WORKING FREQUENCY
RANGE KV
RANGE MAS
A.T.PILOTAGE
RIPPLE
TOTAL FILTRATION
RISING TIME
Doc. 7748 Page 7 of 14
LP (Low Power) HP (High Power) LP (Low Power) HP (High Power)
7.5 kW 15 kW 7.5 kW 30 kW
150 mA 375 mA 150 mA 425 mA
3 ms ÷ 1.3 s 1 ms ÷ 0.6 s 3 ms ÷ 1.3 s 1 ms ÷ 0.5 s
Selected by the processor according to the mAs
100 kHz
40÷125(Step of 1 kV)
0.5 ÷200 in 25 values
Inverter driven by IGBT
≤ 3% at Max Power
> 2.7 mmAl
≤ 1 ms
Section 1
User’s Manual
GENERAL DESCRIPTION
X-Ray Tube Head
TYPE OF ANODE
FOCAL SPOTS
All the other information relevant to the X-Ray Tube Head and to the X-Ray Tube can be found in the X-Ray Tube Head
Technical Data Sheet
Rotation with speed 3000 RPM
0.6 mm
1.3 mm
Collimator (optional)
SHUTTERS TO MULTIPLE PLANSParallels and perpendicular with manual movement
All the other information relevant to the Collimator can be found in the relative Technical Data Sheet
Dosimeter (optional)
MODELKermax-plus VacuDap 2000 with printer optional
ACTIVE AREA146 x 146 mm2 147 x 147 mm2
MINIMAL DOSE RESOLUTION 1 mGycm2 1 mGycm2
MAXIMAL MEASURABLE DOSE 9999.9999 mGycm2 9999.9999 mGycm2
Operating modes and functionality
INTERFACE USERPolycarbonate flat keyboard with alphanumeric LCD display for all the operative
parameters and messages of possible anomalous conditions – administrated by a
microprocessor.
OPERATING MODES RADIOGRAPHY
X-RAY CONTROLDistance control with double – click and extensible cable (≥4m)
SAFETY
Two-points techniques (kV-mAs)
40 programmable anatomic technique (20 for LP and 20 for HP)
Filament current
mA
and mA
min
Maximum exposure time
Temperature maximum X-ray tube head
Count thermal units X-ray tube head
Max kV, min kV, max ∆kV, max I
Anode rotation
Microprocessor self – test
max
Transport and storage conditions
MAXIMAL TEMPERATURE
RECOMMENDED TEMPERATURE
RELATIVE HUMIDITY
ATMOSPHERIC PRESSURE
The aim of this manual is to provide qualified radiology technicians, medical and paramedical
staff with operating instructions to make use of the radiographic unit simple and safe.
This apparatus emits X-
RAYS and must only be used in compliance with the safety instructions
indicated in this manual and must not be used for any other purposes than those foreseen.
The system must only be used by personnel with the necessary knowledge in the field of radiation
protection and with the necessary training in use of X-ray apparatus.
Follow the indications given below very carefully:
the apparatus must not be used when there are any electrical and/or mechanical faults;
do not use the system when any signalling or alarm device of the system is not working
properly;
under no circumstances work with the unit
in places saturated with vapours and/or
flammable gases or explosives;
any modification to the system must be authorised in writing by the
MANUFACTURER;
if you want to use the apparatus in combination with other equipment, components or
modules, and when compatibility with the latter is not certain, it is indispensable to make
sure that there is no danger to the patients and/or operating personnel. In this case, consult
the manufacturer of the apparatus in question or an expert in the sector;
like any other apparatus, the System must be used correctly. It also need
PERIODIC
CHECKS AND MAINTENANCE
all maintenance, repair and/or modification work must be carried out by
QUALIFIED AND AUTHORIZED
as specified under § 2.1 in this section;
PERSONNEL
by the Manufacturing Company. The latter declines any
responsibility for malfunctions caused by unauthorised interventions;
the
MANUFACTURING COMPANY of the apparatusdeclines any responsibility for damage
to people and/or things caused by its improper use.
