Siemens 2005C User Manual

SOMATOM Spirit
Application Guide
Protocols Principles Helpful Hints
syngo 3D syngo Fly Through syngo Dental CT syngo Osteo CT syngo Volume Evaluation syngo Dynamic Evaluation
Software Version syngo CT 2005C
The information presented in this Application Guide is for illustration only and is not intended to be relied upon by the reader for instruction as to the practice of medicine. Any health care practitioner reading this information is reminded that they must use their own learning, training and expertise in dealing with their individual patients.
This material does not substitute for that duty and is not intended by Siemens Medical Solutions Inc., to be used for any purpose in that regard. The drugs and doses mentioned are consistent with the approval labeling for uses and/or indications of the drug. The treating physician bears the sole responsibility for the diagnosis and treatment of patients, including drugs and doses prescribed in connection with such use. The Operating Instructions must always be strictly followed when operating the CT System. The source for the technical data is the corresponding data sheets.
The pertaining operating instructions must always be strictly followed when operating the SOMATOM Spirit. The statutory source for the technical data are the corresponding data sheets.
We express our sincere gratitude to the many customers who contributed valuable input.
Special thanks to Heike Theessen, Christiane Bredenhöller, Kristin Pacheco, Karin Ladenburger, and Chen Mahao for their valuable assistance.
To improve future versions of this Application Guide, we would greatly appreciate your questions, suggestions and comments.
Please contact us:
USC-Hotline:
Tel. no.+49-1803-112244
email ct-application.hotline@med.siemens.de
Editor: Ute Feuerlein
2
Overview
User Documentation 14
Scan and Reconstruction 16
Dose Information 28
Workflow Information 36
Application Information 54
Head 70
Neck 88
Shoulder 94
Thorax 98
Abdomen 110
Pelvis 124
Spine 132
Upper Extremities 146
3
Lower Extremities 154
Vascular 162
Specials 176
Children 184
syngo 3D 240
syngo Fly Through 258
syngo Dental CT 268
syngo Osteo CT 274
syngo Volume Evaluation 288
syngo Dynamic Evaluation 306
4
5
Contents
User Documentation 14
Scan and Reconstruction 16
Concept of Scan Protocols 16
Scan Set Up 17
Scan Modes 18
- Sequential Scanning 18
- Spiral Scanning 18
- Dynamic Serioscan 18
Slice Collimation and Slice Width 19
- Slice Collimation and Slice Width for
Spiral Mode and HR Spiral Mode 20
- Slice Collimation and Slice Width for Sequence Mode and HR Sequence Mode 20
Increment 21
Pitch 22
Window values 23
Kernels 24
Image Filters 25
Improved Head Imaging 27
Dose Information 28
CTDI
and CTDI
W
Effective mAs 30
CARE Dose 32
Vol
- How does CARE Dose work? 32
Workflow Information 36
Recon Jobs 36
Examination Job Status 37
Auto Load in 3D and Post-processing
Presets 38
How to Create your own
Scan Protocols 39
- Edit/Save Scan Protocol 39
- Scan Protocol Manager 40
6
28
Contents
Contrast Medium 45
- The Basics 45
-IV Injection 47
-Bolus Tracking 48
- Test Bolus using CARE Bolus 50
- Test Bolus 51
Application Information 54
SOMATOM life 54
-General Information 54
-Key Features 55
- Description 56
- Access to Computer Based Training or
Manuals on CD ROM 57
- SRS Based Services 58
- Download of Files 59
- Contact incl. DICOM Images 60
- Trial Order and Installation 62
Image Converter 64
File Browser 66
Patient Protocol 68
Head 70
Overview 70
Hints in General 71
- Head Kernels 71
HeadRoutine 72
HeadSeq 74
