Sierra QuadraTherm 640i Series, QuadraTherm 780i Series Instruction Manual

Page 1
QuadraTherm® 640i/780i Series
Insertion and In-Line Mass Flow Meters
Preliminary Instruction Manual
Part Number: IM-640i/780i Rev.v1
November 2012
1
Page 2
GLOBAL SUPPORT LOCATIONS: WE ARE HERE TO HELP!
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940 Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
www.sierrainstruments.com
EUROPE HEADQUARTERS
Bijlmansweid 2 1934RE Egmond aan den Hoef The Netherlands Phone +31 72 5071400 Fax +31 72 5071401
ASIA HEADQUARTERS
Second Floor Building 5, Senpu Industrial Park 25 Hangdu Road Hangtou Town Pu Dong New District, Shanghai, P.R. China Postal Code 201316 Phone: + 8621 5879 8521 Fax: +8621 5879 8586
IMPORTANT CUSTOMER NOTICE- OXYGEN SERVICE
Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instruments standard mass flow meters for oxygen gas. You are responsible for determining if this mass flow meter is appropriate for your oxygen application. You are responsible for cleaning the mass flow meter to the degree required for your oxygen flow application.
© COPYRIGHT SIERRA INSTRUMENTS 2012
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change without notice.
TRADEMARKS
QuadraTherm™, iTherm™ and iAnywhere™ are trademarks of Sierra Instruments, Inc. Other product and company names listed in this manual are trademarks or trade names of their respective manufacturers.
2
Page 3
Table of Contents
Chapter 1: Introduction & Product Description
Product Description
Technical Assistance
............................................................................................................................ 5
...................................................................................................................... 7
.................................................................................... 5
Principle of Operation ........................................................................................................................ 8
iTherm™ Electronics Features
Enclosure Options
........................................................................................................................... 10
............................................................................................................ 9
Chapter 2: Installation & Wiring ........................................................................................................ 11
Installation Overview ....................................................................................................................... 11
Installing the Flow Meter .................................................................................................................. 13
Cold Tap Installation Hot Tap Installation
........................................................................................................................ 13
......................................................................................................................... 14
Wiring Connections ......................................................................................................................... 17
Input Power Wiring
Alarm Output Wiring
........................................................................................................................ 18
................................................................................................................... 21
Remote Sensor Wiring .................................................................................................................... 24
Chapter 3: Operation & Programming
............................................................................................ 25
Start-up Routine ......................................................................................................................... 26
Level 1: Main Menu ..................................................................................................................... 26
Level 2: Sub-Menu (Password Protected) ..................................................................................... 28
Calibration (Sub-Menu) ............................................................................................................... 29
Calibration (Sub-Menu): Gas Type –“iTherm Dial-A Gas” ............................................................... 30
Calibration (Sub-Menu): Flow Units .............................................................................................. 32
Calibration (Sub-Menu): Temperature Unit .................................................................................... 33
Calibration (Sub-Menu): MenuTune .............................................................................................. 33
Calibration (Sub-Menu): Low Flow Cut Off .................................................................................... 34
Calibration (Sub-Menu): Full Scale Flow ....................................................................................... 35
Process Pressure (Sub-Menu) ..................................................................................................... 35
Process Pressure (Sub-Menu): Pressure Units .............................................................................. 36
Process Pressure Sub-Menu: Process Pressure ............................................................................ 36
Totalizer (Sub-Menu) .................................................................................................................. 37
Totalizer Sub-Menu: Reset Pulse Count ....................................................................................... 37
Totalizer (Sub-Menu): Units Per Pulse .......................................................................................... 39
Totalizer (Sub-Menu): Totalizer On/Off ......................................................................................... 39
Output Adjust (Sub-Menu): Flow, Pressure and Temperature ......................................................... 42
Reference Conditions (Sub-Menu): Standard and Normal .............................................................. 45
Change Password, Change Language .......................................................................................... 45
Restore Factory .......................................................................................................................... 45
Box Car ..................................................................................................................................... 45
Chapter 4: Troubleshooting & Repair
Troubleshooting the Flow Meter Returning Equipment to Factory
................................................................................................ 49
.................................................................................................... 49
.................................................................................................... 51
Appendix A: Product Specifications ................................................................................................. 52
Appendix B: Smart Interface Installation ........................................................................................... 60
3
Page 4
Warning!
the flow meter nameplate for specific flow meter approvals before any hazardous location installation.
Warning!
permit. The manufacturer of the hot tap equipment and/or the contractor performing the hot tap is responsible for providing proof of such a permit
Warning!
Warning!
when wiring this unit to a power source and to peripheral devices. Failure to do so could result in injury or death. All AC power connections must be in accor dance with published CE directives.
Warning!
over-heating of the sensors and/or damage to the electronics.
Warning!
Warning!
Caution!
to any master control system. Adjustments to the electronics will cause direct changes to flow control settings.
Warnings and Cautions
Agency approval for hazardous location installations varies between flow meter models. Consult
Hot tapping must be performed by a trained professional. U.S. regulations often require a hot tap
All wiring procedures must be performed with the power off.
To avoid potential electric shock, follow National Electric Code safety practices or your local code
Do not power the flow meter with the sensor remote (if applicable)wires disconnected. This could cause
Before attempting any flow meter repair, verify that the line is de-pressurized.
Always remove main power before disassembling any part of the mass flow meter.
Before making adjustments to the device, verify the flow meter is not actively monitoring or reporting
Caution!
higher pressure rating as the main pipeline.
Caution!
accuracy of the flow meter. You cannot add or subtract wire length without returning the meter to the factory for re-calibration.
Caution!
at full gas flow before installing the meter.
Caution!
Caution!
follow these precautions to minimize the risk of damage:
All flow meter connections, isolation valves and fittings for hot tapping must have the same or
Changing the length of cables or interchanging sensors or sensor wiring will affect the
When using toxic or corrosive gases, purge the line with inert gas for a minimum of four hours
The AC wire insulation temperature rating must meet or exceed 80°C (176°F).
Printed circuit boards are sensitive to electrostatic discharge. To avoid damaging the board,
before handling the assembly, discharge your body by touching a
grounded, metal object
handle all cards by their edges unless otherwise required when possible, use grounded electrostatic discharge wrist straps when handling
sensitive components
4
Page 5
Chapter 1: Introduction & Product Description
From Sierra’s beginning over forty years ago, Founder Dr. John G. Olin was driven by the vision of supplying industrial customers with the world’s most accurate thermal mass flow
meter. And, he knew it was a “sensor” game.
The development of an industrialized metal-sheathed sensor in the early 80s was Sierra’s first big step, but Dr. Olin is a driven innovator, and this was only the beginning for
someone who saw “Thermal Mass Flow” as his life’s work. Many successful innovations
followed, but in 1999 Sierra experienced a major breakthrough with the introduction of their patented no-drift DrySense™ thermal mass velocity sensor. Sierra engineers now recognized they were on the cusp of realizing Dr. Olin’s vision.
Realizing the Vision: Thermal technology, by its very nature, uses the physics of heat transfer and and conservation of energy in a closed system to measure mass flow rate. This means that for a thermal mass flow meter to achieve the greatest accuracy, it must solve the First Law of Thermodynamics (Heat Energy In = Heat Energy Out) for each data point.
As you can imagine, solving the First Law in a flow instrument was no easy task. By Dr.
Olin’s own accounting, decades of “hard-nosed dedication to excellence” by himself and
Sierra’s engineering team, years of testing, and his stack of yellow note pads over five feet
high, jammed with his handwritten equations and designs, finally yielded the secret in the form of two revolutionary technologies—QuadraTherm™ and iTherm™, now both patented worldwide.
The QuadraTherm Sensor: Traditional thermal sensors have two sensors-–one temperature sensor and one velocity sensor, each in a separate probe. QuadraTherm (the term “Quad” meaning “four”) introduces four sensors—three precision platinum temperature sensors and one patented DrySense mass velocity sensor. Performance improvements never before possible are gained as the QuadraTherm Technology isolates forced convection (the critical variable for measuring gas mass flow rate) by calculating and then eliminating unwanted heat-transfer components, like sensor stem conduction, one of the major causes of false flow readings.
iTherm, the Brains Behind it: iTherm is the true “Brain” of the instrument and a
revolutionary, living, learning algorithm set made possible by today’s hyper-fast microprocessors and QuadraTherm sensor inputs. iTherm manages changes in gas flow, temperature and pressure, as well as outside temperature, via a comprehensive heat­transfer model. The result of iTherm is a proprietary, fundamentally different gas mass flow rate calculation using all pertinent variables for the most precise, stable and accurate mass flow measurement possible.
Website & Downloads
QuadraTherm microsite:
www.sierrainstruments.com/quadratherm
Download this manual:
www.sierrainstruments.com/quadrathermIM
5
Page 6
Features and Benefits
Accuracy (air) +/- 0.5% of reading from 50% to 100% of full scale +/-0.5% of reading plus 0.5% of full scale from 0% to 50% of full scale
Patented vastly improved thermal sensor element for high accuracy
Patented DrySense™ Technology for long term drift stability; lifetime warranty
Instrument-optimized proprietary iTherm Gas Library for application independence
Sensor design improvements to eliminate downdraft effects from probe
Dial-A-Gas®: change gas and engineering units in the field, on the fly with the push of a button
Dial-A-Pipe™: Use insertion version on different pipe sizes, set up on the fly with the push of a button
Multivariable: simultaneous measurement of mass flow rate, process temperature, and pressure
CE approved
FM approval pending
Digital communications suite pending
Using This Manual
This manual provides information needed to install and operate QuadraTherm 640i and 780i thermal mass flow meters. The four chapters and two appendices of this manual cover the following areas:
     
Chapter 1: Introduction & Product Description Chapter 2 Installation & Wiring Chapter 3: Operation & Programming Chapter 4: Troubleshooting & Repair Appendix A: Product Specifications Appendix B: Smart Interface Program (SIP) software
6
Page 7
Warning!
