Sierra CalTrak 800 Series, SL 800-24, SL 800 Series, SL 800-3, SL 800-10 Instruction Manual

...
Page 1
CalTrak® 800
High Accuracy Modular Primary
Cal Labs & Industry Applications
Instruction Manual
Models: SL800
Part Number: IM-Cal-SL800 Rev. F 4/16
1
Page 2
GLOBAL SUPPORT LOCATIONS: WE ARE HERE TO HELP!
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940 Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
www.sierrainstruments.com
EUROPE HEADQUARTERS
Bijlmansweid 2 1934RE Egmond aan den Hoef The Netherlands Phone +31 72 5071400 Fax +31 72 5071401
ASIA HEADQUARTERS
Second Floor Building 5, Senpu Industrial Park 25 Hangdu Road Hangtou Town Pu Dong New District, Shanghai, P.R. China Postal Code 201316 Phone: + 8621 5879 8521 Fax: +8621 5879 8586
© COPYRIGHT SIERRA INSTRUMENTS 2016
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change without notice.
TRADEMARKS
CalTrak™, CalTrak®, CalSoft®, and SmartTrak® are trademarks of Sierra Instruments, Inc. Other product and company names listed in this manual are trademarks or trade names of their respective manufacturers.
2
Page 3
Warnings and Cautions
Warning!
Caution!
This statement appears with information that is important to protect people and equipment from damage. Pay very close attention to all warnings that apply to your application.
This statement appears with information that is important for protecting your equipment and performance. Read and follow all cautions that apply to your application.
Note and Safety Information
We use caution and warning statements throughout this book to draw your attention to important information.
Warning!
flow meter nameplate for specific flow meter approvals before any hazardous location installation.
Agency approval for hazardous location installations varies between flow meter models. Consult the
Warning!
Warning!
All wiring procedures must be performed with the power off.
To avoid potential electric shock, follow National Electric Code safety practices or your local code when wiring this unit to a power source and to peripheral devices. Failure to do so could result in injury or death. All AC power connections must be in accordance with published CE directives.
Warning!
Do not power the flow meter with the sensor remote (if applicable) wires disconnected. This could cause over­heating of the sensors and/or damage to the electronics.
Warning!
Before attempting any flow meter repair, verify that the line is de-pressurized.
Warning!
Always remove main power before disassembling any part of the mass flow meter/controller.
Caution!
Before making adjustments to the device, verify the flow meter/controller is not actively monitoring or reporting to any master control system. Adjustments to the electronics will cause direct changes to flow control settings.
Caution!
When using toxic or corrosive gases, purge the line with inert gas for a minimum of four hours at full
gas flow before installing the meter.
Caution!
Caution!
The AC wire insulation temperature rating must meet or exceed 80°C (176°F).
Printed circuit boards are sensitive to electrostatic discharge. To avoid damaging the board, follow
these precautions to minimize the risk of damage:
before handling the assembly, discharge your body by touching a grounded, metal object handle all cards by their edges unless otherwise required when possible, use grounded electrostatic discharge wrist straps when handling sensitive component
3
Page 4
Receipt of System Components
When receiving a Sierra mass flow meter, carefully check the outside damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered components are present. Make sure any spare parts or accessories are not discarded with the packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service.
packing
carton for damage incurred in shipment. If the carton is
Technical Assistance
If you encounter a problem with your flow meter, review the configuration information for each step of the installation, operation, and setup procedures. Verify that your settings and adjustments are consistent with factory recommendations. Installation information can be found in Chapter 2 of this manual, and troubleshooting information can be found in Chapter 10.
If the problem persists after following the troubleshooting procedures outlined in this manual, contact Sierra Instruments by fax or by E-mail (see inside front cover). For urgent phone support you may call (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00 p.m. PST. In Europe, contact Sierra Instruments Europe at Instruments Asia at +
86-21-58798521.