InnerEarHR 76
InnerEarHRSeq 78
Sinus 80
SinusSeq 82
Orbita 84
Dental 86
7
Contents
Neck 88
Overview 88
Hints in General 89
-Body Kernels 90
Neck 92
Shoulder 94
Overview 94
Hints in General 95
-Body Kernels 95
Shoulder 96
Thorax 98
Overview 98
Hints in General 99
- Body Kernels 101
ThoraxRoutine/ThoraxRoutine08s 102
ThoraxFast 104
ThoraxHRSeq 106
LungLowDose 108
Abdomen 110
Overview 110
Hints in General 111
- Body Kernels 113
AbdomenRoutine/AbdomenRoutine08s 114
AbdomenFast 116
AbdMultiPhase/AbdMultiPhase08s 118
AbdomenSeq 122
8
Contents
Pelvis 124
Overview 124
Hints in General 125
- Body Kernels 125
Pelvis 126
Hip 128
SI_Joints 130
Spine 132
Overview 132
Hints in General 133
- Body Kernels 135
C-Spine 136
C-SpineSeq 138
Spine 140
SpineSeq 142
Osteo 144
Upper Extremities 146
Overview 146
Hints in General 147
- Body Kernels 148
WristHR 150
ExtrRoutineHR 152
Lower Extremities 154
Overview 154
Hints in General 155
- Body Kernels 156
KneeHR 158
FootHR 160
ExtrRoutineHR 161
9
Contents
Vascular 162
Overview 162
Hints in General 163
- Head Kernels 163
- Body Kernels 163
HeadAngio/HeadAngio08s 164
CarotidAngio/CarotidAngio08s 166
ThorAngio/ThorAngio08s 168
Embolism 170
BodyAngioRoutine/BodyAngioRoutine08s 172
BodyAngioFast 174
Specials 176
Overview 176
- Trauma 176
- Interventional CT 176
- Test Bolus 176
Trauma 177
- The Basics 177
PolyTrauma 178
HeadTrauma 180
Interventional CT 181
Biopsy 182
TestBolus 183
Children 184
Overview 184
Hints in General 187
- Head Kernels 190
- Body Kernels 191
HeadRoutine_Baby 192
HeadRoutine_Child 194
HeadSeq_Baby 196
HeadSeq_Child 198
InnerEar 200
SinusOrbi 202
Neck 204
10
Contents
ThoraxRoutine_Baby 206
ThoraxRoutine_Child 208
ThoraxHRSeq_Baby 210
ThoraxHRSeq_Child 212
Abdomen_Baby 214
Abdomen_Child 216
Spine_Baby 218
Spine_Child 220
ExtrHR_Baby 222
ExtrHR_Child 224
HeadAngio 226
HeadAngio08s 228
CarotidAngio 230
CarotidAngio08s 232
BodyAngio 234
BodyAngio08s 236
NeonateBody 238
syngo 3D 240
- Multi Planar Reconstruction (MPR) 240
- Maximum Intensity Projection (MIP) 240
- Shaded Surface Display (SSD) 241
- Volume Rendering Technique (VRT) 241
- Prerequisites 242
Workflow 242
- Loading the Images 242
- Creating Series 244
- Editing 246
- Documentation of Results 249
Workflow for a CT Extremity Examination 250
- Using MPR/MPR Thick 250
- Using SSD 251
- Using VRT 251
Workflow for a CT Angiography 252
- Using MIP/MIP Thin 252
- Using VRT/VRT Thin/Clip 253
Hints in General 254
- Setting Views in the Volume Data Set 254
- Changing /Creating VRT Presets 255
11
Contents
- Auto Load in 3D and Post-processing
Presets 257
- Blow-up Mode 257
syngo Fly Through 258
Key Features 258
Prerequisites 259
The Basics for CT Virtual Endoscopy 259
- SSD and VRT Presets for Endoscopic Renderings 259
- Endoscopic Viewing Parameters/
Fly Cone Settings 260
- Patient Preparation 262
Workflow 263
- Navigation of the Endoscopic Volume 265
- Fly Path Planning 266
syngo Dental CT 268
The Basics 268
Scan Protocols 269
Additional Important Information 271
syngo Osteo CT 274
The Basics 274
Scanning Procedure 275
Configuration 278
Evaluation Workflow 282
Additional Important Information 287
syngo Volume Evaluation 288
Prerequisites 290
Workflow 291
General Hints 300
Configuration 303
12
Contents
syngo Dynamic Evaluation 306
Prerequisites 308
Workflow 309
- 1. Loading the Images 309
- 2. Inspecting the Input Images 310
- 3. Generation of Parameter Images 310
- 4. Creating a Baseline Image 313
- 5. Evaluation of Region of Interests 314
- 6. Enhancement Curve 315
- 7. Documentation of Results 316
General Hints 317
13

