Lower Section of Software Window– Changing
Parameters
Across the bottom half of the screen you will find a number of white boxes. Each
box allows you to adjust one or more meter functions. The various functions are
reviewed in the following section.
58
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Change Setpoint Value (flow controllers only)
This is the box that allows you to change the setpoint of the mass flow controller.
To make a change to the setpoint value, move your pointer to this box and left click.
You may change a digit or delete the current setpoint and key in a new one. When
you have made your selection, press the enter key to implement it.
If, instead of using the RS-232 link or the Pilot Module, you prefer to supply analog
set point signal to the Smart-Trak 2, activate the pull-down menu. Make your
selection carefully and click. Your selection will appear in the box and, after a few
seconds, it will also appear in the yellow setpoint box.
CAUTION: If you change the source of the setpoint to one of the analog
values, you will not be able to control your Smart-Trak 2 with your computer.
Change Valve Operation –Automatic, Close, Purge
This function enables or overrides any setpoint command given to the Smart-Trak
2. At start-up, the box will show Automatic:
59
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
In this normal operating position, the valve is ready to control flow when it receives
a setpoint from some source. Using this box, you may set the valve to open well
beyond its normal maximum position (“Purge”) or force the valve to remain shut
regardless of setpoint (“Closed”). Use the pull down menu to make this change.
Choose “purge” to fully open the valve regardless of the given setpoint or choose
“closed” to override any setpoint command and close the valve.
CAUTION: Do not activate the valve state pull-down menus unless you are
ready to make this change. Once in the pull-down menu, if you click on one of
the selections, the valve operation will change immediately. Remember that
the Smart-Trak 2 is not a positive shut-off device.
Change Gas (Dial-A-Gas
This box will display the current gas selection. If you wish to change the gas used
in the instrument, activate the pull-down menu. The 10 gases programmed into the
memory of your Smart-Trak 2 will be displayed.
TM
)
Choose the correct gas for your current application and click. The new gas
selection will appear in the box. Within a few seconds, the new gas selection will
also appear in the yellow Flow box and the values of the mass flow, setpoint and the
full scale will be adjusted for this new gas. This is normal. In some cases, where
the thermal properties of one gas are very similar to another, these values may not
change very much.
60
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Change Units
This box will show the current engineering units. Should you wish to change the
engineering units, use the pull-down menu.
Make your selection, noting that each engineering unit is available with 3 different
time constants: seconds, minutes and hours. Choose the combination of units and
time constants you desire, then click. The box will display your new selection.
Within a few seconds, the units will change in the Flow, Setpoint, and Full Scale
boxes and the numerical values will be adjusted into the new units.
Change Analog Out
This box will show the two analog output signals currently selected for your
instrument. To re-configure the analog output signals, use the pull-down and make
your selection. The Smart-Trak 2 will adjust the analog output voltage signal per
your instruction.
61
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
C: ADJUSTMENTS:
Clicking on “adjustments” brings up the meter settings screen.
This screen allows for zeroing, spanning and adjusting the full scale for each of the
10 gases in the gas table individually. The screen will be displayed for the gas
selected in the gas table on the main screen
Caution: the values in this screen can be set individually for each of the 10 gases.
MODIFICATIONS of this screen will only be applicable to the gas displayed. To
modify other gases, change the gas in the pull down menu and repeat.
62
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Change Zero:
This button automatically zeros the meter.
To perform this zero procedure, you must make sure that the actual flow is zero.
Attempting to zero the meter or controller while there is a flow condition will cause
an offset. Note: This MUST be done using the appropriate gas as selected.
Caution: do not perform auto zero when there is no gas present in the meter and or
the gas is FLOWING.
Default zero:
This will set the zero value back to factory conditions. Use this in order to recover
from a mistaken entry.
63
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Span:
This allows the user to enter an offset to the span to adjust the meter outputs to
match another device or field conditions. The span is a MULITPLIER and will
multiply all outputs by the entered amount.
For example, the meter reads 98. The desired reading is 100. A span value of 1.02
will adjust the meter output to read 100.
Caution: this number should not be adjusted beyond 0.8 to 1.2.