Caution!
This statement appears with information that is important to protect people and equipment from damage. Pay very close attention to all warnings that apply to your application.
This statement appears with information that is important for protecting your equipment and performance. Read and follow all cautions that apply to your application.
Note and Safety Information
We use caution and warning statements throughout this book to draw your attention to important information.
Receipt of System Components
When receiving a Sierra mass flow meter, carefully check the outside carton for damage incurred in shipment. If the carton is damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered components are present. Make sure any spare parts or accessories are not discarded with the packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service
packing
.
Technical Assistance
If you encounter a problem with your flow meter, review the configuration information for each step of the installation, operation, and setup procedures. Verify that your settings and adjustments are consistent with factory recommendations. Refer to Chapter 4: Troubleshooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 4, contact Sierra Instruments by fax or by E-mail (see inside front cover). For urgent phone support you may call (800) 866-0200 or (831) 373­0200 between 8:00 a.m. and 5:00 p.m. PST. In Europe, contact Sierra Instruments Europe at +31 20 6145810. In the Asia-Pacific region, contact Sierra Instruments Asia at + Support, make sure to include this information:
 
 
86-21-58798521.
The flow range, serial number, and Sierra order number (all marked on the meter nameplate) The software version (visible at start up)
When contacting Technical
The problem you are encountering and any corrective action taken Application information (gas, pressure, temperature and piping configuration)
7
Page 8
Principle of Operation
Watch a principal of operation video at
Sierra’s unique QuadraTherm™ sensor probe is responsible for the unsurpassed accuracy, ruggedness, and reliability of Sierra industrial flow meters. The immersible sensor consists of four sensing elements–a velocity sensor, a temperature sensor, and two stem conduction sensors that measure heat lost to the outside.
When power is applied to the flow meter, the transducer electronics heats the velocity sensor to a constant temperature differential above the gas temperature and measures the cooling effect of the gas flow. The electrical power required to maintain a constant temperature differential is directly proportional to the gas mass flow rate. The two stem conduction sensors ensure that any heat lost is due to this cooling effect, and not due to conduction to the outside.
The velocity sensor is a reference-grade platinum resistance temperature detector (RTD). The platinum RTD wire is wound on a rugged ceramic mandrel for strength and stability. The temperature sensor is clad in 316 stainless steel. The velocity sensor is clad in a Pt/Ir Alloy.
www.sierrainstruments.com/thermalprincipal
Figure 1-1:640i & 780i Series Sensor Assembly
8
Page 9
iTherm™ Electronics Features
iTherm™ Dial-A-Gas Technology
Select gas to be measured
iTherm™ Dial-A-Pipe Technology
Set-up pipe dimensions
Units
Select units for mass flow, temperature, and pressure
User Full Scale Flow Rate
Field-configure from 50% to 100% of the factory full scale setting
Alarms
Program high & low or window alarm limits independently for mass flow, temperature, pressure and totalizer (select one)
MeterTune™ (Span Adjust)
Change the calibration correction factor to compensate for flow profile disturbances or specific application conditions. The MeterTune™ (span adjust) is a multiplication factor applied to the flow signal
Dual Output Signals
Two 4-20mA separate linear output signals proportional to flow and temperature. Optional 4-20mA proportional to pressure if pressure option selected.
Totalizer
Set up totalizer pulse output
Set Standard Conditions
Chose between, normal, standard and other
Set Password
Set user password
Set Low Flow Cut-off
Force flow to zero at a specified flow rate
View Tag Number
View assigned tag number
Set Language
Set language for display
9
Page 10
View Communications Protocol Parameters
View communications settings: 38,400 Baud rate, no parity, 8 data bits, and 1 stop Bit
ValidCal™ Diagnostics
Review minimum, maximum flow, temperature and pressure and other diagnostics
Enclosure Options
Flow meter electronics are available mounted directly to the flow body, or remotely mounted up to 200 feet (60 meters) away. The electronics housing may be used indoors or outdoors.
Display options include a programmable LCD displaying mass flow, temperature, pressure, totalized flow, Dial-A-Gas, Dial-A-Pipe, MeterTune, and user full scale along with alarm, reference conditions, serial numbers and tag number. Local operation and re­configuration is accomplished using the push buttons on the device. Electronics include non-volatile memory that stores all configuration information. The memory allows the flow meter to function immediately upon power up, or after an interruption in power.
10
Page 11
Warning!
Agency approval for hazard­ous location installations var­ies between flow meter mod­els. Consult the flow meter nameplate for specific flow meter approvals before any
hazardous location installation.
Chapter 2: Installation & Wiring
Installation Overview
The 640i and 780i are easy to set up in the field. The flow meter should be inserted to the centerline of the pipe.
When selecting an installation site, make sure that:
1. Line pressure and temperature will not exceed the flow meter rating. If ambient temperature exceeds 50°C, remove flow meter to a cooler area.
2. The location meets the required minimum number of pipe diameters upstream and downstream of the sensor head (see Figure 2-1 on the next page)
3. Safe and convenient access with adequate clearance.
.
Also, verify the meter is located where the gas is clean and dry.
4. Verify that the cable entry into the instrument meets the FM standard (if required).
5. For remote installations, verify the supplied cable length is sufficient to connect the flow meter sensor to the remote electronics. (Do not extend or shorten the supplied cable between the probe and the electronics.)
6. Also, before installation check your flow system for anomalies such as:
Leaks
Valves or restrictions in the flow path that could create disturbances in the flow profile that might cause unexpected flow rate indications
Heaters that might cause rapid excursions in the measured temperature
11
Page 12
Unobstructed Flow Requirements
640i and 780i Straight Run Requirements
Piping Condition Example
A-Upstream 640i Insertion
(1)
A-Upstream 780i Inline with Flow Conditioning
(1)
B-Downstream
(2)
1
15D
1D
0D
2
20D
3D
0D 3 40D
3D
0D
4
15D
3D
0D
5
30D
3D
0D 6 40D
5D
0D
Notes: (1) Number of diameters (D) of straight pipe required between upstream disturbance and the flow meter (2) Number of diameters (D) of straight pipe required downstream of the flow meter
Select an installation site that will minimize possible distortion in the flow profile. Valves, elbows, control valves and other piping components may cause flow disturbances. Check your specific piping condition against the examples shown below. In order to achieve accurate and repeatable performance install the flow meter using the recommended number of straight run pipe diameters upstream and downstream of the sensor.
Figure 2-1: Recommended Pipe Length Requirements for Installation
12
Page 13
Caution!
When using toxic or cor­rosive gases, purge the line with inert gas for a minimum of four hours
at full gas flow before
installing the flow meter.
Installing the Flow Meter
NOTE: Probe may enter the pipe from any direction as long as the tips are in the center.
When positioning the flow meter, refer to the flow direction indicator at­tached to the probe. For proper operation, install the meter with the flow direction indicator pointing downstream in the direction of flow. Installing the meter opposite this direction will result in inaccurate flow measurement
Cold Tap Installation
1. Confirm that the installation site meets the minimum upstream and downstream pipe diameter requirements shown in Figure 2-1.
2. Turn off the flow of process gas. Verify that the line is not pres-
surized.
3. Use a cutting torch or sharp cutting tool to tap into the pipe. The pipe opening must be at least 0.78 inches in diameter. (Do not attempt to insert the sensor probe through a smaller hole.)
4. Remove all burrs from the tap. Rough edges may cause flow profile distortions that could affect flow meter accuracy. Also, obstructions could damage the sensor assembly when inserting into the pipe.
5. Mount the compression or flange fitting on the pipe. Make sure this connection is within ±5° perpendicular to the pipe centerline as shown at left.
6. When installed, cap the fitting. Run a static pressure check on the connection. If pressure loss or leaks are detected, repair the connec­tion and re-test.
7. Insert the sensor probe through the compression or flange fitting into the pipe. The correct insertion depth places the centerline of the sensor access hole in the probe at the pipe’s centerline.
8. Align the sensor head using the flow direction indicator. Adjust the indicator parallel to the pipe pointing downstream in the direction of flow.
9. Tighten the fitting to lock the flow meter in position. (When a com­pression fitting is tightened, the position is permanent unless using Teflon ferrules.)
13
Page 14
Caution!
All flow meter connec-
tions, isolation valves
and fittings for hot tap-
ping must have the same
or higher pressure rating
as the main pipeline.
Warning!
Hot tapping must be
performed by a trained
professional. U.S. regula-
tions often require a hot
tap permit. The manufac-
turer of the hot tap
equipment and/or the
contractor performing the
hot tap is responsible for
providing proof of such a
Hot Tap Installation
When positioning the flow meter, refer to the flow direction indicator at­tached to the probe. For proper operation, install the meter with the flow direction indicator pointing downstream in the direction of flow. Install­ing the meter opposite this direction will result in inaccurate flow meas­urement. Low pressure hot taps cannot exceed 150 psia (10 bara) maxi­mum. High pressure hot taps cannot exceed 1000 psia (70 bara) maxi­mum. Make sure the pipe pressure does not exceed these limits before beginning this procedure.
1. Confirm that the installation site meets the minimum upstream and downstream pipe diameter requirements shown in Figure 2-1.
2. Calculate the flow meter insertion depth as shown in Figure 2-2 for a low pressure tap or Figure 2-3 for a high pressure tap.
3. Weld the process connection on the pipe. Make sure the process connection is within ± 5° perpendicular to the pipe centerline (see previous page). The pipe opening must be at least 0.88 inches (22 mm) in diameter.