User Documentation

For further information about the basic operation, please refer to the corresponding syngo CT Operator Manual:
syngo CT Operator Manual
Volume 1:
Security Package Basics Preparations Examination CARE Bolus CT
syngo CT Operator Manual
Volume 2:
syngo Patient Browser syngo Viewing syngo Filming syngo 3D
syngo CT Operator Manual
Volume 3:
syngo Data Set Conversion syngo Dental CT syngo Dynamic Evaluation syngo Osteo CT syngo Volume
14
User Documentation
15

Scan and Reconstruction

Concept of Scan Protocols

The scan protocols for adult and children are defined according to body regions – Head, Neck, Shoulder,
Thorax, Abdomen, Pelvis, Spine, Upper Extremities, Lower Extremities, Specials, and Vascular.
The general concept is as follows: All protocols without suffix are standard spiral modes. E.g., “Shoulder” means the spiral mode for the shoulder.
The suffixes of the protocol name are follows:
Routine“: for routine studies
Seq”: for sequence studies
Fast“: use a higher pitch for fast acquisition
HR“: use a thinner slice width (1.0 mm) for High Res­olution studies and a thicker slice width for soft tissue studies
The availability of scan protocols depends on the sys­tem configuration.
16
Scan and Reconstruction

Scan Set Up

Scans can be simply set up by selecting a predefined examination protocol. To repeat any mode, just click the chronicle with the right mouse button for “repeat”. To delete it, select “cut“. Each range name in the chron icle can be easily changed before “load“.
Multiple ranges can be run either automatically with “auto range“, which is denoted by a bracket connecting the two ranges, or separately with a “pause” in between.
-
17
Scan and Reconstruction

Scan Modes

Sequential Scanning

This is an incremental, slice-by-slice imaging mode in which there is no table movement during data acquisi tion. A minimum interscan delay in between each acquisition is required to move the table to the next slice position.

Spiral Scanning

Spiral scanning is a continuous volume imaging mode. The data acquisition and table movements are per­formed simultaneously for the entire scan duration. There is no interscan delay and a typical range can be acquired in a single breath hold.
Each acquisition provides a complete volume data set, from which images with overlapping can be recon­structed at any arbitrary slice position. Unlike the sequence mode, spiral scanning does not require addi­tional radiation to obtain overlapping slices.
-

Dynamic Serioscan

Dynamic serial scanning mode without table feed. Dynamic serio can still be used for dynamic evaluation, such as Test Bolus.
18
Scan and Reconstruction

Slice Collimation and Slice Width

Slice collimation is the slice thickness resulting from the effect of the tube-side collimator and the adaptive detector array design. In Multislice CT, the Z-coverage per rotation is given by the product of the number of active detector slices and the collimation (e.g., 2 x
mm).
1.0
Slice width is the FWHM (full width at half maximum) of the reconstructed image.
With the SOMATOM Spirit, you select the slice collima­tion together with the slice width desired. The slice width is independent of pitch, i.e. what you select is always what you get. Actually, you do not need to be concerned about the algorithm any more; the software does it for you.
The Recon icon on the chronicle will be labeled with “RT”. After the scan, the Real Time displayed image series has to be reconstructed.
The following tables show the possibilities of image reconstruction in spiral and sequential scanning.
19
Scan and Reconstruction

Slice Collimation and Slice Width for Spiral Mode and HR Spiral Mode

1 mm: 1, 1.25, 2, 3, 5 mm
1.5 mm: 2, 3, 5, 6 mm
2.5 mm: 3, 5, 6, 8, 10 mm 4 mm: 5, 6, 8, 10 mm 5 mm: 6, 8, 10 mm

