Siemens UROSKOP Access User manual

Page 1
Camera Installation
Product: Uroskop Access
Title: Specific Hardcopy Camera Information for
Kodak DryView 8200 / PACS Link MIM 200
Affected systems: Image System ASPIA
Replaces: n.a.
1.1 Block Diagram
SCU / Modality
Uroskop Access
Network
1.2 Available Film Sizes
Application ASPIA Film Type Medium Type
14"x17"
14"x14"
11"x14"
8"x10"
others
Released : yes - Film size is available and released Released : no - Film size is available but not released Released : n.a. - Film size is not available
Portrait yes
Landscape yes
Portrait yes
Landscape yes
Portrait yes
Landscape yes
Portrait
Landscape
n.a.
n.a.
n.a. n.a. n.a.
SCP
Kodak
MIM 200
Kodak DVB blue based
Kodak DVB blue based
Kodak DVB blue based
n.a. n.a.
DryView 8200
HCC
Kodak
Siemens AG, Medical Solutions PrintNo.: SPL5-330.814.17.01.02 © Siemens AG, 2002 CS SD 24 12.2002 Page 1 of 5
Page 2
Uroskop Access Camera Installation
Specific Kodak DryView 8200 / PACS Link MIM 200 Information
1.3 Restrictions and known Effects
For HCC-independent restrictions, see also the general restrictions in the document Uroskop Access, General Hardcopy Information.
2 Modality SW Configurations / Parameters
2.1 Configuration
Refer to: Uroskop Access, General Hardcopy Camera Information
Application Entity Properties Display Changes allowed yes/no AE Title Port Number Supported DICOM Services
HC Device / General Settings Display Changes allowed Type Class (Printer spec file name) DICOM Node Filming Properties Display Changes allowed Hold printed film jobs Min. density Pixel Size [1/1000 mm]
Film Sheet Formats Portrait Number of pixels
Film Sheet Formats Landscape Number of pixels
Medium Type Film Destination Color appearance Background Transformation
Note: Do not modify and/or add film sizes.
Do not modify pixel sizes. Delete film sizes that are not available at customer site. (*1) K8200 or dv8200 does not support NEVENT REPORT; The verification of the AE-Title works properly. NER_8200 supports NEVENT REPORT. The verification of the AE-Title failed.
10
0.20
78.0 x 78.0
[columns x rows]
[columns x rows]
blue film No processor No monochrome black Yes nomagnification
K8200 (*1)
5040
Print
DICOM printer
Inches 11x14
Inches 14x14
Inches 14x17
Inches 11x14
Inches 14x14
Inches 14x17
kodak8200
dv8200
3272x4360
4360x4244
4360x5220
4360x3272
4244x4360
5220x4360
In "define new" Do not modify/add Select only Delete only
In "define new" Do not modify/add Select only Delete only
Yes
No No
No No
Yes
Yes Yes
No
No
No
Siemens AG, Medical Solutions PrintNo.: SPL5-330.814.17.01.02 © Siemens AG, 2002 CS SD 24 12.2002 Page 2 of 5
Page 3
Uroskop Access Camera Installation
Specific Kodak DryView 8200 / PACS Link MIM 200 Information
2.2 Error and Warning Messages
Printer Status info Description Message string visible
in the Status Bar
PRINTER -BUSY No description available Printer busy <None>/idle CHECK CHEMICALS No description available Check chemicals <None>/idle EMPTY 4X6 No description available Empty 4x6 <None>/idle EMPTY 4X6 PAPR No description available Empty 4x6 papr <None>/idle EMPTY 85X11 No description available Empty 85x11 <None>/idle EMPTY 85X11 PAPR No description available Empty 85x11 papr <None>/idle EMPTY A4 No description available Empty a4 <None>/idle LOW 4X6 No description available Low 4x6 <None>/idle LOW 4X6 PAPR No description available Low 4x6 papr <None>/idle LOW 85X11 No description available Low 85x11 <None>/idle EMPTY 85X11 PAPR No description available Low 85x11 papr <None>/idle LOW A4 No description available Low A4 <None>/idle
Other action for syngo/camera symbol
3 Hardcopy Camera, Configuration
3.1 Identification
Device Name:
Manufacturer:
Device-Type:
SW Revision:
Image Memory:
Harddisk:
Physical Network Interface:
MIM 200 DryView 8200 Kodak Kodak PACS Link DRY- Laserimager V5.0 V3.1 Minimum 64MB None Minimum 2 GB None 100/10Base-T Ethernet, UTP/STP None
3.2 SCP / Printer Settings (provided by OEM)
file name content remarks
No file available.