4. Bolt an isolation valve on the process connection. The valve’s full
5. Hot tap the pipe.
6. Close the isolation valve. Run a static pressure. If pressure loss or
7. Insert the sensor probe through the isolation valve into the pipe with
8. Tighten the fittings to lock the flow meter in position.
open bore must be at least 0.88 inches (22 mm) in diameter.
leaks are detected, repair the connection and re-test.
the flow direction indicator parallel to the pipe pointing downstream in the direction of flow. The correct insertion depth places the sensor at the pipe’s centerline. Do not force into the pipe.
14
Page 15
Calculating Insertion Depth for a Low Pressure Tap
Variables
L = Nominal probe length D = Duct O.D.
C = Duct I.D.
T = Height of “Threadolet” or customer provided “Weldolet”
Formula
L ≥ 12 + D/2 + T L must be equal or greater than 12 inches plus the height of the
“Threadolet” plus half the duct O.D.
(All dimensions in inches.)
Calculating Insertion Depth for a High Pressure Tap
Figure 2-2: Low Pressure Tap Insertion Depth
Variables
S = Distance from face of mounting flange to outside of duct D = Duct O.D. P = Minimum probe length T = Minimum probe travel R = Allowable probe travel IN = Inserted position (marker location) RE = Retracted position (marker location)
15
Page 16
Formulas
1) P = D/2 + S + 6.75
2) T = D/2 + 0.54
3) R = 28.2 – [actual probe length – S – (D/2)]
4) IN = (actual probe length + 2) – (5.5 + S + D/2)
5) RE = IN + T
(All dimensions in inches.)
(minimum probe length–use next longer whole number length probe)
(must be greater than or equal to T)
Figure 2-3. High Pressure Tap Insertion Depth
16
Page 17
Figure 2-4. Wiring Access
To avoid potential electric
shock, follow National Elec- tric Code safety practices or your local code when wiring
this unit to a power source
and to peripheral devices.
Failure to do so could result
in injury or death. All AC
power connections must be
in accordance with pub-
lished CE directives.
Warning!
Wiring Connections
Use the terminal blocks located inside the cap of the flow meter enclosure for all wiring connections. Make sure to observe all CE compliance requirements for AC wiring connections given on the next page.
17
Page 18
Warning!
All wiring procedures must
be performed with the
power Off.
Caution!
The AC wire insulation tem-
perature rating must meet or
exceed 80 °C (176°F).
Figure 2-5: AC Input Power Connections
Input Power Wiring
AC Power Wiring
The AC power wire size must be 26 to 16 AWG with the wire stripped 1/4 inch (6 mm). Connect 100 to 240 VAC (0.4 Amps RMS at 230 VAC) to the Neutral and Line terminals on the terminal block. Connect the ground wire to the safety ground lug. Torque all connections to 4.43 to 5.31 in-lbs (0.5 to 0.6 Nm).
The Hazardous-Area enclosure has two separate conduit entries to maintain separation between AC input power and output signal wiring. To eliminate the possibility of noise interference, use a separate cable entry for the AC power and signal lines.
18
Page 19
Warning!
All wiring procedures must
be performed with the
power Off.
Figure 2-6: DC Input Power Connections
DC Power Wiring
The DC power wire size must be 26 to 16 AWG with the wire stripped 1/4 inch (6 mm). Connect 24 VDC +/- 10% (1.04A load, maximum) to the terminals marked on the terminal block. Connect the Earth ground wire to the safety ground log. Torque all connections to 4.43 to 5.31 in-lbs (0.5 to
0.6 Nm). If conduit seals are used, they must be installed within 18 inches of the
enclosure.
19
Page 20
Figure 2-7: 4-20mA Output Connections
Warning!
Do not externally power
the 4-20mA output loop.
It is a self-powered
loop.
Figure: 2-8 Ferrite Installation (Ferrite not required for conduit)
Output Signal Wiring
Output signal cable should be completely screened with a 100% shield. You must use metal cable glands that provide cable screen clamping. The cable screen should be connected to the gland and shielded at both ends over 360 degrees. The shield should be terminated to an earth ground. For all installations not using metal conduit two ferrite beads should be added, one on each end of the I/O cable. This is to maintain CE related EMI/RFI protection. Good quality (Highest impedance at 100MHz) Broadband ferrites should be used, a solid cylindrical ferrite (recommended) usually has better performance than a clamp on ferrite. The ferrites should fit as tight as possible to the OD of your cable.
All QuadraTherm 640i/780i Series flow meters are equipped with calibrated 4-20 mA output signals for both T and P, with an optional 4-20 MA output for pressure.
4-20 mA Output Wiring
The 4-20 mA current loop output is non-isolated. Max load 500 ohms
20
Page 21
Alarm Output Wiring
Figure 2-9: AC or DC Power Supply
One alarm output contact is included on the flow meter terminal block. The alarm output is driven by an optical relay that is normally-open single-pole.
The relay is isolated and requires a separate power supply (isolated) the voltage of the alarm output is the same as the voltage supplied to the circuit.
To use an external power supply for an isolated alarm output, connect as shown in Figure 2-8. You may set low, high or window alarms for temperature, pressure, totalizer or mass flow
.
21
Page 22
Figure 2-11:Pulse Output
Figure 2-10: RS-232 Communication
RS-232 Wiring
RS-232 provides serial communication. Wire per figure 2-10.
Pulse Output
QuadraTherm provides an adjustable pulse output with a maximum of 1 Hz. Wire per figure 2-11 below.
22
Page 23
USB Output
Figure 2-12: USB Output (pending)
If supplied, plug your USB adapter into J1, per figure 2-12 (USB pending).
23
Page 24
Figure 2-13: Remote Junction Box
Remote Sensor Wiring
When connecting the sensor probe to a remotely mounted flow meter en­closure, use only factory supplied cables. When connecting more than one meter, do not intermix the sensor probes and electronics. The elec­tronics, sensor probes and interconnecting cables supplied by Sierra In­struments are calibrated as a complete precision mass flow circuit.
To make wiring connections from a sensor probe junction box to a remotely mounted enclosure, see Figure 2-13.
24
Page 25
Chapter 3: Operation & Programming
General Navigation
In general terms, the menu system consists of a main menu, the set-up menu, a sub­menu to program each item in the set-up menu, and a series of data entry or pull­down screens with to enter set-up data for each parameter.
Menus may be adjusted using the six buttons on the front of the unit: up , down , left , right , enter and escape/cancel , or by using the Smart Interface Program (SIP) software provided free with the instrument.
Pressing the left button will move the menu selection to the left, or the data entry field to the left if updating a menu value.
Pressing the right button will move the menu selection to the right, or the data entry field to the right if updating a menu value.
Pressing the up button will move the menu selection up, or increment the data entry field if you’re updating a menu value.
For example, if you’re updating a value, and that value is currently set to “0”, pressing
the up key will increase the value to the next logical value, in our example it would be “1”, then “2”, “3” and so on to “9” then back to “0”
Pressing the down button will move the menu selection down, or decrement the data
entry field if you’re updating a menu value.
For example, if you’re updating a value, and that value is currently set to “9”, pressing
the down key will decrease the value to the next logical value, in our example it would be “8”, then “7”, “6” and so on to “0” then back to “9”
Pressing the enter key accepts the current value.
Pressing the escape/cancel key returns to the last previous menu, and if you’re currently editing a value, will cancel any changes you’ve made.
Note: The screen data “flashes” to show that you are actively editing. Pressing the enter key permanently writes the data to the meters memory.
25
Page 26
640i/780i V1.0.X
Serial 1234XXXX
Full Scale
100.00 SCFM
Dial-A-Gas Carbon Dioxide
Tag 1234XXXX
Start-up Routine
Then the meter is first powered up, it will cycle through set-up data. See an example below of the type of screens you will see.
1. Product Name and Firmware Version. Left is main PCA / right is display PCA
2. Serial Number
3. Full Scale
4. Gas
5. Tag Number
NOTE: All of these values are also displayed on the provided SIP (Smart Interface Program) software.
Level 1: Main Menu
Table 3-1: Main Menu Level 1
26
Page 27
0.11 SCFM
97.66 F
Pressure
0.00/psia
Total Units SCF
0.00
Active Alarm Flow – L (On)
Full Scale Flow
100.00 SCFM
Auto Scroll
Off
Auto Scroll
On
Once set-up data is displayed, the meter will display the flow variables. The instrument will Auto Scroll through each flow variable. You can turn OFF the Auto Scroll feature by hitting the down arrow.
Turn Auto Scroll back ON again by hitting the up arrow. Auto Scroll will default back to on if power is cycled.
The following flow variable screens will Auto Scroll. The data shown below is for example only:
1. Flow and Temperature
2. Pressure (N/A if option was not ordered)
3. Totalizer (only displayed if turned ON)
You can now cycle through more main menu screens. By entering >, you will see the additional screens. You can also go back using < at any time or exit to the main screen by pressing (X).
4. Alarm
5. Full Scale Flow
27
Page 28
Dial-A-Gas Carbon Dioxide
Ref: Standard
14.695949 / 70.0
Serial 1234XXXX
Tag 1234XXXX
Password
0000
Dial-A-Pipe ID
6. iTherm Dial-A-Gas (Gas Type)
7. iTherm Dial-A-Pipe
8. Reference Conditions
9. Serial
10. Tag
And finally complete the cycle back to the flow and temperature screen.
Level 2: Sub-Menu (Password Protected)
At this point, you can access various other menus .To enter the next level press the enter key. You will be requested to enter a password. Default is 0000 and can be reset in a later screen or using SIP:
Use the up and down buttons to cycle through numbers and left and right buttons to move to the next digit. Once the correct password has been entered, hit enterand you will enter the Level 2 Sub-Menu shown in Table 3-2.