The flow range, serial number, and Sierra order number (all marked on the meter nameplate)
The software version (visible at start up)
The problem you are encountering and any corrective action taken
Application information (gas, pressure, temperature and piping configuration)
+31 (0)72-5071400
When contacting Technical Support, make sure to include this information:
. In the Asia-Pacific region, contact Sierra
4
Page 5
Table of Contents
Table of Contents ............................................................................................................................... 5
Chapter 1: Introduction ....................................................................................................................... 6
Chapter 2: Using Your CalTrak 800 ................................................................................................... 8
Chapter 3: Using the CalTrak 800 Touchscreen ................................................................................ 9
Chapter 4: Installing and Removing Flow Cells ................................................................................ 14
Chapter 5: Factory Default Settings ................................................................................................. 15
Chapter 6: Connecting the CalTrak 800 to Gas Flow Source ........................................................... 16
Chapter 7: Taking Readings ............................................................................................................. 17
Chapter 8: Using Sensor Factors for MFC and MFM Calibration ..................................................... 18
Chapter 9: Error Messages .............................................................................................................. 19
Chapter 10: Annual Maintenance and Calibration ............................................................................ 20
Chapter 11: Storage ......................................................................................................................... 22
Chapter 12: Product Specifications .................................................................................................. 23
Appendix A: Warranty Policy ............................................................................................................ 24
5
Page 6
Chapter 1: Introduction
2
Congratulations! You’ve chosen the CalTrak SL 800, featuring proven CalTrak® technology from Sierra. The CalTrak SL 800 has five interchangeable cells that offer ±0.25% to ± 0.15% of reading standardized accuracy within a portable primary standard that’s ideal for both
volumetric and standardized flow applications where a high degree of measurement accuracy is required. Before you get started, we recommend taking a few moments to review this manual and familiarize yourself with the CalTrak SL 800. If at any time you have questions regarding its operation, please contact Sierra through our web site (www.sierrainstruments.com), or call us at SL 800-866-0200 to speak with one of our knowledgeable customer support representatives.
Operation
The CalTrak SL 800 is a positive displacement primary piston prover for gas flow measurements in either pressure or vacuum applications. Using Proven CalTrak Technology, it combines the accuracy of a primary standard with unequaled speed and convenience. Volumetric or standardized flow readings are obtained with the push of a button. The CalTrak SL 800 can be set to take flow readings manually, one reading at a time, or automatically in the continuous auto-read mode. The CalTrak SL 800 can be programmed for up to 100 readings in an averaging sequence.
The versatile CalTrak SL 800 can be used to measure the gas flow rates at atmospheric pressure ± 7mmHg. Its Proven CalTrak Technology features a near-frictionless piston and flow cell design that provides highly-accurate, liquid-free flow measurements of unequaled speed and convenience. The CalTrak SL 800 consists of two primary components: The base and the flow cell. There are five interchangeable flow cell models, each with a designated flow range. The flow cell plugs into the base unit in order to create a complete, functional CalTrak SL 800 system. The base component houses the main computer and timing circuitry. The flow cell is easily fitted into the base unit’s interface connector and two guide pins.
The flow cell performs the actual flow measurements. For this purpose, it houses a “piston and glass assembly” comprised of a borosilicate glass tube housing a precision-machined graphite piston. The flow cell also contains an integrated temperature sensor and barometric pressure transducer in the gas flow stream for instant conversion of the volumetric readings into standardized flow.
CalTrak Pro software captures flow data from your CalTrak instrument to a PC.
The CalTrak Pro software allows you to:
View and plot the flow data in real-time. Export your data to a Windows environment. Enter flow data from a flow meter and compare the flow measurements from your CalTrak
Upload the latest version of the firmware to your CalTrak
®
.
®
precision calibrator.
6
Page 7
CalTrak SL 800 Electronic Base Power Supply Adapter RS-232 Serial Cable USB Cable Certificate of Calibration
CalTrak SL 800 Flow Cell Dust Caps Certificate of Calibration External Inlet Filter (SL 800-3 only)
Warnings!
The CalTrak SL 800 is not rated intrinsically safe and is not for use with explosive gasses or for
use in explosive environments.
