The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
0 - 2Revision
ChapterPageRevision
AllAll04
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 2 of 4Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
4 _______Changes to previous versions____________________________________4 - 1
Siemens AGSPR8-220.031.04Page 3 of 4MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
0 - 4Contents
This page intentionally left blank.
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 4 of 4Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
Prerequisites1
Purpose1
This document describes how to install a DAP measuring system (Dose Area Product
measuring system) on a MOBILETT Plus/ Plus E/ Plus HP.
General1
The DAP equipment is designed for the measurement of the area kerma product
(kerma = kinetic energy released in matter) during an X-ray examination. It is especially
used to record the radiation dose to which the patient is exposed.
Systems / products1
This installation instruction is only valid for:
MOBILETT Plus with Serial No. higher than 2350
MOBILETT Plus E with Serial No. higher than 10200
MOBILETT Plus HP with Serial No. higher than 30001
Important note1
1 - 1
Store the document "Supplement to the Instructions for Use" together with the MOBILETT
Instructions for Use as a reference on how to use the DAP meter after installation.
NOTE
To facilitate possible troubleshooting, store this document along
with the Technical Manual
.
Components included1
The kit (Part No. 66 13 082) contains the following parts:
Siemens AGSPR8-220.031.04Page 1 of 4MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
1 - 2Prerequisites
Pos No. Quantity Part No.Name
14166 11 107Shrinking tubes, length 150 mm
15165 82 407Shrinking tube for twisted cables, length 40 mm
16166 11 115Shrinking tube for K9, length 50 mm
17166 13 558Flat tie holder
18165 84 531Mounting bracket/wedge for board D70
19265 84 796Spacer for the transport safeguard
20165 84 853Power supply
21165 61 745Cable K71 (board D70)
22160 22 966Screw (MC6S 3x8 mm) for the D70 wedge
23260 32 965Stop screw (SK6SS 6x16mm)
24360 23 485Screw (MC6S 5x16mm)
25160 23 162Screw (MC6S 4x8mm)
26262 29 400Screw (MRX-TT 5x10mm)
27665 91 536Screw (PT 4x8mm)
28360 34 573Nut (M6NY 5mm)
291060 32 366Washer (KB 5.1mm)
30161 68 269Washer for board D70 (BRB 4.3mm)
31265 84 663Lock strip for Ionization chamber
32166 31 001This instruction
33167 96 544Flat connector
34165 28 108Label kit
35166 08 330Supplement to the Instructions for Use (English)
36166 08 983Supplement to the Instructions for Use (German)
37166 08 967Supplement to the Instructions for Use (French)
38166 08 975Supplement to the Instructions for Use (Spanish)
39166 08 991Supplement to the Instructions for Use (Swedish)
40166 09 007Supplement to the Instructions for Use (Italian)
41166 09 015Supplement to the Instructions for Use (Finnish)
NOTE
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 2 of 4Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
Some components will not be used during installation, depending
on the type of unit the kit is installed on. Some components also
have spare parts.
Prerequisites1 - 3
Tools / documents required1
• Standard installation tools
• Soldering iron
• The document Service Instructions for the relevant MOBILETT
• Set of ring spanners / open end keys, including a key for 5.5 mm diameter
• Drill and drill bit for 5 mm diameter
Time required1
• Approximately 8 hours for one customer support engineer.
Safety information1
WARNING
Dangerous x-ray radiation and/or dangerous electrical voltage
during checks and adjustment work steps!
If not observed, death or serious physical injury can occur.
The following items must be complied with when performing service work and tests:
- The product-specific safety information in the documents,
- The general safety information contained in ARTD Part 2.
• Connect the MOBILETT only to a mains power supply outlet
(receptacle) that corresponds to the installation requirements of
VDE 0107 or to country-specific regulations.
• Remove or install boards only when the generator is switched
OFF . Adhere to the ESD guidelines.
• Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol . Radiation
protective clothing m ust be worn for this type of adjustment (see
ARTD, P art 2).
Siemens AGSPR8-220.031.04Page 3 of 4MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
1 - 4Prerequisites
This page intentionally left blank.
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 4 of 4Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
Installation2
Preparatory work2
5
2
4
2 - 1
1
Fig. 1 Removal of MOBILETT covers
1.System OFF.