28
Page 29
Calibration
(Sub-Menu)
Table 3-2: Level 2 Sub-Menu
Calibration (Sub-Menu)
Navigate to the Calibration Sub Menu, use Table 3-3 below for guidance if necessary.
The calibration menu allows you to set up the meter to match the application.
To start, move down to the next level by pressing the enter key (). From this level, you can set up Dial-A-Gas and Dial-A-Pipe, as well as flow and temperature units (pressure is optional and has its own set-up menu to be covered later).
In addition you can adjust the MeterTune (span adjust), full scale flow, low flow cut-off and view the date the meter was last calibrated. Note that you can also use the Quick Keys or the function selector of the Smart Interface Program (SIP) provided free with the meter.
Table 3-3: Calibration Sub-Menu
29
Page 30
Dial-A-Gas
Carbon Dioxide
Calibration (Sub-Menu): Gas Type –“iTherm Dial-A Gas”
Use this menu to change and select the gas to be measured. This is called our Dial-A-Gas feature and is a powerful and unique feature of the 640i/780i.
The display will show gas type and list the currently selected gas.
Navigate to the “Dial-A-Gas” screen. Use Table 3-3: Calibration Sub-Menu for guidance if necessary.
Once on the “Dial-A-Gas” screen, press enter key  once again until the current selected gas begins to flash. It may take a few seconds to begin to flash.
Use the up and down keys to scroll through the available iTherm Dial-A-Gas menu. The meter can be pre-configured with four gases from Sierra’s iTherm Gas Library. There are eighteen gases and gas mixes available from the Sierra iTherm Gas Library. Sierra will continually update the library as more gases and mixtures become available. New or updated gases can be downloaded from the library using the SIP program.
Default gases: Air Carbon Dioxide Methane Nitrogen
Once you have selected the desired gas, press enter key to make the change permanent. The selection will stop flashing.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Dial-A-Pipe ID. We discuss this next.
Calibration (Sub-Menu): Dial-A-Pipe ID, Dial-A-Pipe ID, Dial-A-Type
This instrument can be moved to different pipe sizes and types as needed and perform precision flow measurement. This capability is called Dial-A-Pipe and is a powerful and unique feature of the QuadraTherm 640i immersible thermal mass flow meter.
The internal pipe diameter (ID) and pipe type are very important factors when considering flow profile. The 640i uses internal diameter (ID) and the pipe type to mathematically model the flow profile and enable Dial-A-Pipe.
Adjustment can also be done using the SIP (Smart Interface Program) Software. It is recommended to use the SIP to perform Dial-A-Pipe since the process is more automated.
The Dial-A-Pipe process requires 2 steps:
1) Setting Pipe Inner Diameter (ID)
2) Setting Pipe Type
30
Page 31
Dial-A-Pipe ID
6.065000 / In
Dial-A-Pipe ss-smooth
Step 1(Dial-A-Pipe): Navigate to the “Dial-A-Pipe ID” screen. Use Table 3-3: Calibration Sub-Menu for guidance if necessary.
The display will show pipe inner diameter in the units selected. In the example below the inside diameter of the pipe is 6.065000 inches.
Press enter to cause the first digit (the 6 in 6.065000 in our example) to begin flashing. Use the up and down arrows to select the value 0 to 9. Once done, use the right key to move on to the selected significant digit. Once the I.D. value is selected, use the up and down keys to select the desired units.
IMPORTANT: It is vital to use the correct I.D. and units of the pipe where the instrument is located. If the schedule of the pipe is known, the I.D. can be looked up from pipe tables. If the pipe outer diameter and wall thickness is known, the I.D. can be calculated. If you do not have a round pipe, you must enter the equivalent round I.D. There are various methods of calculating this. Here’s an easy calculation to give you the hydraulic diameter (HD):
Hydraulic Diameter = HD
HD = 4A A= Cross sectional area of duck P P = Wetted perimeter of duct
I.D. Units: in: inches m: meters mm: millimeters ft: feet
Step 2 (Dial-A-Pipe Type): Pipe roughness does matter. The friction losses of the gas
moving through the pipe are a major factor in the instruments calculation of the flow profile. Navigate to the “Pipe Type” screen. Use Table 3-3: Calibration Sub-Menu for guidance if necessary.
Use this menu screen to complete entering Dial-A-Pipe data. The display will show pipe roughness and the “type” of pipe. See example below.
Press enter and the type of pipe will begin flashing. Go to the next level down with the button and select the proper pipe type.
Choices Available: ss-smooth ss-normal ss-rough
31
Page 32
cs-smooth
Flow Units
SCFM
cs-normal cs-rough c-fiber cast-iron concrete PVC Glass ss-UHP
where: ss = stainless steel cs = carbon steel c-fiber = carbon fiber UHP = ultra-high purity (5 – 10 Ra finish)
Again, use the up and down keys to select. Choice will flash until the enter key is pressed. Once the selection is made, the instrument Dial-A-Pipe feature is complete.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Flow Units.
Calibration (Sub-Menu): Flow Units
Navigate to the “Flow Units” screen. Use Table 3-3: Calibration Sub-Menu for guidance if necessary. Use this menu to enter the units for mass velocity or flow. The display will show the previously selected flow units. For example:
Press enter and the units will begin flashing. Go to the next level with the button and select the proper pipe type. Again, use the up and down keys to select. The current choice will flash until the enter key is pressed.
Available Choices:
SCFS, SCFM, SCFH, SCFD, SCFY: Standard Cubic Feet per Second, Minute, Day,
Hour or Year. Standard conditions are set in the Reference Conditions menu.
NCFS, NCFM, NCFH, NCFD, NCFY: Normal Cubic Feet per Second, Minute, Day,
Hour or Year. Normal conditions are set in the Reference Conditions menu.
SM3/sec; min; hr; day; yr: Standard cubic meters per unit time. Standard conditions
are set in the Reference Conditions menu.
NM3/ sec; min; hr; day; yr: Normal cubic meters per unit time. Normal conditions
are set in the Reference Conditions menu.
SLPS,SLPM, SLPH, SLPD, SLPY: standard Liters per unit time. Standard
conditions are set in the Reference Conditions menu.
NLPS,NLPM, NLPH, NLPD, NLPY: normal Liters per unit time. Normal conditions
are set in the Reference Conditions menu.
Lbs/sec; min; hr; day; year: Pounds per unit time. Kg/sec; min; hour; day; year: Kilograms per unit time.
32
Page 33
SFPS, SFPM, SFPH, SFPD, SFPY: Standard Feet per Second, Minute, Day, Hour or
Temperature Unit
F
MenuTune
1.000000
Year. Note this is a point VELOCITY.
SMPS, SMPM, SMPH, SMPD, SMPY: Standard Meters per Second, Minute, Day,
Hour or Year. Note this is a point VELOCITY.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Temperature Units.
Calibration (Sub-Menu): Temperature Unit
Use this menu to enter the units for temperature. Use Table 3-3: Calibration Sub-Menu for guidance if necessary. The display will show the currently selected temperature unit.
Press enter and the units will begin flashing. Use the up and down keys to select. The units will flash until the enter key is pressed again.
Choices Available:
F (Fahrenheit) C (Celsius) K (Kelvin) R (Rankine)
Once the temperature units have been selected press enter () to save. You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Span Value.
Calibration (Sub-Menu): MenuTune
The MenuTune multiplies the meter output by a fixed value, from 0.5000000 to 2.000000. Displayed MeterTune is for the currently selected gas. This is useful if the meter displays a known offset from another device or expectation (calculated flow rate for example), and you want to make the meters “match”.
For instance, assume your process needs a less accurate rotameter that reads 100 SCFM, and the more accurate 640i or 780i that reads 95 SCFM to align. Enter a span factor of 100/95 =
1.052632. The meter will then multiply by this factor so that the meter output equals the rotameter output of 100 SCFM. Default is of course 1.000000.
Press enter () and the units will begin flashing. As in previous examples, use the  buttons to enter the desired value, and the enter key () to save the value.
33
Page 34
Low Flow Cut-Off
5.000000
Calibration Date
08/13/2012
You can press the exit several times to climb back up to the main menu level or press to
Full Scale Flow
100.00 SCFM
Low Flow Cut-Off
0.000000
proceed to the next Calibration Sub-Menu called Low Flow Cut Off.
Calibration (Sub-Menu): Low Flow Cut Off
The Low Flow Cut Off value forces both the digital display and the analog output values to zero/4 mA at a percentage of full scale from 0 to 10% of the full scale value. This is useful because pipes can show flow when all valves are shut.
The QuadraTherm 640i and 780i are extremely sensitive flow instruments that can pick up natural convection flows, where thermal gradients cause gas movement inside the piping system. Phenomenon’s like natural convection will lead to totalization errors.
Displayed value is for the currently selected gas.
Enter the desired low flow cut off as a percentage of the full scale flow. The full scale flow is calibrated at the factory and is found in the main menu. You can press the exit several times to climb back up to the main menu . For example, if the full scale flow as found in the main menu is:
And the meter consistently shows a flow of 2 to 3 SCFM even with all valves closed (and you are sure there are no leaks), an example is to set the low flow cut off just above to:
This is +/- 5% of the full scale of 100 SCFM. The analog output and the digital display be
0.000000 for any flow below 5.000000 SCFM. The default value for low flow cut off is
0.000000.
Once the low flow cut off has been entered, use the enter key () to save the value. You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Calibration Date.
Calibration (Sub-Menu): Calibration Date
The Calibration Date is a read-only field that shows when the meter was last factory-calibrated in MM/DD/YYYY format.