The CalTrak SL 800 is not designed for gas flows above the rated specifications of the flow
cell in use. Please consult the product specification on page 23 of the manual for more information regarding acceptable gas flow ranges
For use with clean laboratory air or other inert gasses only.
Unpacking Checklist
Your CalTrak 800 has been packaged with care and includes all components necessary for complete operation. Please take a moment to check that you have received the following items. If you believe you have not received a full shipment or if you have any questions, please contact Sierra immediately.
Your CalTrak comes with the following:
Your SL800 Base with the following: Your SL800 Flow Cell Includes
Carrying cases and accessories are available for purchase from Sierra or your local Sierra Partner.
7
Page 8
Chapter 2: Using Your CalTrak 800
Getting Started
Ensure that the included 12V 3A AC power supply is connected to the base. Install the desired cell onto the base. All cells, except the -75 cell, should be installed with the blue adapter ring in place. If using the -75 cell, remove the adapter ring by rotating counter clockwise and lift off.
Turning the CalTrak 800 On and Off
Flip the power switch in the rear of the CalTrak SL 800 base to start the unit. An opening screen will appear followed by the splash screen.
The CalTrak SL 800 is designed to operate using only AC power, this unit does not have a battery. Ensure that your unit is plugged in prior to turning on.
Communications Ports
The CalTrak SL 800 includes a USB-B port and RS-232 serial port for bidirectional communication, including data download and operation using the CalTrak Pro software. The USB port includes a driver and does not require any additional driver download.
New firmware, as it becomes available, can be easily uploaded through the USB or RS-232 serial ports.
The side of the base contains a gold port that is used for pressure calibration at our facility. Nothing should be inserted into this port by the user.
On the back on the base is the clock signal port. This is used for verifying clock timing.
8
Page 9
Chapter 3: Using the CalTrak 800 Touchscreen
CalTrak 800 Touchscreen
The CalTrak SL 800 uses a resistive touchscreen display for easy navigation through menu options and data entry. Numeric values can be changed using the blue arrows on the right of the screen. At any time, the user may return to the home screen by pushing the home icon in the lower left of the screen.
Home Screen
The Home screen allows access to the Measure screen and all submenus. You can return to this screen at any time during operation by pressing the Home button in the lower left corner of the screen.
Measure Screen
The Measure screen allows you to capture and observe instantaneous flow data. During operation, the measured flow rate, reading number, standardization information, and average readings are displayed at the top. Gas pressure and temperature inside the tube are displayed on a constant basis regardless of whether the unit is operating.
By pressing on the blue and grey portions of the screen allow you to access additional user-defined parameters. By pressing the blue flow
9
Page 10
measurement section, you can change the sensor factor, PTVM, flow standardization, standardization temperature, and reading average
5
number. Pressing the grey gas temperature/pressure section allows you to change the pressure and temperature units. This opti on is only available through this menu.
Operations Screen
The Operation screen allows you to change measurement parameters including sensor factor, PTVM, standardized or volumetric flow, standardization temperature, and reading average number. When you are done changing these values, pressing CONFIRM will take you to the Measure screen. These input parameters can also be accessed by pressing the ‘plus’ symbol in the blue portion of the Measure screen.
Date/Time Delay Screen
The Date/Time/Delay screen allows you to change the date, time, and delay between individual measurements. To edit these fields, press on the gray box of the field you would like to change. The value highlighted will change by using the arrows on the right.
10
Page 11
About Screen
The About screen displays the base serial number, firmware revision, date of last cell calibration, cycles run since last calibration, calibration number, model of installed cell and cell serial number. Provide these numbers when contacting customer service or technical support.
Diagnostics Screen
11
Page 12
The Diagnostics screen shows important information regarding the base and cell pressure, cell temperature, and optical sensor reading. These are continuously monitored when the base is operating.
Setting User Preferences
The CalTrak SL 800 offers enhanced electronics options to allow the user to define parameters specific to an application.
These parameters may be entered from the Operation screen or by pressing the blue section of the Measure screen.
“Sensor Factor”
As applicable, change the sensor factor from its default value of 1.000 to the value provided by the MFC or MFM manufacturer. See Using Sensor Factors.