Change full scale:
This setting allows you to change the full scale of your instrument
64
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
The instrument will display its current full-scale value. If your instrument is new or
this value has never been modified, the value displayed will be 100% of the
maximum flow rate your Smart-Trak 2 can measure or control without factory recalibration. This is the number listed here:
The instrument’s maximum flow rate is also recorded on the data label (at the back
of the instrument) and on the calibration certificate.
You may select any full-scale value between 100% and 50% of the maximum fullscale value. If you select a new full-scale value, it will re-define the analog outputs
of the instrument. The 20 mA signal and the corresponding voltage signal (5 or 10
VDC) will now represent this new full-scale value.
To change the full-scale value of your device, modify one digit or erase the current
value displayed and type in your desired full-scale value. When you are satisfied
with your selection, press the enter key. The new full-scale value will appear in the
white box. A few seconds later, the same value will appear in the yellow box in the
upper half of the screen titled “Full Scale.”
65
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CAUTION: The Change Full Scale function allows you to re-range your
instrument, but it is important to be aware of the limitations of this function.
Changing beyond recommended values may require a new LFE or even flow
body be installed. If in doubt, select factory reset and both zero and span
values are reset to factory values
Here are some guidelines to remember if you choose to modify the Full Scale
Value: Changing the full-scale value of the instrument does not affect the accuracy
of the measurement. Instrument accuracy is a percentage of the original factory
full-scale value
If you enter a full-scale value beyond the instrument’s calibrated maximum range,
the Smart-Trak 2 will automatically modify the full-scale to equal the factory fullscale value. Below 1% of the original factory full-scale value for a meter and 2%
for a controller, your Smart-Trak 2 is programmed to read zero because the error in
the measurement may exceed the measurement itself.
Other Useful Features
There are several additional features that you may utilize via the Smart-Trak 2
Software. Across the top of the Smart-Trak 2 Master Screen, you can see these
options:
Com Port
Read Parameters
Info
To use any of these functions, simply place the mouse pointer onto the name of the
feature you wish and left click.
66
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Com Port
To change the communication port for the Smart-Trak 2, click on this feature. Use
the pull down menu to choose the Comm Port that your Smart-Trak 2’s serial
connector is plugged into. Remember that the software asks you to confirm your
Comm Port every time you re-start. This feature is designed for situations when you
have multiple Comm Ports available and you wish to move the Smart-Trak 2 from
one to another or you are controlling more than one Smart-Trak 2 at one time.
Read Controller Parameters
Use this function if you believe your computer and your Smart-Trak 2 instrument
may have stopped communicating. It is essentially a re-boot command for the
Smart-Trak 2 microprocessor. When you switch Comm Ports, it is recommended
that you use this function. If your computer requires a re-boot and you do not
unplug your Smart-Trak 2, the communication link will be broken. Use this
function to re-establish communication. If at any time, the instrument does not
appear to be sending correct information to the Smart-Trak 2 software, try this
command to ensure communication.
Info
This gives information about the meter serial and software revisions.
67
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
CHAPTER 6: TECHNICAL SUPPORT & SERVICE
Technical Support
If you encounter any problem with your instrument, review the configuration
information for each step of the installation, operation, and set up procedures as
explained in this manual. Verify that your settings and adjustments are consistent
with factory recommendations.
If the problem persists, Sierra is eager to help you. You may contact us at any of the
following Technical Support Centers. It may also help to call your Sierra Sales
Agent, who is also well trained in the operation of the product.
IMPORTANT: When contacting Technical Support, make sure you have
included the following information:
The flow range, serial number, Sierra order number and model number (all
marked on the instrument data label).
The problem you are encountering and any corrective action taken.
Application information (gas, pressure, temperature, pipe and fitting
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Sierra Websites
Sierra Instruments, Inc. (USA Headquarters):www.sierrainstruments.com
Sierra Instruments b.v. (Europe): www.sierrainstruments.nl
Sierra Asia Instruments (China): www.sierra-asia.com
69
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Returning Equipment to the Factory
Factory Calibration—All Models
Sierra Instruments maintains a fully-equipped calibration laboratory. All measuring
and test equipment used in the calibration of Sierra transducers are traceable to
NIST Standards. Sierra is ISO-9001 registered and conforms to the requirements of
ANSI/NCSL-Z540 and ISO/IEC Guide 25.