2.Disconnect the mains cable.
3.Remove the system upper cover (pos. 1 / Fig. 1). With MOBILETT Plus HP,
disconnect K106.
6
3
MOB00453
4.Switch S10 to discharge position.
5.Wait 5 minutes to allow the capacitor banks to discharge. Then remove the side
covers (pos. 2 / Fig. 1), the column cover (pos. 2 / Fig. 2) and the cover with the
cassette compartment (pos. 3 / Fig. 1).
6.For MOBILETT Plus HP only: Remove the protective cover in front of the
batteries. Disconnect one of the blue battery connector pairs.
Siemens AGSPR8-220.031.04Page 1 of 30MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
2 - 2Installation
7.Make sure the capacitor bank is discharged.
WARNUNGWARNING
8.Remove the lower and upper arm covers (pos. 4 / Fig. 1).
NOTICE
The capacitor bank might still contain up to 350 VDC.
Life-threatening hazard of electric shock.
Follow the safety procedures described in the docu-
ment Service Instructions for detailed instructions on
how to ensure the capacitor bank is discharged.
For MOBILETTs with remote exposure switch system
installed:
When removing the lower arm cover, be sure to unplug
the module connector for the remote control from the
electronics package alongside the main MOBILETT
CPU board (see pos. 1 / Fig. 18). Dismount the IR sensor from the lower arm cover.
Remember how the cable was attached for future steps
in section "Cable assembling"!
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 2 of 30Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
Installation2 - 3
Removal of covers2
1.Remove both the right and left circular covers (pos. 5 / Fig. 1). Unplug K1 from K1
on D1 (pos. 2 / Fig. 18) and remove the cover of the articulated shaft (pos. 6 / Fig.
1).
C
1
A
2
B
Fig. 2 Removal of collimator covers
2.Turn the single tank so that the collimator is facing upwards (Fig. 2).
3.Secure the arm and the single tank with the belts (Fig. 2).
4.Remove covers “A” and “B” (Fig. 2).
5.Remove the adjustment knobs according to Fig. 3.
1.Remove collimator cover “C” (Fig. 2) and the shadowcross inside.
MOB00396
Fig. 3 Removal of collimator cover “C”
Siemens AGSPR8-220.031.04Page 3 of 30MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
2 - 4Installation
Cable assembling2
1.Feed the ends of the cable with pin plugs marked K21 and K22 (65 61 992) from
the left side of the MOBILETT (pos. 1 / Fig. 2). Insert them as shown in the figure
below (pos. 1 / Fig. 4
)
NOTICE
To facilitate further cable attachment work:
Be sure to leave enough cable so that pin plug K9a
reaches the position K9 on the main CPU board D1
(see Fig. 18)
M
C
B
MOBILETT Plus
2
WARNING!
1
MOB00615
Fig. 4 Insertion of cables, cutting of cable tie
2.Cut and remove the upper cable tie (pos. 2 / Fig. 4).
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 4 of 30Siemens AG
Rev. 0409.03CS PS 24Medical Solutions
Installation2 - 5
3.Feed the pin plugs (K21 and K22) underneath the bar and between the two thick
cables (Fig. 5).
MOB00520
Fig. 5 Feeding cables
4.Insert the cable end with pin plug K22 into the joint between the upper and lower
arm as shown in Fig. 6. Make sure the cable runs between the two thick cables.
Fig. 6 Cables in joint
MOB00455
Siemens AGSPR8-220.031.04Page 5 of 30MOBILETT Plus/ -E/ -HP
Medical SolutionsRev. 0409.03CS PS 24
2 - 6Installation
5.Fasten the cable with a cable tie (2.5x100mm) according to pos. 2 / Fig. 7. The pin
plug K22 must have the same position as shown in Fig. 7, in line with the arm
pieces (pos. 3 / Fig. 7).
2
3
1
MOB00450
Fig. 7 Cable fastening on upper arm.
6.Place two cable ties (2.5x100mm) according to pos. 5 / Fig. 8.
7.Move the arm to see that the cable runs freely throughout the arm and is not
stretched or trapped.
MOBILETT Plus/ -E/ -HPSPR8-220.031.04Page 6 of 30Siemens AG
Rev. 0409.03CS PS 24Medical Solutions