UROSKOP D2/D3 (SX)
CS
Update Instructions
Title:
Reason for update: Safety
Urgency: Immediate Within 3 months
Update materials required? Yes No
Mater ials free of ch arge? Yes No
Return of parts? Yes No
Estimated completion time: 2.5 h Number of CSE’ s: 1
Customer application training? Yes No
Systems/Products affected/System identifying IVK
Name Material No. Serial No.
Software Update
X
X
x
x
SP013/02/S
x
x
UROSKOP D2/D3 (SX) 97 85 015 G5354 1050, 1053 - 1186
Remar k : n. a.
Components affected/to be modified/IVK
Name Material No. Serial No. Component status
affected
n.a. n.a. n.a. n.a.
Remar k : n. a.
© Siemens AG 2002
Chg. Ref. No.: 112 559
Name: Erlwein
Dept. : CS PS 24
Print No.: R LL5-310.896.0 5. 01.02 Doc. Gen. Date: 10.02
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model _or_ design,_are_
reserved.
Replaces: n.a. Version 2.0; August 1, 2002
Page 1 of 10
2 Update Instructions SP013/02/S
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authoriz ed by Siemens o r o ne of its af f iliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products Affected 0
UROSKOP D2 / D3 with POLYDOROS SX.
Reason for the Update 0
With UROSKOP D systems, unforeseen, unco ntro l led system movements or unwanted
automatic movements of the tabletop may occur e.g. when raising the unit out of its horizontal position.
Prerequisites 0
In order to be able to make the changes on boards, the servicetechnician must have the
knowledge and experience required for soldering on boards. If this is missing, a changed
board must be ord ere d.
Special Tools / Documents 0
Service PC with WIN 95/98 operating system.
Serial PC connection cable 99 00 440.
UROSKOP D2/D3 (SX) RLL5-310.896.05. Page 2 of 10 Siemens AG
Rev. 01 10.02 CS PS 24 Medical Solutions
Update Instructions SP013/02/S 3
Ordering Information 0
The following update kit has to be ordered from CSML (SAP Distribution Channel, factory
2050) or BU Logistics (factory BU) (only with an order via BU Logistics):
Update Kit 11 56 624
List the system Serial Number on the order!
NOTE
If a second
firmware 11 56 558 must be ordered in addition.
HBF (tableside control) is installed on the system,
Contents of the Update Kit 0
Update kit 11 56 624 contains the following parts:
Pos. Quan. Part No. Name
1 1 16 61 300 H ost Software VA00F
2 1 16 28 523 Firmware ABF VA00C
3 1 16 61 326 Firmware BBC VA00C
4 1 11 56 558 Firmware HBF VA00D
5 74 62 901 Generator Host Software Update VE00D
6 2 Connecting cables
70.5 m Wire
8 0.5 m Solder tin
9 1 Technical document
Tab. 1
Return of Parts 0
n.a.
Siemens AG RLL5-310.896.05. Page 3 of 10 UROSKOP D2/D3 (SX)
Medical Solutions Rev. 01 10.02 CS PS 24