The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
Register 4English
Print No.: RLL5-310.034.04.04.02 Doc. Gen. Date:12.98
12 ______Changes to previous version____________________________________12 - 1
Siemens AGRegister 4RLL5-310.034.04Page 5 of 6UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
0 - 6Contents
Page
This page intentionally left blank.
UROSKOP D3Register 4RLL5-310.034.04Page 6 of 6Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
CAUTION
General information1
Safety information1
When performing service work and tests, please adhere to the
product-specific safety informati on contained in the document, as
well as the general safety information contained in Register 2 of
the TI binder.
Tests and adjustments performed with radiation ON are identified with the radiation warning symbol . Radiation protection must be worn during these types of adjustments.
General information and information on documentation1
These instructions contain the checks necessary for system startup.The sequence in
which they are given must be maintained.
The unit movements necessary for these procedures must be carri ed out very carefully
to avoid damaging the equipment in the event of a malfunction.
Since the unit has been adjusted completely at the factory, only the functioning and factory settings (accuracy of tolerance) are checked; the unit must be adapted to on-site
requirements, if necessary.
1 - 1
If any mechanical or electrica l malfunctions oc cur or if tolerances are exceeded (damages
caused by transport etc.), perform ser vi ce work and adjust ments acc ording to adjustm ent
instructions RLL5-310.071.01... or the service instructions for the corresponding components.
Measurement results marked”” must be entered in the UROSKOP test certificate.
For the acceptance test which is requ ired i n Germany in accordance wit h §16 of the Rönt-
genverordnung (x-ray ordinance) and for the acceptance test required in t he USA, the following checks have already been performed in the factory test area and documented in
the test certificate:
• Visual check of the filt er values.
• Check of the SID display.
• Brightness of the light localizer.
• Coincidence of light fiel d and radiation field.
• Accuracy of the manual format col limation.
• Coincidence of radiatio n field and film center.
• Accuracy of positive beam li mitation (PBL) system for cassett e formats.
• Fluoroscopic field limitat ion.
• Centering of the radiation f ield and monitor image center.
• Function check of the Iontomat measurement fields.
• Unit attenuation facto r.
• Check of the tomographic device.
Siemens AGRegister 4RLL5-310.034.04Page 1 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
1 - 2General information
NOTICE
Enter the measurement values and the additional values determined during system start-up from the test certificate into the
acceptance certificate (Germany) or in the acceptance test certi fi cate (USA).
Preliminary information on image quality test1
Purpose:1
• The image quality test provides objec tive data as a basis for image quality and faci litates
system start-up.
• The test certificate is val id for system acceptance and for futur e service and mainte-
nance work.
Sequence of work:1
• Follow the sequence of measurements and checks ex actly as describes in order to
ensure correct measurements.
• Enter the results of each measurement i nto the test certificates.
• Compare the measured values with the values indi cated in the list of nominal values .
• If the limit values are exceeded, y ou must determine the cause and correct the er ror or
the erroneous setting before cont inuing the test. Otherwise the measurement s that
follow may be incorrect.
Requirements for start-up1
• The system must be cabled completely.
• Switch on high voltage only if radi ation is necessary.
• Setting Instructions POL YDOROS SX RX63-050.032.02...
• System configurationPOL YDOROS SX RX63- 050.034.06...