Instructions for Returning Your Instrument for Service
The following information will help you return your instrument to Sierra
Instruments' Factory Service Center and will ensure that your order is processed
promptly. Prices may vary depending on the flow range, type of gas and operating
pressure of your unit. To request detailed pricing contact your local Sierra
Instruments distributor or contact one of our offices directly. Our expedite fees are:
three-day turnaround 25%, two-day turnaround 40%.
Please follow these easy steps to return your instrument for factory
service:
1. Obtain a Return Materials Authorization (RMA) number from Sierra
Instruments. You may obtain this from the factory by calling (800) 866
0200 between 8:00 a.m. and 5:00 p.m. PST Monday through Friday. You
may also obtain this number via e-mail by contacting
service@sierrainstruments.com.
2. Once you have obtained an RMA number, complete the RMA form. If you
require service beyond calibration, but do not know which service(s) will be
required, describe the symptoms as accurately as possible on the RMA form.
Submit electronically or by fax to (831) 373-2414.
3. Pack your instrument carefully. Use the original packaging and foam or
bubble wrap (packing peanuts NOT recommended) and include a copy of
the RMA form (complete with Sierra supplied RMA number) with the
unit(s). This is particularly important when shipping the medium and high
flow versions. Due to their weight, they can be damaged in transit if not
packed properly.
70
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
4. Ship the unit(s) to the following address:
RETURN ADDRESS:
Sierra Instruments, Inc.
Attention: Factory Service Center
5 Harris Court, Building L
Monterey, CA 93940 USA
71
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
IMPORTANT SAFETY NOTE ABOUT PURGING
WARNING: When toxic or corrosive gases are used, purge unit
thoroughly with inert dry gas before disconnecting from the gas
line to prevent personnel from being injured when coming in
contact with the instrument.
WARNING: If an instrument used with a toxic or corrosive gas is
returned to the factory, a Material Safety Data Sheet (MSDS)
must be enclosed & attached to the outside of the box to alert
Sierra personnel of the potential hazard. Also, make sure the
inlet & outlet are solidly plugged off.
72
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Appendix A: Gas Tables
Smart-Trak 2 Pre-programmed Gases: Dial-A-Gas
The following gases have been programmed into the Smart-Trak 2 instrument in
this order. If you are using one of these gases, you may use the Dial-A-Gas feature
in either the Pilot Module or the Software Package and the instrument will adjust
the outputs automatically.
1. Air
2. Argon
3. CO2
4. CO
5. Helium
6. Hydrogen
7. Methane
8. Nitrogen
9. Nitrous Oxide
10. Oxygen
K-Factor Calculations-Using Smart-Trak 2 with Other Gases
If you will be using Smart-Trak 2 with a gas not on this list, you may use the tables
below. They provide K-factors and thermodynamic properties of gases commonly
used with mass flow meters and controllers. This is particularly useful if the actual
gas is not a common gas or if it is toxic, flammable, corrosive, etc. The tables can
also be used to interpret the reading of a flow meter or flow controller that has been
calibrated with a gas other than the actual gas.
Before applying the tables, set the instrument for Air. Then, the following fundamental
relationship may be used:
Q
Where:
Q = The volumetric flow rate of the gas referenced to standard
conditions of 0°C and 760 mm Hg (sccm or slm),
K = The K-factor from the following tables, referenced to Air
( )
= Refers to the “actual” gas, and
1
= Refers to the “reference” gas, Air in this case.
( )
N
= K1/KN
1/QN
73
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
If no O-ring material is specified then O-ring to be used is Viton. NEO is neoprene or equivalent. KR is
DuPont Kalrez or equivalent. Valve Seat applies only to controllers.
76
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Appendix B: Product Specifications
Performance Specifications:
Accuracy
Standard Calibration: ±1.0% of full scale at operating conditions
High Accuracy Calibration: ±0.7% of reading plus 0.3% of full scale at
operating conditions.
