Siemens 2005C User Manual

4 (1)

SOMATOM

Spirit

Application Guide

Protocols

Principles

Helpful Hints

syngo 3D

syngo Fly Through syngo Dental CT syngo Osteo CT

syngo Volume Evaluation syngo Dynamic Evaluation

Software Version syngo CT 2005C

The information presented in this Application Guide is for illustration only and is not intended to be relied upon by the reader for instruction as to the practice of medicine. Any health care practitioner reading this information is reminded that they must use their own learning, training and expertise in dealing with their individual patients.

This material does not substitute for that duty and is not intended by Siemens Medical Solutions Inc., to be used for any purpose in that regard. The drugs and doses mentioned are consistent with the approval labeling for uses and/or indications of the drug. The treating physician bears the sole responsibility for the diagnosis and treatment of patients, including drugs and doses prescribed in connection with such use. The Operating Instructions must always be strictly followed when operating the CT System. The source for the technical data is the corresponding data sheets.

The pertaining operating instructions must always be strictly followed when operating the SOMATOM Spirit. The statutory source for the technical data are the corresponding data sheets.

We express our sincere gratitude to the many customers who contributed valuable input.

Special thanks to Heike Theessen, Christiane Bredenhöller, Kristin Pacheco, Karin Ladenburger, and Chen Mahao for their valuable assistance.

To improve future versions of this Application Guide, we would greatly appreciate your questions, suggestions and comments.

Please contact us:

USC-Hotline:

Tel. no.+49-1803-112244

email ct-application.hotline@med.siemens.de

Editor: Ute Feuerlein

2

Overview

 

User Documentation

14

 

Scan and Reconstruction

16

 

 

Dose Information

28

 

 

 

Workflow Information

36

 

 

 

Application Information

54

 

 

 

Head

70

 

 

 

Neck

88

 

 

 

Shoulder

94

 

 

 

Thorax

98

 

 

 

Abdomen

110

 

 

 

Pelvis

124

 

 

 

Spine

132

 

 

 

Upper Extremities

146

 

 

 

 

 

3

 

Lower Extremities

154

 

Vascular

162

 

 

Specials

176

 

 

 

Children

184

 

 

 

syngo 3D

240

 

 

 

syngo Fly Through

258

 

 

 

syngo Dental CT

268

 

 

 

syngo Osteo CT

274

 

 

 

syngo Volume Evaluation

288

 

 

 

syngo Dynamic Evaluation

306

 

 

 

 

 

4

5

Contents

 

User Documentation

14

 

Scan and Reconstruction

16

 

 

Concept of Scan Protocols

16

 

 

Scan Set Up

17

 

Scan Modes

18

-

Sequential Scanning

18

-

Spiral Scanning

18

-

Dynamic Serioscan

18

 

Slice Collimation and Slice Width

19

 

- Slice Collimation and Slice Width for

 

 

 

Spiral Mode and HR Spiral Mode

20

 

- Slice Collimation and Slice Width for

 

 

 

Sequence Mode and HR Sequence Mode

20

 

Increment

21

 

Pitch

22

 

Window values

23

 

Kernels

24

 

Image Filters

25

 

Improved Head Imaging

27

 

Dose Information

28

 

 

CTDIW and CTDIVol

28

 

 

Effective mAs

30

 

CARE Dose

32

 

- How does CARE Dose work?

32

 

Workflow Information

36

 

 

Recon Jobs

36

 

 

Examination Job Status

37

 

Auto Load in 3D and Post-processing

 

 

Presets

38

 

How to Create your own

 

 

Scan Protocols

39

-

Edit/Save Scan Protocol

39

-

Scan Protocol Manager

40

6

Contents

 

Contrast Medium

45

-

The Basics

45

-

IV Injection

47

-

Bolus Tracking

48

-

Test Bolus using CARE Bolus

50

-

Test Bolus

51

 

Application Information

54

 

 

SOMATOM life

54

 

-

General Information

54

-

Key Features

55

-

Description

56

-

Access to Computer Based Training or

 

 

 

Manuals on CD ROM

57

-

SRS Based Services

58

-

Download of Files

59

-

Contact incl. DICOM Images

60

-

Trial Order and Installation

62

 

Image Converter

64

 

File Browser

66

 

Patient Protocol

68

 

Head

 

70

 

 

 

Overview

70

 

 

Hints in General

71

-

Head Kernels

71

 