Siemens AG, Medical Solutions PrintNo.: SPL5-330.814.17.01.02 © Siemens AG, 2002 CS SD 24 12.2002 Page 3 of 5
MIM200 Setup:
on Configure, Network: Node Name=MIM, IPAddr=accordingly Subnet Mask=accordingly, Gateway Addr=accordingly on Configure, Input, Network, Add:
Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
General: Log.Name=8200, IPAddr=100.100.100.40,
General: Database Name=8200, Model=8200, Memory=32
Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
Supported Item: Collation=On, Common Text=On, Tonescaling=On
Supported Item: Remote Image Deletion=On, Rotation=On
Interpolation: No Magnification=On, Bilin=On, Repl=On
Keypad configuration
Page 4
Uroskop Access Camera Installation
Specific Kodak DryView 8200 / PACS Link MIM 200 Information
Interpolation: Cubic Spline=On, Cubic Convolution 2=On
Format Types: Custom=On, Row Symmetric=On
Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
Formatting: DICOM Annotation Box=On, Req.Size Scale=On
Formatting: Req.Size Mix in Row=On, Trim=On
Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
Formatting: Req.Image Size=On, Req.Size Mix on Page=On
Formatting: True Landscape=On
Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
Formatting: Trim Width Enabled=2, Trim Width Disabled=0
Formatting: Pixel Pitch=12.8205
Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
SCU: Density Space= On, Variable Border=On, PLUT Service= On
SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
on Configure, SCP Services, select the specific modality, Modify:
PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
TFT/ULUT= Wrkstn2a.w87 (Always=Y), Contrast (dry)=3 (Always=Y)
Image Dmin=0.00 (Always=N), Image Dmax=2.80 (Always=N)
ICCProfile=Off (Always=N), Modality=OT (Always=N)
BodyPart=Default (Always=N), Modality Group=6
NeventReports=Off (Always=Y), Warning Status=On
Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
Modality ID=Off, Annotation Override=Off (Always=N)
Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
Conv.Group=1, Gamma Table=1, Smoothing Type=0 (Always=N)
Border=Black (Always=N), Film Base=Blue Film (Always=N)
Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
8200 Setup on CONFIGURATION Menu:
on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87 on LOCAL PANEL Menu: Contrast=6, Density=3.00 on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025
Maximum number of associations supported: 8
Siemens AG, Medical Solutions PrintNo.: SPL5-330.814.17.01.02 © Siemens AG, 2002 CS SD 24 12.2002 Page 4 of 5
Page 5
Uroskop Access Camera Installation
Specific Kodak DryView 8200 / PACS Link MIM 200 Information
4 Image Quality and Optical Density
4.1 Reference Film and LUT
Refer to: Uroskop Access, General Hardcopy Camera Information
4.2 Real Size Scaling
n.a.
5 Abbreviations
DCS DICOM Conformance Statement DICOM Digital Imaging and Communication in Medicine HW / SW Hardware / Software HCC Hardcopy Camera LUT Look Up Table OEM Original Equipment Manufacturer SCP Service Class Provider SCU Service Class User TCP/IP Transfer Control Protocol / Internet Protocol
6 Changes to Previous Version
n.a.
Siemens AG, Medical Solutions PrintNo.: SPL5-330.814.17.01.02 © Siemens AG, 2002 CS SD 24 12.2002 Page 5 of 5