Dial-A-Gas: ±1.0% of full scale in all 10 gases
Repeatability
± 0.2% of full scale
Temperature Coefficient
±0.025% of full scale per °F (±0.05% of full scale per °C), or better
Pressure Coefficient
±0.01% of full scale per psi (±0.15 of full scale per bar), or better
Response Time
300 millisecond time constant--2 seconds (typical) to within ±2% of final
value (includes settling time)
Operating Specifications:
Gases
All clean gases including corrosives; specify when ordering. The following
10 gases make up the standard Dial-A-Gas feature. Alternate gases may be
substituted at time of ordering. Flow rates in slpm.
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Mass Flow Rates
101 Micro-Trak: 0 to 4 sccm
100L low flow body: 0 to 10 sccm to 0 to 50slpm
100M medium flow body: 0-20 to 0-300 slpm
100H high flow body: 0-100 to 0-1000 slpm
Flow ranges specified are for nitrogen at 760 mm Hg and 21°C (70°F) or for an equivalent;
other engineering units are available (e.g., nlpm, scfh, nm
Note: tests performed at 21 degrees C, outlet at ambient pressure
*Larger fittings recommended for these flow rates because these fittings reduce overall
performance.
78
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Differential Pressure Requirement for Controllers
(lower or higher available upon request)
Optimum: 30-60 psi (2-4 bar)
Minimum: See chart below. Note that required pressure increases with flow rate.
Minimum Differential Pressure Requirement for Air, Mass Flow Controllers
Note: tests performed at 21 degrees C, outlet at ambient pressure
*Larger fittings recommended for these flow rates because these fittings reduce overall
performance
Gas & Ambient Temperature
32 to 122°F (0 to 50°C)
Leak Integrity
5 X 10
-9
standard cc/sec of helium maximum
Power Requirements
Ripple should not exceed 100 mV peak-to-peak
For Mass Flow Meters: 15-24 ±10%, VDC, (230mA, regulated)
For Mass Flow Controllers: C100L: 24 VDC ±10%, (500mA, regulated)
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Control Range
2–100% of full scale flow; automatic shut-off at 1.9%.
Output Signals--Analog:
Linear 4–20 mA, 500 ohms maximum loop resistance plus one of the following
(user selectable):
Linear 0–5 VDC, 1000 ohms minimum load resistance
Linear 0-10 VDC, 1000 ohms minimum load resistance
Linear 1-5 VDC, 1000 ohms minimum load resistance
Output Signals--Digital:
RS-232 (standard)
Pilot module (optional)
Command Signal--Analog (choice of one):
Linear 4–20 mA
Linear 0–5 VDC
Linear 0-10 VDC
Linear 1-5 VDC
Command Signal--Digital:
RS-232
Pilot module (optional)
Physical Specifications:
Wetted Materials
316 stainless steel or equivalent, 416 stainless steel, Viton® “O”-rings and valve
seat (standard). Other elastomers are available upon request.
Optional Parts & Accessories:
The following optional accessories are available from your Authorized Sierra
Instruments Representative or directly from Sierra Instruments. They are provided
to improve your experience with the Smart-Trak 2 Series 100 Mass Flow Meters
and Controllers. Call or write for current pricing and availability.
80
Sierra Instruments Instruction Manual Smart-Trak® 2 Series 100
Description
Remote Pilot Module RDO
Compod without display CM-NR
Compod with display CM-DD
Compod with Relays CM-Relays
Power Supply with D-connector attached, USA plug 100 T8D
Power Supply with fly leads (bare wires), USA plug 100 T8F
Power Supply with D-connector for C100H units 100 T10D
Power Supply with fly leads (bare wires) for C100H units 100 T10F
Multi-adapter with voltage regulator (to permit operation
15 pin mini-D mating connector only C0
Communication cable—6 inch C1
Communication cable—3 foot C3
Communication cable—10 foot C10
Communication cable—Custom length C ( ) specify length
RS-232 cable, 10 foot, with CAT 5 on one end and
Standard DB-9 connector on the other CRN
10 micron filter for M100L or C100L 100L Filter
10 micron filter for M100M or C100M 100M Filter
10 micron filter for M100H or C100H 100H Filter
Note: Pins 1, 3, and 10 are connected together inside the instrument. Do
not tie these grounds together outside the instrument. Must have one
connection per analog ground. Recommended use listed.