Section 2
Page 2 of 10 Doc. 7749
User’s Manual
SAFETY AND MAINTENANCE
1.2 RESPONSIBILITY DECLARATION
Encosed to the presente manual the MANUFACTURER RESPONSIBILITY DECLARATION (SEE
ANNEX
1).
1.3 COMPLIANCE AND REFERENCE ADDRESS
For information relevant to the compliance make always reference to ANNEX 1.
1.4 SYSTEM SAFETY
1.4.1 MECHANICAL SAFETY
For moving the unit only use the proper handle
for tilting operations of the unit only use the proper handles and relative tail rotor control
panel
avoid hitting the unit against obstacles
do not remove the protective casing of the apparatus except for maintenance work
expressly foreseen and described in this U
SER’S MANUAL or in the SERVICE MANUAL.
1.4.2 ELECTRICAL SAFETY
Make sure that the power supply socket where the apparatus is to be connected is
approved for the foreseen voltage and current for use of the system
the radiology system must not be used in rooms where there is the risk of explosion
disconnect the installation from the mains before carrying out any cleaning, disinfection
and/or sterilisation
the cleaning and disinfection products for the installation can form explosive
gaseous mixtures. It is compulsory to only use products which comply with the
corresponding Standards in force
take care not to spill conductive liquids on the apparatus as they would jeopardise
operation and safety were they to penetrate to the inside
always turn the apparatus off after use.
Section 2
Doc. 7749 Page 3 of 10
User’s Manual
SAFETY AND MAINTENANCE
1.5 PROTECTION AGAINST IONISING RADIATION
Before carrying out any X-ray exposure, make sure that all the necessary precautions against
radiation have been taken.
During radiation emission, the personnel in the X-ray room must respect the regulations in force
regarding protection against radiation. For this, bear in mind the following rules:
where necessary, use protective accessories against radiation;
always use the special X-ray protective coats: an X-ray protective material equivalent to
0.35 mm of lead (0.35 mm Pb) attenuates the radiation produced at 50 kV by 99.95%
and at 100 kV by 94.5%;
the best protection against radiation is distance: therefore keep as far away as possible
both from the source of radiation and from the exposure object, also by using the suited
exposure-push-button with its extensible cable;
avoid moving or standing in the path of the rays;
always use the smallest exposure range possible: the radiation dispersed largely depends
on the volume of the object X-rayed;
keep the patient’s examination area the furthest away as possible from X-ray source.
Section 2
Page 4 of 10 Doc. 7749
User’s Manual
SAFETY AND MAINTENANCE
1.6 RESIDUAL RISKS
The system is designed and constructed according to the strictest principles of compliance with
safety requirements. However, there are residual risks due either to incorrect use of the apparatus
or to deficiencies in the protective measures taken.
With regard to the risks due to incorrect use of the apparatus, please see the instructions and
recommendations given in the points above and we
The mobile unit has been designed and constructed so that it will not tilt over up to an
angle of 10° to horizontal in transport position (
advisable:
underline that:
SEE SECTION 3 - § 1). It is therefore
• not to stand, move or place the mobile unit on surfaces with a slope of more than 10°
• not to try to move the mobile unit with the brakes on
• to take care to avoid any obstacles on the floor when moving the mobile unit (cables,
steps and uneven levels of all kinds)
For the residual risks due to any defect in the protection measures taken, it must be remembered
that:
Protection against electric shocks is carried out by means connecting the metallic
covering parts of the apparatus to earth: it is therefore necessary to periodically check –
according to the N
correct operation of the whole earthing circuit.
ORMAL MAINTENANCE PLAN described under § 2.1of this section –
Not taking notice of the unit alarms could cause overheating of the X-
: this overload could lead to loss of the means of insulation in the X-ray tube
!
!
!
Doc. 7749 Page 5 of 10
HEAD
itself at very high temperatures.
head
During apparatus movements, take care that the parts do not hit the patient or
the operator.
Avoid very fast movements: the kinetic energy accumulated could be a hazard
for personnel near the unit.
