Siemens SX 50 80 User manual

POLYDOROS SX 50/80
Start-up Instructions
SIRECON 2 Connection
RX
© Siemens AG 1996
The reproduction, transmission or use of th is do cu ment or its con ten ts is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 4 English Print No.: RX63-050.034.12.01.02 Doc. generation Date: 05.96
Replaces: n.a.
Chapter 0Revision
Page 2 of 4
Chapter Page Rev .
01 to 201 11 to 401 21 to 601 3 1 to 22 01 4 1 to 10 01 51 to 201 61 to 401
Letzte Seite des Kapitel
POLYDOROS SX 50/80 Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
Contents Chapter 0
Page 3 of 4
Page
1 _______Prerequisites __________________________________________________1 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Required Tools and Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Safety-technical Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Remarks Regarding Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Image Quality in Fluoroscopy Systems . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
2 _______Preparations for dose rate setting_________________________________2 - 1
Notes on the VIDOMATIC P. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Suppression of light distributor (220 V, not for I.I.40) . . . . . . . . . . . . . . . . . . 2 - 1
Fluoroscopic dose rate
(Excerpt from "Test Certificate for Fluoro and DFR/DR Systems") . . . . . . . . . . . 2 - 2
3 _______Startup _______________________________________________________3 - 1
Programming the POLYMATIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Fluoroscopy Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Additional Information :. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Excerpt from the "Reference Value List for Fluoro and DFR/DR Systems":. . . . . . . 3 - 3
Fluoroscopy dose rate at the I.I. input. . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
DR Programming (if Configured) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Excerpt from the Reference Value List for "Fluoro and DFR/DR Systems" :. . . . . 3 - 4
SIRCAM Programming (if Configured) . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
Excerpt from the Reference Value List for "Fluoro and DFR/DR Systems: . . . . . 3 - 5
Configuration" / "IONTOMAT..." . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
"Data" / "Restore" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
IONTOMAT P: Checking Drift and Hum (Ripple) duri ng F luor oscopy. . . . . . . . . . 3 - 7
Adjusting the Dose Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Preadjustment - "PDA Preadjust" . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 10
Carrying Out the Sensitivity Adjustment . . . . . . . . . . . . . . . . . . . . . . 3 - 11
Checking the Dose Rate for Fluoroscopy . . . . . . . . . . . . . . . . . . . . . . . 3 - 13
Preparations (Test Setup) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 13
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 14
Checking the Dose/Image for SIRCAM . . . . . . . . . . . . . . . . . . . . . . . . 3 - 15
Setting the Maximum Skin Dose Rate. . . . . . . . . . . . . . . . . . . . . . . . . 3 - 16
Preparations (Test Setup) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 16
Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 16
Final Check of the Skin Dose Rate . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 17
Calibrating the Acquisition (Poly-I.I. Adaption) . . . . . . . . . . . . . . . . . . . . 3 - 18
Preparations (Test Conditions) . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX 50/80 Medical Engineering Rev. 01 05.96
Chapter 0Contents
Page 4 of 4
Page
Adjusting the TV Iris Diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 19
Adjusting the TV Iris Diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Preparations (Test Setup) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 21
Checking the B-Signal Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 21
Preparations (Test Setup) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 21
4 ______ Fluoro and Exposures Curves ____________________________________4 - 1
Programming the Fluoro Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Generating Your Own Fluoro Curves (Brief Description). . . . . . . . . . . . . . . . 4 - 4
Final Recording of Fluoroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Preparations (Test Setup) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Recording. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Exposure curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
General Remarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
Generating Your Own Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 8
Allocating the Exposure Curves to the Organ Keys . . . . . . . . . . . . . . . . . . 4 - 9
5 ______ Final works____________________________________________________5 - 1
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Important: Saving the Programming . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
6 ______ Appendix, SDM Service__________________________________________6 - 1
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Checking the PDA Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
PDA measured at SIMOMED H - 44 cm . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Adjusting the Size of the PDA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Checking the Selected ADC Measuring Field . . . . . . . . . . . . . . . . . . . 6 - 2
Checking the Superimposed ADC Measuring Field with FLUROSPOT H . . . . . . . 6 - 3
Measurement field sizes in zoom mode (RX unit with 40-4 HDR I.I.) . . . . . . . . . 6 - 4
Letzte Seite des Kapitel 0
POLYDOROS SX 50/80 Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering

Prerequisites 1

General Remarks 1

This document is applicable for the followi ng generators:
POLYDOROS SX 50/80
The SIRECON 2 connection is an expansion to the basic generator (see generator instal­lation instructions) and is comprised of the D270 board, the D271 assembly, the multiplier extension cable as well as the connections parts and installation materials.

