GCX Mounting Assembly Operation/Installation Manual
®
Dräger Model Cato
The purpose of these instructions is to:
1. Describe attachment of GCX mounting hardware to the Cato anesthesia machine.
2. Describe attachment of the Siemens monitor to the GCX mounting hardware.
WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE
DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE
ANESTHESIA MACHINE.
EXTENDED WHEEL ADAPTER UPGRADE SET FROM DRÄGER (PART NUMBER: 86 018 68)
THIS CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH
Installer: When installation is completed, provide these instructions to the end user for future reference.
INSTALLATION
Tools Required: Philips head screw driver, Hex key (incl.).
Mounting the Channel to Cato Edition
Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2).
Brass Plate with inserts from Dräger (M29110)
1. To attach the Channel to the Cato left side accessory track,
locate the brass plate in the accessory track, align the
mounting holes in the Channel with the threaded holes on
the brass plate then run the M5 flat head screws (2) partly
in. (Note: Fixed Stop is at the bottom of the Channel)
2. Set the height of the Channel in the accessory track and
tighten the M6 screws to hold in place.
Prepare M Series Arm
1) Install the Tilt Limiting Stop screw.
Patient Monitor – Left Side on Pivot Arm / Power Supply on Right Side
5/16-18 x ½” SHCS (1)
with Siemens SC9000 Patient Monitor
Accessory Track
M6 x 16 mm (2) FHMS
Fixed Stop
DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 1 of 4
Install M Series Arm
1. The Slide has (2) ¼-20 x ¾” SHSS set screws that, when tightened, secure the Arm at the desired location
within the Channel.
2. While supporting the bottom of the Arm with one hand, guide the Slide (at the rear of the arm) into the top
of the Channel.
3. Once the Arm is at the desired height, tighten both set screws using the supplied 1/8” (3 mm) hex key.
Slide Pick & Go Base into Mounting Plate
1) Align the mounting adapter located at the bottom of the Pick & Go base with the edge of Mounting Plate. Slide the
Pick & Go base into the plate until it stops at the Spring Plunger (located below the Mounting Plate).
Pull the Spring Plunger and continue to slide the base until the mounting adapter engages the Spring Plunger. Tighten
the (4) nylon thumb screws on the back of the Mounting Plate.
Install 4” Cato Channel on Right Side
As described on Page 1 of this document.
Install Power Supply Mounting Adapter to Siemens Power Supply Mount (Siemens P/N: 4720129)
1) Install according to the instructions included with the adapter.
Slide Power Supply Mount into Channel & Insert Power Supply
1) Align the Slide of the mount with edge of the Channel insert into channel.
DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 2 of 4