Slice Collimation and Slice Width for Sequence Mode and HR Sequence Mode

1.0 mm: 1, 2 mm
1.5 mm: 1.5, 3 mm
2.5 mm: 2.5, 5 mm
4.0 mm: 4, 8 mm
5.0 mm 5, 10 mm
20
Scan and Reconstruction

Increment

The increment is the distance between the recon­structed images in the Z direction. When the increment chosen is smaller than the slice thickness, the images are created with an overlap. This technique is useful to reduce partial volume effect, giving you better detail of the anatomy and high quality 2D and 3D post-process ing.
Slice Thickness = 10 mm
-
Increment = 10 mm
Reconstruction Increment
Increment = 5 mm
Increment = 3 mm
21
Scan and Reconstruction

Pitch

In single slice CT: Pitch = table movement per rotation/slice collimation
E.g.,: slice collimation = 5 mm, table moves 5 mm per rotation, then pitch = 1.
With the Siemens Multislice CT, we differentiate between:
Feed/Rotation, the table movement per rotation
Volume Pitch, table movement per rotation/single
slice collimation.
Pitch Factor, table movement per rotation/complete slice collimation.
E.g., slice collimation = 2 x 5 mm, table moves 10 mm per rotation,
then Volume Pitch = 2, Pitch Factor = 1.
With the SOMATOM Spirit, the pitch, slice, collimation, rotation time, and scan range can be adjusted. The pitch factor can be selected from 0.5-2.
Pitch 1
Pitch Models
22
Pitch 1.5
Pitch 2
Scan and Reconstruction

Window values

The Scale of the CT Hounsfield Units is from -1024 to +3071.
The displayed window values have to correspond to the anatomical structure.
Windowing is used to optimize contrast and brightness of images.
Hounsfield Units
Spleen
Fat
Window center C
Kidneys
Pancreas
Lung
Adrenal Glands
Blood
Heart
Window width W
Bones
Water
Breast
Air
Organ specific window values
Gray scale
white
CT-window values
Liver
Tumo r
Bladder
Colon
black
23
Scan and Reconstruction

Kernels

There are 3 different types of kernels: “H“ stands for Head, “B“ stands for Body, “C“ stands for ChildHead.
The image sharpness is defined by the numbers – the higher the number, the sharper the image; the lower the number, the smoother the image.
A set of 18 kernels is supplied, consisting of:
• 6 body kernels: smooth (B20s), medium smooth (B31s), medium (B41s), medium sharp (B50s), sharp (B60s), high res (B70s)
• 7 head kernels: smooth (H21s), medium smooth (H31s), medium (H41s), medium sharp (H50s), sharp (H60s), high res (H70s), ultra high res (H80s)
• 3 child head kernels: smooth (C20s), medium (C30s), sharp (C60s)
• 2 special kernels: S80s, U90s
Note: Do not use different kernels for body parts other than what they are designed for.
For further information regarding the kernels, please refer to the “Hints in General” of the corresponding body region.
24
Scan and Reconstruction

Image Filters

There are 3 different filters available:
LCE: The Low-contrast enhancement (LCE) filter
enhances low-contrast detectability. It reduces the image noise.
• Similar to reconstruction with a smoother kernel
• Reduces noise
• Enhances low-contrast detectability
• Adjustable in four steps
• Automatic post-processing
25
Scan and Reconstruction
HCE: The High-contrast enhancement (HCE) filter
enhances high-contrast detectability. It increases the image sharpness, similar to reconstruction with a sharper kernel.
• Increases sharpness
• Faster than raw-data reconstruction
• Enhances high-contrast detectability
• Automatic post-processing
ASA: The Advanced Smoothing Algorithm (ASA) fil­ter reduces noise in soft tissue, while edges with high contrast are preserved.
• Reduces noise without blurring of edges
• Enhances low-contrast detectability
• Individually adaptable
• Automatic post-processing
26
Scan and Reconstruction