83
Sierra Instruction Manual Smart-Trak® 2 Series 100
Appendix E: Product Dimensions and Mounting
All dimensions are in inches with millimeters in (brackets). Dimension “L” is listed in the Table at
the end of Appendix E.
84
Sierra Instruction Manual Smart-Trak® 2 Series 100
85
Sierra Instruction Manual Smart-Trak® 2 Series 100
86
Sierra Instruction Manual Smart-Trak® 2 Series 100
Appendix F: Micro-Trak
The information included in this section applies to the Model 101 Micro-Trak mass flow
meters and controllers only.
Introduction
The Micro-Trak series of flow instruments from Sierra are designed to measure and
control ultra low flows of clean gas. Special considerations must be taken to obtain
optimum performance from your Model 101 Micro-Trak instrument.
Installation
Because the gas flows used with Micro-Trak are extremely small, the plumbing of these
instruments has a dramatic impact on their performance. Sierra recommends all
installations use tubing of 1/8 inch diameter or smaller. Larger tubing may be used, but a
reduction in the response time of the instruments may be noticed. It is also important to
minimize large volumes within the gas systems because these will act as manifolds and
introduce delays in flow measurement and control.
Micro-Trak must be securely mounted in a location away from vibrations or it may not
function normally. Periodic impacts will upset the delicate equilibrium of Micro-Trak
and should be avoided.
Any gas used with Micro-Trak must be extremely clean and dry so it does not
contaminate the micro passages inside the instrument. An upstream 10 micron particle
filter coupled with water and oil traps are suggested.
Inlet gas pressure must be very well regulated for optimum performance from MicroTrak. The M101 flow meter can be over-ranged by a pressure change of 10 in H20 (0.36
psi or 25 mbar). The C101 flow controller can temporarily lose control over a similar
pressure fluctuation. In either case, the Micro Trak will recover within several seconds,
but a delicate process could be affected. For these reasons, caution must be taken to
maintain stable pressure to obtain the excellent performance Micro-Trak was engineered
to deliver.
Electrical installation is identical to other 100 Series instruments. See Chapter 2 for
details.
Operation
Operation of Micro-Trak is similar to all other 100 Series instruments outlined in this
manual. All functions of the Smart-Trak 2 will function with Micro-Trak and can be
used with analog or digital signals, per the instructions in Chapters 3, 4 and 5.
87
Sierra Instruction Manual Smart-Trak® 2 Series 100
Performance Specifications—Micro-Trak
Accuracy
1% of full scale including linearity at operating conditions
+
Dial-A-Gas
+
1% of full scale in all 10 standard gases
Repeatability
0.2% of full scale
+
Temp Coeff
+
0.025% of full scale per ºF (+0.05% of full scale per ºC) or better
Pressure coeff
+0.01% of full scale per psi (+0.15% of full scale per bar) or better
Response Time
Governed by total volume of installation. If suggestions above are followed, response
time will be similar to all other Smart-Trak 2 instruments.
Operating Specifications
All clean gases, including corrosives and toxics. The following 10 gases make up the
standard Dial-A-Gas feature. Alternate gases may be substituted at time of order.
500 psig (34 barg) maximum pressure; burst tested to 750 psig (52 barg)
Pressure Drop Across a Micro-Trak Meter
0.36 psi (24.5 mbar) with 1/8 inch compression fittings
Differential Pressure Requirement for Micro-Trak Flow Controllers
30 psi (2040 mbar) optimum
1 psi (68 mbar) minimum with 1/8 inch compression fittings
88
Sierra Instruction Manual Smart-Trak® 2 Series 100
Leak Integrity
5 X 10-9 atm cc/sec of helium maximum
Power Requirements
Ripple should not exceed 100 mV peak-to peak
For Mass Flow Meters: 15-24 ±10%, (230mA, regulated)
For Mass Flow Controllers: C100L: 24 VDC ±10%, (500mA, regulated)
NOTE: The Compod™ adds an additional 100mA to the ST2 current ratings.