RAY TUBE
Section 2
User’s Manual
SAFETY AND MAINTENANCE
1.7 SIGNALS
1.7.1 SYMBOLS USED
Apart from the symbol on the Control console, others have been used on the unit, as illustrated
below.
SYMBOLMEANINGPOSITIONING
C
AUTION: SEE THE ATTACHED
DOCUMENTATION
ADIOGRAPHY
R
RAY FOCUS POSITION
X-
TRANSPORT POSITION
Cassette holder and panel of
protection of the electronic group
Radiography pushbutton
X-ray tube head covering
Pantograph arm
LOCKED MOVEMENT
Pantograph arm
UNLOCKED MOVEMENT
HIGH VOLTAGE
PROTECTIVE EARTH
Pantograph arm
Panel of protection of the
electronic group
Equipotential bar
1.7.2 LABELLING
Sample of the IDENTIFYING LABELS sticked on the X-ray unit is attached to the present
manual
Section 2
Page 6 of 10 Doc. 7749
(SEE ANNEX 2).
User’s Manual
SAFETY AND MAINTENANCE
1.7.3 SIGNALLING AND ALARM MESSAGES
In the presence of the following alarms visualized on the display, the X-ray exposition is
disqualified and the console’s alarm red led light up. In case the alarm is also present after the
execution of the suggested intervention, please contact the service assistance department.
MESSAGEMEANINGINTERVENTION
SUPPLY FAULT
KV FAULT
MA FAULT
FILAMENT FAULT
THERMIC SAFETY
STARTER FAULT
X-RAY LACKING
MAN STOP RX
MAX TIME
X-RAY TUBE TOO HOT
X-RAY COMMAND ACTIVE
INVERTER FAULT
SWITCH OFF FOR 1 MIN
BATTERY FAULT
OVERVOLTAGE BATTERY
WAIT CONNECTION
POTTER FAULT
(only with Potter installed)
Error in the electronic system Contact technical service
During a radiograph the effective
kV are less than 85% of those set:
fault on the power circuit
During an exposure the mA value
is lower than the allowed limit
No current in the filament Switching off the unit, switching it on
X-ray tube head too higt Wait for the X-ray tube head to cool
Error in the x-ray tube stator power
supply circuit
Error in the high voltage generation
circuit
During a radiograph with cassette,
the control pushbutton for X-ray
command has been released early
The unit has interrupted
radiography exposure as the
maximum exposure time allowed
has been reached
It is not possible to begin exposure
since the remaining thermal units
available are too few
The operator has pressed the
radiography command before the
system had finished the initial
control stage
Fault into the inverter Switching off the unit, switching it on
Capacitors bank still loads With the unit off wait 1 minute before
Power circuit fault Contact technical service
Battery circuit fault Contact technical service
The keyboard does not communicate with the unit
Potter fault Switching off the unit, switching it on
Switching off the unit, switching it on
and repeat the X-ray
Switching off the unit, switching it on
and repeat the X-ray
and repeat the X-ray
down.
Contact technical service
Switching off the unit, switching it on
and repeat exposure
Assess the validity of the image
obtained and, if necessary, repeat the
exposure.
Switching off the unit, switching it on
and repeat exposure
Wait for the X-ray tube to cool down.
Release the radiography pushbutton
and wait until the system is ready
and repeat exposure
switching it on
Switching off the unit, switching it on
and repeat exposure
and repeat exposure
Section 2
Doc. 7749 Page 7 of 10
User’s Manual
SAFETY AND MAINTENANCE
2 MAINTENANCE
This manual only refers to routine maintenance. For special maintenance operations, interventions
in the case of faults and/or replacement of components, the S
must be consulted.
ERVICE MANUAL - SECTION 3 - § 2
2.1 ROUTINE MAINTENANCE
2.1.1 GENERAL RACOMMENDATIONS
The radiological system requires regular checks and maintenance. The following
recommendations have the aim of helping the operator to keep the apparatus in good working and
safe conditions during service.