“PTVM”
PTV stands for Piston Tare Value, which is the amount of gas that passes around the piston during measurement. All CalTrak calibration equipment has a factory set PTV that is stored in the memory of the CalTrak cell. The value is typically very small: 0.008 for the ultra-low flow cells, 0.1 ccm for low flow cells, 0.2 ccm for medium flow cells, and 1.4 ccm for the high flow cells. We adjust for this leakage by adding PTV to the measurements.
On our highest accuracy instruments, we allow for the adjustment of the Piston Tare Value with the Piston Tare Value Multiplier (PTVM). When using the instrument with gas species other than air or nitrogen, the molecular behaviors of these gases may
degrade the PTV. For highest accuracy, the instrument’s PTV can be adjusted. Adjusting the PTV is accomplished by entering
a new PTVM. The PTV is multiplied by the PTVM to adjust the measurement. The default value for air and nitrogen is
1.000. The PTVM can be set to any value from 3.000 to 0.200.
For flows above 20 ccm, a new PTVM can be calculated by using the viscosity of the gas being measured and accurate results will be obtained. Calculate the PTVM by taking the ratio of the viscosity of nitrogen to the viscosity of the gas under test. For example, to calibrate hydrogen, consider the following: at 0°C, the viscosity of nitrogen is 165.31 microPoise, and the viscosity of hydrogen is 83.21 microPoise. Express these as 165.31/83.21, or 1.987, and enter 1.987 as the PTVM for this cell.
When measuring alternate gases at flows below 20 ccm, or for absolute best accuracy, it may be necessary to perform a dynamic leak test using the gas under test. Contact Sierra for information on this test.
12
Page 13
7
“Standardized or Volumetric”
Choose whether you would like to conduct standardized or volumetric flow readings. The user can select the standardization temperature. The standardization pressure is set at 760mmHg (14.7 psi = 1 atm = 1013.25 mbar). Change the standardization temperature by selecting this option and changing the value using the arrows.
“Number in Average”
Changes the quantity in an average sequence (consecutive readings) from 1-100. Default value is set at 10.
13
Page 14
Chapter 4: Installing and Removing Flow Cells
The CalTrak accepts interchangeable CalTrak SL 800 specific flow cells for different flow ranges. When a cell is installed, the home screen displays the model number and flow range of the installed cell. If no cell is installed, this information will be blank. The base includes a blue friction fit adapter ring that should be removed when using the -75 cell due to its larger diameter. Replace this adapter ring when using the other cells.
Installing Flow Cells
1. Ensure blue adapter ring is installed for using all cells except the -75. Remove the ring if installing the -75 cell.
2. Position the selected flow cell into the base opening, its top label facing you.
3. Carefully align the cell with the guide pins; when the guide pins are engaged, press down.
4. When powered on, the CalTrak SL 800 senses which cell is installed and displays the appropriate units for that
cell.
Removing Flow Cells
Grasp the flow cell firmly, hold the base in place, and lift upwards.
14
Page 15
Chapter 5: Factory Default Settings
Parameters
Factory Settings
Optional Settings
No. of Readings in an Averaging Sequence
10
1-100
Atmospheric Pressure
mm Hg
mBar, kPa, PSI
Temperature
°C
°F
Standardized Temperature Setting
0°C
0.0 – 50.0°C
The SL800 has a number of user-definable features and settings. To return to factory default settings at any time, press Reset followed by Save from the “Main Menu”.
15
Page 16
Chapter 6: Connecting the CalTrak 800 to Gas Flow Source
The accuracy of the CalTrak SL 800 is dependent upon its source being stable. An unstable flow source may produce inconsistent readings. Flow direction is indicated on the cell. To use a pressure flow source, connect to the inlet fitting, or to use a vacuum flow source, connect to the outlet fitting.
The CalTrak SL 800 is designed to be used at ambient pressures only. This is easily accomplished by leaving the outlet of the flow cell open to atmosphere for pressure applications or the inlet open to atmosphere in vacuum scenarios. If tubing is needed to exhaust test gases to a fume hood insure that the exhaust tubing is of sufficient diameter such that the pressure to the CalTrak SL 800 calibrator does not exceed 5 inches of water column (WC) above ambient pressure.