• IQ T est c ertificateRXD0-000.037.01... Logbook, Reg. 9
• T est certi ficateUROSKOP D3 RLL5-310.035.01...Logbook, Reg.3
• Test certificate POLYDOROS SX X2075Logbook, Reg. 4
UROSKOP D3Register 4RLL5-310.034.04Page 2 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
General information1 - 3
CAUTION
Required measuring instruments1
• 3-Phase field rotation meter , elect ronic 97 02 713 Y7933
• Internal line impedance meter84 28 104 RE999
• Service PC
• Oscilloscope > 50 MHz, e.g. Tektronix 2232
with delayed time base
• Multimeter 8060 A (Fluke)97 02 101 Y4290
• mAs meter81 60 400 RE999
• Water level28 69 436 RE999
• Set of Cu filters44 06 120 RV090
• 17 µm Cu strips (included in the LOG book) 1 1 67 663 G5247
• cm scale
• T omogr aphic cube with resolution test 44 06 054 RV 090
• Protective conductor meter44 15 899 RV090
• 8x magnifying glass44 14 850 RH090
• Precision radiation filter99 00 598 XE999
• Centering cross or lead rul er 96 60 051 RE999 or 28 63 025 RE999
• Mavo monitor , direct ordering97 02 432 Y0526
• Resolution test patterns, types 4128 71 820 RE999
• Dynamic test case37 90 156 X1963
• heart contour diaphragm37 90 172 X1963
• Capillary test37 90 180 X1963
• DSA displacement test (with DSA capi llary
test)
• Aperture plate for tomograp hy
97 02 234 Y3155
97 50 019 X1963 (only with DSA)
Successful system start-up can only be ensured by using the
above measurement instruments.
Siemens AGRegister 4RLL5-310.034.04Page 3 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
1 - 4General information
This page intentionally left blank.
UROSKOP D3Register 4RLL5-310.034.04Page 4 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
CAUTION
Start-up preparation2
Protective measures2
• Observe the protective measures described in Register 1 of the UROSKOP D Logbook.
• Prior to performin g any service work on the generator, swi tch it off with the power OFF
switch on D200.
• In order to deenergize all parts of the system, set the system switc h to the OFF position.
With the generator switched off, line voltage is still present at
transformer T1 and fuse panel D200 (refer to wiring diagram
X2075-1 1 ).
After switching off the generator, approx. 600 V DC voltage are
still present at the inverter; indicated by LEDs V8 and V9 on bo ard
D250 (see X2075-17).
Within approx. 1 1/2 minutes the voltage drops to 0 V; the LEDs go
out at approx. 30 V.
• To prevent unintentional rel ease of high voltage or radiation, set swi tch SS on D200 to
OFF (no actuation of the inverters).
• Remove or insert boards only wit h the generator switched off. Wh en doing so, observe
the ESD guidelines (see TI 219, Info RA.5. RA0- 000.012.19... on microfilm).
2 - 1
• On D211: switch S1 INT OFF (position 2) .
• On D200: switch SS OFF.
Measuring the line voltage2
• Remove fuses F1, F2 and F3 from the fuse board i n M16 (generator).
• System contactor ON.
• Measure the line voltage.
The voltage must correspond to the val ues indicated in the test certific ate (logbook,
Reg. 3).
Checking the on-site emergency shutdown button 2
• Actuate the emergency shutdown butt on installed by the customer. The entire syst em
must be disconnected from the power line.
• System contactor OFF.
Siemens AGRegister 4RLL5-310.034.04Page 1 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
2 - 2Start-up preparation
Checking the phase connection2
• Check the correct phase connecti on of the power lines at M16.F1, F2, F3 (X2075-11) ,
using the rotating field instru ment:
• In M16, remove the cover from the fu se holders.
• Connect the3-phase field rotat ion meter to the lower connections of F1, F2, F3 (on the
side of the power supply).
• System contactor ON.
• Perform the measurement.
• System contactor OFF.
• Disconnect the rotating field i nstrument.
Measuring the internal line impedance2
• Measure the internal line impedan ce at M16. F1. F2. F3 (X2075-11):
- Connect the internal line impedance meter bet ween two phases of F1, F2, F3 (at the
lower connections of the fuse holders ).
- System contactor ON.
- Perform the measurement.
- System contactor OFF.
• Record the measured values in the test certificate (l ogbook Reg. 3)
• Enter the measured values into the ”LINE PARAMETER” module of the system
programming.