The system contains mechanical parts subject to wear according to use: following prolonged
use, wear on parts may decrease safety during use. For this reason, it is essential for the checking
and maintenance operations indicated below to be carried out consistently to protect the operators
and patients against any damage caused by mechanical breakdowns.
Correct adjustment of the electrical and electronic systems has a direct influence on the
operation of the system, on the quality of the image and on the electrical safety of the system, as
well as on the level of exposure to radiation the operators and patients are subjected to.
The M
and interventions to be carried out by specialised personnel authorised by the manufacturer. All
maintenance operations are the responsibility of the owner of the apparatus.
AINTENANCE PROGRAMME, described in the following paragraphs, consists of controls
Should it be necessary to replace components or parts which may in any way
!
condition the safety of the machine, only use original spare parts.
Section 2
Page 8 of 10 Doc. 7749
User’s Manual
SAFETY AND MAINTENANCE
2.1.2 FREQUENT CHECKS AND INSPECTIONS
The operating personnel must be suitably trained to be able to carry out the daily and weekly
checks indicated in T
ABLE 1.
The other controls described in this chapter and the interventions described in the following
chapters are reserved for qualified and authorised personnel of the technical assistance service.
T
ABLE 1
INTERVALCHECK
DAILY
CHECKS
WEEKLY
CHECKS
6-MONTHLY
CHECKS
Operation of the signals, displays and LEDs
Operation of the stationary brake
Integrity of the warning and danger labels
Absence of oil leaks from the X-ray tube head
Absence of unusual noises in the X-ray tube head during X-ray emission
Correct operation and the value of the whole earthing circuit
Power supply voltage value
Value of the continuous voltages generated inside the system
Fixing and general state (dust and corrosion) of the boards
Centering of the X-ray tube head-collimator assembly
2.2 CLEANING AND DISINFECTION
Products with a high content of alcohol, corrosive and/or abrasive detergents or solvents must
not be used to clean the surfaces of the apparatus.
To disinfect the system, only use methods in compliance with the laws in force regarding
disinfection and protection procedures against explosion.
To carry out the cleaning and disinfection operations, take the following precautions:
turn the system off and disconnect the mains power supply cable
make sure that no liquid gets into the apparatus so as to avoid any short-circuits or
corrosion of the electrical and electromechanical parts.
The unit has not to be used in presence of anaesthetic and/or infiammable
disinfectant and cleaning products.
!
Doc. 7749 Page 9 of 10
If, producing explosive gaseous mixture, are used, make sure that gases are
dispersed before switching on the unit.
Section 2
User’s Manual
SAFETY AND MAINTENANCE
2.3 DISPOSAL OF DEVICE
During the disposal of a device you have to take case of the components that may present risks
connected to their elimination:
the X-ray tube head holds an insulating dielectric means and lead protections that have
to be eliminated taking into account the standards and the laws in force (consult the
documentation released by the manufacture)
the collimator holds leads shutters that have to be eliminated in respect of the standards
in force (consult the documentation released by the manufacture)
the electrolytic capacitors holds an insulating dielectric means that have to be eliminated
taking into account the standards and the laws in force.
The other elements of the unit are composed of:
ferrous material (chassis, screws, etc)
plastic material (covering)
electrical cables
electronic boards.
These elements do not represent a direct source of risk during the elimination phasis of the devices.
The disposal of all components have to be carried out in compliance with the
!
local laws in force at the moment of the disposal.
Section 2
Page 10 of 10 Doc. 7749
User’s Manual
Use of the
unit
Section 3
Doc. 7750 Page 1 of 18
User’s Manual
USE OF THE UNIT
1 TRANSPORT
For the transport of the unit follow what mentioned indicated below: the figure shows the transport
configuration (for the numerical references S
¯ wind up the power supply cable to its special power supply cable holder
EE SECTION 1-§2.1):
(12)
¯ insert the pantograph arm lock (15)
¯ move the unit just after the release of the positional brakes by pressing the special pedal
(19)
¯ to overcome any small differences in level, press the tilting support (14) with one foot
and meanwhile pull the tilting handle (3-optional)
in the direction shown in the figure.