Required Documents 1

TV-I.I. A Connection Wiring Diagram: X1768
POLYDOROS SX Wiring Diagram: X2075
POLYDOROS SX Start-up Instructions: RX63-050.034.1 0...
For SIRCAM:
SIRCAM 103 Wiring Diagram G5129 or
SIRCAM 106 Wiring Diagram G5091
Chapter 1
Page 1 of 4
The following documents are need ed to car ry out t he image qualit y acceptance t est for Fluoro and DFR/DR systems:
Measuring Instructions RXD0-000.074.01.
T est Certificate RXD0-000.037.01.
List of nominal values RXD0-000.075.01.
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Chapter 1Prerequisites
CAUTION
NOTICE
CAUTION
Page 2 of 4

Required Tools and Test Equipment 1

- Standard installation tool ki t
- Service PC per ARTD, Parts 1, 3, 1.0
- PC connection cable, 5 m 99 00 440 RE999
- 2-channel memory oscilloscope, TEKTRONIX 2232 97 02 234 Y3155
- Precision copper filter, 2.1 mm 99 00 598 XE999
- Cu filters, set 44 06 120 RV090
- Resolution test, Type 41/42 28 71 820 RE999
µm copper strips 11 67 662 G5247
- 17
- Digital multimeter, FLUKE 8060A 97 02 101 Y4290
- mAs meter 81 60 400 RE999
- kV/dose meter, PTW-NOMEX 97 08 637 Y0388
- Ground wire test instrument 44 15 899 RV090
When using the oscilloscope, the ground wire in the power cable may not be disconnected under any circumstances. For measure­ments in which any ground loops may be present which may cause the measurement result to be adversely affected, use the TEK isolating amplifier and the triggering option.

Safety-technical Information 1

The "Safety-technical Informati on" contained in generator ins tallation instru ctions must be observed.
Checks or adjustments which must be carried out under radi ation are labeled in the
instructions with the radiat ion warning symbol . While these work steps are being carried out, observe the appropr iate radiation protection measur es (see TI 216)

Remarks Regarding Start-up 1

With systems which have been preassembled and tested at the fact ory per the BZ (can
be seen from the accompanying "Test Certi ficate, Jumper Protocol” in the Regist er labeled ”Protocol”), only ver y brief checks and adjustments need to be ca rried out.
Enter the results from the measurements labeled wit h ” ” in the Test Certificate in the
column marked ”Start-up Measurement Value” .
POLYDOROS SX Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
Switch on the dose meter at least 15 minutes prior to the meas­urement and then calibrate it.
No additional filter may be selected at the collimator.
Prerequisites Chapter 1
Page 3 of 4

Image Quality in Fluoroscopy Systems 1

Prerequisites and General Remarks Regarding Fill ing Out the IQ Test Certificate for Fluoroscopy Systems:
Beam geometry (collimation and centering) of the applicat ion equipment must be in order. When filling out the "Test Certificate / Fluoro and DFR/DR Systems",
RXD0-000.037.01... observe the foll owing:
In Chapter 1, enter the Setting Instruct ions containing the data for f luoroscopy in the IQ
tests used for start-up .
In Chapter 1.1, enter the customer data and the BZ Number.
In Chapter 1.2, enter the configured co mponents along with their designation. Part No.
and Serial No.
In Chapter 1.3, list the test equipment used al ong with designation and Serial No. and
the date until which the calib ration is still valid.
In Chapter 2, enter the data from the configur ed image intensifier.
After each measurement, enter the results i n the "Test Certificate/Fluoro and DFR/DR-
Systems".
Compare the measurement results with the reference val ues.
If tolerances are exceeded, determine the r eason and the error or correct the incor rect
setting before continuing with the chec ks. Otherwise it is possible that subsequen t measurements will be adversely infl uenced.
Enter the BZ number and the data on each page of the "Test Certi ficate/Fluoro and DFR/
DR Systems" to assure that there will be cl ear organization of the certific ate pages.
There may not be any empty fields in the complete d certificate. Enter n.a. (not
applicable) in fields in whic h no entry can be made. Fields which can only be f illed in during customer start-up are mark ed in the certificate as such. Pencil may not be used to fill out the cer tificate; do not use Tippex to make corr ections.
Record the start-up instructions whi ch are used for customer start-up. These cont ain
information pertaining to perf ormance of IQ texts and must be recorded in t he "Test Certificate/Fluoro and DFR/DR Systems" in Chapter 1.
The tester at the factory and the tester ca rrying out the start-up must si gn and date the
"Test Certificate/Fluoro and DFR/DR Systems" in Chapter 1-1.
Cross out incorrect entries and wri te the correct value next to it al ong with the date and
the tester’s initials.
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Chapter 1Prerequisites
Page 4 of 4
At the Conclusion of the Test:
Leave the "Test Certificate/Fluoro and DFR/DR Syst ems" RXD0-000.037.01... with the
system and file it in Register 9 of the Sys tem Binder. If the System Binder is not availabl e, file the Test Certificat e in Register 9 of the Generator Binder.
POLYDOROS SX Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
NOTICE