34
Page 35
Full Scale Flow
100.00 SCFM
Sierra recommends recalibration as required by your metrology policy or when the meter validation routine found in the Smart Interface Program software detects a problem. See Meter Validation in XXXXXX).
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Calibration Sub-Menu called Full Scale.
Calibration (Sub-Menu): Full Scale Flow
The full scale value is the flow value that corresponds to 20mA on the Flow 4-20mA analog output. This value can be adjusted between 50-100% of the original factory calibration full scale value found on the data tag. Note that accuracy is still based on the original factory calibration full scale value.
Press enter () and the units will begin flashing. As in previous examples, use the buttons to enter the desired value, and the enter key () to save the value. This concludes our review of the Calibration Sub-Menu. Press the exit climb back up to the main menu.
Process Pressure (Sub-Menu)
Navigate to the Process Pressure Sub Menu, use Table 3-4 below for guidance if necessary.
The QuadraTherm 640i and 780i is a multivariable thermal mass flow meter. It always measures mass velocity (from which the mass flow rate is derived) and temperature (VT) and has the optional ability to measure process pressure along with VT, this is called (VTP). The Process Pressure Sub-Menu allows you to input the desired pressure units.
Table 3-4: Process Pressure Sub-Menu
35
Page 36
Pressure Units
Psia
Process Pressure
14.700154
Process Pressure (Sub-Menu): Pressure Units
Use this menu to enter the units for pressure.
Use the enter key () to drill down to Pressure Units selection. Press enter and the units will begin flashing. Use the buttons to enter the desired information, and the enter key () to save the value.
Available units are:
Psia Psia Bar A Bar G KPa A KPa G Kg/CM2 A Kg/CM2 G In H2O A In H2O G MM H2O A MM H2O G
Once the pressure units have been selected press the enter key () to save. You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Process Pressure Temperature Units.
Process Pressure Sub-Menu: Process Pressure
Use this menu to enter the process pressure.
This process pressure is used by the instruments iTherm mathematical model to improve gas properties calculations. Gas properties have a very slight dependence on pressure, so if process pressure changes appreciably over time this value should be updated to reflect process conditions (about 0.01% per psi).
The units of the pressure in this menu are the same as what was set in the previous menu. In the example above, the units are 14.700154 Psia since this is what was set in the pressure units sub-menu.
This concludes our review of the Process Pressure Sub-Menu. Press the exit climb back up to the main menu.
36
Page 37
Totalizer (Sub-Menu)
Reset Totalizer
Yes
Reset Pulse Cnt.
349
The Totalizer Sub Menu controls all aspects of the totalizer function. Totalizers are used to monitor accumulated flow, often for billing and cost allocation purposes. The totalizer sub­menu is used to turn the totalizer ON or OFF, to reset, and to tell the meter when to send out a totalizer pulse and what that pulse should look like.
Totalizer (Sub-Menu): Reset Totalizer
Navigate to the Reset Totalizer screen, use Table 3-5 above for guidance if necessary.
Use this menu to reset the totalizer to zero. Press enter () and you will see a flashing “YES”. Pressing enter () again to reset the totalizer to zero and bring you back to the Totalizer (Sub-Menu).
In addition, from the screen above you can press the down button to a flashing “NO”. Pressing enter will return back to the Totalizer (Sub-Menu).
Totalizer Sub-Menu: Reset Pulse Count
If you need to actually set up the totalizer (rather than just reset it to zero), navigate to the Reset Pulse Count screen, use Table 3-5 above for guidance if necessary.
Use this menu to reset the pulse counter. The pulse count is the count of pulses that went to the pulse relay since the last reset. This count will match your external counter if they were both reset at the same time and can be used to verify an external counter. Pulse count X units per pulse= totalizer count within +/- 1 units per pulse value.
The screen example below, there were 349 pulses since the last reset.
Table 3-5: Totalizer Sub-Menu
37
Page 38
Pulse Width
50 ms
Reset Pulse Cnt.
YES
Reset Pulse Cnt.
0
To reset this to zero, press enter () and a “YES” will appear and flash.
Press enter () again and the pulse counter will be reset. In addition, from the screen above you can press the down button to a flashing “NO”. Pressing enter () will return back to the Reset Pulse Count screen. If you answered YES, the count will read zero “0”.
Totalizer (Sub-Menu): Pulse Width
Use this menu to set the width of the pulse in milliseconds. Navigate to the Pulse Width screen, use Table 3-5 above for guidance if necessary.
The pulse output is an isolated relay contact, so one pulse of for example 50ms closes the normally open relay for 50ms. Some counters work straight off of the relay contacts closing and some pull up the relay so when it closes it is a negative 50ms pulse. The 640i and 780i can use just the contacts or an approximately 10K pull-up resistor.
Choices Available:
On (for testing: closes the normally open relay; default is 50 msec) Off (for testing, opens the relay; default is 50 msec) 50 ms 100 ms 250 ms
In the screen example above, the pulse width will close the isolated normally open contacts for 50ms, the frequency of the pulse varies with flow rate and units per pulse value (see below) but cannot exceed 1Hz maximum.
Press enter () and the units will begin flashing. Press the up button or down button to choose the desired pulse. Next, press the enter key () to save. You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub­Menu called Units Per Pulse.
38
Page 39
Units per pulse
1.700000
Totalizer ON
Totalizer (Sub-Menu): Units Per Pulse
Use this menu to select the units per pulse. Navigate to the Units Per Pulse screen, use Table
3-5 above for guidance if necessary. Note the maximum frequency at which the totalizer
can count is 1 pulse per second, or 1 Hz. This puts a limit on the units per pulse and makes resolution dependent on flow rate.
For example: Assume the maximum flow rate is 100 SCFM and you want the totalizer to pulse every 5 SCF. Since the flow is 100 SCFM, it will take 5/100 minutes, or 3 seconds to totalize 5 SCF. That is acceptable since the totalizer takes longer than 1 second to accumulate SCF.
Assume that now you want more resolution, so you would like the totalizer to pulse every 1 SCF. Since the flow is 100 SCFM, it will take 1/100 minutes, or 0.6 seconds to totalize 1 SCF. Since the instrument’s totalizer is 1 Hz. It can’t pulse any faster than a maximum rate of once per second (1 Hz).
In this example, the best units per pulse would be 1.7 SCF; that is just over the 1 HZ maximum sampling rate.
Press enter () and the units will begin flashing. As in previous examples, use the buttons to enter the desired value, and the enter key () to save the value.
Press to proceed to the Totalizer On/Off screen or press the exit several times to climb back up to the main menu level.
NOTE: The units of the totalizer are the same as entered in the units menu (minus the time since this is an accumulation and not a rate).
Totalizer (Sub-Menu): Totalizer On/Off
Use this menu to turn the totalizer On or Off. Use Table 3-5 for guidance if necessary. Press enter () and the variable will begin flashing. As in previous examples, use the buttons to select On or Off, and press enter () to save the value.
Note: The Totalizer can only be used when mass units are being used. For example, Lbs/Hr or Kg/Hr.
39
Page 40
Totalizer set up in now complete. This concludes our review of the Totalizer (Sub-Menu).
Active Alarm
Flow
Press exit to climb back up to the main menu.
Alarm (Sub-Menu)
The alarm sub menu controls all aspects of the 640i and 780i Alarm feature. The instrument can be set to alarm on one of the following variables: Flow, Pressure, Temperature or Totalizer. The instrument alarm can be set on one of the following conditions: Low, High or Window. The Window alarm becomes active when the parameter reads higher OR lower than the set point (when it is outside of its operating window).
Table 3-6: Alarm Sub Menu
Alarm (Sub-Menu): Active Alarm
Navigate to the Active Alarm screen, use Table 3-6 above for guidance if necessary. Press enter and the variable will begin flashing. In the screen example below, the active alarm is Flow.
Use the buttons to select On or Off, and press enter () to save the value.
Alarm Choices Available:
Flow Pressure Temperature Totalizer Off On (use only to test the alarm relay)
40
Page 41
NOTE: Only one alarm can be active at any one time. The alarm relay is a normally open
Active Mode
Low
isolated contact. The alarm is tripped when the relay closes.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Alarm Mode.
Alarm (Sub-Menu): Alarm Mode
Navigate to the Alarm Mode screen, use Table 3-6 above for guidance if necessary. Press enter and the variable will begin flashing. In the screen example below, the active alarm condition is “Low”.
Use the buttons to select On or Off, and press enter () to save the value.
Available Choices:
High Low Window Off On
Alarms (Sub-Menu)
The Alarms (Sub Menu) allows you to set the Low and High alarm values for Flow, Pressure, Temperature, and Totalizer.
Table 3-6.1: Flow Alarms (Sub Menu)
41
Page 42
We will set a low flow alarm to demonstrate the Alarms function. Navigate to the Flow Alarms
Low (SCFM)
000010.000000
(Sub Menu) screen, use Table 3-6.1 above for guidance if necessary. Assume that you already
set the Active Alarm to “Flow” and the alarm condition is set to go off on a “Low” flow. The
Flow Alarms (Sub Menu) allows you to set the actual alarm setpoint along with hysteresis.
Press enter () and the units will begin flashing. As in previous examples, use the buttons to enter the desired numeric value, and the enter key () to save the value.
In the example screen below, 10 SCFM has been entered. The units were already set in the units menu.
If you wanted to now set a High alarm, press the right key and the High alarm setup screen will appear.
Note: If you wanted to set the active alarm on pressure, temperature or totalizer (instead of flow), press the right key to reach the appropriate sub-menu and use the buttons to enter the desired numeric value, and the enter key () to save the value
This concludes our review of the Calibration Sub-Menu. Press the exit climb back up to the main menu.