3
PULL
12
PUSH
14
19
15
Section 3
Page 2 of 18 Doc. 7750
User’s Manual
USE OF THE UNIT
N
OTE: the mobile unit is normally braked and cannot be moved unless the pedal brake, shown in
the figure (19), is applied.
Section 3
Doc. 7750 Page 3 of 18
User’s Manual
USE OF THE UNIT
2 UNIT MOVEMENTS
Do not try to move the system when the brakes are applied.
!
To move the apparatus, use the special handles provided.
2.1 ARM POSITIONING
¯ Release the pantograph arm lock (15)
¯ using the X-ray tube head positioning handle
the X-ray tube head/collimator assembly
required height. By releasing the handle, the system
stays in the position it has been put, thanks to a clutch
incorporated in the tube arm.
device
(7), lift
up to the
15
15
7
Section 3
Page 4 of 18 Doc. 7750
User’s Manual
USE OF THE UNIT
2.2 X-RAY TUBE HEAD/COLLIMATOR ASSEMBLY MOVEMENTS
The X-RAY TUBE HEAD/COLLIMATOR assembly can be rotated in all directions, as shown in
the figures below. For positioning, use the X-
The angle of the X-
(6) and lateral (5) position respectively.
RAY ASSEMBLY is indicated by the two GONIOMETERS positioned in the front
RAY TUBE HEAD POSITIONING HANDLE (7).
6
5
7
Section 3
Doc. 7750 Page 5 of 18
User’s Manual
USE OF THE UNIT
6
5
Section 3
Page 6 of 18 Doc. 7750
3 SWITCHING ON
Make sure that the power supply socket is approved for the values reported on
User’s Manual
USE OF THE UNIT
!
the unit label and provided with earthing terminal
Do not carry out continuous switching on and off of the unit.
!
To make the system operational, follow the indications given below:
¯ connect the unit to the power supply mains: on unit the LED of the
control console remain lit
¯ turn the unit on
¯ the initial automatic test of the unit will be activated immediately:
• for about 4 seconds, the displayshows the version of control
software installed; all the leds light up in sequence
¯ successively will appear on the display the message W
the capacitors battery loading are ready, the value of the battery
voltage and the job technique
N
OTE: Before carrying out any other operation, wait until the wording “WAITING”disappears
from the display.
Doc. 7750 Page 7 of 18
30E100 R. X.XX
UNIT CONTROL CONSOLE–MESSAGES SHOWN ON SWITCHING ON
AITING until
Section 3
User’s Manual
USE OF THE UNIT
WAITING 250VMAN
UNIT CONTROL CONSOLE–MESSAGES SHOWN ON SWITCHING ON
¯ the display shows the preset radiography values and the working
technic (technic preset manual), indicates the percentage of thermal
units still available (when the X-ray tube head is under starting
conditions, the display will indicate a percentage of H.U. of 100%)
¯ in this situation the machine is ready for X-ray command
¯ select the type of exposure required according to the description on
the following pages.
400.5100%MAN
UNIT CONTROL CONSOLE–MESSAGES SHOWN WHEN SYSTEM IS READY
Section 3
Page 8 of 18 Doc. 7750
User’s Manual
USE OF THE UNIT
4 CONFIGURATION
4.1 CHANGING THE LANGUAGE FOR THE UNIT CONSOLE
The signalling on the unit console can be displayed in Italian, English, French or Spanish.
Selection of the language is made when the system is turned on as shown below.
¯ Turn the unit on using the ON/OFF button (S
EE §3): the initial test
of the unit will be activated automatically, during which the
display will show the version of the control software installed
¯ keep immediately the relevant button pressed until the word
AITING" appears on the display in the language selected.
