Siemens UROSKOP D2, UROSKOP D3 Update Instructions

Page 1
SIEMENS
UROSKOP Dl UROSKOP D2 UROSKOP D3
Service
Unit/Generator
Repair Instructions Removal and installation of the
spotfilm device
C
&!!!!A
Register 1
RLL5-310.091 .Ol .01.02
Replaces:
English
06.93
Weltergabe sowie Vervelfaltlgung dew Unterlage. Verwertung und Mlttellung lhres Inhalts mcht gestattet, soweit nocht ausdrirckllch zugestanden. Zuwiderhandlungen verpfhchten zu Schadenersatz. ~lle Rechte ftir den Fall der Patenterteilung oder GM-Eintragung vorbehalten
Copyright (C) Siemens AG 1993
Propretry data, company, confidential. All “ghts reserved. Conf4 d twe de wcret d’ entreprise Tous droits w&v& Conflado coma secrete Industrial. Nor resewamos todos lorderechor.
Page 2
2
Replacement of the spotfilm device

Contents

Cover page Contents, revision status
1
General
Safety note Documents required Measuring instruments required
2
3
Removal of the spotfilm device
Removal and installation of the installation plate
Page
1 2
l-l
l-l
l-l
2-l
3-1
Rev.
01 01
01 01 01
01
01
7
Installation of the spotfilm device
4-l
01
Siemens AG Medical Erqneerq Grolq
RLL5-310.091.01.01.02
Rev. 01
Page 3

General

Safety notes, documents and measuring instruments required

Safety note

l-l
Note! When performing work and tests, observe the product-specific
tes
contained in
of the TI folder.
Tests and settings which need to be made under radiation are marked with the radiation symbol ‘2 During all procedures marked with this symbol, radiation protective precautions must be observed strictly.
this document and the genera/ safety notes in register 2
safety no-

Documents required

Operating instructions, UROSKOP D... Wiring diagram

Measuring instruments required

Service tools
Slemens AG Medical Engineering Grolrp
RLL5-310.091.01.01.02
Rev.
01
Page 4
l-2
General
Siemens AG Medical Enqinrterlng Gro~rp
RLL5310.091.01.01.02
Rev. 01
Page 5

Removal of the spotfilm device

Fig. 1
2-1
l
Unscrew the four-part cover of the unit support on the bottom side of the unit support (rlFig.1).
l
Unscrew the M6 socket head cap screw under the Emergency Shutdown switch (PiFig.1).
l
Completely detach the cover of the swivel frame or of the fixed frame (liFig.2).
l
Set the adjustment disk and
Siemens AG Medical Engineering Grocq
the
L.,R lever to the center position (Fig.3).
RLL5-310.091.01.01.02
Fig. 3
Rev.
01
Page 6
Removal of the spotfilm device
Fig. 5
Support the weight of the image intensifier using a box that is 30 to 50 cm high and stable
Switch the system on and lower the unit until the 1.1. rests on a prepared stable base.
Switch the system off.
Detach the 1.1. (eight M8 socket head cap screws, /)/Fig.4). Switch the system on and lift up the table carefully. Cover the 1.1. screen to protect it from damage (e.g.: cassette). Tilt the unit to approx. 80”.
(In this position it will later be easier to remove and install the spotfilm device).
Move the table all the way to the front towards the operator. Switch the system off. Pull the cover (AIFig.5) of the spotfilm device approx.10 cm out. Undo the plug connection N16.Xl (unloading and Emergency Shutdown
switch) and jumper pin 1 and pin 2 at connector N16.Xl (liFig.5) (safety cir­cuit).
Pull the spotfilm device cover all the way out and set it down laterally.
Siemens AG Medical Engineering Gro~tp
RLL5310.091.01.01.02
Rev.
01
Page 7
Removal of the spotfilm device
2-3
Fig. 6
l
Switch the system on
l
Load a small cassette (18 x 24) ,
+ The clamping arms move together
and the upper fastening screws of the spotfilm device become accessible
l
Then actuate the
EMERGENCY SHUTDOWN switch
3 The clamping arms remain in their
present position so that the upper fastening screws of the spotfilm devi-
ce remain accessible even after the system has been switched off.
l
Switch the system off.
l
Pull out the following plugs of the spotfilm device cable harness (Fig.6): NlG.JKl, D50.X58, D50.X59.
l
Unscrew the protective ground wire of the spotfilm device cable harness.
0 Pull out the protective ground wire.
l
Loosen the bottom fastening screw of the spotfilm device (ISK M6, dIFig.7).
Sternens AG Medical Engmeermg Group
RLL5-310.091 .Ol .01.02
Rev. 01
Page 8
2-4
Removal of the spotfilm device
Fig. 8
/ -
l
Remove the spotfilm device top fastening screws (4 x M6 Allen screws Fig.8).
Fig. 9
l
If not already done, set the adjustment disk and the lever to the center positi­on (Fig.3).
l
Unscrew the bracket across the spotfilm device ( T jFig.9) at the rear on the right and press it upwards so that the adjustment disk can be passed through.
l
Carefully pull out the spotfilm device towards the front (2 people required!); In doing so, press the lever (Fig.3) against the adjustment disk, if necessary.
l
Set the spotfilm device down laterally on a suitable base or on its bottom side
(not on the motor!) (Fig. 10).
Siemens AG Medical Engineering Group
RLL5-310.091.01.01.02
Rev. 01
Page 9