HeadRoutine

72

 

HeadSeq

74

 

InnerEarHR

76

 

InnerEarHRSeq

78

 

Sinus

80

 

SinusSeq

82

 

Orbita

84

 

Dental

86

7

Contents

 

Neck

88

 

Overview

88

 

Hints in General

89

 

- Body Kernels

90

 

Neck

92

 

Shoulder

94

 

 

Overview

94

 

 

Hints in General

95

 

- Body Kernels

95

 

Shoulder

96

 

Thorax

98

 

 

Overview

98

 

 

Hints in General

99

 

- Body Kernels

101

 

ThoraxRoutine/ThoraxRoutine08s

102

 

ThoraxFast

104

 

ThoraxHRSeq

106

 

LungLowDose

108

 

Abdomen

110

 

 

Overview

110

 

 

Hints in General

111

 

- Body Kernels

113

 

AbdomenRoutine/AbdomenRoutine08s

114

 

AbdomenFast

116

 

AbdMultiPhase/AbdMultiPhase08s

118

 

AbdomenSeq

122

8

Contents

 

Pelvis

124

 

Overview

124

 

Hints in General

125

 

- Body Kernels

125

 

Pelvis

126

 

Hip

128

 

SI_Joints

130

 

Spine

132

 

 

Overview

132

 

 

Hints in General

133

 

- Body Kernels

135

 

C-Spine

136

 

C-SpineSeq

138

 

Spine

140

 

SpineSeq

142

 

Osteo

144

 

Upper Extremities

146

 

 

Overview

146

 

 

Hints in General

147

 

- Body Kernels

148

 

WristHR

150

 

ExtrRoutineHR

152

 

Lower Extremities

154

 

 

Overview

154

 

 

Hints in General

155

 

- Body Kernels

156

 

KneeHR

158

 

FootHR

160

 

ExtrRoutineHR

161

9

Contents

 

Vascular

162

 

Overview

162

 

Hints in General

163

-

Head Kernels

163

-

Body Kernels

163

 

HeadAngio/HeadAngio08s

164

 

CarotidAngio/CarotidAngio08s

166

 

ThorAngio/ThorAngio08s

168

 

Embolism

170

 

BodyAngioRoutine/BodyAngioRoutine08s 172

 

BodyAngioFast

174

 

Specials

176

 

 

Overview

176

 

-

Trauma

176

-

Interventional CT

176

-

Test Bolus

176

 

Trauma

177

-

The Basics

177

 

PolyTrauma

178

 

HeadTrauma

180

 

Interventional CT

181

 

Biopsy

182

 

TestBolus

183

 

Children

184

 

 

Overview

184

 

 

Hints in General

187

-

Head Kernels

190

-

Body Kernels

191

 

HeadRoutine_Baby

192

 

HeadRoutine_Child

194

 

HeadSeq_Baby

196

 

HeadSeq_Child

198

 

InnerEar

200

 

SinusOrbi

202

 

Neck

204

10

Contents

 

ThoraxRoutine_Baby

206

 

ThoraxRoutine_Child

208

 

ThoraxHRSeq_Baby

210

 

ThoraxHRSeq_Child

212

 

Abdomen_Baby

214

 

Abdomen_Child

216

 

Spine_Baby

218

 

Spine_Child

220

 

ExtrHR_Baby

222

 

ExtrHR_Child

224

 

HeadAngio

226

 

HeadAngio08s

228

 

CarotidAngio

230

 

CarotidAngio08s

232

 

BodyAngio

234

 

BodyAngio08s

236

 

NeonateBody

238

 

syngo 3D

240

 

 

- Multi Planar Reconstruction (MPR)

240

 

 

- Maximum Intensity Projection (MIP)

240

 

- Shaded Surface Display (SSD)

241

 

- Volume Rendering Technique (VRT)

241

-

Prerequisites

242

 

Workflow

242

-

Loading the Images

242

-

Creating Series

244

-

Editing

246

-

Documentation of Results

249

 

Workflow for a CT Extremity Examination

250

-

Using MPR/MPR Thick

250

-

Using SSD

251

-

Using VRT

251

 

Workflow for a CT Angiography

252

-

Using MIP/MIP Thin

252

-

Using VRT/VRT Thin/Clip

253

 

Hints in General

254

 

- Setting Views in the Volume Data Set

254

 

- Changing /Creating VRT Presets

255

11

Contents

 