"W
ENGLISH
FRENCH
ITALIAN
SPANISH
Section 3
Doc. 7750 Page 9 of 18
User’s Manual
USE OF THE UNIT
4.2 PROGRAMMING THE ANATOMICAL TECHNIQUES
Programming an ANATOMICAL TECHNIQUE means associating the KV and mAs values
considered suitable with the identifier (i.e. with the name, such as "S
will then be possible to call up an
ANATOMICAL TECHNIQUE by means of its name and immediately
KULL", "SHOULDER"…). It
find the radiographic parameters already set to the values memorised previously. To associate the
KV and mAs values with an identifier, proceed as follows.
¯Press the A
NATOMICAL TECHNIQUE SELECTION button: the LED
in the corner of the button itself will light up
¯ using the P
NATOMICAL TECHNIQUE for which the KV and mAs values are to
A
ROGRAMSELECTION up-down type buttons, select the
inserted; also select the power LP or HP where the values will be
memorised
¯ select the KV and mAs values considered suitable for that type of
examination using the KV S
ELECTION and mAs SELECTION
buttons (if necessary see the § 6.2)
¯Press the D
ATA MEMORISATION button until the acoustic signal is
heard. At this point, the KV and mAs values relative to the selected
NATOMICAL TECHNIQUE have been memorised
A
¯ It is therefore possible to select a new A
NATOMICAL TECHNIQUE
to associate specific radiographic parameters to, by repeating the
same procedure
¯ For exit to the programming the
NATOMICAL TECHNIQUE SELECTION button
the A
ANATOMICAL TECHNIQUES press
Section 3
Page 10 of 18 Doc. 7750
User’s Manual
5 ADJUSTMENT OF THE X-RAY FIELD SIZE
¯ Adjust the Focus / Film distance - which can be measured using
ETRACTILE METER (53) - moving the X-Ray Assembly (SEE
the R
§ 2.2)
¯ once the X-ray tube head/Collimator Assembly has been
positioned, switch the Collimator light on by pressing the
Collimator Light On Button (50) on the Control Panel or by
pressing the Collimator Light On Button on the Collimator itself
¯ at this point, adjust the luminous field (which coincides with the X-
Ray Field) by means of the Longitudinal Diaphragm Adjustment
USE OF THE UNIT
(54) and Transverse Diaphragm Adjustment (51) knobs.
51
50
54
53
52
Section 3
Doc. 7750 Page 11 of 18
User’s Manual
USE OF THE UNIT
Section 3
Page 12 of 18 Doc. 7750
6 EMISSION OF X-RAYS
Before carrying out any exposure, make sure that all the necessary precautions
User’s Manual
USE OF THE UNIT
!
!
against radiation have been taken.
Whatever the exposure mode chosen, X-ray passage is signalled by
ED lighting up.
the L
6.1 EMISSION OF X-RAYS USING THE MANUAL TECHNIQUE
¯ After being switched on (SEE § 3 of this section), the unit
¯ the preset
¯ using the
(1)
¯ pull out the X-
automatically goes into M
you like select High Power HP use the button
KV and mAs values appear on the display
KVSELECTION and mAsSELECTION up-down buttons,
set the required values for these parameters
RAY CONTROL pushbutton. This device is a double-
ANUAL TECHNIQUE on Low Power LP; if
click button (Click A and Click B – see figure) and must be used as
indicated below
Doc. 7750 Page 13 of 18
A
B
Section 3
User’s Manual
USE OF THE UNIT
¯ use the X-RAY CONTROL pushbutton and make the first click (A);
k
eep the button in this position until on the display appears the
wordings R
EADY FOR RADIOGRAPHY and the X-ray ready LED
lights up
¯ make the second click (B) with the X-
RAY CONTROL pushbutton
and keep it in that position. At that moment X-ray emission starts
and the X-ray on LED lights up
¯ at the end of X-ray emission, the X-ray ON LED turns off and the
acoustic signal stops. At this point both of the X-ray control clicks
can be released and the display shows the actual emission time.
Releasing the first click (A) too soon has no effect.
Releasing the second click (B) too soon interrupts X-ray emission; the display will
!
show the message “M
alarm sound (S
EE SECTION 2-§ 1.7.3).