16
Page 17
Chapter 7: Taking Readings
The CalTrak SL 800 default is preset for ten (10) readings in an averaging sequence. This parameter is user-definable from the Operation menu.
1. Press Measure on home screen.
Press the Mode button to cycle through options for single, burst, or continuous readings. Press Start button to begin measurements.
Press the Stop button to stop current flow reading and open valve.
Setting User Preferences
The CalTrak SL 800 offers enhanced electronics options to allow the user to define parameters specific to an application.
These parameters may be entered from the Operation screen or by pressing the blue section of the Measure screen
17
Page 18
Chapter 8: Using Sensor Factors for MFC and MFM Calibration
MFCs (mass flow controllers) and MFMs (mass flow meters) are originally calibrated by the manufacturer with a specific gas. When you use an MFC or MFM with a different gas than the one used in the last calibration, you must apply a flow correction factor to account for different gas properties that affect flow. The MFC or MFM manufacturer can provide the appropriate sensor factor value. The user can also apply this flow correction factor to the CalTrak SL 800 readings.
To enable your CalTrak SL 800 to scale its actual flow measurements to match the adjusted flow from the MFC or MFM, input the sensor factor into the CalTrak SL 800 before calibrating the MFC or MFM.
The Sensor Factor is displayed on the run screen of the CalTrak SL 800 as SF X.XXX. The default setting is
1.000. If you input a sensor factor other than 1, red text will display on the Measure screen showing the sensor factor applied.
18
Page 19
Chapter 9: Error Messages
“No Cell” - This message means that the cell is not present. Hit the home button to reset and install the cell.
“Piston Error” - This message means that the piston did not return to the bottom of the cell properly during operation. Hit the
home button to reset.
If you continue to receive error messages after following these steps. Please contact customer service for assistance.
19
Page 20
Chapter 10: Annual Maintenance and Calibration
RMA Note – Returning Unit for Service
Sierra will not evaluate or service your instrument without an RMA number. Go to
http://www.sierrainstruments.com/rma to complete an RMA.
Your CalTrak SL 800 is engineered to provide years of reliable service, with appropriate care and maintenance. Sierra recommends an annual calibration, to help ensure the best possible flow measurements and to provide an audit trail for those applications subject to regulatory requirements. If you should encounter any problems with your CalTrak SL 800, immediately contact Customer Service and provide a detailed description of your situation, including CalTrak SL 800 model and serial number, information about the flow source and the current calibration setup, environmental conditions during the test, the flow point or points that you’re checking and an explanation of the issue you’re experiencing.
Recertification
Your CalTrak SL 800 primary piston prover is a precision measuring standard comprised of moving parts that are ma ­chined to extremely close tolerances. Additionally, various environmental factors, product wear, drift of the temperature sensors and pressure transducers, or inadvertent damage may adversely affect your measurement accuracy or general performance.
For these reasons, Sierra highly recommends having your CalTrak SL 800 annually verified in order to ensure its measurement integrity. For those applications subject to regulatory or ISO requirements, verification by oan accredited laboratory provides you with a defensible audit trail of the highest quality.
Returning Equipment to Factory
Please contact Sierra for a service quote before sending your CalTrak SL 800 to our factory.
If you are sending in your CalTrak for repair or evaluation (rather than elective re-certification), contact Sierra for technical support or troubleshooting assistance prior to shipping the unit. Provide us a detailed description of your issues. If we are unable to resolve the situation by phone or email, we will issue you an RMA (return merchandise authorization) number. Follow the instructions for returning your instrument for service as noted below.
Shipping
When shipping your CalTrak, be sure to follow some simple guidelines to avoid costly damage to your property.
Pack your instrument carefully. Use adequate packing material. Whenever possible, use the original
packing and foam or bubble wrap that came with your CalTrak (packing peanuts NOT recommended). Or use a Sierra CalTrak Pelican carrying case, which provides a hard case shell for protection of your valuable equipment. If you do not already have a Pelican case, visit us at www.sierrainstruments.com for more information on obtaining one.