Preparations for dose rate setti ng 2

Notes on the VIDOMATIC P 2

For KK version systems, 2 fluoroscopy characte ristic curves can be programmed
(Fluoro 1 and Fluoro 2).
For systems in RAS version, up to 4 fluoroscopic characteristic curves can be
programmed if they can be selected from the uni t (Fluoro 1. . . Fluoro 4).
Dose-saving zoom can be programmed with existing iris di aphragm.
Customer-specific fluoroscopy curves can be generated within the range of C80...C99.
Normal fluoroscopy (max. 450 W)
Can be selected from fluoroscopy characteri stic curves in the firmware (C00, C01, C02,
C03, C10, C11)
High-contrast fluoroscopy (max. 1200 W, with modified high-volta g e generator, as from Serial No. 01056, max. 2000 W)
Can be selected from fluoroscopy characteri stic curves in the firmware (C40, C41, C42,
C50, C60, C61, C62, C63, C64, C65)
Therapy simulation
Chapter 2
Page 1 of 4
It is possible to program 2 fluoroscopy c urves which can be selected from:
- fluoroscopy characteristi c curves in the firmware (C40, C41, C42, C50, C60, C61, C62, C63, C64, C65)
Suppression of light distributor (220 V, not for I.I.40) 2
Voltage peaks caused by switching off the motor voltage of the mirror motor in the light distributor result in generator interfer­ences
For the reason, generally incorporate two RC element s, Part No.31 33 147 B2003
parallel to the motor.
For 220V mirror motor voltage, reconnect jumper x7 on board D271
M 8.7
M
M 8.8
100
0.22µF
x7
3
24V mirror (40cm I.I.)
2
220V mirror
1
M 8.3
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Chapter 2Preparations for dose rate setting
Page 2 of 4
Fluoroscopic dose rate (Excerpt from "Test Certificate for Fluoro and DFR/DR Systems")
Basic info on dose/dose rate measurement on image intensifier workstations For dose measurement at the I.I. input (K
( ), attach the dose measuring chamber centrally to the radiation field, as close as pos-
K
B
sible to the I.I.; in the case of spotfil m devices, place it in the cassette shaft. If this is not possible, measure the dose (K
) or the dose rate ( ) in front of the attenuating layers:
Tg
- With OT units on the tabletop
- With UT units on the spotfilm device rear panel
- With C-arm units on the grid
Using the correction factor m (unit attenuation factor), the value at the I.I. input is calcu­lated.
The unit attenuation factor contains all attenuating layers between the patient and the image receptor , inc luding t he distan ce fac tor; i t is i ndic ated in th e unit tes t certif icat es and can also be calculated by multiplication of the attenuation factors of all individual layers.
The radiation attenuation of 6% resulting from the dose measuring chamber must be taken into account as well (factor 0.94).
Determination of the dose/dose rate directly at the I.I. input
) or dose rate measurement at the I.I. input
B
K
Tg
2
Dose Dose rate
Meas. chamber in the I.I. plane K
Meas. chamber in front of K
=KBg x 0,94 = x 0,94
B
x 0,94 x 0,94
K
TG
==
B
K
B
K
B
K
Bg
K
Tg
attenuating layers mm
Special case 1: Dose rate measurement if the ADC measuring field is larger than the dose measuring chamber:
Allow the ADC to adjust with 2.1 mm Cu prefi ltration, chan ge over to ST OP , now place the dose measuring chamber in the beam path and measure the dose rate. The factor of
0.94 must not be taken into account in this case.
Special case 2: Dose measurement if the dominant is larger than the dose measur­ing chamber:
Place 2.1 mm Cu and the dose measuring chamber in the beam path, release an expo­sure and measure the dose (KBg or KTg) and the mAs value (Qg1). Remove the dosime­ter from the beam path, release an exposure and measure the mAs value (Qg2). Calculate the dose at the I.I. input (KB)