AN STOP RX” for about 10 seconds and the unit will give an
Section 3
Page 14 of 18 Doc. 7750
User’s Manual
USE OF THE UNIT
6.2 X-RAY EMISSION USING THE ANATOMICAL TECHNIQUE
The X-ray unit allows up to 20 anatomical techniques to be memorized for each power (LP
and HP). A
NATOMIC TECHNIQUE means preset KV and mAs values which have been assigned a
more easily understood name for the user. The identifiers (i.e. the names) of each of the 20
Anatomical techniques which can be programmed on the unit are shown in table.
Programmable anatomical techniques
N°
1 SKULL 75 40
2 CERVICAL 65 16
3 BACK-BONE 90 50
4 LUMBAR 95 50
5 LUNG 110 1
6 SHOULDER 70 20
7 COLLAR-BONE 70 10
8 HUMERUS 50 10
9 ELBOW 48 10
Exam KV mAs
10 INFERIOR ARM 46 10
11 WRIST 42 10
12 FINGER 40 5
13 PELVIS 75 20
14 FEMORAL JOINT 70 25
15 ABDOMEN 85 16
16 FEMUR 70 16
17 KNEE 60 6.3
18 INFERIOR LEG 55 10
19 ANKLE 55 8
20 FOOT 48 6.3
Section 3
Doc. 7750 Page 15 of 18
User’s Manual
USE OF THE UNIT
It is possible to select one of the ANATOMICAL TECHNIQUES programmed previously at any time
and to use the memorized radiographic parameters. This procedure is described in the following
steps:
¯ select the power and consequently the filament which you desire to
use
¯press the A
NATOMICAL TECHNIQUE SELECTION: the LED in the
corner of the button itself will light up
¯using the P
ROGRAMME SELECTION up-down type buttons, select
the anatomical technique to be used
OTE: to make a radiography is possible to use the stored dates or change the value using the KV
N
SELECTION and mAsSELECTION up-down buttons. If you desire to memory the variation of the value
compare with the § 4.2 of this section; differently, when you go out from the modality “Anatomical
technique” the value variation will be lost.
¯ make the exposure as described in §
MANUAL TECHNIQUE
starting from point (1).
6.1X-RAY EMISSION IN
Section 3
Page 16 of 18 Doc. 7750
7 OPERATIONS AFTER USE
W
On completion of use, always proceed as specified below:
User’s Manual
USE OF THE UNIT
¯ Put the unit in the parking position (S
EE §1)
¯ Turn the unit off using the OFF button
¯ Disconnect the cable and wind it up over the special supports.
Do not remove the connector from the power supply socket if the U
!
been turned off.
NIT has not
ounding cable
Section 3
Doc. 7750 Page 17 of 18
User’s Manual
USE OF THE UNIT
Section 3
Page 18 of 18 Doc. 7750
Annex 1
RESPONSIBILITY DECLARATION
DECLARATION OF RESPONSIBILITY
IMD is only responsible for the safety of its products when their maintenance, repair
and/or modification has been carried out by IMD or by personnel expressly authorised to
do so by IMD itself.
IMD shall not be held responsible in any way for malfunction, damage and/or danger due
to incorrect use of the system or to disregard of the maintenance Regulations.
The user of the installation where the system is connected is responsible for making sure
that the installation itself is only used by suitably trained and qualified operators.
C
OMPLIANCE AND REFERENCE ADDRESS
The BASIC MOBILE RADIOLOGICAL UNIT is in compliance with the national and international
standard in force. Information relevant to the compliance may be required to:
Via Aldo Moro, 5/7 I-24020 Scanzorosciate (BG) ITALY
+39 035 66.81.63 FAX +39 035 66.81.66
P. IVA e Codice Fiscale IT 02602490167
Doc. 5987 Page 1 of 1
SYSTEM IDENTIFICATION PLATES
Plate placed
on the U
NIT
ALIMENTAZIONE / MAI NS VOLTAGE
CLASSIFICAZIONE / CLASSIF ICAT ION : 1-B IEC 601-1
S