Output Adjust (Sub-Menu): Flow, Pressure and Temperature
The QuadraTherm® 640i and 780i comes standard with linear 4 to 20 mA analog outputs for flow and temperature. If the pressure option is selected, a third linear 4 to 20 mA analog output is provided for pressure.
Use the output adjust menu to view, test, or adjust the 4 and 20 mA raw digital DAC counts value for these parameters as well as to view or adjust the pressure and temperature values that
Table 3-7: Output (Sub Menu)
42
Page 43
correspond to 4 and 20mA. Zero flow always corresponds to 4mA and full scale flow always
Flow 4 mA
11373
Flow 4 mA
11373 - Testing
Flow 20 mA
56643
corresponds to 20mA.
All values in the output menu have been calibrated and set at the factory. You only need to enter these screens if you wish to view, adjust, or test any of the values.
Setting Up: 4-20 mA Outputs
We will set Flow 4-20 mA to demonstrate this function. The same routines apply for setting up Pressure 4-20 mA and Temperature 4-20 mA.
Flow 4 mA : View, Test, Adjust
Navigate to the Flow 4 mA screen in the Output (Sub Menu), use Table 3-7 above for guidance if necessary. From this menu, press enter to get to the flow 4 mA setup screen.
This numeric value is the DAC value that produces 4 mA on the analog flow output. In the example screen DAC is 11373. Press enter () and the units will begin flashing. Increasing this value will raise the 4 mA output and decreasing this value will lower the 4 mA output. Use the buttons to enter the desired numeric value, and the enter key () to save the value.
You can then use the down button to force/test flow 4 mA output and hit the up button to stop the force/test. It will flash testing until you stop the test. You can repeat this process, if desired, until you achieve the desired result.
Flow 20 mA : View, Test, Adjust
Navigate to the Flow 20 mA screen in the Output (Sub Menu), use Table 3-7 above for guidance if necessary. From the Flow 4 mA screen, pressing the right button will send you to the view, test, or adjust screen for 20 mA value.
This value is the DAC value that produces 20 mA on the analog flow output. In the example screen DAC is 56643. Press enter ()and the units will begin flashing. Increasing this value will raise the 20 mA output and decreasing this value will lower the 20 mA output. Use the buttons to enter the desired numeric value, and the enter key to save the value.
You can then use the down button to force/test flow 20 mA output and hit the up button to stop the force/test. It will flash testing until you stop the test. You can repeat this process, if
43
Page 44
desired, until you achieve the desired result.
Pressure Low
0.00
Pressure High
30.00
Temperature Low
32.00
Flow 20 mA
56643 - Testing
NOTE: The engineering units are as set in the Calibration (Sub Menu). Go to that section for more detailed information.
In the example above, assume full scale flow is 100 SCFM. As a result, a 20 mA output equals 100 SCFM and 4 mA output equals 0.00000 SCFM. These values are fixed for flow. The output is linear between these two points, in this example, if we had a flow of 50 SCFM, the output would be 12 mA.
Once you have completed viewing, testing, and adjusting the flow, pressure, and temperature 4-20 mA DAC values press the right key to reach the appropriate sub-menu and use the buttons to enter the desired numeric value, and the enter key () to save the value
Getting 4 to 20 mA for Temperature and Pressure
If you wanted to set the 4 to 20 mA active alarm on pressure and temperature (instead of flow).
The pressure units above are the current pressure units you ordered or that you have changed in the process pressure sub-menu. In the example above if your pressure units are psia then 0.00 psia would correspond to the pressure 4mA output. Press enter key () and use the left right up and down keys to modify this value if desired and then press the enter key () to accept. Press the right key to continue to the pressure high value that corresponds to the 20mA output.
In the example above, if your pressure units are psia then 30.00 psia would correspond to the pressure 20mA output. Press enter key () and use the left right up and down keys to modify this value if desired and then press the enter key () to accept. If you change this value be sure not to enter a Pressure High value greater than the full scale of the pressure option you ordered or you would be over the range of the transducer. If you change these pressure low/high values the pressure accuracy is still based on the original +/-1% of full scale of the Pressure transducer option ordered. Press the right key to continue to the temperature low value that corresponds to the 4mA output.
The temperature units above are the current temperature units you ordered or that you have changed in the calibration sub-menu. In the example above, if your temperature units are °F then 32.00 °F would correspond to the temperature 4mA output. Press enter key () and use
44
Page 45
the left right up and down keys to modify this value if desired and then press the enter key ()
Temperature High
212.00
Ref. Con.
(Sub Menu)
Standard Temp.
70.000000 / F
Standard Press.
14.695949 / Psia
to accept. Press the right key to continue to the temperature high value that corresponds to the 20mA output.
In the example above, if your temperature units are °F then 212.00 °F would correspond to the temperature 20mA output. Press enter key () and use the left right up and down keys to modify this value if desired and then press the enter key () to accept. If you change these temperature low/high values the temperature accuracy is still based on the original +/-1°C specification.
In the above examples, from 32.00° F to 212.00° F or from 0 to 30 psia would correspond to the 4 and 20mA analog outputs for temperature and pressure. These values are dependent upon your process. Finally, exit back to the set-up menu and right to reference conditions.
Reference Conditions (Sub-Menu): Standard and Normal
Use this menu to establish the reference conditions as specified in the units menu. Navigate to the Reference Conditions (Sub Menu), use Table 3-8 below for guidance if necessary.
Press enter () once and then use the  buttons to select “Standard” or “Normal” .
45
Page 46
Table 3-8: Reference Conditions (Sub Menu)
Standard Temp.
70.000000 / F
Standard Press.
14.695949 / Psia
Normal Temp.
0.000 / C
Normal Press.
101.325/KPa A
For example, SCFM (Standard Cubic Feet per Minute) is commonly used in the United States. The S = Standard in SCFM and is referenced to 70°F (21.1°C) and 14.696 Psia (or 1 atmosphere).
In Europe, Nm3/hr (Normal Cubic Meters per Hour) is commonly used. The N = Normal and is generally referenced to 0°C and 101.32 KPa A(or 1 atmosphere).
Once a “Standard” or “Normal” selection has been made press enter () again to go to the adjust screen.
From here you can adjust instrument reference conditions . To adjust, press () again and the units will begin flashing. As in previous examples, use the buttons to enter the desired value, and press enter () to save the value.
Once you are done press the enter key () to save. You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Diagnostics.
Diagnostics (Sub-Menu)
In the Diagnostics Sub-Menu, you can read the maximum and minimum: flow, pressure & temperature.
Navigate to the Diagnostics (Sub Menu), use Table 3-9 below for guidance if necessary.
Use the left and right buttons to view these values. These are the maximum and minimum values that the flow meter has measured. Press the exit button several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Change Password (Sub Menu).
NOTE: Extensive diagnostics are available via the Smart Interface Program (SIP) software.
46
Page 47
Table 3-9: Diagnostics (Sub Menu)
Change Password
1234
Change Language
US-English
Change Password (Sub-Menu)
Use this menu to set or change a password. Navigate to the Change Password (Sub Menu), use
Table 3-9 above for guidance if necessary.
A password is required to access the set-up sub menus. Press enter () and the four digit password units will begin flashing. As in previous examples, use the buttons to enter the desired value, and press enter () to save the value.
The default password is 0000. If you have lost your password and can’t gain needed access to the setup screens, you can contact our service department for a password that would allow you to access the setup screens and setup a new password.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Change Language.
Change Language (Sub-Menu)
Navigate to the Change Language (Sub Menu), use Table 3-9 above for guidance if necessary.
Use this menu to change the displayed language. Use SIP program to download latest languages to your device.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Restore Factory.
47
Page 48
Caution!
Before attempting any flow meter repair, verify that the line is not pressurized.
Caution!
Always remove main power before disassembling any part of the mass flow meter.
Restore Factory
YES
Box Car
020
Restore Factory (Sub-Menu):
Use this menu to restore all parameters to the original factory conditions. This will undo any changes you have made.
Navigate to the Box Car (Sub Menu), use Table 3-9 above for guidance if necessary.
Press enter () and the YES will begin flashing. As in previous examples, use the
buttons to enter YES or NO, and press the enter key () to save.
You can press the exit several times to climb back up to the main menu level or press to proceed to the next Sub-Menu called Box Car.
Box Car (Sub-Menu):
Use this menu to view or change your current box car value. Navigate to the Change Language (Sub Menu), use Table 3-9 above for guidance if necessary.
This value is a smoothing or averaging function of the indicated flow rate. This value can be from 1 to 200. Increasing this value will smooth-out the indicated flow rate more and slow down the time response slightly.
Decreasing this value will make the indicated flow noisier, but will also speed up the response time slightly.
Press enter () and the units will begin flashing. As in previous examples, use the buttons to enter the desired value, and the enter key () to save the value.
Hit the exit key to return to the main top menu.
48
Page 49
Chapter 4: Troubleshooting & Repair
Troubleshooting the Flow Meter
Begin hardware troubleshooting by verifying the following facilities is­sues are correct. These areas impact system operation and must be corrected prior to performing any flow meter inspections.
1. Verify the incoming power to the flow meter is present and of the correct voltage and polarity.
2. Check the flow meter wiring for correct connections as described in Chapter 2.
3. Verify the flow meter is installed with the correct number of up­stream and downstream pipe diameters as shown on page 2-2.
4. Verify the flow direction indicator is correctly aligned pointing downstream of flow.
5. Make sure there are no leaks in the line being measured.
After verifying the factors above, follow the troubleshooting procedures outlined on the next page. If you need to return the flow meter to the factory, see page 4-3 for return shipping instructions
.
