Siemens Uroskop D Service Instructions

UROSKOP D
Service Instructions
SP
Standard table pads for metal table extensions
© Siemens AG 1997
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: RLL5-310.076.08.01.02 Doc. Gen. Date: 02.97
Replaces: n.a.
2
NOTICE
Standard table pads1

General 1

Table pads, Item nos. 3750 366 (27 cm) and 3750 374 (83 cm), can be attached to the
older tabletop extensions (metal construction ) using Velcro pads.
Table pads can be attached to the carbon fiber table extensions Item nos. 1676 910 (83 cm) and 1676 928 (27 cm) as well as metal table extensions using this procedure.

Materi al 1

4 pieces (fuzzy) Velcro pads, I tem no. 3838 344 per table pad

Service equipment 1

Measuring tape.

Procedure 1

To determine the position for the adhesive:
- Move the table in the horizontal posi tion and attach the tabletop extension.
- Center the pads on the table extensi on.
- Mark the locations for the adhesi ve that correspond to the Velcro st rips attached to the underside of the tablepad (refer t o the drawing).
Clean the positions where adhesive will be appli ed.
Remove the protective backing from the (fuzzy) Velcro pads and attach them to the
marked positions on the upper side of the table extension pressing them down fir mly.
(fuzzy) Velcro pads 60 x 50 mm
tabletop extension 27 cm
(fuzzy) Velcro pads 60 x 50 mm
tabletop extension 83 cm
top view
TD SP 2 / Fleischter SMS Iselin / O’Donnell
UROSKOP D RLL5-310.076.08 Page 2 of 2 Siemens AG Rev. 01 02.97 TD SP 1 Medical Engineering