Removal and installation of the installation plate

3-1
Removal
Fig.1 Fig.2
Prepare the spotfilm device;
The AM51 motor must face downwards
Pull out connectors X52 (encoder) and X53 on PC board D50 (Fig.1). Disconnect the 24V cable of the AM52 collimator plate motor on PC board
D50 (terminal X65) (plFig.1).
Unscrew the IONTOMAT connector NlG.JKl (Fig.2) on the bottom side of the spotfilm device and remove the IONTOMAT cable attached to the frame of the spotfilm device by means of a cable tie.
Remove the securing screws of the installation plate:
- 14 M4 Allen screws with washers (LIFig.2) One of the screws close to the AM52 motor is longer!
- 3 M4 socket head cap screws (LiFig.2).
Pull the previously disconnected cables out from behind PC board D50. Take off the installation plate;
In doing, so, carefully pull the disconnected cables and the AM52 motor
through the frame openings.
Sremens AG Medrcal Eqneering Groclp
RLL5310.091 .Ol .01.02
Rev.
01
Page 10
3-2
Removal and installation of the installation plate
Removal
Put the installation plate in place; In doing so, carefully feed the cables and the AM52 motor through the frame openings.
Fasten the screws of the installation plate
- 14 M4 Allen screws with washers (LIFig.3) One of the screws close to the AM52 motor is longer!
- 3 socket head cap screws (UFig.3).
Tighten the IONTOMAT connector NlG.JKl (Fig.2) back in place and fasten the IONTOMAT cable to the frame of the spotfilm device by means of cable ties.
Run the other cables behind PC board D50. Plug connectors X52 (encoder) and X53 on PC board D50 (Fig.2). Reconnect the 24V cable of the AM52 collimator plate motor on PC board
D.50 (terminal X65) ( - = blue, + = brown, 7iFig.l).
Siemens AG Medical Engineerq Group
RLL5-310.091 .01.01.02
Rev. 01
Page 11

Installation of the spotfilm device

l
Set the adjustment disk and
the L/R lever to the center po-
sition (Fig. 1).
4-1
l
Push the bracket across the spotfilm device at the rear on the right upwards ( 7 that the adjustment passed through.
l
Insert the spotfilm device into
the unit from the front.
Proceed very carefully and do not tilt the spotfilm device.
This procedure requires 2
persons.
l
Connect the NlG.JKl,
D50.X58, D50.X59 connectors and the ground cable (Fig.3).
l
Tighten the bottom fastening screw of the spotfilm device (M6 Allen screw, Fig.3).
Fig. 1
/Fig. 2) so
disk can be
Fig. 2
Siemens
Medical Engineering Group
AG
Fig. 3
RLL5310.091 .Ol .01.02
Rev.
01
Page 12
4-2
Fig. 4
Installation of the spotfilm device
Switch the unit on Load a small cassette (18 x 24),
+ the clamping arms move
together
Then actuate the EMERGENCY SHUTDOWN switch
+ the upper four fastening screws
of the spotfilm device become accessible.
Switch the system off.
Tighten the upper fastening screws of the spotfitm device (four M6 Allen screws, Fig.4).
Insert the cover (AjFig.5) of the spot­film device up to approx. 10 cm short of the stop.
____
i
Fig. 5
Remove the jumper at connector
N16.Xl
Make the plug connection N16.Xl (to the unloading and Emergency Shut­down switch) (VFig.5) .
Completely insert the cover of the
spotfilm device (AiFig.5).
Tighten the bracket (Fig.2) back on
again on the right back side. Attach the covers of the swivel or the
fixed frame. Take the protective plate off the 1.1.
input screen. Release the
Emergency SHUTDOWN Switch. Switch the system on and tilt the unit
into the 0” position.
(LIFig.5).
Stemens AG Medc4 Engineering Grotrp
Carefully lower the unit until it rests
on the 1.1. Switch the system off.
RLL5310.091.01.01.02
Rev.
01
Page 13
Installation of the spotfilm device
4-3
Fig. 7
l
Attach the 1.1. (using eight M8 socket head cap screws, Fig.7).
l
Reinstall the four-part cover of the unit support on the bottom side of the unit (Fig.8).
l
Tighten the M6 socket head cap screw under the Emergency Shutdown
switch (LIFig.8).
l
Switch the system on.
l
Lift the table and remove the box supporting the weight of the image intensi­fier.
l
Check the functions of the spotfifm device.
l
Switch the system off.
Fig. 8
Siemens AG
Medical Encpeerq Grot~p
TDRL2IWendt TDRL3Sauter TDRLlSchlee TDU 6Steinbrink
RLL5-310.091.01.01.02
Rev. 01