UROSKOP D
Maintenance Instructions
SP
The maintenance protocol
RLL5-310.105.02.02.02
is required for this instructions
English
Print No.: RLL5-310.101.02.02.02 Doc. Gen. Date: 01.97
Replaces: RLL5-310.101.02.01.02
© Siemens AG 1993
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
0 - 2Revision
Chapter Page Revision
01 - 402
11 - 402
2 1 - 16 02
31 - 202
UROSKOP D RLL5-310.101.02 Page 2 of 4 Siemens AG
Rev. 02 01.97 TDSP 1 Medical Engineering
Contents 0 - 3
Page
1 _______General information ____________________________________________1 - 1
Required documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Required tools, measurement and auxiliary equipment. . . . . . . . . . . . . . . . . 1 - 1
Required lubricant. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Text emphasis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Safety instructions and protective measures . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Explanation of abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
2 _______Inspection and maintenance _____________________________________2 - 1
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Safety inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
System base. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Lifting drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Tilt drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Rotating column drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Longitudinal table drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Transverse movement drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
System general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Safety devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Checking radiation protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Periodic maintenance / parts replacement . . . . . . . . . . . . . . . . . . . . . . . 2 - 7
Inspection for wear and tear. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 8
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Vertical drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Tilt drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Rotation drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Longitudinal table drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Spotfilm device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
System general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Inspecting the internal heat dissipation . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Cabinets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Operating values - inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Lifting base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Spotfilm device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Collimator blades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Cassette drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Function inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Component quality inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Concluding steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 16
3 _______Changes to previous version_____________________________________3 - 1
Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D
Medical Engineering Rev. 02 01.97 TDSP 1
0 - 4 Contents
This page intentionally left blank.
UROSKOP D RLL5-310.101.02 Page 4 of 4 Siemens AG
Rev. 02 01.97 TDSP 1 Medical Engineering
General information 1 - 1
1 General information
The sequence for complete inspection and maintenance is
described on the following pages.
The safety inspection may be performed alone; however, the
operating values check must be performed in conjunction with
the safety inspection.
1.1 Required documentation
depending on the system)
• UROSKOP D1 circuit diagram G5353
• UROSKOP D2 circuit diagram G5354
• UROSKOP D3 circuit diagram G5403
• Adjustment instructions RLL5-310.071.01.02.01
• Uroskop D1/D2 start-up instr uctions RLL5-310.034.01.04.01
• Uroskop D3 start-up instr uctions RLL5-310.034.02.03.01
• Maintenance protocol RLL5-310.105.02.02.01
1.2 Required tools, measurement and auxiliary equipment
• Protective conductor meter 44 15 899 RV090
• DVM e.g. Fluke 8060 84 27 684 RE999
• T or que wrench, to 100 Nm
• Spring scale, to 2.5 N
• Spring scale, to 100 N
• Standard service equipment
1.3 Required lubricant
• Grease gun for Longtime PD2 71 09 010
• Longtime PD2 - grease 20 g tube 34 91 271
1 kg can 73 95 445
• Optimol Optiflex ELM paste 1 kg can 15 90 293
• Silicon paste 100 g tube 20 49 914
Siemens AG RLL5-310.101.02 Page 1 of 4 UROSKOP D
Medical Engineering Rev. 02 01.97 TDSP 1
1 - 2 General information
1.4 Text emphasis
"WARNINGs" are information provided with special emphasis
when there is the potential for injury to operator or patients.
"CAUTIONs" are information provided with special emphasis
when there is the potential for damage to the equipment.
"NOTICEs" are information provided with special emphasis to
facilitate proper use of the equipment or proper execution of
a procedure.
1.5 Safety instructions and protective measures
• When performing service work and tests adhere to:
- the product-specific safety informati on in the document,
- the safety instructions in RA0-000.012. 40... in the UROSKOP D
- Logbook, as well as
- the general safety information contai ned in the TI folder /
ARTD Part 2 (CD-ROM) .
• T ests and adjustments thamust be performed with r adiation
ON, are identified by the radiation warning symbol .
Radiation protection must be worn during these types of
adjustments.
UROSKOP D RLL5-310.101.02 Page 2 of 4 Siemens AG
Rev. 02 01.97 TDSP 1 Medical Engineering
General information 1 - 3
1.6 Explanation of abbreviations
Abbrev. Explanation
SI Safety Inspection
SIE Electrical Sa fe t y Inspection
SIM Mechanical Safety Inspection
PM Preventive Maintenance
PMP Preventive Maintenance, Preventive Parts Replacement, Visual Inspection, etc.
PMA Preventive Maintenance Adjustments
PMF Preventive Maintenance, Function Check, Operating Value Check
Q Quality Check
QIQ Image Quality Check
QSQ System Quality Check
SW Software Maintenance
The items identified by these abbreviations are contained in the maintenance certificate
and should be checked off upon completion.
Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D
Medical Engineering Rev. 02 01.97 TDSP 1
1 - 4 General information
This page intentionally left blank.
UROSKOP D RLL5-310.101.02 Page 4 of 4 Siemens AG
Rev. 02 01.97 TDSP 1 Medical Engineering