Improved Head Imaging

An automatic bone correction algorithm has been included in the standard image reconstruction. Using a new iterative technique, typical artifacts arising from the beam-hardening effect, e.g., Hounsfield bar, are minimized without any additional post-processing. This advanced algorithm allows for excellent images of the posterior fossa, but also improves head image quality in general. Bone correction is activated auto matically for body region “Head”.
In order to optimize image quality versus radiation dose, scans in the body region “Head” are provided within a maximum scan field of 300 mm with respect to the iso-center. No recon job with a field of view exceeding those limits will be possible. Therefore, patient positioning has to be performed accurately to ensure a centered location of the skull.
Head image without correction.
-
Head image with cor­rections.
27

Dose Information

CTDIW and CTDI
Vol
The average dose in the scan plane is best described by the CTDIW for the selected scan parameters. The CTDI is measured in the dedicated plastic phantoms – 16 cm diameter for head and 32 cm diameter for body (as defined in IEC 60601 –2 – 44). This dose number gives a good estimate for the average dose applied in the scanned volume as long as the patient size is similar to the size of the respective dose phantoms.
Since the body size can be smaller or larger than 32 cm, the CTDIW value displayed can deviate from the dose in the scanned volume.
The CTDIW definition and measurement is based on sin­gle axial scan modes. For clinical scanning, i.e. scan­ning of entire volumes in patients, the average dose will also depend on the table feed in between axial scans or the feed per rotation in spiral scanning. The dose, expressed as the CTDI
, must therefore be cor-
W
rected by the Pitch Factor of the spiral scan or an axial scan series to describe the average dose in the scanned volume.
For this purpose the IEC defined the term “CTDIVol“ in September 2002:
W
CTDI
= CTDIW/Pitch Factor
Vol
This dose number is displayed on the user interface for the selected scan parameters.
28
Dose Information
The CTDI tion of the radiation risk associated with CT examina­tion. For the purpose, the concept of the “Effective Dose“ was introduced by ICRP (International Commis sion on Radiation Protection). The effective dose is expressed as a weighted sum of the dose applied not only to the organs in the scanned range, but also to the rest of the body. It could be measured in whole body phantoms (Alderson phantom) or simulated with Monte Carlo techniques.
The calculation of the effective dose is rather compli­cated and has to be done by sophisticated programs. These have to take into account the scan parameters, the system design of individual scanner, such as x-ray filtration and gantry geometry, the scan range, the organs involved in the scanned range and the organs affected by scattered radiation. For each organ, the respective dose delivered during the CT scanning has to be calculated and then multiplied by its radiation risk factor. Finally, the weighted organ dose numbers are added up to get the effective dose.
The concept of effective dose allows the comparison of radiation risk associated with different CT or x-ray exams, i.e. different exams associated with the same effective dose would have the same radiation risk for the patient. It also allows comparing the applied x-ray exposure to the natural background radiation, e.g., 2 – 3 mSv per year in Germany.
value does not provide the entire informa-
vol
-
29
Dose Information

Effective mAs

In sequential scanning, the dose (Dseq) applied to the patient is the product of the tube current-time (mAs) and the CTDI
D
= D
seq
In spiral scanning, however, the applied dose (Dspiral) is influenced by the “classical“ mAs (mA x Rot Time) and in addition by the Pitch Factor. For example, if a Multislice CT scanner is used, the actual dose applied to the patient in spiral scanning will be decreased when the Pitch Factor is larger than 1, and increased when the Pitch Factor is smaller than 1. Therefore, the dose in spiral scanning has to be corrected by the Pitch Factor:
D
= (D
spiral
To make it easier for the users, the concept of the “effective“ mAs was introduced with the SOMATOM Multislice scanners.
The effective mAs takes into account the influence of pitch on both the image quality and dose:
Effective mAs = mAs/Pitch Factor
per mAs:
w
x mAs
CTDIw
x mA x Rot Time)/Pitch Factor
CTDIw
To calculate the dose you simply have to multiply the CTDIw per mAs with the effective mAs of the scan:
D
spiral
= D
x effective mAs
CTDIw
30
Loading...
+ 291 hidden pages