49
Page 50
Flow rate below meter’s minimum
flow rating Flow has exceeded the maximum
range of the flow meter
Contact factory for instructions
Set the user full scale to equal the fac­tory full scale
Reduce flow below the maximum range shown on the meter’s nameplate or con­tact the factory for re-calibration advice
No power Turn on power to the flow meter
Problem Possible Cause Solution
Velocity measurement is Very erratic or non-uniform flow Follow installation requirements shown erratic or fluctuating in Chapter 2
Velocity measurement seems too high or low
Flow meter installed with less than required minimum pipe diameters upstream and downstream of the sensor
Insertion sensor probe not mounted securely
Sensor component broken Return to factory for replacement
Malfunction in system Return to factory for evaluation electronics
Ground loop Check wiring, see Chapter 2
Moisture present in gas flow Install a water trap or filter upstream of
Sensor assembly not aligned cor­rectly to flow
Sensor probe not inserted to the proper depth
Follow installation requirements shown in Chapter 2
Sensor probe must be mounted se­curely without vibration
the flow sensor Correct alignment with the flow indicator
pointing downstream in the direction of flow
Verify sensing point is located on the centerline of the pipe
No response to flow from sensor assembly
using the Smart Interface software.
Low flow cutoff too high Correct low flow cutoff programming
Flow profile distortions Try to find another location for the meter Extremely turbulent flow Do not place the meter near a ventilator,
static mixer or valve
Sensor failure Return to factory for evaluation Printed circuit assembly defective Return to factory for evaluation
50
Page 51
Returning Equipment to Factory
Factory CalibrationAll Models
Sierra Instruments maintains a fully-equipped calibration laboratory. All measuring and test equipment used in the calibration of Sierra transducers are traceable to NIST Standards. Sierra is ISO-9001 registered and conforms to the requirements of ANSI/NCSL-Z540 and ISO/IEC Guide 25.
Instructions for Returning Your Instrument for Service
The following information will help you return your instrument to Sierra Instruments' Factory Service Center and will ensure that your order is processed promptly. Prices may vary depending on the flow range, type of gas and operating pressure of your unit. To request detailed pricing, contact your local Sierra Instruments distributor or contact one of our offices directly.
Please follow these easy steps to return your instrument for factory service:
1. To obtain a Return Materials Authorization (RMA) number from the Sierra
Instruments to http://www.sierrainstruments.com/rma/new.php to create a Sierra Account.
2. Once you have created an account, click on the Submit New RMA tab and fill in the
RMA form and follow the instructions. You will receive an email confirmation once you have submitted your RMA.
3. Print a copy of the RMA (that now includes RMA #) and send a copy of the RMA
form along with your meter back to the factory.
If you require service beyond calibration, but do not know which service(s) will be required, describe the symptoms as accurately as possible on the RMA form.
Pack your instrument carefully. Use the original packaging and foam or bubble wrap (packing peanuts NOT recommended) and include a copy of the RMA form (complete with Sierra supplied RMA number) with the unit(s).
Ship the unit(s) to the following address:
Sierra Instruments, Inc.
Attention: Factory Service Center
5 Harris Court, Building L
Monterey, CA 93940 USA
RE: RMA# (your number)
51
Page 52
Appendix A: Product Specifications
52
Page 53
53
Page 54
54
Page 55
55
Page 56
56
Page 57
57
Page 58
58
Page 59
59
Page 60
60
Page 61
Appendix B: Smart Interface Installation
Install the RS-232 Cable
Install the provided RS-232 pigtail to the terminal board (TB) as shown below, then install power to the meter.
Installing the Program
System Requirements:
Operation System: Windows
QuadraTherm 640i & 780i Sierra Interface Program (SIP): Click on the file (640i & 780i Sierra Interface Program). This file is self-extracting. Follow instructions for installing the files. Process takes about 1-10 minutes total depending on speed of your computer.
QuadraTherm 640i & 780i Sierra Interface Program Windows Compatible
Instructions: Click on Setup let it install the 640i & 780i Smart Interface Program Installation Disc. If you have any questions about downloading the files please contact Rory Ross at:
r_ross@sierrainstruments.com.
61
Page 62
Program Start-up
The smart interface program allows the user to configure the unit and view all unit parameters.
NOTE: The current version of the program. You may obtain the latest version from
www.sierrainstruments.com.
Make sure your meter is powered up and the meter is connected via the serial or USB port to the computer running this SIP program. Click on the 640i and 780i SIP Icon to start the program.
Note select Port. To begin, use the pull down menu to select the appropriate com port. Once the port is selected, the meter data is read into the SIP program. Note how all data fields become populated. This may take several seconds. You will see a progress bar in the lower right as the meter data is collected. Be aware that all the data fields are read only. Entering and writing this data is done via the Function Selector or quick keys. The quick keys and will be discussed in those sections.
62
Page 63
Meter Data
The meter data shows the basic identifying data of the meter hardware, firmware and manufacture date. The calibration date represents the last time the meter was factory calibrated.
Latest PCA and Firmware revisions are posted at www.sierrainstruments.com. The iTherm™ math model is continuously evolving, with new gases added to the library periodically, so check the firmware revision periodically in order to take advantage of these refinements. Factory recalibration is at the discretion of the end user. With Dial-A-Pipe and Dial-A-Gas technology the 640i/780i can be reconfigured to fit into a large variety of applications.
Meter Display
The meter display shows the primary variables (mass flow, temperature, pressure and totalized flow, as well as what gas and pipe have been selected. Both Dial-A- Gas and Dial-A-Pipe are field-configurable, offering unparalleled flexibility. Alarm status is also shown.
63
Page 64
In this example:
The gas being measured is air The calibrated full scale (FS) is 1440.40 SCFM The inside diameter of the pipe is 4.0260 Inches The meter is currently reading 0.0000 SCFM (there in no flow in the pipe) The meter has totalized 839.440 SCFM (there WAS flow in the pipe at some point) The temperature is 85.68 °F The meter was purchased without the pressure option, so pressure is not active The low alarm is set to energize at 10.0000 SCFM Since flow is 0.0000 SCFM, the alarm is ON indicating flow is below the 10.000 SCFM
set point
Meter Control
A series of quick keys, represented by the red keys, are provided to allow for rapid configuration of application conditions and alarm set-up
More detailed configuration can also be done using the function selector (in the above example
currently showing alarm setpoint)
64
Page 65
Quick Keys: Dial-A-Pipe
The Dial-A-Pipe feature makes field configuration very convenient. You must know the type of pipe and either the pipe schedule or the inside diameter.
If inside diameter is known, select the units from the diameter units menu and type in the number in the Inside diameter field.
If the pipe schedule is known, select it from the pipe schedule pull down menu. The inside diameter will then be auto-populated with the correct value (in the units as selected in the Diameter Units drop-down menu). Currently only ANSI schedule 40 is supported but future updates will address other schedules and types.
Finally, use the pipe type pull down menu and select the type.
In the example below we have selected 4-inch ANSI Schedule 40. The program knows this has an inside diameter of 4.0260 inches. We have selected ss-smooth (stainless steel-smooth inside wall) as the closest match to our application.
65
Page 66
Quick Keys: Dial-A-Gas and Units
As with Dial-A-Pipe, Dial-A-Gas is another powerful feature of the 640i/780i. Simply select the gas to be measured from the pull-down menu. The menu can support 4 different gases and can be updated from Sierra’s gas-properties library as new gases and mixtures are added.
The Gas and Units quick key also allows the selection of gas, temperature and pressure units. Simply select the desired units from the pull down menus. If you need other gases in the future, contact the factory and we can download updates from our gas library.
iTherm Gas Library
Selecting iTherm Gas Library from the Dial-A-Gas pull down menu allows users to access Sierra’s unique Dial-A-Gas library. Our library is continually growing and expanding. As Sierra calibrates more and more gases with the 640i, we are able to improve the accuracy of our current iTherm gases and add new ones. These new or improved gases are available for download from our iTherm Gas Library, and can then be uploaded into your 640i as a replacement gas for any of the three slots available (remember Air is always in slot 1 and is NOT replaceable).
66
Page 67
To start the iTherm gas library, select and save iTherm Gas Library from the drop down menu. This will bring up the following screen:
Local Gases
All Gas Files saved on your computer are displayed in the ‘Local Gases’ window. These are not necessarily the gas files in your meter, but ones you have downloaded from the Sierra Server previously. If you have never done so, this area will be blank.
Most recent version files are saved in {Application Directory}\640i\gas Older versions are stored in {Application Directory}\640i\gas\archive
Remote Gases
By clicking the ‘List Remote Files’ button, the Remote Gas window is populated with gas files available for download. You must have an internet connection to connect to the Sierra server. If the gas file is a newer version than a corresponding local file or is not present locally, then it appears in red. Files that are already saved locally are black.
67
Page 68
Download Gas File (Arrow Button)
To download a remote gas file, the file is selected in the Remote Gas window and the arrow button is clicked. Once the download is completed, the gas file will appear in the Local Gases window and will no longer appear as red in the Remote gases window. If a new version of a gas file is downloaded, the older local version is moved into the Archive directory.
Load Gas
To load a gas into one of the 3available slots (Air is in Slot 1 and cannot be changed), select the gas to be replaced in ‘Gas’ dropdown on the left and click the ‘Load Gas’ button, then press load gas. You will see a progress bar indicating the upload status. Upon completion your old iTherm gas will have been replaced!
68
Page 69
Quick Keys: Alarm Setpoint
Pressing the Alarm Setpoint Quick Key brings up the following details about the alarms:
Only one alarm can be active at any time. Use the Active pull down menu to select the active alarm or to turn off the alarm function entirely.
Once you have selected the active alarm, use the Mode pull down menu to select whether the meter should alarm on low, high or to set an alarm window. A window alarm will turn on once the parameter drops below the low alarm or exceeds the high alarm setpoints, so it serves as a combination low and high alarm (the alarm is off when the active parameter is within the “window”).
