Shimadzu Sonialvision Safire Operation Manual

O P E R A T I O N G U I D E
OPERATION GUIDE
M517-E077A
Notation
The following notation is used in this operation guide.
Denotes the contents of the operation
1, 2, 3
Denotes the contents of the operation step by step
Denotes the consequences of the operation Caution Denotes a special notice of the operation Note
Denotes the information which helps operation
Denotes points and notices of the operation
This operation guide is intended to compliment each "Operation Manual" of Sonialvision Sa re II / Sa re 17 and concisely describes its basic operations. Prior to using this guide, make sure to carefully read all "Operation Manual" booklets of Sonialvision Sa re II / Sa re 17 and fully understand them.
In case of Sonialvision Sa re II,
• Digital Radiography System DAR-8000f Operation Manual (M517-E060)
• X-ray High Voltage Generator UD150B-40 Operation Manual (M501-E052)
• Remote-Controlled X-ray Diagnostic Table ZS-100I/IR Operation Manual (M506-E034) In case of Sonialvision Sa re 17,
• Digital Radiography System DAR-8000f Operation Manual (M517-E060)
• X-ray High Voltage Generator D150BC-40, GSC-2002L Operation Manual (M501-E093)
• Remote-Controlled X-ray Diagnostic Table ZS-100I/IR Operation Manual (M506-E034)
M517-E077A
OPERATION GUIDE
System Overview........................... 4
Configuration................................................................ 4
Diagnostic table console.............................................. 5
X-ray high voltage generator console..........................6
Digital image processor main screen...........................6
Operation flow chart.....................................................7
1
2 3
4
5
6
7
System Startup and Shutdown .... 8
Startup ......................................................................... 8
Shutdown..................................................................... 9
FPD calibration .......................................................... 10
Enter Study .................................. 13
Defining usual study................................................... 13
Defining patient information in an emergency............ 14
Choosing types of study............................................. 15
Fluoroscopy/Radiography.......... 16
Setting radiography condition .................................... 16
Radiography............................................................... 17
· SPOT radiography/SERIAL radiography …… 17
· DSA radiography (Option) ……………………18
· Digital tomography (Option)…………………… 19
Process Images........................... 20
Printing images.......................................................... 20
Sending by DICOM function ...................................... 23
Close Study.................................. 24
Closing active study................................................... 24
Troubleshooting .......................... 25
Emergency stop/Recovery.........................................25
Error messages.......................................................... 26
Actions after power failure ......................................... 30
4
M517-E077A
1 System Overview
Automatic collimator
FPD
(Flat panel detector)
Hand switch
Speaker unit
Microphone
Diagnostic table console
Foot switch
X-ray high voltage generator console
Image monitor
Exposure switch
Mouse Keyboard
Control cabinet
(Diagnostic table)
Subsystem control cabinet
(Flat panel detector)
Diagnostic table
Selector
(attached mouse, monitor and keyboard)
PCU
(FPD control)
Control cabinet
(Flat panel detector)
Control cabinet
(X-ray high voltage generator)
X-ray tube unit
Control cabinet
(Digital image processor)
1
Configuration
System Overview
5
M517-E077A
1 System Overview
R1 - R4
Selects the APR.
Table to the vertical position
T
able to the
horizontal position
SID selector switches
Display panel
Exposure indicator
Exposure switch
Reverse the image vertically
Reverse the image horizontally
Stop switch
Adjusts the contrast of the image.
Adjusts the brightness of image.
Joystick
Table/imaging unit control lever
F1 - F5
Sets up the Subdivisional Acquisition for SPOT radiography. Sets up the tomography parameter for tomography (Option).
Fluoroscopy selection switch
Oblique projection center switch
L1 - L4
Selects the magnifi cation size of FPD.
Diagnostic table console
6
M517-E077A
1 System Overview
1 2 3
+5
0
-5
Focus button
kV up/down button (Adjust in small increments)
kV shuttle (Adjust in large increments)
Power ON button
Power OFF button
Home button
Log button
Setup button
mA/mAs up/down button (Adjust in small increments)
mA/mAs shuttle (Adjust in large increments)
sec up/down button (Adjust in small increments)
sec shuttle (Adjust in large increments)
AEC button
Hand switch
Density up/down button
X-ray radiography button
Radiography preparation button
mA/mAs switch button
Standard control panel
Expanded control panel
Standard control panel/thumbnail display button
Patient information
Radiography program button
Touch panel
Digital image processor main screen
X-ray high voltage generator console
7
M517-E077A
1 System Overview
Start study P.13
Choose type of study P.15
Set Fluo/Rad condition P.16
Fluoroscopy/Radiography P.17
Print image P.20
Send by DICOM P.23
Close study P.24
Shut down system P.9
Repeat as required.
Start up system P.8
Calibrate FPD P.10
Operation flow chart
8
M517-E077A
2 System startup and shutdown
1
Verify
Image monitor of digital image processor and peripherals are turned OFF.
2
Press
will be illuminated after approx.
2 seconds.
Image monitor, peripherals, and
control cabinet are turned ON.
1 2 3
+5
0
-5
Digital image processor
While the system is running the
power up self-testing routine, a splash screen appears.
Startup is completed when the
following screen appears.
X-ray high voltage generator console
Stand-by indicator
Main image area
2
System startup and shutdown
Startup
9
M517-E077A
2 System startup and shutdown
1
Close active study
1 Click on the Standard control
panel.
A confirmation dialog appears.
2 Click [OK].
The study is closed.
2
Close the application program
1 Click on the Standard control
panel.
A confirmation dialog appears.
2 Click [Yes].
The system will be automatically
shut down and the control cabinet will be turned OFF.
X-ray high voltage generator console
3
Turn OFF the X-ray high voltage generator
Press OFF.
ON lamp is turned off, and then
OFF lamp is turned ON.
After pressing the OFF button on the X-ray high voltage generator console, do not press the ON button for at least 10 seconds. If you do, the system may not operate properly.
Caution
Digital image processor
1 2 3
+5
0
-5
Shutdown
10
M517-E077A
2 System startup and shutdown
This operation maintains the quality of fl uoroscopy/radiography images. Be sure to perform the FPD calibration once a day after system startup. Verify that no substance is located on the diagnostic table during FPD calibration. Approx. 15 minutes is required to complete the FPD calibration.
Auto FPD calibration on system startup
Perform when the system starts up.
1
Start up the system
After system startup, an FPD
calibration dialog appears.
A on diagnostic table console
blinks, and the message appears.
2
Press and hold the [SET] on the diagnostic table console
The diagnostic table moves to
calibration position.
It sounds beep when the diagnostic
table arrives in calibration position.
An FPD Calibration start dialog
appears.
FPD calibration
Press [F5 : SET] to move Table to
calibration starting position.
Press [F5 : SET] to move Table to
calibration starting position.
[SET]
Loading...
+ 22 hidden pages