Control Range
2–100% of full scale flow; automatic shut-off at 2%.
Output Signals
Analog:
• Linear 4–20 mA, 500 ohms maximum loop resistance plus one of the following
(user selectable):
• Linear 0–5 VDC, 1000 ohms minimum load resistance
• Linear 0-10 VDC, 1000 ohms minimum load resistance
• Linear 1-5 VDC, 1000 ohms minimum load resistance
Digital:
• RS-232
• Pilot module controller
Command Signal
Analog (choice of one):
• Linear 4–20 mA
• Linear 0–5 VDC
• Linear 0-10 VDC
• Linear 1-5 VDC
Digital:
• RS-232
• Pilot module controller
Physical Specifications--Wetted Materials
316 stainless steel; 416 stainless steel; synthetic ruby, Viton® “O”-rings and valve seat
standard. Other elastomers are available, consult factory.
Dimensions
Same as Model C100L, M100L. See Appendix E for details.
89
Sierra Instruction Manual Smart-Trak® 2 Series 100
Appendix G: Smart-Trak Basic Commands
IMPORTANT CUSTOMER NOTICE
This document is for firmware version 2.044. Sierra Instruments, Inc. reserves the right to
change the command set without notification to the user. However, Sierra will also make every
effort to maintain backwards compatibility with previously published command sets for Smart-
®
Trak
. There is no implied warranty or guarantee regarding the use of this commands set. Sierra
Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the
use of this command set.
PROTOCOL
The Smart-Trak communication is based on a standard RS232 port. All bytes are ASCII except
the two CRC bytes used for a redundancy check. Each command starts with four bytes of
command type, followed by a variable length value section. The two CRC bytes follow the
value string and a carriage return ends the string. Each command string will vary in length but,
can not exceed 26 bytes.
Command String = {4 command bytes + variable length value + 2 CRC bytes + carriage return}
< 26 bytes.
The RS232 port does not depend on hardware handshaking and uses only three wires on the port:
transmit, receive and ground. The port on the host needs to be configured to (9600,n,8,1) : 9600
baud, no parity, eight bit characters, one stop bit.
DESCRIPTION
Commands for the Meter. There are old commands Smart-Trak (version 1.xx) and new
commands Smart-Trak 2 (version 2.xx). The 2.xx commands will support some of the
1.xx commands. If a 1.xx commands is not shown below, it is not supported in the 2.xx
commands. CMNDS is a class included in the class of Meter. This page shows all types,
fields, and values as derived from Meter. Version 2.xx commands that start with a '?' are
read commands and commands that start with a '!' are write commands. This is list of the
most commonly used commands, which are relative safe.
We can supply a list of more advanced commands, however you will need to agree to
some special terms. Misuse of the Advanced commands can cause irreversible damage to
the unit which would void your warranty.
90
Sierra Instruction Manual Smart-Trak® 2 Series 100
FLOW
DESCRIPTION
FLOW makes the streaming mode data = flow only. If meter is in streaming mode, this
data will stream until a FLOWwSETPOINT, FLOWwSETPOINTwDAC or
TOTALwDIFFwDAC is sent to the meter. If not in streaming mode then one flow will be
return with each FLOW command sent. ReadMeter, WriteMeter both will set flow as the
streaming data.
ASCII STRING COMMANDS
'?Flow + CRC + cr' Version 2.xx read command
'!Flow + value + CRC + cr' Version 2.xx write command
'?Flow + CRC + cr' Version 1.xx read command
Returns ASCII string 'Flow + value + CRC + cr'
Values are a string of digits and an optional decimal place: '10.00'
SETPOINT
DESCRIPTION
SETPOINT sets the control setpoint. The set point is store in the flash memory. This
command is not recommended for version 2.xx,. Use SETPOINT_FLASH.
ASCII STRING COMMANDS
'?Sinv + CRC + cr' Version 2.xx read command
'!Sinv + value + CRC + cr' Version 2.xx write command
'?Sinv + CRC + cr' Version 1.xx read command
'Sinv + value + CRC + cr' Version 1.xx write command
Values are a string of digits and an optional decimal place: '10.00'
CRC=redundancy check bytes; cr=carrage return byte.