Important: In order to obtain the full gener ator power, the internal line impedance
UROSKOP D3Register 4RLL5-310.034.04Page 2 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
Start-up preparation2 - 3
CAUTION
Checking unacceptable ground connections2
(connection between and 0VL ).
• Disconnect the connection l ead between the upper and lower grounding bus (in the
generator power cabinet).
• Using the multimeter, measure the resistance between and 0V
The measured resistance must be approx. > 50 k
Ω.
:
L
• Reconnect the upper and the lower grou nding bus.
Switching on the generator2
• System contactor ON:
- LED V26 (via fuse F11) must light up on fuse board D200.
In case of an error, check the fuses in M16 and on D200
(X2075-11 and 12).
• Generator ON:
- On fuse board D200, press key and actuate button .
- The LED’s must light up as follows on the service panel of D211:
The +5 V,
If the +5 V voltage is missing, the generator is disabled.
In this case, check the fuses on D200 ( X2075-12). All LED’ s on D200 must l ight
up.
- Following connection of the intermediat e circuit (activation of charging contactor and
power contactor), LEDs V9 and V8 on board D250 must light up (X2075-17).
± 15 V, +24 V supply voltages are immediately present.
A DC voltage > 500 V is applied to the intermediate circuit.
- Following correct initializati on (after approx. 10 sec.), the fol lowing LEDs must light on
the service panel of D211:
() ( ) TU1 (after system programming)
GEN-OK,+24 V, +15 V, -15 V, +5 V
If no error has occurred, the 7-segment displays on D210 indi cate ”E000”.
In case of an error, the three-digit error code is displayed.
- When programing the generator, "ALP" appears on the kV di splay of the control deck.
• Check the primary voltage for the s upply voltages (X2075-12) at the coil co nnections of
the GS contactor (X2075-11) in M16:
230 V
250 V
± 10 % at 50 Hz
± 10 % at 60 Hz
• Check the blocking:
On D200: Generator OFF
The generator cannot be switched on again f rom the deck.
Siemens AGRegister 4RLL5-310.034.04Page 3 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
2 - 4Start-up preparation
This page intentionally left blank.
UROSKOP D3Register 4RLL5-310.034.04Page 4 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
NOTICE
Generator functions3
Checking the emergency stop switches3
• System ON.
• Press the three system emergenc y stop switches consecutively:
With the emergency stop switch pr essed in,
- error code ’’U01’ ’ must appear in the
- all system movements must be blocked.
area dose product
Checking the control panel3
• System ON.
• Select from 40 kV to the max. kV value on t he kV control console by pressing the ± kV
keys.
• Check the mAs display by pressing the ± mAs keys.
display on the control console;
3 - 1
Function check of the anode starter3
• Generator ON : On D95, display1→2→3
• Set release button S27 to preparati on: the rotating anode boosts.
Display 3
Let go of the release button S27 again: the rotating anode should brake.
Display 6
Key to 7-segment display on board D95:3
1=Initialization4=Start-up
2=Init-brakes5=Continued run
3=Stand-by6=Brakes
If the ERROR message flashes, please refer to the information contained in the
service so ft ware.
LED status display3
→4→5
→3
The rotational speed must be checked with the tube assembly
warmed up.
light up depending on the
operating condition
Tube assy. 2 selected
Siemens AGRegister 4RLL5-310.034.04Page 1 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
*
Tube assy. 1 selected
*
3 - 2Generator functions
NOTICE
Checking the fluoroscopic and exposure release circuit3
Fluoroscopic circuit3
• On board D200: SS switch OFF.
• On D211: FL switch ON.
• On D211: the following LED’s must light up (X 2075-23):
- TU1/TU2TU 1/ TU 2 (acc. to selected workstation/tube assembly)
-small focus
-large focus
- GEN-OKGenerator enable state
- DLFl uoroscopy (FL) "on" (FL request)
- EXT-HWRadiation enable state/safety circuit
- SWRStart/inverter
SWR "start/inverter" is not illuminated as long as Kathotest
(module J05) is programmed.