Q
KB=K
POLYDOROS SX Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
x ;KB= x
Bg
Q
g2 g1
K m
Tg
Q Q
g2 g1
Preparations for dose rate setti ng Chapter 2
s
Page 3 of 4
Determining the unit attenuation factor from the indi vidual attenuation factors
m
unit
attenuation
factor
Attenuation factor of tabletop, spotfilm device rear panel etc. m Attenuation factor of scattered radiation grid m Attenuation factor of Iontomat chamber m
2
r
Attenuation factor distance factor mA =
B
r
T
Unit attenuation factor m = mT x mR x m? x mA . . . . . . .
Cu filter
r
B
r
T
K
Tg
Measuring chamber in front of attenuating layer Tabletop or spotfilm device rear panel Grid Iontomat chamber
K
Bg
K
B
Measuring chamber in I.I. plane (cassette shaft) I.I. input
T R I
/ = Dos e /dose rate directly at the I.I. input
K
K
B
B
/ = Measured dose/dose rate in the I.I. plane
K
K
Bg
Bg
/ = Measured dose/dose rate in front of the attenuating layers
K
K
Tg
Tg
r
B
r
T
= Source-image receptor distan ce = Distance between source and beginning of the attenuati ng layers
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Chapter 2Preparations for dose rate setting
Page 4 of 4
Formulas for calculating the dose/dose rate in the measuring plane for the dose/ dose rate setting
During setting, the dose or dose rate required directly at the I.I. input must be converted into the measuring plane (I.I. plane or in front of the attenuating layers). For this, use the modified fo rmulas. .
Dose
K
Meas.chamb. in I.I.plane KBg==
B
0.94
Dose rate
K
B
K
Bg
0.94 Meas.chamber in front of attenuating layers K
K
==
Tg
B
0.94
K
Tg
K
B
0.94
x m
x m
Remark: The radiation attenuation resulting from the dose measuring chamber is already
taken into account during setting
(for adjustment: remove the dose measuring chamber from the beam path). Do not use the f actor of 0.94 for calculating the dose r ate to be set.If the dose/ dose rate is checked (not set), the factor of 0.94 must be taken into account.
Example for determining the correction factor
Correction factor m (total) = (source-I.I. distance/source-measuring chamber distance)
2
x m-factor (Iontomat camber) x m-factor (Raster) x m-factor (tabletop etc., if between dose measuring chamber and I.I.) etc.
- Distance correction factor (m
source-I.I. distance( rB) source-meas. chamber distance(rT)
) =
A
2
Example:
100
2
=1.17
92.5
- Grid correction factor (m
- IONTOMAT chamber correct. factor (m
- Tabletop correct ion factor (m
) = 1.7
R
)= 1.1
I
) = 1.1
T
When setting the dose/dose rate, multiply the dose/dose rate required at the I.I. input by the correction factors.
e.g. 174 nGy/s x (m
A
x m
R
x m
x mT)
I
174 nGy/s x (1.17 x 1.7 x 1.1 x 1.1) = 174 nGy/s x 2.4 = 408 nGy/s
µR=8.7 nGy)
(1
Record the correction factor m in Chapter 4.1 of the "Test Certificate for DL and DFR/DR Systems". The test certificate is filed in Register 9 of the System or Generator Binder.
POLYDOROS SX Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
NOTICE

Startup 3

Programming the POLYMATIC 3

General Remarks 3

In system shipments, the programming has already been carried out at the factory. In subsystems, the expansions to be connected during start-up must be programmed
appropriately.
File the test certificate"Fluoro and DFR/DR Systems", RXD0-
000.037.01... in Register 9 of the System or the Generator Binder.
A help function is available in the Service Software beginni ng with Service Software
VE00A.
By clicking on the "HELP" box in the particular window or
by pressing the "F1" function key on the PC or
in the "Help" / "Index" in the main menu
valuable information and the required programming and entries can be selected.
Chapter 3
Page 1 of 22