Include a copy of the RMA form (complete with Sierra supplied RMA number) with the unit(s).
Use a major freight carrier (e.g., FedEx, UPS) that supplies tracking numbers. Insure your CalTrak. Sierra is not responsible for damage occurred during transit. Understand our mutual shipping obligations. Sierra is responsible for shipping cost only if the issue is
product related and the CalTrak is still under warranty.
20
Page 21
Ship the unit(s) to the following address:
Attention: Factory Service Center
Sierra Instruments, Inc.
5 Harris Court, Building L Monterey, CA 93940 USA
RE: RMA# (your number)
21
Page 22
Chapter 11: Storage
Protecting Your CalTrak When Not In Use
If you need to store your CalTrak SL800 for an extended period, please follow these guidelines: Always store it in a clean, dry place.
22
Page 23
Chapter 12: Product Specifications
Model Number
SL 800-3
SL 800-10
SL 800-24
SL 800-44
SL 800-75
Flow Range
0.550 mL/min
5500 mL/min
50—5,000 mL/min
500—50,000 mL/min
1-100 L/min
Accuracy, Standardized
±0.25% 0.5-50 sccm ±
0.002 sccm
±0.15%
±0.15%
±0.15%
±0.15%
Time Per Reading
160 seconds
3135 seconds
390 seconds
135 seconds
1-50 seconds
Weight
80 oz/2300 grams
85 oz / 2413 g
86 oz / 2439 g
88 oz / 2507 g
160 oz / 4535 g
Inlet Fitting
1/8 in
1/4 in
1/4 in
1/4 in
1/2 in
Outlet Fitting
1/8 in
1/4 in
1/4 in
1/2 in
1/2 in
Interchangeable Flow Cell Models:
Specifications based on averaged readings. Accuracy is stated as a percent of reading (including standardization, if applicable).
Basics:
Dimensions: Cell Model 3: 9.0 x 4.0 in / 229 x 102 mm Dimensions: Cell Models 10, 24, 44: 13.3 x 4.0 in / 337 x 102 mm Dimensions: Cell Model 75: 14.8 x 5 in / 375 x 102 mm Dimensions: Base: (H x W x D) 3.9 x 8.5 x 12.6 in / 99 x 216 x 320 mm
Weight: Base 5 lbs / 2268 g Gas Compatibility: Non-corrosive, non-condensing, non-combustible gasses. Less than 70% humidity.
Usage:
Flow Mode: Suction and Pressure Configuration: Base with modular, interchangeable flow cells (five) Temperature & Pressure Sensors: Yes, in the flow stream of each interchangeable flow cell Reading Modes: Single, Auto or User-Specified Burst AC Adapter/Plug: 12V DC, > 3.0A, 2.5 mm, center positive, North American standard, others available Inlet and Outlet Fittings: SwagelokTM Operating Temperature: 15–30° C Storage Temperature: 0– 70° C Operating Humidity: 0–70%, non-condensing Display: Resistive touchscreen Operating Pressure (Absolute): 15 PSI Data Port: RS-232 Serial and USB-B
Sierra highly recommends annual Recertification (product maintenance, pre and post dynamic flow comparisons and full dimensional calibration). This is elective and is not included as a warranty item.
All specifications are subject to change. Please contact Sierra
CalTrak Pro Software:
Visit Sierra’s website to download your copy of CalTrak Pro software CalTrak Pro captures flow data from your CalTrak SL 800 directly to a pre-configured table. The data can be exported to selectable Windows office environment.
To run CalTrak Pro, you must have Windows® XP or 7, Microsoft Excel® 2003 and up, and either a USB port or a RS-232 serial port.
23
Page 24
Appendix A: Warranty Policy
LIMITED WARRANTY POLICY- REGISTER ONLINE
All Sierra products are warranted to be free from defects in material and workmanship and will be repaired or replaced at no charge to Buyer, provided return or rejection of product is made within a reasonable period but no longer than one (1) year for calibration and non-calibration defects, from date of delivery. To assure warranty