- Auto Load in 3D and Post-processing

 

 

 

Presets

257

-

Blow-up Mode

257

 

syngo Fly Through

258

 

 

Key Features

258

 

 

Prerequisites

259

 

The Basics for CT Virtual Endoscopy

259

 

- SSD and VRT Presets for Endoscopic

 

 

 

Renderings

259

-

Endoscopic Viewing Parameters/

 

 

 

Fly Cone Settings

260

-

Patient Preparation

262

 

Workflow

263

 

- Navigation of the Endoscopic Volume

265

-

Fly Path Planning

266

 

syngo Dental CT

268

 

 

The Basics

268

 

 

Scan Protocols

269

 

Additional Important Information

271

 

syngo Osteo CT

274

 

 

The Basics

274

 

 

Scanning Procedure

275

 

Configuration

278

 

Evaluation Workflow

282

 

Additional Important Information

287

 

syngo Volume Evaluation

288

 

 

Prerequisites

290

 

 

Workflow

291

 

General Hints

300

 

Configuration

303

12

Contents

 

syngo Dynamic Evaluation

306

 

Prerequisites

308

 

Workflow

309

 

- 1. Loading the Images

309

 

- 2. Inspecting the Input Images

310

- 3.

Generation of Parameter Images

310

- 4.

Creating a Baseline Image

313

- 5.

Evaluation of Region of Interests

314

 

- 6. Enhancement Curve

315

- 7.

Documentation of Results

316

 

General Hints

317

13

User Documentation

For further information about the basic operation, please refer to the corresponding syngo CT Operator Manual:

syngo CT Operator Manual Volume 1:

Security Package

Basics

Preparations

Examination

CARE Bolus CT

syngo CT Operator Manual Volume 2:

syngo Patient Browser syngo Viewing

syngo Filming syngo 3D

syngo CT Operator Manual Volume 3:

syngo Data Set Conversion syngo Dental CT

syngo Dynamic Evaluation syngo Osteo CT

syngo Volume

14

User Documentation

15

Scan and Reconstruction

Concept of Scan Protocols

The scan protocols for adult and children are defined according to body regions – Head, Neck, Shoulder,

Thorax, Abdomen, Pelvis, Spine, Upper Extremities, Lower Extremities, Specials, and Vascular.

The general concept is as follows: All protocols without suffix are standard spiral modes. E.g., “Shoulder” means the spiral mode for the shoulder.

The suffixes of the protocol name are follows:

Routine“: for routine studies

Seq”: for sequence studies

Fast“: use a higher pitch for fast acquisition

HR“: use a thinner slice width (1.0 mm) for High Resolution studies and a thicker slice width for soft tissue studies

The availability of scan protocols depends on the system configuration.

16

Scan and Reconstruction

Scan Set Up

Scans can be simply set up by selecting a predefined examination protocol. To repeat any mode, just click the chronicle with the right mouse button for “repeat”. To delete it, select “cut“. Each range name in the chronicle can be easily changed before “load“.

Multiple ranges can be run either automatically with “auto range“, which is denoted by a bracket connecting the two ranges, or separately with a “pause” in between.

17

Scan and Reconstruction

Scan Modes

Sequential Scanning

This is an incremental, slice-by-slice imaging mode in which there is no table movement during data acquisition. A minimum interscan delay in between each acquisition is required to move the table to the next slice position.

Spiral Scanning

Spiral scanning is a continuous volume imaging mode. The data acquisition and table movements are performed simultaneously for the entire scan duration. There is no interscan delay and a typical range can be acquired in a single breath hold.

Each acquisition provides a complete volume data set, from which images with overlapping can be reconstructed at any arbitrary slice position. Unlike the sequence mode, spiral scanning does not require additional radiation to obtain overlapping slices.

Dynamic Serioscan

Dynamic serial scanning mode without table feed. Dynamic serio can still be used for dynamic evaluation, such as Test Bolus.

18

Scan and Reconstruction

Slice Collimation and

Slice Width

Slice collimation is the slice thickness resulting from the effect of the tube-side collimator and the adaptive detector array design. In Multislice CT, the Z-coverage per rotation is given by the product of the number of active detector slices and the collimation (e.g., 2 x 1.0 mm).

Slice width is the FWHM (full width at half maximum) of the reconstructed image.