SETPOINT_FLASH
DESCRIPTION
SETPOINT_FLASH returns the current flash memory value setpoint with read
command. The write command sets the flash memory value and makes it the active
setpoint. This would be the setpoint after a power down.
91
Sierra Instruction Manual Smart-Trak® 2 Series 100
ASCII String commands
'?Setf + CRC + cr' Version 2.xx read command
'!Setf + value + CRC + cr' Version 2.xx write command
Values are a string of digits and an optional decimal place: '10.00'
CRC=redundancy check bytes; cr=carrage return byte.
SETPOINT_RAM
DESCRIPTION
SETPOINT_RAM returns the current ram memory value setpoint with read command.
The write command sets the ram memory value and makes it the active setpoint.
ASCII STRING COMMANDS
'?Setr + CRC + cr' Version 2.xx read command
'!Setr + value + CRC + cr' Version 2.xx write command
unitIndex is an integer string value of 1 to 30
CRC=redundancy check bytes; cr=carrage return byte.
VALVE_INDEX
DESCRIPTION
VALVE_INDEX selects the state of the valve. Automatic = 1, Closed = 2, Purge = 3.
This is a write to flash memory command. For version 2.xx, this command is not
recommended. See VALVE_FLASH_INDEX and VALVE_RAM_INDEX.
ASCII STRING COMMANDS
'?Vlvi + CRC + cr' Version 2.xx read command
'!Vlvi + valveIndex + CRC + cr' Version 2.xx write command
'?Vlvi + CRC + cr' Version 1.xx read command
'Vlvi + valveIndex + CRC + cr' Version 1.xx write command
valveIndex is an integer string value of 1 to 3
CRC=redundancy check bytes; cr=carrage return byte.
GAS_INDEX
DESCRIPTION
GAS_INDEX selects the current gas.
ASCII STRING COMMANDS
'?Gasi + CRC + cr' Version 2.xx read command
'!Gasi + gasIndex + CRC + cr' Version 2.xx write command
'?Gasi + CRC + cr' Version 1.xx read command
'Gasi + gasIndex + CRC + cr' Version 1.xx write command
Returns ASCII string 'Gasi + gasIndex + CRC + cr'
gasIndex value is 1 through 10
CRC=redundancy check bytes; cr=carrage return byte.
STREAM
DESCRIPTION
Communications can be in one of three mode set by this STREAM command. In off
mode, the meter will respond when queried with a read command. In echo mode, the
meter will respond with either a read or a write command. In stream mode, the meter will
continuously send data back. If a read or write command needs to update more than one
93
Sierra Instruction Manual Smart-Trak® 2 Series 100
value, all values with be sent back on write command.
Sync returns the following events, VersionNumberEvent, SerialNumberEvent,
DataEvent, ManufactureNumberEvent, TypeEvent, PassWordEvent, SetpointEvent,
FullScaleEvent, GasNameEvent(10 times for all gas names), GasIndexEvent,
94
Sierra Instruction Manual Smart-Trak® 2 Series 100
GasSpanEvent(current gas only), SetpointIndexEvent, OuputIndexEvent,
UnitIndexEvent, ValveFlashEvent, StreamEvent, SyncEvent
ASCII STRING COMMANDS
'?Sync + CRC + cr' Version 2.xx read command
Returns ASCII string 'Sync + "" + CRC + cr'
SyncEvent returned last to indicate all events have been sent
CRC=redundancy check bytes; cr=carrage return byte.
ZERO
DESCRIPTION
ZERO sets the flow offset value to a zero flow reading. Warning: All flow must be shut
off, the unit needs to be at pressure with the gas being used.
Below is the routine used to calculate the CRC bytes in C#.