The "radiation indicator" lights up only with radia tion ON.
• FL switch OFF
• SS switch ON
• FL ON
• Measure the minimum and maximum kV and mA values and enter them int o the
generator certificate.
Exposure release circuit3
With preparation:
• Set the release button S27 to preparati on .
• On D211, the following LEDs must light up:
- TU1/TU2TU1/TU2 (acc. to selected workstation/tube assembly)
- / (depending on focus selection)
- GEN-OKGenerator enable state
- ZBZB "ON" (ex posure preparation (ZB) request)
With exposure release:
• Press the release button S27 all the way down.
• On D211, the following LED’s must light up
- VH:Exposure request to unit
- AR:Exposure enable state from unit
- EXT-HW:Radiation enable state/safety circuit
UROSKOP D3Register 4RLL5-310.034.04Page 2 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
Generator functions3 - 3
NOTICE
SWR "start/inverter" is not illuminated as long as Kathotest
(module J05) is programmed.
The "radiation indicator" lights up only with radiation on.
• Release radiation release butt on S27.
Controls and indicator lamps for radiation protection3
At the customer’s request, the following controls and indica tors can be connected in the
power cabinet according to the installation instructions RLL5-310.031.03. .., page 4-6.
Door contact for radiation disable function3
(e.g. at the door of the examination room)
• Check:When the door is open, radiation release must not be possible.
Indicator lamps for tube assembly selection3
• Check:For fluoroscopy or exposure preparation, the respect ive indicator lamp
associated with the tube assembly must light up.
Checking the kV/mA values3
• System OFF.
• Connect oscilloscope to test points MP.kV
• System ON.
• SS switch ON.
• On the control console, sel ect 77 kV, 32 mAs for and focus respectively.
• Release one exposure in each case and check t he kV/mA values of test points
MP.kV
and MP.mA
act
with the oscilloscope.
act
• Enter the measured kV values in to the test generator test certi ficate and compare them
with the values recorded at the fa ctory. They must be within the specifi ed tolerances.
and MP.mA
act
on service board D211.
act
Siemens AGRegister 4RLL5-310.034.04Page 3 of 4UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
3 - 4Generator functions
NOTICE
NOTICE
Checking the mAs values3
• System OFF.
• On PC board D220 (over H1), remove the jumper from the mAs jacks.
• Connect the mAs meter to the mAs jacks.
• System ON.
• On the control console, select , 77 kV, 80 mAs.
• Release an exposure and check the mAs value on the mAs meter .
• Enter the measured value into the gener ator test certificate.
Start-up of the automatic exposure control3
Check the functions of the IONTOMAT P and PLANI-IONTOMAT P automatic exposure
control systems according to the generator setting instructions RX63-050.032....
The installation procedures described i n the start -up inst ruct ions
may be disregarded since both subassemblies have alr eady been
installed at the factory.
§16
Since the UROSKOP has no selection for the film-screen keys
H-D-U, identical values must be programmed for H-D-U in the
IONTOMAT.
Checking the reproducibility of the automatic exposure control
• Attach 2.1 mm Cu to the multileaf colli mator.
• Set 77 kV,
100%,
large focus.
• Select Iontomat.
• Release 5 exposures consecutively.
• Enter the measured dose values into the generat or test certificate.
UROSKOP D3Register 4RLL5-310.034.04Page 4 of 4Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
Unit movements4
Preparation and limit values for tabletop lift4
• Connect the service PC to connec tor X10 of board D1.
• System (When the emergency stop switch is actuated,” U01” appears in the display).
• If necessary, release th e emergency stop switch.
• Service PC .
• Start the service progra m
(refer to the service soft ware instructions RLL5-310.113.02...) .
