RLL5-310.076.03.01.02
UROSKOP D2 / UROSKOP D3
Replacmg:
Polydoros SX
RX-speed-Info-No.:
Service information
In the Uroskop D2/D3, only one ring dominant is available as ADR measuring field. If in the
event of service the sensitivity is readjusted, incorrect measurement results occur in the
subsequent check of the dose rate.
The cause of this is that the ring dominant is larger in the 1.1. full format than the dose
measuring chamber.
The dose rate can be set as described in the UROSKOP D start-up instructions.
Check the dose rate as follows:
Deconfigure spotfilm device according UROSKOP D
adjustments instructions
Configure FC-Curve 00 (antiisowatt curve) for Automatic 1
Select 1.1. full format and Automatic 1
Attach 2.1 mm Cu to the collimator
FLUORO ON
Let the dose rate settle down
Push Automatic STOP button on the unit console
FLUORO OFF
Bring dose measuring chamber into the beam path centrally and close to
the 1.1.
FLUORO ON
Check the dose rate set in the Vidomatic P configuration
FLUORO OFF
Perform measurement for all 1.1. formats in the same way
Configure spotfilm device again
TDRX 4 / Seef