With the SOMATOM Spirit, you select the slice collimation together with the slice width desired. The slice width is independent of pitch, i.e. what you select is always what you get. Actually, you do not need to be concerned about the algorithm any more; the software does it for you.

The Recon icon on the chronicle will be labeled with “RT”. After the scan, the Real Time displayed image series has to be reconstructed.

The following tables show the possibilities of image reconstruction in spiral and sequential scanning.

19

Scan and Reconstruction

Slice Collimation and Slice Width for Spiral Mode and HR Spiral Mode

1 mm:

1, 1.25, 2, 3, 5 mm

1.5 mm:

2, 3, 5, 6 mm

2.5 mm:

3, 5, 6, 8, 10 mm

4 mm:

5, 6, 8, 10 mm

5 mm:

6, 8, 10 mm

Slice Collimation and Slice Width for Sequence Mode and HR Sequence Mode

1.0 mm:

1, 2 mm

1.5 mm:

1.5, 3 mm

2.5 mm:

2.5, 5 mm

4.0 mm:

4,

8 mm

5.0 mm

5,

10 mm

20

Scan and Reconstruction

Increment

The increment is the distance between the reconstructed images in the Z direction. When the increment chosen is smaller than the slice thickness, the images are created with an overlap. This technique is useful to reduce partial volume effect, giving you better detail of the anatomy and high quality 2D and 3D post-process- ing.

Slice Thickness = 10 mm

 

Increment = 3 mm

Increment = 10 mm

Increment = 5 mm

 

Reconstruction Increment

21

Scan and Reconstruction

Pitch

In single slice CT:

Pitch = table movement per rotation/slice collimation

E.g.,: slice collimation = 5 mm,

table moves 5 mm per rotation, then pitch = 1.

With the Siemens Multislice CT, we differentiate between:

Feed/Rotation, the table movement per rotation

Volume Pitch, table movement per rotation/single slice collimation.

Pitch Factor, table movement per rotation/complete slice collimation.

E.g., slice collimation = 2 x 5 mm, table moves 10 mm per rotation,

then Volume Pitch = 2, Pitch Factor = 1.

With the SOMATOM Spirit, the pitch, slice, collimation, rotation time, and scan range can be adjusted. The pitch factor can be selected from 0.5-2.

Pitch 1

Pitch 1.5

Pitch 2

Pitch Models

22

Scan and Reconstruction

Window values

The Scale of the CT Hounsfield Units is from -1024 to +3071.

The displayed window values have to correspond to the anatomical structure.

Windowing is used to optimize contrast and brightness of images.

Hounsfield Units

 

Gray scale

 

 

 

 

 

 

white

Window

Window

 

 

width W

center C

 

 

 

 

 

black

 

 

CT-window values

 

Blood

Liver

 

Spleen Kidneys

Tumor

Heart

 

Bones

Pancreas Adrenal

Colon

Bladder

 

Glands

 

 

 

 

Water

Breast

Fat

Air Lung

Organ specific window values

23

Scan and Reconstruction

Kernels

There are 3 different types of kernels: “H“ stands for Head, “B“ stands for Body, “C“ stands for ChildHead.

The image sharpness is defined by the numbers – the higher the number, the sharper the image; the lower the number, the smoother the image.

A set of 18 kernels is supplied, consisting of:

6 body kernels: smooth (B20s), medium smooth (B31s), medium (B41s), medium sharp (B50s), sharp (B60s), high res (B70s)

7 head kernels: smooth (H21s), medium smooth (H31s), medium (H41s), medium sharp (H50s), sharp (H60s), high res (H70s), ultra high res (H80s)

3 child head kernels: smooth (C20s), medium (C30s), sharp (C60s)

2 special kernels: S80s, U90s

Note: Do not use different kernels for body parts other than what they are designed for.

For further information regarding the kernels, please refer to the “Hints in General” of the corresponding body region.

24

Scan and Reconstruction

Image Filters

There are 3 different filters available:

LCE: The Low-contrast enhancement (LCE) filter enhances low-contrast detectability. It reduces the image noise.

Similar to reconstruction with a smoother kernel

Reduces noise

Enhances low-contrast detectability

Adjustable in four steps

Automatic post-processing

25

Scan and Reconstruction

HCE: The High-contrast enhancement (HCE) filter enhances high-contrast detectability. It increases the image sharpness, similar to reconstruction with a sharper kernel.