Comments are between the /* */.
privatestaticuint CalcCRC(byte[] cmnd)
/* cmnd is a byte array containing the command ASCII string … cmnd[]=”Sinv2.000” */
/* An unsigned 32 bit integer is return to the calling program */
95
Sierra Instruction Manual Smart-Trak® 2 Series 100
/* only the lower 16 bits contain the crc */
{
int i,j; /* interating indexes for the for loops */
uint crc; /* crc variable that will be returned */
crc=0xffff; /* initialize crc to hex value 0xffff */
for (i=0; i<cmnd.Length; i++)
/* this for loop starts with ASCCII ‘S’ and loops through to the last ASCII ‘0’ */
{
crc=crc^((uint)(cmnd[i]*0x0100));
/* the ASCII value is times by 0x0100 first then XORED to the current crc value */
for(j=0; j<8; j++)
/* the crc is hashed 8 times with this for loop */
/* if the 15th bit is set (tested by ANDING with hex 0x8000 and testing for 0x8000 result)
then crc is shifted left one bit (same as times 2) XORED with hex 0x1021 and ANDED
to hex 0xffff to limit the crc to lower 16 bits. If the 15th bit is not set then the crc is shifted
left one bit and ANDED with hex 0xffff to limit the crc to lower 16 bits. */
{
if((crc&0x8000)==0x8000)
crc=((crc<<1)^0x1021)&0xffff; else
crc=(crc<<1)&0xffff;
/* end of j loop */ }
/* end of i loop */ }
/* There are some crc values that are not allowed, 0x00 and 0x0d */
/* These are byte values so the high byte and the low byte of the crc must be checked and
incremented if the bytes are either 0x00 0r 0x0d. */
if((crc&0xff00)==0x0d00) crc +=0x0100;
if((crc&0x00ff)==0x000d) crc +=0x0001;
if((crc&0xff00)==0x0000) crc +=0x0100;
if((crc&0x00ff)==0x0000) crc +=0x0001;
return crc;
}
/* If the string Sinv2.000 is sent through this routine the crc = 0x8f55 */
/* The complete command “Sinv2.000” will look like this in hex
0x53 0x69 0x6E 0x76 0x32 0x2e 0x30 0x30 0x30 0x8f 0x55 0x0d */
Below is the c# routine with no comments.
privatestaticuint CalcCRC(byte[] cmnd)
{
int i,j;
uint crc;
crc=0xffff;
96
Sierra Instruction Manual Smart-Trak® 2 Series 100
Public Sub SendPacket(packet As String)
Dim crcBytes() As Byte
Dim crcLen, vpacket
Dim crcWord As Long
Dim i, j
packet = Replace(packet, ",", ".") ' if commas are used as decimal place holder
crcLen = Len(packet) - 1 ' zero based counter
ReDim crcBytes(0 To crcLen)
For i = 0 To crcLen ' convert string to bytes
crcBytes(i) = Asc(Mid(packet, i + 1, 1)) ' get character
Next
' calc crc
crcWord = (&H1FFFE) / 2 ' not -1
For j = 0 To crcLen
crcWord = crcWord Xor CLng(crcBytes(j)) * &H100
For i = 1 To 8
If (crcWord And &H8000) Then
crcWord = (crcWord * 2) Xor &H1021 ' prime number
Else
crcWord = crcWord * 2
End If
crcWord = (((crcWord * 2) And &H1FFFE)) / 2
Next i
97
Sierra Instruction Manual Smart-Trak® 2 Series 100
Next j
If (crcWord And &HFF00) = &HD00 Then ' carriage char not allowed
crcWord = crcWord + &H100 ' make this byte &h0exx
End If
If (crcWord And &HFF00) = &H0 Then ' zero char is death
crcWord = crcWord + &H100 ' make this byte &h01xx
End If
If (crcWord And &HFF) = &HD Then ' carriage char not allowed
crcWord = crcWord + 1 ' make this byte &hxx0e
End If
If (crcWord And &HFF) = &H0 Then ' zero char is death
crcWord = crcWord + 1 ' make this byte &hxx01
End If
ReDim Preserve crcBytes(0 To crcLen + 3) ' make room for crc and carriage return
crcBytes(crcLen + 1) = Int(crcWord / &H100) ' high byte
crcBytes(crcLen + 2) = (crcWord And &HFF) ' low byte
crcBytes(crcLen + 3) = 13 ' carriage return
vpacket = crcBytes
commQueue.Add vpacket
End Sub
98
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.