69
Page 70
You may now enter the alarm low and high values. To enter, click on the field and type in the numeric values for the desired alarm. Hit save and the entered value will be written into memory.
In the example above, flow is selected as the Active alarm. We have indicated we want the alarm to activate in Low Flow mode, and we want the alarm to activate at 10.0000. The units as set up in the units menu are in SCFM, so once flow drops below 10.000 SCFM we should see an alarm condition.
Quick Keys: ValidCal™ Diagnostics
ValidCal™ Diagnostics is a complete meter diagnostics suite. It tests all analog outputs, relays and primary variable measurements.
ValidCal™ Diagnostics is designed to be performed in AIR under ambient conditions (room temperature and pressure) It must be performed under no flow conditions. The meter must be removed from the pipe if possible, and the Sierra end cap placed over the sensor. The test may be performed in the pipe provided these same conditions are met while the probe tip is inside the pipe.
70
Page 71
If any of the tests fail, follow the troubleshooting tips below and the more general ones contained
in chapter four of this manual.
4-20 mA Output Validations
The 4-20 mA output validation allows users to force a 4 or 20 mA output on all three analog outputs. To test:
Hook a DVM (select DC current) to the +/- terminals of the output you are validating:
71
Page 72
Press Test 4 mA and enter the value from your DVM (should be around 4.000 mA)
The program will show passed provided the reading is within tolerance. If it shows failed, please contact the factory for assistance. Repeat this procedure for all remaining 4- 20 mA tests. The most common failure is that the 4- 20 mA wires have become loose or that the DVM is incorrectly set up (must be set up for measuring direct DC current in the mA range).
The 4 to 20 mA DAQ conts may be adjusted by going to the “Flow, Pressure and Temperature 4- 20mA Output Range Setup” in the function selector.
Flow Sensor Validation
This routine tests the flow sensor and sensor drive circuitry by measuring the Watts required to heat the velocity sensor to a value of 50°C above the current temperature of the sensor.
The reference flow Watts is the watts required at the time of calibration to attain this 50°C overheat. Barring damage or drift, this value should remain relatively constant over time.
72
Page 73
Selecting validate Flow will bring up the following screen:
The test will automatically begin. Look at actual temperature. It will begin increasing as watts
are added to the flow sensor, heating it up to 50 °C (122 °F). Once this temperature reaches set point, the test will complete itself, usually within 5 minutes. If the Reference Flow is within tolerance of actual flow, the meter will pass validation. If the test fails, ensure you have allowed adequate time for the sensor to cool down before starting the test, that the probe tip is covered and that you began the test with no flow present and at room temperature and pressure. True sensor failures are rare and will be covered by Sierra’s lifetime sensor warranty.
Temperature Validation
This routine checks the QuadriTherm gas temperature sensors (T2, T3 and T4). Select Validate
Temperature, then Enter the temperature in the pipe as read by another device (not the Sierra flow meter) in the pop-up window. The readings should be roughly the same. Note it can take up to 20 minutes for the probe tip to cool to the surrounding temperature once it has been removed from the pipe. This is the most common “failure”. Allow the probe tip to cool and repeat the test.
73
Page 74
Pressure Validation
This routine validates the gas pressure sensor. Select validate pressure, then enter the pressure in
the pipe or test area as read by another device (not the Sierra flow meter) in the pop-up window.
The readings should be roughly the same. If there is no pressure transducer, the Screen will
appear as below.
The typical failure for a pressure transducer is if it has been over pressurized and the pressure
sensor membrane has burst. You pressure transducer has a burst pressure of 5 times its operational rating. For example, an MP2 100 psia sensor has a burst pressure of 500 psia.
Alarm/Totalizer Relay Validation
This routine tests the Alarm and Totalizer relays.
Test Alarm Relay
Hook up a DVM as shown:
74
Page 75
Use the pull down menu to select relay Open and close and use your DVM to check the relay.
Test Totalizer Relay
Hook up DVM s shown:
Open and close the relay and use your DVM to check the relay.
75
Page 76
Low/High FTP
Pressing the Low/High FTP button (located at the bottom left of the main ValiCal screen brings
up the following data screen. It shows the lowest and highest flow, temperature and pressure the
meter has experienced since the last reset. This information is valuable in documenting events
that might have caused meter damage, such as blowing out the pressure transducer or severely
overheating the sensor.
Quick Keys: Meter Tune (Span)
Meter Tune (Span Adjust) allows you to fine tune your meter reading to match other
devices or expectations. Simply enter the reading from the other device and the meter span will be auto adjusted to make the readings the same.
76
Page 77
Quick Keys: Exit
Exit will immediately close the program. Any unsaved data will be lost.
Function Selector
The function selector does all of the same things the quick keys do, while giving deeper access to
some functions. To operate the selector, pick the desired function from the drop-down menu and
hit enter or left click.
77
Page 78
Dial-A-Pipe
See Dial-A-Pipe Quick Key Description
Dial-A-Gas and Units
See Dial-A-Gas Quick Key Description
iTherm Gas Library
See iTherm Gas Library information on page 65
Alarm Setpoint
See Alarm Setpoint Quick Key Description
ValidCal Diagnostics
See ValidCal Quick Key Description
MeterTune (Span Adjust)
See MeterTune Quick Key Description
Reference Conditions
78
Page 79
This function allows the user to enter the normal and standard temperatures and pressures to
which gas MASS flows are referenced. Typically this is 70°F and 1 atmosphere in the USA (f.i.
standard conditions as used in scfm: standard cubic feet per minute) or 0°C and 1 atmosphere in
Europe (f.i. normal conditions as used in nlpm: normal liters per minute).
Totalizer Setup/Reset
If units are not mass flow, it will not turn on. The totalizer will totalize accumulated flow if it is
turned on. Turn it on or off with the Totalizer Status drop down menu.
NOTE: The totalizer operates at a maximum frequency of 1HZ (1 pulse per second). This
limits the maximum units per pulse.
For example, you have a flow of 60 scfm with a units-per-pulse set to 1 SCF This means 1 pulse
per second or 1 Hz. This is within the ability of the totalizer.
Reducing the flow to 30 scfm means the totalizer will pulse once every two seconds. Again this is
OK.
Increasing the flow to 120 scfm means the totalizer would pulse once every 0.5 sec, but this is too
fast.
The totalizer will not count (all the pulses will run together).
This puts a limit on the resolution of the totalizer at higher flows. You must convert the flow to
units per second and ensure this number is equal to our bigger than one.
The pulse itself is a negative going square wave pulse. The duration of the pulse can be set with
the pulse width pull down menu.
79
Page 80
Low Flow Cut Off Value
Due to thermal gradients, movement of gas molecules in the pipe can occur even with all valves
shut. The Zero Cut Off forces both analog and digital outputs to zero at a percentage of the full
scale. The Zero Cut off can be set to from 0 to 10.00% FS (full scale), meaning anytime the
measured flow drops below this value the analog and digital outputs will be forced to zero.
Totalization will not occur.
Full scale is the maximum flow rate for which the meter was calibrated. Exceeding it will not
damage the meter, but flow readings will not be accurate. The calibrated full scale is found in the
calibration menu.
Flow/Pressure/Temperature 4 – 20 mA Output Setup
This function allows you to set up the analog outputs. The 640i and 780i comes standard with
linear 4-20 mA analog outputs for flow and temperature. If the meter is orders with pressure, a
third analog output for pressure is available. For flow, 4 mA is always = 0 flow. The user may
80
Page 81
select 20 mA to equal any value up to the calibrated full scale. Full scale is the maximum flow rate
for which the meter was calibrated. Exceeding it will not damage the meter, but flow readings
will not be accurate. The calibrated full scale is found in the calibration menu. Note the meter
will still read digital flow values up to the calibrated full scale, but analog outputs will be scaled
per the user settings.
Pressure and temperature may be scaled as the user desires.
All units are as set in the units menu. The 4 and 20 mA analog outputs can be adjusted by
adjusting the DAQ counts until 4 or 20 mA is observed. For example if the flow 4 mA output is actually reading 3.98 mA, adjust the DAQ counts up from 1130 and press test and repeat until the readings has increased to 4.00 mA,
Pressure Math Value
This is the pressure used by the iTherm math to calculate gas properties. This makes gas
properties calculations more accurate.
81
Page 82
Save/Restore Data
This screen allows users to:
Save any new data Restore the meter to factory default (undoing any changes) Reboot (restart) the meter
Meter Information
User information allows you to view data about the meter and to customize some aspects of it.
82
Page 83
Display Language
Sierra’s iTherm Language Library is continually updated as our meter firmware is
translated into other languages. These languages can be downloaded from the Sierra server.
Local Languages
All Language Files saved lo on your computer are displayed in the ‘Local Languages’ window
Most recent version files are saved in {Application Directory}\640i\lang
Older versions are stored in {Application Directory}\640i\lang\archive
Remote Languages
By clicking the ‘List Remote Files’ button, the Remote Languages window is populated with
language files available for download. You must be hooked up to the internet so you can
download these files from the Sierra Server. If the language file is a newer version than a
corresponding local file or is not present locally, then it appears in red. Files that are already saved
locally are black.
Download Language File (Arrow Button)
To download a remote language file, the file is selected in the Remote Language window and the
arrow button is clicked. Once the download is completed, the language file will appear in the
Local Languages window and will no longer appear as red in the Remote Languages window.
If a new version of a language file is downloaded, the older local version is moved into the
Archive directory.
83
Page 84
Load Language
Three language slots are available. To upload a new language to the meter, select the language
you want to replace from the drop down menu, then select the language to upload from the local
languages list, then press load Language. The language is uploaded to the meter and is available
for use.
84
Loading...