Increases sharpness

Faster than raw-data reconstruction

Enhances high-contrast detectability

Automatic post-processing

ASA: The Advanced Smoothing Algorithm (ASA) filter reduces noise in soft tissue, while edges with high contrast are preserved.

Reduces noise without blurring of edges

Enhances low-contrast detectability

Individually adaptable

Automatic post-processing

26

Siemens 2005C User Manual

Scan and Reconstruction

Improved Head Imaging

An automatic bone correction algorithm has been included in the standard image reconstruction. Using a new iterative technique, typical artifacts arising from the beam-hardening effect, e.g., Hounsfield bar, are minimized without any additional post-processing. This advanced algorithm allows for excellent images of the posterior fossa, but also improves head image quality in general. Bone correction is activated automatically for body region “Head”.

In order to optimize image quality versus radiation dose, scans in the body region “Head” are provided within a maximum scan field of 300 mm with respect to the iso-center. No recon job with a field of view exceeding those limits will be possible. Therefore, patient positioning has to be performed accurately to ensure a centered location of the skull.

Head image without correction.

Head image with corrections.

27

Dose Information

CTDIW and CTDIVol

The average dose in the scan plane is best described by the CTDIW for the selected scan parameters. The CTDIW is measured in the dedicated plastic phantoms – 16 cm diameter for head and 32 cm diameter for body (as defined in IEC 60601 –2 – 44). This dose number gives a good estimate for the average dose applied in the scanned volume as long as the patient size is similar to the size of the respective dose phantoms.

Since the body size can be smaller or larger than

32 cm, the CTDIW value displayed can deviate from the dose in the scanned volume.

The CTDIW definition and measurement is based on single axial scan modes. For clinical scanning, i.e. scanning of entire volumes in patients, the average dose will also depend on the table feed in between axial scans or the feed per rotation in spiral scanning. The dose, expressed as the CTDIW, must therefore be corrected by the Pitch Factor of the spiral scan or an axial scan series to describe the average dose in the scanned volume.

For this purpose the IEC defined the term “CTDIVol“ in September 2002:

CTDIVol = CTDIW/Pitch Factor

This dose number is displayed on the user interface for the selected scan parameters.

28

Dose Information

The CTDIvol value does not provide the entire information of the radiation risk associated with CT examination. For the purpose, the concept of the “Effective Dose“ was introduced by ICRP (International Commission on Radiation Protection). The effective dose is expressed as a weighted sum of the dose applied not only to the organs in the scanned range, but also to the rest of the body. It could be measured in whole body phantoms (Alderson phantom) or simulated with Monte Carlo techniques.

The calculation of the effective dose is rather complicated and has to be done by sophisticated programs. These have to take into account the scan parameters, the system design of individual scanner, such as x-ray filtration and gantry geometry, the scan range, the organs involved in the scanned range and the organs affected by scattered radiation. For each organ, the respective dose delivered during the CT scanning has to be calculated and then multiplied by its radiation risk factor. Finally, the weighted organ dose numbers are added up to get the effective dose.

The concept of effective dose allows the comparison of radiation risk associated with different CT or x-ray exams, i.e. different exams associated with the same effective dose would have the same radiation risk for the patient. It also allows comparing the applied x-ray exposure to the natural background radiation,

e.g., 2 – 3 mSv per year in Germany.

29

Dose Information

Effective mAs

In sequential scanning, the dose (Dseq) applied to the patient is the product of the tube current-time (mAs) and the CTDIw per mAs:

Dseq = DCTDIw x mAs

In spiral scanning, however, the applied dose (Dspiral) is influenced by the “classical“ mAs (mA x Rot Time) and in addition by the Pitch Factor. For example, if a Multislice CT scanner is used, the actual dose applied to the patient in spiral scanning will be decreased when the Pitch Factor is larger than 1, and increased when the Pitch Factor is smaller than 1. Therefore, the dose in spiral scanning has to be corrected by the Pitch Factor:

Dspiral = (DCTDIw x mA x Rot Time)/Pitch Factor

To make it easier for the users, the concept of the “effective“ mAs was introduced with the SOMATOM Multislice scanners.

The effective mAs takes into account the influence of pitch on both the image quality and dose:

Effective mAs = mAs/Pitch Factor

To calculate the dose you simply have to multiply the CTDIw per mAs with the effective mAs of the scan:

Dspiral = DCTDIw x effective mAs

30

Loading...
+ 291 hidden pages