Main menu
Info database
4 - 1
Unit parameters
Unit state
Unit errors
Calibration
Options
Download
Unit parameters
Load from unit
Load from file
Show
Modify
Send to unit
Load from unit
Load from file
Tomography
Memoskop 2K/100
Write to file
Output
ceiling height
• For ceiling height enter the act ual room height available
• Press return.
• Press <F4>. Initial questions with Yes.
Siemens AGRegister 4RLL5-310.034.04Page 1 of 6UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
4 - 2Unit movements
CAUTION
NOTICE
Lifting movement
Rotating axis
Lifting movement4
Move into the end positions in gradual increments.
• Tilt the unit into the 0° posit ion.
• Successively move the unit into both end pos itions.
• Measure the distance between the floor or top surface of the installation pl ate and the
top surface of the tabletop or rota ting axis for both system positions :
Nominal distance between the floor or top surface of the installation plate
and top surface of the tabletopand rotating axis
H
u
82 cm ± 5 mm128 cm ± 5 mm62 cm ± 5 mm108 cm ± 5 mm
for I.I. 4086 cm
! minimum distance I.I.
Read the values displayed on the PC and compare them with the factory test values on
the test certificate.
± 5 mm128 cm± 5 mm66 cm ± 5 mm108 cm ± 5 mm
⇔ floor: 4 cm !
= 128 cm, possible only for room heights 2.60 m and above.
H
o
H
o
A
u
A
o
• Initial the comparison as conf irmation.
Lowest point: 62 cm to center of rot ation with tilted table (+88°).
• Correction: UROSKOP D adjustment instructions
UROSKOP D3Register 4RLL5-310.034.04Page 2 of 6Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
Unit movements4 - 3
CAUTION
Oblique projection4
Move into the end positions in gradual increments.
• Swivel the tube assembly support arm i n each case from the 0° position into bot h end
positions, -15° and +15°.
The 15° values must be indicated on the system cont rol console; at the same time, the
safety limit switch must not be pres sed.
• Read the values displayed on the PC under menu it em [
compare them with the factory test v alues in the test certifi cate.
• Initial the comparis on as confirmation.
• Correction: UROSKOP D adjustment instructi ons.
Unit state
], [Current] and
Siemens AGRegister 4RLL5-310.034.04Page 3 of 6UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
4 - 4Unit movements
CAUTION
Tilting movement4
Requirement: The installation plate must be level.
Move into the end positions in gradual increments.
• Tilt the unit in each case from bo th directions into the 0° positi on.
• Check the 0° position in each case with the p recision water level after the automati c stop.
Nominal: 0°
• Swivel from the 0° position in to the +88° end position; the switchi ng cam must not touch
the limit switch.
• Check the +88° end position with th e precision water level.
Nominal: 88°
• Swivel from the +88° end positio n into the -15° end position; the swit ching cam must not
touch the limit switch.
• Check the -15° end position with th e precision water level.
Nominal: - 15°
• Read the values displayed on the PC and compare them with t he factory test values in
the test certificate.
± 0.3°
± 0.5°
± 0.5°
• Initial the comparison as conf irmation.
• Correction: UROSKOP D adjustment instructions .
UROSKOP D3Register 4RLL5-310.034.04Page 4 of 6Siemens AG
Rev. 0412.98TD PS 24Medical En gineering
Unit movements4 - 5
CAUTION
Wall side
Physician‘s side
Head end
Longitudinal movement of the tabletop4
Move to the end positions in gradual increments.
• Move the tabletop successivel y into end positions A and K.
• Measure the distance between the l eft tabletop edge and the left front of the front guide
rail for both table pos itions.
Nominal:L
= 573 mm ± 5 mm
a
= 573 mm ± 5 mm
L
k
• Read off the values displayed on the PC and compare them with the factory test value s
of the test certificate.
• Initial the comparis on as confirmation.
• Correction: UROSKOP D adjustment instructi ons.
Siemens AGRegister 4RLL5-310.034.04Page 5 of 6UROSKOP D3
Medical EngineeringRev. 0412.98TD PS 24
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.