Programming 3

1 Establishing the Generator - Service PC Connection 2 "System" / "Connect" 3 "Configuration" / "Options..." / "Page 2..."
Set the dose and dose rate units of measurement.
4 "Configuration" / "Unit Selection" / "Button 1..."
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Chapter 3Startup
Page 2 of 22

Fluoroscopy Programming 3

The Test Certificate "Fluoro and DFR/DR Systems", RXD0-000.037.01... is filed in Register 9 of the System or Generator Binder.
Select "Fluoroscopy" (if not already sel ected!)
Select "FC-Details..."
Appropriately program the "Primary FC-Data" and con firm with "OK".
Appropriately program the "VIDEOMATIC P" and confirm with "OK".
Record the values programmed in the Test Certificate:
SID (focus-I.I. distance) in Chapters 4 and 4.2, Over-table units: 115 cm Under-table units: max. SID (do not move out the Distator) C-Arm units: 100 cm
I.I. formats in Chapter 4.1.2 Zoom dose rate (Zoom dose factors) in Chapters 4.1.2 and 4.2 Dose rate in Chapters 4.1 and 4.2 The fluoroscopy curves used by the customer in Chapter 4.2.

Additional Information: 3

"maximum fluoro rate":
- Therapy Simulation: OPTILIX 150/12/50 C: max. 1000 W
- High Contrast Fluoro: MEGALIX 125/40/82 C: max. 1200 W (with HV generators beginning with Ser. No. 01056 max. 2000 W) with the fluoro footswitch wit h the 2nd pressure point
- "Normal Fluoro" all other X- ray tubes max. 450 W
When the image intensifier model is selected, al l input boxes are assigned default
values (see also the IQ reference value l ist).
The dose rate selected is always relative to the full format
(largest I.I. form at!)!
Program the same dose rate values for Automatic 1 and 2
Fluoroscopy blocking: "Block on"
according to regulations in Germany and in DHHS countries, fluoroscopy must be blocked after 5 minutes.
If fluoroscopy is configured after the "test shots" of the X-ray tube, the X- ray tube test
shots must be carried out again ("Adjustment"/"Tube 1... "/"PT Emulation...": Module J05, J0 6 and J07")!
POLYDOROS SX Register 4 RX63-050.034.12 TD AX 6 Siemens AG Rev. 01 05.96 Medical Engineering
Startup Chapter 3
Page 3 of 22
Excerpt from the "Reference Value List for Fluoro and DFR/DR Systems":

Fluoroscopy dose rate at the I.I. input 3

2.1 mm Cu filter, X-ray tube voltage 65 - 85 kV
Zoom Dose
Factors
PL SX
50/80
1:
2.0:
4.0:
5.0 1:
1.8:
3.3:
5.5 1:
2.1:
3.8
BV-Typ /
I.I. Type
40 cm HD 40
40 cm HDR 40
33 cm HD 33
Formate /
Formats
28 20 14
30 22 17
23 17
1, 2
Dosisleistg.
Dose rate
(nGy/s)
87 160 315 (435)
87 120 255 440
113 240 450
3
Hochkontrast /
High cont rast
(nGy/s)
139 255 505
(695)
139 195 405 700
3
33 cm HDR 33
22 17 13
27 cm 27
17 14
Sirec.Komp. (alt/old)
23 cm HDR/HDLE Sirec.Komp. (neu/new)
U
Multiplier
Tolerance: I.I. full format Tolerance of indirect dose rate control Values in brackets ( ) apply if TV iris diaphragm is configured. Urology: With TV iris diaphragm for Automatic 2, set doubl e the dose rate.
= 700 – 950 V
22 17
23 17 13
± 10%; at Zoom formats approximate value only.
113 235 405 680 (430)
174 445 655 (435)
261 435
261 440 730 (470)
*1
1 nGy/s = 0.115 µR/s / 1 µR/s = 8.7 nGy/s
±1 kV, mA: ±10%
1:
2.3:
3.8:
6.4 (3.8) 1:
2.5:
3.7 (2.5) 1:
1.7 1:
1.8:
3.1 (1.8)
*1 effecti ve beginning 7.93 (possible only with masked Multiplier)
Siemens AG TD AX 6 RX63-050.034.12 Register 4 POLYDOROS SX Medical Engineering Rev. 01 05.96
Loading...
+ 35 hidden pages